Anritsu 37211B Software Users Guide

Page 1
Model 360X/372XX(A /B)/373XXA
Vector Network Analyzer
Software User’s Guide
Software Revision 1.04
This manual applies to the
WILTRON 360X/372XX(A/B)/373XXA Vector Network Analyzer
Performance Verification Software,
WILTRON Part Number 2300-175, Revision 1.04
490 JARVIS DRIVE MORGAN HILL, CA 95037-2809
P/N: 10410-00162
REVISION: B
PRINTED: JUNE 1996
COPYRIGHT 1995 WILTRON CO.
Page 2
WARRANTY
The WILTRON product(s) listed on the title page is (are) warranted against defects in materials and workmanship for one year from the date of shipment.
WILTRON’s obligation covers repairing or replacing products which prove to be defective during the warranty period. Buyers shall prepay transportation charges for equipment returned to WILTRON for warranty repairs. Obligation is limited to the original purchaser. WILTRON is not liable for consequential damages.
LIMITATION OF WARRANTY
The foregoing warranty does not apply to WILTRON connectors that have failed due to normal wear. Also, the warranty does not apply to defects resulting from improper or inadequate maintenance by the Buyer, unauthorized modification or misuse, or operation outside of the environmental specifications of the product. No other warranty is expressed or implied, and the remedies provided herein are the Buyer’s sole and exclusive remedies.
TRADEMARK ACKNOWL EDGEMENTS
V Connector and K Connector are registered trademarks of WILTRON Company. GPC-7 is a registered trademark of Amphenol Corporation. ANACAT is a registered trademark of EEsof, Inc. Ink Jet and Think Jet are registered trademarks of Hewlett-Packard Co. MS-DOS is a registered trademark of Microsoft Corporation.
NOTICE
WILTRON Company has prepared this manual for use by WILTRON Company personnel and customers as a guide for the proper installation, operation and maintenance of WILTRON Company equipment and computer programs. The draw­ings, specifications, and information contained herein are the property of WILTRON Company, and any unauthorized use or disclosure of these drawings, specifications, and information is prohibited; they shall not be reproduced, copied, or used in whole or in part as the basis for manufacture or sale of the equipment or software programs without the prior written consent of WILTRON Company.
Page 3
CONTENTS
Page
Chapter 1 General Information . . . . . . . . 1
Chapter 2 Required Equipment . . . . . . . . 3
Chapter 3 Configuring the System . . . . . . . 5
Chapter 4 Running the Program . . . . . . . . 6
Chapter 5 In Case of Trouble . . . . . . . . . 10
Appendix A Example of Tabular
Test Results . . . . . . . . . . . . A1
i/ii
Page 4
Chapte r 1
General Infor mation
Contents Para Name Page
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Format Of The Verification Software Disks . . . . . . . . . . 1
3 Capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
4 Data Output . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1 Introduction
This manual supports the WILTRON 2300-175 VNA Performance Veri- fication Software. This software is used with the WILTRON 360X/372XX(A/B)/373XXA Vector Network Analyzer Systems. The manual is organized into five chapters, as shown below.
General Information—provides an overview of the product.
Required Equipment—describes the test equipment, compo-
nents, hardware and software required to use this product. Configuring the System—describes how the equipment is setup
and interconnected. Running the Program—provides step-by-step instructions for
running the software. In Case of Trouble—contains troubleshooting suggestions and
service information.
2 Format of the Verification Software Disks
The Performance Verification Software is offered on an IBM compatible
1.44MB floppy disk.
3Capability
This software provides for automating measurements of the test com­ponents contained in a WILTRON 3666, 3667, 3668, 3669, or 3669B Verification Kit.
It compares the measurements made on your 360X/372XX(A/B)/373XXA with the test component data provided in each verification kit. This will aid in determining if the measurement values obtained for the components are within the uncertainty limits provided on the data disk packaged with each verification kit.
4 Data Output
The test data and results are output in the form of four files to a directory (C:\verify.vna\data) on your computers hard drive. The de­fault file names, depending on the type of test being performed, are:
Page 1
Page 5
20db.dat 40db.dat or 50db.dat airline.dat beatty.dat
NOTE:
The Performance Verification Software allows you to rename these files to file names of your choice with default ".dat" extension.
The tabular data in each file is given at discrete frequencies (1.0, 1.5, or 3.0 GHz intervals—depending on the connector type and correspond­ing frequency range).
