Ametek DTI-1000, STS-100 Specifications

Page 1
t e m p e r a t u r e
Wide temperature range
DTI-1000: -200 to 750°C (-328 to 1382°F) STS-100: -150 to 650°C (-238 to 1202°F)
Improve your accuracy
DTI-1000: accuracy to ±0.005°C (±0.009°F) DTI-1000 + STS probe: ±0.03°C (±0.054°F) ± LSD
Reliable temperature readings
The measuring principle is a 4-wire True Ohm Measurement, which eliminates the EMF in cables, sockets, and sensors
Reference Digital Temperature Indicator
DTI-1000 and STS-100
Dual channel inputs
Sensor 1 and sensor 2, with differential value 1 - 2, peak hold etc.
Minimize paperwork
RS232 communication, special calibration and data storage software are included
Fast response time
Ensures correct monitoring of the tempera­ture stability
Specified low drift
Maintains a minimum uncertainty budget over the entire period between re-calibra­tion intervals
Wide selection of probes
Including 90° angled or cable probe, and 4 mm / 0.16 in or 6.35 mm / 0.25 in, offer­ing flexibility in test methods and sensors­under-test
Wherever there is a demand for reliable and accurate temperature mea­surement, you can rely on the JOFRA DTI-1000 reference thermometer and the JOFRA STS reference probes; backed on more than 50 years of experience.
The DTI-1000 with a STS-probe is a fully traceable thermometer recom­mended as the reference instrument to verify the true temperature in any type of temperature calibrator, liquid bath, or dry-block calibrator.
Use the DTI-1000 and the STS probes as your working temperature reference in any calibration application or use the set-up directly in cus­tody transfer applications where high accuracy (low uncertainty) means money.
The superior specifications combined with a long history of reliability and low drift have made the DTI-1000 and the STS probes the working stan­dard in many national laboratories worldwide.
The STS industrial temperature reference probes are built to last. All Superior Temperature Standard probes are economical and offer fast response times, low immersion depths, compact physical sizes, and specified low drift rates; even at high temperatures.
ISO 9001 Manufacturer
Specification Sheet, SS-DTI-1000
Page 2
The ON/OFF key automatically initiates a self-test routine and the display indicates tested param­eter results plus the date of the last calibration.
Sensor 1 connections
Sensor 2 connections
The MENU/ENTER key includes intelligent prompts that guide the user through setup and operation. This key allows the user to specify measuring units, resolution, sensor identification, and coefficients.
The RESET/SELECT key allows the user to enter peakholds and to change settings in the MENU mode.
Display with 2 * 20 characters, which continuously read out the two inputs.
The MODE key allows the user to specify temperature modes on the dual channel inputs; sensor 1 and sensor 2, with differential values 1-2, peak hold etc.
2 www.jofra.com
Carrying case (Optional) - 124944
The optional protective carrying case ensures safe transportation and storage of the instrument and associated equipment.
Digital Temperature Indicator • DTI-1000 & STS-100
Page 3
True Ohm Measurement
The DTI-1000 has been designed to employ state-of-the­art DC measuring techniques in combination with powerful microprocessor technology. To achieve high accuracy, the measuring principle used by the DTI-1000 is True Ohm Measurement thus eliminating the EMF from cables, sock­ets, and sensors.
Conversion to temperature
DTI-1000 will accept Callendar van Dusen coefficients R0, A, B and C or ITS-90 coefficients for converting resistance to temperature. If you do not have these coefficients avail­able from the certificate for the sensor, but have measure­ment data (temperature and resistance), JOFRACAL can help you calculate the coefficients required.
True Ohm Measurement is a proven method to achieve accurate compensation for errors induced by thermal effects.
The resistance is measured through the 4-wire system at 1 mA, after which the instrument takes a reading with­out any applied current; this second reading is the “error EMF”.
Combining a DTI-1000 and a STS reference sensor with the use of ITS-90 coefficients ensures the best overall accuracy.
Simplified operation
The DTI-1000 features an easy-to-read VFD display and dedicated function keys. All operations are performed from the instrument’s front panel.
Reference resistors
To minimize the effect of any drift caused by ambient temperature, humidity and /or aging, the DTI series is designed with built-in high precision and extremely stable reference resistors. This technique minimizes drift.
