Thank you for purchasing an MPX series magnetostrictive level sensor from APG. We appreciate your
business and your trust. Please take a few minutes to familiarize yourself with your MPX and this manual.
The MPX level sensors provide highly accurate and repeatable level readings in a wide variety of liquid level
measurement applications. The MPX-F is certied for installation in hazardous areas in the US and Canada
by CSA for Class I, Division 2 and Class I, Zone 2 environments. The MPX-F’s exible stainless steel stem
allows for accurate measurements in environments that are not straight-forward.
Reading your label
Every APG instrument comes with a label that includes the instrument’s model number, part number,
and serial number. Please ensure that the part number on your label matches your order. The following
electrical ratings and approvals are also listed on the label. Please refer to the CSA Certicate of Compliance
at the back of this manual for further details.
Electrical ratings
Rated 12 - 24 VDC, 4-20 mA, or 80 mA
Class I, Division 2, Groups C, and D
Ex nA IIB
Class I, Zone 2; AEx nA IIB
C
Non-Incendive Wiring Requirements: Vmax Ui= 28VDC, Imax Ii = 200ma, Ci = 0nF, Li = 0μH
US
IMPORTANT: MPX-F level sensor MUST be installed according to drawing 9003468
(Hazardous Installation and Non-Incendive Wiring Drawing) on page 30 to meet listed
approvals. Faulty installation will invalidate all safety approvals and ratings.
DANGER: OPEN CIRCUIT BEFORE REMOVING COVER or KEEP COVER TIGHT WHILE CIRCUITS ARE ALIVE;
AVERTISSEMENT -- OUVRIR LE CIRCUIT AVANT D’ENLEVER LE COUVERCLE, or GARDER LE COUVERCLE BIEN
FERME TANT QUE LES CIRCUITS SONT SOUS TENSION.
DANGER: WARNING -- EXPLOSION HAZARD -- SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY
FOR CLASS I, DIVISION 2;
AVERTISSEMENT -- RISQUE D’EXPLOSION -- LA SUBSTITIOND E COMPOSANTSP EUTR ENDRE CE MATERIEL
INACCEPTABLE POUR LES EMPLACEMENTS DE CLASSE I, DIVISION 2.
DANGER: WARNING -- EXPLOSION HAZARD -- DO NOT DISCONNECT EQUIPMENT UNLESS POWER HAS
BEEN SWITCHED OFF OR THE AREA IS KNOWN TO BE NON-HAZARDOUS;
AVERTISSEMENT -- RISQUE D’EXPLOSION -- AVANT DE DECONNECTER L’EQUIPEMENT, COUPER LE COURANT
OU S’ASSURER QUE L’EMPLACEMENT EST DESIGNE NON DANGEREUX.
Note: stem RTDs default to 6” from bottom of probe
C. Housing Type
All Housing Die-cast Aluminum, NEMA 4X, IP68, Blue
▲
□ __
Large Housing
□ B Large Housing with window
D. Float 1 (Top Float)
□ Z 5.5h x 3d in. Red Polyurethane (0.65 SG)
□ Y 5.5h x 3d in. Blue Polyurethane (0.94 SG)
□ X 5 in. Round 316L SS (0.52 SG)
□ W 5 in. Round 316L SS (0.92 SG)
□ V 6h x 3d in. Oval 316L SS (0.58 SG)
□ U 6h x 3d in. Oval 316L SS (0.94 SG)
□ T 3 in. Round 316L SS (0.60 SG)
□ S 3 in. Round 316L SS (0.94 SG)
□ R 5.5h x 2.8d in. Red Polyurethane (0.59 SG)
□ P 5.5h x 2.8d in. Blue Polyurethane (0.94 SG)
□ M 5.5h x 2d in. Red Polyurethane (0.57 SG)
□ L 5.5h x 2d in. Blue Polyurethane (0.94 SG)
□ N None
□ O Other
E. Float 2 (optional)
□ N None
□ Y 5.5h x 3d in. Blue Polyurethane (0.94 SG)
□ W 5 in. Round 316L SS (0.92 SG)
□ U 6h x 3d in. Oval 316L SS (0.94 SG)
□ P 5.5h x 2.8d in. Blue Polyurethane (0.94 SG)
□ L 5.5h x 2d in. Blue Polyurethane (0.94 SG)
□ O Other
F. Mounting Type
□ F Flat Face ANSI Flange 150#
(Sizes: 2, 2.5, 3, 4, 5, 6)
□ S Tri Clamp (Sizes: 2, 2.5, 3)
□ P NPT Plug 150# (Sizes: 2, 2.5, 3, 4)
□ N None
□ O Other
G. Mounting Size
□ __ See Mounting Type for available sizes
H. Mounting Connection
□ W Welded (xed)
□ S Slide with Compression Fitting (adjustable)
I. Stem Material
□ B 316L SS
J. Total Stem Length in Inches
□ __ Min. 120 in. - Max. 384 in.
