Aquametrix Model P/R60C-5
by
Water Analytics
Aqua
Metrix
Differential pH or ORP Probes
New
Features
• Replaceable Salt Bridge
• Low Maintenance Cost
• Field-proven
Encapsulated Preamplier
• Transmits up to 3000 ft.
Options
Description
The P60C-5 pH and the R60C-5
ORP probes are the latest additions
to the legendary Aquametrix differential probe family. It lls the gap
for a 1” xed insertion, MNPT probe
and uses the same electrodes and
salt bridge as the 1” variable insertion P60C-6 probe.
All of the P60 and P65 probes are
dependable industrial grade sensors
designed to provide accurate measurement and longer service life under the most demanding conditions.
Some features of these probes
include: differential measurement
technology, replaceable salt bridge
and encapsulated preamp.
The P/R60C-5 incorporates all the
benets of differential measurement found in other models of the
60 series, eld proven in thousands
of installations. This technique uses
two glass electrodes to make the
measurement differentially with respect to a third metal electrode. The
domed glass process electrode is
specially designed for tough applications. The second electrode is
immersed in a pH 7 buffer encapsu-
lated in the probe.
This second electrode is protected
from the process by a double junction replaceable salt bridge. The resulting true differential measurement
has several advantages over conventional probes: ground loop problems are virtually eliminated, and the
salt bridge is easily replaced.
If the internal solution becomes contaminated, the probe can be rejuvenated at modest cost by replacing
the salt bridge and reference solution. Automatic temperature compensation is accomplished through
the use of a thermistor at the tip of
the probe. This thermistor placement provides rapid response for
process temperature variations.
The encapsulated preamplier provides an output signal which can
be transmitted 3000 feet over inexpensive cable. Another version
encapsulates a blind 4-20 mA two
wire transmitter which can transmit
a virtually unlimited distance over a
twisted pair cable. (See P/R65 data
sheet).
• 4-20 mA two-wire blind transmit-
ter version available (model P/R65)
• ORP available with gold electrode
Applications
• Process Control
• Industrial and Municipal Waste
Treatment and Neutralization
• Fume Scrubbers
• Plating
• Circuit Board Manufacturing
• Food and Beverage
• Chemical Processing
• Pulp and Paper
• Mining
• Power Generation
by
Andover, MA 01810
Toll free - 855-747-7623
www.WaterAnalytics.net
Metrix
100 School Street
978-749-9949
Model P/R60C-5 Differential pH or ORP Probes
Technical Data
Measuring Range
pH 0 to 14 pH
(Consult factory for applications below 2 or
above 12).
ORP -2000 to +2000 mV
Flow Rate 10 ft./sec maximum (3 metres/sec)
Flow should be as low as possible in
low conductivity water and in solutions
with high suspended solids
Wetted Materials CPVC, Kynar, glass, titanium palladium al-
loy and EPDM (platinum for ORP probe);
opt. Viton and ceramic
Transmission Distance 3000 ft. (900 m)
Sensitivity
pH 0.001 pH
ORP 0.1 mV
Stability
pH 0.03 pH per day, non- cumulative
ORP 2 mV per day, non-cumulative
Temperature Copensation
Automatic -5 to 95°C (23 to 203°F)
Pressure Limit 100 psig at 65°C maximum
Temperature Limits
CPVC -5 to 95°C (23 to 203°F)
The temperature limit of probes in owthrough applications is limited by pressure
and by the pipe tting material.
Probe Cable 5 Conductor plus shield, 15 ft. (4.6m) long
Dimensions
Ordering Information
P60C-5 pH probe, 1” NPT threaded body
P60C-5F pH probe with at electrode
P60C-5A pH probe with Antimony electrode
P60C-5H pH probe with Hard glass electrode
R60C-5 ORP probe, 1” NPT threaded body
R60C-5F ORP probe with at electrode
R60C-5-H ORP probe with Hard glass electrode
R60C-5-G ORP probe with Gold electrode
Related Products
CABLES & ACCESORIES
JB1 NEMA 4X junction box
STC60-5 Mounting kit for submersion applications includes
1” NPT x 1” reducer, 4 feet of 1” CPVC pipe with
watertight strain relief tting and securing assembly
C42-5PXXX Interconnect cable; specify length
C35-17 Salt bridge kit w/ kynar outer junction (Package of 3)
C35-17K Salt bridge kit w/ ceramic outer junction (Package of 3)
Protector-6 Protection shroud for submersion applications
CALIBR ATION SOLUTIONS
A35-13 pH 4 Buffer, 500 mL.
A35-14 pH 7 Buffer, 500 mL
A35-24 pH 10 Buffer, 500 mL
A35-40 ORP Buffer, 200 mV, 500 mL
A35-41 ORP Buffer, 600 mV, 500 mL
Differential Measurement Technique
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)