Amprobe DM7C User Manual [en, de, fr, es, it]

Contents
Introduction..........................................................................................2
Contents ..........................................................................................3
Transport and Storage ..........................................................................3
Safety ..........................................................................................3
Appropriate Usage ..............................................................................4
Feature Diagram ..................................................................................4
Voltage Measurement................................................................5
DC Voltage Measurement ..........................................................5
AC Voltage Measurement ..........................................................5
Resistance Measurement ..........................................................6
Battery Test................................................................................6
Diode Test..................................................................................7
Current Measurement ..............................................................7
Current Measurement 200 mA ..................................................8
Changing the batteries..........................................................................8
Fuse replacement..................................................................................9
Maintenance ........................................................................................9
Cleaning ....................................................................................9
Calibration Interval ....................................................................9
Specifications ......................................................................................10
Warranty ..........................................................................................11
Repair ..........................................................................................12
1
DM7C Multimeter
Users Manual
References marked on instrument or in users manual
The users manual contains information and references, necessary for safe operation and maintenance of the instrument. Prior to using the instrument the user is kindly requested to thoroughly read the users manual and comply with it in all sections. Failure to read the users manual or to comply with the warnings and references contained herein can result in serious bodily injury or instrument damage
Introduction
The Amprobe Multimeter DM7C is a universal, multi-purpose electrical measuring in­strument. It comply with the standards DIN VDE 0411 and EN 61010, and p r o vi d e safe, reliable operation. The multimeter is a valuable tool for all sorts of measurements in both trade and industry.
• 3
1
/2
digit LC Display, max. 1999 digits
• manual measurement range selection
• AC and DC voltage measurement up to 600 V
• DC Current measurement up to 200 mA
• Resistance measurement up to 2 MOhm
• Diode test
• Battery test for 1.5 V and 9 V batteries
After unpacking, check that the instrument is complete, and that all accessories are present.
2
Warning of a potential danger, comply with users manual.
Equipment protected throughout by double insulation or reinforced insulation.
Caution! Dangerous voltage. Danger of electrical shock.
Canadian Standards Association.
Conformity symbol, the instru­ment complies with the valid di­rectives. It complies with the EMC Directive (89/336/EEC) and the Low Voltage Directive (73/23/EEC) with their valid standards.
Reference. Please use utmost attention.
Symbol for the marking of elec­trical and electronic equipment (WEEE Directive 2002/96/EC).
Contents
1 pc. Amprobe Multimeter DM7C 1 pc. Holster 2 pcs. Test Leads (1ea. red, 1ea. black) 1 pc. Battery 9 V IEC 6F22 or 6LR61 1 pc. Users Manual
Transport and Storage
• Please keep the original packaging for later transport, e.g. for calibration. Any
transport damage due to faulty packaging will be excluded from warranty claims.
• In order to avoid instrument damage, it is advised to remove batteries when not using
the instrument over a certain time period. However, should the instrument be contaminated by leaking battery cells, you are kindly requested to return it to the factory for cleaning and inspection.
• Instruments must be stored in dry and closed areas. In the case of an instrument
being transported in extreme temperatures, a recovery time of minimum 2 hours is required prior to instrument operation.
Safety
• The Amprobe multimeter DM7C has been manufactured and tested to comply with
the safety regulations for electronic measuring equipment contained in IEC61010 and EN 61010, and left the factory in a safe condition. To maintain this condition, the user must observe the safety instructions contained in this users manual.
• To avoid electric shock, safety measures must be observed when working with volt-
ages higher than 120 V (60 V) DC or 50 V (25 V) RMS AC. These are the values of threshold contact voltages given by DIN VDE.
• Before each measurement make sure that the test leads and the instrument are un-
damaged.
• Only handle test leads and probes on the grips provided. Avoid touching probes under
any circumstances.
• Measurements in dangerous proximity of electrical installations are only to be exe-
cuted when instructed by a responsible electrical specialist, and never alone.
• The relevant safety regulations for electrical plant and equipment must be
observed during all operations.
• The instrument must only be used in the specified ranges.
• Before opening the instrument, it must be disconnected from all circuits.
• Protect the instrument from prolonged exposure to direct sunlight.
3
Appropriate Usage
• The instrument may only be used under those conditions and for those purposes for
which it was conceived. For this reason, in particular the safety references, the tech­nical data including environmental conditions and the usage in dry environments must be followed.
