Analog Devices AN532 Application Notes

AN-532
a
APPLICATION NOTE
One Technology Way • P.O. Box 9106 • Norwood, MA 02062-9106 • 781/329-4700 • World Wide Web Site: http://www.analog.com
Applying Model 5B38 with 120 V Strain Gages
INTRODUCTION
The Model 5B38 isolated 10 V excitation voltage is rated to drive strain gage and load cell impedances as low as 300 . However, the 5B38 excitation voltage can easily be applied to drive a 120 bridge with the simple addi­tion of two resistors in series with the excitation supply, as shown in Figure 1. This approach is based on the un­derstanding that the 5B38 excitation voltage is a regu­lated supply voltage within its rated load from 10 k to 300 Ω. It also recognizes that 120 strain gages typically operate with an excitation less than 10 V to reduce ex­cessive internal power dissipation that would cause bridge self-heating effects with resulting thermal insta­bility during operation.
Procedure To Calculate Series Resistance:
1. Using the maximum load resistance rating of the 5B38 excitation voltage (300 ), and the load resis­tance of the strain gage (120 ), the total added series resistance is calculated as follows:
R
MAX LOAD
R
=
BRIDGE
300 –120 = 180
R
SERIES TOTAL
Strain Gage Excitation Voltage
The resulting bridge excitation voltage (e.g., the excita­tion voltage seen directly across the 120 bridge), is reduced from the 5B38 10 V to: [120 /300 ] × [10 V] =
4.0 V. Applications with low impedance bridges below 350 , such as the 120 strain gage in this example, typically work well with an excitation voltage of less than 10 volts. Note the power dissipation of a 120 bridge will decrease by a factor of six—from 830 mW to 130 mW—when the excitation voltage is reduced from 10 V to 4.0 V. The benefit is reduced thermal, or self­heating effects of the strain gage itself. However, it is always best to contact the strain gage manufacturer and verify that the 4.0 excitation voltage is acceptable for a specific strain gage or load cell model.
+EXC
HI
5B38
LO
–EXC
90V
90V
R
R
B
B
R
R
B
B
E3282–0–3/98
2. The calculated total series resistance is then divided in two equal value resistors—[180 /2 = 90]—each of which is placed in series with the 5B38 excitation supply terminals, +EXC and –EXC, and the 120 strain gage.
The calculated 180 series additional resistance is
split into two equal parts, to provide the strain gage with an excitation voltage that is balanced between the common or midpoint of the bridge voltage. This will eliminate common-mode voltage (CMV) error by establishing the bridge input CMV at 0 V.
With this arrangement, the maximum load resistance rating of the 5B38 excitation voltage is maintained: e.g., 90 + 90 + 120 = 300 .
Figure 1. Resist or Connections to 5B38 and 120 Ω Bridge
PRINTED IN U.S.A.
Loading...