Your Amprobe product will be free from defects in material and workmanship for 1 year from the
date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident,
neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers
are not authorized to extend any other warranty on Amprobe’s behalf. To obtain service during the
warranty period, return the product with proof of purchase to an authorized Amprobe Test Tools
Service Center or to an Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY
IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES - WHETHER EXPRESS, IMPLIED OR STAUTORY INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY,
ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY.
Since some states or countries do not allow the exclusion or limitation of an implied warranty or of
incidental or consequential damages, this limitation of liability may not apply to you.
Repair
All test tools returned for warranty or non-warranty repair or for calibration should be accompanied
by the following: your name, company’s name, address, telephone number, and proof of purchase.
Additionally, please include a brief description of the problem or the service requested and include
the test leads with the meter. Non-warranty repair or replacement charges should be remitted in the
form of a check, a money order, credit card with expiration date, or a purchase order made payable to
Amprobe® Test Tools.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During the
warranty period any defective test tool can be returned to your Amprobe® Test Tools distributor for
an exchange for the same or like product. Please check the “Where to Buy” section on www.amprobe.
com for a list of distributors near you. Additionally, in the United States and Canada In-Warranty repair
and replacement units can also be sent to a Amprobe® Test Tools Service Center (see next page for
address).
Non-Warranty Repairs and Replacement – US and Canada
Non-warranty repairs in the United States and Canada should be sent to a Amprobe® Test Tools
Service Center. Call Amprobe® Test Tools or inquire at your point of purchase for current repair and
replacement rates.
In USA In Canada
Amprobe Test Tools Amprobe Test Tools
Everett, WA 98203 Mississauga, ON L4Z 1X9
Tel: 888-993-5853 Tel: 905-890-7600
Fax: 425-446-6390 Fax: 905-890-6866
Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe® Test Tools distributor for a nominal
charge. Please check the “Where to Buy” section on www.amprobe.com for a list of distributors near
you.
Amprobe® Test Tools Europe
In den Engematten 14
79286 Glottertal, Germany
tel: +49 (0) 7684 8009 - 0
*(Correspondence only – no repair or replacement available from this address. European customers
please contact your distributor.)customers please contact your distributor.)
ACD-45PQ 600A Power Quality Clamp
1
Transformer Clamp Jaws for AC
1
22
10
3
9
4
8
4
7
56
current magnetic field pick up
Jaw marking lines for ACA (&
2
thus Power) position error
indication
Hand/Finger Barrier to indicate
3
the limits of safe access to
the jaws during current
measurements
Push-buttons for special
4
functions & features
Input Jack for all functions
5
EXCEPT non-invasive ACA
current (& thus Power) function
Common (Ground reference)
6
Input Jack for all functions
EXCEPT non-invasive ACA
current (& thus Power) function
Do not dispose of this product as unsorted municipal waste.
Contact a qualified recycler for disposal
Safety Information
This manual contains information and warnings that must be followed for
operating the instrument safely and maintaining the instrument in a safe
operating condition. If the instrument is used in a manner not specified by the
manufacturer, the protection provided by the instrument may be impaired.
The meter is intended only for indoor use.
The meter protection rating, against the users, is double insulation per
IEC61010-1 2nd Ed., EN61010-1 2nd Ed., UL61010-1 2nd Ed. and CAN/CSA C22.2
No. 61010.1-0.92 to Category III 1000 Volts AC & DC and Category IV 600 Volts
AC & DC.
Per IEC61010-1 2nd Ed. (2001) Measurement Category
Measurement Category IV (CAT IV) is for measurements performed at the source
of the low-voltage installation. Examples are electricity meters and measurements
on primary overcurrent protection devices and ripple control units.
1
Measurement Category III (CAT III) is for measurements performed in the
building installation. Examples are measurements on distribution boards,
circuit- breakers, wiring, including cables, bus-bars, junction boxes, switches,
socket-outlets in the fixed installation, and equipment for industrial use
and some other equipment, for example, stationary motors with permanent
connection to the fixed installation.
Measurement Category II (CAT II) is for measurements performed on circuits
directly connected to the low voltage installation. Examples are measurements
on household appliances, portable tools and similar equipment.
