Amprobe AM-510 User Manual

Page 1
AM-510
Commercial / Residential Multimeter
Users Manual
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Page 2
Page 3
AM-510
Commercial / Residential Multimeter
Users Manual
July 2011, Rev.1 ©2011 Amprobe Test Tools. All rights reserved. Printed in China
English
Page 4
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for 1 year from the date of purchase, unless local laws require otherwise. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Amprobe’s behalf. To obtain service during the warranty period, return the product with proof of purchase to an authorized Amprobe Test Tools Service Center or to an Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES - WHETHER EXPRESS, IMPLIED OR STAUTORY - INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Repair
All test tools returned for warranty or non-warranty repair or for calibration should be accompanied by the following: your name, company’s name, address, telephone number, and proof of purchase. Additionally, please include a brief description of the problem or the service requested and include the test leads with the meter. Non-warranty repair or replacement charges should be remitted in the form of a check, a money order, credit card with expiration date, or a purchase order made payable to Amprobe® Test Tools.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During the warranty period any defective test tool can be returned to your Amprobe® Test Tools distributor for an exchange for the same or like product. Please check the “Where to Buy” section on www.amprobe. com for a list of distributors near you. Additionally, in the United States and Canada In-Warranty repair and replacement units can also be sent to a Amprobe® Test Tools Service Center (see address below).
Page 5
AM-510
Commercial / Residential Multimeter
1
2
3
6
4
5
10
9
Flash light
1
LCD Display
2
Function Buttons
3
Input Terminal for voltage, frequency, diode,
7
capacitance, resistance and continuity measurement
COM (return) terminal for all measurements
8
Input Terminal for battery test and AC/DC mA or μA
9
measurement
10
Input Terminal for AC/DC A measurement to10A
Flash light Button
4
Rotary Switch
5
SELECT Button
6
7
8
Page 6
Screen Display
5
4 3
2
1
10 11 12 13
76 8 9
The Meter selects the range with best resolution
1
Negative reading
2
Alternate Current
3
Low battery indicator
4
Data hold
5
Diode test
6
Continuity test
7
Relative zero mode
8
Non-Contact Voltage
9
Measurement units Duty Cycle Measurement unit for voltage Battery Test
10
11
12 13
Page 7
AM-510
Commercial / Residential Multimeter
CONTENTS
SYMBOL .................................................................................................................2
SAFETY INFORMATION .........................................................................................2
UNPACKING AND INSPECTION .............................................................................3
FEATURES ...............................................................................................................4
MAKING MEASUREMENT .....................................................................................5
Measuring AC and DC Voltage .........................................................................6
Measuring AC and DC Current .........................................................................7
Measuring Resistance .......................................................................................8
Measuring Continuity .......................................................................................9
Measuring Diode ...............................................................................................9
Measuring Capacitance ....................................................................................10
Measuring Frequency .......................................................................................10
Non-Contact Voltage Sensing ..........................................................................11
Battery Test ........................................................................................................12
SPECIFICATION .......................................................................................................12
MAINTENANCE ......................................................................................................17
BATTERY AND FUSE REPLACEMENT .....................................................................18
1
Page 8
SYMBOLS
Caution ! Risk of electric shock.
Caution! Refer to the explanation in this Manual
Alternating Current (AC)
Direct Current (DC)
The equipment is protected by double insulation or reinforced
insulation
Earth (Ground)
Audible tone
Battery
Complies with European Directives
Conforms to relevant Australian standards
Canadian Standards Association (NRTL/C)
Do not dispose of this product as unsorted municipal waste. Contact
a qualified recycler.
SAFETY INFORMATION
The Meter complies with: IEC/EN 61010-1 3rd Edition, UL61010-1 2nd Ed. and CAN/CSA C22.2 No.
61010.1-0.92 to Category III 600 Volts, Pollution degree 2 IEC/EN 61010-2-030 IEC/EN 61010-2-31 for test leads EMC IEC/EN 61326-1
Measurement Category III (CAT III) is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit- breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to the fixed installation.
2
Page 9
CENELEC Directives
The instruments conform to CENELEC Low-voltage directive 2006/95/EC and Electromagnetic compatibility directive 2004/108/EC
Warning: Read Before Using
• To avoid possible electrical shock or personal injury, follow these instructions and use the Meter only as specied in this manual.
• Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. If in doubt, have the Meter serviced.
• Always use the proper function and range for measurements.
• Before rotating the function range selection switch, disconnect test probe from circuit under test.
• Verify the Meter’s operation by measuring on a known voltage source.
• Do not apply more than the rated voltage, as marked on the Meter, between the test probe or between any test probe and earth ground.
