P/R60 & P/R65 Page 6
3.0 INSTALLATION
3.1 General Instructions
3.1.1 Specific instructions for each type of probe are given in the following pages.
Common to all probes are the following instructions:
a) If the distance between the probe and the instrument is such that a direct
connection is not possible, the probe cable should be routed to a junction box
with a terminal strip (AquaMetrix Part No. JB1). The box should be well sealed
and away from corrosion danger. Be sure that you have sufficient slack cable
to allow for probe removal for calibration and servicing.
b) Route the interconnect cable from the junction box to the instrument, preferably
in metal conduit. Do not run the power cable or control cables in the same
conduit with the probe interconnect cable.
c) Remove the protective plastic caps from the end of the probe before placing in
service.
d) For best results probes should always be mounted vertically with electrodes
down. If this is not possible, the probe must be at least 15° above horizontal.
3.2 ORP 4-20 mA Probes, R65
3.2.1 a) ORP 4-20 mA probes have four wires; black, red, green and white. The red
wire is to be connected to the +24 Vdc terminal and the back wire to the 24 Vdc
common terminal via the load in the loop.
b) For an instrument with a range of 0 to 1000 mV the green and white wires are
to be shorted.
c) For an instrument with a range of –500 to 500 mV the green and white wires
are to be isolated from each other.
3.3 Submersion Mounting Series 8 Differential Probes
Refer to Dwg# N105-72
3.3.1 a) A submersion mounting kit, STC60L is available from Water Analytics which
includes 4 ft. of 1” pipe, 1-1/2” x 1” reducer, a strain relief fitting and wire
mounting bracket. Proceed as follows, either with the kit or with your own
hardware.
b) Apply a thread sealant to the thread on the cable end of the probe and screw a
1-1/2” x 1” NPT reducer onto the probe. Route the sensor cable through an
appropriate length of 1” pipe and using thread sealant, screw the pipe into the
reducer. The cable end of the probe should not be exposed to the process. A
cable strain relief fitting should be used on the upper end of the pipe. In the kits
a wire bracket is provided to aid in supporting the assembly.
NOTE: An optional protective shroud, Part No. PROTECTOR-3 should be used on
the electrode end of the probe to protect the electrodes from accidental
contact with the tank bottom, sides or objects in the process.