APG MPX-F Operating Manual

APG
MPX Magnetostrictive Level Sensors
User Manual
For The MPX-F
R
Doc #9004829
Rev B, 03/2017
Table of Contents
Introduction ................................................................................................................ iii
Warranty and Warranty Restrictions .................................................................... iv
Chapter 1: Specications and Options.....................................................................1
Dimensions ....................................................................................................................................1-2
Specications ...................................................................................................................................3
Model Number Congurator .......................................................................................................... 4
Electrical Connections and System Wiring Diagrams ........................................................ 5-6
Chapter 2: Installation and Removal Procedures and Notes ..............................7
Tools Needed .....................................................................................................................................7
Physical Installation Notes ...........................................................................................................7
Physical Installation Instructions ............................................................................................... 8
Electrical Installation .....................................................................................................................9
Removal Instructions .....................................................................................................................9
Chapter 3: Programming ..........................................................................................10
Modbus Programming .................................................................................................................. 10
Modbus Programming with RST-6001 and APG Modbus Software .................................... 10
4-20 mA Programming with RST-4100 and APG Modbus Software ..................................... 11
Modbus Register Lists for MPX-F4 ....................................................................................... 11-12
MPX-F4 Modbus Sensor Parameters .................................................................................... 13-17
APG Modbus Register Lists for MPX-F2 and MPX-F3 ....................................................... 17-18
MPX-F2 and MPX-F3 APG Modbus Sensor Parameters ...................................................19-23
MPX-F Modbus Application Type Parameters .................................................................. 24-28
Chapter 4: Maintenance ...........................................................................................29
General Care .................................................................................................................................... 29
Repair and Returns ........................................................................................................................29
Chapter 5: Hazardous Location Installation and Certication .........................30
Hazardous Location and Non-Incendive Wiring Diagram .................................................... 30
CSA Certicate of Compliance ...............................................................................................31-34
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Introduction
237484
Thank you for purchasing an MPX series magnetostrictive level sensor from APG. We appreciate your business and your trust. Please take a few minutes to familiarize yourself with your MPX and this manual.
The MPX level sensors provide highly accurate and repeatable level readings in a wide variety of liquid level
measurement applications. The MPX-F is certied for installation in hazardous areas in the US and Canada
by CSA for Class I, Division 2 and Class I, Zone 2 environments. The MPX-F’s exible stainless steel stem
allows for accurate measurements in environments that are not straight-forward.
Reading your label
Every APG instrument comes with a label that includes the instrument’s model number, part number, and serial number. Please ensure that the part number on your label matches your order. The following
electrical ratings and approvals are also listed on the label. Please refer to the CSA Certicate of Compliance
at the back of this manual for further details.
Electrical ratings
Rated 12 - 24 VDC, 4-20 mA, or 80 mA
Class I, Division 2, Groups C, and D Ex nA IIB Class I, Zone 2; AEx nA IIB
C
Non-Incendive Wiring Requirements: Vmax Ui= 28VDC, Imax Ii = 200ma, Ci = 0nF, Li = 0μH
US
IMPORTANT: MPX-F level sensor MUST be installed according to drawing 9003468
(Hazardous Installation and Non-Incendive Wiring Drawing) on page 30 to meet listed approvals. Faulty installation will invalidate all safety approvals and ratings.
DANGER: OPEN CIRCUIT BEFORE REMOVING COVER or KEEP COVER TIGHT WHILE CIRCUITS ARE ALIVE;
AVERTISSEMENT -- OUVRIR LE CIRCUIT AVANT D’ENLEVER LE COUVERCLE, or GARDER LE COUVERCLE BIEN FERME TANT QUE LES CIRCUITS SONT SOUS TENSION.
DANGER: WARNING -- EXPLOSION HAZARD -- SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY
FOR CLASS I, DIVISION 2; AVERTISSEMENT -- RISQUE D’EXPLOSION -- LA SUBSTITIOND E COMPOSANTSP EUTR ENDRE CE MATERIEL INACCEPTABLE POUR LES EMPLACEMENTS DE CLASSE I, DIVISION 2.
DANGER: WARNING -- EXPLOSION HAZARD -- DO NOT DISCONNECT EQUIPMENT UNLESS POWER HAS
BEEN SWITCHED OFF OR THE AREA IS KNOWN TO BE NON-HAZARDOUS; AVERTISSEMENT -- RISQUE D’EXPLOSION -- AVANT DE DECONNECTER L’EQUIPEMENT, COUPER LE COURANT OU S’ASSURER QUE L’EMPLACEMENT EST DESIGNE NON DANGEREUX.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
iii
Warranty and Warranty Restrictions
This product is covered by APG’s waranty to be free from defects in material and workmanship under
normal use and service of the product for 24 months. For a full explanation of our Warranty, please visit
https://www.apgsensors.com/about-us/terms-conditions. Contact Technical Support to recieve a Return
Material Authorization before shipping your product back.
Scan the QR code below to read the full explanation of our Warranty on your tablet or smartphone.
iv
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Chapter 1: Specications and Options
• Dimensions
MPX-F Sensor Dimensions
OPENINGS
COND UIT
HOUSING
HOUSING
ADAPTER
[110mm]
4.34"
[179mm]
7.03"
THRU
TOP DEADBAND
[254mm]
10.0"
[126mm]
2x 3/4" NPT
2x ø2.76"
GROUND
SCREW
MOUNTING OPTION FLOAT( S)
4.98"
[110mm]
4.34"
MEASURABLE LENGTH (PER ORDER)
TOTAL LENGTH
MPX-F STEM
ASSEMBLY
FLOAT
STOP
BOT DEADBAND
[152mm]
6.0"
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
1
MPX-F Floats
Float Options
5.90”
5.00”
U & V
3.03”
W & X
5.10”
Float Ref. S1=10.63” S2=6”
Float Ref. S1=12.25” S2=4.25”
S & T
2.80”
Float Ref. S1=12.25” S2=4.25”
3.07”
Y & Z / P & R / L & M
Float Ref.
