APG warrants its products to be free from defects of material and workmanship
and will, without charge, replace or repair any equipment found defective
upon inspection at its factory, provided the equipment has been returned,
transportation prepaid, within 24 months from date of shipment from factory.
THE FOREGOING WARRANTY IS IN LIEU OF AND EXCLUDES
ALL OTHER WARRANTIES NOT EXPRESSLY SET FORTH HEREIN,
WHETHER EXPRESSED OR IMPLIED BY OPERATION OF LAW OR
OTHERWISE INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE.
No representation or warranty, express or implied, made by any sales
representative, distributor, or other agent or representative of APG which is
not specifi cally set forth herein shall be binding upon APG. APG shall not be
liable for any incidental or consequential damages, losses or expenses directly
or indirectly arising from the sale, handling, improper application or use of the
goods or from any other cause relating thereto and APG’s liability hereunder, in
any case, is expressly limited to the repair or replacement (at APG’s option) of
goods.
Warranty is specifi cally at the factory. Any on site service will be provided at
the sole expense of the Purchaser at standard fi eld service rates.
All associated equipment must be protected by properly rated electronic/
electrical protection devices. APG shall not be liable for any damage due
to improper engineering or installation by the purchaser or third parties.
Proper installation, operation and maintenance of the product becomes the
responsibility of the user upon receipt of the product.
Returns and allowances must be authorized by APG in advance. APG will
assign a Return Material Authorization (RMA) number which must appear
on all related papers and the outside of the shipping carton. All returns are
subject to the fi nal review by APG. Returns are subject to restocking charges as
determined by APG’s “Credit Return Policy”.
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
The PG10 has 4 standard operating modes to choose from:
Mode 000: Full Access
- Provides access to all menu settings. If no buttons are pushed for 1 minute,
the gauge will revert back to Mode 003 (factory default).
Mode 002: Limited Access
- Menu is locked.
ENTR
- zeros the reading (press and hold for 3 seconds).
- or cycles between the Max and Min readings.
MODE
- On/Off functions only on battery powered gauges.
Mode 003: Factory Default
- Full access except Full Scale Calibration is locked.
Mode 005: Locked Access
- All buttons locked except will power on/off battery powered gauges.
MODE
Accessing the Operating Mode Setting:
Step 1: Simultaneously press and hold and for approximately
MODE
3 seconds. This will bring up the 3 digit mode number.
Step 2: Enter the desired mode number (see Operating Modes above) by
using or to change the value of the fl ashing digit, and to accept the value and advance to the next digit.
ENTR
ENTR
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Pressing the or button while in operating mode will cycle between
displaying the present pressure reading, the maximum pressure reading and
the minimum pressure reading. The maximum and minimum readings will be
stored until the gauge is powered down or the max/min readings are reset.
Resetting the Max/Min readings:
Step 1: Press once to enter the main setup menu.
MODE
Step 2: Press or cycle through the options until MAXMIN is
displayed.
Step 3: Press to access the Max/Min reset options.
ENTR
Step 4: Press or to toggle between YES and NO until YES is
displayed.
Step 5: Press to reset the Max/Min readings and return to the main setup
ENTR
menu.
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Allows the user to select the unit of measure to be displayed as the pressure
reading.
Options:
For gauges over 120 psi:
PSI (pounds per square inch)
bAR (bar)
KPA (kilopascals)
*CUSTOM (see “Using Custom Units” on next page)
KGCM^2 (kilograms per cubic centimeter)
MPA (megapascals)
FTH20 (feet of water @ 60 F)
cmHG (centimeters of mercury)
For gauges less than 120 psi:
PSI (pounds per square inch)
bAR (bar)
KPA (kilopascals)
*CUSTOM (see “Using Custom Units” on next page)
mbAR (millibar)
INHG (inches of mercury)
INH20 (inches of water @ 60 F)
mmHG (millimeters of mercury)
Setting the Unit of Measure:
Step 1: Press once to enter the main setup menu.
MODE
Step 2: Press orto cycle through the options until UNITS is displayed.