L. Float Stop Options
▲
□ __
1 in. Stem, 316L SS, 1 piece, 1.5 in. OD
□ C 1 in. Stem, 316L SS, 2 piece, 1.75 in. OD
□ D 1 in. Stem, Titanium 2, 1 piece, 1.5 in. OD
M. Weights
□ W1 5lb weight, Standard for MPX-F 12 ft or shorter
□ W2 10lb weight, Standard for MPX-F longer than 12 ft
N. Temperature Sensor Options
MPX-F4
□ T__ Specify location of stem RTD in inches from bottom of
probe (6” is standard location)
□ 1T_ Digital Temperature Sensor A (no RTD) and location
from bottom of probe in inches
□ 2T_ Digital Temperature Sensors A (no RTD), B and loca-
tions from bottom of probe in inches
□ 3T_ Digital Temperature Sensors A (no RTD), B, C and loca-
tions from bottom of probe in inches
□ 4T_ Digital Temperature Sensors A (no RTD), B, C, D and
locations from bottom of probe in inches
□ 5T_ Digital Temperature Sensors A (no RTD), B, C, D, E and
locations from bottom of probe in inches
N. 4-20 Output Set Points
MPX-F2/F3
□ A__ 4mA set point location, in inches from bottom of probe
□ B__ 20mA set point location, in inches from top of probe
• Electrical Connections and System Wiring Diagrams
Modbus System Wiring For MPX-F4 Sensors
Power
Supply
Master
Device
+12-24 Vdc
GND
RS-485 A
RS-485 B
Use Shielded Cable
120 Ω
terminating
resistor
Note: Earth Ground terminal is factory
connected to internal housing ground
lug.
MPX-F4
series sensor
Sensor 1
V+
B
A
GND
EARTH
GND
MPX-F4
series sensor
Sensor 2
V+
B
A
GND
EARTH
GND
Modbus System Wiring with RST-6001 For MPX-F4 Sensors
Power
Supply
+12-24 Vdc
GND
Use Shielded Cable
MPX-F4
series sensor
Sensor 3
V+
B
A
GND
EARTH
GND
120 Ω
terminating
resistor at
last sensor
RST-6001
Modbus
Controller
USB to computer
with APG Modbus
software
Note: An independent +12-24 Vdc
power supply is required when
using an RST-6001 Modbus Controller.
The RST-6001 can only supply ±5 Vdc,
not the +12-24 Vdc required by the MPX.
Note: Earth Ground terminal is factory
connected to internal housing ground lug.
IMPORTANT: Refer to Chapter 5 for Hazardous Location and Non-Incendive Wiring
A
B
-5V
+5V
Equivalent 120 Ω
terminating resistor
internal to RST-6001
MPX-F4
series sensor
Sensor 1
V+
B
A
GND
EARTH
GND
MPX-F4
series sensor
Sensor 2
V+
B
A
GND
EARTH
GND
MPX-F4
series sensor
Sensor 3
V+
B
A
GND
EARTH
GND
diagram.
IMPORTANT: For lightning and surge protection on an MPX-F4, either connect the
grounding screw (see page 1) to an earth ground, or ensure that tank mounting of the
MPX-F4 is grounded.
properly on the stem, or sensor readings
will be inaccurate and unreliable. Untapered
UP
Figure 2.1
oats will have a sticker indicating the top of
the oat. Remove sticker prior to use.
IMPORTANT: MPX-F level sensor MUST be installed according to drawing 9003468
(Hazardous Installation and Non-Incendive Wiring Drawing) on page 30 to meet listed
approvals. Faulty installation will invalidate all safety approvals and ratings.
