
Data Sheet
THWD-10W
RH Meter with Wireless
Perform precision temperature and humidity measurement for 
HVAC, Indoor Air Quality and environmental applications. The 
large LCD display, full suite of measuring features and two 
temperature probes (internal and external) make this product a 
preferred choice for a professional.
 Wireless two ways transmission 
 Highly stable humidity measurement ±2.5%
 Processing relative humidity and temperature measurements 
  simultaneously
 Two temperature sensors (internal and external)
 Wet/dry bulb, dew point measurement
 Easy selection between K/NTC/K-DP
No hassle 
warranty
No waiting. 
No shipping  
charges.
Our commitment to 
high-quality products 
and customer service 
is demonstrated by 
our industry exclusive  
“No Hassle” warranty.  
In the unlikely event 
that an Amprobe Test 
Tool requires warranty 
service, any of our 
local dealers are 
authorized to replace 
it, on the spot.
(note: $500 MSLP limit)
 MAX/MIN/Average/Relative Zero and HOLD functions
Amprobe® Test Tools 
www.Amprobe.com
 

THWD-10W RH Meter with Wireless
General Specifications
Display  5 digit liquid crystal display (LCD). 
Polarity  Automatic, positive implied, negative polarity indication. 
Low Battery Indication  The “ 
Operating Environment  0°C to 50°C (32°F to 122°F) <80% R.H.  
Auto Power Off  5 minutes 
Storage Environment  -20°C to 60°C (-4°F to 140°F) <70% R.H. with battery removed from meter 
Accuracy  Stated accuracy at 23°C ±5°C, <75% R.H. 
Battery  1.5Vx4pcs AAA size. 
Battery Life  200 hours typical 
Dimensions  L241.3 X W157.98 XH76.2mm (9.5 X 6.22 X3 in). 
Weight  Approx. 304g (0.67 lb) including batteries. 
WIRELESS Features 
Frequency Range  910~920MHz. 
     Low current consumption less than 1mA. 
     The transmitting distance can reach 25M without magnetic interference.
” is displayed when the battery voltage drops below the operating level.
Data Sheet
Amprobe® Test Tools 
www.Amprobe.com
 

THWD-10W RH Meter with Wireless
Data Sheet
Electrical Specifications
Temperature Scale   Celsius or Fahrenheit user-selectable  
Measurement Range  K-Type -50°C to 1000°C, (-50°F to 1832°F)  
Resolution  0.1°C or 0.2°F  
Accuracy   Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1  
  year, not including thermocouple error. ±(0.1% rdg + 1°C) on -50°C to 1000°C ±(0.1% rdg + 2°F) on 
  -50°F to 1832°F ±(0.05% rdg + 1.4°F) -58°F to -328°F  
Sensor   Thermistor temperature sensor  
Range   0°C to 60°C, (32°F to 140°F)  
Resolution   0.1°C/°F  
Accuracy   ±2°C on 0°C to 10°C ±0.5°C on 10°C to 45°C ±2°C on 45°C to 60°C ±4°F on 32°F to 50°F ±1°F on 50°F to 
  113°F ±4°F on 113°F to 140°F  
RELATIVE HUMIDITY 
Sensor   Capacitive Humidity Sensor  
Range   0% to 100% RH  
Accuracy   ±2.5% at 25°C (77°F), 10% to 90% RH,  
  ±5% at 25°C (77°F), 0% to 10% RH,  
  90% to 100% RH  
Sensor Response Time      60 sec typical    
for 90% of Total Range               
Sensor Hysteresis   ±1%RH typical 
(excursion of 10% to 90% to 10% RH)
Temperature   0.1 times the applicable accuracy  
Coefficient   Specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).  
Input Protection   24V dc or 24V ac rms maximum input Voltage on any combination of input pins.  
Input Connector   Accepts standard miniature thermocouple connectors (flat blades spaced 7.9mm, center to center). 
Amprobe® Test Tools 
www.Amprobe.com
 

THWD-10W RH Meter with Wireless
Data Sheet
Amprobe® Test Tools
website: www.Amprobe.com 
email: info@amprobe.com 
Everett, WA 98203 
Tel: 877-AMPROBE
Amprobe® Test Tools 
Amprobe® Test Tools Europe
In den Engematten 14 
79286 Glottertal, Germany 
Tel.: +49 (0) 7684 8009 - 0
©2010 Amprobe Test Tools . All righ ts reserved.
6/2010 TH WD-10W_ DS Rev A
www.Amprobe.com