Top: Hot-Tap probe P60C-7 with
P60-HTC ball valve assembly
Bottom: Variable insertion probe P60C-6
Description
The 60 series pH and ORP probes are
dependable industrial grade sensors
designed to provide accurate measurement and longer service life under
the most demanding conditions. The
P60C-6 pH probes and R60C-6 ORP
probes incorporate all the benets of
differential measurement found in oth-
er models of the 60 series, eld proven
in thousands of installations. This technique uses two glass electrodes to
make the measurement differentially
with respect to a third metal electrode.
This second electrode is protected
from the process by a double junction
replaceable salt bridge. The resulting true differential measurement has
several advantages over conventional
probes: ground loop problems are virtually eliminated, and the salt bridge is
easily replaced. If the internal solution
becomes contaminated, the probe can
be rejuvenated at modest cost by replacing the salt bridge and reference
solution. Automatic temperature compensation is accomplished through the
use of an embedded thermistor inside
the tip of the ground rod. This thermistor placement provides rapid response
for process temperature variations.
The unique P60C-6 probe mounts in
Aquametrix Models P/R60C-6 and
P/R60C-7 Differential pH/ORP Probes
a standard 1-1/4” NPT tee, or in a tank,
through a specially designed CPVC
compression tting. This technology
provides several important advantages: the probe enters straight into
the tting eliminating twisting of the
cable; the pipe size is only 1-1/4” NPT;
the insertion length is variable
between 7/8” and 5”; one mod-
el is used for both ow-through
and submersion tank-mount.
Installation does not require a pipe
wrench, strap wrench or thread sealant on the probe. This helps to reduce down time for calibration.
The P/R60C-7 probes have extended
length for a hot-tap applications. It designed to be used with P60-HTC ball
valve assembly with a variable insertion
depth between 7/8” and 14”. The probe
can be easily removed for cleaning and
calibration without emptying the tank.
The encapsulated preamplier provides
an output signal which can be transmitted 3000 feet over inexpensive cable.
For greater distances, see the P65 data
sheet for the probe that encapsulates
a blind 4-20 mA two-wire transmitter
which can transmit a virtually unlimited distance over a twisted pair cable.
Features
• Differential Measurement
• Replaceable Salt Bridge
• Low Maintenance Cost
• Easy in: no twisted cable
• Easy out: no special tools
• 1-1/4” NPT tting
• Adjustable insertion depth
• Easy Cleaning
• Encapsulated Preamp
Transmits up to 3000 ft
• 4-20 mA two-wire blind
transmitter version available
• Hot-Tap hardware available
• Flat-faced and Hardened
glass version available for pH
• ORP available with
gold electrode
Applications
• Process Control
• Industrial and Municipal Wa
ter Treatment
• Industrial and Municipal
Waste Treatment and Neu
tralization
• Fume Scrubbers, Circuit
Board Manufacturing, Plating,
Food and Beverage, Chemical Processing, Pulp and Paper, Mining, Power Generation,
Pharmaceuticals
100 School Street
Andover, MA 01810
978-749-9949
Toll free - 855-747-7623
www.WaterAnalytics.net
Model P60C-6 and P60C-7 pH or
R60C6 and R60C-7 ORP Differential Probes
Technical Data
Measuring Range
pH 0 to 14 pH
(Consult factory for applications below 2 or
above 12).
ORP -2000 to +2000 mV
Flow Rate 10 ft./sec maximum (3 metres/sec)
Flow should be as low as possible in low
conductivity water and in solutions with
high suspended solids
Wetted Materials CPVC, kynar, glass, titanium palladium al-
loy and EPDM (platinum for ORP probe)
Transmission Distance 3000 ft. (900 m)
Sensitivity
pH 0.01 pH
ORP 0.1 mV
Stability
pH 0.03 pH/day, non-cumulative
ORP 2 mV/day, non-cumulative
Temperature Copensation
300 Ohm NTC Automatic pH compensation
Pressure Limit 100 psig at 65°C maximum
Temperature Limits
CPVC -5 to 95°C (23 to 203°F)
The temperature limit of probes in ow-
through applications is limited by pressure
and by the pipe tting material.
Probe Cable 5 Conductor shielded, default length 15 ft.
(4.6m)
Dimensions
P60C6
Related Products
CABLES & ACCESORIES
AM-JB-1 NEMA 4X junction box
AM-HTC-7 Hot Tap Ball Valve assembly (Probe not included)
C42-5PXXX Interconnect cable, dressed both ends. Specify length.
AM-SBK-567 Salt bridge Kit for all -6 and -7 series probes, package of 3.
Ordering Information
P60C-6-015 pH probe complete with universal installation tting for ow-through or submersion applications
P60C-7-015 pH Probe for use with Hot-Tap hardware
P60C-6F-015 Flat-face pH probe
R60C-6-015 ORP probe complete with installation tting for low-through or submersion applications
R60C-6-G-015 ORP probe with gold electrode
R60C-7-015 ORP Probe for use with Hot-Tap hardware
P60C7