Features
Description
For years the MS series of epoxy body probes have been known as the most extensive conductivity probes on the market. The new AS series represent a major upgrade.
The AS conductivity probes feature a choice of 316 stainless steel or graphite electrodes and Noryl shells that will satisfy just about any industrial, power plant or laboratory application. This technology, insures that, with reasonable maintenance, years of drift-free performance can be expected.
Integral automatic temperature compensation is standard on all types and provides compensation to 25°C when used with a compatible analyzer. Users can choose between a 1000 Ω RTD or 3 kΩ thernistor. Either elementi is place to give close thermal contact with the liquid being mea­sured to assure fast response to changing temperature.
Submersion mounting is achieved by reversing the direction of the compression tting, allowing the cell to be threaded into the MHMSC mounting hardware. Mounting hardware is available which includes a 2 foot length of CPVC pipe and a simple but versatile adjustable bracket for hanging on the rim of the tank.
316 SS or graphite Electrodes
Economically Priced
Automatic Temperature Com-
pensation
Adjustable Insertion Length
Ball Valve “Hot Tap” Types
Constants 0.01, 0.1, 0.5, 1.0, 10
and 50.
Choice of 3 kΩ thermistor or
1000 Ω RTD for automatic temper­ature compensation.
Applications
Cooling towers
High purity water applications
Pharmaceutical Applications
Reverse osmosis systems
Food and Beverage
by
Andover, MA 01810
Toll free - 855-747-7623
www.WaterAnalytics.net
Metrix
100 School Street
978-749-9949
Model AS Conductivity Cells
Technical Data
Cell Constants 0.01, 0.1, 0.5, 1.0, 10 and 50
Thread size 3/4” NPT
Temperature Compensation* Automatic, 0 to 1050C
Cable Length 20 ft. (6.1 m)
Inser tion Length Variable. See table below
Wetted Materials 316 stainless steel, polypropylene,
Epoxy and EPDM. Compression tting: CPVC or 316 stainless steel.
Temperature and Pressure
Type Max pres. (psi) Max temp. (0C)
P 50 60
100 50
150 20
S 200 105
Safety Feature Shoulder on cells provides
Temperature and Pressure 3 metres (10 ft.) coaxial, terminated
an integral safety stop
with a spade lug for active electrode, other wires tinned
316 stainless steel, polypropylene, epoxy and EP
Dimensions
Related Products
MS Series of probes
MHMSC, 24 inch (610 mm) long 1” CPVC pipe w/ cable clamp
MS-HTC, 3/4” socket-weld CPVC ball valve assembly
AM3422, AM3444 and AM3455 316 SS body conductivity sensors
Ordering Information
Temperature Compensa-
Cell Constant Selection
Constant Code
0.01 001
0.1 01
0.5 05
1.0 1
10 10
tion Device Selection
Description Code
PT1000 RTD 1
3 kΩ ohm NTC 2
Cell with 316 stainless steel tting S
Cell for Ball Valve Assembly HT
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)
Type Selection
Description Code
Cell with PVC tting P
Loading...