Now with larger electrode and ground rod for
better signal-to-noise and faster response.
Description
The 60 series pH and ORP probes are
dependable industrial grade sensors
designed to provide accurate measurement and longer service life under
the most demanding conditions. The
P60C-6 pH probes and R60C-6 ORP
probes incorporate all the benets of
differential measurement found in other models of the 60 series, eld proven
in thousands of installations. This technique uses two glass electrodes to
make the measurement differentially
with respect to a third metal electrode.
This second electrode is protected
from the process by a double junction
replaceable salt bridge. The resulting true differential measurement has
several advantages over conventional
probes: ground loop problems are virtually eliminated, and the salt bridge is
easily replaced. If the internal solution
becomes contaminated, the probe can
be rejuvenated at modest cost by replacing the salt bridge and reference
solution. Automatic temperature compensation is accomplished through
the use of an embedded thermistor
inside the tip of the ground rod. This
thermistor placement provides rapid
response for process temperature
variations. This unique probe mounts
in a standard 1 1/4” NPT tee, or in a
tank, through a specially designed
CPVC compression tting. This technology provides several important advantages: the probe enters straight into
the tting eliminating twisting of the
cable; the pipe size is only 1 1/4” NPT;
the insertion length is variable
between 7/8” and 5”; one model is used for both ow-through
and submersion tank-mount.
Installation does not require a pipe
wrench, strap wrench or thread sealant on the probe. This helps to reduce down time for calibration. The
P/R60C-6 may be ordered with a
at face electrode which fouls less
frequently and is easily cleaned.
The encapsulated preamplier provides an output signal which can
be transmitted 3000 feet over inexpensive cable. For greater distances, see the P65C data sheet
for the probe that encapsulates a
blind 4-20 mA two-wire transmitter
which can transmit a virtually unlimited distance over a twisted pair cable.
Aquametrix Models P/R60C-6 and -7
P/R60C- Differential pH/ORP Probes
Features
• Differential Measurement
• Replaceable Salt Bridge
• Low Maintenance Cost
• Easy in: no twisted cable
• Easy out: no special tools
• 1 1/4” Compression Fitting
• Adjustable insertion depth
• Easy Cleaning
• Semi-ush face for
reduced buildup
• Encapsulated Preamp
• Transmits up to 3000 ft
• 4-20 mA two-wire blind
transmitter version available
• Hot-Tap version available
• Flat-faced version
available for pH
• ORP available with
gold electrode
Applications
• Process Control
• Industrial and Municipal Wa-
ter Treatment
• Industrial and Municipal
Waste Treatment and Neutralization
• Fume Scrubbers, Circuit
Board Manufacturing, Plating,
Food and Beverage, Chemical Processing, Pulp and Paper, Mining, Power Generation,
Pharmaceuticals
100 School Street
Andover, MA 01810
978-749-9949
Toll free - 855-747-7623
www.WaterAnalytics.net

Model P60C-6 pH or R60C-6 ORP Differential Probes
Technical Data
Dimensions
Measuring Range
pH 0 to 14 pH
(Consult factory for applications below 2 or
above 12).
ORP -1000 to +1000 mV
Flow Rate 10 ft./sec maximum (3 metres/sec)
Flow should be as low as possible in low
conductivity water and in solutions with
high suspended solids
Wetted Materials CPVC, kynar, glass, titanium palladium al-
loy and EPDM (platinum for ORP probe)
Transmission Distance 3000 ft. (900 m)
Sensitivity
pH 0.01 pH
ORP 0.1 mV
Stability
pH 0.03 pH per day, non-cumulative
ORP 2 mV per day, non-cumulative
Temperature Copensation
Automatic -5 to 95°C (23 to 203°F)
Pressure Limit 100 psig at 65°C maximum
Temperature Limits
CPVC -5 to 95°C (23 to 203°F)
The temperature limit of probes in ow-
through applications is limited by pressure
and by the pipe tting material.
Probe Cable 5 Conductor plus shield, 15 ft. (4.6m) long
1.125 (29)
1.00 (25.4)
5.08 (129)
1.125 (28.6)
1.00 (25.4)
9.25 (235)
14.0 (356)
18.1 (460)
Related Products
CABLES & ACCESORIES
JB-1 NEMA 4X junction box
STC60-6 Mounting kit for submersion applications includes 1-1/4” NPT x 1” reducer, 4
feet of 1” CPVC pipe with water-tight strain relief tting and securing assembly
P60HTC Hot Tap Ball Valve assembly (Probe not included)
C42-5PXXX Interconnect cable, dressed both ends. Specify length.
C35-17 Salt bridge Kit for all -6 series probes, package of 3.
C35-79 Compression tting for all -6 series probes.
C34-51 Adapter for GLI Installations.
Protector-6 Protection shroud for submersion applications.
CALIBRATION SOLUTIONS
A35-13 pH 4 Buffer, 500 mL.
A35-14 pH 7 Buffer, 500 mL
A35-24 pH 10 Buffer, 500 mL
A35-40 ORP Buffer, 200 mV, 500 mL
A35-41 ORP Buffer, 600 mV, 500 mL
Ordering Information
P60C-6 pH probe complete with universal installation
tting for ow-through or submersion applica-
tions
P60C-7 pH Probe for use with Hot-Tap hardware
P60C-6F Flat-face pH probe
R60C-6 ORP probe complete with installation tting
for low-through or submersion applications
R60C-6-G ORP probe with gold electrode
R60C-7 ORP Probe for use with Hot-Tap hardware
P60C6
P60C7
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)