• Physical Installation Instructions
Ensure that all components have been received, including:
• MPX-F sensor head and stem
• Float (or oats, if two oats purchased)
• Optional Slide Mount
• Weight, Clevis Pin, and Cotter Pin
• Assembly drawing
Assemble sensor mounting, oat(s), weight and pins at installation location, if possible.
• Slide mounting option onto stem. Loosen compression cap so it will slide easily on stem.
• Note: If the oats do not t through the tank/vessel mounting hole, mount them on the stem from
inside the vessel being monitored. Then secure the sensor to the vessel.
• Slide oats onto stem. If using two oats, slide the lighter oat on rst. Tops of oats will be indicated
by sticker. After ensure top of oat is toward MPX-F sensor head, remove sticker.
• For sensors with oat stops, refer to the assembly drawing included with the sensor for oat stop
installation locations.
• Insert weight on end of stem. Hold in place with clevis pin.
• Insert cotter pin in clevis pin. Bend ends of cotter pin back to lock pins into place.
Install MPX-F sensor on tank
• When lifting and installing the sensor be sure to minimize the bending angle between the rigid stem at
the top and bottom of the sensor and the exible stem in-between. Sharp bends at those points could
damage the sensor.
• If your sensor’s stem and oats t through the mounting hole, insert the weight and the oats into the
mount opening.
• Carefully unroll and feed the MPX-F sensor stem into the tank. Slide the mount up to the top of the
stem.
• When the weight is on the bottom of the tank, secure the mounting option to the vessel.
• Take any slack out of the exible stem. Tighten the compression tting to hold stem in place.
• Install conduit tting(s) to 3/4” NPT conduit opening(s) on housing. Fittings must be UL/CSA Listed for
CSA installation.
• Feed system wires into MPX through conduit.
• Connect wires to MPX terminals. Use crimped ferruls on wires, if possible.
• Replace the housing cover.
• For lightning and surge protection on an MPX-F4, either connect the grounding screw (see page 1) to an
earth ground, or ensure that tank mounting of the MPX-F4 is grounded.
See Electrical Connections and System Wiring Diagrams (pages 5-6) for Modbus and 4-20 mA wiring
examples.
• Removal Instructions
Removing your MPX level sensor from service should be done with care.
• If the oats on your sensor t through the mounting hole, carefully lift the entire sensor assembly out
of and away from the vessel.
• If the oats on your sensor do not t through the mounting hole, they will need to be removed from
the stem before the sensor can be removed. Be sure to drain the vessel being monitored to allow access to the oats and stem for removal.
• Clean the stem and oats of any build up or debris and inspect for damage.
• Store your sensor in a dry place, at a temperature between -40° F and 180° F.
MPX-F4 series sensors use standard Modbus RTU protocol (RS-485). The sensors can only operate as slave
devices. Sensor default transmission settings are 9600 Baud, 8 Bits, 1 Stop Bit, No Parity, and require a
minimum delay of 300 ms between transactions. See MPX-F4 Modbus Register Lists on pages 11 and 12.
NOTE: For more information about Modbus RTU, please visit www.modbus.org.
• Modbus Programming with RST-6001 and APG Modbus Software
An APG RST-6001 Modbus Controller can be used in tandem with APG Modbus software to program and
control up to 20 MPX-F4 series sensors. Through APG Modbus, you can monitor the raw readings from the
sensor, congure the data for distance, level, volume, or weight, and enter measurements for a strapping
chart. See MPX-F4 Modbus Register Lists on pages 11 and 12.
NOTE: For APG Modbus programming instructions, or to download APG Modbus soft-
• 4-20 mA Programming with RST-4100 and APG Modbus Software
An APG RST-4100 Programming Module can be used in tandem with APG Modbus software to program
a single MPX-F2/3 series sensor. Through APG Modbus, you can congure the 4 mA and 20 mA output
setpoints and calibration settings. If your monitoring equipment (PLC, etc.) can be congured to interpret
the 4-20 mA output(s) of the MPX as volume, then the MPX can be congured accordingly via APG Modbus.
See MPX-F2 and F3 Modbus Register Lists on pages 17 and 18.
However, the RST-4100 is not designed to be used for continuous monitoring of a sensor. After
programming your MPX sensor, the RST-4100 must be removed and the wiring returned to normal. See 4-20
mA Loop Wiring and 4-20 mA Programming Wiring on page 6.