5.50”
3.08”
2.80”
2.00”
S1=10” S2=6.5”
Y & Z P & R L & M
2
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Specications
Performance
Resolution 4-20 mA: 14 bit DAC Modbus: 0.04 in. (1 mm) Accuracy ±0.05% of Full Scale or 1mm (whichever is larger)
Environmental
Operating Temperature -40° to 185° F (-40° to 85° C) Enclosure Protection NEMA 4X, IP65
Electrical
Supply Voltage 12-24 VDC on sensor Current Draw Modbus (RS-485): 28 mA
4-20 mA: 22 mA single / 44 mA dual (Max)
Materials of Construction
Housing Cast aluminum, epoxy coated Stem 7/8” ø 316L SS Flexible Tubing with Braid
Mounting (slide) 316L SS Compression Fitting (slide) Aluminum with Neoprene bushing
Connectivity
Output Modbus RTU (RS-485) 2 wire, loop-powered 4-20 mA 3 wire, loop-powered dual 4-20 mA
Programming
RS-485 Optional RST-6001 USB-to-RS-485 converter
4-20 mA Optional RST-4100 programming module
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
3
Model Number Congurator
Model Number: MPX - _F_ _____ _____ - _____ _____ - _____ _____ _____ _B_ - _____ - _____ _____ - ______ A B C D E F G H I J L M N
A. Stem Type
F 7/8 in. diameter Flexible Tubing with Braid, 316L SS
B. Output
2 Single oat, 4-20 mA (loop powered, 2 wire)
□ □ 3 Dual oat, 4-20 mA (loop powered, 3 wire) 4 Modbus RTU, surge/lightning protection, stem RTD
temperature sensor
Note: stem RTDs default to 6” from bottom of probe
C. Housing Type
All Housing Die-cast Aluminum, NEMA 4X, IP68, Blue
__
Large Housing
B Large Housing with window
D. Float 1 (Top Float)
Z 5.5h x 3d in. Red Polyurethane (0.65 SG) Y 5.5h x 3d in. Blue Polyurethane (0.94 SG) X 5 in. Round 316L SS (0.52 SG) W 5 in. Round 316L SS (0.92 SG) V 6h x 3d in. Oval 316L SS (0.58 SG) U 6h x 3d in. Oval 316L SS (0.94 SG) T 3 in. Round 316L SS (0.60 SG) S 3 in. Round 316L SS (0.94 SG) R 5.5h x 2.8d in. Red Polyurethane (0.59 SG) P 5.5h x 2.8d in. Blue Polyurethane (0.94 SG) M 5.5h x 2d in. Red Polyurethane (0.57 SG) L 5.5h x 2d in. Blue Polyurethane (0.94 SG) N None O Other
E. Float 2 (optional)
N None Y 5.5h x 3d in. Blue Polyurethane (0.94 SG) W 5 in. Round 316L SS (0.92 SG) U 6h x 3d in. Oval 316L SS (0.94 SG) P 5.5h x 2.8d in. Blue Polyurethane (0.94 SG) L 5.5h x 2d in. Blue Polyurethane (0.94 SG) O Other
F. Mounting Type
F Flat Face ANSI Flange 150#
(Sizes: 2, 2.5, 3, 4, 5, 6)
S Tri Clamp (Sizes: 2, 2.5, 3) P NPT Plug 150# (Sizes: 2, 2.5, 3, 4) N None O Other
G. Mounting Size
__ See Mounting Type for available sizes
H. Mounting Connection
W Welded (xed) S Slide with Compression Fitting (adjustable)
I. Stem Material
B 316L SS
J. Total Stem Length in Inches
__ Min. 120 in. - Max. 384 in.
L. Float Stop Options
__
1 in. Stem, 316L SS, 1 piece, 1.5 in. OD
C 1 in. Stem, 316L SS, 2 piece, 1.75 in. OD D 1 in. Stem, Titanium 2, 1 piece, 1.5 in. OD
M. Weights
W1 5lb weight, Standard for MPX-F 12 ft or shorter W2 10lb weight, Standard for MPX-F longer than 12 ft
N. Temperature Sensor Options
MPX-F4
T__ Specify location of stem RTD in inches from bottom of
probe (6” is standard location)
1T_ Digital Temperature Sensor A (no RTD) and location
from bottom of probe in inches
2T_ Digital Temperature Sensors A (no RTD), B and loca-
tions from bottom of probe in inches
3T_ Digital Temperature Sensors A (no RTD), B, C and loca-
tions from bottom of probe in inches
4T_ Digital Temperature Sensors A (no RTD), B, C, D and
locations from bottom of probe in inches
5T_ Digital Temperature Sensors A (no RTD), B, C, D, E and
locations from bottom of probe in inches
N. 4-20 Output Set Points
MPX-F2/F3
A__ 4mA set point location, in inches from bottom of probe B__ 20mA set point location, in inches from top of probe
This option is standard
4
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Electrical Connections and System Wiring Diagrams
Modbus System Wiring For MPX-F4 Sensors
Power Supply
Master Device
+12-24 Vdc
GND
RS-485 A
RS-485 B
Use Shielded Cable
120 Ω
terminating resistor
Note: Earth Ground terminal is factory connected to internal housing ground lug.
MPX-F4 series sensor
Sensor 1
V+
B A
GND
EARTH
GND
MPX-F4 series sensor
Sensor 2
V+
B A
GND
EARTH
GND
Modbus System Wiring with RST-6001 For MPX-F4 Sensors
Power Supply
+12-24 Vdc
GND
Use Shielded Cable
MPX-F4 series sensor
Sensor 3
V+
B A
GND
EARTH
GND
120 Ω
terminating resistor at last sensor
RST-6001
Modbus
Controller
USB to computer with APG Modbus software
Note: An independent +12-24 Vdc power supply is required when using an RST-6001 Modbus Controller. The RST-6001 can only supply ±5 Vdc, not the +12-24 Vdc required by the MPX.
Note: Earth Ground terminal is factory connected to internal housing ground lug.
IMPORTANT: Refer to Chapter 5 for Hazardous Location and Non-Incendive Wiring
A B
-5V
+5V
Equivalent 120 Ω
terminating resistor internal to RST-6001
MPX-F4 series sensor
Sensor 1
V+
B A
GND
EARTH
GND
MPX-F4 series sensor
Sensor 2
V+
B A
GND
EARTH
GND
MPX-F4 series sensor
Sensor 3
V+
B A
GND
EARTH
GND
diagram.
