The WILTRON product(s) listed on the title page is (are) warranted against defects
in materials and workmanship for one year from the date of shipment.
WILTRON’s obligation covers repairing or replacing products which prove to be
defective during the warranty period. Buyers shall prepay transportation charges
for equipment returned to WILTRON for warranty repairs. Obligation is limited to
the original purchaser. WILTRON is not liable for consequential damages.
LIMITATION OF WARRANTY
The foregoing warranty does not apply to WILTRON connectors that have failed
due to normal wear. Also, the warranty does not apply to defects resulting from
improper or inadequate maintenance by the Buyer, unauthorized modification or
misuse, or operation outside of the environmental specifications of the product. No
other warranty is expressed or implied, and the remedies provided herein are the
Buyer’s sole and exclusive remedies.
TRADEMARK ACKNOWLEDGEMENTS
V Connector and K Connector are registered trademarks of WILTRON Company.
Ink Jet and Think Jet are registered trademarks of Hewlett-Packard Co.
MS-DOS is a registered trademark of Microsoft Corporation.
NOTICE
WILTRON Company has prepared this manual for use by WILTRON Company
personnel and customers as a guide for the proper installation, operation and
maintenance of WILTRON Compan y equipment and co mputer progr ams. The drawings, specification s, and infor mation contain ed herein are the prope rty of WILTRON
Company, and any unauthorized use or disclosure of these drawings, specifications,
and information i s pr oh ibited; they sh all n ot be reproduced, c opied, or used in w ho le
or in part as the basis for manufacture or sale of the equipment or software
programs without the prior written consent of WILTRON Company.
This manual supports the WILTRON 2300-175 VNA Performance Veri-
fication Software. This software is used with the WILTRON
360X/372XXA Vector Networ k Analy zer Sy stems. The man ual is org anized into five chapters, as shown below.
General Information—provides an overview of the product.
•
Required Equipment—describes the test equipment, compo-
•
nents, hardware and software required to use this product.
Confi guring the S yste m—describes how the equipment is setup
•
and interconnected.
Running the Program—provides step-by-step instructions for
•
running the software.
In Case of Trouble—contains troubleshooting suggestions and
•
service information.
2Format of the Verification Software Disks
The Performan ce Ve rif ication Sof tware is o ffer ed on an IBM co mpatible
720KB floppy disk.
3Capability
This software provides for automating measurements of the test components contained in a WILTRON 3666, 3667, 3668, 3669, or 3669B
Verification Kit.
It compares the measurements made on your 360X/372XXA with the
test component data provided in each verification kit. This will aid in
determining if the measurement values obtained for the components
are within the uncertainty limits provided on the data disk packaged
with each verification kit.
4Dat a Ou tpu t
The test data and results are output in the form of fou r files to th e root
directory of your computer’s drive C:\ (in the root directory.) The
default file names are:
The Performanc e Verification Softwar e allow s you to ren ame
these files to file names of your choice with default “.dat”
extension.
The tabular data in each file is given at discrete frequencies (1.0, 1.5,
or 3.0 GHz inter vals—de pendin g on the con nector type and co rrespond ing frequency range).
This section describes the equipment required to use the Performance
Verification Software.
6Computer (System Controller)
You need an IBM AT or compatible computer w ith 640K of RAM , a h ard
disk configured as C: drive, a 3.5 inch floppy disk drive, and MS-DOS
3.3 (or higher) Operating System. This computer system is also used
as the system controller (see below for GPIB configuration.)
7GPIB Interface Card
Your computer must have a National Instruments IEEE-488 GPIB
Interface Adapter running driver version 2.1 or later. Ensure your
GPIB adapter is setup to default configuration as per National’s Instruct ions.
8Vector Network Analyzer
Yo u need a WILTRON Model 360X/372XXA Ve ctor Network Analyzer
System, with the appropriate 3650, 3651, 3652, 3654, or 3654B Calibration Kit. A 360X system consists of a 360/360A/360B VNA, a 36XXSeries Test Set, and a Wiltron RF signal source.
The 360X/372XXA V ector Netw or k Analyzer will be referr ed to as VNA
throughtout this User’s Guide.
9Verification Kit
You need a WILTRON 3666, 3667, 3668, 3669, or 3669B Verification Kit,
with data disk.
