Analog Devices AN369 Application Notes

AN-369
a
APPLICATION NOTE
One Technology Way • P.O. Box 9106 • Norwood, MA 02062-9106 • 781/329-4700 • World Wide Web Site: http://www.analog.com
Thermocouple Signal Conditioning Using the AD594/AD595
by Joe Marcin
INTRODUCTION
One of the most widely used devices for temperature measurement is the thermocouple. Whether in an indus­trial, commercial or scientific application, a thermo­couple offers a cost effective solution to temperature measurements in many environments over wide tem­perature ranges. Unfortunately, their basic principles are often misunderstood resulting in serious measure­ment errors. This application note will review thermo­couple fundamentals and illustrate circuit designs for thermocouple signal conditioning using the AD594/ AD595 monolithic IC.
BACK TO BASICS
The basic principles of the thermocouple were discov­ered in 1821 by Thomas Seebeck. When two dissimilar metals are joined at both ends and one end is heated, a current will flow. If the loop is broken at the center, an open circuit voltage (the Seebeck Voltage) is generated and is proportional to the difference in temperature be­tween the two junctions. Therefore, in determining the temperature of the measuring junction, the reference junction temperature must be known.
V
– V
T1
T2
MEASURING
JUNCTION
T1 T2
A
V
T1
B
A
V
T2
B
REFERENCE
JUNCTION
Figure 1a. Thermocouple Loop
An ice bath provides a well defined temperature of 0°C
for the reference junction. This has become a standard reference point for the thermocouple output voltage vs. temperature tables for various metal combinations.
V
(V
= 0)
T1
T2
T2 = 08C
V
T2
MEASURING
JUNCTION
T1
A
V
T1
B
A
B
These combinations have been characterized and classi­fied by the National Institute of Standards and Technol­ogy (formerly the National Bureau of Standards). The table below lists the types, composition and characteris­tics of the more commonly used thermocouples.
Table I. Thermocouple Properties
ANSI Alloy Temperature mV Code Combination Range Output
B Platinum/Rhodium 0°C to +1700°C 0 to +12.426 E Chromel/Constantan –200°C to +900°C –8.824 to +68.783 J Iron/Constantan 0°C to +750°C 0 to +42.283 K Chromel/Alumel –200°C to +1250°C –5.973 to +50.633 N Nicrosil/Nisil –270°C to +1300°C –4.345 to +47.502 R Platinum/Rhodium 0°C to +1450°C 0 to +16.741
Platinum
S Platinum/Rhodium 0°C to +1450°C 0 to +14.973
Platinum
T Copper/Constantan –200°C to +350°C –5.602 to +17.816
Maximum
A voltmeter is commonly used to measure the Seebeck voltage; however, great care must be exercised when in­terconnecting it to the thermocouple. Referring to Fig­ure 1c, two additional junctions, J2 and J3, are formed at the connection between the thermocouple and meter. These two junctions produce opposing voltages within the thermocouple loop. Using an isothermal block at the point of connection keeps these junctions in thermal equilibrium and produces equal but opposite emfs. The measured voltage now is the difference in potential be­tween the measuring junction and the isothermal block which serves as the reference junction.
T2
V
A
V
B
B
(VT2 = VBA = V+VB)
Cu
Cu
– V
V
T1
T2
MEASURING
JUNCTION
T1
V
T1
Figure 1c. Measuring a Thermocouple Voltage with a Voltmeter
Figure 1b. Ice Point Reference
AN-369
PRACTICAL THERMOCOUPLE MEASUREMENT
For most applications, it is impractical to use an ice bath for the reference junction. By compensating for the volt­age developed at the reference junction, the ice point reference may be eliminated. This is performed by add­ing a voltage into the thermocouple loop, equal but op­posite to that of the reference junction. A circuit that provides cold junction compensation along with ampli­fication and open thermocouple detection is included in the AD594/AD595 family of thermocouple signal condi­tioning ICs.