Linearity
To obtain high accuracy, it is necessary to know the char­acteristics of the Pt100 sensor to be used with the DTI­1000 e.g. one of the JOFRA STS probes. The DTI-1000 provides 3 different linearity setups:
Certified data pairs of ohm and reference temperature.
Use JOFRACAL for calculating and downloading of individual coefficients from a certified sensor e.g. a STS JOFRA probe based on “best curve fit”
Coefficients, according to IEC 751 (Callender van Dusen)
Coefficients according to ITS-90
DTI-1000 & STS-100 • Digital Temperature Indicator
www.jofra.com 3
Page 4
JOFRACAL CALIBRATION SOFTWARE
FUNCTIONAL SPECIFICATIONS
JOFRACAL calibration software ensures easy calibration of RTD´s, thermocouples, transmitters, thermoswithes, pressure gauges and pressure switches.
JOFRACAL can be used with DPC-500, HPC and IPI pres­sure calibrators, all JOFRA temperature calibrators, as well as JOFRA AMC910, ASC300 multi signal calibrator and ASM-800 signal multi scanner. When used with JOFRA ASM-800 signal multi scanner, JOFRACAL can perform a simultaneous semi automatic calibration on up to 24 pres­sure and/or temperature devices under test in any combi­nation.
JOFRACAL software controls the complete calibration procedure, stores the results and provides a calibration audit trail through hard-copy certificates. All calibration data are stored for each sensor to monitor drift and opti­mise recalibration intervals. A scheduler feature allows planning of future calibrations.
REQUIREMENTS JOFRACAL
Minimum hardware requirements:
Intel® Pentium® lI 1.4 GHz processor.
64MB RAM (128MB recommended)
80MB free disk space on hard disk (120MB
recommended) prior to installation
Standard VGA (800x600, 256 colours). 1024x768
recommended.
CD-ROM drive for installation of program
1 or 2 free RS-232 serial ports, depending on
configuration
Minimum software equirements:
Microsoft Windows® 98, Microsoft Windows® NT 4.0,
Microsoft Windows® 2000, Microsoft Windows® ME, Microsoft Windows® XP, Vista.
System fonts: MS Sans Serif and Arial
Input interface
DTI-1000 A (Pt100 / Pt25) ........................................ 0-360Ω
DTI-1000 B (Pt25) ..........................................................0-95
Accuracy, 12 months use
DTI-1000 A ......................................±(6 ppm rdg + 1.4 mΩ)
DTI-1000 B ...................................... ±(6 ppm rdg + 0.7 mΩ)
For accuracies in °C and °F please see table below
Power supply
Mains .........................................................9 VDC / 200 mA
Batteries ................................................ 8 x 1.5 V (type AA)
Battery life ...............................................15 hours typically
Mains specifications
Temperature range ............-200 to 750°C / -328 to 1382°F
Temperature coeffecient: .. outside 23°C ±3°C / 73°F ±5°F
.................................................... 0.8 ppm/°C / 0.45 ppm/°F
Input channels ................................................................... 2
Termination Goldplated LEMO / 4 mm banana test sockets Resolution (user-selectable) .. 0.1, 0.01 or 0.001 °C/°F/K/Ω
Update rate (0.1 / 0.01 / 0.001°) ............2 / 3 / 12 seconds
Measuring units .......................................... °C, °F, K and Ω
Measuring current .........................................................1mA
Instrument dimensions
L x W x H: ............ 225 x 135 x 195 mm / 8.9 x 5.3 x 7.7 in
Weight ............................................................2.2 kg / 4.9 lb
Shipping (+ std. accessories + carrying case)
L x W x H ...... 600 x 380 x 310 mm / 23.6 x 15.0 x 12.2 in
Weight ........................................................10.4 kg / 22.9 lb
Shipping (+ std. accessories but no carrying case)
L x W x H ........ 320 x 240 x 270 mm / 12.6 x 9.5 x 10.6 in
Weight: ..........................................................3.5 kg / 7.7 lb
Shipping (carrying case only)
L x W x H ...... 600 x 380 x 310 mm / 23.6 x 15.0 x 12.2 in
Weight: ........................................................6.9 kg / 15.2 lb
Miscellaneous
Serial data interface ................................................. RS232
Display: ..........VFD, two lines, 20 characters alphanumeric
Operating (ambient) temperature ....0 to 50°C / 32 to 122°F
Storage (ambient) temperature .... -20 to 60
Humidity ......................................................... 0 to 90% RH
o
C / -4 to 140oF
4 www.jofra.com
Digital Temperature Indicator • DTI-1000 & STS-100
Page 5
JOFRA STS REFERENCE SENSORS
To get an ideal reference system, JOFRA offers a range of reference sensors. JOFRA STS Superior Temperature reference Sensors are based on more than 50 years of in­dustrial temperature sensor manufacturing experience. The main requirement of a reference sensor is stability: The less the sensor drifts, the lower the measurement uncertainty. All JOFRA Superior Temperature Standard sensors are economical and offer fast response times, low immersion depths, compact physical sizes, and specified low drift rates: even at high temperatures. These are all important considerations when selecting a reference sensor.