Page 3
Page 7
10Printer
A printer is not required since the verification results and data will be
stored in four files on your computer’s hard disk drive (C:). These files
are saved in ASCII format for easy viewing and printing later if desired.
11Cables
You need a GPIB cable (Wiltron PN: 2100-2) and a two-foot RF Test
Port Cable (WILTRON PN: 3670A50-2, 3670K50-2, 3670V50-2,
3671A50-2, 3671K50-2, or 3671V50-2.)
This chapter describes how the various system elements are interconnected and the preliminary steps required for operation of the verification software.
13Hardware Interconnection
Connect the computer’s GPIB port to the VNA system using the GPIB
cable.
If the VNA is a 360X Network Analy zer, connect to th e GPIB co nnec ter
labeled “360 GPIB” not the one labeled “System Bus.”
If the VNA is a 372XXA Network Analyzer, connect to the GPIB connecter labeled “IEEE 488.2 GPIB” not the one labeled “Dedicated
GPIB.”
Connect the two-foot test port cable female end to the VNA Port 2.
Connect a female-female Phase Equal Instertable Adapter from the
Calibration Kit to the VNA Port 1. This does not apply to GPC-7
Calibration.
The PV software can either be run directly from the floppy disk or
copied to and run from your C: drive. The Verification data disk, when
requested by the PV software must be installed in drive A:.
NOTE:
The PV distribution disk contains batch files that show examples of how to set up your computer so that B: drive can
be used.
Ensure your GPIB Adapter driver software is installed and tested as
per National Instrument’s Instructions.
The Performance Verification software runs in DOS text mode as described below.
16S tart in g the Pr ogr am
With th e equipment an d softwar e config ured as desc ribed in Ch apter 3,
turn your computer on and allow it to boot up to the DOS prompt (C:\.)
Once the computer has finished booting up to the DOS prompt, insert
the performance verification software disk in drive A: or start it from
your C: drive.
At the DOS prompt, Type:
vnaveri
Press: <ENTER> to invoke the performance verification program.
The following message will be displayed on the computer screen:
WILTRON VNA VERIFICATION SOFTWARE
* * VERSION 1.03 * *
Press any key to continue.
At this point you can in stall the 3.5-in ch data di sk from the Verification
Kit into drive A: and leave it in for the duration of the Performance
Verification.
Page 6
Page 10
17E nt e ri n g S ys te m Da ta and C alibrat ion
The program now prompts you to enter the following data:
Start Frequency
•
Verification Kit Type
•
Stop Frequency
•
Enter the information as it is requested.
Next, you will be prompted to verify that the calibration kit coeff icients
have been loaded into the VNA system. To do this:
a. Insert the data disk from your WILTRON Calibration Kit (not the
verification kit) in the Network Analyzer’s disk drive.
b. Press the UTILITY MENU key on the front panel of the network
analyzer. (The Utility Menu is displayed on the netwo rk analyze r’s
display screen.)
c.Select CALIBRATION COMPONENT UTILITIES on the Utility
Menu and press <ENTER>. (The Calibration Component Utilities
Menu is displayed.)
d. Select INSTALL CALIBRATION COMPONENT INFORMAT ION
FROM DISK on the Calibration Component Utility Menu and
press <ENTER>. (Loading of the calibration kit coefficients from
the data disk begins.)
When the calibration kit coefficients have been loaded into the VNA
system, press any key on the computer keyboard to continue.
The program automatically sets up the proper frequency points for
your measurement based upon your responses to the earlier program
prompts. The progr am then automatic ally sets up the VNA so that you
can manually perform a full, 12-term calibration of the VNA system.
Using the two-foot test cable, set up the VNA system for calibration as
shown in Figure 1. Perform the full, 12-term calibration following the
prompts presented on the network analyzer’s display screen and using
the calibration components from your WILTRON Calibration Kit.
Completion of the calibration is indicated by the lighting of the “FULL
12-TERM” CALIBRATION LED indicator on the 360X VNA and the
“APPLY CAL” LED indicator only on the 372XXA VNA front panel.
Press any key on the computer keyboard to continue the verification
program.
Next, the program prompts you to enter header information for the
data printouts, as follows:
VNA Serial Number (found on the Network Analyzer’s rear
•
panel).
