AMETEK Programmable Power, Inc., a Division of AMETEK, Inc., is a global leader in the design
and manufacture of precision, programmable power supplies for R&D, test and measurement,
process control, power bus simulation and power conditioning applications across diverse
industrial segments. From bench top supplies to rack-mounted industrial power subsystems,
AMETEK Programmable Power is the proud manufacturer of Elgar, Sorensen, California
Instruments and Power Ten brand power supplies.
AMETEK, Inc. is a leading global manufacturer of electronic instruments and electromechanical
devices with annualized sales of $2.5 billion. The Company has over 11,000 colleagues working
at more than 80 manufacturing facilities and more than 80 sales and service centers in the United
States and around the world.
Trademarks
AMETEK is a registered trademark of AMETEK, Inc.
Other trademarks, registered trademarks, and product names are the property of their respective
owners and are used herein for identification purposes only.
UNLESS SPECIFICALLY AGREED TO IN WRITING, AMETEK PROGRAMMABLE POWER,
INC. (“AMETEK”):
(a) MAKES NO WARRANTY AS TO THE ACCURACY, SUFFICIENCY OR SUITABILITY OF
ANY TECHNICAL OR OTHER INFORMATI ON PROVIDED IN ITS MANUALS OR
OTHER DOCUMENTATION.
(b) ASSUMES NO RESPONSIBILITY OR LIABILITY FOR LOSSES, DAMAGES, COSTS OR
EXPENSES, WHETHER SPECIAL, DIRECT, INDIRECT, CONSEQ UENTIAL OR
INCIDENTAL, WHICH MIGHT ARISE OUT OF THE USE OF SUCH INFORMATION. THE
USE OF ANY SUCH INFORMATION WILL BE ENTIRELY AT THE USER’S RISK, AND
(c) RE MINDS YOU THAT IF THIS MANUAL IS IN ANY LANGUAGE OTHER THAN
ENGLISH, ALTHOUGH STEPS HAVE BEEN TAKEN TO MAINTAIN THE ACCURACY OF
THE TRANSLATION, THE ACCURACY CANNOT BE GUARANTEED. APPROVED
AMETEK CONTENT IS CONTAINED WITH THE ENGLISH LANGUAGE VERSION,
WHICH IS POSTED AT WWW.PROGRAMMABLEPOWER.COM
Date and Revision
February 2012 Revision A
Part Number
M370499-01
Contact Information
Telephone: 800 733 5427 (toll free in North America)
858 450 0085 (direct)
The XDL series II provides the highest performance levels available in a laboratory power supply.
Voltage and current are controlled using 16 bit DACs enabling voltages to be set to 1mV resolution
even at full output. Indeed, the accuracy is sufficient for the PSU to be used as a calibration
source for some hand-held DMMs.
The XDL series II uses pure linear technology and offers unrivalled performance in terms of
regulation, output noise and dynamics. Line and load regulation are close to the limit of
measurement. Output noise is less than 350µV rms in CV mode and down to 20µA rms in CI
mode. Recovery time from transient current pulses is better than 50µs.
It provides full remote sense capability via dedicated sense terminals. Remote sense is essential
to maintain precise regulation at the load. When remote sense is not required, internal local
sensing can be selected at the touch of a button.
Multiple Ranges for Greater Flexibility
The XDL series II provides multiple ranges for increased current capability at lower voltages. The
main range offers 0 to 35 Volts at up to 3 Amps (XDL 35-5) or 0 to 56 Volts at up to 2 Amps (XDL
56-4). The higher current range provides up to 5 Amps for voltages up to 15V (XDL 35-5) or 4
Amps for voltages up to 25V (XDL 56-4). A further low current range provides enhanced current
setting and measurement resolution of 0·1mA.
The product of voltage and current can be displayed at any time by pressing the VxA button. The
power is displayed to a resolution of 0·01 Watts.
Introduction
Fast, Simple and Safe to use
The user interface of the XDL series II has been carefully designed to provide rapid control whilst
guarding against any possibility of error.
Voltage and current setting can be performed either by direct numeric entry or, for applications
where the voltage or current must be gradually changed, by using the quasi-analogue Jog control.
To enable the current limit to be set before connecting the load, the limit setting is displayed when
the output is off. Pressing the View Limits key at any time provides a temporary display of the limit
values allowing precise adjustment to also be made with the output on.
Setting Memories for Added Convenience
The XDL series II provides storage of up to 50 power supply sets-ups in non-volatile memory for
each main output, plus (T models only) a further 50 set-ups for linked mode operation, plus 10 setups for the auxiliary output. Upon mains switch-off, the set -up of the PSU is saved and is
automatically restored at switch-on.
OVP and OCP Trips with 'Alarm' Output
The XDL series II provides fully adjustable over-voltage and over-current trips which can be used
both as a fail-safe against accidental mis-setting and as a protection against inappropriate load
conditions. In addition to turning the output off, a trip condition switches the rear panel alarm signal
enabling other equipment to be controlled.
For complete protection of the power supply, the trip will also be operated by over-temperature or
excess voltage on the sense terminals.
Auxiliary Output with Fully Variable Voltage (T models)
The XDL series II triple output power supplies incorporate an auxiliary output which is fully
variable between 1 volt and 6 volts to a resolution of 0.01V, and has a current capability of 3
amps.
A front panel button enables to voltage and current for the auxiliary output to be viewed on the
Output 1 display whenever required.
4
Fully Programmable via GPIB, RS 232, USB or LAN
The programmable ‘P’ models incorporate a full bus interface permitting remote control and
readback via either GPIB (IEEE-488), RS232, USB or LAN.
The GPIB interface conforms fully with IEEE-488.2 and IEEE-488.1.
The RS232 interface uses a standard 9-pin D-connector and has a Baud rate variable from 600 to
19200.
The USB interface is compatible with USB 2.0 and USB 1.x.
The LAN interface is LXI version 1.2, Class C, compliant.
The XDL series II uses simple and consistent command structures which make programming
particularly easy regardless of which interface is used.
An IVI driver for Windows is included. This provides support for common high-level applications
such as LabView*, LabWindows*, and HP/Agilent VEE*.
All power supply settings can be controlled via the bus. Voltage and curre nt can be set to a
resolution of 1mV or 0·1mA (main outputs). Actual voltage and current can be read back together
with the power supply status.
* LabView LabWindows is a trademark of National Instruments Corp. Agilent VEE is a trademark of Agilent Technologies inc.
Normal mode current: <0·2mArms; <20µArms on 500mA range.
Specification applies for sense lead resistance <0·5Ω.
Current <0·01% + 250µA; <0.01%+ 50µA on 500mA range.
General specifications apply for the temperature range 5°C to 40°C. Accuracy specifications
apply for the temperature range 18°C to 28°C after 1 hour warm-up with no load and calibration at
23°C. Typical specifications are determined by design and are not guaranteed.
MAIN OUTPUTS
Specification
0V to 35V/0·001A to 3A
0V to 35V/0·1mA to 500mA
Voltage Setting: Resolution 1mV
Current Setting: Resolution 1mA; 0·1mA on 500mA range
Output Mode: Constant voltage or constant current with automatic cross-over.
Output Switch: Electronic, non-isolating. Switch illuminated when Output on.
Output T e rminals: Universal 4mm safety binding posts on 19mm (0·75”) pitch for Output;
screwless terminals for Sense.
Transient Response:
Voltage Programming
<50µs to within 15mV of set level for a change in load current from full
Maximum time required for output to settle within 1% of its total excursion
Full Load
No
0V to 56V/0·001A to 2A
0V to 56V/0·1mA to 500mA
Full Load No
Up
Up
Down
Down
Ripple and Noise
Load Regulation: For any load change, measured at the output terminals, using
Line Regulation: Voltage <0·01% + 2mV for 10% line change.
Temperature Coefficient: Voltage: typically <(50ppm + 0·5mV)/°C
15V 5A
35V 3A
15V 5A
35V 3A
Normal mode voltage: <0·35mVrms and 2mVp-p
remote sense.
Voltage <0·01% + 2mV.
Current <0·01% + 250µA; <0.01% +50µA on 500mA range.
Add typically 2·5mV for a 0·5V drop in the positive output lead.
Current: typically <(100ppm + 1mA)/°C;
typically<(100ppm + 0·1mA)/°C on 500mA range.
6ms
20ms
6ms
25ms
6ms
7ms
250ms
600ms
25V/4A
56V/2A
25V/4A
56V/2A
10ms
40ms
10ms
50ms
6ms
15ms
400ms
800ms
6
Response time typically 100µs
Response time typically 35ms
Protection Functions:
Output trips off for OVP, OCP, ove r -temperature and Sense miswiring
Output Protection: Output will withstand forward voltages of up to 20V above rated output
voltage. Reverse protection by diode clamp for currents up to 3A.