Step 3: Press to access the Units options.
ENTR
Step 4: Press or to cycle through setting options until the desired unit of measure is displayed.
Step 5: Press to apply the setting and return to the main setup menu.
APG...Providing tailored solutions for measurement applications
When the Peak-Hold is enabled, the gauge will display the “peak” or maximum
pressure reading since the gauge was powered on or the Max/Min value was
reset.
NOTE 1: When the Peak-Hold feature is enabled, the icon will be
PEAK
displayed in the upper left area of the display.
Options: Off or On
Enabling the Peak-Hold feature:
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press orto cycle through the options until P HOLd is displayed.
Step 3: Press to access the Peak-Hold setting options.
ENTR
Step 4: Press orto toggle between OFF and ON.
Step 5: Press to apply the displayed setting and return to main setup menu.
ENTR
NOTE 2: The peak value can be reset by pressing
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
The Advanced Settings menu is used to customize the LCD display and to setup
any optional features, such as an analog output.
Auto-Off (AUTO):
This function is applicable to battery powered units only. The Auto-Off feature
allows the user to designate the time of inactivity (no buttons pushed) until the
gauge automatically powers down.
Options: OFF, 2 MIN, 4 MIN, 8 MIN, 16 MIN, 32 MIN and LIGHT
NOTE 1: Selecting OFF disables the Auto-Off feature; the gauge will then
remain powered indefi nitely so long as suffi cient voltage is being supplied
(>1.8V).
NOTE 2: Selecting LIGHT will keep the gauge continuously powered as
long as the light sensor in the PG10 detects adequate light levels. When the
environment is suffi ciently dark, the PG10 will automatically shut down until
the lighting once again reaches the threshold of the light sensor.
Setting the Auto-Off feature:
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the menu options until AdVSET is displayed.
Step 3: Press to enter the Advanced Settings menu.
ENTR
Step 4: Press or to cycle through the menu options until AUTO is displayed.
Step 5: Press to access the Auto-Off setting options.
ENTR
Step 6: Press or to cycle through setting options until the desired setting is displayed.
Step 7: Press to apply the setting and return to advanced setup menu.
ENTR
Step 8: To exit the advanced setup menu, press or until EXIT is displayed and press to exit to the main setup menu.
APG...Providing tailored solutions for measurement applications
The reading can be set to display in High Resolution (HI RES), Medium
Resolution (MEdRES) or Low Resolution (LO RES) mode. Switching between
resolutions will shift the displayed reading by one decimal place position.
NOTE: Gauges without a decimal place position will display a dummy zero (or
zeros) when the resolution is changed to medium or low.
Options: HI RES (high resolution), MEdRES (medium resolution) or LO RES
(low resolution)
Setting the Decimal Place feature:
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until AdVSET is displayed.
Step 3: Press to enter the Advanced Settings menu.
ENTR
Step 4: Press or to cycle through the options until dEC PL is
displayed.
Step 5: Press to access the Decimal Place setting options.
ENTR
Step 6: Press orto cycle through the resolution settings.
Step 7: Press to apply the displayed setting and return to advanced setup
ENTR
menu.
Step 8: To exit the advanced setup menu, press or until EXIT is displayed and press to exit to the main setup menu.
ENTR
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Adjusts the rate at which the gauge takes sample readings.
NOTE: Setting the Sample Rate to “SLOW” will help preserve battery life
(when applicable) and will also help to smooth rapidly fl uctuating
readings.
Options: SLOW (4x/second), MEdIUM (8x/second), FAST (16x/second)
Setting the Sample Rate feature:
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until AdVSET is
displayed.
Step 3: Press to access the Advanced Settings menu.
ENTR
Step 4: Press or to cycle through the options until SAMPLE is
displayed.
Step 5: Press to access the Sample Rate setting options.
ENTR
Step 6: Press or to cycle through the setting options until the desired setting is displayed.
Step 7: Press to apply the displayed setting and return to advanced setup
ENTR
menu.
