Amprobe RTD-10 Data Sheet

Data Sheet
RTD-10
RTD Thermometer
The RTD probes are preferred for industrial process applications. They typically provide higher accuracy then thermocouples and maintain stability for many years.
RTD dual input function
Pt3916/Pt3926/Pt385 RTD standard
°C/°F user-selectable
MAX/MIN with elapsed time, AVG, REL, HOLD functions
Warning beeper with Hi/Lo setting
Resolution 0.1°C to 0.2°F
No hassle warranty
No waiting.
No shipping charges.
Our commitment to high-quality products and customer service is demonstrated by our industry exclusive “No Hassle” warranty. In the unlikely event that an Amprobe Test Tool requires warranty service, any of our local dealers are authorized to replace it, on the spot.
(note: $500 MSLP limit)
Amprobe® Test Tools
RTD-10 RTD Thermometer
Data Sheet
General Specifications
Display 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Polarity Automatic, positive implied, negative polarity indication. Low Battery Indication The “ Measurement Rate 1 times/second. Auto Power Off 30 minutes Operating Environment 0°C to 50°C (32°F to 122°F) at <70% R.H. Storage Environment -20°C to 60°C (-4°F to 140°F), 0 to 80% R.H. with battery removed from meter. Accuracy Stated accuracy at 23°C ±5°C, <75% R.H. Battery 1.5V x 4pcs AAA size Dimensions L241.3 X W158 X 76.2mm (9.5 X 6.22 X 3 in) Battery Life 200 Hours typical Weight Approx. 285.76g (0.63 lb) including batteries and packaging box.
” is displayed when the battery voltage drops below the operating level.
Electrical Specifications
Temperature Scale Celsius or Fahrenheit user-selectable Measurement Range Pt385 (100Ω) -200°C to 800°C (-328°F to 1472°F) Pt3916/Pt3926 (100Ω) -200°C to 630°C
(-328°F to 1166°F)
Resolution 0.1°C or 0.2°F Accuracy Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1
year, not including thermocouple error. ±(0.05%rdg+0.2°C) on °C scale ±(0.05%rdg+0.4°F) on °F scale
Temperature 0.1 times the applicable accuracy Coefficient Specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F). Input Protection 24V dc or 24V ac rms maximum input Voltage on any combination of input pins. Input Accepts for 3 wires or 4 wires Mini-Din Connector Plug and the meter is designed for standard. The other probes 2.3 wires will cause measuring
tolerance due to cable resistance. Included Accessories RTD Connectors (2); users manual; batteries (4)
Amprobe® Test Tools
©2010 Amprobe Test Tools . All righ ts reserved.
6/2010 RTD-10_DS Rev A