Over-voltage Protection:
(OVP)
Over-current Protection:
(OCP)
Range 1V to 40V (XDL 35-5), 1V to 60V (XDL 56-4)
Resolution 0·1V; accur acy ± (2% + 0·5V)
Range 0·01A to 5·5A (XDL 35-5), 0·01A to 4·4A (XDL 56-4)
Resolution 0·01A; accuracy ± (0·2% + 0·01A)
METER SPECIFICATIONS (Main Outputs)
Voltage (CI mode): Resolution 10mV
Current (CV mode): Resolution 0·001A; 0·1mA on 500mA range
V x A: Resolution 0·01W; 0·001W on 500mA range
AUXILIARY OUTPUT (T models only)
Voltage Setting: Resolution: 10mV
Output T e rminals: Universal 4mm safety binding posts on 19mm (0·75”) pitch. Duplicate
Output Protection: Output will withstand up to 16V forward voltage. Diode clamp reverse
Ripple & Noise:
Status Indication: Current limit lamp. Current overload trip indication.
Meter Specifications:
(use SET/VIEW button)
Voltage Programming
<2mV rms, 10mVp-p
Voltage Meter: Resolution 10mV, accuracy ± 0.5% ±10mV
Current Meter: Resolution 10mA, accuracy ± 0.5% ±10mA
Maximum time required for output to settle within 1% of its total
KEYBOARD & ROTARY CONTROL
All functions, including the selection and set-up of the remote control interfaces, can be set from
the keyboard. The rotary jog control can be used to adjust output voltage and current settings in
a quasi-analogue mode.
7
General
RS232:
Standard 9-pin D-connector. Variable Baud rate (600 to 19200).
GPIB:
Conforming with IEEE488.1 and IEEE488.2
USB:
Standard USB 2.0 hardware connection. Operates as a virtual COM port.
MDIX. LXI V1.2, Class C compliant.
change.
Status Indication:
Remote mode and LAN status indicators
Main Outputs
Voltage Setting:
16-bit; Resolution 1mV, accuracy ± (0·03% +5mV)
Resolution 0.01mA, Accuracy ± (0·2% + 0·5mA) on 500mA r ange.
Readback V & I
See meter specifications.
Auxiliary Output (T models only)
Voltage Setting:
Resolution 10mV, accuracy ± 0.5% ±10mV
Current Setting:
Resolution 10mA, accuracy ± 0.5% ±10mA
Readback V & I
See meter specifications
Installation Category II
Power Consumption:
Single output: 250VA max; Triple output: 500VA max.
Operating Range:
+5ºC to +40ºC, 20% to 80% RH
Storage Range:
−40ºC to + 70ºC
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
output if internal temperatures exceed predetermined thresholds.
interfaces.
Safety:
Complies with EN61010−1
EMC:
Complies with EN61326
Triple output: 280 x 160 x 290mm (WxHxD), excl. feet & terminals
ALARM OUTPUT
Isolated rear-panel open-collector output signal. User can select output to be activated for either
OVP, OCP, Overtemperature or Sense miswiring, or for any of those four faults.
DIGITAL INTERFACES (P models only)
Full digital remote control facilities are available through the RS232, USB, LAN and GPIB interfaces.
LAN: Ethernet 100/10base-T hardware connection. Auto speed detect, auto
Remote Command
Processing Time:
Current Setting: 16-bit; Resolution 0.1mA, accuracy ± (0·2% + 5mA)
GENERAL
AC Input: 230V AC or 115V AC ± 10%, 50/60Hz
Typically <25ms between receiving the command terminator for a step
voltage change at the instrument and the output voltage beginning to
Cooling: Intelligent variable-speed fan. Over-temperature trip shuts down
Store/Recall: Up to 50 set-ups each main output, 50 linked set-ups, and 10 auxiliary
output settings can be saved and recalled via the keyboard or remote
Size: Single output: 140 x 160 x 290mm (WxHxD), excl. feet & terminals.
Weight: Single: 5.5kg; Triple: 10·5kg
8
l
mains supply ON.
Safety
This power supply is a Safety Class I instrument according to IEC classification and has been
designed to meet the requirements of EN61010-1 (Safety Requirements for Electrical Equipment
for Measurement, Control and Laboratory Use). It is an Installation Category II instrument
intended for operation from a normal single phase supply.
This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe
condition. This instruction manual contains some information and warnings which have to be
followed by the user to ensure safe operation and to retain the instrument in a safe condition.
This instrument has been designed for indoor use in a Pollution Degree 2 environment in the
temperature range 5°C to 40°C, 20% - 80% RH (non-condensing). It may occasionally be
subjected to temperatures between +5°C and –10°C without degradation of its safety. Do not
operate while condensation is present.
Use of this instrument in a manner not specified by these instructions may impair the safety
protection provided. Do not operate the instrument outside its rated supply voltages or
environmental range.
WARNING! THIS INSTRUMENT MUST BE EARTHED
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
When the instrument is connected to its supply, terminals may be live and opening the covers or
removal of parts (except those to which access can be gained by hand) is likely to expose live
parts. The apparatus shall be disconnected from all voltage sources before it is opened for any
adjustment, replacement, maintenance or repair.
Capacitors inside the power supply may still be charged even if the power supply has been
disconnected from all voltage sources but will be safely discharged about 10 minutes after
switching off power.
Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided
as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of
the hazard involved.
If the instrument is clearly defective, has been subject to mechanical damage, excessive moisture
or chemical corrosion the safety protection may be impaired and the apparatus should be
withdrawn from use and returned for checking and repair.
Make sure that only fuses with the required rated current and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuse holders is prohibited.
Do not wet the instrument when cleaning it.
The following symbols are used on the instrument and in this manual:-
Earth (ground) terminal.
9
mains supply OFF.
alternating current (ac)
direct current (dc)
for 230V operation:
1.6A (T) 250V HBC
for 115V operation:
3.15A (T) 250V HBC
for 230V operation:
4A (T) 250V HBC
for 115V operation:
8A (T) 250V HBC
for 230V operation:
1.6A (T) 250V HBC
for 115V operation:
3.15A (T) 250V HBC
BROWNBLUE
BROWNBLUEBROWN
115V230V
Installation
Check that the instrument operating voltage marked on the rear panel is suitable for the local
supply. Should it be necessary to change the operating voltage, proceed as follows:
1) Disconnect the instrument from all voltage sources.
2) Remove the screws which retain the top cover and lift off the cover.
3) Change the connections on both transformers following the appropriate diagram below:
4) Refit the cover and the secure with the same screws.
5) To comply with safety standard requirements the operating voltage marked on the rear panel
must be changed to clearly show the new voltage setting.
6) Chang e all three fuses to ones of the correct rating, see below.
Fuse
The AC inlet fuse is located in the fuse drawer in the lower part of the IEC inlet connector. To
change the fuse remove the line cord and open the fuse drawer with a suitable tool.
XDL 35-5, XDL 35-5P, XDL 56-4 & XDL 56-4P
The correct mains fuse type is 20 x 5mm 250V HBC time-lag with the following rating:
XDL 35-5T, XDL 35-5TP, XDL 56-4T & XDL 56-4TP
The correct mains fuse type is 20 x 5mm 250V HBC time-lag with the following rating:
In addition, the two transformers are individually fused inside the power supply. To access these
fuses remove the cover as described above; both fuses are clipped to the small PCB which is
fitted directly onto the pins of the IEC inlet connector itself.
The correct fuse type is 20 x 5mm 250V HBC time-lag with the following rating:
Make sure that only fuses with the required current rating and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuseholders are prohibited.
10
Mains Lead
Connect the instrument to the AC su pply using the mains lead prov id ed. Should a mains plug be
required for a differ ent m ains outlet socket, a suitably rated and approved mai ns lea d s et should be
used which is fitted with the required wall plug and an IEC60320 C13 connector for the inst r ument
end. To determine the minimum current r at ing of the lead-set for the inten ded AC supply, refer to the
power rating information o n the equipment or in the Specification.
Any interruption of the ma i ns ear t h c onductor inside or outside t he instrument will make the instrument
dangerous. Intentional i nt er r upt io n is pr ohibited.
Mounting
This instrument is suitable both for bench use and rack mounting. It is delivered with feet for
bench mounting. The front feet include a tilt mechanism for optimal panel angle.
A rack kit for mounting XDL Series power supplies is available from the Manufacturers or their
overseas agents. The rack will accommodate 1, 2 or 3 single units or a triple and single unit; a
blanking piece is also available for unused positions in the rack.
Ventilation
The power supply is cooled by an intelligent multi-speed fan which vents at the rear. Take care
not to restrict the air inlets at the side panels or the exit at the rear. In rack-mounted situations
allow adequate space around the instrument and/or use a fan tray for forced cooling.
WARNING! THIS INSTRUMENT MUST BE EARTHED.
Front Panel Connec tions
The loads should be connected to the positive (red) and negative (black) terminals marked
OUTPUT 1, OUTPUT 2, or AUX.