Verification Kit Serial Number (found on the label on the side of
•
the verification kit wooden box).
Date (dd/MONTH/yy format). For example, if the date is Sep-
•
tember 1, 1993 then enter 01/SEPT/93.
Operator’s Name (up to 12 characters long-no spaces).
•
Page 7
Page 11
Enter the i n form ation as it is re q u ested , p r essin g <ENTER> after each
NETWORK
ANALYZER
SIGNAL
SOURCE
TEST SET
2 FOOT
TEST CABLE
INSERT
CALIBRATION
COMPONENTS
HERE
PORT 1
PORT 2
entry.
The program now prompts you to verify that the Verification Kit data
disk is installed in drive A: of your computer. Install the disk if you
have not already done so.
18Measuring the Verification Devices
The program now presents the follow ing menu on the computer scree n:
(1) Measure 20 dB Attenuator
(2) Measure 40 or 50 dB Attenuator
(3) Measure Beatty Standard
(4) Measure Airline
(5) Exit
Enter the number corresponding to the action you wish to take.
At this point you can begin measuring your verification kit devices. As
you select a verification device, the program prompts you to enter the
serial number of the device and to orient the device with the label
facing the user. Such orientation places port 1 on your left. It is at this
measurement point that the device is installed and measured on the
VNA. See Figure below showing the measurement setup.
The computer will now read data from the Verification Kit data disk
(installed in drive A:.) The verification software then automatically
measures the device with the VNA and compares the measured data
with the data stored on the Verification Kit data disk. The results and
measurement data is stored to data files on your C: hard drive (see
Section 1-4, Data Output for file names.) The data on each file will
Page 8
Figure 1. VNA Calibration and Measurement Setup Shown on a 360X VNA.
Page 12
contain an asterisk (*) next to measrement data that is not within the
specified limits.
The above steps can be repeated for each of the three remaining devic es
in the verification kit: the 40/50 dB attenuator, the Beatty standard,
and the air line.
NOTE:
The software will wr ite the d ata to the file name you choose,
or the default file names in Sec tio n 1-4. If these files alr eady
exist, they will be overwritten. Rename the files if you do
not wish them overwritten with the new data.
To e xit the program, enter 5 at the main menu.
To view or print the data files, simply use any text editor. You can also
view the data files using the batch file called veridata.bat included on
your Performance Verification Software distribution disk.
If you have difficulty getting the program to run properly:
a. Check your GPIB interconnection.
b. Check cables and addresses, ensure that the GPIB-PC directory is
present on the boot drive and contain s the GPIB.COM file, and tha t
the CONFIG.SYS file contains the line DEVICE=GPIB.COM.
c.Ensure that the performance verification software disk is in drive
A: and that it contains the VNAVERI.EXE file.
d. Ensure that after starting-up your Perfomance Verification Soft-
ware, the Verification Kit data disk is installed in drive A: and that
it contains the following files:
e. Ensure ANSI.SYS is a device in your CONFIG.SYS file. For exam-
ple, if your Disk Operating System is in a directory called DOS on
your C drive , you would add the f ollowin g line to y our CONFI G.SY S
file: DEVICE = C:\DOS\ANSI.SYS.
If, after checking the above, you still have difficulty, contact WILTRON
Customer Service at (408) 778-2000 (fax : 408 778 0239) an d ask for th e
Vector Network Analyzers Support Engineer for further assistance.
20Difficulty Meeting System Specifications
If the verification software appears to run properly but the results are
not within the measu remen t limits a ssociated with th e verific ation k it:
a. Check both the verific ation kit and calibra tion kit devic es for signs
of physical damage. Make sure that the connectors are clean.
Page 10
b. Ensure that the serial number of the verification kit data disk
matches that shown on the verification kit.
Page 14
c.Repeat the proc ess with a f resh calibration . Save the results of both
measurements as an aid in troubleshooting, if you require factory
assistance.
d. When installing calibration devices, and when measuring
verification devices, pay particular attention to proper connector
alignment and torque. Torque the connector using the torque
wrench supplied with the calibration kit.
If, after following the above steps, you still have difficulty, please
contact WILTRON Customer Service at (408) 778-2000 (fax: 408 778
0239) and ask for the Vector Network Analyzers Support Engineer for
further assist ance.