Page 2
Page 6
Chapte r 2
Required Equ ipment
Contents Para Name Page
5 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
6 Computer (System Controller) . . . . . . . . . . . . . . . . . 3
7 GPIB Interface Card . . . . . . . . . . . . . . . . . . . . . . . 3
8 Vector Network Analyzer . . . . . . . . . . . . . . . . . . . . 3
9 Verification Kit . . . . . . . . . . . . . . . . . . . . . . . . . . 3
10 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
11 Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
5 Introduction
This section describes the equipment required to use the Performance Verification Software.
6 Computer (System Controller)
You need an IBM AT or compatible computer with 640K of RAM, a hard disk configured as C: drive, a 3.5 inch floppy disk drive, and MS-DOS
3.3 (or higher) Operating System. This computer system is also used as the system controller (see below for GPIB configuration.)
7 GPIB Interface Card
Your computer must have a National Instruments IEEE-488 GPIB Interface Adapter running driver version 2.1 or later. Ensure your GPIB adapter is setup to default configuration as per National’s In­structions.
8 Vector Network Analyzer
You need a WILTRON Model 360X/372XX(A/B)/373XXA Vector Network Analyzer System, with the appropriate 3650, 3651, 3652, 3654, or 3654B Calibration Kit. A 360X system consists of a 360/360A/360B VNA, a 360X-Series Test Set, and a Wiltron RF signal source.
The 360X/372XX(A/B)/373XXA Vector Network Analyzer will be re­ferred to as VNA throughout this User’s Guide.
9 Verification Kit
You need a WIL T RON 3666, 3667, 3668, 3669, or 3669B Verification Kit, with data disk.
Page 3
Page 7
10 Printer
A printer is not required since the verification results and data will be stored in four files on your computer’s hard disk drive (C:). These files are saved in ASCII format for easy viewing and printing later if de­sired.
11 Cables
You need a GPIB cable (Wiltron PN: 2100-2) and a two-foot RF Test Port Cable (WILTRON PN: 3670A50-2, 3670K50-2, 3670V50-2, 3671A50-2, 3671K50-2, or 3671V50-2.)
Page 4
Page 8
Chapte r 3
NETWORK ANALYZER
SIGNAL SOURCE
TEST SET
2 FOOT TEST CABLE
INSERT
CALIBRATION
COMPONENTS
HERE
PORT 1
PORT 2
Configuring the System
Contents Para Name Page
12 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
13 Hardware Interconnection . . . . . . . . . . . . . . . . . . . . 5
14 Software Installation and Configuration . . . . . . . . . . . . 5
12 Introduction
This chapter describes how the various system elements are intercon­nected and the preliminary steps required for operation of the verifi­cation software.
13 Hardware Interconnection
Connect the computer’s GPIB port to the VNA system using the GPIB cable.
If the VNA is a 360X Network Analyzer, connect to the GPIB connector labeled "360 GPIB" not the one labeled "System Bus."
If the VNA is a 372XX(A/B)/373XXA Network Analyzer, connect to the GPIB connector labeled "IEEE 488.2 GPIB" not the one labeled "Dedi- cated GPIB."
Connect the two-foot test port cable female end to the VNA Port 2. Connect a female-female Phase Equal Insertable Adapter from the
Calibration Kit to the VNA Port 1. This does not apply to GPC-7 Calibration.
Figure 1. VNA Calibration and Measurement Setup Shown on a 360X VNA.
Page 5
Page 9
14 Performance Verification (PV) Software Installation
The installation is very simple with the vinstall program. Just put the disk into a floppy drive, switch to that floppy drive and type vinstall then press the enter key. The install program creates a directory called verify.vna on your computers C: hard disk. Unlike the previous ver­sion, the Verification data disk can be in drive A or B. Ensure your GPIB Adapter driver software is installed and tested as per National Instruments Instructions.
NOTE:
The program can be run from a floppy disk as before, however, you must have a directory on the C: hard drive called "VERIFY.VNA\DATA".
Page 6
Page 10
Chapte r 4
Running the Program
Contents Para Name Page
15 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
16 Starting The Program . . . . . . . . . . . . . . . . . . . . . . 6
17 Entering System Data And Calibration . . . . . . . . . . . . . 7
18 Measuring the Verification Devices . . . . . . . . . . . . . . . 8
15 Introduction
The Performance Verification software runs in DOS text mode as de­scribed below.