• Modbus Register Lists for MPX-F4
Input Registers (0x04)
Register Returned Data
30300 Raw Top Float Reading (in mm, unsigned)
30301 Raw Bottom Float Reading (in mm, unsigned)
30302 Temperature Reading (in 0C, signed)
30303-30304 Calculated Top Float Reading (in selected Units)
30305-30306 Calculated Bottom Float Reading (in selected Units)
30307 Version
NOTE: The Calculated Readings will be returned without a decimal place. In order to
obtain the true result, the Decimal Place setting must be taken into account.
40403 Volume Units 1, 2, 3, 4, 5, 6, 7
40404 Decimal Place 0, 1, 2, 3
40405 †Max Distance 0 to 11,278 mm
40406 Full Distance 0 to 10,364 mm
40407 Empty Distance 0 to 11,278 mm
40408 †Sensitivity 0 to 100
40409 †Pulses 0 to 20
40410 †Blanking 0 to 10,364 mm
40411 NA NA
40412 †Averaging 1 to 31
40413 †Filter Window 0 to 10,364 mm
40414 †Out of Range Samples 1 to 255
40415 †Sample Rate 10 to 1,000 msec.
40416 †Multiplier 1 to 1,999 (1000 = 1.000)
40417 †Oset -10,364 to 10,364 mm
40418 †Pre lter 0 to 10,364 mm
40419 †Noise limit 0 to 255
40420 Temperature Select 0 to 8*
40421 †RTD Oset (0C) -100 to 100
40422 †Float Window 0 to 1,000 mm 0=1 oat
40423 †1st Float Oset -10,364 to 10,364
40424 †2nd Float Oset -10,364 to 10,364
40425 †Gain Oset 0 to 255
40426 4 mA Set Point NA*
40427 20 mA Set Point NA*
40428 4 mA Calibration NA*
40429 20 mA Calibration NA*
40430 t1d NA*
40431 t1w NA*
40432 t1t NA*
40433 t2d NA*
40434 t2w NA*
40435 t2t NA*
40436-40437 Parameter 1 Data 0 to 1,000,000 mm
40438-40439 Parameter 2 Data 0 to 1,000,000 mm
40440-40441 Parameter 3 Data 0 to 1,000,000 mm
40442-40443 Parameter 4 Data 0 to 1,000,000 mm
40444-40445 Parameter 5 Data 0 to 1,000,000 mm
*These registers are not used by the MPX-F4, even though they are labeled in the APG Modbus software.
†Setting is factory calibrated. Do not adjust.
Determines the units of measure for the Calculated Reading when Application Type is set to 0, 1, or 7.
1 = Feet 2 = Inches 3 = Meters
40402 - Application Type
Determines the type of Calculated Reading performed by the sensor.
0 = Distance
1 = Level
2 = Standing Cylindrical Tank with or without Hemispherical Bottom
3 = Standing Cylindrical Tank with or without Conical Bottom
4 = Standing Rectangular Tank with or without Chute Bottom
5 = Horizontal Cylindrical Tank with or without Spherical Ends
Determines the number of decimal places included in the Calculated Reading(s). The Calculated Reading will
always be returned as a whole number.
For example, a Calculated Reading of 1126.658 (gallons, ft3, etc.) will be returned as follows:
Decimal Place = 0 Volume = 1127 (rounded to nearest whole number)
Decimal Place = 1 Volume = 11267 (divide by 10 to get true result)
Decimal Place = 2 Volume = 112666 (divide by 100 to get true result)
Decimal Place = 3 Volume = 1126658 (divide by 1000 to get true result)
Sets the distance (beginning from the Zero Reference) to the point where the sensor will stop looking for
oat signals, usually the bottom of the stem. A oat beyond the Maximum Distance value will not be detected.
40406 - Full Distance
Sets the positive distance (beginning from the sensor Zero Reference) to the point where the monitored
vessel is considered full.
40407 - Empty Distance
Sets the positive distance (beginning from the Zero Reference) to the point where the monitored vessel is
considered empty (usually the bottom of the stem).
40408 - Sensitivity (Factory Calibrated)
Sets the level of gain that is applied to the returning oat signal.
40409 - Pulses (Factory Calibrated)
Controls the duration of the signal being sent down the magnetostrictive wire.