Step 8: To exit the advanced setup menu, press or until EXIT is displayed and press to exit to the main setup menu.
ENTR
14
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Bar Graph 0% (bAR 0) & Bar Graph 100% (bAR100) Settings:
Allows the user to defi ne the reading values associated with 0% and 100% on
0
the 270
display bar graph. Bars will appear/disappear in 10% increments of the
total span between the two values.
NOTE: The 0% reference does not have to be the lower pressure setting;
0% can be set as the higher pressure setting, thereby causing the bar graph to
increase as the pressure decreases. Negative pressure settings can also be used
as either the 0% or 100% reference points.
Setting the Display Bar Graph:
Step 1: Press once to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until AdVSET is displayed.
Step 3: Press to access the Advanced Settings menu.
ENTR
Step 4: Press or to cycle through the options until BAR 0 is displayed.
Step 5: Press to access the Bar Graph 0% value. A 5-digit number will
ENTR
appear with the fi rst digit fl ashing.
Step 6: Press or to change the value of the fi rst fl ashing digit.
Step 7: Press to accept the value of the fl ashing digit and advance to the
ENTR
next digit......repeat steps 6 and 7 until the desired 0% reading is fully
entered.
Step 8: Press or to cycle through the options until BAR100 is displayed.
Step 9: Repeat Steps 5-7 to enter the Bar Graph 100% value.
Step 10: To exit the advanced setup menu, press or until EXIT is
displayed and press to exit to the main setup menu.
ENTR
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Allows the user to adjust the reading at full-scale pressure. The reading can be
adjusted by +/-10% full-scale.
NOTE: The pressure reading must be within 5% of the full-scale value in order
to make Range adjustments. For example, a 1000 psi gauge would need
to be reading between 950 psi and 1050 psi in order to adjust the Range
feature. If the reading is not within 5% of full scale, NOAdJU (No
Adjustment) will be displayed when trying to adjust the Range.
Adjusting the Full-Scale Range:
Step 1: Ensure that the pressure reading is within 5% of full-scale.
Step 2: Press to enter the main setup menu.
MODE
Step 3: Press or to cycle through the options until AdVSET is displayed.
Step 4: Press to access the Advanced Settings menu.
ENTR
Step 5: Press or to cycle through the options until RANGE is displayed.
Step 6: Press to enter the Range adjust mode.
ENTR
Step 7: Press to increase the reading, or press to decrease the reading.
Step 8: Press to accept the adjusted reading and return to the advanced
MODE
setup menu.
Step 9: To exit the Advanced Setup menu, press or until EXIT is displayed and press to exit to the main setup menu.
ENTR
16
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Allows the user to perform a fast calibration of the zero and span.
NOTE: Both zero pressure and full-scale pressure must be applied to the gauge
in order to complete the quick calibration process.
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the menu options until AdVSET is displayed.
Step 3: Press to enter the Advanced Settings menu.
ENTR
Step 4: Press or to cycle through the options until CALIBR is displayed.
Step 5: Press to access the CALIBR mode. The gauge will display the
ENTR
word ZERO, prompting the user to perform the zero pressure quick calibration.
Step 6: Ensure no pressure is applied to the gauge. Press to zero the gauge.
The gauge will then display the word SPAN, prompting the user to perform the full-scale pressure quick calibration.
Step 7: Apply full pressure to the gauge and press to complete the quick
ENTR
calibration and return to advanced setup menu.
Step 9: To exit the advanced setup menu, press or until EXIT is displayed and press to exit to the main setup menu.
ENTR
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
When datalogging is enabled, the PG10 will log the pressure readings, along
with the associated date/time stamps, at an interval determined by the Logging
Sample Rate setting (see page 20).
The PG10 can be confi gured to log continuously, based solely on the Logging
Sample Rate setting, or to log only when user defi ned pressure conditions exist
(event logging). The gauge will continue to log readings at the Logging Sample
Rate interval for as long as the event conditions persist.