• When modifying or changing the instrument, the operational safety is no longer en-
sured.
• The instrument may only be opened by an authorised service technician, e.g. for fuse
replacement.
Feature Diagram
1. LCD
2. Measurement function selection switch
3. Input socket for measurement ranges
4. Ground connection for all measurement ranges
5. Input socket for current measurement range 200 mA
4
1
2
3
4
5
Operation
Measurements in dangerous proximity of electrical systems are only to be carried out in compliance with the instructions of a responsible electronics technician, and never alone. Test leads and test probes may only be touched at handle surfaces provided. Ab­solutely avoid the direct contact of the test probes. Prior to switching to a new measurement range or a new type of measurement, remove all connections from UUT. Measurements have to be carried out by respecting the standards.
Voltage Measurement
To avoid electrical shock, the valid safety measures and VDE directives strictly have to be met concerning excessive contact voltage when working with voltages exceeding 120V (60V) DC or 50V (25V)rms AC.
DC Voltage Measurement
1) Select measurement range (2V, 20V, 200V or 600V) via measurement
function selection switch.
2) Connect the black test lead to the COM socket and the red test lead to the
socket.
3) Connect test leads to UUT.
4) Read the measurement result displayed on the screen.
If display shows ”1” on the left, change selection to next highest voltage range.
AC Voltage Measurement
1) Select measurement range (200V or 600V) via measurement function
selection switch.
2) Connect the black test lead to the COM socket and the red test lead to the
socket.
3) Connect test leads to UUT.
4) Read the measurement result displayed on the screen.
If display shows ”1” on the left,, change selection to next highest voltage range.
5
Resistance Measurement
Prior to any resistance measurement it has to be ensured that the resistor to be tested does not have voltage across it. Failure to comply with this prescription can lead to dangerous user injuries or cause instrument damage. Addi tion ­ally, foreign voltages falsify the measurement result.
1) Select measurement range (2000!, 20k!, 200 k! or 2 M!) via measurement
function selection switch.
2) Connect the black test lead to the COM socket and the red test lead to the
socket.
3) Connect the test leads to UUT.
4) Read the measurement result displayed on the screen.
If display shows ”1” on the left, change selection to next highest resis.-range.
Battery Test
1) Select measurement range 9V or 1.5V via measurement function selection switch
.
2) Connect the black test lead to the COM socket and the red test lead to the
BATTERY TEST socket.
3) Connect test leads to UUT/Battery.
4) Read the measurement result displayed on the screen.
6
Diode Test
Prior to any diode test, it must be ensured, that the diode to be tested does not have voltage across it. Failure to comply with this prescription can lead to dan­gerous user injuries or cause instrument damage. Additionally, foreign voltages falsify the measurement result.
Resistors and semiconductor paths in parallel to the diode cause falsified meas­urement results.
1) Position measurement function selection switch to measurement range.
2) Connect the black test lead to the COM socket and the red test lead to the
socket.
3) Connect test leads to UUT.
4) Read the measurement result displayed on the screen.
Current Measurement
Ensure that the measurement circuit is not live when connecting the measurement instrument.
The instruments may only be used in current circuits protected at 16 A up to a nominal voltage of 600 V.
The nominal cross section of connecting line has to be respected and a safe con­nection has to be ensured.
After instrument fuse tripping eliminate the cause for the tripping prior to fuse re­placement.
7
Current Measurement 200 mA
1) Position measurement function selection switch to 200m measurement range.
2) Connect the black test lead to the COM socket and the red test lead to the 200mA
socket.
3) Connect test leads to UUT.
4) Read the measurement result displayed on the screen.
Changing the batteries
Prior to battery replacement, disconnect the instrument from any circuits.
Only use batteries as described in the specifications. If the symbol for Low-battery appears in the upper left corner of the display, the battery must be changed.
This is carried out as follows:
1) Separate the Amprobe multimeter DM7C from any circuit, and remove the test leads.
2) Switch the instrument off
3) Open the housing by removing the 4 screws on the rear face.
4) Remove the old battery.
5) Insert a new battery (type 1 x 9 V IEC 6F22 or 6LR61), taking care that the polarity
is correct. Make sure that no wires are trapped between the 2 halves of the housing, and close it again.