• Use the Meter with caution for voltages above 30 Vac rms, 42 Vac peak, or 60 Vdc. These voltages pose electrical shock hazards.
• Disconnect circuit power and discharge all high-voltage capacitors before testing resistance.
• Do not use the Meter around explosive gas or vapor.
• When using the test leads, keep your ngers behind the nger guards.
• Remove test leads from the Meter before opening the Meter case or battery door.
UNPACKING AND INSPECTION
Your shipping carton should include: 1 AM-510 Multimeter 1 Pair of test leads 1 9V (6F22) battery (installed) 1 Users manual If any of the items are damaged or missing, return the complete package to
the place of purchase for an exchange.
3
Page 10
FEATURES
The digital multimeter designed for advanced residential applications. Rewire an electrical panel, install heated oors or new light fixtures, troubleshoot and repair home appliances, electrical sockets and automotive electrical problems with this easy-to-use multimeter. The AM-510 features a built-in ashlight to see wires in the dark, a kick stand, and probe holder to give you the “third-hand” you need while taking measurements and non-contact voltage detection for quick go-no-go checks without the need for an additional tool. Compact yet tough, this multimeter is built to last through all your electrical projects.
• Measurements: Voltage up to 600V AC/DC, AC/DC Current and Resistance
• Frequency, Capacitance, Duty Cycle for troubleshooting applications
• Special Functions:
- Non-contact Voltage Detection
- Audible continuity
- Diode Test
• Backlit LCD display
• Events:
- Data hold
- Relative zero mode
• Built in work light (ashlight)
• Built in test leads storage and “third hand” probe holder
• Auto and Manual ranging
• Auto power off
• Low battery warning
• Safety: CAT III 600V
4
Page 11
MAKING MEASUREMENT

1. Use the proper function and range for measurements.
2. To avoid possible electrical shock, personal injury or damages to the Meter, disconnect circuit power and discharge all high-voltage capacitors before testing resistance and diode.
3. Connecting test leads:
• Connect the common (COM) test lead to the circuit before connecting the live lead;
• After measurement, remove live lead before removing the common (COM) test lead from the circuit
4. Symbol “OL” is displayed on LCD when the measurement is out of range.
Rotary Switch Positions
Switch Position Measurement Function
V
Hz Frequency measurement
% Duty cycle
NCV Non-contact voltage
mA A
μA
AC or DC voltage measurement (use SELECT button for switching to AC or DC).
Resistance measurement Voltage measurement of diode PN junction
Continuity measurement Capacitance measurement
9V For measurement of dry batteries of not exceeding 15Vdc
1.5V For measurement of dry batteries of not exceeding 2Vdc AC or DC current measurement (use SELECT button for
switching to AC or DC).
Rotary Switch Positions
Button Measurement Function
SELECT
Press the yellow SELECT button to select alternate measurement functions on the rotary switch.
5
Page 12
HOLD /
Press
REL
RANGE
Hz / %
to enable the function when at relevant rotary switch function.
Display freezes present reading / press 2 sec to turn on LCD backlight.
Relative zero mode Manual or Auto range switching. The default setting
is Auto ranging, press to switch to manual ranging (selectable resolutions). Press for 2 sec to return to auto ranging.
Frequency / Duty Cycle. Press to turn on Frequency measurement mode; press again for duty cycle measurement.
Flash light
Auto Power OFF
Auto power off: approx. 30 minutes. When the Meter is in auto power off mode, press any button to resume normal operation.
Measuring AC and DC Voltage
To avoid personal injury or damage to the Meter, do not apply voltage
higher than 600V.
Auto Range /Manual Range
6
Page 13
Measuring AC and DC Current
Press SELECT button to select AC or DC current measurement function.
To avoid personal injury or damage to the Meter:
1. Do not attempt to make an in-circuit current measurement when the open-circuit potential to earth ground exceeding 600V.
2. Switch to proper function and range for your measurement.
3. Do not place the test probe in parallel with a circuit when the test leads are connected to the current terminals.
4. Connect the test leads to the correct input A/mA μA current terminal and to the circuit before powering the circuit under test.
5. For current range from 8-10A, do not measure current for more than 20 minutes. Wait for 10 minutes before taking another measurement
6. After measurement, switching OFF the circuit’s power before removing test leads from the circuit.
AC/DC
RL
Auto Range /Manual Range
RL
7
Page 14
AC/DC
RL
Auto Range /Manual Range
RL
Measuring Resistance
Disconnect circuit power and discharge all high-voltage capacitors
before testing resistance.
