Model IP80 Series
by
Water Analytics
Aqua
Metrix
Impeller Flow Sensor
Features
• Low-friction, long-life jewel bearings
• Single moving part
• Six bladed impeller
• Fully eld-repairable
• Choice of materials for compatibility
with variety of chemicals
• Integrated Tee and sensor
• Fits 1/2” to 8” pipe sizes
• Fixed depth in tting ensures proper
placement in pipe
• High level signal can be sent up to
2000 feet
• Each tting individually calibrated
and marked with K-factor (Pulses per
Gallon)
Description
The IP80 series are impeller type (or “paddlewheel”) insertion meters designed for use with a wide variety of liquids in pipe sizes 1/2” to 8”. Sensors
are available in brass, 316 stainless steel, PVC, and polypropylene. Bodies
are machined from a solid rod for maximum precision. High-quality jewel
bearings and nickel-bound tungsten carbide shafts are used for extreme low
friction and long life.
The rotation of the rotor is detected by a non-drag Hall-effect sensor. Output
is a current-sinking pulse (square wave), which can be sent long distances
(up to 2,000 feet) without a transmitter. This signal can be connected directly
to PLC’s, counters, and computer cards, as well as a variety of AquaMetrix
controls and displays.
For rate and total display, the Shark-120/240 controller or Shark TX/P transmitter (loop powered) models are compatible with IP80 series sensors. They
can be mounted in either surface, pipe
or panel congurations.
The IP80 series require special ttings
that ensure correct depth placement in
the pipe. Fittings come in a variety of
materials for compatibility with specic
applications. Tee ttings are individually wet-calibrated at the factory and
marked with the K-factor (pulses per
gallon). Saddle ttings must be eldinstalled on the pipe and do not come
wet-calibrated. K-factors for saddles
are based on factory testing.
Applications
• Waste water ow
• Water Treatment systems
• Irrigation
• Groundwater Remediation
• Chemical Mixing
• Process cooling systems
• Neutralizaton Systems
• Process ow control
• Filtration systems
by
Toll free - 855-747-7623
www.WaterAnalytics.net
Metrix
Water Analytics
100 School Street
Andover, MA 01810
978-749-9949

Model IP80 Series Impeller Flow Sensors Technical Data
Materials
Sensor Body PVC, Polypro, Brass, or 316 SS
Rotor PVDF
Shaft Nickel-bonded tungsten carbide
(Ceramic optional)
Bearings Ruby jewel
O-Ring EPDM (Viton optional)
Pipe Size
IP81 1/2” to 3” (12 to 76 mm)
IP82 4” to 8” (100 to 200 mm)
Maximum Pressure
PVC or Polypro 175 PSI (12 bar) at 75º
Brass 200 PSI (14 bar)
316 SS 250 PSI (17 bar)
Maximum Temperature
PVC or Polypro 130ºF (55ºC)*
Brass or SS 200ºF (93ºC)
Flow Rate 0.3 to 30 FPS (0.1 to 9 M/s)
Accuracy 1-1/2% of full scale
Signal Hall effect current sinking pulse
Power 6-24 VDC, 2 mA
Maximum Current
Cable #22 AWG 3-con, 18 ft. (6m)
20 mA
Flow Range (GPM)
IP81
1/2” 3/4” 1” 1-1/2” 2” 3” 4” 6” 8”
0.28 0.5 0.8 1.9 3.1 6.9 12 27 47
Min
28 50 80 190 314 691 1200 2700 4700
Max
PVC/Polypro Working Pressure vs. Temperature
60º7 0º 80º9 0º 100º 110º 120º 130º 140º
200
150
P.S.I.
100
50
IP82
Order Information
Dimensions
PVC Tee
Size Dim. A
1.5” 19.2”
2” 19.9”
PVC Block Tee
Size Dim. B Dim. C
1/2” 1.50” 3.80”
3/4” 1.90” 4.00”
1” 1.88” 4.00”
Sensor Style
Impeller sensor ( 1/2” - 3”) 81
Impeller sensor ( 4” - 6”) 82
Sensor Material
PVC P
Stainless Steel S
Polypropylene Y
Brass B
Mounting Style
1/2” Tee tting 050
3/4” Tee tting 075
1” Tee tting 100
1.5” Tee tting 150
2” Tee tting 200
3” Saddle/Weldolet tting 300
4” Saddle/Weldolet tting 400
6” Saddle/Weldolet tting 600
Mounting Tee Material
PVC P
304 Stainless Steel (available only for 0.5” - 2”) S
Bronze (Not available for 6” pipes) B
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)