APG PT-405 Specifications

Page 1
APG
Explosion Proof, Heavy Duty Pressure Transducer Series: PT-405
®
C
US
hazardous locations. With a CSA certication for Class I, Division 1, Explosion Proof locations, an IP65 rated sensor
body, and single seal construction for the highest safety factor, the PT-405 can survive in rough environments. With
a standard accuracy of 0.25% BFSL, and NIST certicates available for 0.5%, 0.25%, and 0.1%, the PT-405 is sensitive
enough to give you precise pressure measurements over its 30,000 psi range. Durability, reliablity, and sensitivity, all in one tough little package.
Features
Available ranges from -15 - 30,000 psi
• Standard Outputs: 4-20 mA, 0-5 VDC, 0.5-4.5 VDC,
1-5 VDC, 0-10 VDC, RS-485
• High overpressure capability
• Zero and Span adjustments
• cCSAus hazardous location approved (up to 30,000 psi)
• Single seal for highest safety protection factor
Automation Products Group, Inc. | 888.525.7300 | sales@apgsensors.com | www.apgsensors.com
Rev. B, 07/24/19
Page 2
APG
PT-405 Specications
2.50”
63.5 mm
2.50”
63.5 mm
*Overall length may vary depending on process connection.
3.90”
99.06 mm
1.50”
38.10 mm
0.55”
13.97 mm
Performance
• Accuracy (linearity & hysteresis): ±0.25% of full scale (BFSL)
• Standard Pressure Ranges: -15 - 30,000 psi
• Stability - One Year Zero Drift:
17-4 / 316L: <±0.5% FS
• Overpressure: 1.5x full scale
• Burst Pressure: up to 3x full scale or limit of
process connection
• Frequency Response:
Less than 5ms
Connectivity
• Output: 4-20 mA (2 wire, loop-powered)
0-5, 0.5-4.5, 1-5, 0-10 VDC (non-isolated 3 wire) Modbus/RTU (RS-485) with temp. output
Environmental
• Standard Compensated Temp.:
≤ 10 psi: 0º - 60ºC / 32º - 140ºF 10 < x ≤ 1000 psi: -10º - 70ºC / 14º - 158ºF > 1000 psi: -17º - 54ºC / 0º - 130ºF
• Storage Temp: -40º - 82ºC / -40º - 180ºF
• Operating Temp: -40º - 85ºC / -40º - 185ºF
Electrical
• Supply Voltage:
4-20 mA, 0-5 / 0.5-4.5 / 1-5 VDC: 9-28 VDC 0-10 VDC: 12.5-28 VDC Modbus/RTU (RS-485): 9-28 VDC
• Electrical Connection:
Pigtail with ying leads (4-foot minimum length)
• Electrical Protection: Protected against reverse polarity, surge per
IEC 61000-4-5
Certication
• CSA/cCSAus Contract #237484
Ambient: -40º to 85º C
Max. Working Pressure: 30,000 psi Single Seal
4-20mA, 0-5/0.5-4.5/1-5 VDC, 0-10 VDC, RS-485/Modbus
- Explosion Proof: Class I, Div. 1, Groups C & D
Physical
• Weight: 0.9 - 1.5 lbs. (408 - 680 g) typical
• Wetted Materials: 316L Stainless Steel ( ≤ 1,000 psi)
17-4 Stainless Steel ( > 1,000 psi) Incoloy ( > 10,000 psi)
• Incoloy conforms to NACE MR-01-75 and ISO 1516-3
Automation Products Group, Inc. | 888.525.7300 | sales@apgsensors.com | www.apgsensors.com
®
2
Rev. B, 07/24/19
Page 3
APG
Common Model Congurations
Short can, 1/2” NPTM 5 ft ying leads, Standard Temperature Range
Model Number Model Description
PT-405-L1-15-PSIA-E60-5-P0-N12-M1-S10 4-20 mA, 0-15 PSI, Absolute Pressure, 316L SS, 0.5% Accuracy, -10° - 70°C
PT-405-L1-5-PSIS-E60-5-P0-N12-M1-S9 4-20 mA, 0-5 PSI, Sealed Gauge Pressure, 316L SS, 0.5% Accuracy, 0° - 60°C
PT-405-L1-15-PSIS-E60-5-P0-N12-M1-S10 4-20 mA, 0-15 PSI, Sealed Gauge Pressure, 316L SS, 0.5% Accuracy, -10° - 70°C
PT-405-L1-30-PSIS-E60-5-P0-N12-M1-S10 4-20 mA, 0-30 PSI, Sealed Gauge Pressure, 316L SS, 0.5% Accuracy, -10° - 70°C
PT-405-L1-50-PSIS-E60-5-P0-N12-M1-S10 4-20 mA, 0-50 PSI, Sealed Gauge Pressure, 316L SS, 0.5% Accuracy, -10° - 70°C
PT-405-L1-100-PSIS-E60-5-P0-N12-M1-S10 4-20 mA, 0-100 PSI, Sealed Gauge Pressure, 316L SS, 0.5% Accuracy, -10° - 70°C