IMPORTANT: For lightning and surge protection on an MPX-F4, either connect the
grounding screw (see page 1) to an earth ground, or ensure that tank mounting of the MPX-F4 is grounded.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
120 Ω
terminating resistor at last sensor
5
4-20 mA Loop Wiring
4-20 mA Dual Float Loop Wiring
4-20 mA Single Float Loop Wiring
(MPX-F2 Series)
+
Sensor
12-24 Vdc
Out
Power Source
(12-24 Vdc)
Receiver
Input Com
IMPORTANT: Refer to Chapter 5
for Hazardous Location and Non-In­cendive Wiring diagram.
4-20 mA Programming Wiring
(MPX-F3 Series)
Sensor
12-24 Vdc
Out 1 Out 2
+
Power Source
(12-24 Vdc)
Receiver
Input 1 Input 2 Com
OR
+
Sensor
12-24 Vdc
Out 1 Out 2
Power Source
Receiver 1
Input Com
(12-24 Vdc)
Receiver 2
Input Com
Programming configuration is
for programming ONLY. After programming, sensor must be reintegrated to 4-20 mA loop for proper system operation.
RST-4100
MPX-F2 or MPX-F3 series sensor.
Sensor
12-24 Vdc
Out (1)
Out 2
Out 24V
Load Resistor
(150 Ω to 700 Ω)
Wired in series for
programming.
USB to computer with APG Modbus software
Programming Module and APG Modbus software required for programming MPX-F2 and MPX-F3 series sensors.
Note: a minimum of 14 Vdc is required to establish communications with the RST-4100
+
Power
Source
(14- 24 Vdc)
NOTE: For MPX-F3 series sensors,
- Vdc from power source must be connected to Output1 on sensor for correct sensor programming.
6
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Chapter 2: Installation and Removal Procedures and Notes
• Tools Needed
You will need the following tools to install your MPX level sensor:
Wrench sized appropriately for MPX mounting
Wrench sized appropriately for conduit connections
Flat-head screwdriver for wire terminals
Channel lock pliers for tightening compression tting
1/8” Hex Allen wrench for screws on oat stop(s).
Pliers for bending cotter pin.
• Physical Installation Notes
The MPX-T should be installed in an area--indoors or outdoors--which meets the following conditions:
Ambient temperature between -40°C and 85°C (-40°F to +185°F)
Relative humidity up to 100%
Altitude up to 2000 meters (6560 feet)
IEC-664-1 Conductive Pollution Degree 1 or 2
IEC 61010-1 Measurement Category II
No chemicals corrosive to stainless steel (such as NH3, SO2, Cl2)
Ample space for maintenance and inspection
Additional care must be taken to ensure:
The probe is located away from strong magnetic elds, such as those produced by motors, transform- ers, solenoid valves, etc.
The medium is free from metallic substances and other foreign matter.
The probe is not exposed to excessive vibration.
The oat(s) t through the mounting hole. If the oat(s) does/do not t, it/they must be mounted on
the stem from inside the vessel being monitored.
The oat(s) is/are oriented properly on the stem (See Figure 2.1). MPX-F oats are installed by custom- er.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
7
Taper
IMPORTANT: Floats must be oriented
properly on the stem, or sensor readings will be inaccurate and unreliable. Untapered
UP
Figure 2.1
oats will have a sticker indicating the top of the oat. Remove sticker prior to use.
IMPORTANT: MPX-F level sensor MUST be installed according to drawing 9003468
(Hazardous Installation and Non-Incendive Wiring Drawing) on page 30 to meet listed approvals. Faulty installation will invalidate all safety approvals and ratings.
• Physical Installation Instructions
Ensure that all components have been received, including:
MPX-F sensor head and stem
Float (or oats, if two oats purchased)
Optional Slide Mount
Weight, Clevis Pin, and Cotter Pin
Assembly drawing
Assemble sensor mounting, oat(s), weight and pins at installation location, if possible.
Slide mounting option onto stem. Loosen compression cap so it will slide easily on stem.
Note: If the oats do not t through the tank/vessel mounting hole, mount them on the stem from
inside the vessel being monitored. Then secure the sensor to the vessel.
Slide oats onto stem. If using two oats, slide the lighter oat on rst. Tops of oats will be indicated by sticker. After ensure top of oat is toward MPX-F sensor head, remove sticker.
For sensors with oat stops, refer to the assembly drawing included with the sensor for oat stop
installation locations.
Insert weight on end of stem. Hold in place with clevis pin.
Insert cotter pin in clevis pin. Bend ends of cotter pin back to lock pins into place.
Install MPX-F sensor on tank
When lifting and installing the sensor be sure to minimize the bending angle between the rigid stem at
the top and bottom of the sensor and the exible stem in-between. Sharp bends at those points could
damage the sensor.
If your sensor’s stem and oats t through the mounting hole, insert the weight and the oats into the
mount opening.
Carefully unroll and feed the MPX-F sensor stem into the tank. Slide the mount up to the top of the stem.
When the weight is on the bottom of the tank, secure the mounting option to the vessel.
Take any slack out of the exible stem. Tighten the compression tting to hold stem in place.
8
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
• Electrical Installation
• Remove the housing cover of your MPX.
Install conduit tting(s) to 3/4” NPT conduit opening(s) on housing. Fittings must be UL/CSA Listed for CSA installation.
• Feed system wires into MPX through conduit.
• Connect wires to MPX terminals. Use crimped ferruls on wires, if possible.
• Replace the housing cover.
• For lightning and surge protection on an MPX-F4, either connect the grounding screw (see page 1) to an earth ground, or ensure that tank mounting of the MPX-F4 is grounded.
See Electrical Connections and System Wiring Diagrams (pages 5-6) for Modbus and 4-20 mA wiring
examples.
• Removal Instructions
Removing your MPX level sensor from service should be done with care.
If the oats on your sensor t through the mounting hole, carefully lift the entire sensor assembly out
of and away from the vessel.
If the oats on your sensor do not t through the mounting hole, they will need to be removed from the stem before the sensor can be removed. Be sure to drain the vessel being monitored to allow ac­cess to the oats and stem for removal.
Clean the stem and oats of any build up or debris and inspect for damage.
Store your sensor in a dry place, at a temperature between -40° F and 180° F.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
9
Chapter 3: Programming
Modbus Programming
MPX-F4 series sensors use standard Modbus RTU protocol (RS-485). The sensors can only operate as slave devices. Sensor default transmission settings are 9600 Baud, 8 Bits, 1 Stop Bit, No Parity, and require a minimum delay of 300 ms between transactions. See MPX-F4 Modbus Register Lists on pages 11 and 12.
