Amprobe ACD-14 Users Manual

INSTRUCTION MANUAL

ACD-14 TECH CLAMP

AMPROBE

®
1
Congratulations! Your new instrument has been crafted according to quality standards and contains quality components and workmanship. It has been inspected for proper operation of all of its functions and tested by qualified factory technicians according to the long-established standards of AMPROBE.
1) SAFETY
This manual contains information and warnings that must be followed for operating the instrument safely and maintaining the instrument in a safe operating condition. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired.
This meter meets the requirements for double insulation to IEC61010-2-032(1994), EN61010-2-032(1995), UL3111-2-032(1999): Category III 600 Volts ac and dc.
OVERVOLTAGE INSTALLATION CATEGORIES, PER IEC61010:
OVERVOLTAGE CATEGORY II: Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III: Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV: Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. Note – Examples include electricity meters and primary over-current protection equipment.
TERMS IN THIS MANUAL
WARNING: Identifies conditions and actions that could result in serious injury or
even death to the user.
CAUTION: Identifies conditions and actions that could cause damage or
malfunction in the instrument.
2
WARNING: To reduce the risk of fire or electric shock, do not expose this product to rain or moisture. The meter is intended only for indoor use.
To avoid electrical shock hazard, observe the proper safety precautions when working with voltages above 60 VDC or 30 VAC rms. These voltage levels pose a potential shock hazard to the user.
Inspect test leads, connectors, and probes for damaged insulation or exposed metal before using the instrument. If any defects are found, replace them immediately.
Do not touch test lead tips or the circuit being tested while power is applied to the circuit being measured. To avoid accidental short circuit of bare (un-insulated) hazardous live conductors or bus bars, switch them off before insertion and removal of the current clamp jaws. Contact with the conductor could result in electric shock.
Keep your hands/fingers behind the hand/finger barriers that indicate the limits of safe access of the meter and the test leads during measurement.
CAUTION: Disconnect the test leads from the test points before changing meter functions.
INTERNATIONAL ELECTRICAL SYMBOLS
Caution! Refer to the explanation in this Manual
Caution! Risk of electric shock
Earth (Ground)
Double Insulation or Reinforced insulation
Fuse
AC--Alternating Current
DC--Direct Current
2) CENELEC Directives
This instrument conforms to CENELEC Low-voltage directive 73/23/EEC and Electromagnetic compatibility directive 89/336/EEC
3) PRODUCT DESCRIPTION
Note: Please refer to Specification Section for function details.
3
10
1) Clamp Jaw for AC current magnetic field
1
pick up
2) Hand/Finger Barrier to indicate the limits of safe access of the meter during measurement
2
3) Power ON/OFF buttons for ACA current function and MAX/Hold features
4) Push-button for frequency selection
5) Input Jack for all functions EXCEPT
3
clamp-on ACA current function
8
7
9
6
6) Common (Ground reference) Input Jack for all functions EXCEPT clamp-on ACA current function
7) Slide-switch Selector to turn the lower display ON/OFF and to Select functions
8) Lower Display: for Input Jack functions 3-3/4 digits 4000 counts LCD
4
9) Upper Display: for ACA current Clamp-on 3-3/4 digits 4000 counts LCD
5
10) Jaw trigger for opening the ACA current clamp jaw
4) OPERATION
4
DC Voltage, AC Voltage, Hz Frequency functions
Inputs are made through the test lead terminals. Move the slide-switch to the “V” position. The unit defaults at DC voltage. Press SELECT button momentarily to select AC voltage.
To activate the Hz Frequency function, press the Hz button momentarily.
Note: 400.0mV range is designed with 1000MΩ high input impedance for least current drain in measuring small signals, and can cope better with most commercially available voltage output transducers/adapters. The non-zero display reading is normal when the meter inputs are open circuit, which will not affect actual measurement accuracy. The