T2 A B
Cu
V
COMP
Cu Cu
VT1 – VT2 + V (V
– VT2)
COMP
COMP
MEASURING
JUNCTION
T1
V
T1
Figure 1d. Cold Junction Compensation
THE AD594/AD595 CIRCUIT DESCRIPTION
Figure 2 is a block diagram of the AD594/AD595 thermo­couple signal conditioner IC. A Type J (for the AD594) or Type K (for the AD595) thermocouple is connected to Pins 1 and 14, the inputs to an instrumentation amplifier differential stage. This input amplifier is contained in a loop that uses the local temperature as its reference. With the IC also at the local temperature, an ice point compensation circuit develops a voltage equal to the de­ficiency in the locally referenced thermocouple loop. This voltage is then applied to a second preamplifier whose output is summed with the output of the input amplifier. The resultant output is then applied to the in­put of a main output amplifier with feedback to set the gain of the combined signals. The ice point compensa­tion voltage is scaled to equal the voltage that would be produced by an ice bath referenced thermocouple mea­suring the IC temperature. This voltage is then summed with the locally referenced loop voltage, the result being a loop voltage with respect to an ice point.
–IN –ALM +ALM V+ COMP V
14 13 12 11 10 9 8
OVERLOAD
DETECT
AD594/ AD595
+A
FB
O
Through the feedback path, the main amplifier main­tains a balance at its inputs. In the event of a broken thermocouple or open circuit at the device’s input, these inputs become unbalanced, the fault is detected, and the overload detection circuit drives a current limited n-p-n transistor that may be interfaced as an alarm.
Although these ICs are specifically calibrated for a Type J or K thermocouple, other thermocouple types may be used with recalibration. Pin connections to internal nodes for the temperature controlled voltages and feed­back are provided to perform recalibration.
INTERPRETING AD594/AD595 OUTPUT VOLTAGES
To produce a temperature proportional output of
10 mV/°C, and provide an accurate reference junction
over the rated operating temperature range, the AD594/ AD595 is gain trimmed at the factory to match the trans­fer characteristics of Type J and K thermocouples at
+25°C. At this calibration temperature, the Seebeck coef-
ficient, the rate of change of thermal voltage with respect to temperature at a given temperature, is
51.70 µV/°C for a Type J thermocouple and 40.44 µV/°C
for a Type K. This corresponds to a gain of 193.4 for the
AD594 and 247.3 for the AD595 to realize a 10 mV/°C out-
put. Although the device is trimmed for a 250 mV output
at +25°C, an input offset error is induced in the output amplifier resulting in offsets of 16 µV and 11 µV for the
AD594/AD595 respectively. To determine the actual out­put voltage from the AD594/AD595, the following equa­tions should be used:
AD594 Output
AD595 Output
= (
Type J Voltage
= (
Type K Voltage
+ 16 µV) × 193.4 + 11 µV) × 247.3
where the Type J and K voltage are taken from the thermocouple voltage tables referred to zero degrees Celsius.
It is important to note that a thermocouple’s output is linear over a narrow temperature range. Over a wide temperature range, the Seebeck coefficient introduces nonlinearity. Linearization is not provided by the AD594/ AD595, and any linearization techniques must be per­formed externally. This entails calculating thermo­couple temperature using high order polynomials. The National Institute of Standards and Technology offers tables of polynomial coefficients for a given thermo­couple type which may be used in this process.
G
1234567
+IN +C +T COM V––T –C
G
+TC
ICE POINT COMP.
