878-03-02
Pay attention to the enclosed documentation
THIS MANUAL MUST BE CONSIDERED AS AN INTEGRATING PART OF THE UNIT AND MUST ALWAYS STAY NEAR IT.
THE FOLLOWING ADVICES MUST BE KEPT IN THE HIGHEST CONSIDERATION IN ORDER TO AVOID DAMAGES TO
PEOPLE AND/OR TO THE UNIT ITSELF.
THE MANUFACTURER IS NOT RESPONSIBLE IN CASE THE USER DOES NOT FOLLOW THE INSTRUCTIONS
CONTAINED IN THIS MANUAL.
THE EQUIPMENT IN ITS OPERATIVE CONFIGURATION IS AN X-RAY SOURCE, SO PLEASE FOLLOW X-THE X-RAY
PROTECTION REGULATIONS IN FORCE.
Reference number : 3W -40-002 Rev 0 Page 1 of 42
KRISTAL – User’s Manual
CONTENTS
1. GENERAL DESCRIPTION...............................................................................................................................3
1.1 General Technical Features..................................................................................................................4
2. SPOT FILM DEVICE......................................................................................................................................15
2.1 Functions on push button panel........................................................................................................15
2.2 Instructions of use.................................................................................................................................15
5.2 Errors that imply to restart the unit because of movement anomalies........................................41
5.3 Movement Errors ...................................................................................................................................41
5.5 Communication Errors.........................................................................................................................41
Page 2 of 42 Reference number : 3W -40-002 Rev 0
KRISTAL – User’s Manual
1. GENERAL DESCRIPTION
TILTING 90/20 & TILTING 90/90 TABLES
Compact and versatile Unit, tilting table remote controlled for fluoroscopy, angio-seriografy and
standard examinations.
Proper for any type of examination of routine and speciality, without any limitation, with wide versatility
and operational simplicity.
Thanks to its dimensions and weight reduced, can be installed in smalls diagnostic rooms
The wide longitudinal runs of column, gives the possibility to execute the patient exploration-without moving
it. All movements are silent, and guarantee to the patient a great safety in the behavior of the
examinations.
Tilting movement with stop in any position and automatic stop in horizontal, vertical and
Trendelemburg.
Spot film device programs allow the execution of automatic tomography and universal tomography
can be carried out with any inclination of the table, allowing oblique projections with incidences to 30°
besides.
The table tilting allows you to make examinations on external accessories only on one side (for TILTING
90°/20° Table) or on both sides (for TILITING 90°/90° Table).
All the movements are under microprocessor control.
End of runs have double safety and self diagnosis.
Operational console with ergonomic position of the controls, visualization on display graphic VGA of all
the functions of the system.
Are displayed:
- the focus film distance,
- the angle of tilting,
- the height of tomo layer,
- the dimension of the cassette with the relative S.F.D. programs and all other parameters connected with
the system.
- all the movement controls are also available on the table front end keyboard.
- the S.F.D. with auto-centering cassette system allows the possibility to use mm. or inches cassettes, from
18x24 cm to 35x43 cm with free division format.
- predisposition for digital system or cine with possibility to install an image-intensifier from 6 to 16 inches.
Reference number : 3W -40-002 Rev 0 Page 3 of 42
KRISTAL – User’s Manual
1.1 General Technical Features
Type: 90/90° TILTING TABLE
90/20° TILTING TABLE
Power supply: 3 x 2307kVA – (400V) 7kVA
Power output of auxiliary mains socket -outlets 230 V- 1Kva 50 Hz
1.1.1 Technical Data
Tilting table with progressive starting + 90°/-90°, +90°/-20°
Tilting speed 4 °/sec
Table top 73 x 240 cm
Absorption coefficient <_ 0,8 mm Al.
