Anritsu 37225A Software Users Guide

Page 1
Model 360X/372X XA
e
Vector Network Analyzer
Software User ’s Guid
Software Revision 1.03
This manual applies to the
WILTRON 360X/372XXA Vector Network Analyzer
Performance Verification Software,
WILTRON Part Number 2300-175, Revision 1.03
490 JARVIS DRIVE MORGAN HILL, CA 95037-2809
P/N: 10410-00162
REVISION:A
PRINTED: FEBRUARY 1995
COPYRIGHT 1995 WILTRON CO.
Page 2
WARRANTY
The WILTRON product(s) listed on the title page is (are) warranted against defects in materials and workmanship for one year from the date of shipment. WILTRON’s obligation covers repairing or replacing products which prove to be defective during the warranty period. Buyers shall prepay transportation charges for equipment returned to WILTRON for warranty repairs. Obligation is limited to the original purchaser. WILTRON is not liable for consequential damages.
LIMITATION OF WARRANTY
The foregoing warranty does not apply to WILTRON connectors that have failed due to normal wear. Also, the warranty does not apply to defects resulting from improper or inadequate maintenance by the Buyer, unauthorized modification or misuse, or operation outside of the environmental specifications of the product. No other warranty is expressed or implied, and the remedies provided herein are the Buyer’s sole and exclusive remedies.
TRADEMARK ACKNOWLEDGEMENTS
V Connector and K Connector are registered trademarks of WILTRON Company. Ink Jet and Think Jet are registered trademarks of Hewlett-Packard Co. MS-DOS is a registered trademark of Microsoft Corporation.
NOTICE
WILTRON Company has prepared this manual for use by WILTRON Company personnel and customers as a guide for the proper installation, operation and maintenance of WILTRON Compan y equipment and co mputer progr ams. The draw­ings, specification s, and infor mation contain ed herein are the prope rty of WILTRON Company, and any unauthorized use or disclosure of these drawings, specifications, and information i s pr oh ibited; they sh all n ot be reproduced, c opied, or used in w ho le or in part as the basis for manufacture or sale of the equipment or software programs without the prior written consent of WILTRON Company.
Page 3
Table of Contents
Chapter 1 General Information
1 Introduction . . . . . . . . . . . . . . . . . . . . 1
2 Format of the Ver ific ation Software Disks . . . . 1
3 Capability . . . . . . . . . . . . . . . . . . . . . 1
4 Data Output . . . . . . . . . . . . . . . . . . . . 1
Chapter 2 Required Equipment
5 Introduction . . . . . . . . . . . . . . . . . . . . 3
6 Computer (System Controller) . . . . . . . . . . 3
7 GPIB Interface Card . . . . . . . . . . . . . . . 3
8 Vector Network Analyzer . . . . . . . . . . . . . 3
9 Verification Kit . . . . . . . . . . . . . . . . . . 3
10 Printer . . . . . . . . . . . . . . . . . . . . . . . 4
11 Cables . . . . . . . . . . . . . . . . . . . . . . . 4
Chapter 3 Configuring the System
12 Introduction . . . . . . . . . . . . . . . . . . . . 5
13 Hardwar e Interconne ction . . . . . . . . . . . . 5
14 Performance Verification
(PV) Software Installat ion . . . . . . . . . . . 5
Chapter 4 Running the Program
15 Introduction . . . . . . . . . . . . . . . . . . . . 6
16 Starting the Program . . . . . . . . . . . . . . . 6
17 Entering System Data and Calibration . . . . . 7
18 Measuring the Verification Devices . . . . . . . 8
Chapter 5 In Case of Trouble
19 Difficulty Running the Program . . . . . . . . . 10
20 Difficulty Meeting System Specifications . . . . 10
i/ii
Page 4
Chapter 1
General Information
Contents Para Name Page
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Format Of The Verification Software Disks . . . . . . . . . . 1
3 Capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
4 Data Output . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1 Introduction
This manual supports the WILTRON 2300-175 VNA Performance Veri- fication Software. This software is used with the WILTRON 360X/372XXA Vector Networ k Analy zer Sy stems. The man ual is org an­ized into five chapters, as shown below.
General Information—provides an overview of the product.