NOTE: For more information about Modbus RTU, please visit www.modbus.org.
Modbus Programming with RST-6001 and APG Modbus Software
An APG RST-6001 Modbus Controller can be used in tandem with APG Modbus software to program and control up to 20 MPX-F4 series sensors. Through APG Modbus, you can monitor the raw readings from the
sensor, congure the data for distance, level, volume, or weight, and enter measurements for a strapping
chart. See MPX-F4 Modbus Register Lists on pages 11 and 12.
NOTE: For APG Modbus programming instructions, or to download APG Modbus soft-
ware, please visit www.apgsensors.com/suppport.
10
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
4-20 mA Programming with RST-4100 and APG Modbus Software
An APG RST-4100 Programming Module can be used in tandem with APG Modbus software to program
a single MPX-F2/3 series sensor. Through APG Modbus, you can congure the 4 mA and 20 mA output setpoints and calibration settings. If your monitoring equipment (PLC, etc.) can be congured to interpret the 4-20 mA output(s) of the MPX as volume, then the MPX can be congured accordingly via APG Modbus.
See MPX-F2 and F3 Modbus Register Lists on pages 17 and 18.
However, the RST-4100 is not designed to be used for continuous monitoring of a sensor. After
programming your MPX sensor, the RST-4100 must be removed and the wiring returned to normal. See 4-20 mA Loop Wiring and 4-20 mA Programming Wiring on page 6.
Modbus Register Lists for MPX-F4
Input Registers (0x04)
Register Returned Data
30300 Raw Top Float Reading (in mm, unsigned)
30301 Raw Bottom Float Reading (in mm, unsigned)
30302 Temperature Reading (in 0C, signed) 30303-30304 Calculated Top Float Reading (in selected Units)
30305-30306 Calculated Bottom Float Reading (in selected Units)
30307 Version
NOTE: The Calculated Readings will be returned without a decimal place. In order to
obtain the true result, the Decimal Place setting must be taken into account.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
11
Holding Registers (0x03)
Register Function Value Range
40400 Device Address 1 to 247 40401 Units 1, 2, 3
40402 Application Type 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11
40403 Volume Units 1, 2, 3, 4, 5, 6, 7 40404 Decimal Place 0, 1, 2, 3 40405 †Max Distance 0 to 11,278 mm 40406 Full Distance 0 to 10,364 mm 40407 Empty Distance 0 to 11,278 mm 40408 Sensitivity 0 to 100 40409 †Pulses 0 to 20 40410 †Blanking 0 to 10,364 mm 40411 NA NA 40412 Averaging 1 to 31 40413 †Filter Window 0 to 10,364 mm 40414 †Out of Range Samples 1 to 255 40415 †Sample Rate 10 to 1,000 msec. 40416 †Multiplier 1 to 1,999 (1000 = 1.000) 40417 †Oset -10,364 to 10,364 mm 40418 †Pre lter 0 to 10,364 mm 40419 †Noise limit 0 to 255 40420 Temperature Select 0 to 8* 40421 †RTD Oset (0C) -100 to 100 40422 †Float Window 0 to 1,000 mm 0=1 oat 40423 †1st Float Oset -10,364 to 10,364 40424 †2nd Float Oset -10,364 to 10,364 40425 †Gain Oset 0 to 255 40426 4 mA Set Point NA* 40427 20 mA Set Point NA* 40428 4 mA Calibration NA*
40429 20 mA Calibration NA*
40430 t1d NA* 40431 t1w NA* 40432 t1t NA* 40433 t2d NA* 40434 t2w NA* 40435 t2t NA* 40436-40437 Parameter 1 Data 0 to 1,000,000 mm
40438-40439 Parameter 2 Data 0 to 1,000,000 mm
40440-40441 Parameter 3 Data 0 to 1,000,000 mm 40442-40443 Parameter 4 Data 0 to 1,000,000 mm 40444-40445 Parameter 5 Data 0 to 1,000,000 mm
*These registers are not used by the MPX-F4, even though they are labeled in the APG Modbus software.Setting is factory calibrated. Do not adjust.
12
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
MPX-F4 Modbus Sensor Parameters
40401 - Units
Determines the units of measure for the Calculated Reading when Application Type is set to 0, 1, or 7. 1 = Feet 2 = Inches 3 = Meters
40402 - Application Type
Determines the type of Calculated Reading performed by the sensor. 0 = Distance 1 = Level
2 = Standing Cylindrical Tank with or without Hemispherical Bottom 3 = Standing Cylindrical Tank with or without Conical Bottom 4 = Standing Rectangular Tank with or without Chute Bottom 5 = Horizontal Cylindrical Tank with or without Spherical Ends
6 = Spherical Tank 7 = Pounds (Linear Scaling) 8 = N/A
9 = Vertical Oval Tank 10 = Horizontal Oval Tank
11 = Strapping Chart
See MPX-F Modbus Application Type Parameters pages 25-29.
40403 - Volume Units
Determines the units of measure for the Calculated Reading when Application Type is set to 2 - 6 or 9 -11.
1 = Feet3 5 = Liters 2 = Million Feet3 6 = Inches3
3 = Gallons 7 = Barrels
4 = Meters3
40404 - Decimal Place
Determines the number of decimal places included in the Calculated Reading(s). The Calculated Reading will always be returned as a whole number.
For example, a Calculated Reading of 1126.658 (gallons, ft3, etc.) will be returned as follows:
Decimal Place = 0 Volume = 1127 (rounded to nearest whole number) Decimal Place = 1 Volume = 11267 (divide by 10 to get true result) Decimal Place = 2 Volume = 112666 (divide by 100 to get true result) Decimal Place = 3 Volume = 1126658 (divide by 1000 to get true result)
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
13
40405 - Maximum Distance (Factory Calibrated)
Sets the distance (beginning from the Zero Reference) to the point where the sensor will stop looking for oat signals, usually the bottom of the stem. A oat beyond the Maximum Distance value will not be detect­ed.
40406 - Full Distance
Sets the positive distance (beginning from the sensor Zero Reference) to the point where the monitored vessel is considered full.
40407 - Empty Distance
Sets the positive distance (beginning from the Zero Reference) to the point where the monitored vessel is considered empty (usually the bottom of the stem).
40408 - Sensitivity (Factory Calibrated)
Sets the level of gain that is applied to the returning oat signal.