The gauge can store a maximum of 60 records. Each record is logged in
a FIFO (fi rst-in, fi rst-out) buffer. This means that once the buffer is full (60
records), the oldest record will be overwritten by the next logged reading.
18
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Used to set the date and time, which will be used as the reference for the time
stamp of the logged readings.
NOTE: Hours are set using the 24 hour system (e.g. 1:00 p.m. is 13:00).
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the menu options until dATLOG is
displayed.
Step 3: Press to enter the Datalogging menu options.
ENTR
Step 4: Press or to cycle through the menu options until DATE is displayed.
Step 5: Press to access the Date settings. A 2-digit month number will
ENTR
appear with the fi rst digit fl ashing.
Step 6: Press orto cycle the fi rst digit, and to accept the value
ENTR
and advance to the second digit. Repeat the same process for the second digit and press to accept and advance to the day of the
ENTR
month setting, which will appear with the fi rst digit fl ashing. Repeat the process one last time by entering the year.
Step 7: Press to apply the setting and return to Datalogging menu.
ENTR
Step 8: Press or to cycle through the menu options until TIME is displayed.
Step 9: Press to access the time settings. Repeat the process used to set the
ENTR
date to set the time (hour, minute, seconds).
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Used to enable/disable datalogging, and to clear all logged records.
Setting the Logging Type (continuous or event):
The PG10 can be set to continuously log readings or to only log readings based
on the Trip Relay status.
NOTE 1: Gauges not equipped with relay outputs will still use the Trip Relay
settings to defi ne the pressure range(s) for the event logging (see pages 30-32
programming details).
NOTE 2: Whenever the PG10 is logging, whether in continuous or event mode,
the interval between logged readings is determined by the Logging Sample Rate
setting (see page 20).
Enabling Continuous or Event Logging
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until dATLOG is displayed.
Step 3: Press to enter the Datalogging menu.
ENTR
Step 4: Press orto cycle through the options until LOG is displayed.
Step 5: Press to access the Logging Control options.
ENTR
Step 6: Press or to cycle through the setting options until CONTIN or
EVENT is displayed.
Step 7: Press to enable logging and return to logging control menu.
ENTR
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
When Event Logging is enabled, the gauge will log readings based on the status
of one or both of the Trip Relays (see Trip Relay programming on pages 30-32).
NOTE: Gauges not equipped with relay outputs will still use the Trip Relay
settings for event logging.
Event logging can be set to initiate on one of the following conditions:
When Trip 1 is On (T1ON)
When Trip 2 is On (T2ON)
When both Trip 1 & Trip 2 are On (T1T2ON)
When both Trip 1 & Trip 2 are Off (T1T2OF)
When Trip 1 is Off (T1OFF)
When Trip 2 is Off (T2OFF)
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until dATLOG is displayed.
Step 3: Press to enter the Datalogging menu.
ENTR
Step 4: Press orto cycle through the options until EVENT is displayed.
Step 5: Press to access the Event options.
ENTR
Step 6: Press or to cycle through the options until the desired Event condition is displayed.
Step 7: Press to accept the event condition and return to the Datalogging
ENTR
menu.
22
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Disabling Logging (STOP) and Clearing Logged (CLEAR) Readings:
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until dATLOG is displayed.
Step 3: Press to enter the Datalogging menu.
ENTR
Step 4: Press orto cycle through the options until LOG is displayed.
Step 5: Press to access the Logging Control options.
ENTR
Step 6: Press or to cycle through the setting options until STOP or CLEAR is displayed.
Step 7: Press to apply the displayed option (stop or clear) and return to
ENTR
Datalogging menu.
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
The latest reading will be the fi rst record displayed when viewing the logged
readings.
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until dATLOG is displayed.
Step 3: Press to enter the Datalogging menu.
ENTR
Step 4: Press orto cycle through the options until VIEW is displayed.
Step 5: Press to access the logged readings.
ENTR
Step 6: Press to cycle back through progressively older readings. will cycle forward through progressively newer records.
Step 7: Press to exit the logged readings.
ENTR
NOTE: When viewing a logged record, the time stamp of the record will toggle
momentarily to display the date stamp once every 8 seconds.