Required Equipment—describes the test equipment, compo-
nents, hardware and software required to use this product. Confi guring the S yste m—describes how the equipment is setup
and interconnected. Running the Program—provides step-by-step instructions for
running the software. In Case of Trouble—contains troubleshooting suggestions and
service information.
2 Format of the Verification Software Disks
The Performan ce Ve rif ication Sof tware is o ffer ed on an IBM co mpatible 720KB floppy disk.
3Capability
This software provides for automating measurements of the test com­ponents contained in a WILTRON 3666, 3667, 3668, 3669, or 3669B Verification Kit.
It compares the measurements made on your 360X/372XXA with the test component data provided in each verification kit. This will aid in determining if the measurement values obtained for the components are within the uncertainty limits provided on the data disk packaged with each verification kit.
4 Dat a Ou tpu t
The test data and results are output in the form of fou r files to th e root directory of your computer’s drive C:\ (in the root directory.) The default file names are:
Page 1
Page 5
20db_pad.dat 40db_pad.dat (or 50db_pad.dat) air_line.dat rho__std.dat
NOTE:
The Performanc e Verification Softwar e allow s you to ren ame these files to file names of your choice with default “.dat” extension.
The tabular data in each file is given at discrete frequencies (1.0, 1.5, or 3.0 GHz inter vals—de pendin g on the con nector type and co rrespond ­ing frequency range).
Page 2
Page 6
Chapter 2
Required Equipment
Contents Para Name Page
5 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
6 Computer (System Controller) . . . . . . . . . . . . . . . . . 3
7 GPIB Interface Card . . . . . . . . . . . . . . . . . . . . . . . 3
8 Vector Network Analyz er . . . . . . . . . . . . . . . . . . . . 3
9 Ve rification Kit . . . . . . . . . . . . . . . . . . . . . . . . . . 3
10 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
11 Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
5 Introduction
This section describes the equipment required to use the Performance Verification Software.
6 Computer (System Controller)
You need an IBM AT or compatible computer w ith 640K of RAM , a h ard disk configured as C: drive, a 3.5 inch floppy disk drive, and MS-DOS
3.3 (or higher) Operating System. This computer system is also used as the system controller (see below for GPIB configuration.)
7 GPIB Interface Card
Your computer must have a National Instruments IEEE-488 GPIB Interface Adapter running driver version 2.1 or later. Ensure your GPIB adapter is setup to default configuration as per National’s In­struct ions.
8 Vector Network Analyzer
Yo u need a WILTRON Model 360X/372XXA Ve ctor Network Analyzer System, with the appropriate 3650, 3651, 3652, 3654, or 3654B Cali­bration Kit. A 360X system consists of a 360/360A/360B VNA, a 36XX­Series Test Set, and a Wiltron RF signal source.
The 360X/372XXA V ector Netw or k Analyzer will be referr ed to as VNA throughtout this User’s Guide.
9 Verification Kit
You need a WILTRON 3666, 3667, 3668, 3669, or 3669B Verification Kit, with data disk.
Page 3
Page 7
10 Printer
A printer is not required since the verification results and data will be stored in four files on your computer’s hard disk drive (C:). These files are saved in ASCII format for easy viewing and printing later if de­sired.
11 Cables
You need a GPIB cable (Wiltron PN: 2100-2) and a two-foot RF Test Port Cable (WILTRON PN: 3670A50-2, 3670K50-2, 3670V50-2, 3671A50-2, 3671K50-2, or 3671V50-2.)
Page 4
Page 8
Chapter 3
Configuring the System
Contents Para Name Page
12 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
13 Hardware Interconnection . . . . . . . . . . . . . . . . . . . . 5
14 Software Installation and Configuration . . . . . . . . . . . . 5
12 Introduction
This chapter describes how the various system elements are intercon­nected and the preliminary steps required for operation of the verifi­cation software.
13 Hardware Interconnection
Connect the computer’s GPIB port to the VNA system using the GPIB cable.
If the VNA is a 360X Network Analy zer, connect to th e GPIB co nnec ter labeled “360 GPIB” not the one labeled “System Bus.”
If the VNA is a 372XXA Network Analyzer, connect to the GPIB con­necter labeled “IEEE 488.2 GPIB” not the one labeled “Dedicated GPIB.”