Terms In This Manual
WARNING - identifies conditions and actions that could result in
serious injury or even death to the user.
CAUTION - identifies conditions and actions that could cause damage
or malfunction in the instrument.
WARNING - To reduce the risk of fire or electric shock, do not expose
this product to rain or moisture.
To avoid electrical shock hazard, observe the proper safety precautions
when working with voltages above 60 VDC or 30 VAC rms. These voltage
levels pose a potential shock hazard to the user.
Keep your fingers behind the finger guards of the test leads during
measurement. Inspect test leads, connectors, and probes for damaged
insulation or exposed metal before using the instrument. If any defects
are found, replace them immediately.
This Clamp-on meter is designed to apply around or remove from
uninsulated hazardous live conductors. But still, individual protective
equipment must be used if hazardous live parts in the installation where
measurement is to be carried out could be accessible.
CAUTION - Disconnect the test leads from the test points before
changing meter functions. Only use the accompanied test leads, or
replace with the same rating or better.
CENELEC Directives
The instruments conform to CENELEC Low-voltage directive 2006/95/EC and
Electromagnetic compatibility directive 2004/108/EC
2
UNPACKING AND INSPECTION
Your shipping carton should include:
1 ACD-45PQ 600A Power Quality Clamp
1 Test Leads (Black x 1; Red x 1)
1 Bag
1 Users Manual
2 1.5V AAA LR03/NEDA 24A Batteries (Installed)
If any of the items are damaged or missing, return the complete package to
the place of purchase for an exchange.
OPERATION
CAUTION
Before and after hazardous voltage measurements, test the voltage
function on a known source such as line voltage to determine proper
meter functioning.
3
AutoVATM Function
Set the slide-switch function-selector to the position.
• With no input, the meter displays “ ” when it is ready.
• With no ACA current input via the jaws but a voltage signal above the
nominal threshold of DC 2.4V or AC 30V (40Hz ~ 500Hz) up to the rated
600V is present on V-COM terminals, the meter displays the voltage
value in appropriate DC or AC, whichever larger in peak magnitude.
Annunciators “Auto” “dc” and “Auto” “
• On the contrary, with no voltage signal present on V-COM terminals
but a ACA current signal above the nominal threshold of AC 1A (40Hz ~
500Hz) up to the rated 600A is input via the jaws, the meter displays the
ACA current value. Annunciators “
• The Auto-VA feature stays at the auto-selected function as long as
its signal remains above the specified threshold. Press SELECT button
momentarily to manually select and lock (annunciator “Auto” turns off)
thru the functions ACA, ACV, DCV and then goes back to Auto-VA.
CAUTION
• For non-invasive ACA current measurements, press the jaw trigger
and clamp the jaws around only one single conductor of a circuit
for load current measurement. Make sure the jaws are completely
closed, or else it will introduce measurement errors. Enclosing more
than one conductor of a circuit will result in differential current (like
identifying leakage current) measurement. Locate the conductor(s) at
the Jaws center as much as possible to get the best measuring accuracy.
For removal, press the jaw trigger and remove the jaws from the
conductor(s).
• Adjacent current-carrying devices such as transformers, motors and
conductor wires will affect measurement accuracy. Keep the jaws away
from them as much as possible to minimize influence.
” turn on respectively.
” and “Auto” turn on accordingly.
4
Peak-rms Mode
Peak-rms compares and displays the maximum RMS value of surge voltage
or current with durations as short as 65ms. When ACV or ACA function is autoselected or manual-selected, press and hold Peak-rms
or more toggles to this mode. The annunciators “P-” “Max” turn on. APO
(Auto Power Off) feature is disabled automatically accordingly.
In ACA function, the Peak-rms mode starts at the highest 600A range to
maximize measuring dynamic range. Before making measurement, press
the Peak-rms
measuring dynamic range 400.0A or 40.00A for higher measuring resolutions.
button momentarily again can manually select thru lower
button for one second
Line-level Frequency (Hz) Function
When ACV or ACA function is auto-selected or manual-selected, press Hz
button momentarily toggles to Line-level Frequency (Hz) function. The Hz
trigger level is determined by the selected function-range from where the Hz
function is activated.
