Amprobe MAN30 User manual

MAN02 MAN15 MAN30 Manometer
Users Manual
• Mode d’emploi
• Bedienungshandbuch
• Manual d’Uso
• Manual de uso
Find Quality Products Online at: sales@GlobalTestSupply.com
English
MAN02 MAN15 MAN30 Manometer
Users Manual
MAN02_15_30 Rev001 © 2009 Amprobe Test Tools. All rights reserved.
Find Quality Products Online at: sales@GlobalTestSupply.com
3
Limited Warranty and Limitation of Liability
Your Amprob e produc t will be free from de fects in material a nd workma nship for 1 y ear from the date of purchase. T his warrant y does not cover fu ses, di sposab le batte ries or damage from a ccident , neglec t, misuse, alt eration, c ontamination, or ab normal co ndition s of operati on or handli ng. Resell ers are no t authorize d to exten d any other w arranty o n Amprob e’s behalf. To obtain ser vice during th e warrant y period , return th e produc t with pro of of purcha se to an auth orized Amp robe Test Tools Ser vice Center or to an A mprobe d ealer or di stribu tor. See Repai r Sectio n for deta ils. THIS WAR RANT Y IS YOUR ONLY REMEDY. ALL OTHER WAR RANT IES - WHE THER EXP RESS, IM PLIED OR STAUTORY - I NCLUDIN G IMPLIE D WARRANTIES OF FI TNESS FO R A PARTICU LAR PU RPOSE OR MERCHA NTABILI TY, ARE HEREBY D ISCL AIMED. MA NUFACTU RER SHAL L NOT BE LIA BLE FOR AN Y SPECIA L, INDI RECT, INC IDENTAL OR CO NSEQUEN TIAL DAM AGES OR LOSS ES, ARI SING FROM A NY CAUSE O R THEORY. Since so me stat es or count ries do not a llow the excl usion or lim itation o f an implied warranty o r of incide ntal or con sequential damag es, this limitatio n of liabilit y may not apply to you.
Repair
All test t ools retu rned for wa rranty o r non-wa rranty re pair or for c alibratio n should be a ccompan ied by the foll owing: your n ame, company’s nam e, addre ss, tele phone number, and pro of of purcha se. Additio nally, plea se includ e a brief des criptio n of the prob lem or the se rvice re quest ed and incl ude the tes t leads wit h the meter. Non-warra nty repair or repla cement ch arges sh ould be remitted in th e form of a check, a mon ey order, cre dit card wi th expirat ion date, or a purchas e order mad e payable to Ampro be® Test Tools .
In-Warranty Re pairs an d Replac ement – All Countri es
Please r ead the warr anty st atement a nd check yo ur batte ry befo re reque sting rep air. During the warrant y period a ny defec tive tes t tool can be r eturned t o your Ampr obe® Test Tool s distrib utor for an exch ange for th e same or like p roduc t. Pleas e check the “ Where to B uy” sec tion on ww w.
Warrant y repair an d replace ment unit s can also b e sent to a Am probe® Tes t Tools Ser vice Cente r (see addre ss below).
Non-Warranty R epairs a nd Repla cement – US and Cana da
Non- warrant y repairs i n the United S tates and Canada sh ould be se nt to a Ampro be® Test Tools Serv ice Cente r. Call Amprob e® Test Tools or i nquire at yo ur point of purchase for current r epair and replacement rates.
Non-Warranty R epairs a nd Repla cement – Europe
Europea n non-wa rranty un its can be replace d by your Amp robe® Test Tool s distrib utor for a no minal
*(Corres ponde nce only – no re pair or rep lacemen t available f rom this ad dress. European cu stome rs please c ontac t your dist ributor.)
