Anritsu 11410-00276 User Manual

Measurement Methods
During network deployment, mainte­nance, and trouble shooting phases, insertion loss can be measured by dis­connecting the antenna and connecting an enclosed short at the end of the transmission line. If a Tower Mounted Amplifier (TMA) is used in the transmis­sion feed line system, it is best to remove the TMA and antenna from the system configuration to perform an insertion loss measurement. It is best to always dis­connect the cable at the same location so the measured data can be compared to the historical data for accuracy and repeatability.
Using Site Master, cable insertion loss can be measured in CABLE LOSS or RETURN LOSS mode. In Cable Loss mode, Site Master automatically consid­ers the signal traveling in both directions thus making the measurement easier for the user in the field.
The following section explains the pro­cedure to measure insertion loss in cable loss mode and return loss mode. The measurement setup and equipment required is the same for both modes.
Insertion Loss Measurement Methods
APPLICA TION NOTE
Introduction
Transmission feed line system performance plays an important role in wireless network coverage. Insertion loss measurement is one of the critical measurements used to analyze transmission feed line installation and performance quality. This application note explains how Site Master is used to measure cable insertion loss with different test methods and how to predict the maxi­mum allowable cable insertion loss through manual calculations.
In wireless communication systems, the transmit and receive antennas are connected to the radio through coaxial cable and/or waveguide transmission lines (Figure 1).
Insertion loss measures the energy absorbed by the transmission line in the direction of the signal path in dB/meter or dB/feet. Transmission line losses are dependent on cable type, operating frequency and the length of the cable run. Insertion loss of a cable varies with frequency; the higher the frequency, the greater the loss.
Insertion loss measurements help troubleshoot the network by verifying the cable installation and cable performance. High insertion loss in the feedline or jumpers can contribute to poor system performance and loss of coverage. Measuring insertion loss using Site Master assures accurate and repeatable measurements.
SITE MASTER™
Verify Cable Inser tion Loss From Ground Level
Figure 1. Typical transmission line system
Antenna
Jumper
Main Feeder
Connectors
Antenna Jumper
2
Measuring Insertion Loss Using CABLE LOSS MODE
Application Note
Required Equipment
• Site Master Model S11xx, S33xx, or S251x
• Precision Open/Short, Anritsu 22N50 or Precision Open/Short/Load, Anritsu OSLN50LF
• Precision Load, Anritsu SM/PL
• Test Port Extension Cable, Anritsu 15NNF50-1.5C
• Optional 510-90 Adapter, DC to 7.5 GHz, 50 Ω, 7/16(f)-N(m)
Figure 3. Insertion loss measurement setup when TMA is in line.
Step 4. Connect the phase stable test
port extension cable to the RF port.
Step 5. Calibrate the Site Master at the
end of the phase stable test port extension cable. (See the section entitled “Calibrating Site Master with the Phase Stable Test Port Extension Cable” for details.)
Step 6. Disconnect the antenna and
connect an enclosed precision "short" at the end of the transmission line.
Step 7. Connect the other end of the
transmission line to the phase stable cable of the Site Master. A trace will be displayed on the screen when the Site Master is in continuous sweep mode.
Step 8. Press the
AMPLITUDE key and
set the TOP and BOTTOM values of the display. In Figure 4, the TOP is set to 2 dB, and the BOTTOM is set to 5 dB.
Step 9. Press the
MARKER key.
Step 10.Set
M1 to MARKER TO PEAK.
Step 11.Set M2 to MARKER TO VALLEY. Step 12.Calculate the insertion loss by
averaging M1 (MARKER TO PEAK) and M2 (MARKER TO VALLEY) values as follows:
Step 13. Press
SAVE DISPLAY and name
the trace using the soft keys and press
ENTER.Saving the
display is recommended for historical documentation. Saved traces can be used in the future to compare, check or verify changes in the trans­mission feedline performance.
Procedure
Step 1. Power On the Site Master and
press the
MODE key.
Step 2. Select FREQ-CABLE LOSS using
the Up/Down arrow key and press
ENTER.
Step 3. Set the start and stop frequen-
cies F1 and F2. As an example, F1=750 MHz and F2=850 MHz for a typical cellular frequency band.
Insertion Loss Measurement Setup
The insertion loss measurement set up for a typical transmission feed line system is shown in Figure 2. Remove the antenna and connect an enclosed precision “short” at the end of the transmission line.
If a Tower Mounted Amplifier (TMA) is in the transmission feed line system, remove the TMA and antenna and connect an enclosed short at the end of the transmission line. Insertion loss measurement for a transmission feed line system with a tower mounted amplifier is shown in Figure 3.