Remote sense connections to the loads on Outputs 1 or 2, if required, are made from the
corresponding positive (+) and negative (−) REMOTE SENSE terminals. Remote sense
operation is selected from the keyboard or via a remote control interface (P models only); the
REMOTE SENSE lamp is lit when remote sense is selected. Switching off remote sense returns
the instrument to local sensing at the output terminals.
The terminal marked
is connected to the chassis and safety earth ground.
Rear Panel Connections
Main Output Terminals (P models only)
The output and sense terminals are duplicated on the rear panel screw-terminal block marked
Output +, Output −, Sense + and Sense − ; these connections are paralleled with their front panel
equivalents.
Connections
Remote sense operation is selected from the keyboard or via a remote control interface. When
the rear panel terminals are used, remote sense should always be selected to ensure that output
regulation is maintained within specification.
Auxiliary Output Terminals (T models only)
The front panel AUX OUTPUT terminals are duplicated on the rear panel with screwless terminals
marked AUXILIARY OUTPUT.
11
3
RXD
Received data to instrument
Alarm Outputs
Associated with each main output are recessed 2-pin connectors marked Alarm. These provide
access to an opto-isolated NPN switching transistor, the function of which can be set from the
keyboard, see the Alarm Output section of this manual.
The maximum operating voltage that can be applied across the terminals is 20VDC and the
maximum sink current for the switch 'closure' is 1mA.
Do not apply external voltages between the terminals exceeding 30VDC.
RS232 (P models only)
9−pin female D−connector with pin connections as shown below. Can be connected to a
standard PC port using a fully wired 1:1 male-female cable without any cross-over connections.
PinNameDescription
1 RI
2 TXD Transmitted data from inst rument
4 CTS
5 GND Signal ground
6 RTS
7 DSR No internal connection
8 DTR
9 CD No internal connection
Passively asserted (+V through 10kΩ)
Passively asserted (+V through 10kΩ)
Signal ground is connected to instrument ground.
USB (P models only)
The USB port is connected to instrument ground. It conforms with USB 2.0 (Full Speed) and
accepts a standard USB cable. The Windows plug-and-play functions should automatically
recognise that the instrument has been connected. If the correct driver is not found, follow the
Windows on-screen prompts and install the required files from the CD supplied.
LAN (P models only)
The LAN interface is designed to meet LXI ( Lan eXtensions for Instrumentation) version 1.2; the
instrument is Class C compliant. Remote control using the LAN interface is possible using a
TCP/IP Socket protocol. The instrument also contains a basic Web server which provides
information on the unit and allows it to be configured. Since it is possible to misconfigure the LAN
interface, making it impossible to communicate with the instrument over LAN, a LAN
Configuration Initialise (LCI) mechanism is provided via a recessed switch on the rear panel
(marked LAN RESET) to reset the unit to the factory default.
Further details are given in the Remote Operation chapter. For more information on LXI
standards refer to www.lxistandard.org/home
GPIB (P models only)
The GPIB signal grounds are connected to the instrument ground. The implemented subsets are:
SH1 AH1 T6 TE0 L4 LE0 SR1 RL2 PP1 DC1 DT0 C0 E2
The GPIB address is set from the front panel.
12
This section of the manual is a general introduction to the controls and operation of the instrument
and is intended to be read before using the power supply for the first time.
In this manual front panel keys, connections and display indicators are shown in capitals,
e.g. STORE, ESCAPE, OUTPUT, JOG. Messages shown on the 7-segment display are printed
in a different type-font, e.g.
characters as they are shown on the 7-segment display.
StorE, GPIb, triP in upper or lower case to represent the
Switching On, Output On/Off
The power switch is located at the bottom left of the front panel.
At power-up the default behaviour is for the instrument's settings to be restored to those
automatically saved when it was switched off, but with all outputs always off. However, the user
can change the default setting such that selected outputs are restored at power-up to their status
at power-down, see the Extra Functions section.
The DC outputs are switched electronically with their respective ON/OFF keys; the key illuminates
when the output is on. In addition, all outputs can be switched on and off together using the ALL
ON and ALL OFF keys.
Pressing ALL OFF at any time will synchronously turn off any outputs that are on; under the same
load conditions outputs will typically turn off within 1ms of each other. With all outputs off the ALL
OFF key is illuminated green. Pressing the ALL ON key when all the outputs are off will turn all
the outputs on synchronously; outputs with identical settings and load conditions will typically turn
on within 1ms of each other. However, if one output is already on, pressing ALL ON will turn the
remaining outputs on but the turn-on delay between the outputs will be up to 80ms, even with the
same output setting and load conditions.
Initial Operation
Keypad
Only the principles of operation are outlined here; the setting of individual parameters is given in
detail in later sections.
The paramount consideration in designing the user interface has been to make changing settings
as 'safe' as possible (i.e. with minimal risk of accidentally applying excessive voltages to a target
system) whilst achieving ease of use. This has been achieved by requiring the user to confirm
(OK) new numeric settings, with the option to ESCAPE at any point or even to simply pause until
the operation times-out and the instrument returns to its original settings.
In addition a buzzer, illuminated keys, LED indicators and display messages prompt, guide or
warn the user such that entry or control errors are minimised. Where some of these features (e.g.
beeps or flashing indicators) are considered unnecessary by regular users, the option exists to
disable them, see the Extra Functions section.
On triple output (T) models the ability to change settings from the keypad or by using the Jog
controls is assigned to Output 1, Output 2 or both by using the 1, 2 or LINK CONTROL keys
respectively. The key (1 or 2) associated with the selected output illuminates to show which
output is under control. In LINK mode (both keys lit) both outputs are controlled at the same time,
including some of the shifted operations (RANGE, STORE, RECALL and V x A). The further
descriptions that follow apply to either or both main outputs as appropriate to the setting indicated
by the illumination of the CONTROL keys.
Under normal conditions the numeric keypad is disabled; pressing any key will cause the buzzer
to make a double beep, indicating an illegal operation. To set a voltage or current with the keypad
press the V or I NUMERIC SET key; the appropriate display shows 0·000V or 0·000A with the
digit to the left of the decimal point flashing. Digits are entered in response to the flashing prompt,
together with the decimal point at the appropriate time, and the entry is confirmed with the OK
key. If OK is not pressed within 10 seconds of the last numeric key the entry is cancelled and the
display returns to its original setting. If ESCAPE is pressed anywhere in the entry procedure,
entry is cancelled and the display returns to its original setting.
13
The OK key is used to confirm most keypad entries. At all other times it becomes the VIEW V/I
LIMITS key and pressing it will cause the display to show preset output voltage and current limit
for 3 seconds; during this period the LIM indicator in the display flashes.
Pressing SHIFT illuminates the key and gives the numeric keys the functions marked above them
(e.g. STORE, RECALL, etc.). When a function is selected by pressing one of these keys SHIFT
is cancelled (the SHIFT key is no longer lit). The further key presses required to complete the
selected function are described in detail in the sections that follow; if no key is pressed within 10
seconds to complete the function, the function will terminate as if ESCAPE has been pressed.
SHIFT is a toggle key; pressing SHIFT again when it has been selected will cancel SHIFT. SHIFT
is also cancelled by ESCAPE, or by pressing SET V or SET I. Note that in LINK mode settings
accessed by STORE and RECALL are specific to the LINK mode and are in addition to settings
accessible when STORE and RECALL are used on individually selected outputs.
Jog Control
The rotary 'jog' control permits the output voltage or current limit to be incremented or
decremented in steps with a resolution set by the JOG SET keys; the output immediately follows
the setting, i.e. no OK is required.
At power-up jog is always off. To jog the voltage or current setting press the V or I JOG SET key;
the key will illuminate and the JOG indicator under the digit that was last jogged will flash. Whilst
the V or I JOG SET key is lit, each further press of the V or I key moves the JOG indicator one
digit to the left; the selection 'wraps-round' such that when the largest value of jog increment has
been reached the next press returns it to the lowest. The default position at power-up is under
the LSD, i.e. the lowest jog increment is selected.
Turning the rotary jog control clockwise/anti-clockwise increments/decrements the selected digit;
digits to the left of the one being jogged are automatically incremented/decremented when the
decade overflow/underflow point is reached. Digits to the right of the one being jogged remain
unchanged unless the jog step overflows/underflows the range maximum/minimum in which case
they are set to zero. On the XDL 35-5, for example, 33·65V goes to 34·65V goes to 35·00V for
the 35V range and a 1V jog increment; 0·160A goes to 0·060A goes to 0·001A for a 0·1A jog
decrement.
The jog steps that can be selected for the main range are 1mV, 10mV, 100mV and 1mA, 10mA,
100mA; if the 500mA range has been selected the current increment steps are 0·1mA, 1mA,
10mA.
To disable the jog rotary control press the JOG SET OFF key; reselecting JOG SET V or I will
enable jog on the last used digit position. Jog is not cancelled by using numeric entry or any of
the SHIFT functions but it is disabled whilst that function is enabled.