Find Quality Products Online at: sales@GlobalTestSupply.com
4
➊ ON/OFF Pushbutton
Record pushbutton
Data Hold pushbutton
➍
Differential pushbutton
Unit Selector pushbutton
Backlight pushbutton
➊
Primary data screen displays pressure value
“-” Minus pressure display
MAX MIN pressure recorded
➍
REC starts recording mode and displays
max./min. pressure recorded
➎DC power input jack
➏
USB output port (USB cable not included)
➐
H/M/S 88:88:88 displays time in Hour/ Minute/Second
➑
Pressure unit indication
➒
BAT Low battery indicator
DIF Differential pressure mode
“+” Positive pressure connection input
“-” Negative pressure connection input
Data Hold freezes pressure reading
Indicator
HOLD DIF
DAT A LOGGER
REC MAX MIN BAT
inH O psi mbar bar mmH2O
2
MANOMETER
1
2
3
4
5
6
7
8
9
1 1
1 0
1 2
1 3
HOLD DIF
DAT A LOGGER
REC MAX MIN BAT
inH O psi mbar bar mmH2O
2
MANOMETER
1
2
3
4
5
6
Find Quality Products Online at: sales@GlobalTestSupply.com
1
MAN02 / MAN15 / MAN30 Manometer
CONTENTS
INTRODUCTION ...................................................................................................................................2
FEATURES .............................................................................................................................................2
STANDARD ACCESSORIES ...................................................................................................................2
QUICK START .......................................................................................................................................3
AUTO POWER OFF ...............................................................................................................................3
OPERATION MODE ..............................................................................................................................3
CALIBRATION MODE ...........................................................................................................................4
Calibration point reference ...........................................................................................................5
MANUAL ZERO SETTING .....................................................................................................................5
TROUBLE SHOOTING ...........................................................................................................................5
Replacing The Battery ...................................................................................................................6
USB PC INTERFACE CAPABILITIES .......................................................................................................6
Download Suite Software installation .........................................................................................6
Operation .......................................................................................................................................6
MAINTENANCE ....................................................................................................................................7
Cleaning .........................................................................................................................................7
SPECIFICATIONS ...................................................................................................................................7
Operating Conditions ....................................................................................................................7
Find Quality Products Online at: sales@GlobalTestSupply.com
2
Congratulations on your purchase of the Amprobe MAN02 MAN15 MAN30 manometers. These instruments are portable, battery operated pressure measuring device. They are ideal for HVAC/R technicians measuring pressure level, Medical equipment, Computer peripherals, pneumatic controls.
INTRODUCTION
The meter will display all LCD segments when it is first turned on for approx. 3 seconds. •
The LCD is divided into two distinct sections : One large (primary) top screen and one smaller • right bottom screen (relative Clock). The two display areas will constantly update with the pressure measurements.
FEATURES
The meter measures:
Gauge pressure - • a measurement of pressure that is referred to ambient pressure.
Differential pressure • - a measurement of difference of two pressures .
5 selectable units of measure: in H2o, psi, mbar, bar, mmH2o. (For MAN02)•
11 selectable units of measure: bar, mmHg, ozin2, Kgcm2, psi, inH2O, kpa, ftH2O, inHg, • cmH2O, mbar (For MAN15 and MAN30)Standard Accessories.
Please check that the tubing is not leaking or damaged before use.•
STANDARD ACCESSORIES
The package contains:
Manometer•
One 9V Battery•
Users manual•
2 connection hoses 4mm(ID) x 6mm(OD) x 500mm•
Carrying Case•
Find Quality Products Online at: sales@GlobalTestSupply.com
3
QUICK START
Unscrew battery compartment on the rear of the instrument and insert the battery. Replace cover and secure with screw.
Press 1. to switch the instrument on.
Press 2.
to select unit of pressure measurement required. For zeroing, press for
three seconds.
Press 3.
for differential pressure measurement.
Press 4.
to freeze the reading on the display. Press again to cancel feature.
Press 5.
to start a recording; press again to display MAX value of the recording session; press again to display MIN value of the recording session; and press again to return to real time recording mode.
The Instrument records only the Max and the Min values for each session. To record all
values for a recording session, use the Download Suite Software.
Press
and hold for 3 seconds to turn clock feature off.
Note: Clock feature available with gauge pressure only, not differential. The instrument will
automatically switch off after 20 minutes unless sleep mode is disabled.
Press 6. to turn ON the back-light. It automatically switches off after 30 seconds.
AUTO POWER OFF
(Sleep mode function)
This instrument will turn off automatically in approx. 20 minutes after power-on if there are no key activations. For recording or operating over longer periods of time, you can disable the sleep mode by pressing and
simultaneously before powering on.
An “n” will appear in the middle of the screen at which time you can release the button. (See Fig. 1) The disable sleep mode will be invalid after power off.
OPERATION MODE
(See Fig. 18.)
Turns instrument on and off.1.
Press momentarily and relative clock starts in the lower right screen.2.
REC is displayed in the middle left of screen (Fig.2).The other functions are locked out except
Power and Backlight.