5
In ACA function, activating the Hz function during significant measurements
can get the most appropriate trigger level to avoid electrical noises in
most cases. Activating the Hz function at AC 40.00A range (before making
significant measurements) can get the lowest trigger level (highest sensitivity).
HOLD Mode
When any function is auto-selected or manual-selected, press HOLD button
momentarily toggles to Hold mode. The annunciator “
mode freezes the display for later viewing.
Notes on Displacement Power Factor & Total Power Factor
• Introduction: Power is the rate of change of energy with respect to time
(in terms of voltage V and current A). Instantaneous (real) power w = vi
where v is the instantaneous voltage and i the instantaneous current.
The average (real) power is the mean of vi and is given by:
W = ω/2π∫vi dt , over the interval from 0 to 2π/ω
• Displacement Power Factor (more traditional): Assuming V and A are
pure sinusoidal waveforms without harmonics (as in most traditional
cases), that is, v = V sinωt and i = I sin (ωt -θ), the expression can be
simplified to:
W = 1/2 x V x I x Cosθ where V and I are the peak values, θ is the
displacement power factor angle, and Cosθ is the displacement power
factor. Using RMS values, it is written as:
W = Vrms x Arms x Cosθ
Practically, in such cases without harmonics, θ is also called the phaseshift angle of the current A to the voltage V. An inductive circuit is said
to have a lagging power factor since current A lags voltage V (phaseshift angle θ and thus Sinθ are both “+”), and a capacitive circuit is said
to have a leading power factor since current A leads voltage V (phaseshift angle θ and thus Sinθ are both “-”).
• Total Power Factor (encountering harmonics): When encountering
distorted waveforms with the presence of harmonics, however, the
simplified power expression should not be used since substituting the
above mentioned pure sinusoidal V and A functions cannot fulfill the
actual conditions. Cosine of phase-shift angle (Cosθ), or the displacement
” turns on. Hold
6
power factor, is no longer the only component constituting the overall
power factor. Harmonics do increase apparent power and thus decrease
the overall power factor.
That is the Total Power Factor is actually affected by both phase-shift
angle and harmonics, and is given by the expression:
Total Power Factor (PF) = Real Power (W) / Apparent Power (VA)
In order to improve overall system power factor, nowadays power-system
engineer needs to address both phase-shift and harmonics problems.
Practically, harmonics should be dealt with (e.g. filtering out) before
phase-shift to be corrected (e.g. installing capacitors in parallel with
inductive loads).
7
Single-Phase Power & 3-Phase Balanced-Load Power Functions
Set the slide-switch function-selector to the “3~Bal • 1~” Power position.
• Default at last selected function.
• Press “3~Bal • 1~” button momentarily to toggle between “Single-
Phase” and “3-Phase Balanced Load” Power functions. Annunciators “
” and “3~” turn on respectively.
• Press SELECT button momentarily selects between W (real power), VAR
(reactive power), VA (apparent power) & kWHr (real-time readings or
stored result) functions. In W (real power), VAR (reactive power), or VA
(apparent power) function:
1. PF (Total Power Factor) is displayed automatically in the secondary mini
display.
2. Annunciator “A-lags-V” turns on to indicate an inductive circuit is
being measured. That is, the Current waveform is lagging the Voltage
waveform, and the phase-shift angle θ is “+”.
3. On the contrary, together with significant PF values, WITHOUT
“A-lags-V” being turned on indicates that a capacitive circuit is being
measured. That is, the Current waveform is leading the Voltage
waveform, and the phase-shift angle θ is “-”.
8
Note:
1. Under proper measurement setups for load circuits, the W (real power)
readings are always positive. Negative W readings indicate reversed
clamp-on jaws direction or test leads polarities, or even incorrect voltage
lines are being measured as in 3-phase measurement setups. Correct
them for proper “A-lags-V” indications.
