Manual Calibration Procedure
Step 1. Power on the Site Master.
Step 2. Select the appropriate frequency range.
Step 3. Connect the phase stable test port extension cable to
the RF port.
Step 4. Press the START CAL key. The message "Connect Open to
RF OUT port or connect InstaCal module and press ENTER"
will appear in the display.
Step 5. Connect the OPEN precision calibration component
to the end of the test port extension cable. Press the ENTER
Key.
Step 6. The message "Measuring OPEN" will appear, and after the
measurement "Connect SHORT to RF OUT" will appear.
Step 7. Remove the “open” and connect the “short” precision cali-
bration component to the test port extension cable.
Press the ENTER key.
Step 8. The message "Measuring SHORT" will appear, and after the
measurement "Connect TERMINATION to RF OUT" will
appear.
Step 9. Remove the “short” and connect the “precision termination”
at the end of the test port extension cable. Press the ENTER
key.
Step 10.The message "Measuring TERMINATION" will appear. After
the measurement, the “CAL OFF” message will change to
"CAL ON" on the upper left-hand corner of the display.
Step 11.Remove the precision termination from the text port
extension cable.
Calculating Transmission Line Insertion Loss
Cables have different insertion losses at different frequencies.
For example LDF4-40A attenuation at 1 GHz is 0.022 dB/ft
(0.073 dB/m) and at 2 GHz it is 0.0325 dB/ft (0.107 dB/m).
As the frequency increases or the length of the cable run increases,
the amount of cable insertion loss increases.
To verify cable insertion loss measurements are reasonable, the
expected insertion loss can be calculated manually using the
following procedure:
• Calculate the estimated worst loss of each component in the
transmission line system.
• Add all the component’s estimated worst losses together to
calculate total insertion loss in the transmission line system.
Insertion loss of the = Bottom Jumper Loss
transmission system + Main Cable Loss
+ Top Jumper Loss
+ Connector Losses
= 0.408 + 2.3 +0.204 +1.12
= 4.03 dB
Compare the measured insertion loss to
the calculated insertion loss to verify
transmission line performance. The
measured cable insertion loss should be
lower than the calculated cable insertion loss.
Summary
The preferred method to measure
Cable Insertion Loss using Site Master
is Cable Loss mode. Cable loss mode
automatically considers the signal
traveling in both directions and thus
makes it easier to measure the cable
insertion loss in the field. Measured
Insertion Loss should always be compared to the calculated loss to verify
accuracy thus assuring transmission line
performance. The calculated insertion
loss is usually a “worst case scenario”.
Cable insertion loss may be difficult to
measure on excessivly long or highly
lossy cables. When the cable insertion
loss is greater than 20 dB, it will be hard
to measure.
All trademarks used are the property of their
respective owners.
Cable
Type
Cable
Attenuati
on (dB/ft)
Cable
Length
Insertion
Loss (dB)
Bottom
Jumper
LDF4-50A 0.0204 X 20 = 0.408
Main
Cable
LDF5-50A 0.0115 X 200 = 2.30
Top
Jumper
LDF4-50A 0.0204 X 10 = 0.204
Number of
Connector Pairs
Loss per Pair
in dB
Total Connector
Loss (dB)
4 X 0.28 = 1.12
Note: For Cable Loss-One Por t
measurements, Site Master S251C
requires only one port calibration.
Note: If the phase stable cable is
removed from the test port, the
calibration is not valid.
11410-00276 Rev. A, ©Anritsu March 2003, Data subject to change without notice
.