24
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Enabling the Tare function will set the current pressure reading as the zero
reference pressure in order to measure a net change in pressure as opposed to
measuring the gross pressure.
Enabling the Tare feature:
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until TARE is displayed.
Step 3: Press to access the Tare setting options.
ENTR
Step 4: Press or to toggle between OFF and ON.
Step 5: Press button to apply the displayed option and return to main setup
ENTR
menu.
NOTE 1: When tare function is enabled, the icon will appear in the upper
TARE
center of the display.
WARNING: Do NOT disconnect the gauge from the pressure fi tting while
the tare function is enabled. The gauge could still be under pressure even
though the reading shows 0.
NOTE 2: If the maximum gross full-scale pressure value is reached while the
Tare feature is enabled, the PG10 will automatically disable the Tare feature
and return to reading the gross pressure in order to help prevent the user from
accidently overpressuring the gauge.
NOTE 3: Pressing and holding with the tare function enabled will re-tare
ENTR
the gauge at the current pressure value.
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Used to confi gure any optional gauges outputs, such as an analog signal or trip
point relays.
Analog Low (AL SET) & Analog High (AH SET) Setpoints:
Allows the user to defi ne the reading values associated with the Low Analog
signal (i.e. 4mA or 0V) and the High Analog signal (i.e. 20mA, 2V or 5V).
NOTE: The analog setpoints must be entered using the gauge’s base unit of
measure (typically PSI).
Setting the Analog Signal Span:
Step 1: Press to enter the main setup menu.
MODE
Step 2: Press or to cycle through the options until OUTPUT is displayed.
Step 3: Press to access the Output Settings menu.
ENTR
Step 4: Press or to cycle through the options until AL SET is displayed.
Step 5: Press to access the Analog Low setpoint value. A 5-digit number
ENTR
will appear with the fi rst digit fl ashing.
Step 6: Press or to change the value of the fi rst fl ashing digit.
Step 7: Press to accept the value of the fl ashing digit and advance to the
ENTR
next digit......repeat steps 6 and 7 until the desired Analog Low reading
is fully entered.
Step 8: Press to reenter the Output Settings menu.
ENTR
Step 9: Press or to cycle through the options until AH SET is displayed.
Step 10: Repeat Steps 5-7 to enter the Analog High setpoint value.
Step 11: To exit the output menu, press or until EXIT is displayed and
press to exit to the main setup menu.
28
ENTR
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Analog Low (AL CAL) & Analog High (AH CAL) Calibration:
Allows the user to calibrate or “trim” the endpoints of the analog signal output
(i.e. 4mA & 20mA or 0V & 2V/5V)
Calibrating the Analog Signal End-Points:
Step 1: Use a calibrated meter to monitor the analog output signal.
Step 2: Force a low analog output signal (i.e. 4mA or 0V) either by adjusting the applied pressure or by adjusting the Analog Setpoints (see “AnalogSetpoints”on page 18 for details).
Step 3: Press to enter the main setup menu.
MODE
Step 4: Press or to cycle through the options until Output is displayed.
Step 5: Press to access the Output Settings menu.
ENTR
Step 6: Press or to cycle through the setup options until AL CAL is
displayed.
Step 7: Press to access the Analog Low Calibration value. A 5-digit
ENTR
number will appear with the fi rst digit fl ashing.
Step 8: Press or to change the value of the fl ashing digit.
NOTE: Changing the digit farthest to the left will produce the coarsest
adjustment, while each successive digit moving to the right will cause
subsequently fi ner adjustments.
Step 9: Press to accept the value of the fl ashing digit and advance to the
ENTR
next digit......repeat steps 7 and 8 until the desired Analog Output
value is displayed on the meter.
Step 10: Press to reenter the Output Settings menu.
ENTR
Step 11: Press or to cycle through the options until AH CAL is
displayed.
Step 12: Repeat Steps 7-9 to adjust the Analog High Calibration value.