6) The instrument is now ready for further use.
Please consider your environment when you dispose of. They belong in a dispos­al area for hazardous waste.
Please, comply with the respective valid regulation regarding the return, recycling and disposal of used batteries.
If an instrument is not used over an extended time period, the batteries must be removed. Should the instrument be contaminated by leaking battery cells, the in­strument has to be returned for cleaning and inspection to the factory.
8
Fuse replacement
Prior to fuse replacement, disconnect the instrument from any circuits.
Only use fuses of voltage and current values as described in the specifications.
Using auxiliary fuses, in particular short-circuiting fuse holders is prohibited and can cause instrument destruction or serious bodily injury of operator.
1) Separate the Amprobe multimeter DM7C from any circuit, and remove the test leads.
2) Switch the instrument off and remove the holster.
3) Open the housing by removing the 4 screws on the rear face.
4) Remove the defective fuse
5) Insert a new fuse. Make sure that no wires are trapped between the 2 halves of the
housing, and close it again.
6) The instrument is now ready for further use.
Maintenance
Provided it is used in accordance with the users manual, the instrument needs no spe­cial maintenance.
Cleaning
If the instrument is dirty after daily usage, it is advised to clean it by using a damp cloth and a mild household detergent.
Prior to cleaning, ensure that instrument is switched off and disconnected from external voltage supply and any other instruments connected (such as UUT, con­trol instruments, etc.).
Never use acid detergents or dissolvants for cleaning.
Calibration Interval
We suggest a calibration interval of one year. If the instrument is used very often or if it is used under rough conditions we recommend shorter intervals. If the instrument is used only a few times a year, the calibration interval can be extended to 3 years.
9
Specifications (at 23° C ± 5° C, max. 75 % rel. humidity)
Display: 3
1
/2 digits, LC-Display
Total display: 1999 digit Overload display: the left ”1” is displayed Polarity display: automatic Battery status display: Battery symbol appears (7.2 V) Overvoltage class: CAT II 600 V Degree of contamination: 2 Power supply: Battery 9V IEC 6F22 or 6LR61 Power consumption: approx. 1,5 mA (typical) Dimensions: 170 x 85 x 50 mm incl. Holster Weight: approx.. 410 g (incl. Holster)
Ambient conditions: Operation temperature: 0...50°C (0...80% rel. humidity) Storage temperature: -10...60°C (0...80% rel.humidity)(without battrie) Height above sea level: up to 2000 m Overload protection: Fuse (mA) : FF200mA / 700V / 50kA , 6x32mm Temperature coefficient: 0,15 x specified accuracy per 1°C (<18°C and >28°C)
Voltage DC
Range Resolution Accuracy Overload protection 2 V 1 mV ± (1.2% 20 V 10 mV rdg. + 3 digit) 600 V DC/ACeff 200 V 100 mV 600 V 1 V Input Impedance 1 MOhm
Voltage AC (40...1 kHz)
Range Resolution Accuracy
40...100 Hz 100Hz...1kHz 200 V 100 mV ±(1.5% rdg.+3digit) ±(2.5% rdg.+ 3digit) 600 V 1 V Overload protection: 600 V DC/ACeff Input Impedance: 460 kOhm
Battery Test
Range Resolution Accuracy Overload protection
1.5V 10 mV ±(2.5% rdg 600 V DC/ACeff 9 V 10 mV +3 digit)
10
Current DC
Range Resolution Accuracy Overload protection 200 mA 100 µA ± (2,.5% rdg. 600 V DC/ACeff
+ 3 digit)
Resistance
Range Resolution Accuracy Overload protection 2000 Ohm 1 Ohm ± (1.5% rdg. 20 kOhm 10 Ohm + 3 digit) 600 V DC/ACeff 200 kOhm 100 Ohm 2 MOhm 1 kOhm
Test Voltage < 3,2 V
Diode Test
Range Resolution Test Voltage Test Current Diode 1 mV < 3.2 V < 1 mA
Overload protection 600 V DC/ACeff
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for 1 year from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Amprobe’s behalf. To obtain service during the warranty period, return the product with proof of purchase to an authorized Amprobe Test Tools Service Center or to an Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES - WHETHER EXPRESS, IMPLIED OR STAUTORY - INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SH ALL NOT B E LIABL E FOR A NY SPEC IAL , INDIR ECT, INC IDENTAL O R CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
11
Repair
All test tools returned for warranty or non-warranty repair or for calibration should be accompanied by the following: your name, company’s name, address, telephone number, and proof of purchase. Additionally, please include a brief description of the problem or the service requested and include the test leads with the meter. Non-warranty repair or replacement charges should be remitted in the form of a check, a money order, credit card with expiration date, or a purchase order made payable to Amprobe® Test Tools.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During the warranty period any defective test tool can be returned to your Amprobe® Test Tools distributor for an exchange for the same or like product. Please check the “Where to Buy” section on www.amprobe.com for a list of distributors near you. Additionally, in the United States and Canada In-Warranty repair and replacement units can also be sent to a Amprobe® Test Tools Service Center (see below for address).