2. When encountering largely distorted waveforms, “A-lags-V” detection
might be affected due to the influence of harmonics. It is recommended
to manage (e.g. filter out) harmonics problems before measuring/
dealing with phase-shift problems.
kWHr (kilo-Watt-Hour) Recording function
9
Set the slide-switch function-selector to the “3~Bal • 1~” Power position. Setup
power measurements as mentioned in the previous “Single-Phase Power & 3-Phase Balanced-Load Power functions” section
• To start (“
buttons at the same time. Annunciator “
accumulated time (in Hour) is displayed automatically in the secondary
mini display.
• To pause (“
stops flashing and is always on.
• To continue (“
Annunciator “
• To stop (“”), press the “3~Bal • 1~” and “HOLD” buttons at the
same time again. Annunciator “
is then displayed on the LCD for immediate viewing. Annunciator “
turns on & flashes.
• When the low battery annunciator “
stop (“
Recording result as in above.
Note:
• During kWHr Recording session, real-time W, VAR, VA as well as kWHr
accumulated readings can be selected by pressing the SELECT button
momentarily. A flashing “
going. An always on “
• When kWHr Recording is not activated, kWHr stored result instead
of accumulated readings is displayed when selected as in above.
Annunciator “
• The meter separately stores one Single-Phase and one 3-Phase-Balanced-
Load kWHr result for later viewing. When they are being viewed, press
“3~Bal • 1~” button momentarily to toggle between them.
• When the display readings exceed 9999kWHr/999hours, exponential
readings are displayed. “2.3E4” kWHr represents 2.3 x 104 kWHr, or
23000 kWHr for example,.
”) kWHr recording session automatically and display kWHr
” denotes that kWHr Recording is still under-
” denotes that kWHr Recording is being paused.
” turns on & flashes.
10
• After the kWHr Recording session is stopped (“”) properly, the new
result will supersede the previous one stored in the non-volatile memory.
You can then switch off the meter for transportation, storage, or even
battery changing with memory remained.
• To avoid mis-storage to memory, it is important to properly stop (“
kWHr Recording session before sliding the slide-switch function-selector
to any other function positions.
”)
3-Phase 3-Wire (3
• Set the slide-switch function-selector to the “3~Un-Bal” Power position.
Press “3W•4W” button momentarily to select 3-Wire measurements.
Annunciator “3W” turns on.
• Clamp the jaws around “Line 1” as reminded by annunciators “
and connect Black test probe (COM terminal) to “Line 3” and Red test
probe (+ terminal) to “Line 1” as reminded by annunciators “
mini-display.
• When the reading is stable, press “
first measuring value.
• Then clamp the jaws around “Line 2” as reminded by annunciators “L2”, and connect Black test probe (COM terminal) to “Line 3” and Red
test probe (+ terminal) to “Line 2” as reminded by annunciators “
” on mini-display.
• When the reading is stable, press “ ” button momentarily to enter the
second measuring value. The meter will then calculate, store and display
the total 3-Phase Power result automatically. Annunciators “
turn on.
• Press “
• Press “SELECT” button momentarily to view (“”) the last stored
result. Annunciator “
momentarily again to continue (“
3W) Unbalanced-Load Power Function
” button momentarily to enter the
” button momentarily again for new measurements.
” turns on & flashes. Press “SELECT” button
”).
11
L1”,
” on
L1 L2 L3”
3-Phase 4-Wire (34W) Unbalanced-Load Power Function
• Set the slide-switch function-selector to the “3~Un-Bal” Power position.
Press “3W•4W” button momentarily to select 4-Wire measurements.
Annunciator “4W” turns on.
• Clamp the jaws around “Line 1” as reminded by annunciators “
and connect Black test probe (COM terminal) to “Line n (neutral)” and
Red test probe (+ terminal) to “Line 1” as reminded by annunciators “
” on mini-display.
• When the reading is stable, press “ ” button momentarily to enter the
first measuring value.
• Then clamp the jaws around “Line 2” as reminded by annunciators “L2”, and connect Black test probe (COM terminal) to “Line n (neutral)”
and Red test probe (+ terminal) to “Line 2” as reminded by annunciators
” on mini-display.
“
• When the reading is stable, press “ ” button momentarily to enter the
second measuring value.