Press momentarily again and the unit cycles through MAX (Fig. 3) and MIN (Fig.4) and back
to current pressure reading; the record mode is displayed on the LCD.
Press and hold for 3 seconds to turn off the record function and return to normal mode.
Press momentarily to freeze the pressure reading (Fig. 5).3.
Press momentarily, 4. DIF appears on top of the LCD and the display indicates the relative zero (relative zero causes the value of the display to show as “0.0”) - only the amount of pressure change will be indicated. Press momentarily again and the unit returns to the normal mode of pressure differential (see Fig.6).
Find Quality Products Online at: sales@GlobalTestSupply.com
4
Differential Pressure: a measurement of the difference between two pressures, i.e. use differential pressure sensor to measure gauge pressure by leaving one process connection open to atmosphere and connecting the second sensor port to your system.
Press momentarily and the units will cycle through 5. “InH2o”,”psi”, “mbar”, “ 5. bar”,“mmH2o” which are indicated on the bottom of the display. (For MAN02)
Press momentarily and the units will cycle through bar, mmHg, ozin2, Kgcm2, psi, inH2O, kpa, 6. ftH2O, inHg, cmH2O, mbar which are indicated on the bottom of the display (See fig.7 & 8). (For MAN15 and MAN30)
Press momentarily and the backlight illuminates for approx. 30 seconds, then turns off 7. automatically.
Press momentarily to decrease the figure when calibration is being performed.
CALIBRATION MODE
Calibration mode is only applicable for a standard manometer calibrator or any qualified meter calibration facility for annual calibration.
First, manually set the display to zero (no pressure applied to the connector ), refer to the 1. manual zero procedure.
Turn the meter off.2.
Press 3. &
simultaneously, “CA” appears on the display, (See Fig. 9 ) the meter enters to the calibration mode, make sure the units setting on “PSI” to start positive (+) pressure calibration.
The meter has defaulted as 1.6 psi (for MAN02); 12.00 psi (for MAN15); 24.00 psi (for MAN30) 4. calibration point, the adjustable pressure range is from 1.5 to 1.7 (for MAN02); 11.70 to
12.30 (for MAN15); 23.40 to 24.60 (for MAN30). If calibration pressure source is not 12.00 psi, increase by pressing
key, or decrease by pressing L key to set calibration point as
required.
Save the calibration point by pressing 5.
key, “SA” and small “CA” appears on the display in 2 seconds (See Fig. 10). The meter auto-skips to the negative pressure (-) point for next calibration mode.
Follow the same procedure as step 4 for the negative pressure calibration point.6.
The LCD now displays” -12.00 “ and small”CA” (See Fig.11 ). Perform the necessary calibration
figure. Refer to your pressure standard if needed.
Again save the calibration point by pressing 7.
key, “SA” and “CA” appears in 2 seconds
and then press
;”End”and “CA” appear in another 2 seconds. The meter turns back to
the normal mode (See Fig. 12 ).
If you can’t save by pressing
key, i.e. no “SA” appeared, please check the followings:
 •Thecalibrationpressuresourcecouldbereferredtothefollowingtable.
 •ifyouenteredtherightpositivepressure(+)ornegativepressure(-).
Find Quality Products Online at: sales@GlobalTestSupply.com
5
If you want to skip positive (+) calibration when in the calibration mode, press key to skip to negative (-) calibration point.
Calibration point reference
Model Psi Range Calibration point (+/-) Recommended (+/-)
MAN02 0~+/-2 1.6 1.5 t ~ 1.7
MAN15 0~+/-15 12.00 11.70 ~ 12.30
MAN30 0~+/-30 24.00 23.40 ~ 24.60
MANUAL ZERO SETTING
When you set the display to zero( no pressure applied to the connector), press the
key for 2 seconds. The meter should display “- 0 .000” from right to left (See Fig.13). Then, the LCD display shows normal mode.
TROUBLE SHOOTING
Cannot Power on.
Check the battery connections. Replace the battery or attach Optional AC adaptor.
Low BAT indication.
Replace the battery when LCD displays BAT.
No Display.
Make sure the battery is functional.• check if sleep mode is active. Refer to disable sleep mode on page.6 to turn it off. • check if the tubing is connected to the meter tightly.•
Err.1.
For the pressure value exceeding the maximum range, “ Err.1” appears on the display (See Fig.
14). Do not exceed rated over pressure range of manometer. Sensor may be damaged.