40410 - Blanking (Factory Calibrated)
Sets the blanking distance, which is the zone from the Zero Reference of the sensor to the point from which
the rst signal will be valid. Signals from a oat in the blanking area will be ignored.
40412 - Averaging (Factory Calibrated)
Sets the number of qualied received oat signals to average for the raw reading. Qualied received signals
are placed in a rst-in, rst-out buer, the contents of which are averaged for the raw reading. The larger
the number of qualied received signals being averaged, the smoother the reading will be, and the slower
the reading will be to react to quickly changing targets.
Determines the physical range (0 - 10,364 mm) of qualied received signals, based on the current raw
reading. Signals beyond the +/- Filter Window range of the current reading will not qualify unless the
average moves. Signals outside the extents of the Filter Window are written to the Out of Range samples
buer (Holding Register 40414). See Figure 3.1.
Window = 300 mm
Out of Range Samples = 10
All samples
Samples are rejected within
this area unless they persist
for 10 consecutive samples
Min. ReadingMax. Reading
Current value
of Distance
are accepted
within this area
150 mm150mm
Figure 3.1
Samples are rejected within
this area unless they persist
for 10 consecutive samples
40414 - Out of Range Samples (Factory Calibrated)
Sets the number of consecutive samples outside the Filter Window (Holding Register 40413) necessary to
automatically adjust the current reading and move the Filter Window.
40415 - Sample Rate (Factory Calibrated)
Sets the update rate of the sensor (between 10 - 1000 ms). Shorter time delays allow for quicker sensor
response times to changing levels. Typical setting is 200 ms. Settings under 200 ms are not recommended.
40416 - Multiplier (Factory Calibrated)
Calibrates the distance reading span. The Multiplier is shown by the values 1 - 1999, but these values are
understood to represent 0.001 - 1.999. The default of 1000 (i.e. 1.000) is used for most applications.
Sets the Zero Reference of the sensor, the point from which the calculated distance is measured.
40418 - Pre lter
Denes the physical range (0 - 10,364 mm) of the start up (pre-lter) window. Four sample readings must be
found within the Pre lter window for the MPX sensor to successfully start up.
This register is used for factory diagnostics only.
40419 - Noise limit
Sets the limit for number of signals (0-255) outside the Pre lter range for the MPX at start up. If the Noise
Limit is reached before four readings register within the Pre lter window, the MPX will not start up.
This register is used for factory diagnostics only.
40420 - Temperature Select
Selects the temperature sensor reading to be displayed in Input Register 30302.
MPX-F4 sensors can accomodate up to ve digital temperature sensors in the stem. Options 0-5 and 8 will
work for MPX-F4 sensors.
0 = RTD 5 = Digital Temperature Sensor E
1 = Digital Temperature Sensor A 6 =
2 = Digital Temperature Sensor B 7 =
3 = Digital Temperature Sensor C 8 = Digital Temperature Sensor on Circuit Board
Sets the distance (0 - 1000 mm) between the rst (i.e. top) oat and the point at which the sensor will begin
looking for the second (bottom) oat. 0 indicates a single oat.
40423 - 1st Float Oset (Factory Calibrated)
Used to calibrate top oat reading (-10,364 - 10,364 mm).
40424 - 2nd Float Oset (Factory Calibrated)
Used to calibrate bottom oat reading (-10,364 - 10,364 mm).
40425 - Gain Oset (Factory Calibrated)
Used to move the centerline of the oat response signal to optimize signal strength (0 - 255).
• APG Modbus Register Lists for MPX-F2 and MPX-F3
Input Registers (0x04)
Register Returned Data
30300 Raw Top Float Reading (in mm, unsigned)
30301 Raw Bottom Float Reading (in mm, unsigned)
30302 Temperature Reading (in 0C, signed)
30303-30304 Calculated Top Float Reading (in selected Units)
30305-30306 Calculated Bottom Float Reading (in selected Units)
30307 Version
NOTE: Input Register values for MPX-F2 and MPX-F3 are only visible while program-
40403 Volume Units 1, 2, 3, 4, 5, 6, 7
40404 Decimal Place 0, 1, 2, 3*
40405 †Max Distance 0 to 10,364 mm
40406 Full Distance 0 to 10,364 mm
40407 Empty Distance 0 to 10,364 mm
40408 †Sensitivity 0 to 100
40409 †Pulses 0 to 20
40410 †Blanking 0 to 10,364 mm
40411 Fail Safe 0 = Disable, 1 = 3.8 mA, 2 = 22 mA
40412 †Averaging 1 to 31
40413 †Filter Window 0 to 10,364 mm
40414 †Out of Range Samples 1 to 255
40415 †Sample Rate 10 to 1,000 msec.