40409 - Pulses (Factory Calibrated)
Controls the duration of the signal being sent down the magnetostrictive wire.
40410 - Blanking (Factory Calibrated)
Sets the blanking distance, which is the zone from the Zero Reference of the sensor to the point from which
the rst signal will be valid. Signals from a oat in the blanking area will be ignored.
40412 - Averaging (Factory Calibrated)
Sets the number of qualied received oat signals to average for the raw reading. Qualied received signals are placed in a rst-in, rst-out buer, the contents of which are averaged for the raw reading. The larger the number of qualied received signals being averaged, the smoother the reading will be, and the slower
the reading will be to react to quickly changing targets.
14
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
40413 - Filter Window (Factory Calibrated)
Example:
Determines the physical range (0 - 10,364 mm) of qualied received signals, based on the current raw
reading. Signals beyond the +/- Filter Window range of the current reading will not qualify unless the
average moves. Signals outside the extents of the Filter Window are written to the Out of Range samples buer (Holding Register 40414). See Figure 3.1.
Window = 300 mm Out of Range Samples = 10
All samples
Samples are rejected within this area unless they persist
for 10 consecutive samples
Min. Reading Max. Reading
Current value
of Distance
are accepted
within this area
150 mm 150mm
Figure 3.1
Samples are rejected within this area unless they persist
for 10 consecutive samples
40414 - Out of Range Samples (Factory Calibrated)
Sets the number of consecutive samples outside the Filter Window (Holding Register 40413) necessary to
automatically adjust the current reading and move the Filter Window.
40415 - Sample Rate (Factory Calibrated)
Sets the update rate of the sensor (between 10 - 1000 ms). Shorter time delays allow for quicker sensor response times to changing levels. Typical setting is 200 ms. Settings under 200 ms are not recommended.
40416 - Multiplier (Factory Calibrated)
Calibrates the distance reading span. The Multiplier is shown by the values 1 - 1999, but these values are understood to represent 0.001 - 1.999. The default of 1000 (i.e. 1.000) is used for most applications.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
15
40417 - Oset (Factory Calibrated)
Sets the Zero Reference of the sensor, the point from which the calculated distance is measured.
40418 - Pre lter
Denes the physical range (0 - 10,364 mm) of the start up (pre-lter) window. Four sample readings must be found within the Pre lter window for the MPX sensor to successfully start up.
This register is used for factory diagnostics only.
40419 - Noise limit
Sets the limit for number of signals (0-255) outside the Pre lter range for the MPX at start up. If the Noise Limit is reached before four readings register within the Pre lter window, the MPX will not start up.
This register is used for factory diagnostics only.
40420 - Temperature Select
Selects the temperature sensor reading to be displayed in Input Register 30302.
MPX-F4 sensors can accomodate up to ve digital temperature sensors in the stem. Options 0-5 and 8 will
work for MPX-F4 sensors.
0 = RTD 5 = Digital Temperature Sensor E 1 = Digital Temperature Sensor A 6 =
2 = Digital Temperature Sensor B 7 = 3 = Digital Temperature Sensor C 8 = Digital Temperature Sensor on Circuit Board
4 = Digital Temperature Sensor D
40421 - RTD Oset C° (Factoy Calibrated)
Calibrates the RTD termperature sensor.
16
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
40422 - Float Window (Factory Calibrated)
Sets the distance (0 - 1000 mm) between the rst (i.e. top) oat and the point at which the sensor will begin looking for the second (bottom) oat. 0 indicates a single oat.
40423 - 1st Float Oset (Factory Calibrated)
Used to calibrate top oat reading (-10,364 - 10,364 mm).
40424 - 2nd Float Oset (Factory Calibrated)
Used to calibrate bottom oat reading (-10,364 - 10,364 mm).
40425 - Gain Oset (Factory Calibrated)
Used to move the centerline of the oat response signal to optimize signal strength (0 - 255).
APG Modbus Register Lists for MPX-F2 and MPX-F3
Input Registers (0x04)
Register Returned Data
30300 Raw Top Float Reading (in mm, unsigned)
30301 Raw Bottom Float Reading (in mm, unsigned)
30302 Temperature Reading (in 0C, signed) 30303-30304 Calculated Top Float Reading (in selected Units)
30305-30306 Calculated Bottom Float Reading (in selected Units)
30307 Version
NOTE: Input Register values for MPX-F2 and MPX-F3 are only visible while program-
ming via the RST-4100.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
17
Holding Registers (0x03)
Register Function Value Range
40400 Device Address 1 to 247* 40401 Units 1, 2, 3
40402 Application Type 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11
40403 Volume Units 1, 2, 3, 4, 5, 6, 7 40404 Decimal Place 0, 1, 2, 3* 40405 †Max Distance 0 to 10,364 mm 40406 Full Distance 0 to 10,364 mm 40407 Empty Distance 0 to 10,364 mm 40408 Sensitivity 0 to 100 40409 †Pulses 0 to 20 40410 †Blanking 0 to 10,364 mm 40411 Fail Safe 0 = Disable, 1 = 3.8 mA, 2 = 22 mA 40412 Averaging 1 to 31 40413 †Filter Window 0 to 10,364 mm 40414 †Out of Range Samples 1 to 255 40415 †Sample Rate 10 to 1,000 msec. 40416 †Multiplier 1 to 1,999 (1000 = 1.000) 40417 †Oset -10,364 to 10,364 mm 40418 †Pre lter 0 to 10,364 mm 40419 †Noise limit 0 to 255 40420 NA NA 40421 †RTD Oset (0C) -100 to 100 40422 †Float Window 0 to 1,000 mm 0=1 oat 40423 †1st Float Oset -10,364 to 10,364 40424 †2nd Float Oset -10,364 to 10,364 40425 †Gain Oset 0 to 255 40426 4 mA Set Point 0 - 10,364 mm 40427 20 mA Set Point 0 - 10,364 mm 40428 †4 mA Calibration 0 - 1,000 40429 †20 mA Calibration 0 - 1,000 40430 t1d NA* 40431 t1w NA* 40432 t1t NA* 40433 t2d NA* 40434 t2w NA* 40435 t2t NA* 40436-40437 Parameter 1 Data 0 to 1,000,000 mm
40438-40439 Parameter 2 Data 0 to 1,000,000 mm
40440-40441 Parameter 3 Data 0 to 1,000,000 mm 40442-40443 Parameter 4 Data 0 to 1,000,000 mm 40444-40445 Parameter 5 Data 0 to 1,000,000 mm
*These registers are not used by the MPX-F2 or MPX-F3, even though they are labeled in the APG Modbus
software.