Non-Warranty Repairs and Replacement – US and Canada
Non-warranty repairs in the United States and Canada should be sent to a Amprobe® Test Tools Service Center. Call Amprobe® Test Tools or inquire at your point of purchase for current repair and replacement rates.
In USA In Canada Amprobe Test Tools Amprobe Test Tools Everett, WA 98203 Mississauga, ON L4Z 1X9 Tel: 877-993-5853 Tel: 905-890-7600 Fax: 425-446-6390 Fax: 905-890-6866
Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe® Test Tools distributor for a nominal charge. Please check the “Where to Buy” section on www.amprobe.com for a list of distributors near you.
European Correspondence Address* Amprobe® Test Tools Europe P.O. Box 1186 5602 BD Eindhoven
The Netherlands *(Correspondence only – no repair or replacement available from this address. European customers please contact your distributor.)
12
Inhalt
Einleitung ..........................................................................................2
Lieferumfang ........................................................................................3
Transport und Lagerung........................................................................3
Sicherheitsmaßnahmen ........................................................................3
Bestimmungsgemäße Verwendung ......................................................4
Bedienelemente ..................................................................................4
Durchführen von Messungen................................................................5
Spannungsmessung ................................................................5
Gleichspannungsmessung DC ..................................................5
Wechselspannungsmessung AC................................................5
Widerstandmessung..................................................................6
Batterietest ................................................................................6
Dioden-Test ..............................................................................7
Strommessung ........................................................................7
Strommessung 200 mA ............................................................8
Batteriewechsel ....................................................................................8
Sicherungswechsel ..............................................................................9
Wartung ..........................................................................................9
Reinigung ..................................................................................9
Kalibrierintervall ........................................................................9
Technische Daten ................................................................................10
Gewährleistung ....................................................................................13
Reparatur ..........................................................................................13
1
Bedienungdsanleitung
DM7C Multimeter
Auf dem Gerät und in der Bedienungsanleitung vermerkte Hin­weise:
Die Bedienungsanleitung enthält Informationen und Hinweise, die zu einer sicheren Bedienung und Nutzung des Gerätes notwendig sind. Vor der Verwen­dung des Gerätes ist die Bedienungsanleitung aufmerksam zu lesen und in allen Punkten zu befolgen. Wird die Anleitung nicht beachtet oder sollten Sie es ver­säumen, die Warnungen und Hinweise zu beachten, können ernste Verletzungen des Anwenders bzw. Beschädigungen des Gerätes eintreten.
Einleitung / Lieferumfang
Die Amprobe Multimeter DM7C ist ein universell einsetzbares Messinstrument. Das Messgerät wird nach den neuesten Normen gebaut und gewährleistet ein sicheres und zuverlässiges Arbeiten. Das Multimeter ist im handwerklichen und industriellen Bere­ich eine wertvolle Hilfe.
• 3
1
/2 stellige digitale Anzeige, maximal 1999 Digits
• Manuelle Messbereichswahl
• Gleich- und Wechselspannungsmessung bis 600 V
• Gleichstrommessung bis 200 mA
• Widerstandsmessung bis 2 MOhm
• Diodentest
• Batterietest für 1,5V- und 9V-Batterien
Bitte überprüfen Sie nach dem Auspacken des Gerätes den Lieferumfang auf Vollständigkeit.
2
Ac htung : Warn ung vor e iner Gefahr enstelle, Bedienun gsan­leitung beachten.