40416 †Multiplier 1 to 1,999 (1000 = 1.000)
40417 †Oset -10,364 to 10,364 mm
40418 †Pre lter 0 to 10,364 mm
40419 †Noise limit 0 to 255
40420 NA NA
40421 †RTD Oset (0C) -100 to 100
40422 †Float Window 0 to 1,000 mm 0=1 oat
40423 †1st Float Oset -10,364 to 10,364
40424 †2nd Float Oset -10,364 to 10,364
40425 †Gain Oset 0 to 255
40426 4 mA Set Point 0 - 10,364 mm
40427 20 mA Set Point 0 - 10,364 mm
40428 †4 mA Calibration 0 - 1,000
40429 †20 mA Calibration 0 - 1,000
40430 t1d NA*
40431 t1w NA*
40432 t1t NA*
40433 t2d NA*
40434 t2w NA*
40435 t2t NA*
40436-40437 Parameter 1 Data 0 to 1,000,000 mm
40438-40439 Parameter 2 Data 0 to 1,000,000 mm
40440-40441 Parameter 3 Data 0 to 1,000,000 mm
40442-40443 Parameter 4 Data 0 to 1,000,000 mm
40444-40445 Parameter 5 Data 0 to 1,000,000 mm
*These registers are not used by the MPX-F2 or MPX-F3, even though they are labeled in the APG Modbus
Determines the units of measure for the Calculated Reading when Application Type is set to 0, 1, or 7.
1 = Feet 2 = Inches 3 = Meters
For MPX-F2 and MPX-F3, this is seen only when using APG Modbus to program the MPX. This setting does
not affect the 4-20 mA output.
40402 - Application Type
Determines the type of Calculated Reading performed by the sensor.
0 = Distance
1 = Level
2 = Standing Cylindrical Tank with or without Hemispherical Bottom
3 = Standing Cylindrical Tank with or without Conical Bottom
4 = Standing Rectangular Tank with or without Chute Bottom
5 = Horizontal Cylindrical Tank with or without Spherical Ends
11 = Strapping Chart
See MPX-F Modbus Application Type Parameters pages 24-28.
For the MPX-F2 and MPX-F3, the 4-20 mA output can be scaled for linear output over distance/level or scaled
for linear output over volume. When setup in any of the volumetric application types, the 4-20mA output
becomes linear with regards to the volume (linear mA change per gallon, liter, etc.)
40403 - Volume Units
Determines the units of measure for the Calculated Reading when Application Type is set to 2 - 6 or 9 -11.
Determines the number of decimal places included in the Calculated Reading(s). For MPX-F2 and -F3, this is
seen only when using APG Modbus to program the MPX. This setting does not affect the 4-20 mA output.
Sets the distance (beginning from the Zero Reference) to the point where the sensor will stop looking for
oat signals, usually the bottom of the stem. A oat beyond the Maximum Distance value will not be detected.
40406 - Full Distance
Sets the positive distance (beginning from the sensor Zero Reference) to the point where the monitored
vessel is considered full.
40407 - Empty Distance
Sets the positive distance (beginning from the Zero Reference) to the point where the monitored vessel is
considered empty (usually the bottom of the stem).
40408 - Sensitivity (Factory Calibrated)
Sets the level of gain that is applied to the returning oat signal.
40409 - Pulses (Factory Calibrated)
Controls the duration of the signal being sent down the magnetostrictive wire.
40410 - Blanking (Factory Calibrated)
Sets the blanking distance, which is the zone from the Zero Reference of the sensor to the point from which
the rst signal will be valid. Signals from a oat in the blanking area will be ignored.
40411 - Fail Safe
Sets the output condition that the MPX-F will revert to in the event of a loss of return signal condition.