When in constant voltage mode, with the output on, the right-hand display will show actual current
rather than current limit. If JOG SET I is selected the JOG indicator under the selected digit will
flash at half-speed ('lazy' flash). To observe the effect of jogging the current limit it will be
necessary to either turn the output off (so that the display permanently shows the current limit) or
to press VIEW V/I LIMITS which causes the current limit to be displayed until 3 seconds after
movement of the jog control ceases. The 'lazy' flash is also used when JOG SET V has been
selected and actual voltage is being shown because the supply has gone into current limit.
The factory default is to flash the JOG indicator under the selected digit for the whole time that jog
is selected so that the user is constantly reminded which parameter can be incremented/
decremented. Where this degree of reminding is considered inadequate the user can select,
using the Extra Functions capability, t o flash the digit itself; conversely, where the flashing is
considered intrusive the user can select to not flash the JOG indicator (except when the 'lazy'
flash is shown).
14
Display
The displays shows the voltage on the left (5 digits) and the current on the right (4 digits) for both
the main outputs. These 7-segment displays are also used to show prompts during some of the
function settings (e.g. memory store/recall or remote control address setting) using the limited
'character set' that can be achieved with a 7-segment display; these are necessarily a mixture of
upper and lower case letters.
Above and below the 7-segment display are several secret-until-lit annunciators.
To the right, above the current display, are the indicators which show the selected operating
range: 35V/3A, 15V/5A or 35V/500mA (XDL 35-5T & TP) or 56V/2A, 25V/5A or 56V/500mA (XDL
56-4T & TP); the indicators light beneath the range printed immediately above them and, in the
case of the 500mA range, the indicator is marked mA to emphasise that the current display is now
showing mA.
The other annunciators above the displays are:
CI, indicating that the instrument is in constant current mode;
LIM, which flashes when the VIEW V/I LIMITS key is pressed to show the set voltage/set current
limit in the display;
REM, which lights when the instrument is under control from a remote interface (P models only)
and LAN, which indicates the status of the LAN interface connection (P models only).
Below the three least significant digits of both the voltage and current displays are the JOG
indicators; the appropriate indicator flashes when the jog function is being used, see the Jog
Control section above.
The display of Output 1 can alternatively be used to show the voltage and current of the AUX
output (T models only). Full details are given in the Auxiliary Out put section.
15
Main Outputs
New users should first read the Initial Operation chapter which describes the operating principles
of the keypad and rotary jog control. The following paragraphs describe the independent
operation of either Main Output. To select which output is to be controlled by the keypad/Jog
controls it is first necessary to select that output by pressing the appropriate CONTROL key
(1 or 2); the key lights to show that it is the selected output.
The additional features available on the triple output (T) models in LINK mode (both Main Outputs
selected) are described in the Main Outputs – Link Mode section later in this manual.
Set Voltage
The left-hand display shows the set voltage to a resolution of 1mV, except when the instrument is
in constant current (CI) mode. In CI mode the actual output voltage (which will be less than the
set voltage) is shown and the display resolution is 10mV; the least significant digit (1mV
resolution) is always displayed as a zero.
The voltage can be set directly from the numeric keypad: press the NUMERIC SET V key, enter
the new value using the numeric keys and confirm by pressing OK. The broad principles of
keypad entry are explained in the Initial Operation chapter, which should be read by new users.
When SET V is pressed the display shows 0·000; a new voltage is then entered (e.g. 12·345V is
entered as 1, 2, ·, 3, 4, 5) and confirmed by OK. The position of the decimal point in the display is
fixed to reduce the risk of entering a wrong value. As a consequence, and to avoid the need to
enter leading zeroes (e.g. 2·345V is entered as 2, ·, 3, 4, 5, OK), numbers to the left of the
decimal point are shown slightly differently to the numbers to the right of the decimal point during
number entry; this is self-evident during number entry.
Manual Operation
The minimum voltage setting is 0·000V; the maximum setting for XDL 35-5 is 35·000V (15·000V
on the 15V/5A range) or 56.000V (25.000V on the 25V/4A range) for XDL 5 6-4.
Pressing OK at any point will set the voltage entered with any remaining digits set to zero,
e.g. 1, 2, ·, 3, OK will set 12·300V; 1, OK will set 1·000V; pressing OK immediately after SET V
(while the display shows 0·000V) will set 0·000V.
Pressing ESCAPE at any time during the sequence, or making no further key press within
10 seconds of the previous one will cause the display to return to its original reading before
SET V was pressed.
Entering a voltage outside the range maximum (including trying to enter 3 digits before the
decimal point) or trying to enter more than 5 digits will cause the buzzer to beep twice; the last
key entry will be ignored.
The voltage can also be set using the Jog control. Pressing JOG SET V will illuminate the V key
and the JOG indicator under the digit that was last jogged will flash. Whilst the V key is lit, each
further press will move the JOG indicator one digit to the left; the selection 'wraps round' such that
when the largest value of jog increment has been reached the next press returns it to the lowest.
The default position at power-up is under the LSD, i.e. the lowest jog increment is selected. The
jog steps that can be selected are 1mV, 10mV and 100mV.
With jog enabled the output voltage can be incremented or decremented with the rotary jog
control with a step resolution indicated by the position of the flashing JOG indicator. The output
immediately follows the setting, i.e. no OK is required. If the output goes into constant current
mode (indicated by the CI indicator flashing) the left-hand display shows actual voltage not set
voltage. If JOG SET V is selected the JOG indicator under the selected digit will flash at half
speed ('lazy' flash). To observe the effect of jogging the set voltage it will be necessary to either
turn the output off (so that the display permanently shows the set voltage) or to press
VIEW V/I LIMITS which causes the set voltage to be displayed until 3 seconds after movement of
the jog control ceases.
16
Note that in constant current mode the actual voltage is measured and displayed to only 10mV
resolution; the 1mV digit permanently displays zero.
Further details on the jog control can be found in the Initial Operation chapter.
Set Current Limit
With the output off, the right-hand display shows the current limit to a resolution of 1mA (0·1mA on
the 500mA range).
The current limit can be set directly from the numeric keypad: press the NUMERIC SET I key,
enter the new value using the numeric keys and confirm by pressing OK. The broad principles of
keypad entry are explained in the Initial Operation chapter, which should be read by new users.
When SET I is pressed the display shows 0·000; a new current is then entered (e.g. 1·234A is
entered as 1, · , 2, 3, 4,) and confirmed by OK. The position of the decimal point in the display is
fixed to reduce the risk of entering a wrong value. As a consequence, and to avoid the need to
enter or display leading zeroes (e.g. 0·234A is entered as ·, 2, 3, 4, OK), numbers to the left of
the decimal point are shown slightly differently to the numbers to the right of the decimal point
during number entry; this is self-evident during number entry.
The minimum current setting is 0·001A (0·1mA on the 500mA range); the maximum setting is
3·000A, 5·000A or 500·0mA (XDL 35-5) or 2·000A, 4·000A or 500·0mA (XDL 56-4), according to
range, i.e. there is no over-range capability.
Pressing OK at any point will set the current entered with any remaining digits set to zero,
e.g. 1, ·, 2, OK will set 1·200A; 1, OK will set 1·000A; pressing OK immediately after SET I (while
the display shows 0·000A) will set 0·00IA.
Pressing ESCAPE at any time during the sequence, or making no key press within 10 seconds of
the previous one will cause the display to return to its original reading before SET I was pressed.
Entering a value outside the range maximum (including trying to enter 2 digits before the decimal
point) or trying to enter more than 4 digits will cause the buzzer to beep twice; the last key entry
will be ignored.
The current limit can also be set using the rotary jog control. Pressing JOG SET I will illuminate
the key and the JOG indicator under the digit that was last jogged will flash. Whilst the I key is lit,
each further press will move the JOG indicator one digit to the left; the selection 'wraps round'
such that when the largest value of jog increment has been reached the next press returns it to
the lowest. The default position at power-up is under the LSD, i.e. the lowest jog increment is
selected. The jog steps that can be selected are 1mA, 10mA and 100mA (0·1mA, 1mA and 10mA
on the 500mA range).
With jog enabled the current limit can be incremented or decremented with the rotary jog control
with a step resolution indicated by the position of the flashing JOG indicator. The output
immediately follows the setting, i.e. no OK is required. With the output on, the right-hand display
shows actual current, not current limit (except in constant current mode). If JOG SET I is selected
the JOG indicator under the selected digit will flash at half speed ('lazy' flash). To observe the
effect of jogging the current limit it will be necessary to either turn the output off (so that the
display permanently shows the current limit) or to press VIEW V/I LIMITS which causes the
current limit to be displayed until 3 seconds after movement of the jog control ceases.