R
Note: On a higher resistance measurement (>1M), the measurement may take a few seconds to get stable reading. Over range or open circuit indication: OL
8
Page 15
Measuring Continuity
Disconnect circuit power and discharge all high-voltage capacitors
before testing continuity.
R 1 0
R 150
Measuring Diode
Disconnect circuit power and discharge all high-voltage capacitors
before testing diode.
D
9
Page 16
Measuring Capacitance
Disconnect circuit power and discharge all high-voltage capacitors
before testing capacitance.
Auto Range
C
Measuring Frequency
Press Hz/% button to select Frequency / Duty Cycle measurement function.
To avoid personal injury or damage to the Meter, do not apply voltage
higher than 600V.
Auto Range
10
Page 17
Non-Contact Voltage Sensing

1. To avoid personal injury or damage to the Meter, do not test on
un-insulated high voltage wires.
2. Buzzer will sound when detecting voltage higher than AC 90V. Screen
displays“OL”.
3. Do not test on hazardous live wires higher than AC 600V.
0
9
11
Page 18
Battery Test
Applying a voltage source or incorrect battery type under battery test
may cause personal injury or damage to the Meter. Battery 1.5V range is for dry battery not exceeding 2Vdc. The resistance load is around 30Ω. Battery 9V range is for dry battery not exceeding 15Vdc. The resistance load is around 1KΩ.
SPECIFICATION
Ambient temperature: 23°C ±5°C (73.4°F ±9°F) Relative temperature: 75% Accuracy: ±(% of reading + digits) Maximum voltage between input terminal and earth ground: AC 600Vrms or
DC 600V
Fuse for mA μA input: 0.5A H 600V fast-fuse, ( 6×32)mmFuse for 10A input: 10A H 600V fast-fuse, ( 6×32)mm.
Maximum display: Digital 3999 counts, updates 3/sec. Frequency: 4999 counts. Over-range indication: OL Range: Automatic Altitude: Operating 2000m Operating temperature: 0°C ~ +40°C (32°F ~ 104°F)
12
Page 19
Relative humidity: 0°C ~ +30°C (32°F ~ 86°F) 75%; +30°C ~ +40°C (86°F ~ 104°F) 50%
Storage temperature: -10°C ~ +50°C (14°F ~ 122°F) Electromagnetic compatibility: In an RF filed of 1V/m = Specified accuracy ±5% Battery: 9V, 6F22, NEDA1604 or equivalent Low battery indication: Dimensions (L x W x H): 182 mm x 90 mm x 45 mm (7.2 in x 3.5 in x 1.8 in) Weight: Approx. 354g (0.78lb) with batteries installed
1. DC Voltage Measurement
Range Resolution Accuracy
4.000V 1mV
±(0.8%+1dgt)40.00V 10mV
400.0V 100mV
600V 1V ±(1.0%+3dgt)
Input impedance: around 10MΩ; (Input impedance > 3GΩ for DC 400mV range)
Overload protection: ±600V
2. AC Voltage Measurement
Range Resolution Accuracy
400.0mV 0.1mV ±(1.2%+3dgt)
4.000V 1mV
±(1.0%+3dgt)40.00V 10mV
400.0V 100mV
600V 1V ±(1.2%+3dgt)
Note: 400.0mV range is available for manual range only. Input impedance: around 10MΩ Frequency response: 45Hz ~ 400Hz
Average sensing, rms indication. Overload protection: 600Vrms
13
Page 20
3. Resistance Measurement
Range Resolution Accuracy
400.0Ω 0.1Ω ±(1.2%+2dgt)
4.000kΩ 1Ω
±(1.0%+2dgt)40.00kΩ 10Ω
400.0kΩ 100Ω
4.000MΩ 1kΩ ±(1.2%+2dgt)
40.00MΩ 10kΩ ±(1.5%+5dgt)
400Ω range: Measured value = (Measured display value) – (Short-circuiting value of probe)
Open circuit voltage: around 0.5V Overload protection: 600Vrms
4. : Continuity : Diode measurement
Range Resolution Accuracy
Open circuit voltage is around 0.5V.
0.1Ω
Overload protection: 600Vrms
1mV
Resistance >150Ω, buzzer will not sound. Resistance 10Ω, buzzer will sound.
Open-circuit voltage is around 1.5V. Normal voltage is around 0.5V to 0.8V for silicon PN junction.