0 = Disable (no fail safe output)
1 = 3.8 mA
2 = 22 mA
Sets the number of qualied received oat signals to average for the raw reading. Qualied received signals
are placed in a rst-in, rst-out buer, the contents of which are averaged for the raw reading. The larger
the number of qualied received signals being averaged, the smoother the reading will be, and the slower
the reading will be to react to quickly changing targets.
40413 - Filter Window (Factory Calibrated)
Determines the physical range (0 - 10,364 mm) of qualied received signals, based on the current raw
reading. Signals beyond the +/- Filter Window range of the current reading will not qualify unless the
average moves. Signals outside the extents of the Filter Window are written to the Out of Range samples
buer (Holding Register 40414). See Figure 3.2.
Example:
Window = 300 mm
Out of Range Samples = 10
All samples
Samples are rejected within
this area unless they persist
for 10 consecutive samples
Min. ReadingMax. Reading
Current value
of Distance
are accepted
within this area
150 mm150mm
Samples are rejected within
this area unless they persist
for 10 consecutive samples
Figure 3.2
40414 - Out of Range Samples (Factory Calibrated)
Sets the number of consecutive samples outside the Filter Window (Holding Register 40413) necessary to
automatically adjust the current reading and move the Filter Window.
40415 - Sample Rate (Factory Calibrated)
Sets the update rate the sensor (10 - 1000 ms). Shorter time delays allow for quicker sensor response times
to changing levels. Typical setting is 200 ms. Settings under 200 ms are not recommended.
Calibrates the distance reading span. The Multiplier is shown by the values 1 - 1999, but these values are
understood to represent 0.001 - 1.999. The default of 1000 (i.e. 1.000) is used for most applications.
40417 - Oset (Factory Calibrated)
Sets the Zero Reference of the sensor, the point from which the calculated distance is measured.
40418 - Pre lter
Denes the physical range (0 - 10,364 mm) of the start up (pre-lter) window. Four sample readings must be
found within the Pre lter window for the MPX sensor to successfully start up.
This register is used for factory diagnostics only.
40419 - Noise limit
Sets the limit for number of signals (0-255) outside the Pre lter range for the MPX at start up. If the Noise
Limit is reached before four readings register within the Pre lter window, the MPX will not start up.
This register is used for factory diagnostics only.
40421 - RTD Oset C° (Factory Calibrated)
Calibrates the RTD termperature sensor.
40422 - Float Window (Factory Calibrated)
Sets the distance (0 - 1000 mm) between the rst (i.e. top) oat and the point at which the sensor will begin
looking for the second (bottom) oat. This essentially functions as a secondary blanking distance for the
minimum depth of the top uid. Set to 0 for single oat.
Used to calibrate top oat reading (-10,364 - 10,364 mm).
40424 - 2nd Float Oset (Factory Calibrated)
Used to calibrate bottom oat reading (-10,364 - 10,364 mm).
40425 - Gain Oset (Factory Calibrated)
Used to move the centerline of the oat response signal to optimize signal strength (0 - 255).
40426 - 4mA Set
Used to set the distance which will correspond to an output of 4 mA. For Application 1 (Distance), this is
measured from the Zero Reference. For all other applications (Level & Volumetric) this is measured from the
bottom of the probe.
40427 - 20mA Set
Used to set the distance which will correspond to an output of 20 mA. For Application 1 (Distance), this is
measured from the Zero Reference. For all other applications (Level & Volumetric) this is measured from the
bottom of the probe.
40428 - 4mA Cal (Factory Calibrated)
Used to calibrate the 4 mA output of the MPX-F2 or -F3.
40429 - 20mA Cal (Factory Calibrated)
Used to calibrate the 20 mA output of the MPX-F2 or -F3.
Your MPX level sensor is designed to be low maintenance. However, in general, you should:
• Periodically inspect your MPX to ensure the stem and oats are free of any heavy buildup that might
impede the movement of the oats.
• Ensure the housing cover is snuggly secured. If the cover becomes damaged or is misplaced, order a
replacement immediately.
• Repair and Returns
Should your MPX level sensor require service, please contact the factory via phone, email, or online chat. We
will issue you a Return Material Authorization (RMA) number with instructions.
• Phone: 888-525-7300
• Email: sales@apgsensors.com
• Online chat at www.apgsensors.com
Please have your part number and serial number available. See Warranty and Warranty Restrictions for
more information.