Instantaneous Current Output
The current limit control can be set to limit the continuous output current to levels down to 1mA
(0·1 mA on 500mA range). However, in common with all precision bench power supplies, a
capacitor is connected across the output to maintain stability and good transient response. This
capacitor charges to the output voltage and short-circuiting of the output will produce a current
pulse as the capacitor discharges which is independent of the current limit setting.
17
Range Selection
The instrument has three ranges: 35V/3A, 15V/5A and 35V/500mA (XDL 35-5) or 56V/2A, 25V/4A
and 56V/500mA (XDL 56-4). The selected range is shown by an illuminated indicator below the
appropriate legend at the top right-hand side of the instrument; when the 500mA range is selected
the indicator legend is mA to emphasise that the current meter now shows milliamps not amps.
To change range press SHIFT followed by RANGE or RANGE; each press of RANGE
selects the next range to the left; each press of RANGE selects the next range to the right;
there is no 'wrap-round'. When the range is changed the indicator that represents the new range
and the OK key both flash; pressing OK sets the new range. To exit without changing range
press ESCAPE. Pressing any other key whilst in range change mode causes the warning buzzer
to beep twice; no other action is taken. If OK is not pressed within 10 seconds of the last range
change key press the range selection remains unchanged.
The range can only be changed when the output is off. Pressing the RANGE or RANGE
keys with the output on will cause the output ON/OFF key (as well as the OK key) to flash. The
output may be turned off with the ON/OFF key and the range then changed by pressing OK, or
OK may be pressed directly in which case the output is automatically turned off and the range
then changed.
If a range change causes a voltage or current limit setting to exceed the corresponding maximum
of the new range the range change is accepted but the setting is made equal to the maximum of
the new range.
Note that the OVP setting is not changed when the range is changed (e.g., for XDL 35-5, an OVP
setting of 38V remains valid on the 15V range); it is left to the user to independently change the
OVP setting if required.
Output Settings – Front Panel Lock
To avoid accidental changes to the output settings in a bench or rack set-up, front panel control of
Range, Voltage, Current Limit, OVP and OCP can be ‘locked’ and ‘unlocked’ with alternate uses
of the #33 function, see Extra Functions section. The output ON/OFF key remains unlocked, as
does selection of remote sense. Front panel lock still operates in remote control mode (P models
only) but is ignored by the remote commands.
On triple output models the AUX output voltage is also locked when Output 1 settings are locked
but, additionally, the CONTROL keys remain unlocked. All outputs are locked and unlocked
together if #33 is used in Link mode, see Auxiliary Output and Main Outputs-Link Mode sections.
Connection to the Load
The load should be connected to the positive (red) and negative (black) OUTPUT terminals. Both
are fully floating and either can be connected to ground.
Remote Sensing
The instrument has a very low output impedance, but this is inevitably increased by the resistance
of the connecting leads. At high currents this can result in significant differences between the
indicated source voltage and the actual load voltage (two 20mΩ connecting leads will drop 0·2V
at 5 Amps, for instance). This problem can be minimised by using short, thick, connecting leads,
but where necessary it can be completely overcome by using the remote sense capability. This
requires the sense terminals to be connected to the output at the load instead of at the source;
insert wires into the spring-loaded REMOTE SENSE terminals and connect directly to the load.
Select remote sense by pressing SHIFT, SENSE; the OK key flashes and the lamp above the
remote sense terminals lights to show that remote sense will be selected when OK is pressed.
Press OK to confirm; press ESCAPE to exit without changing state. Remote sense is turned off
by pressing SHIFT, SENSE again; the OK key flashes and the remote sense lamp goes off to
indicate that local sense will be restored when OK is pressed. Press OK to confirm; press
ESCAPE to exit without changing state.
18
To avoid instability and transient response problems, care must be taken to ensure good coupling
between each output and sense lead. This can be done either by twisting the leads together or by
using coaxially screened cable (sense through the inner). An electrolytic capacitor directly across
the load connection point may also be beneficial.
The voltage drop in each output lead must not exceed 0·5 Volts.
The P models have rear panel output and sense terminals, appropriate for when the instrument is
used in a rack. The rear panel sense terminals should always be used with the rear panel output
connections.
Sense Miswiring Trip
The output will be tripped off if the voltage between an output terminal and its corresponding
sense terminal exceeds approximately 1V; this will happen if the sense wires are wired at the load
to the wrong output or if an attempt is made to draw power from the sense wires.
If the sense terminals are miswired in this way the display shows the message
and the output is turned off. Pressing ESCAPE at this point removes the message and the
display now shows the preset voltage and current limit. When the cause of the trip has been
corrected the output can be turned on again.
Series or Parallel Connection with Other Outputs
The outputs of the power supply are fully floating and may be used in series with other power
supply units to generate high DC voltages up to 300V DC. It should be noted that the unit can
only source current and cannot sink it, thus units cannot be series connected in anti-phase.
SENSE triP
The maximum permissible voltage between any terminal and earth ground () is 300VDC.
WARNING! Such voltages are exceedingly hazardous and great care should be taken to shield
the output terminals for such use. On no account should the output terminals be touched when
the unit is switched on under such use. All connections to the terminals must be made with the
power switched off on all units.
The unit can be connected in parallel with others to produce higher currents. Where several units
are connected in parallel, the output voltage will be equal to that of the unit with the highest output
voltage setting until the current drawn exceeds its current limit setting, upon which the output will
fall to that of the next highest setting, and so on. In constant current mode, units can be
connected in parallel to provide a current equal to the sum of the current limit settings.
Note that the output terminals are rated at 30A maximum; if several outputs are operated in
parallel to source higher currents than this the junction should be made at a separate point, not
one of the terminals.
Over-Voltage Protection
Over-Voltage Protection (OVP) can be set from 1·0V to 40V (XDL 35-5) or to 60V (XDL 56-4). If
the output voltage exceeds the set OVP the output is immediately shut down (typically within
100µs), thus avoiding damage to the circuit under test. The OVP circuit will protect against
accidental excessive voltage settings from the front panel or via the remote control interfaces,
external voltages impressed across the output terminals, or a failure in the control circuitry of the
instrument itself.
To set OVP press SHIFT, OVP; the 100mV step JOG indicator will start flashing and the jog rotary
control can be used to increment/decrement the OVP setting in 100mV steps. Press OK to
confirm the new setting; to exit without entering a new value press ESCAPE. The factory default
settings are 40·0V (XDL 35-5) and 60V (XDL 56-4).
If the OVP is tripped the display shows the message
Pressing ESCAPE at this point removes the message and the display now shows the preset
voltage and current limit. When the cause of the OVP has been removed (or the OVP limit
changed) the output can be turned on again.
Note that the OVP setting is not changed when the range is changed (e.g., for XDL 35-5, an OVP
setting of 38V remains valid on the 15V range); it is left to the user to independently change the
OVP setting if required. Note also that it is possible and valid to set OVP below the set voltage. If
the supply is in constant current mode the output voltage will be below the set voltage; OVP could
OUP triP and the output is turned off.
19
be set such that is was above the actual output voltage but below the set voltage. This could be
sed to trip the output under a fault condition which caused the load impedance to increase and
the actual output voltage to therefore rise above the OVP point.
Over-Current Protection
Over-Current Protection (OCP) can be set from 0·01A to 5·5A (XDL 35-5) or to 4.4A (XD L 56-4).
If the output current exceeds the set OCP the output is shut down (typically within 35ms).
To set OCP press SHIFT, OCP; the 10mA step JOG indicator will start flashing and the jog rotary
control can be used to increment/decrement the OCP setting in 10mA steps. Press OK to confirm
the new setting; to exit without entering a new value press ESCAPE. The factory default setting
is 5·50A (XDL 35-5) or 4.4A (XDL 56-4).
If the OCP is tripped the display shows the message
Pressing ESCAPE at this point removes the message and the display now shows the preset
voltage and current limit. When the cause of the OCP has been removed (or the OCP limit
changed) the output can be turned on again.
Note that as with OVP, the OCP setting is not changed when the range is changed.
Note also that is possible and valid to set OCP below the set current limit. For example, the
power supply may be used to repetitively test a unit under test (UUT) which normally takes a peak
current of, say, 2 Amps. However, a f aulty UUT would take a current of more than 2 Amps and
would be damaged by being left in a 2 Amp current-limited state. In this case the current limit
could be set to 2·1A, say, and the OCP set to 2·0A to ensure that a faulty UUT will trip the supply
off.
OCP triP and the output is turned off.
Output Protection
In addition to OVP and OCP for forward over-voltage and over-current protection, the output is
protected from reverse voltages by a diode; the continuous reverse current must not exceed 3
Amps although transients can be much higher.
Output Power (V x A)
If SHIFT, V x A is pressed the voltage display shows the product of measured output voltage x
measured current and the current display shows
updated at the normal measurement rate. Output Power mode is cancelled by pressing either
ESCAPE or V x A again. Jog is temporarily disabled (and the JOG indicators are turned off)
during the V x A display.