16 Starting the Program
With the equipment and software configured as described in Chapter 3 Section 14, turn on your computer and allow it to boot up to the DOS prompt (C:\).
At the DOS prompt, Type : vnaveri Press: <ENTER> to invoke the performance verification program. If you are going to run the Verification program from a floppy without
installing the software do the following : after your computer boots up switch to the c drive type mkdir verify.vna then press the enter key type mkdir verify.vna\data then press the enter key put the verification software disk in the floppy drive of your choice,
switch to that floppy drive an Type : vnaveri.exe Press: <ENTER> to invoke the performance verification program. The following message will be displayed on the computer screen:
WILTRON VNA VERIFICATION SOFTWARE
* * VERSION 1.04 * *
Press any key to continue. At this point you can install the 3.5-inch data disk from the Verification
Kit into an available floppy drive and leave it in for the duration of the Performance Verification.
Page 7
Page 11
17 Entering System Data and Calibration
The program now prompts you to enter or confirm the following data:
The VNA’s GPIB Address. The default Wiltron address is 6
The VNA’s Family (360X, 272XX[A or B], 373XXA)
If a 360X was selected you will be prompted to see if option 5 is installed.
Start Frequency
Verification Kit Type
Stop Frequency
Verify the calibration kit coefficients have been loaded into the
VNA system.
To do this: a. Insert the data disk from your WILTRON Calibration Kit (not the
verification kit) in the Network Analyzer’s disk drive.
b. Press the UTILITY MENU key on the front panel of the network
analyzer. (The Utility Menu is displayed on the network analyzer’s display screen.)
c. Select CALIBRATION COMPONENT UTILITIES on the Utility
Menu and press <ENTER>. (The Calibration Component Utilities Menu is displayed.)
d. Select INSTALL CALIBRATION COMPONENT INFORMATION
FROM DISK on the Calibration Component Utility Menu and press <ENTER>. (Loading of the calibration kit coefficients from the data disk begins.)
When the calibration kit coefficients have been loaded into the VNA system, press any key on the computer keyboard to continue.
The program automatically sets up the proper frequency points for your measurement based upon your responses to the earlier program prompts. The program then automatically sets up the VNA so that you can manually perform a full, 12-term calibration of the VNA system. Using the two-foot test cable, set up the VNA system for calibration as shown in Figure 1 (chapter 3 section 13).
Perform the full, 12-term calibration
Follow the prompts presented on the network analyzer’s display screen and using the calibration components from your WILTRON Calibration Kit.
Completion of the calibration is indicated by the lighting of the "FULL 12-TERM" CALIBRATION LED indicator on the 360X VNA and the "APPLY CAL" LED indicator only on the 372XX(A/B)/373XXA VNA front panel.
Press any key on the computer keyboard to continue the verification program.
Page 8
Page 12
Enter the Letter of the floppy drive the Verification Data Disk
is in
Next, the program prompts you to enter header information for the data printouts, as follows:
VNA Serial Number (found on the Network Analyzer’s rear
panel). Verification Kit Serial Number (found on the label on the side of
the verification kit wooden box). Date (dd/MONTH/yy format). For example, if the date is Sep-
tember 1, 1993 then enter 01/SEPT/93. Operator’s Name (up to 12 characters long-no spaces).
Enter the information as it is requested, pressing <ENTER> after each entry.
18 Measuring the Verification Devices
The program now presents the following menu on the computer screen:
(1) Measure 20 dB Attenuator (2) Measure 40 or 50 dB Attenuator (3) Measure Beatty Standard (4) Measure Airline (5) Restart Program
(6) Exit
Enter the number corresponding to the action you wish to take. At this point you can begin measuring your verification kit devices. As
you select a verification device, the program prompts you to enter the serial number of the device and to orient the device with the label facing the user. Such orientation places port 1 on your left. It is at this measurement point that the device is installed and measured on the VNA.
The computer will now read data from the Verification Kit data disk installed in the floppy drive you entered earlier. The verification soft­ware then automatically measures the device with the VNA and com­pares the measured data with the data stored on the Verification Kit data disk. The results and measurement data is stored to data files on the hard drive directory C:\VERIFY.VNA\DATA (see Section 1-4, Data Output for file names.) The data on each file will contain an asterisk (*) next to measurement data that is not within the specified limits.