Connect the two-foot test port cable female end to the VNA Port 2. Connect a female-female Phase Equal Instertable Adapter from the
Calibration Kit to the VNA Port 1. This does not apply to GPC-7 Calibration.
14 Performance Verification (PV) Software Installation
The PV software can either be run directly from the floppy disk or copied to and run from your C: drive. The Verification data disk, when requested by the PV software must be installed in drive A:.
NOTE:
The PV distribution disk contains batch files that show ex­amples of how to set up your computer so that B: drive can be used.
Ensure your GPIB Adapter driver software is installed and tested as per National Instrument’s Instructions.
Page 5
Page 9
Chapter 4
Running the Program
Contents Para Name Page
15 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
16 Starting The Program . . . . . . . . . . . . . . . . . . . . . . 6
17 Entering System Data And Calibration . . . . . . . . . . . . . 7
18 Measuring the Verification Devices . . . . . . . . . . . . . . . 8
15 Introduction
The Performance Verification software runs in DOS text mode as de­scribed below.
16 S tart in g the Pr ogr am
With th e equipment an d softwar e config ured as desc ribed in Ch apter 3, turn your computer on and allow it to boot up to the DOS prompt (C:\.)
Once the computer has finished booting up to the DOS prompt, insert the performance verification software disk in drive A: or start it from your C: drive.
At the DOS prompt, Type: vnaveri Press: <ENTER> to invoke the performance verification program. The following message will be displayed on the computer screen:
WILTRON VNA VERIFICATION SOFTWARE
* * VERSION 1.03 * *
Press any key to continue. At this point you can in stall the 3.5-in ch data di sk from the Verification
Kit into drive A: and leave it in for the duration of the Performance Verification.
Page 6
Page 10
17 E nt e ri n g S ys te m Da ta and C alibrat ion
The program now prompts you to enter the following data:
Start Frequency
Verification Kit Type
Stop Frequency
Enter the information as it is requested. Next, you will be prompted to verify that the calibration kit coeff icients
have been loaded into the VNA system. To do this: a. Insert the data disk from your WILTRON Calibration Kit (not the
verification kit) in the Network Analyzer’s disk drive.
b. Press the UTILITY MENU key on the front panel of the network
analyzer. (The Utility Menu is displayed on the netwo rk analyze r’s display screen.)
c. Select CALIBRATION COMPONENT UTILITIES on the Utility
Menu and press <ENTER>. (The Calibration Component Utilities Menu is displayed.)
d. Select INSTALL CALIBRATION COMPONENT INFORMAT ION
FROM DISK on the Calibration Component Utility Menu and press <ENTER>. (Loading of the calibration kit coefficients from the data disk begins.)
When the calibration kit coefficients have been loaded into the VNA system, press any key on the computer keyboard to continue.
The program automatically sets up the proper frequency points for your measurement based upon your responses to the earlier program prompts. The progr am then automatic ally sets up the VNA so that you can manually perform a full, 12-term calibration of the VNA system. Using the two-foot test cable, set up the VNA system for calibration as shown in Figure 1. Perform the full, 12-term calibration following the prompts presented on the network analyzer’s display screen and using the calibration components from your WILTRON Calibration Kit.
Completion of the calibration is indicated by the lighting of the “FULL 12-TERM” CALIBRATION LED indicator on the 360X VNA and the “APPLY CAL” LED indicator only on the 372XXA VNA front panel.
Press any key on the computer keyboard to continue the verification program.
Next, the program prompts you to enter header information for the data printouts, as follows:
VNA Serial Number (found on the Network Analyzer’s rear
panel). Verification Kit Serial Number (found on the label on the side of
the verification kit wooden box). Date (dd/MONTH/yy format). For example, if the date is Sep-
tember 1, 1993 then enter 01/SEPT/93. Operator’s Name (up to 12 characters long-no spaces).
Page 7
Page 11
Enter the i n form ation as it is re q u ested , p r essin g <ENTER> after each
NETWORK ANALYZER
SIGNAL SOURCE
TEST SET
2 FOOT TEST CABLE
INSERT
CALIBRATION
COMPONENTS
HERE
PORT 1
PORT 2
entry. The program now prompts you to verify that the Verification Kit data
disk is installed in drive A: of your computer. Install the disk if you have not already done so.