Temperature Trip
If the safe internal temperature limit is exceeded because, for example, the fan vents have been
blocked, the output is turned off and the display will show
this point will do one of two things:
i. If the over-temperature condition has already cleared the message will be removed and the
display will show preset voltage and current limit. Assuming the cause of the over-temperature
has been rectified the output can be turned on again.
ii. If the instrument is still above the safe temperature limit the
slowly ('lazy' flash) until the instrument has cooled, at which point the display will show preset
voltage and current limit again. Assuming the cause of the over-temperature has been rectified
the output can be turned on again.
UA ; the output power reading is continuously
OtP triP. Pressing ESCAPE at
OtP triP message will flash
Alarm Output
The recessed 2-pin connector on the rear panel is directly connected to an opto-coupled NPN
switching transistor (pin 1 emitter, pin 2 collector) which is turned on (i.e. switch 'closure')
according to the conditions specified in the Extra Functions section, see later. The default
condition is switch closure for any trip condition (OVP, OCP, SENSE or OTP). The maximum
open-circuit voltage permitted across the switch is 30VDC and the nominal sink current for switch
closure is 1mA.
20
•
Store Settings
The instrument can store 50 set-ups for each output in non-volatile memory; the parameters
stored are range, voltage, current limit, OVP and OCP. The output state and remote sense
setting are not stored. In addition, a further 50 Link Mode set-ups can be saved, see Main
Outputs – Link Mode section.
To store a set-up, first press SHIFT, STORE; the display shows
number (
right. The SHIFT function is cancelled (the light goes off) at this point. The store number (0 to
49) can be set directly from the keypad or by using the Jog control to increment/decrement the
displayed number; the JOG indicator beneath the store number flashes to indicate that the Jog
control is active. Set the required store number by either method and press OK to store the
settings and return the display to showing V & I. The store function can still be used when the
output settings have been ‘locked’ using the #33 function. A full store can be overwritten with new
settings. At any time before the OK key is pressed the store function can be exited without saving
a set-up by pressing ESCAPE or by waiting 10 seconds from the last key entry.
0 to 49) and store status (either E for store Empty or F for store Full) on the
Deleting Stored Settings
Any store can be returned to 'empty' as follows: press SHIFT, STORE, and set the required store
number via keypad or Jog control as described above; at that point press
shows
store. All of the output’s stored set-ups can also be deleted simultaneously by using the #98
function, see Extra Functions section.
dEL in place of Sto , e.g. dEL 29 F ; pressing OK deletes the content of the
Recall Settings
To recall a set-up, first press SHIFT, RECALL; the display now shows rEC. on the left with the
store number (
the right as each store is either selected in turn using the Jog control, or is set from the keypad
(as for storing set-ups, see above). The SHIFT function is cancelled (the light goes off) at this
point. If the store selected is full (
of that store; press OK to recall the settings of the store and return the display to showing V & I. If
the store selected is empty (
possible to recall an ‘empty’ store (pressing OK gives a warning beep); either select a full store or
press ESCAPE to exit Recall. Recall cannot be used if the output has been ‘locked’.
At any time before the OK key is pressed the Recall function can be exited without recalling a setup by pressing ESCAPE or by waiting 10 seconds from the last key entry.
Settings may be recalled with the output on or off. However, if the recalled setting involves a
range change the output is turned off to avoid any 'glitches'. After pressing SHIFT, RECALL, and
setting the store number, the ON/OFF key will flash (as well as the OK key) if completing the
recall involves a range change. The output may be turned off with the ON/OFF key and the recall
then completed by pressing OK, or OK may be pressed directly in which case the output is
automatically turned off and the recall completed.
0 to 49) and store status (either E for store Empty or F for store Full) on
Sto. on the left with the store
. The display now
F), the display changes to a flashing preview of the V & I values
E), the display flashes ----- ---- to indicate this. It is not
Extra Functions
Variations on some of the factory default functions can be set by the user by using the # extra
functions facility. Each function change, detailed in the list below is accessed by pressing SHIFT,
#, nn, when nn is the 2-digit num ber in the list below; the display changes to
SHIFT, # and the buzzer gives a confirmation beep when the 2-digit number entry is complete. As
indicated in the opening paragraph of this section, the # functions can be set independently (i.e.
differently) for each main output; note, however, that the #02, #03 and #21 functions which apply
to the Auxiliary Output can only be set when CONTROLis assigned to Output 1.
The settings of each Main Output can be ‘locked’/’unlocked’ individually by using #33 with the
CONTROL assigned to the respective output; on T models the AUX settings are also
locked/unlocked when Output 1 is locked/unlocked. The settings of all outputs are
locked/unlocked together if #33 is used with CONTROLset to LINK mode.
21
HASH No._ after
# Code
Function
00
Main Output always off at power-up (factory default)
01
Main Output status at power-up the same as at last power-down
Set with control assigned to Output 1.
Set with control assigned to Output 1.
20
Alarm output 'open' for main Output off, 'closed' for main Output on
Set with control assigned to Output 1; applies to Output 1 alarm only.
22
Alarm output 'closed' when over-temperature trip occurs
23
Alarm output 'closed' when sense trip occurs
24
Alarm output 'closed' when over-current trip occurs
25
Alarm output 'closed' when over-voltage trip occurs
26
Alarm output 'closed' when any trip occurs (factory default)
30
Buzzer off
indicates a wrong entry.
33
Lock/Unlock settings. Note that the AUX output settings are locked with Output1.
40
Jog digit flashes, JOG indicator only flashes when jog is 'hidden'
41
JOG indicator always flashes, except when 'hidden' (factory default)
42
JOG indicator doesn't flash, except when 'hidden' (lazy flash)
91
Loads default calibration parameters. Refer to Service Guide
92
Shows firmware version number in the display
93
Sets these # settings to their factory default
94
Loads Factory Default settings (see below)
In LINK mode, clears all memories in all modes (O/P1, O/P2, AUX & LINK).
99
Enter calibration mode. Refer to Service Guide.
Range:
XDL 35-5: 35V/3A
XDL 56-4: 56V/2A
Voltage:
XDL 35-5: 1.000V
XDL 56-4: 1.000V
Current Limit:
XDL 35-5: 1.000A
XDL 56-4: 1.000A
OVP:
XDL 35-5: 40V
XDL 56-4: 60V
OCP:
XDL 35-5: 5.5A
XDL 56-4: 4.4A
Output:
Output off; local Sense
# Settings:
00 Main Output always off at power-up
02 Aux Output always off at power up (an Output 1 # function only).
26 Alarm output 'closed' when any trip condition occurs
31 Buzzer on
41 JOG indicator always flashes; 'lazy' flash when hidden
02 Aux Output (T models only) always off at power up (factory default).
03 Aux Output (T models only) status at power up the same as at last power down.
21 Alarm output 'closed' when Aux Output is in Current Limit.
31 Buzzer on (factory default). A single beep indicates confirmation, a double beep
98 Clears all the output’s set-up memories. AUX output always cleared with O/P1.
Factory Default Settings
The ex-factory default settings (which will apply at first power-up) are as follows:
22
RS232:
9,600 Baud (P versions only)
Address:
11 (P versions only)
Error No.
Error Description
Action on pressing OK
1
Calibration constants corrupted at power-up
Loads default calibration parameters
2
# functions corrupted at power-up
Loads default # settings
power up
settings
Error Messages
The following hardware errors are indicated by showing the appropriate error number in the display.
The OK key will flash and if pressed the error will be ignored and operation will continue as described.
3 Power-do wn settings not correctly loaded at
Switching the instrument off with the error message showing will leave all settings unchanged.
Loads factory default power-up
Main Outputs – Link M ode (T models onl y)
In Link mode, selected by pressing the LINK key, the key parameters of the two Main Outputs are
adjusted together; when Link mode has been selected both CONTROLkeys (1 and 2) are lit
to show that both outputs are selected.
The following paragraphs only describe the differences between independent and linked
operation; they should be read in conjunction with the corresponding paragraphs in the Main
Outputs section.
Link Mode Operation − Overview
Control of the two main outputs can be "linked" so that changes are applied to both outputs
simultaneously. There can be several reasons for wanting to do this:
1. Series or Parallel Wiring The user may wish to create an output with either twice the voltage or twice the current
capability, see the Series or Parallel Connection with Other Outputs section on page 19.
Link mode provides a convenient means for controlling the two outputs when they are series
or parallel connected.
2. Tracki ng Voltages (or Currents)
When in Link mode, using Numeric Set will set equal voltages and/or currents on the two
outputs. Control of the outputs can also be linked with different voltages and/or currents set
on the outputs. Use of the Jog control will then make equal changes to the two outputs.
3. Simultaneous Recall of Stored Settings
Each output has its own set of 50 memories. However, when in Link mode, a further set of
50 memories is available which can store settings for both outputs. Voltages and currents
can be set individually for each output and the control put into Link mode before storing. The
stored settings can then be recalled to both outputs simultaneously. Using #98 in Link
mode will clear all the memories of all 3 outputs in all modes, see Extra Functions section.