Max patient weight 140Kg
Lateral movement ±11,5 cm (Max speed 3 cm/sec)
Plane-film distance 7,5 cm
Film - I.I. distance 7.5 cm
Movement column 1400 mms (Max speed 13 cm/sec)
X-ray tube ass. rotation + 90° - 180° manually
Focus - film distance from 107 to 153 cm (Max speed 3 cm/sec)
Tube angulation ±30°
Compression Motorized
Angiography Step by step
Prepared to receive Image Intensifier 9", 12", 14",16"
Tomographic function
Linear tomography plain-plain
Tomographic angles 8°, 15°, 30°, 40°
Tomographic speeds 4 speeds
Layer height adjustment 1?300 mm
Fixed SID 110 cm
SFD
S.F.D. electronic automatic
Film divisions only in vertical sense;
Division 2/3/4/5/6
Prepared for 3 field ionization camera ( 5 mm. max thick.)
S.F.D. allows rapid series exposures
Cassettes from 18x24 to 35x43 cm in both directions
Transition time between fluoro and radiography about 1 sec.
Grid Motorized, scattered
On a single film it is possible to have radio and tomo exposures
Tube Features
Max. weight of X-ray tube 24 Kg
Page 4 of 42 Reference number : 3W -40-002 Rev 0
KRISTAL – User’s Manual
Weight 950 Kg
Optional
- Compression band
- Lateral cassette holder
Electric Classification
Class: I with applied part of B type
Required environmental features
Temperature Operative +10°C to +35°C
Storage –10°C to +40°C
Maximum humidity 80% with dew
Maximum altitude for use 3500 meters
NOTE: Do not use the equipment in flammable or explosive environment.
This equipment is not protected against water.
Operating Conditions
Continuous functioning with intermittent charge.
Manufactured by
IMAGO Radiology S.r.l.
For APELEM-DMS Group
Parc Scientifique Georges Besse
175, Allée Von Neumann
30035 NIMES CEDEX 1
Reference number : 3W -40-002 Rev 0 Page 5 of 42
KRISTAL – User’s Manual
1.1.2 Classification
- Class I equipment with applied part of "B" type.
- Common equipment (equipment with cover not protected against water penetration).
- Equipment not to be used with an anaesthetic mixture flammable with air, oxygen or nitrous oxide.
- Continuous functioning with continuous intermittent rated load.
ELECTROMAGNETIC COMPATIBILITY
This equipment is built in conformity to EN 60601 –1 -2 regulations: electro-medical equipments electromagnetic compatibility.
The conformity is guaranteed only in case the unit is installed and used accordingly to the instruction
manual.
In order to avoi d any malfunctioning, we recommend you to use the unit at an adequate distance from
disturbances sources such as monitors, loudspeakers, transceivers or any other radio-frequency
source.(Also mobile phones or normal phones can cause problems.)
Page 6 of 42 Reference number : 3W -40-002 Rev 0
KRISTAL – User’s Manual
1.2 DIMENSIONS
1.2.1 Lateral view 90/90° TILTING TABLE
Reference number : 3W -40-002 Rev 0 Page 7 of 42
KRISTAL – User’s Manual
1.2.2 Front view 90/90° TILTING Table
Page 8 of 42 Reference number : 3W -40-002 Rev 0
KRISTAL – User’s Manual
1.2.3 Front view 90/90° TILTING Table
Reference number : 3W -40-002 Rev 0 Page 9 of 42
KRISTAL – User’s Manual
1.2.4 Lateral view 90/20° TILTING Table
Page 10 of 42 Reference number : 3W -40-002 Rev 0
KRISTAL – User’s Manual
1.2.5 Front view 90/20° TILTING Table
Reference number : 3W -40-002 Rev 0 Page 11 of 42
KRISTAL – User’s Manual
1.2.6 Front view 90/20° TILTING Table
Page 12 of 42 Reference number : 3W -40-002 Rev 0
KRISTAL – User’s Manual
1.2.7 Console – Rack 90/90 & 90/20 TILTING Tables
Reference number : 3W -40-002 Rev 0 Page 13 of 42