Figure 2. An insertion loss measurement setup after antenna is removed.
Antenna Removed
TMA Jumper
Antenna Jumper
Antenna Jumper
Precision Short
Antenna Removed
Test Port
Jumper Main Feeder
Cable
Site Master S251C
ESCAPE CLEAR
1
2
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
56
MARKER7LIMIT
ENTER
8
SAVE
RECALL
RUN
DISPLAY
DISPLAY
HOLD
0
9
+/-
PRINT
ON
SYS
OFF
.
SWEEP
MODE
AMPLITUDE
FREQ/DIST
TMA Removed
Test Port
Jumper Main Feeder
Cable
Site Master S251C
ESCAPE
CLEAR
1
2
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
56
MARKER7LIMIT
ENTER
8
SAVE
RECALL
RUN
DISPLAY
DISPLAY
HOLD
0
9
+/-
PRINT
ON
SYS
OFF
.
SWEEP
MODE
AMPLITUDE
FREQ/DIST
Tx/Rx
Precision Short
Insertion Loss = (M1 dB + M2 dB)
2
= 3.23 + 3.95 = 3.59 dB
2
In Cable Loss mode, the Site Master automatically considers the signal path in both directions when calculating the cable insertion loss. As such, Cable Loss mode is recommended when making cable insertion loss measurements.
3
Step 7. Connect the other end of the transmission
line to the phase stable cable of the Site Master. A trace will be displayed on the screen when the Site Master is in continuous sweep mode.
Step 8. Press the AMPLITUDE key and set the TOP
and BOTTOM values of the display. In Figure 5, the TOP is set to 4 dB, and the
BOTTOM is set to 10 dB. Step 9. Press the MARKER key. Step 10.Set M1 to MARKER TO PEAK. Step 11.Set M2 to MARKER TO VALLEY. Step 12.Calculate the insertion loss by averaging M1
(MARKER TO PEAK) and M2 (MARKER
TO VALLEY) values and dividing by two as
follows:
Step 13.Press
SAVE DISPLAY and name the trace
using the soft keys and press ENTER.
Calibrating Site Master with the Phase Stable Test Port Extension Cable
The phase stable test port extension cable is used as an extension cable to the test port of the Site Master, and ensures accurate and repeatable measurements. The phase stable cable can be moved and bent while making a measurement without causing errors in the measurement. When poor quality cables are used as an extension test port cable, large error will be introduced in the measurements when the cable is moved.
For accurate results, Site Master should be calibrated at the ambient temperature before making any measurements. The Site Master must be re-calibrated whenever the setup frequency changes, the tempera­ture exceeds the calibration temperature window or when the test port extension cable is removed or replaced.
Connect the phase stable test port extension cable to the Site Master RF Port. The Site Master with phase stable cable combination can be calibrated manually using Open, Short and Load (OSL) precision components (Figure 6), or by using the InstaCal
®
module. Manual calibration is explained here. For the InstaCal procedure, refer to the Site Master user guide.
Note: The InstaCal module is not compatible with the Site Master S251C Model.
Figure 4. Typical Insertion Loss Display in Cable Loss Mode.
LIM ON
Figure 5. Typical Insertion Loss Display in Return Loss Mode
Measuring Insertion Loss using RETURN LOSS MODE
Step 1. Power on the Site Master and press the
MODE key.
Step 2. Select the FREQ-RETURN LOSS using the
Up/Down arrow key and press ENTER.
Step 3. Set the start and stop frequencies F1 and F2.
As an example, F1=750 MHz and F2=850 MHz for a typical cellular frequency band.
Step 4. Connect the phase stable test port extension
cable to the RF port.
Step 5. Calibrate the Site Master at the end of
the phase stable test port extension cable. (See the section entitled “Calibrating Site Master with the Phase Stable Test Port Extension Cable” for details.)
Step 6. Disconnect the antenna and connect an
enclosed precision "short" at the end of the transmission line.
Application Note
CL - CABLE LOSS
517 POINTS
M3=OFF M4=OFF
RECALL
M1
M2
M3
M4
M5
M6
CAL ON
2.00 dB
5.00 dB
750 dB FREQ (MHz) 850.0 M1:3.23,750.4 MHz M2:3.95,848.4 MHz
LIM ON
RL - RETURN LOSS
517 POINTS
M3=OFF M4=OFF
RECALL
M1
M2
M3
M4
M5
M6
CAL ON
4.00 dB
10.00 dB
750 dB FREQ (MHz) 850.0 M1:6.47,750.4 MHz M2:7.89,848.4 MHz
M1 dB + M2 dB
Insertion Loss =
2 2
6.47 + 7.89
=
2 2
= 3.59 dB
Figure. Calibrating at the end of the Phase Stable Test Port Extension Cable.
