One Technology Way • P.O. Box 9106 • Norwood, MA 02062-9106 • 781/329-4700 • World Wide Web Site: http://www.analog.com
Thermocouple Signal Conditioning Using the AD594/AD595
by Joe Marcin
INTRODUCTION
One of the most widely used devices for temperature
measurement is the thermocouple. Whether in an industrial, commercial or scientific application, a thermocouple offers a cost effective solution to temperature
measurements in many environments over wide temperature ranges. Unfortunately, their basic principles
are often misunderstood resulting in serious measurement errors. This application note will review thermocouple fundamentals and illustrate circuit designs for
thermocouple signal conditioning using the AD594/
AD595 monolithic IC.
BACK TO BASICS
The basic principles of the thermocouple were discovered in 1821 by Thomas Seebeck. When two dissimilar
metals are joined at both ends and one end is heated, a
current will flow. If the loop is broken at the center, an
open circuit voltage (the Seebeck Voltage) is generated
and is proportional to the difference in temperature between the two junctions. Therefore, in determining the
temperature of the measuring junction, the reference
junction temperature must be known.
V
– V
T1
T2
MEASURING
JUNCTION
T1T2
A
V
T1
B
A
V
T2
B
REFERENCE
JUNCTION
Figure 1a. Thermocouple Loop
An ice bath provides a well defined temperature of 0°C
for the reference junction. This has become a standard
reference point for the thermocouple output voltage vs.
temperature tables for various metal combinations.
V
(V
= 0)
T1
T2
T2 = 08C
V
T2
MEASURING
JUNCTION
T1
A
V
T1
B
A
B
These combinations have been characterized and classified by the National Institute of Standards and Technology (formerly the National Bureau of Standards). The
table below lists the types, composition and characteristics of the more commonly used thermocouples.
BPlatinum/Rhodium0°C to +1700°C0 to +12.426
EChromel/Constantan –200°C to +900°C–8.824 to +68.783
JIron/Constantan0°C to +750°C0 to +42.283
KChromel/Alumel–200°C to +1250°C–5.973 to +50.633
NNicrosil/Nisil–270°C to +1300°C–4.345 to +47.502
RPlatinum/Rhodium0°C to +1450°C0 to +16.741
Platinum
SPlatinum/Rhodium0°C to +1450°C0 to +14.973
Platinum
TCopper/Constantan–200°C to +350°C–5.602 to +17.816
Maximum
A voltmeter is commonly used to measure the Seebeck
voltage; however, great care must be exercised when interconnecting it to the thermocouple. Referring to Figure 1c, two additional junctions, J2 and J3, are formed at
the connection between the thermocouple and meter.
These two junctions produce opposing voltages within
the thermocouple loop. Using an isothermal block at the
point of connection keeps these junctions in thermal
equilibrium and produces equal but opposite emfs. The
measured voltage now is the difference in potential between the measuring junction and the isothermal block
which serves as the reference junction.
T2
V
A
V
B
B
(VT2 = VBA = V+VB)
Cu
Cu
– V
V
T1
T2
MEASURING
JUNCTION
T1
V
T1
Figure 1c. Measuring a Thermocouple Voltage with a
Voltmeter
Figure 1b. Ice Point Reference
AN-369
PRACTICAL THERMOCOUPLE MEASUREMENT
For most applications, it is impractical to use an ice bath
for the reference junction. By compensating for the voltage developed at the reference junction, the ice point
reference may be eliminated. This is performed by adding a voltage into the thermocouple loop, equal but opposite to that of the reference junction. A circuit that
provides cold junction compensation along with amplification and open thermocouple detection is included in
the AD594/AD595 family of thermocouple signal conditioning ICs.
T2
A
B
Cu
V
COMP
Cu
Cu
VT1 – VT2 + V
(V
– VT2)
COMP
COMP
MEASURING
JUNCTION
T1
V
T1
Figure 1d. Cold Junction Compensation
THE AD594/AD595 CIRCUIT DESCRIPTION
Figure 2 is a block diagram of the AD594/AD595 thermocouple signal conditioner IC. A Type J (for the AD594) or
Type K (for the AD595) thermocouple is connected to
Pins 1 and 14, the inputs to an instrumentation amplifier
differential stage. This input amplifier is contained in a
loop that uses the local temperature as its reference.
With the IC also at the local temperature, an ice point
compensation circuit develops a voltage equal to the deficiency in the locally referenced thermocouple loop.
This voltage is then applied to a second preamplifier
whose output is summed with the output of the input
amplifier. The resultant output is then applied to the input of a main output amplifier with feedback to set the
gain of the combined signals. The ice point compensation voltage is scaled to equal the voltage that would be
produced by an ice bath referenced thermocouple measuring the IC temperature. This voltage is then summed
with the locally referenced loop voltage, the result being
a loop voltage with respect to an ice point.
–IN–ALM+ALMV+COMPV
141312111098
OVERLOAD
DETECT
AD594/
AD595
+A
FB
O
Through the feedback path, the main amplifier maintains a balance at its inputs. In the event of a broken
thermocouple or open circuit at the device’s input, these
inputs become unbalanced, the fault is detected, and the
overload detection circuit drives a current limited n-p-n
transistor that may be interfaced as an alarm.