The above steps can be repeated for each of the three remaining devices in the verification kit: the 40/50 dB Attenuator, the Beatty standard, and the air line.
Page 9
Page 13
NOTE:
The software will write the data to the file name you choose, or the default file names in Section 1-4. If these files already exist, they will be overwritten. Rename the files if you do not wish them overwritten with the new data.
To exit the program, enter 6 at the main menu. To view or print the data files, simply use any text editor. You can also
view the data files using the batch file called veridata.bat included on your Performance Verification Software distribution disk.
Page 10
Page 14
Chapte r 5
In Case o f Trouble
Contents Para Name Page
19 Difficulty Running the Program . . . . . . . . . . . . . . . . 10
20 Difficulty Meeting System Specifications . . . . . . . . . . . 10
19 Difficulty Running the Program
If you have difficulty getting the program to run properly: a. Check your GPIB interconnection. b. Check cables and addresses, ensure that the GPIB-PC directory is
present on the boot drive and contains the GPIB.COM file, and that the CONFIG.SYS file contains the line DEVICE=GPIB.COM.
c. If you are running the program from the hard disk it should be
installed with the install program. If you are running from a floppy drive with or without installing the program create a directory on the C: drive called VERIFY.VNA\DATA, change to the floppy drive with the program disk, type VNAVERI.EXE and press the enter key to start the program.
d. Ensure that after starting-up your Performance Verification Soft-
ware, the Verification Kit data disk is installed in drive A or B (the floppy drive you told the program it was in) and that it contains the following files:
DEV1S11.DAT DEV3S11.DAT UNCDEV1.DAT DEV1S12.DAT DEV3S12.DAT UNCDEV2.DAT DEV1S21.DAT DEV3S21.DAT UNCDEV3.DAT DEV1S22.DAT DEV3S22.DAT UNCDEV4.DAT DEV2S11.DAT DEV4S11.DAT VALDEV1.DAT DEV2S12.DAT DEV4S12.DAT VALDEV2.DAT DEV2S21.DAT DEV4S21.DAT VALDEV3.DAT DEV2S22.DAT DEV4S22.DAT VALDEV4.DAT
If, after checking the above, you still have difficulty, contact WILTRON Customer Service at (408) 778-2000 (fax: 408 778 0239) and ask for the Vector Network Analyzer Support Engineer for further assistance.
20 Difficulty Meeting System Specifications
If the verification software appears to run properly but the results are not within the measurement limits associated with the verification kit:
a. Check both the verification kit and calibration kit devices for signs
of physical damage. Make sure that the connectors are clean.
b. Ensure that the serial number of the verification kit data disk
matches that shown on the verification kit.
Page 11
Page 15
c. Repeat the process with a fresh calibration. Save the results of both
measurements as an aid in troubleshooting, if you require factory assistance.
d. When installing calibration devices, and when measuring
verification devices, pay particular attention to proper connector alignment and torque. Torque the connector using the torque wrench supplied with the calibration kit.
If, after following the above steps, you still have difficulty, please contact WILTRON Customer Service at (408) 778-2000 (fax: 408 778
0239) and ask for the Vector Network Analyzer Support Engineer for further assistance.
Page 12
Page 16
Appendix A A1
Appendix A
Example of Tabular Test Results
WILTRON VNA Verification SW Version 1.04
37200 Serial Number: 456001 Verification Kit Model Number: 3667 Verification Kit Serial Number: 89001 Date: 04/01/96 Device Serial Number: 050021 Operator: VnaTester
AIR LINE
—————————————————-S12—————————————————­Freq
(GHz)
0.04 0.0013 0.1013 -0.1000 0.1100 -0.9 71.3 -72.2 180.0
1.50 0.0050 2.2862 -2.2812 0.0903 131.8 -173.4 54.9 180.0
3.00 0.0052 0.4899 -0.4847 0.0906 -171.7 -99.0 -72.6 180.0
4.50 0.0045 2.6800 -2.6755 0.0909 -119.0 -110.6 -8.5 180.0
6.00 0.0013 0.4873 -0.4860 0.0912 -79.8 83.3 -163.1 180.0
7.50 0.0033 0.4291 -0.4258 0.0915 130.0 -78.9 151.2 180.0
STD (db)
MEAS (db)
DIFF UNC
(+/-)
STD (DEG)
MEAS (DEG)
DIFF UNC
(+/-)