AquaMetrix P575, R575 Data sheet

Aquametrix Combination Style
Water Analytics
Aq ua
by
Water Analytics
Aqua
Metrix
pH and ORP Sensors
Model P575 (pH) Model R575 (ORP)
Features
• Fixed insertion depth
• Industrial-grade quality
• Convertible design allows for
flow-through and submersion use
• 1000Ω RTD temperature
element included
• Mounts from either end in
standard 1” FNPT fittings
• Durable CPVC shell provides
excellent chemical resistance
Description
These industrial combination pH or ORP sensors use the conventional measurement technique, employing a process pH glass electrode (platinum for ORP) which is com­pared to a reference electrode in KCL solu­tion.
For relatively clean water they are an eco­nomical alternative to our flagship differential 60/65 series probe.
Our combination sensors have been com­pletely modernized with several major up­grades.
A durable CPVC shell material provides ex­cellent chemical resistance, and temperature compensation (PT1000RTD) is now a stan­dard option.
The temperature element is now embed­ded in the glass electrode for a faster re­sponse. One cable now includes both pairs of wires for process and temperature output.
The reference chamber now contains a rugged ceramic double junction, replacing the single junction used in the past. This re­sults in an approximate doubling of the sen­sor’s lifespan.
The dual-threaded, 1” fixed-insertion de­sign of these probes allows them to be used in both flow-through and submersion ap­plications, minimizing specialized inventory
requirements.
These sensors are available with an op­tional screw-on electrode protector for sub­mersion mount use (AM-PTR-5). It is easily removed for flow-through installations when the probe is mounted to a tee. The sensor can also be mounted on a submersion arm (AM-ARM-5) for hard-to-reach applications.
The 575 series probes can be directly con­nected to the AM-2250 TX transmitter or AM­2250 controller, as well as to all legacy lines of Aquametrix analyzers. For ease of installation, the analyzer must be located within reach of the 15 ft (4.5 m) standard sensor cable. Non­standard cable lengths are also available; however, a preamplifier is required for trans­mission distances greater than 50ft.
An alternative model is installed with a 3/4” MNPT compression fitting. With this design, the probe does not screw into the process line, but is simply inserted through the com­pression fitting. Probe cleaning and system calibration has never been easier. To order this style, ask for data sheet P/R585.
For industrial or municipal wastewater ap­plications, AquaMetrix differential sensors still provide the lowest lifetime cost of ownership.
Applications
• Process Control
• Industrial and Municipal Water
Treatment
• Circuit Board Manufacturing
• Mining
• Power Generation
• Food and Beverage
• Aquaculture
• Agriculture
by
100 School Street
Andover, MA 01810
Toll free - 855-747-7623
www.WaterAnalytics.net
Me trix
978-749-9949
Model P575 and R575 Combination Style pH or ORP Sensors
Technical Data
Measuring Range
pH 0 to 14 pH (Consult factory for ap-
plications below 2 or above 12).
ORP -1000 to +1000 mV
Wetted Materials CPVC body, polycarbonate (or epoxy), ceramic
junction, silicone rubber, glass electrode, EPDM Optional: platinum for ORP, teflon for HF-resistant pH, antimony for HFA
Temperature Limits 23 to 176°F (-5 to 80°C) at amospheric pressure
Maximum Pressure 100 psig at 150°F (65°C)
Maximum Flow Rate 10 ft. (3 meters) per second
Automatic Tempera­ture Compensation
Standard PT1000Ω RTD Optional ATC: 300 NTC, pt100Ω RTD, 3000 NTC
Dimensions
Stability
pH 0.05 pH per day, non-cumulative
ORP 3.0 mV per day, non-cumulative
Sensitivity
pH 0.01 pH unit
ORP 0.1 mV
Accuracy
pH ± 0.02 pH
ORP ± 5 mV
Sensor Cable Default length 15 ft (4.5m)
dual coaxial, tinned wires Up to 50ft cable available Optional BNC connectors
Related Products
ANALYZERS
AM-2250 Multi-Parameter Controller
AM-2250TX Multi-Parameter Transmitter
ACCESORIES
AM-ARM-5 Submersion mounting arm, 4ft
AM-PTR-5 Electrode protector, CPVC
CALIBRATION SOLUTIONS
AM-PH4-1P pH 4 Buffer, 1 pint (500 mL) AM-PH7-1P pH 7 Buffer, 1 pint (500 mL) AM-PH10-1P pH 10 Buffer, 1 pint (500 mL) AM-R200-1P ORP Buffer, 200 mV, 1 pint (500 mL) AM-R600-1P ORP Buffer, 600 mV, 1 pint (500 mL)
DIFFERENTIAL SENSORS
P60R8-015 Differential pH sensor with 1-1/2” NPT R60R8-015 Differential ORP sensor with 1-1/2” NPT
Ordering Information
P575T1-015 Standard Industrial Combination pH probe
for flow-through or submersion application with pt1000RTD temperature compensa­tion device and 15 ft dual coaxial cable
R575T1-015 Standard Industrial Combination ORP probe
for flow-through or submersion application with pt1000RTD temperature compensa­tion device and 15 ft dual coaxial cable
P575HF1-015
P575HFA-015 Industrial Combination pH probe for ap-
Industrial Combination pH probe for applica­tions with low concetration (<2%) of HF acid. Flow-through or submersion application with pt1000RTD temperature compensa­tion device and 15 ft dual coaxial cable
plications with <5% HF acid. Flow-through or submersion application. Single junc­tion with no temperature compensation device and 15 ft dual coaxial cable
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)