5. Capacitance Measurement
Range Resolution Accuracy
40.00nF 10pF ±(3%+10dgt) under REL status
400.0nF 100pF
4.000uF 1nF
40.00uF 10nF ±(3%+5dgt)
100.0uF 100nF ±(4%+5dgt)
Overload protection: 600Vrms
±(3%+5dgt) under REL status
14
Page 21
6. Measurement of frequency/duty cycle
Range Resolution Accuracy
10Hz~10MHz 0.01Hz~0. 01MHz ±(0.1%+4dgt)
0.1%~99.9% 0.1% --
Overload protection: 600Vrm Input amplitude: (DC level is 0.) 10Hz~1MHz: 300mV a 30Vrms >1MHz~10MHz: 600mV a 30Vrms Input amplitude and frequency response must meet following conditions when reading frequency or duty cycle during AC voltage or current measurement
• Input amplitude Range × 30%
• Frequency response: 400Hz
7. Battery Test
Range Internal load resistance Accuracy
1.5V About 30Ω
9V About 1kΩ
Overload protection:
±(1.0%+3dgt)
F1 fuse, 0.5A H 600V fast-fuse, ( 6×32)mm For 1.5V range: Load resistance is around 30Ω. For 9V range: Load resistance is around 1kΩ
8. DC Current Measurement
Range Resolution Accuracy
400.0μA 0.1μA
μA
mA
4000μA 1μA
40.00mA 10μA
400.0mA 0.1mA
4.000A 1mA
A
10.00A 10mA
±(1.0%+2dgt)
±(1.2%+3dgt)
15
Page 22
Overload protection:
mA /μA input:
F1 fuse, 0.5A H 600V fast-fuse, ( 6×32)mm
10 A input:
F2 fuse, 10A H 600V fast-fuse, ( 6×32)mm
9. AC Current Measurement
Range Resolution Accuracy
400.0μA 0.1μA
μA
mA
Frequency response: 45Hz ~ 400Hz Average sensing, rms indication Overload protection:
4000μA 1μA
40.00mA 10μA
400.0mA 0.1mA
4.000A 1mA
A
10.00A 10mA
mA /μA input:
F1 fuse, 0.5A H 600V fast-fuse, ( 6×32)mm
10 A input:
F2 fuse, 10A H 600V fast-fuse, ( 6×32)mm
±(1.2%+2dgt)
±(1.5%+3dgt)
16
Page 23
MAINTENANCE AND REPAIR
If the Meter fails to operate, check battery, test leads, etc., and replace as necessary. Double check the followings:
1. Replace the fuse or battery if the meter does not work.
2. Review the operating instructions for possible mistakes in operating procedure.
Quick check on 0.5A FUSE:
Step 1: Turn the rotary switch to Ω function. Step 2: Connect test lead to /V/Ω/Hz terminal and mA/μA terminal.
Resistance reading 10MΩ: the fuse is OK Resistance reading “OL”: the fuse is open. Replace the fuse as specified.
Quick check on 10A FUSE:
Step 1: Turn the rotary switch to Ω function. Step 2: Connect test lead to /V/Ω/Hz terminal and mA/μA terminal.
Resistance reading 0.5Ω: the fuse is OK. Resistance reading “OL”: the fuse is open. Replace the fuse as specified.
Auto Range /Manual Range
17
Page 24
Except for the replacement of the battery, repair of the meter should be performed only by a Factory Authorized Service Center or by other qualified instrument service personnel.
The front panel and case can be cleaned with a mild solution of detergent and water. Apply sparingly with a soft cloth and allow to dry completely before using. Do not use aromatic hydrocarbons, Gasoline or chlorinated solvents for cleaning.
BATTERY AND FUSE REPLACEMENT
WARNING
To avoid shock, injury, or damage to the Meter: Disconnect test leads before opening case. Use ONLY fuses with the amperage, interrupt, voltage, and speed ratings specied.
Replacing BATTERY follow below steps:
1. Disconnect the test lead probe from measuring circuit.
2. Turn the Meter to OFF position.
3. Remove the screws from the battery cover and open the battery cover
4. Remove the batteries and replace with one 9V (6F22) or equivalent. The battery cover provides the correct polarity fitting construction design. Install the battery in the battery cover.
5. Put the battery cover back and re-fasten the screw.
Battery: 9V (6F22) Battery or equivalent
Replacing FUSE follow below steps:
1. Disconnect the test lead probe from measuring circuit.
2. Turn the Meter to OFF position.
3. Remove the screws from the enclosure and open the enclosure.
4. Remove the broken fuse and replace with new specified fuse.
5. Put the enclosure back and re-fasten the screw.
Fuse ratings: mA /μA input terminal:
F1 fuse, 0.5A H 600V fast-fuse, (
10 A input terminal:
F2 fuse, 10A H 600V fast-fuse, ( 6×32)mm
6×32)mm
18
Page 25
9V Battery
19
Page 26
Loading...