Amprobe BT-250/KIT User manual

Page 1
Circuit Tracer
Transmitter
BT-120, BT-250
Breaker Tracer
Circuit Tracer
Receiver
www.GlobalTestSupply.com
User Manual
ENG
Page 2
BT-120 BT-250
Breaker Tracer
User Manual
8/2018, 6011458B ©2018 Amprobe All rights reserved. Printed in China
English
www.GlobalTestSupply.com
Page 3
BT-120 BT-250 Breaker Tracer
1
6
2
Circuit Tracer
Receiver
7
Circuit Tracer
Transmitter
4
3
Sensor Location
1
Identification Arrow LED Indicator
2
Battery Compartment
3
ON/OFF Switch
4
Adaptor Prongs
5
Transmitter Power LED Indicator
6
7
Receiver Power LED Indicator
5
www.GlobalTestSupply.com
Page 4
BT-120 BT-250 Breaker Tracer
CONTENTS
SYMBOLS AND WARNINGS ..................................................................................2
Safety Information ...........................................................................................2
Warnings and Precautions ................................................................................3
UNPACKING AND INSPECTION .............................................................................3
INTRODUCTION .....................................................................................................4
OPERATION ...........................................................................................................4
SPECIFICATION ......................................................................................................5
BATTERY REPLACEMENT .......................................................................................7
MAINTENANCE AND REPAIR ................................................................................7
1
www.GlobalTestSupply.com
Page 5
SYMBOLS
Battery
Double Insulated
Alternating Current
Direct Current
Application around and removal from hazardous
live conductors is permitted
Do not dispose of this product as unsorted municipal waste. Contact
=
a qualified recycler for disposal.
Underwriters Laboratories.
[Note: Canadian and US.]
Caution ! Refer to the
Manual
Dangerous Voltage
Earth Ground
Fuse
Complies with European
Directives
Conforms to relevant Australian standards
Audible tone
Safety Information
• The BT-120 Breakers Tester is conformed to UL 61010-1; CAT II 120 V, class II and pollution degree 1 or 2.
• The BT-250 Breakers Tester is conformed to UL 61010-1; CAT II 250 V, class II and pollution degree 1 or 2.
Warning and precaution
• Before and after hazardous voltage measurements, test the voltage function on a known source such as line voltage to determine proper meter functioning.
• Do not touch the membrance.
• Inspect the receiver and the transmitter before every use. Do not use any damaged part.
• Never ground yourself when taking measurements. Do not touch exposed circuit elements or test probe tips.
• Do not operate the instrument in an explosive atmosphere.
2
www.GlobalTestSupply.com
Page 6
• To reduce the risk of fire or electric shock, do not expose this product to rain or moisture.
• The meter is intended only for indoor use. To avoid electrical shock hazard, observe the proper safety precautions when working with voltages above 60 VDC, 42.4 Vpk, or 30 VAC rms. These voltage levels pose a potential shock hazard to the user.
UNPACKING AND INSPECTION
Your shipping carton should include:
1 Receiver 1 Transmitter 1 9V alkaline battery 1 User manual 1 Carrying case (BT-250) 1 Connection cable (BT-250) 1 Light fixture adapter (BT-250)
If any of the items are damaged or missing, return the complete package to the place of purchase for an exchange.
INTRODUCTION
The Amprobe BT Series Breaker Tracers work on powered systems from 90 to 120 V AC (BT-120) and 90 to 250 V AC (BT-250) and are designed for use in residential and light commercial environments. Both kits comes complete with a Transmitter and Receiver.
Features:
• Identifies circuit breaker location
• Works on all electrical systems within the voltage rating range
• Perfect for office, residential and HVAC applications
• Automatic sensitivity adjustment
• Microprocessor controlled
• Extremely accurate reading always finds the right breaker
• Durable and dependable
3
www.GlobalTestSupply.com
Page 7
OPERATION
Breaker and Fuse Identification
1. Plug the Transmitter into an energized wall outlet (Figure 1).
Transmitter
Circuit Tracer
Figure 1
2. Make sure the red LED is ON to indicate that the outlet is energized.
3. Push the Power switch on the Receiver to turn it ON. The Receiver will beep and the Power LED will be lit.
4. At the breaker panel or fuse box, hold the Receiver perpendicular to the breakers. Scan slowly all breakers once to calibrate the Receiver. During this scan, the Receiver may beep and flash at several breakers. This is a normal part of the identification process (Figure 2).
Receiver
Circuit Tracer
Figure 2
5. Without touching the ON/OFF button, begin scanning again to identify the right breaker or fuse. When the Receiver beeps, the correct breaker has been identified.
6. Push and hold the Power switch for three seconds to turn the Receiver off. Beeping and flashing during the shutdown is normal.
7. Unplug the Transmitter from the wall outlet.
4
www.GlobalTestSupply.com
Page 8
Using the BT-LFA Incandescent Light Fixture Adapter (BT-250 only)
1. Remove the light bulb exercising the proper caution.
2. Screw in the BT-LFA adapter.
3. Plug the Transmitter into the BT-LFA (Figure 3).
Figure 3
4. Preform scan following directions in step 4 under “Operation.”
Using the BT-VLA High Voltage Leads Adapter (BT-250 only)
The BT-250 Transmitter can be used alone on 110V circuits. For circuits up to 250V or exposed circuits, use the BT-VLA and follow the directions listed below.
1. Plug Transmitter into high voltage lead receptacle. Be sure that the Transmitter is completely seated and that no part of the Transmitter prongs are exposed.
2. Attach the high voltage leads to the terminal or conductor while exercising extreme caution (Figure 4).
3. Preform scan following directions in step 4 under “Operation.”
Figure 4
5
www.GlobalTestSupply.com
Page 9
SPECIFICATIONS
Operating & Detection BT-120 BT-250
Voltage for Transmitter
Operating Frequency 50 Hz to 60 Hz 40 Hz to 70 Hz Storage Temperature 32 °F to 104 °F (0 °C to 40 °C) Operating Temperature 32 °F to 104 °F (0 °C to 40 °C) Sensitivity Adjustment Automatic Power Supply 9 V alkaline battery (Receiver) Transmitter Polarity Automatic Humidity 50% RH (non condensing)
Environmental Conditions
Caution
90-120 VAC, Not to exceed ±10% (inclusive in rating)
Indoor use, altitudes up to 2000 m, Pollution Degree 1 or 2, Installation Category II
BT-VLA cord assembly to be used with Transmitter only according to instructions.
90-250 VAC, Not to exceed ±10% (inclusive in rating)
BATTERY REPLACEMENT
Power is supplied by one 9 V alkaline battery (Receiver). The Identification Arrow LED on the Receiver will not turn ON when
replacement is needed. To replace the battery, remove the screw from the back of the meter and slide the battery door outward. Remove the battery from case bottom and replace it with a fresh 9 V alkaline battery.
Slide battery
door outward
Remove screw
6
www.GlobalTestSupply.com
Page 10
MAINTENANCE AND REPAIR
If there appears to be a malfunction during the operation of the tracer, the following steps should be performed in order to isolate the cause of the problem.
1. Check the battery. Replace the battery immediately if the Receiver LED doesn’t turn ON.
2. Review the operating instructions for possible mistakes in operating procedure.
Except for the replacement of the battery, repair of the tracer should be performed only by a Factory Authorized Service Center or by other qualified instrument service personnel.
The front panel and case can be cleaned with a mild solution of detergent and water. Apply sparingly with a soft cloth and allow to dry completely before using. Do not use aromatic hydrocarbons or chlorinated solvents for cleaning.
7
www.GlobalTestSupply.com
Loading...