IMPORTANT: All repairs and adjustments of the MPX level sensor must be made by
the factory. Modifying, disassembling, or altering the MPX on site is strictly prohibited.
Certificate: 2397437 (237484) Master Contract: 237484
Project: 70022593 Date Issued: 2016-04-22
Issued to: Automation Products Group Inc
1025 West 1700 North
Logan, Utah 84321
USA
Attention: Karl Reid
The products listed below are eligible to bear the CSA Mark shown
with adjacent indicators 'C' and 'US' for Canada and US or with adjacent
indicator 'US' for US only or without either indicator for Canada only.
Ravindra
(Ravi) Kanthe
Ravindra (Ravi)
Kanthe
PRODUCTS
CLASS - C225802 - PROCESS CONTROL EQUIPMENT-For Hazardous LocationsCLASS - C225882 - PROCESS CONTROL EQUIPMENT-For Hazardous Locations - Certified to US
Standards
Class I, Division 1 & 2, Groups C and D
Ex d IIB
Ex nA IIB
Class I, Zone 1; AEx d IIB
Class I, Zone 2; AEx nA IIB
xFloat Level Sensors, Model MPX- E, R, G &T (MPX- abc-de-fghi-jjj), rated 12 - 24 Vdc, 80mA, or
rated 12 to 24 Vdc, 4-20mA; operating ambient Ta is 85°C; Temperature Code T4; Ingress protection
IP65; Field wiring is non-incendive when installed per drawing 9003468.
Class I, Division 2, Groups C and D
Ex nA IIB
Class I, Zone 2; AEx nA IIB
General Requirements – Canadian Electrical Code, Part II
– Tenth Edition
CSA C22.2 No 30-M1986
(Reaffirmed 2007)
Explosion-Proof Enclosures for Use in Class I Hazardous
Locations Industrial Products – Third Edition
CSA C22.2 No 142-M1987
(Reaffirmed 2014)
Process Control Equipment Industrial Products – Third
Edition
Non-incendive Electrical Equipment for Use in Class I,
Edition
Electrical apparatus for explosive gas atmospheres – Part
0: General requirements – First Edit ion
Electrical apparatus for explosive gas atmospheres – Part
1: Flameproof enclosures "d" – First Edition
CSA E60079-15-02
(Reaffirmed 2006)
Electrical Apparatus for Explosive Gas Atmospheres –
Part 15: Type of Protection "n" – Second Edition
Industrial Control Equipment - Seventeenth Edition;
2010
Explosion-Proof and Dust-Ignition-Proof Electrical
Including October 28, 2009
Nonincendive Electrical Equipment for Use in Class I and
(Classified) Locations
Explosive atmospheres - Part 0 Equipment - General
requirements
70022593
Master Contract: 237484
Date Issued: 2016-04-22
xFloat Level Sensors, Model MPX- F (model MPX- abc-de-fghi-jjj), rated 12 - 24 Vdc, 80mA, or rated
12 to 24 Vdc, 4-20mA; operating ambient Ta is 85°C; Temperature Code T4; Ingress protection IP65;
Field wiring is non-incendive when installed per drawing 9003468
Notes for all equipment:
1. The model code breakdown is as follows: a= E, R, F, G or T; b= 1, 2, 3 or 4; c= A,B or C, d= A, B, C, D, E,
F, G, Z, X, V, T, R, M, or J; e= N, B, D, Y, W, U, S, P, L, K, or I; f= F, R, P, S, N, or O; g= 1, 1.5, 2, 2.5, 3,
3.5, 4, 5, 6; h=W or S; i= A, B, C or D; and j= 12–153 for the 1/2" stem Type E or 48–300 for the 1" stem
Type R, Type G and Type T or 120-456 for Flex stem Type F.
2. The equipment is intended to be installed as required by the applicable electrical code (CEC, NEC) and as
specified by the manufacturers Installation Instructions.
3. The installation will be inspected by the authority with jurisdiction in the area where installed.
Update to Report 2397437 to include 3 new Probe types MPX
(for Class 1, Div. 1 & 2 and Zones 1 & 2) and MPX
and Zone 2), new PCB design, and update product label material and printing
method.
2440956
2011
Update to report 2397437 to correct misspellings and add note in Factory
Tests section of report.
2397437
2011-05-18
Original certification for the model MPX-ab-cd-efg-hhh float level sensors.