Sc hutzi solie run g; Ger ät der Schutzklasse II entsprechend IEC
61140.
Spannung: Vorsicht! Gefährliche Spannung, Gefahr des elektrischen Schlages.
Canadian Standards Association.
CE K onf ormit äts -Zeic hen, bestätigt die Einhalten der gülti­gen Richtlinien. Die Anforderun­ge n de r EM V-Ric htlin ie 89/ 336/E WG und der Niede r­spannungsrichtlinie 73/23/EWG mit den jeweils betreffenden Nor­men werden ebenfalls eingehal­ten.
Hinweis: Bitte unbedingt beacht­en.
Kennzeichnung elektrischer und elekt ronischer Geräte (WEEE Richtlinie 2002/96/EG)
Im Lieferumfang ist enthalten: 1 St. Amprobe Multimeter DM7C 1 St. Schutzhülle 2 St. Messleitungen (1x rot, 1x schwarz) 1 St. Batterie 9V IEC 6F22 oder 6LR61 1 St. Bedienungsanleitung
Transport und Lagerung
• Bitte bewahren Sie die Originalverpackung für eine spätere Versendung, z.B. zur Kali-
bration auf. Transportschäden aufgrund mangelhafter Verpackung sind von der Garantie ausgeschlossen.
• Um Beschädigungen zu vermeiden, sollten die Batterien entnommen werden, wenn
das Messgerät über einen längeren Zeitraum nicht benutzt wird. Sollte es dennoch zu einer Verunreinigung des Gerätes durch ausgelaufene Batteriezellen gekommen sein, muß das Gerät zur Reinigung und Überprüfung ins Werk eingesandt werden.
• Die Lagerung des Gerätes muß in trockenen, geschlossenen Räumen erfolgen. Sollte
das Gerät bei extremen Temperaturen transportiert worden sein, benötigt es vor dem Einschalten eine Aklimatisierung von mindestens 2 Stunden.
Sicherheitsmaßnahmen
• Das Amprobe Multimeter DM7C wurde gemäß den Sicherheitsbestimmungen IEC
61010, EN 61010 für elektronische Messgeräte gebaut, geprüft und hat das Werk in sicherheitstechnisch einwandfreiem Zustand verlassen. Um diesen Zustand zu erhal­ten, muß der Anwender die Sicherheitshinweise in dieser Anleitung beachten.
• Um einen elektrischen Schlag zu vermeiden, sind Vorsichtsmaßnahmen zu beacht-
en, wenn mit Spannungen größer 120 V (60 V) DC oder 50 V (25 V) eff AC gearbeit­et wird. Diese Werte stellen nach DIN VDE die Grenze der noch berührbaren Span­nungen dar (Werte in Klammern gelten für z.B. medizinische oder landwirtschaftliche Bereiche).
• Vor jeder Messung vergewissern, dass die Messleitungen und das Prüfgerät in ein-
wandfreiem Zustand sind.
• Die Messleitungen und Prüfspitzen dürfen nur an den vorgesehenen Handgriffen
(Griffflächen) angefaßt werden. Das Berühren der Prüfspitzen ist unter allen Umstän­den zu vermeiden.
• Messungen in gefährlicher Nähe elektrischer Anlagen sind nur nach Anweisung einer
verantwortlichen Elektrofachkraft und nicht alleine durchzuführen.
• Bei sämtlichen Arbeiten müssen die Unfallverhütungsvorschriften der gewerblichen
Berufsgenossenschaften für elektrische Anlagen und Betriebsmittel beachtet wer­den.
• Das Prüfgerät darf nur in den spezifizierten Messbereichen eingesetzt werden.
• Vor dem Öffnen des Batteriefachs muss das Gerät von allen Messkreisen getrennt
werden.
3
• Eine Erwärmung des Gerätes durch direkte Sonneneinstrahlung ist zu vermeiden.
Nur so kann eine einwandfreie Funktion und eine lange Lebensdauer gewährleistet werden.
Bestimmungsgemäße Verwendung
• Das Gerät darf nur unter den Bedingungen und für die Zwecke eingesetzt werden, für
die es konstruiert wurde. Hierzu sind besonders die Sicherheitshinweise, die Tech­nischen Daten mit den Umgebungsbedingungen und die Verwendung in trockener Umgebung zu beachten.