Step 13: To exit the output menu, press or until EXIT is displayed and
press to exit to the main setup menu.
APG...Providing tailored solutions for measurement applications
Trip Points: The PG10 can be confi gured with dual relay trip-point outputs.
The relays can either be solid-state or SPDT mechanical in form. Each trip point
(T1 and T2) can be confi gured to perform one of six different logic functions as
described below and on the chart on the next page.
NOTE:The or icon will appear in the upper left area of the
T1T2
display whenever the associated trip point is active.
Trip Relay Output Types (T1TYPE or T2TYPE): determines the logic
function of the output relays (see chart on next page).
“Type 0” or Normally Closed setting will close the trip relay whenever the
pressure is less than the Trip Pressure setting.
“Type 1” or Exclusive setting will close the trip relay whenever the pressure is
less than the Trip Pressure or greater than the Trip Pressure + Trip Window.
“Type 2” or Normally Closed with Hysteresis setting will close the trip relay
when the pressure drops below the Trip Pressure. Once closed, the trip relay
will remain closed until the pressure becomes greater than the Trip Pressure +
Trip Window, at which point the trip relay will open. Once open the trip relay
will remain open until the pressure once again drops below the Trip Pressure
setting.
“Type 3” or Normally Open setting will close the relay whenever the pressure
is greater than the Trip Pressure setting.
“Type 4” or Inclusive setting will close the relay whenever the pressure is
within the Trip Window pressure range (greater than the Trip Pressure, less than
the Trip Pressure + Trip Window).
“Type 5” or Normally Open with Hysteresis is used for pump up applications,
or for a low level alarm with hysteresis function. This is the opposite function
of the Type 2.
“Type 6” Trip Disable
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Step 2: Press or to cycle through the options until OUTPUT is displayed.
Step 3: Press to access the Output Settings menu.
ENTR
Step 4: Press or to cycle through the options until T1TYPE (or
T2TYPE) is displayed.
Step 5: Press to access the Trip Type setting.
ENTR
Step 6: Press or to change the setting to the desired trip type function.
Step 7: Press to accept the trip type function and return to the Output
ENTR
Settings menu.
Step 8: Press or to cycle through the Output menu options until T1PRES (or T2PRES) is displayed.
Step 9: Press to enter the Trip Pressure setting. A 5-digit number will
ENTR
appear with the fi rst digit fl ashing.
Step 10: Press or to change the value of the fi rst fl ashing digit.
Step 11: Press to accept the value of the fl ashing digit and advance to the
ENTR
next digit......repeat steps 10 and 11 until the desired Trip Pressure is
fully entered.
Step 12: Press or to cycle through the Output menu options until T1WIN
(or T2WIN) is displayed.
Step 13: Press to enter the Trip Window setting. A 5-digit number will
ENTR
appear with the fi rst digit fl ashing.
Step 14: Press or to change the value of the fi rst fl ashing digit.
Step 15: Press to accept the value of the fl ashing digit and advance to the
ENTR
next digit......repeat steps 14 and 15 until the desired Trip Window is
fully entered.
Step 11: To exit the output menu, press or until EXIT is displayed and
press to exit to the main setup menu.
ENTR
32
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
4-20 mA Output:
Input Voltage (Excitation): 9 VDC min (no load) to 28 VDC max
Input Current: 3-30 mA max
Signal Variance: +/-0.16 mA at set points
Output/Input: 2 wire loop powered
Resolution: 14 bit
Protection: Reversed polarity
0-2 VDC Output:
Input Voltage (excitation): Battery powered
Output: Zero set point is +/-0.15 V with a 2 VDC span +/-0.02 VDC
Output/Input: 2 wire
Resolution: 14 bit
0-5 VDC Output:
Input Voltage (Excitation): 9 to 28 VDC
Input Current: 6 mA max
Output: 0-5 VDC / +/-0.5 VDC at set points
Output/Input: Non-isolated 3 wire
Resolution: 14 bit
Protection: Reversed polarity
Solid State Relay Trip Points:
Maximum Switched Voltage: 120 V AC/DC