Notes: The existing settings for Output 1 can be duplicated on Output 2 using the Copy function
before or after linking.
When in Link mode the control functions are limited to Set Range, Set Volts and Set Current
(Numeric Set and Jog set), plus Store and Recall. OVP, OCP and Sense cannot be
changed while in linked mode.
Control of on/off for each output remains separate when in Link mode. To switch the
outputs on or off together the ALL ON/ALL OFF buttons must be used which are
independent of Link mode.
Using #33 in Link mode will lock/unlock front panel control of all 3 outputs simultaneously.
Selecting Link Mode
The only constraint on selecting Link mode is that both Main Outputs must already be set to the
same Range; in particular, the outputs may be linked even if their output voltage and current limit
23
settings are different. Pressing LINK when different ranges are set will cause the buzzers to
sound twice and the Range indicator of the previously unselected output to flash for 2 seconds.
Selecting Link mode will cancel any Jog selection set on either Main Output.
Set Voltage and Set Current Limit
Setting the output voltage and current limit by numeric entry or Jog control is essentially the same
as for the outputs in independent mode. Using numeric entry the two outputs will be set to
exactly the same new voltage and current limit, irrespective of the settings at the time the outputs
were linked. Note that there can be a finite time difference between the changes on the two
outputs, even if they are changing from the same initial setting; typically this time difference
should be no more than 40ms (80ms max). However, if the settings were different at the time the
outputs were linked, changing the voltage or current limit using the Jog control will maintain the
difference between the two outputs by incrementing/decrementing each output by the same step,
i.e. the outputs will 'track' each other. Tracking will be maintained until one of the outputs reaches
the range limit, at which time each further Jog step will cause the buzzer to sound for that output
(but with the output remaining at the range limit) whilst the in-range output continues to change,
i.e. 'tracking' ceases and the outputs converge with each further step. If the increment/decrement
is reversed the new (smaller) difference between the outputs is maintained until one output
reaches the range limit in the other direction.
If Link mode is exited whilst Jog is selected, Jog continues to be active on the selected output.
Store and Recall
In Link mode a further 50 non-volatile memories are available which are quite separate from the
50 memories for each output in independent mode. The parameters stored are Range, Voltage,
Current Limit, OVP and OCP. Operation in Link mode is exactly as described in the Store, Recall
and Delete paragraphs of the Main Outputs section; the display messages described in these
sections appear in both displays when Link mode is selected.
OVP, OCP and Sense
OVP, OCP and Sense can only be set when either channel is independently selected. The
settings are maintained when Link mode is selected; OVP and OCP can be saved as part of a
Link mode set-up but the Sense setting cannot. If an attempt is made to change OVP, OC P or
Sense whilst in Link mode the display of Output 1 will flash SELCt 1or2 to remind the user
that these parameters must be set independently for each channel. Press ESCAPE to cancel the
flashing display, assign control to the appropriate channel using the CONTROL 1 or 2 keys
and set OVP, OCP, or Sense as described for the individual outputs.
Output Power
Pressing SHIFT, V x A causes the output power of both outputs to be displayed simultaneously in
their respective displays (V x A in the voltage display,
described for independent operation.
Extra Functions
The # functions described in the Extra Functions paragraph of the Main Outputs section can also
be set when the instrument is in Link mode; both displays will show
SHIFT, #. Any # function set in this way will of course be the same for both outputs. However,
the # functions can be different for each output if they are separately set whilst in independent
mode and the function selection will be maintained for each output even when the outputs are in
Link mode.
Bus Type and Address/Baud Rate
The Bus Type, Address and Baud Rate can only be selected when control is assigned to Output
2. The Output 2 display is used to show the parameters being set, exactly as described for
independent operation. If an attempt is made to set these parameters in LINK mode, or with
Output 1 selected, the display of Output 1 flashes
Output 2. Press Escape to clear the display prompt (or wait for it to time out), then select
Output 2.
UA in the current display) exactly as
HASH No._ after pressing
SELCt P.U._2 as a prompt to sel ect
24
Main Outputs – Copy Function (T models only)
The principle settings of Output 1 can be copied to Output 2 using the Copy function, irrespective
of the CONTROL mode (1, 2 or LINK) currently selected. The parameters copied are Range,
Voltage, Current Limit, OVP and OCP; the status of Sense, ON/OFF, Jog and the Store contents
are not copied.
Pressing SHIFT, COPY 1>2 causes the Voltage, Current and Range settings of Output 1 to
appear on the displays of Output 2 in flashing mode; the OK key also flashes. Pressing OK
confirms and implements the Copy operation; pressing ESCAPE at that point abandons the
operation.
If the Copy operation will cause a range change to Output 2, and if Output 2 is ON, the Output 2
ON/OFF key also flashes and the output is turned OFF when OK is pressed; the output can also
be turned off directly with its ON/OFF key before OK is pressed.
Auxiliary Output (T models only)
The AUX output can provide up to 3 Am p at an output voltage of 1.00V to 6.00V. The output
voltage is set by the Jog control (only) with a fixed stepping resolution of 10mV; the current limit is
fixed at ≥3A.
The AUX output voltage can be set, and the voltage and current monitored, on the Main Output 1
display with alternate presses of the SET/VIEW button beside the AUX terminals. Press once to
show the AUX V & I (
& I of Output 1.
A shows in front of the V setting to indicate this), press again to show the V
The AUX output is switched on and off with the AUX ON/OFF key; the key is lit when AUX is on.
With the AUX output off, and the AUX V & I shown on the display of Output 1, the AUX output
voltage can be previewed and set using the Jog control; the current limit is fixed and the preview
shows 3.00A. With the AUX output turned on, the Output 1 display shows actual AUX output
voltage and current.
The AUX output is protected against current overload and momentary short-circuit as follows.
When the load current exceeds ~3.00A the I
lamp lights to show that regulation is no longer
LIMIT
maintained. If this overload condition persists for more than approximately 5s the output will trip
off; the AUX ON/OFF key is no longer lit, the I
(if set to show AUX V & I) shows the message
Pressing the AUX ON/OFF key then resets both the I
lamp now flashes, and the display of Output 1
LIMIT
triP in place of the current measurement.
lamp and the Output 1 display to show
LIMIT
preset AUX voltage and current limit. Once the overload condition has been removed the AUX
output can be switched on again.
A further 10 non-volatile memories, separate from the 50 memories for each Main output and Link
mode operation, are available to store AUX output voltage set-ups. Press the SET/VIEW key to
show AUX V & I in the display of Output 1; operation is then exactly as described in the Store,
Recall and Delete paragraphs of the Main Outputs section.
The output terminals are duplicated on the rear panel (screwless terminals) for rack use; there is
no remote sense capability.
The following features of the AUX output can be set/monitored via the remote interfaces:
Set output voltage; readback set output voltage.
Readback actual output voltage and current.
Set a deltaV increment; readback the set increment.
Increment/decrement the output voltage by deltaV.
Switch AUX output on and off.
Readback output on/off status.
Current limit and current trip status (via LSR2? command, see Status Model).
Further details can be found in the Remote Commands section.
25
The instrument can be remotely controlled via its RS232, USB, LAN or GPIB interfaces.
USB remote control operates in a similar way to RS232 but via the USB connector. Software
supplied with the instrument sets up the controlling computer to treat the USB connection as a
virtual COM port. Application software on the computer can then access the instrument via that
COM port.
The LAN interface is designed to meet LXI (Lan eXtensions for Instrumentation) version 1.2; the
instrument is Class C compliant. Remote control using the LAN interface is possible using the
TCP/IP Sockets protocol. The instrument also contains a basic Web server which provides
information on the unit and allows it to be configured from a web browser. Simple command line
control from the browser is also possible.
All interfaces are, by default, live at all times (a LXI requirement) but access to individual
interfaces may be restricted using the configuration options on the web pages.
Interface Locki ng
All interfaces are live at all times; this removes the need to select the active interface and is also a
LXI requirement. To reduce the risk of the instrument being inadvertently under the control of two
interfaces at once a simple lock and release mechanism is provided in the instruction set. The
lock is automatically released where it is possible to detect disconnection and when the local
button is pressed. Access to the interfaces may also be restricted using the web pages.
Remote Operation (P models only)
Any interface may request to have exclusive control of the instrument by sending an “IFLOCK”
command. The lock may only be released by sending an “IFUNLOCK” command from the
interface instance that currently has the lock and may be queried from any interface by sending
an “IFLOCK?” command. The reply to any of these commands will be “-1” if the lock is owned by
another interface instance, “0” if the interface is free and “1” if the lock is owned by the requesting
interface instance. Sending any command from an interface without control privileges that
attempts to change the instrument status will set bit 4 of the Standard Event Status Register and
put 200 into the Execution Error Register to indicate that there are not sufficient privileges for the
required action.