• Die Betriebssicherheit ist bei Modifizierung oder Umbauten nicht mehr gewährleis-
tet.
• Das Gerät darf nur vom autorisierten Servicetechniker, geöffnet werden.
Bedienelemente
1. LC-Anzeige
2. Wahlschalter Messfunktion
3. Eingangsbuchse für die Messbereiche
4. Masseanschluss für alle Messbereiche
5. Eingangsbuchse für Strommessbereich 200 mA
4
1
2
3
4
5
Durchführen von Messungen
Messungen in gefährlicher Nähe elektrischer Anlagen sind nur nach Anweisung einer verantwortlichen Elektrofachkraft und nicht alleine durchzuführen. Die Messleitungen und Prüfspitzen dürfen nur an den dafür vorgesehenen Griff­flächen gehalten werden. Das Berühren der Prüfspitzen ist unter allen Umständen zu vermeiden. Vor dem Umschalten in einen anderen Messbereich oder auf eine neue Messart müssen die Anschlüsse vom Prüfobjekt getrennt werden. Messungen müssen entsprechend den jeweils geltenden Normen durchgeführt werden.
Spannungsmessung
Um einen elektrischen Schlag zu vermeiden, sind unbedingt die geltenden Sicher­heits- und DIN VDE-Bestimmungen bezüglich zu hoher Berührungsspannung zu beachten, wenn mit Spannungen größer 120 V (60 V) DC oder 50 V (25 V)eff AC gearbeitet wird. Die Werte in Klammern gelten für eingeschränkte Bereiche (wie z.B. Medizin, Landwirtschaft).
Gleichspannungsmessung DC
1) Entsprechenden Messbereich mit Wahlschalter Messfunktion anwählen.
2) Die schwarze Messleitung mit der COM-Buchse und die rote Messleitung mit der
-Buchse verbinden.
3) Messleitungen mit dem Prüfobjekt verbinden.
4) Messergebnis vom Display ablesen.
Wechselspannungsmessung AC
1) Entsprechenden Messbereich mit Wahlschalter Messfunktion anwählen.
2) Die schwarze Messleitung mit der COM-Buchse und die rote Messleitung mit der
-Buchse verbinden.
3) Messleitungen mit dem Prüfobjekt verbinden.
4) Messergebnis vom Display ablesen.
5
Widerstandmessung
Vor jeder Widerstandsmessung muss sichergestellt werden, dass der zu prüfende Widerstand spannungsfrei ist. Bei Nichtbeachtung können schwerwiegende Ver­letzungen des Anwenders bzw. Beschädigungen des Gerätes verursacht werden. Zusätzlich verfälschen Fremdspannungen das Messergebnis.
1) Entsprechenden Messbereich ! mit Wahlschalter Messfunktion anwählen.
2) Die schwarze Messleitung mit der COM-Buchse und die rote Messleitung mit der
-Buchse verbinden.
3) Messleitungen mit dem Prüfobjekt verbinden.
4) Messergebnis vom Display ablesen.
Batterietest
1) Messbereich 1,5V oder 9V mit Wahlschalter Messfunktion anwählen.
2) Die schwarze Messleitung mit der COM-Buchse und die rote Messleitung mit
der BATTERY-TEST (200mA)-Buchse verbinden.
3) Messleitungen mit dem Prüfobjekt verbinden.
4) Messergebnis vom Display ablesen.
6
Dioden-Test
Vor jeder Diodenprüfung muss sichergestellt werden, dass die zu prüfende Diode spannungsfrei ist. Bei Nichtbeachtung können schwerwiegende Verletzungen des Anwenders bzw. Beschädigungen des Gerätes verursacht werden. Zusätzlich ver­fälschen Fremdspannungen das Messergebnis.
Parallel zur Diode liegende Widerstände und Halbleiterstrecken verfälschen das Messergebnis.
1) Messbereich mit Wahlschalter Messfunktion anwählen.
2) Die schwarze Messleitung mit der COM-Buchse und die rote Messleitung mit der
-Buchse verbinden.
3) Messleitungen mit dem Prüfobjekt verbinden.
4) Messergebnis vom Display ablesen.
Strommessung
Zum Anschluss des Messgerätes muss der Messkreis spannungsfrei geschaltet sein.