Setting is factory calibrated. Do not adjust.
18
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
MPX-F2 and MPX-F3 APG Modbus Sensor Parameters
40401 - Units
Determines the units of measure for the Calculated Reading when Application Type is set to 0, 1, or 7. 1 = Feet 2 = Inches 3 = Meters For MPX-F2 and MPX-F3, this is seen only when using APG Modbus to program the MPX. This setting does not affect the 4-20 mA output.
40402 - Application Type
Determines the type of Calculated Reading performed by the sensor. 0 = Distance 1 = Level
2 = Standing Cylindrical Tank with or without Hemispherical Bottom 3 = Standing Cylindrical Tank with or without Conical Bottom 4 = Standing Rectangular Tank with or without Chute Bottom 5 = Horizontal Cylindrical Tank with or without Spherical Ends
6 = Spherical Tank 7 = Pounds (Linear Scaling) 8 = N/A
9 = Vertical Oval Tank 10 = Horizontal Oval Tank
11 = Strapping Chart See MPX-F Modbus Application Type Parameters pages 24-28.
For the MPX-F2 and MPX-F3, the 4-20 mA output can be scaled for linear output over distance/level or scaled for linear output over volume. When setup in any of the volumetric application types, the 4-20mA output becomes linear with regards to the volume (linear mA change per gallon, liter, etc.)
40403 - Volume Units
Determines the units of measure for the Calculated Reading when Application Type is set to 2 - 6 or 9 -11.
1 = Feet3 5 = Liters 2 = Million Feet3 6 = Inches3
3 = Gallons 7 = Barrels
4 = Meters3
40404 - Decimal Place
Determines the number of decimal places included in the Calculated Reading(s). For MPX-F2 and -F3, this is seen only when using APG Modbus to program the MPX. This setting does not affect the 4-20 mA output.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
19
40405 - Maximum Distance (Factory Calibrated)
Sets the distance (beginning from the Zero Reference) to the point where the sensor will stop looking for oat signals, usually the bottom of the stem. A oat beyond the Maximum Distance value will not be detect­ed.
40406 - Full Distance
Sets the positive distance (beginning from the sensor Zero Reference) to the point where the monitored vessel is considered full.
40407 - Empty Distance
Sets the positive distance (beginning from the Zero Reference) to the point where the monitored vessel is considered empty (usually the bottom of the stem).
40408 - Sensitivity (Factory Calibrated)
Sets the level of gain that is applied to the returning oat signal.
40409 - Pulses (Factory Calibrated)
Controls the duration of the signal being sent down the magnetostrictive wire.
40410 - Blanking (Factory Calibrated)
Sets the blanking distance, which is the zone from the Zero Reference of the sensor to the point from which
the rst signal will be valid. Signals from a oat in the blanking area will be ignored.
40411 - Fail Safe
Sets the output condition that the MPX-F will revert to in the event of a loss of return signal condition. 0 = Disable (no fail safe output) 1 = 3.8 mA 2 = 22 mA
20
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
40412 - Averaging (Factory Calibrated)
Sets the number of qualied received oat signals to average for the raw reading. Qualied received signals are placed in a rst-in, rst-out buer, the contents of which are averaged for the raw reading. The larger the number of qualied received signals being averaged, the smoother the reading will be, and the slower
the reading will be to react to quickly changing targets.
40413 - Filter Window (Factory Calibrated)
Determines the physical range (0 - 10,364 mm) of qualied received signals, based on the current raw
reading. Signals beyond the +/- Filter Window range of the current reading will not qualify unless the
average moves. Signals outside the extents of the Filter Window are written to the Out of Range samples buer (Holding Register 40414). See Figure 3.2.
Example: Window = 300 mm Out of Range Samples = 10
All samples
Samples are rejected within this area unless they persist
for 10 consecutive samples
Min. Reading Max. Reading
Current value
of Distance
are accepted
within this area
150 mm 150mm
Samples are rejected within this area unless they persist
for 10 consecutive samples
Figure 3.2
40414 - Out of Range Samples (Factory Calibrated)
Sets the number of consecutive samples outside the Filter Window (Holding Register 40413) necessary to
automatically adjust the current reading and move the Filter Window.
40415 - Sample Rate (Factory Calibrated)
Sets the update rate the sensor (10 - 1000 ms). Shorter time delays allow for quicker sensor response times to changing levels. Typical setting is 200 ms. Settings under 200 ms are not recommended.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
21
40416 - Multiplier (Factory Calibrated)
Calibrates the distance reading span. The Multiplier is shown by the values 1 - 1999, but these values are understood to represent 0.001 - 1.999. The default of 1000 (i.e. 1.000) is used for most applications.
40417 - Oset (Factory Calibrated)
Sets the Zero Reference of the sensor, the point from which the calculated distance is measured.
40418 - Pre lter
Denes the physical range (0 - 10,364 mm) of the start up (pre-lter) window. Four sample readings must be found within the Pre lter window for the MPX sensor to successfully start up.
This register is used for factory diagnostics only.
40419 - Noise limit
Sets the limit for number of signals (0-255) outside the Pre lter range for the MPX at start up. If the Noise Limit is reached before four readings register within the Pre lter window, the MPX will not start up.
This register is used for factory diagnostics only.
40421 - RTD Oset C° (Factory Calibrated)
Calibrates the RTD termperature sensor.
40422 - Float Window (Factory Calibrated)
Sets the distance (0 - 1000 mm) between the rst (i.e. top) oat and the point at which the sensor will begin looking for the second (bottom) oat. This essentially functions as a secondary blanking distance for the minimum depth of the top uid. Set to 0 for single oat.
22
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
40423 - 1st Float Oset (Factory Calibrated)
Used to calibrate top oat reading (-10,364 - 10,364 mm).
40424 - 2nd Float Oset (Factory Calibrated)
Used to calibrate bottom oat reading (-10,364 - 10,364 mm).
40425 - Gain Oset (Factory Calibrated)
Used to move the centerline of the oat response signal to optimize signal strength (0 - 255).
