Aquametrix P/R60C-6 User Manual

Now with larger electrode and ground rod for better signal-to-noise and faster response.
Description
The 60 series pH and ORP probes are dependable industrial grade sensors designed to provide accurate mea­surement and longer service life under the most demanding conditions. The P60C-6 pH probes and R60C-6 ORP probes incorporate all the benets of differential measurement found in oth­er models of the 60 series, eld proven in thousands of installations. This tech­nique uses two glass electrodes to make the measurement differentially with respect to a third metal electrode.
This second electrode is protected from the process by a double junction replaceable salt bridge. The result­ing true differential measurement has several advantages over conventional probes: ground loop problems are vir­tually eliminated, and the salt bridge is easily replaced. If the internal solution becomes contaminated, the probe can be rejuvenated at modest cost by re­placing the salt bridge and reference solution. Automatic temperature com­pensation is accomplished through the use of an embedded thermistor inside the tip of the ground rod. This thermistor placement provides rapid response for process temperature
variations. This unique probe mounts in a standard 1 1/4” NPT tee, or in a tank, through a specially designed CPVC compression tting. This tech­nology provides several important ad­vantages: the probe enters straight into the tting eliminating twisting of the cable; the pipe size is only 1 1/4” NPT;
the insertion length is variable between 7/8” and 5”; one mod­el is used for both ow-through and submersion tank-mount.
Installation does not require a pipe wrench, strap wrench or thread seal­ant on the probe. This helps to re­duce down time for calibration. The P/R60C-6 may be ordered with a at face electrode which fouls less frequently and is easily cleaned.
The encapsulated preamplier pro­vides an output signal which can be transmitted 3000 feet over in­expensive cable. For greater dis­tances, see the P65C data sheet for the probe that encapsulates a blind 4-20 mA two-wire transmitter which can transmit a virtually unlim­ited distance over a twisted pair cable.
Aquametrix Models P/R60C-6 and -7
P/R60C- Differential pH/ORP Probes
Features
Differential Measurement
Replaceable Salt Bridge
Low Maintenance Cost
Easy in: no twisted cable
Easy out: no special tools
1 1/4” Compression Fitting
Adjustable insertion depth
Easy Cleaning
Semi-ush face for
reduced buildup
Encapsulated Preamp
Transmits up to 3000 ft
4-20 mA two-wire blind
transmitter version available
Hot-Tap version available
Flat-faced version
available for pH
ORP available with
gold electrode
Applications
Process Control
Industrial and Municipal Wa-
ter Treatment
Industrial and Municipal
Waste Treatment and Neutral­ization
Fume Scrubbers, Circuit
Board Manufacturing, Plating, Food and Beverage, Chemi­cal Processing, Pulp and Pa­per, Mining, Power Generation, Pharmaceuticals
100 School Street
Andover, MA 01810
978-749-9949
Toll free - 855-747-7623
www.WaterAnalytics.net
Model P60C-6 pH or R60C-6 ORP Differential Probes
Technical Data
Dimensions
Measuring Range
pH 0 to 14 pH
(Consult factory for applications below 2 or
above 12). ORP -1000 to +1000 mV Flow Rate 10 ft./sec maximum (3 metres/sec)
Flow should be as low as possible in low
conductivity water and in solutions with
high suspended solids Wetted Materials CPVC, kynar, glass, titanium palladium al-
loy and EPDM (platinum for ORP probe)
Transmission Distance 3000 ft. (900 m) Sensitivity
pH 0.01 pH ORP 0.1 mV
Stability
pH 0.03 pH per day, non-cumulative ORP 2 mV per day, non-cumulative
Temperature Copensation
Automatic -5 to 95°C (23 to 203°F)
Pressure Limit 100 psig at 65°C maximum Temperature Limits
CPVC -5 to 95°C (23 to 203°F)
The temperature limit of probes in ow-
through applications is limited by pressure
and by the pipe tting material. Probe Cable 5 Conductor plus shield, 15 ft. (4.6m) long
1.125 (29)
1.00 (25.4)
5.08 (129)
1.125 (28.6)
1.00 (25.4)
9.25 (235)
14.0 (356)
18.1 (460)
Related Products
CABLES & ACCESORIES
JB-1 NEMA 4X junction box STC60-6 Mounting kit for submersion applications includes 1-1/4” NPT x 1” reducer, 4
feet of 1” CPVC pipe with water-tight strain relief tting and securing assembly P60HTC Hot Tap Ball Valve assembly (Probe not included) C42-5PXXX Interconnect cable, dressed both ends. Specify length. C35-17 Salt bridge Kit for all -6 series probes, package of 3. C35-79 Compression tting for all -6 series probes. C34-51 Adapter for GLI Installations. Protector-6 Protection shroud for submersion applications.
CALIBRATION SOLUTIONS
A35-13 pH 4 Buffer, 500 mL. A35-14 pH 7 Buffer, 500 mL A35-24 pH 10 Buffer, 500 mL A35-40 ORP Buffer, 200 mV, 500 mL A35-41 ORP Buffer, 600 mV, 500 mL
Ordering Information
P60C-6 pH probe complete with universal installation
tting for ow-through or submersion applica-
tions P60C-7 pH Probe for use with Hot-Tap hardware P60C-6F Flat-face pH probe R60C-6 ORP probe complete with installation tting
for low-through or submersion applications R60C-6-G ORP probe with gold electrode R60C-7 ORP Probe for use with Hot-Tap hardware
P60C6
P60C7
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)