Das Messgerät darf nur in mit 16 A abgesicherten Stromkreisen bis zu einer Nennspannung von 600 V verwendet werden.
Der Nennquerschnitt der Anschlussleitung muss beachtet und für eine sichere Verbindung (z.B. über Krokodilklemmen) gesorgt werden.
Beseitigen Sie nach dem Auslösen von Sicherungen des Messgerätes vor dem Sicherungswechsel zuerst die Ursache für das Auslösen der Sicherungen.
7
Strommessung 200 mA
1) Messbereich 200m mit Wahlschalter Messfunktion anwählen.
2) Die schwarze Messleitung mit der COM-Buchse und die rote Messleitung mit der
200mA-Buchse verbinden.
3) Messleitungen mit dem Prüfobjekt verbinden.
4) Messergebnis vom Display ablesen.
Batteriewechsel
Vor dem Batteriewechsel muss das Gerät von den angeschlossenen Messkabeln getrennt werden. Es dürfen nur die in den Technischen Daten spezifizierten Batterien verwendet wer­den! Wenn das Batteriesymbol in der Anzeige erscheint, mus die Batterie gewechselt werden.
1) Das Gerät von allen Messkreisen trennen.
2) Gerät ausschalten und die Schutzhülle entfernen.
3) Schrauben im Gehäuseunterteil lösen und Gehäuseoberteil von Gehäuseunterteil
vorsichtig abheben.
4) Entleerte Batterie entnehmen.
5) Neue Batterien polrichtig einsetzen.
6) Gehäuseoberteil mit Gehäuseunterteil verschrauben.
Bitte denken Sie an dieser Stelle auch an unsere Umwelt. Werfen Sie verbrauchte Batterien nicht in den normalen Hausmüll, sondern geben Sie die Batterien bei Sondermülldeponien oder Sondermüllsammlungen ab. Meist können Batterien auch dort abgegeben werden, wo neue gekauft werden. Es müssen die jeweils gültigen Bestimmungen bzgl. der Rücknahme, Verwertung und Beseitigung von gebrauchten Batterien und Akkumulatoren beachtet werden. Wird das Gerät über einen längeren Zeitraum nicht benutzt, sollten die Batterien entnommen werden. Sollte es zu einer Verunreinigung des Gerätes durch ausge­laufene Batteriezellen gekommen sein, muss das Gerät zur Reinigung und Über­prüfung ins Werk eingesandt werden
8
Sicherungswechsel
Vor dem Sicherungswechsel muss das Gerät von den angeschlossenen Messk­abeln getrennt werden. Verwenden Sie ausschließlich Sicherungen mit den unter den Technische Daten aufgeführten Spannungs- und Stromwerten. Das Verwenden von Behelfssicherungen, insbesonders das Kurzschließen der Sicherungshalter, ist unzulässig und kann die Zerstörung des Gerätes und schw­erwiegende Verletzungen des Bedienpersonals verursachen.
1) Das Gerät von allen Messkreisen trennen.
2) Gerät ausschalten.
3) Schrauben im Gehäuseunterteil lösen und Gehäuseoberteil von Gehäuseunterteil vorsichtig abheben.
4) Defekte Sicherung entehmen
5) Neue Sicherung einsetzen.
6) Gehäuseoberteil mit Gehäuseunterteil verschrauben.
Wartung
Das Gerät benötigt bei einem Betrieb gemäß der Bedienungsanleitung keine besondere Wartung.
Reinigung
Sollte das Gerät durch den täglichen Gebrauch schmutzig geworden sein, können
die Geräte mit einem feuchten Tuch und etwas mildem Haushaltsreiniger gereinigt
werden.
Vor der Reinigung muss das Gerät von allen Messkreisen getrennt werden.
Niemals scharfe Reiniger oder Lösungsmittel verwenden.
Kalibrierintervall
Um die angegeben Genauigkeiten der Meßergebnisse zu erhalten, muss das Gerät regelmäßig durch unseren Werksservice kalibriert werden. Wir empfehlen ein Kalibrier­intervall von einem Jahr. Bei häufigem Einsatz des Gerätes bzw. bei Anwendungen unter rauhen Bedingungen sind kürzere Fristen zu empfehlen. Sollte das Gerät wenig benutzt werden, so kann das Kalibrierintervall auf bis zu 3 Jahre verlängert werden.
9