18 Measuring the Verification Devices
The program now presents the follow ing menu on the computer scree n:
(1) Measure 20 dB Attenuator (2) Measure 40 or 50 dB Attenuator (3) Measure Beatty Standard (4) Measure Airline (5) Exit
Enter the number corresponding to the action you wish to take. At this point you can begin measuring your verification kit devices. As
you select a verification device, the program prompts you to enter the serial number of the device and to orient the device with the label facing the user. Such orientation places port 1 on your left. It is at this measurement point that the device is installed and measured on the VNA. See Figure below showing the measurement setup.
The computer will now read data from the Verification Kit data disk (installed in drive A:.) The verification software then automatically measures the device with the VNA and compares the measured data with the data stored on the Verification Kit data disk. The results and measurement data is stored to data files on your C: hard drive (see Section 1-4, Data Output for file names.) The data on each file will
Page 8
Figure 1. VNA Calibration and Measurement Setup Shown on a 360X VNA.
Page 12
contain an asterisk (*) next to measrement data that is not within the specified limits.
The above steps can be repeated for each of the three remaining devic es in the verification kit: the 40/50 dB attenuator, the Beatty standard, and the air line.
NOTE:
The software will wr ite the d ata to the file name you choose, or the default file names in Sec tio n 1-4. If these files alr eady exist, they will be overwritten. Rename the files if you do
not wish them overwritten with the new data. To e xit the program, enter 5 at the main menu. To view or print the data files, simply use any text editor. You can also
view the data files using the batch file called veridata.bat included on your Performance Verification Software distribution disk.
Page 9
Page 13
Chapter 5
In Case of Trouble
Contents Para Name Page
19 Difficulty Running the Program . . . . . . . . . . . . . . . . 10
20 Difficulty Meeting System Specifications . . . . . . . . . . . 10
19 Difficulty Running the Program
If you have difficulty getting the program to run properly: a. Check your GPIB interconnection. b. Check cables and addresses, ensure that the GPIB-PC directory is
present on the boot drive and contain s the GPIB.COM file, and tha t the CONFIG.SYS file contains the line DEVICE=GPIB.COM.
c. Ensure that the performance verification software disk is in drive
A: and that it contains the VNAVERI.EXE file.
d. Ensure that after starting-up your Perfomance Verification Soft-
ware, the Verification Kit data disk is installed in drive A: and that it contains the following files:
DEV1S11.DAT DEV3S11.DAT UNCDEV1.DAT DEV1S12.DAT DEV3S12.DAT UNCDEV2.DAT DEV1S21.DAT DEV3S21.DAT UNCDEV3.DAT DEV1S22.DAT DEV3S22.DAT UNCDEV4.DAT DEV2S11.DAT DEV4S11.DAT VALDEV1.DAT DEV2S12.DAT DEV4S12.DAT VALDEV2.DAT DEV2S21.DAT DEV4S21.DAT VALDEV3.DAT DEV2S22.DAT DEV4S22.DAT VALDEV4.DAT
e. Ensure ANSI.SYS is a device in your CONFIG.SYS file. For exam-
ple, if your Disk Operating System is in a directory called DOS on your C drive , you would add the f ollowin g line to y our CONFI G.SY S file: DEVICE = C:\DOS\ANSI.SYS.
If, after checking the above, you still have difficulty, contact WILTRON Customer Service at (408) 778-2000 (fax : 408 778 0239) an d ask for th e Vector Network Analyzers Support Engineer for further assistance.
20 Difficulty Meeting System Specifications
If the verification software appears to run properly but the results are not within the measu remen t limits a ssociated with th e verific ation k it:
a. Check both the verific ation kit and calibra tion kit devic es for signs
of physical damage. Make sure that the connectors are clean.
Page 10
b. Ensure that the serial number of the verification kit data disk
matches that shown on the verification kit.
Page 14
c. Repeat the proc ess with a f resh calibration . Save the results of both
measurements as an aid in troubleshooting, if you require factory assistance.
d. When installing calibration devices, and when measuring
verification devices, pay particular attention to proper connector alignment and torque. Torque the connector using the torque wrench supplied with the calibration kit.
If, after following the above steps, you still have difficulty, please contact WILTRON Customer Service at (408) 778-2000 (fax: 408 778
0239) and ask for the Vector Network Analyzers Support Engineer for further assist ance.