40426 - 4mA Set
Used to set the distance which will correspond to an output of 4 mA. For Application 1 (Distance), this is measured from the Zero Reference. For all other applications (Level & Volumetric) this is measured from the bottom of the probe.
40427 - 20mA Set
Used to set the distance which will correspond to an output of 20 mA. For Application 1 (Distance), this is measured from the Zero Reference. For all other applications (Level & Volumetric) this is measured from the bottom of the probe.
40428 - 4mA Cal (Factory Calibrated)
Used to calibrate the 4 mA output of the MPX-F2 or -F3.
40429 - 20mA Cal (Factory Calibrated)
Used to calibrate the 20 mA output of the MPX-F2 or -F3.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
23
Bottom Radius
MPX-F Modbus Application Type Parameters
Application 0 - Distance
Register Function Value Range
40400 Device Address 1 to 247 40401 Units 1 = Feet, 2 = Inches, 3 = Meters 40402 Application Type 0 40403 Volume Units -­40404 Decimal (Calculated) 0 - 3
Application 1 - Level
Register Function Value Range
40400 Device Address 1 to 247 40401 Units 1 = Feet, 2 = Inches, 3 = Meters 40402 Application Type 1 40403 Volume Units -­40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
Application 2 - Volume of Standing Cylindrical Tank ± Hemispherical Bottom
Register Function Value Range
40400 Device Address 1 to 247 40401 Units -- 40402 Application Type 2 40403 Volume Units 1 - 7 40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
40436-40437 Tank Diameter 0 - 1,000,000 (mm)
40438-40439 Radius of Bottom Hemisphere 0 - 1,000,000 (mm)
24
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Full
Level
or
Diameter
Application 3 - Volume of Standing Cylindrical Tank ± Conical Bottom
Register Function Value Range
40400 Device Address 1 to 247 40401 Units -- 40402 Application Type 3 40403 Volume Units 1 - 7 40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
40436-40437 Tank Diameter 0 - 1,000,000 (mm)
40438-40439 Cone Diameter (at bottom of cone) 0 - 1,000,000 (mm)
40440-40441 Length (height) of Cone 0 - 1,000,000 (mm)
Application 4 - Volume of Standing Rectangular Tank ± Chute Bottom
Register Function Value Range
40400 Device Address 1 to 247 40401 Units -- 40402 Application Type 4 40403 Volume Units 1 - 7 40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
Full
Level
Full
Level
Tank X
Diameter
Cone
Diameter
Cone
Length
Tank Y
Chute Length
Chute Y
40436-40437 Tank X Dimension 0 - 1,000,000 (mm)
or
40438-40439 Tank Y Dimension 0 - 1,000,000 (mm)
40440-40441 Chute X Dimension 0 - 1,000,000 (mm) 40442-40443 Chute Y Dimension 0 - 1,000,000 (mm)
Chute X
40444-40445 Length (height) of Chute 0 - 1,000,000 (mm)
NOTE: For all applications other than Distance, Empty Distance is usually the same as
Max Distance.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
25
Application 5 - Volume of Horizontal Cylindrical Tank ± Hemispherical Ends
Level
Register Function Value Range 40400 Device Address 1 to 247 40401 Units -- 40402 Application Type 5 40403 Volume Units 1 - 7 40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
40436-40437 Tank Length 0 - 1,000,000 (mm)
40438-40439 Tank Diameter 0 - 1,000,000 (mm) 40440-40441 Radius of End Hemispheres 0 - 1,000,000 (mm)
End Radius
Diameter
Length
Full
Application 6 - Volume of Spherical Tank
Register Function Value Range
40400 Device Address 1 to 247 40401 Units -- 40402 Application Type 6 40403 Volume Units 1 - 7 40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
40436-40437 Tank Diameter 0 - 1,000,000 (mm)
Full
Level
Diameter
26
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Application 7 - Pounds (Linear Scaling)
Register Function Value Range
40400 Device Address 1 to 247 40401 Units 1 = Feet, 2 = Inches, 3 = Meters 40402 Application Type 7 40403 Volume Units -­40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
40436-40437 Multiplier (linear scalar) 0 - 1,000,000 (1000 = 1.000)
Application 8 - N/A
Application 9 - Volume of Vertical Oval Tank
Register Function Value Range
40400 Device Address 1 to 247 40401 Units --
40402 Application Type 9
40403 Volume Units 1 - 7 40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
40436-40437 Tank Length 0 - 1,000,000 (mm)
40438-40439 Tank Depth 0 - 1,000,000 (mm)
40440-40441 Tank Width 0 - 1,000,000 (mm)
Full
Level
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Width
Depth
Length
27
Application 10 - Volume of Horizontal Oval Tank
Register Function Value Range
40400 Device Address 1 to 247 40401 Units -- 40402 Application Type 10 40403 Volume Units 1 - 7 40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
40436-40437 Tank Length 0 - 1,000,000 (mm)
40438-40439 Tank Depth 0 - 1,000,000 (mm)
40440-40441 Tank Width 0 - 1,000,000 (mm)
Full
Level
Depth
Width
Length
Application 11 - Strapping Chart (Polynomial Values)
Register Function Value Range
40400 Device Address 1 to 247 40401 Units 1 = Feet, 2 = Inches, 3 = Meters 40402 Application Type 11 40403 Volume Units 1 - 7 40404 Decimal (Calculated) 0 - 3
40405 Max Distance (factory set)
40406 Full Distance 0 - 10,364 mm 40407 Empty Distance 0 - 11,278 mm / 0 - 10,364 mm
40436-40437 X^3 Coecient 0 - 1,000,000 40438-40439 X^2 Coecient 0 - 1,000,000 40440-40441 X^1 Coecient 0 - 1,000,000 40442-40443 X^0 Coecient 0 - 1,000,000
28
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Chapter 4: Maintenance
General Care
Your MPX level sensor is designed to be low maintenance. However, in general, you should:
Periodically inspect your MPX to ensure the stem and oats are free of any heavy buildup that might impede the movement of the oats.
Ensure the housing cover is snuggly secured. If the cover becomes damaged or is misplaced, order a replacement immediately.
• Repair and Returns
Should your MPX level sensor require service, please contact the factory via phone, email, or online chat. We will issue you a Return Material Authorization (RMA) number with instructions.
Phone: 888-525-7300
Email: sales@apgsensors.com
Online chat at www.apgsensors.com
Please have your part number and serial number available. See Warranty and Warranty Restrictions for more information.
