Aquametrix Combination Style
by
Water Analytics
Aqua
Metrix
pH and ORP Sensors
Model P575 (pH)
Model R575 (ORP)
Features
• Fixed insertion depth
• Industrial-grade quality
• Convertible design allows for
flow-through and submersion use
• 1000Ω RTD temperature
element included
• Mounts from either end in
standard 1” FNPT fittings
• Durable CPVC shell provides
excellent chemical resistance
Description
These industrial combination pH or ORP
sensors use the conventional measurement
technique, employing a process pH glass
electrode (platinum for ORP) which is compared to a reference electrode in KCL solution.
For relatively clean water they are an economical alternative to our flagship differential
60/65 series probe.
Our combination sensors have been completely modernized with several major upgrades.
A durable CPVC shell material provides excellent chemical resistance, and temperature
compensation (PT1000RTD) is now a standard option.
The temperature element is now embedded in the glass electrode for a faster response. One cable now includes both pairs
of wires for process and temperature output.
The reference chamber now contains a
rugged ceramic double junction, replacing
the single junction used in the past. This results in an approximate doubling of the sensor’s lifespan.
The dual-threaded, 1” fixed-insertion design of these probes allows them to be used
in both flow-through and submersion applications, minimizing specialized inventory
requirements.
These sensors are available with an optional screw-on electrode protector for submersion mount use (AM-PTR-5). It is easily
removed for flow-through installations when
the probe is mounted to a tee. The sensor
can also be mounted on a submersion arm
(AM-ARM-5) for hard-to-reach applications.
The 575 series probes can be directly connected to the AM-2250 TX transmitter or AM2250 controller, as well as to all legacy lines of
Aquametrix analyzers. For ease of installation,
the analyzer must be located within reach of
the 15 ft (4.5 m) standard sensor cable. Nonstandard cable lengths are also available;
however, a preamplifier is required for transmission distances greater than 50ft.
An alternative model is installed with a 3/4”
MNPT compression fitting. With this design,
the probe does not screw into the process
line, but is simply inserted through the compression fitting. Probe cleaning and system
calibration has never been easier. To order
this style, ask for data sheet P/R585.
For industrial or municipal wastewater applications, AquaMetrix differential sensors still
provide the lowest lifetime cost of ownership.
Applications
• Process Control
• Industrial and Municipal Water
Treatment
• Circuit Board Manufacturing
• Mining
• Power Generation
• Food and Beverage
• Aquaculture
• Agriculture
by
100 School Street
Andover, MA 01810
Toll free - 855-747-7623
www.WaterAnalytics.net
Me trix
978-749-9949

Model P575 and R575 Combination Style pH or ORP Sensors
Technical Data
Measuring Range
pH 0 to 14 pH (Consult factory for ap-
plications below 2 or above 12).
ORP -1000 to +1000 mV
Wetted Materials CPVC body, polycarbonate (or epoxy), ceramic
junction, silicone rubber, glass electrode, EPDM
Optional: platinum for ORP, teflon for
HF-resistant pH, antimony for HFA
Temperature Limits 23 to 176°F (-5 to 80°C) at amospheric pressure
Maximum Pressure 100 psig at 150°F (65°C)
Maximum Flow Rate 10 ft. (3 meters) per second
Automatic Temperature Compensation
Standard PT1000Ω RTD
Optional ATC: 300 NTC, pt100Ω RTD, 3000 NTC
Dimensions
Stability
pH 0.05 pH per day, non-cumulative
ORP 3.0 mV per day, non-cumulative
Sensitivity
pH 0.01 pH unit
ORP 0.1 mV
Accuracy
pH ± 0.02 pH
ORP ± 5 mV
Sensor Cable Default length 15 ft (4.5m)
dual coaxial, tinned wires
Up to 50ft cable available
Optional BNC connectors
Related Products
ANALYZERS
AM-2250 Multi-Parameter Controller
AM-2250TX Multi-Parameter Transmitter
ACCESORIES
AM-ARM-5 Submersion mounting arm, 4ft
AM-PTR-5 Electrode protector, CPVC
CALIBRATION SOLUTIONS
AM-PH4-1P pH 4 Buffer, 1 pint (500 mL)
AM-PH7-1P pH 7 Buffer, 1 pint (500 mL)
AM-PH10-1P pH 10 Buffer, 1 pint (500 mL)
AM-R200-1P ORP Buffer, 200 mV, 1 pint (500 mL)
AM-R600-1P ORP Buffer, 600 mV, 1 pint (500 mL)
DIFFERENTIAL SENSORS
P60R8-015 Differential pH sensor with 1-1/2” NPT
R60R8-015 Differential ORP sensor with 1-1/2” NPT
Ordering Information
P575T1-015 Standard Industrial Combination pH probe
for flow-through or submersion application
with pt1000RTD temperature compensation device and 15 ft dual coaxial cable
R575T1-015 Standard Industrial Combination ORP probe
for flow-through or submersion application
with pt1000RTD temperature compensation device and 15 ft dual coaxial cable
P575HF1-015
P575HFA-015 Industrial Combination pH probe for ap-
Industrial Combination pH probe for applications with low concetration (<2%) of HF acid.
Flow-through or submersion application
with pt1000RTD temperature compensation device and 15 ft dual coaxial cable
plications with <5% HF acid. Flow-through
or submersion application. Single junction with no temperature compensation
device and 15 ft dual coaxial cable
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)