Err.2.
For the measurement pressure less than minimum range, “Err. 2” will appear (See Fig. 15).
Err.3.
For a differential pressure value larger than maximum display, “Err.3” appears on the display (See Fig.16 ).
Err.4.
When you set zero, make sure you have disconnected the tubing. If you see an “Err.4” appear on the display, it means the manometer is damaged (See Fig.17)
Note. Err.4 Will be also appear If the tubing is connected during zero set.
Find Quality Products Online at: sales@GlobalTestSupply.com
6
E1OL or E2UL.
When you see these errors while operating computer software, it means pressure source is less or over than the range of the instrument.
Replace your 9-volt battery when:
The • BAT icon appears on the right of the screen.
The meter will not power on.•
Use of the back-light causes the • BAT icon to appear.
Replacing The Battery
Even if the battery was recently replaced, check its voltage level if you get no response from your instrument.
To replace the battery:
Remove the tubing from the instrument.1.
Lay the instrument face-down on a clean, flat surface.2.
Remove battery cover.3.
Remove battery from instruments that you do not plan to use for a month or more.
Do not leave battery in instrument.
USB PC INTERFACE CAPABILITIES
The USB cable and the Download Suite software are required to transfer data to a PC. The USB port is located on the right side of the instrument. The USB Cable is not included. It can be purchased separately as an optional accessory.
Download Suite Software installation
Insert the Download Suite CD into the CD-ROM drive.
To install the software, follow the on-screen instructions.
Operation
To open the program, double-click the Download Suite icon.
To transfer data to a PC, follow the on-screen instructions.
Find Quality Products Online at: sales@GlobalTestSupply.com
7
MAINTENANCE
The meter is calibrated in house before shipping. When properly maintained, the meter will maintain its accuracy as specified.
Cleaning
Use a damp cloth and mild soap to clean the case of the MAN15 Manometer, do not use harsh detergents or abrasives as these may mar the finish or damage the unit’s case with an adverse chemical reaction.
SPECIFICATIONS
Operating Conditions
Compensated temperature range: 0~50 °C (32~122 °F)•
Operating temperature range: 0~50 °C (32~122 °F)•
Storage temperature range: -20~60 °C (-4~140 °F)•
Operating Humidity Max. 80% RH•
Power Supply : 9 volt battery
Exceeding Maximum pressure will cause permanent sensor damage.
Range: 0 ~ ±2 psi (MAN02)
0 ~ ±15 psi (MAN15)
0 ~ ±30 psi (MAN30)
Power Supply :
Conversion & Resolution
1 psi = Resolution
inch of HO
0.401 0.01
psi 0.0145 0.001
mbar 1 0.1
bar 0.001 0.001
mm of HO
10.2 1
Find Quality Products Online at: sales@GlobalTestSupply.com
8
1 psi = Resolution
bar 0.068948 0.001
mmHg 51.715073 0.5
ozin2 16.000844 0.2
Kgcm2 0.070309 0.001
psi 1 0.01
inH2O 27.679904 0.3
kpa 6.894757 0.1
ftH2O 2.306659 0.02
inHg 2.036021 0.01
cmH2O 70.306955 1
mbar 68.94757 1
Accuracy: ±0.3% of full scale at ±25°C (77°F)
Repeatability: ±0.2%~0.5% of full scale
Linearity / Hysteresis: ±0.29% to 1% of full scale
Combined accuracy: ±1.0% of full scale
Dimension: 72 x 182 x 30mm
Unit Weight: Approx. 220 (o.485lb) (with battery)
Response Time: 0.5 seconds
Format: Baud rate: 2400 bit/sec
Data bit: 8, Stop bit: 1
Parity: none
Find Quality Products Online at: sales@GlobalTestSupply.com
9
Fig. 1
Fig. 4
Fig. 7
Fig. 10
Fig. 2
Fig. 5
Fig. 8
Fig. 11
Fig. 3
Fig. 6
Fig. 9
Fig. 12
REC
:
:
H/M/S
DIF
Find Quality Products Online at: sales@GlobalTestSupply.com
10
Fig. 15
Fig. 14
Fig. 13
Fig. 16 Fig. 17
Fig. 18
HOLD DIF
DAT A LOGGER
REC MAX MIN BAT
inH O psi mbar bar mmH2O
2
MANOMETER
1
2
3
4
5
6
Find Quality Products Online at: sales@GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com