PT-405-L1-200-PSIS-E60-5-P0-N12-M1-S10 4-20 mA, 0-200 PSI, Sealed Gauge Pressure, 316L SS, 0.5% Accuracy, -10° - 70°C
PT-405-L1-300-PSIS-E60-5-P0-N12-M1-S10 4-20 mA, 0-300 PSI, Sealed Gauge Pressure, 316L SS, 0.5% Accuracy, -10° - 70°C
PT-405-L1-500-PSIS-E60-5-P0-N0-M1-S10 4-20 mA, 0-500 PSI, Sealed Gauge Pressure, 316L SS, 0.25% Accuracy, -10° - 70°C
PT-405-L1-1000-PSIS-E60-5-P0-N0-M1-S10 4-20 mA, 0-1 kPSI, Sealed Gauge Pressure, 316L SS, 0.25% Accuracy, -10° - 70°C
PT-405-L1-5000-PSIS-E60-5-P0-N0-M2-S0 4-20 mA, 0-5 kPSI, Sealed Gauge Pressure, 17-4 SS, 0.25% Accuracy, -17° - 54°C
PT-405-L1-7500-PSIS-E60-5-P0-N0-M2-S0 4-20 mA, 0-5 kPSI, Sealed Gauge Pressure, 17-4 SS, 0.25% Accuracy, -17° - 54°C
PT-405-L1-10000-PSIS-E60-5-P0-N0-M2-S0 4-20 mA, 0-10 kPSI, Sealed Gauge Pressure, 17-4 SS, 0.25% Accuracy, -17° - 54°C
PT-405-L1-10000-PSIS-E60-5-P56-N12-M7-S0 4-20 mA, 0-10 kPSI, Sealed Gauge Pressure, Incoloy, 0.5% Accuracy, -17° - 54°C
PT-405-L1-15000-PSIS-E60-5-P56-N12-M7-S0 4-20 mA, 0-15 kPSI, Sealed Gauge Pressure, Incoloy, 0.5% Accuracy, -17° - 54°C
Automation Products Group, Inc. | 888.525.7300 | sales@apgsensors.com | www.apgsensors.com
®
3
Rev. B, 07/24/19
Page 4
APG
Model Conguration Options
Model Number: PT-405 - _____ - _____ - _____ - _____ - _____ - _____ - _____ - _____ - _____ - _____
A B C D E F G H I J
®
A. Operation / Output
L1▲ 4 - 20 mA output □ L3 0 - 5 VDC output L10 0 - 10 VDC output L12 1 - 5 VDC output L20 0.5 - 4.5 VDC output
Modbus
L5 RS-485 (Modbus/RTU), 4-wire
Pressure reading (psi) only
L31 RS-485 (Modbus/RTU), 4-wire
Level calculations (mmH2O), tank volume
B. Common Pressure Ranges - PSI*
5 50 200 1000 5000
15 60 300 2000 10000
30 100 500 3000 30000
*Other ranges available. Please consult factory.
C. Units of Measure
psi
O fWC mmH2O inHG
fH
2
bar kPa inH2O inWC
D. Pressure Type
A Absolute ( ≤ 200 psi)
S▲ Sealed ( ≤ 30,000 psi)
F. Electrical Cable Length*
__ Number represents length of ying leads in feet.
4-ft length required minimum. 25-ft maximum.
(ex. E60-10 equals short can, 10 ft leads)
*Note: All cable must be in conduit.
G. Process Connection
P0▲ 1/4 - 18 NPTM ( ≤ 10,000 psi) P1 1/2 - 14 NPTM ( ≤ 10,000 psi) P5 1/4 - 18 NPTF ( ≤ 15,000 psi) P6 1/2 NPTF ( ≤ 10,000 psi) P38 1 1/2 in. tri-clover with 3/4 in. diaphragm ( ≤ 1,000 psi) P52 1 1/2 in. NPTM ( ≤ 1,000 psi) P54 7/16 - 20 UNJF-3A Male w/ Cone ( ≤ 1,000 psi) P56 F250C High Pressure (10,000 psi - 30,000 psi) P57 F560C40 High Pressure (10,000 psi - 30,000 psi)
H. Accuracy
1 - 5,000 PSI
N0*▲ ±0.25% (1% for pressure ≤ 1 psi) □ N1* ±0.25% with NIST certication N2 ±0.1% with NIST certication N12 ±0.5% (required for < 500 psi) N13 ±0.5% with NIST certication
*Note: ±0.25% available up to 10,000 psi for 4-20 mA output only.
Up to 30,000 PSI
N12 ±0.5% N13 ±0.5% with NIST certication
E. Electrical Connection
E60 1/2 in. NPTM with ying leads, short can† E60L 1/2 in. NPTM with lying leads, long can††
Standard for L1, L3, L10, L12, L20. Not available for L5, L31. †† Standard for L5, L31.
This option is standard
I. Materials
M1▲ 316L SS ( ≤ 1,000 psi) □ M2 17-4 SS ( > 1,000 psi) □ M7 Incoloy (10,000 - 30,000 psi; P56/P57 only)
J. Compensated Temperature Range
S0▲ -17º - 54ºC / 0º - 130ºF ( > 1000 psi) □ S1 -40º - 82ºC / -40º - 180ºF ( > 1,000 psi) S3 -34º - 77ºC / -30º - 170ºF ( > 1,000 psi)S9▲ 0º - 60ºC / 32º - 140ºF ( ≤ 10 psi) □ S10▲ -10º - 70ºC / 14º - 158ºF (10 < x ≤ 1000 psi)
Automation Products Group, Inc. | 888.525.7300 | sales@apgsensors.com | www.apgsensors.com
Document #9005321, Part #200125 Rev. B, 07/24/19
Loading...