–TC
Figure 2. AD594/AD595 Functional Block Diagram
–2–
Table II. Calculated Errors at Various Ambient Temperatures
AN-369
Ambient Temp. Rej. Total Temp. Rej. Total Temp. Rej. Total Temp. Rej. Total
AD594C AD594C AD594A AD594A AD595C AD595C AD595A AD595A
Temp. Error Error Error Error Error Error Error Error C C C C C C C C C
–55 4.83 5.83 6.83 9.83 5.28 6.28 7.28 10.28 –25 1.98 2.98 3.23 6.23 2.04 3.04 3.29 6.29 0 0.62 1.62 1.25 4.25 0.62 1.62 1.25 4.25 +25 0.00 1.00 0.00 3.00 0.00 1.00 0.00 3.00 +50 0.62 1.62 1.25 4.25 0.62 1.62 1.25 4.25 +70 1.46 2.46 2.59 5.59 1.38 2.38 2.50 5.50 +85 2.25 3.25 3.75 6.75 1.99 2.99 3.49 6.49 +125 4.90 5.90 7.40 10.40 3.38 4.38 5.88 8.88
NOTE Temp. Rej. Error has two components: (a) Difference between actual reference junction and ice point compensation voltage times the gain; (b) Offset and gain TCs extrapolated from 0°C to +50°C limits. Total error is temp. rej. plus initial calibration error.
OPTIMIZING PERFORMANCE Cold Junction Errors
Optimal performance from the AD594/AD595 is achieved when the thermocouple cold junction and the device are at thermal equilibrium. Avoid placing heat generating devices or components near the AD594/ AD595 as this may produce cold junction related errors. The ambient temperature range for the AD594/AD595 is
specified from 0°C to +50°C, and its cold junction com-
pensation voltage is matched to the best straight line fit of the thermocouple’s output within this range. Opera­tion outside this range will result in additional error. Table II shows the maximum calculated errors at various ambient temperatures.
Circuit Board Layout
The circuit board layout shown in Figure 3 (with the op­tional calibration resistors) achieves thermal equilib­rium between the cold junction and the AD594/AD595. The package temperature and circuit board are ther­mally contacted in the copper printed circuit board tracks under Pins 1 and 14. The reference junction is now composed of a copper-constantan (or copper-alumel) connection and copper-iron (or copper-chromel) con­nection in thermal equilibrium with the IC.
(CHROMEL)
+C+T
IRON
+IN –IN 1
CONSTANTAN
(ALUMEL)
14
–ALM
+ALM
Soldering
Proper soldering techniques and surface preparation are necessary to bond the thermocouple to the PC tracks. Clean the thermocouple wire to remove oxida­tion before soldering. Noncorrosive rosin flux may be used with the following solders: 95% tin-5% antimony, 95% tin-5% silver, or 90% tin-10% lead.
Bias Current Return
The input instrumentation amplifier of the AD594/AD595 requires a return path for its input bias current and may not be left “floating.” If the thermocouple measuring junction is electrically isolated, then Pin 1 of the IC should be connected to Pin 4, the power supply com­mon. In some applications, tying the thermocouple di­rectly to common is not possible. A resistor from Pin 1 to common will satisfy the bias current return path but will, however, generate an additional input offset voltage due to the 100 nA bias current flowing through it. If the thermocouple must be grounded at the measuring junc­tion or if a small common mode potential is present, do not make the connection between Pins 1 and 4.
Noise Suppression
When detecting a low level output voltage from a ther­mocouple, noise reduction is a prime concern. Whether internally generated or induced by radiation from a source, noise becomes one of the limiting factors of dy­namic range and resolution. Solving noise problems in­volves eliminating the source and/or shielding. The latter is more effective when the source cannot be con­trolled or identified.
Noise may be injected into the AD594/AD595 input am­plifier when using a long length of thermocouple. To determine if this noise path is the culprit, disconnect the
COMMON
AD594 AD595
–T – C V– V
Figure 3. PC Board Layout
OUT
87
COMP
V+
thermocouple from the AD594/AD595 and tie Pins 1 and 14 to Pin 4. The output voltage at Pin 9 of the AD594/ AD595 will now indicate ambient temperature (250 mV
at +25°C). If the noise at the output (Pin 9) disappears,
then shielding on the input is required. Shielded ther­mocouple wire with the shield connected to Pin 4 of the IC will provide effective noise suppression. If the output
–3–
Loading...
+ 4 hidden pages