12
L1”,
• Then clamp the jaws around “Line 3” as reminded by annunciators “
L3”, and connect Black test probe (COM terminal) to “Line n (neutral)”
and Red test probe (+ terminal) to “Line 3” as reminded by annunciators
” on mini-display.
“
• When the reading is stable, press “ ” button momentarily to enter the
third measuring value. The meter will then calculate, store and display
the total 3-Phase Power result automatically. Annunciators “
turn on.
• Press “
• Press “SELECT” button momentarily to view (“”) the last stored
result. Annunciator “
momentarily again to continue (“
” button momentarily again for new measurements.
” turns on & flashes. Press “SELECT” button
”).
L1 L2 L3”
13
/ Functions
Set the slide-switch function-selector to the / function position. Default
at last selected function. Press SELECT button to toggle between and
measurement functions.
Backlighted Display
Press the SELECT button for 1 second or more to toggle the display backlight
on or off.
Auto Power Off (APO)
The meter turns off after approximately 30 minutes of neither switch nor
button activity. To wake up the meter from APO, press SELECT button or slide
the function-selector to OFF position and back on again. Always turn the
function-selector to OFF when the meter is not in use.
14
Auto Power Off Quick Test
Press-and-hold the 3W.4W button while powering the meter on. The LCD
displays “
released. Quick test APO timing is 10 seconds after such activation.
” & “” to confirm activation right after the 3W.4W button is
Disabling Auto-Power-Off (APO)
Press-and-hold the HOLD button while powering the meter on. The LCD
displays “
released.
” & “” to confirm activation right after the HOLD button is
Line Frequency Setup
Press-and-hold the Hz button while powering the meter on. LCD displays the
last 50Hz or 60 Hz setup. Press SELECT button momentarily to select 50Hz or
60Hz to cope with your local line frequency. Press Hz button for one second to
store your selection and resume measurements.
PC Computer Interface Capabilities
The instrument equips with an optical isolated data output port at the bottom
case near the battery compartment. Optional purchase PC interface kit USBKIT2 is required to connect the meter to PC computer USB port. The Data
Recording System software equips with a digital meter, an analog meter, a
comparator meter, and a Data Graphical recorder. Refer to the README file
comes with the interface kit for further details.
15
SPECIFICATIONS
General Specications
Display : Voltage functions: 6000 counts LCD display
Special features : Backlighted display; AutoVATM (Auto Selection
on ACV, DCV or ACA functions); selectable
Power parameters of W, VAR & VA with Total
Power Factor in dual-display; kWHr Recording;
Display Hold; PEAK-rms HOLD
Accessories : Test leads (pair), batteries installed, user’s
manual & soft carrying pouch
Electrical Specications
Accuracy is ± (% reading digits + number of digits) or otherwise specified, at
23 ˚C ±5 ˚C & less than 75% R.H.
17
True RMS ACV & ACA clamp-on accuracies are specified from 0% to 100% of
range or otherwise specified. Maximum Crest Factor are as specified below,
and with frequency spectrums, besides fundamentals, fall within the meter
specified AC bandwidth for non-sinusoidal waveforms. Fundamentals are
specified at 50Hz and 60Hz.
ACA AutoVATM Threshold: 1A AC (40Hz ~ 500Hz only) nominal
Crest Factor:
< 3 : 1 at full scale & < 6 : 1 at half scale for 40.00A, 400.0A & 600A ranges
1. Induced error from adjacent current-carrying conductor: < 0.06A/A
2. Specified accuracy is from 1% to 100% of range and for measurements
made at the jaw center. When the conductor is not positioned at the jaw
center, position errors introduced are:
• Add 1% to specied accuracy for measurements made WITHIN jaw
marking lines (away from jaw opening)
• Add 4% to specied accuracy for measurements made BEYOND jaw
marking lines (toward jaws opening)
PEAK-rms HOLD (ACA & ACV only)
Response: 65ms to >90%
Frequency
RangeAccuracy
5Hz ~ 500Hz0.5%+4d
Sensitivity (Sine RMS)
40A range: > 4A
400A range: > 40A
600A: > 400A
600V range: > 30V
19
Ohms
RangeAccuracy
5Hz ~ 500Hz0.5%+4d
Open Circuit Voltage: 0.4VDC typical
Audible Continuity Tester
Audible threshold: between 10 and 300.