Note: it is also possible to configure the privileges for a particular interface to either ‘read only’ or
‘no access’ from the Web page interface.
Address & Baud Rate Selection and Interface Status View
The instrument address capability is strictly required only by the GPIB interface. However, use
can be made of the ADDRESS? command over any of the interfaces to easily identify which
instrument is being controlled by a particular COM port (for RS232 or USB) or TCP socket (for
LAN). Note that the LAN interface also has a separate ‘Identify’ function, accessible from the
instrument’s web pages, that flashes the instrument’s display until the function is cancelled.
To access the interface address or Baud rate the CONTROL mode must first be set to 2 (i.e.
Output 2 only selected). Press SHIFT, Addr/Baud to display the currently selected instrument bus
address. Pressing Addr/Baud again displays the currently selected Baud rate and repeated
presses alternate between the two.
26
To change the address scroll through the available addresses using the Jog control when the
address is displayed. The address can be set between 1 and 31 inclusive. The address can also
be set from the instrument’s web pages. To change the Baud rate scroll through the available
rates (600 to 19200, factory default 9600) using the Jog control when the Baud rate is displayed.
Press OK to select the last displayed address and Baud rate or ESCAPE to retain the previous
selections.
Pin
Name
Description
By default all interfaces are live at all times. It is however possible to configure the privileges for a
particular interface to either “read only” or “no access” from the Web page interface. The status
of the interface privileges may be viewed, but not changed, from the front panel.
To view the interface privileges the CONTROL mode must first be set to 2 (i.e. Output 2 only
selected). Press SHIFT, Bus Ty pe to first display the status of the GPIB interface. To view the
privilege status of another interface, scroll through the interface list using the Jog control until the
required interface is displayed. The status will be shown as either
Remote/Local Operation
At power-on the instrument will be in the local state with the REM indicator off. In this state all
keyboard operations are possible. When the instrument is addressed to listen and a command is
received the remote state will be entered and REM will be turned on. In this state the keyboard is
locked out and remote commands only will be processed. The instrument may be returned to the
local state by pressing the LOCAL key; however, the effect of this action will only remain until the
instrument is addressed again or receives another character from the interface, when the remote
state will once again be entered.
RS232 Interface
RS232 Interface Connect o r
The 9-wa y D-type serial interface connector is located on the instrument rear panel. The pin
connections are as shown below:
Full, oFF or rEAd.
1 RI Passively asserted (+V through 10kΩ)
2 TXD Transmitted data from instrument
3 RXD Received data to ins trument
4 CTS
5 GND Signal ground
6 RTS Passively asserted (+V through 10kΩ)
7 DSR No internal c onnection
8 DTR
9 CD No internal connection
RS232 Connections
The RS232 interface should be connected to a standard PC port using a fully wired 1:1 malefemale cable without any cross-over connections. Alternatively, only pins 2, 3 and 5 need be
connected to the PC, but with links made in the connector at the PC end between pins 1, 4 and 6
and between pins 7 and 8, see diagram.
27
Start Bits: 1
Parity: None
Data Bits: 8
Stop Bits: 1
Baud Rate is set as described above in Addres s and Baud Rate Selection; the other parameters
are fixed as follows:
RS232 Character Set
Because of the need for XON/XOFF handshake it is possible to send ASCII coded data only;
binary blocks are not allowed. Bit 7 of ASCII codes is ignored, i.e. assumed to be low. No
distinction is made between upper and lower case characters in command mnemonics and they
may be freely mixed. In this manual 20H, etc. means 20 in hexadecimal
USB Interface
The USB interface is a virtual COM port which can be controlled by a PC as if it was a RS232
device. The instrument is supplied with a CD containing an .inf file for the standard Microsoft
drivers available in Windows 2000, XP, Vista and Windows 7; the installation wizard will install the
driver (32-bit or 64-bit) appropriate to the PC’s operating system. Any updates are available via
the TTi website, www.tti-test.com.
Installation of the interface driver is achieved by connecting the instrument to a PC via a standard
USB cable. The Windows’ plug and play functions should automatically recognise the addition of
new hardware attached to the USB interface and, if this is the first time the connection has been
made, prompt for the location of a suitable driver. Provided that the standard Windows prompts
are followed correctly Windows will install the appropriate driver and establish a virtual COM port
within the PC. The number of the new COM port will depend upon the number of co-existing COM
ports within the PC. The virtual COM port can be driven by Windows applications in exactly the
same way as a standard COM port, except that the Baud rate setting of the virtual COM port is
ignored.
The driver will remain installed on the PC so that the establishment of a virtual COM port is done
automatically each time the instrument is connected to the PC via USB in the future.
Further virtual COM ports are created for each additional instrument connected to the PC via
USB. Each instrument is assigned a separate virtual COM port when it is first connected and the
same COM port will be assigned each time that instrument is subsequently connected; the PC
software makes use of the unique code embedded in each instrument to link it to the same virtual
COM port irrespective of which physical USB port it is connected to.
Use can also be made of the ADDRESS? command to easily identify which instrument is being
controlled by a particular COM port. Although the addressing capability is ignored in USB
operation the address can still be set and used as an identifier; set each USB-connected
instrument to a different address and send the ADDRESS? command from each virtual COM port
to confirm which instrument is connected to that port.
LAN Interface
The LAN interface is designed to comply with the LXI standard version 1.2 and contains the
interfaces and protocols described below. Since it is possible to misconfigure the LAN interface,
making it impossible to communicate with the instrument over LAN, a LAN Configuration Initialise
(LCI) mechanism is provided via a recessed switch on the rear panel to reset the unit to the
factory default. The default setting is for the instrument to attempt to obtain settings via DHCP if
available or, if DHCP times out (30 seconds), via Auto-IP. In the very unlikely event that an Aut oIP address cannot be found a static IP address of 192.168.0.100 is assigned. Resetting the LAN
removes any password protection.
28
For more information on LXI standards refer to www.lxistandard.org/home
.
LAN Connection
To use the LAN interface, the IP address of the unit must be known. There is a LXI Discovery Tool
on the supplied CD-ROM which can be used to display the IP addresses (and other associated
information) of all connected devices that comply with the VXI-11 discovery protocol. This tool is
a Windows PC application that should be installed and run on the controlling PC with the unit
either connected directly to the PC network connector or via a router. Connecting via a router is
recommended as this is significantly quicker to assign an IP address; connecting directly to the
PC will begin to assign an IP address only after a 30 second DHCP timeout. Double clicking on
any entry in the list of devices discovered will open the PC's web browser and display the Home
page of that device.
There are also tools for LAN discovery included as part of the National Instruments Measurement
and Automation Explorer package and the Agilent Vee application.
The unit will, when first powered up, attempt to obtain settings via DHCP if available or, if DHCP
times out (30 seconds), via Aut o-IP. In the very unlikely event that an Auto-IP address cannot be
found a static IP address of 192.168.0.100 is assigned. During this time the LAN lamp will be lit
and it will stay lit if a LAN connection is successfully made. However, if a connection is still not
made by the end of the above process, or if the LAN connector is physically removed at any time,
the LAN lamp will go off; see LAN Error section for details.
Web Server; Configuration Password Protection
The unit contains a basic web server. This provides information on the instrument and allows it to
be configured. The Configure page can be password protected to deter unauthorised changes to
the remote operation configuration; the default configuration is ‘no password’.
The Configure page itself explains how to set the password. The password can be up to 15
characters long; note that the User Name should be left blank. The password will, however, be
reset to the default (no password) if the rear panel LAN RESET switch is used to reset all the LAN
parameters to their factory default.
The web pages also have an ‘Identify’ function which allows the user to send an identifying
command to the instrument which causes its displays to flash until the command is cancelled.
ICMP Ping Server
The unit contains an ICMP server allowing the instrument to be ‘pinged’ via either its host name or
IP address.
VXI-11 Discovery Protocol
The instrument has very limited support of VXI-11 which is sufficient for the discovery protocol
and no more.
The ins trume nt implements a S un RPC Port-mapper on TCP port 111 and UDP port 111 as
defined in RPC1183. The calls supported are: NULL, GET PORT and DUMP.
On TCP port 1024 a very simple VXI-11 protocol is implemented sufficient only for instrument
discovery. This implements the following ca lls : CREATE LINK, DEVICE_WRITE, DEVICE_READ
and DESTROY_LINK.
Once a link has been created anything written to the device is ignored and any read from the
device returns the identification string as would be expected from a “*IDN?” of the form
‘Manufacturer,Model,Serial No.,X.xx – Y.yy’
for example
SORENSEN, XDL 35-5TP, 279730, 1.00 – 1.00
where ‘X.xx’ is the revision of the main firmware and ‘Y.yy’ is the revision of the interface firmware.
Interface firmware is user field updateable via the USB port.
29
Loading...
+ 155 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.