Although these ICs are specifically calibrated for a Type
J or K thermocouple, other thermocouple types may be
used with recalibration. Pin connections to internal
nodes for the temperature controlled voltages and feedback are provided to perform recalibration.
INTERPRETING AD594/AD595 OUTPUT VOLTAGES
To produce a temperature proportional output of
10 mV/°C, and provide an accurate reference junction
over the rated operating temperature range, the AD594/
AD595 is gain trimmed at the factory to match the transfer characteristics of Type J and K thermocouples at
+25°C. At this calibration temperature, the Seebeck coef-
ficient, the rate of change of thermal voltage with
respect to temperature at a given temperature, is
51.70 µV/°C for a Type J thermocouple and 40.44 µV/°C
for a Type K. This corresponds to a gain of 193.4 for the
AD594 and 247.3 for the AD595 to realize a 10 mV/°C out-
put. Although the device is trimmed for a 250 mV output
at +25°C, an input offset error is induced in the output
amplifier resulting in offsets of 16 µV and 11 µV for the
AD594/AD595 respectively. To determine the actual output voltage from the AD594/AD595, the following equations should be used:
AD594 Output
AD595 Output
= (
Type J Voltage
= (
Type K Voltage
+ 16 µV) × 193.4
+ 11 µV) × 247.3
where the Type J and K voltage are taken from the
thermocouple voltage tables referred to zero degrees
Celsius.
It is important to note that a thermocouple’s output is
linear over a narrow temperature range. Over a wide
temperature range, the Seebeck coefficient introduces
nonlinearity. Linearization is not provided by the AD594/
AD595, and any linearization techniques must be performed externally. This entails calculating thermocouple temperature using high order polynomials. The
National Institute of Standards and Technology offers
tables of polynomial coefficients for a given thermocouple type which may be used in this process.
G
1234567
+IN+C+TCOMV––T–C
G
+TC
ICE
POINT
COMP.
–TC
Figure 2. AD594/AD595 Functional Block Diagram
–2–
Table II. Calculated Errors at Various Ambient Temperatures
NOTE
Temp. Rej. Error has two components: (a) Difference between actual reference junction and ice point compensation voltage times the gain; (b) Offset and
gain TCs extrapolated from 0°C to +50°C limits. Total error is temp. rej. plus initial calibration error.
OPTIMIZING PERFORMANCE
Cold Junction Errors
Optimal performance from the AD594/AD595 is
achieved when the thermocouple cold junction and the
device are at thermal equilibrium. Avoid placing heat
generating devices or components near the AD594/
AD595 as this may produce cold junction related errors.
The ambient temperature range for the AD594/AD595 is
specified from 0°C to +50°C, and its cold junction com-
pensation voltage is matched to the best straight line fit
of the thermocouple’s output within this range. Operation outside this range will result in additional error.
Table II shows the maximum calculated errors at various
ambient temperatures.
Circuit Board Layout
The circuit board layout shown in Figure 3 (with the optional calibration resistors) achieves thermal equilibrium between the cold junction and the AD594/AD595.
The package temperature and circuit board are thermally contacted in the copper printed circuit board
tracks under Pins 1 and 14. The reference junction is now
composed of a copper-constantan (or copper-alumel)
connection and copper-iron (or copper-chromel) connection in thermal equilibrium with the IC.
(CHROMEL)
+C+T
IRON
+IN–IN
1
CONSTANTAN
(ALUMEL)
14
–ALM
+ALM
Soldering
Proper soldering techniques and surface preparation
are necessary to bond the thermocouple to the PC
tracks. Clean the thermocouple wire to remove oxidation before soldering. Noncorrosive rosin flux may be
used with the following solders: 95% tin-5% antimony,
95% tin-5% silver, or 90% tin-10% lead.
Bias Current Return
The input instrumentation amplifier of the AD594/AD595
requires a return path for its input bias current and may
not be left “floating.” If the thermocouple measuring
junction is electrically isolated, then Pin 1 of the IC
should be connected to Pin 4, the power supply common. In some applications, tying the thermocouple directly to common is not possible. A resistor from Pin 1 to
common will satisfy the bias current return path but will,
however, generate an additional input offset voltage
due to the 100 nA bias current flowing through it. If the
thermocouple must be grounded at the measuring junction or if a small common mode potential is present, do
not make the connection between Pins 1 and 4.
Noise Suppression
When detecting a low level output voltage from a thermocouple, noise reduction is a prime concern. Whether
internally generated or induced by radiation from a
source, noise becomes one of the limiting factors of dynamic range and resolution. Solving noise problems involves eliminating the source and/or shielding. The
latter is more effective when the source cannot be controlled or identified.
Noise may be injected into the AD594/AD595 input amplifier when using a long length of thermocouple. To
determine if this noise path is the culprit, disconnect the
COMMON
AD594
AD595
–T– CV–V
Figure 3. PC Board Layout
OUT
87
COMP
V+
thermocouple from the AD594/AD595 and tie Pins 1 and
14 to Pin 4. The output voltage at Pin 9 of the AD594/
AD595 will now indicate ambient temperature (250 mV
at +25°C). If the noise at the output (Pin 9) disappears,
then shielding on the input is required. Shielded thermocouple wire with the shield connected to Pin 4 of the
IC will provide effective noise suppression. If the output
–3–
Loading...
+ 4 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.