IMPORTANT: All repairs and adjustments of the MPX level sensor must be made by
the factory. Modifying, disassembling, or altering the MPX on site is strictly prohibited.
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
29
Chapter 5: Hazardous Location Installation and Certication
Hazardous Location and Non-Incendive Wiring Diagram
30
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
CSA Certicate of Compliance
Issued by:
Certificate of Compliance
Certificate: 2397437 (237484) Master Contract: 237484 Project: 70022593 Date Issued: 2016-04-22 Issued to: Automation Products Group Inc
1025 West 1700 North Logan, Utah 84321 USA
Attention: Karl Reid
The products listed below are eligible to bear the CSA Mark shown
with adjacent indicators 'C' and 'US' for Canada and US or with adjacent
indicator 'US' for US only or without either indicator for Canada only.
Ravindra (Ravi) Kanthe
Ravindra (Ravi) Kanthe
PRODUCTS CLASS - C225802 - PROCESS CONTROL EQUIPMENT-For Hazardous Locations­CLASS - C225882 - PROCESS CONTROL EQUIPMENT-For Hazardous Locations - Certified to US Standards
Class I, Division 1 & 2, Groups C and D
Ex d IIB Ex nA IIB Class I, Zone 1; AEx d IIB Class I, Zone 2; AEx nA IIB
x Float Level Sensors, Model MPX- E, R, G &T (MPX- abc-de-fghi-jjj), rated 12 - 24 Vdc, 80mA, or
rated 12 to 24 Vdc, 4-20mA; operating ambient Ta is 85°C; Temperature Code T4; Ingress protection IP65; Field wiring is non-incendive when installed per drawing 9003468.
Class I, Division 2, Groups C and D Ex nA IIB Class I, Zone 2; AEx nA IIB
DQD 507 Rev. 2012-05-22 Page 1
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
31
Certificate:
2397437
Project:
General Requirements Canadian Electrical Code, Part II Tenth Edition
CSA C22.2 No 30-M1986 (Reaffirmed 2007)
Explosion-Proof Enclosures for Use in Class I Hazardous Locations Industrial Products – Third Edition
CSA C22.2 No 142-M1987 (Reaffirmed 2014)
Process Control Equipment Industrial Products – Third Edition
Non-incendive Electrical Equipment for Use in Class I,
Edition
Electrical apparatus for explosive gas atmospheres – Part 0: General requirements – First Edit ion
Electrical apparatus for explosive gas atmospheres – Part 1: Flameproof enclosures "d" – First Edition
CSA E60079-15-02 (Reaffirmed 2006)
Electrical Apparatus for Explosive Gas Atmospheres – Part 15: Type of Protection "n" – Second Edition
Industrial Control Equipment - Seventeenth Edition;
2010
Explosion-Proof and Dust-Ignition-Proof Electrical
Including October 28, 2009
Nonincendive Electrical Equipment for Use in Class I and
(Classified) Locations
Explosive atmospheres - Part 0 Equipment - General requirements
70022593
Master Contract: 237484
Date Issued: 2016-04-22
x Float Level Sensors, Model MPX- F (model MPX- abc-de-fghi-jjj), rated 12 - 24 Vdc, 80mA, or rated
12 to 24 Vdc, 4-20mA; operating ambient Ta is 85°C; Temperature Code T4; Ingress protection IP65; Field wiring is non-incendive when installed per drawing 9003468
Notes for all equipment:
1. The model code breakdown is as follows: a= E, R, F, G or T; b= 1, 2, 3 or 4; c= A,B or C, d= A, B, C, D, E, F, G, Z, X, V, T, R, M, or J; e= N, B, D, Y, W, U, S, P, L, K, or I; f= F, R, P, S, N, or O; g= 1, 1.5, 2, 2.5, 3,
3.5, 4, 5, 6; h=W or S; i= A, B, C or D; and j= 12–153 for the 1/2" stem Type E or 48–300 for the 1" stem Type R, Type G and Type T or 120-456 for Flex stem Type F.
2. The equipment is intended to be installed as required by the applicable electrical code (CEC, NEC) and as specified by the manufacturers Installation Instructions.
3. The installation will be inspected by the authority with jurisdiction in the area where installed.
APPLICABLE REQUIREMENTS
CSA C22.2 No 0-10
CSA C22.2 No 213-M1987 (Reaffirmed 2008)
CSA C22.2 No 60079-0-07
CSA C22.2 No 60079-1-07
UL 508
UL 1203
ANSI/ISA-12.12.01-2007
UL 60079-0 - 5th Ed (Dec 2009)
DQD 507 Rev. 2012-05-22 Page 2
32
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Division 2 Hazardous Locations Industrial Products – First
Reprint with Revisions Through and Including April 15,
Equipment for Use in Hazardous (Classified) Locations ­Fourth Edition; Reprint with Revisions through and
II, Division 2 and Class III, Divisions 1 and 2 Hazardous
Certificate:
2397437
Project:
Explosive atmospheres - Part 1 Equipment Protection by
Flameproof Enclosures “d”
Electrical Apparatus for Explosive Gas Atmospheres - Part 15 - Electrical Apparatus with Type of Protection n
70022593
Master Contract: 237484
Date Issued: 2016-04-22
UL 60079-1 - 6th Ed (Apr 2009)
UL 60079-15 - 3rd Ed (Oct 2009)
MARKINGS
Please refer to Descriptive report for markings and Instructional material.
DQD 507 Rev. 2012-05-22 Page 3
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
33
Supplement to Certificate of Compliance
Project
Date
Description
70022593
2016
Update to Report 2397437 to include 3 new Probe types MPX (for Class 1, Div. 1 & 2 and Zones 1 & 2) and MPX and Zone 2), new PCB design, and update product label material and printing method.
2440956
2011
Update to report 2397437 to correct misspellings and add note in Factory Tests section of report.
2397437
2011-05-18
Original certification for the model MPX-ab-cd-efg-hhh float level sensors.
Certificate: 2397437 (237484) Master Contract: 237484
The products listed, including the latest revision described below,
are eligible to be marked in accordance with the referenced Certificate.
Product Certification History
-04-22
-07-19
-G, MPX-T
-F (for Class 1, Div. 2
DQD 507 Rev. 2012-05-22 Page 1
34
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
35
APG
R
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • sales@apgsensors.com
Automation Products Group, Inc.
Loading...