Response time: 250µs
Single-Phase & 3-Phase Balanced-Load Power
RANGE 5) 0 ~
600.0kVA
@ PF = 0.99
~ 0.1
F ~ 10th11th ~ 45th46th ~ 51st
2.0%+6d3.5%+6d5.5%+6d
Accuracy 1) 2) 3)
RANGE 5) 0
~ 600.0kW /
kVAR
@ PF = 0.98 ~
0.70
@ PF = 0.70 ~
0.50
@ PF = 0.50 ~
0.30
@ PF = 0.30 ~
0.20
1. Specified accuracy is for ACA clamp measurement at the center of jaws.
When the conductor is not positioned at the jaw center, position errors
introduced are:
• Add 1% to specied accuracy for ACA measurements made WITHIN
jaw marking lines (away from jaw opening)
• Accuracy is not specied for ACA measurement made BEYOND jaw
marking lines (toward jaws opening)
F ~ 10th11th ~ 25th26th ~ 45th46th ~ 51st
2.0%+6d
3.0%+6d
4.5%+6d
10%+6d15%+6d
Accuracy 1) 2) 4)
3.5%+6d
4.5%+6d10%+6d
20
2. Add 4d to specified accuracy for 3-Phase Balanced-Load Power
measurements.
3. Add 1% to specified accuracy @ ACA fundamental < 6A or ACV
fundamental < 90V. Accuracy is not specified @ ACA fundamental < 1A
or ACV fundamental < 30V
4. Add 1% to specified accuracy @ ACA fundamental < 6A or ACV
fundamental < 90V. Accuracy is not specified @ ACA fundamental < 2A
or ACV fundamental < 50V
5. 0 ~ 360.0
Total Power Factor (PF)
RangeAccuracy 1)
0.10 ~ 0.99
1) Specified accuracy @ ACA fundamental > 2A ; ACV fundamental > 50V
F ~ 21st22nd ~ 51st
3d5d
A-lags-V Indication:
LCD annunciator “A-lags-V” turns on to indicate an inductive circuit, or
Current A lags Voltage V (i.e., phase-shift angleθ is “+”).
A-lags-V Indication is specified at 50/60Hz fundamental without the presence
of harmonics, and at ACV > 90V, ACA > 9A and PF < 0.95
kWHr (kilo-Watt-Hour Energy)
Time base accuracy: < 30ppm
Non-volatile memory: Separately stores one 3-Phase-Balanced-Load and one
Single-Phase result
3-Phase Unbalanced-Load Power
This 3-Phase Unbalanced-Load Power measurement is achieved thru the
calculation of discrete single-phase measurements that are taken one at a time
manually. Since it is not real-time on all 3 phases simultaneously, it is intended
only for stable power conditions without significant power fluctuations over
the time of measurements. Result accuracy is hence the accumulated accuracy
of the discrete single-phase measurements plus the associated fluctuations.
21
MAINTENACE AND REPAIR
WARNING
To avoid electrical shock, disconnect the meter from any circuit, remove
the test leads from the input jacks and turn OFF the meter before opening
the case. Do not operate with open case.
Trouble Shooting
If the instrument fails to operate, check batteries and test leads etc., and
replace as necessary. Double check operating procedure as described in this
user’s manual
If the instrument voltage-resistance input terminal has subjected to high
voltage transient (caused by lightning or switching surge to the system) by
accident or abnormal conditions of operation, the series fusible resistors will
be blown off (become high impedance) like fuses to protect the user and the
instrument. Most measuring functions through this terminal will then be open
circuit. The series fusible resistors and the spark gaps should then be replaced
by qualified technician. Refer to the LIMITED WARRANTY section for obtaining
warranty or repairing service.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use
abrasives or solvents. If the meter is not to be used for periods of longer than
60 days, remove the batteries and store them separately
22
Battery Replacement
The meter uses standard 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
Loosen the 2 captive screws from the battery cover case. Lift the battery cover
case. Replace the batteries. Replace battery cover case. Re-fasten the screws.
23
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.