Discover Whats Possible
®
SALES CENTERS: United States (800) ANRITSU Europe 44 (0) 1582-433433 Microwave Measurements Division Canada (800) ANRITSU Japan 81 (03) 3446-1111 490 Jarvis Drive, Morgan Hill, CA 95037-2809 South America 55 (21) 286-9141 Asia-Pacific 65-2822400 http://www.us.anritsu.com
Application Note
For example:
Manual Calibration Procedure
Step 1. Power on the Site Master. Step 2. Select the appropriate frequency range. Step 3. Connect the phase stable test port extension cable to
the RF port.
Step 4. Press the START CAL key. The message "Connect Open to
RF OUT port or connect InstaCal module and press ENTER" will appear in the display.
Step 5. Connect the OPEN precision calibration component
to the end of the test port extension cable. Press the ENTER Key.
Step 6. The message "Measuring OPEN" will appear, and after the
measurement "Connect SHORT to RF OUT" will appear.
Step 7. Remove the “open” and connect the “short” precision cali-
bration component to the test port extension cable. Press the ENTER key.
Step 8. The message "Measuring SHORT" will appear, and after the
measurement "Connect TERMINATION to RF OUT" will appear.
Step 9. Remove the “short” and connect the “precision termination”
at the end of the test port extension cable. Press the ENTER key.
Step 10.The message "Measuring TERMINATION" will appear. After
the measurement, the “CAL OFF” message will change to "CAL ON" on the upper left-hand corner of the display.
Step 11.Remove the precision termination from the text port
extension cable.
Calculating Transmission Line Insertion Loss
Cables have different insertion losses at different frequencies. For example LDF4-40A attenuation at 1 GHz is 0.022 dB/ft (0.073 dB/m) and at 2 GHz it is 0.0325 dB/ft (0.107 dB/m). As the frequency increases or the length of the cable run increases, the amount of cable insertion loss increases.
To verify cable insertion loss measurements are reasonable, the expected insertion loss can be calculated manually using the following procedure:
• Calculate the estimated worst loss of each component in the transmission line system.
• Add all the component’s estimated worst losses together to calculate total insertion loss in the transmission line system.
Insertion loss of the = Bottom Jumper Loss transmission system + Main Cable Loss
+ Top Jumper Loss
+ Connector Losses = 0.408 + 2.3 +0.204 +1.12 = 4.03 dB
Compare the measured insertion loss to the calculated insertion loss to verify transmission line performance. The measured cable insertion loss should be lower than the calculated cable inser­tion loss.
Summary
The preferred method to measure Cable Insertion Loss using Site Master is Cable Loss mode. Cable loss mode automatically considers the signal traveling in both directions and thus makes it easier to measure the cable insertion loss in the field. Measured Insertion Loss should always be com­pared to the calculated loss to verify accuracy thus assuring transmission line performance. The calculated insertion loss is usually a “worst case scenario”.
Cable insertion loss may be difficult to measure on excessivly long or highly lossy cables. When the cable insertion loss is greater than 20 dB, it will be hard to measure.
All trademarks used are the property of their respective owners.
Cable
Type
Cable Attenuati on (dB/ft)
Cable
Length
Insertion
Loss (dB)
Bottom Jumper
LDF4-50A 0.0204 X 20 = 0.408
Main Cable
LDF5-50A 0.0115 X 200 = 2.30
Top Jumper
LDF4-50A 0.0204 X 10 = 0.204
Number of
Connector Pairs
Loss per Pair
in dB
Total Connector
Loss (dB)
4 X 0.28 = 1.12
Note: For Cable Loss-One Por t measurements, Site Master S251C requires only one port calibration.
Note: If the phase stable cable is removed from the test port, the calibration is not valid.
11410-00276 Rev. A, ©Anritsu March 2003, Data subject to change without notice
RF Out/Reflection
Test Port
Site Master S331C
SWEEP
MODE
AMPLITUDE
FREQ/DIST
Site Master S11xx/S33xx/S251x
Open Short Load
Test Port Cable (Optional)
ESCAPE CLEAR
1
2
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
56
MARKER7LIMIT
ENTER
8
SAVE
RECALL
RUN
DISPLAY
DISPLAY
HOLD
0
9
+/-
PRINT
ON
SYS
OFF
.
Calibration
Loading...