All sensors are supplied with an accredited or traceable certificate stating the sensor coefficients.
System accuracy for JOFRA DTI-1000 with JOFRA STS-100/102
STS-100 probe:
-50 to 250°C (-58 to 482°F): ±0.030°C (±0.054°F) ±0.045°C (±0.081°F)
-50 to 320°C (-58 to 608°F): ±0.040°C (±0.072°F) ±0.050°C (±0.090°F)
-50 to 400°C (-58 to 752°F): ±0.060°C (±0.108°F) ±0.070°C (±0.126°F)
-50 to 650°C (-58 to 1202°F): ±0.060°C (±0.108°F)
±0.080°C (±0.144°F)
-90 to 650°C (-130 to 1202°F): ±0.080°C (±0.144°F)
±0.110°C (±0.198°F)
STS-102 cable sensor:
-45 to 155°C (-49 to 311°F): ±0.040°C (±0.072°F) ±0.070°C (±0.126°F)
All sensors are as standard supplied with an accredited certificate stating the sensor coefficients. A traceable certificate is optional.
Note 1: Specified at 95% confidence interval k=2, over full range, including calibration uncertainty, excluding 1 LSD (Least Significant Digit).
Note 2: Excl. sensor drift
Note 3: Incl. sensor drift after 100 hours at max. temperature.
1) 2)
1) 3)
1) 2)
1) 3)
1) 2)
1) 3)
1) 2)
1) 3)
1) 2)
1) 3)
1) 2)
1) 3)
Accuracy - 12 months
Temperature range
-200°C / -328°F ±0.004°C / ±0.006°F ±0.014°C / ±0.026°F ±0.007°C / ±0.013°F
0°C / -32°F ±0.005°C / ±0.009°F ±0.016°C / ±0.028°F ±0.009°C / ±0.015°F
-155°C / -311°F ±0.006°C / ±0.011°F ±0.017°C / ±0.031°F ±0.010°C / ±0.018°F
200°C / 392°F ±0.007°C / ±0.013°F ±0.018°C / ±0.032°F ±0.010°C / ±0.019°F
320°C / 608°F ±0.008°C / ±0.014°F ±0.019°C / ±0.035°F ±0.012°C / ±0.021°F
400°C / 752°F ±0.008°C / ±0.015°F ±0.020°C / ±0.037°F ±0.012°C / ±0.022°F
600°C / 1112°F ±0.010°C / ±0.019°F ±0.023°C / ±0.041°F ±0.014°C / ±0.026°F
650°C / 1202°F ±0.011°C / ±0.020°F ±0.024°C / ±0.043°F ±0.015°C / ±0.027°F
750°C / 1382°F ±0.012°C / ±0.021°F ±0.026°C / ±0.047°F ±0.017°C / ±0.030°F
DTI-1000 & STS-100 • Digital Temperature Indicator
DTI-1000 A with Pt-100 (excl. sensor uncertainty)
DTI-1000 A with Pt-25 (excl. sensor uncertainty)
DTI-1000 B with Pt-25 (excl. sensor uncertainty)
www.jofra.com 5
Page 6
ORDERING JOFRA DTI-1000 AND STS-100
Order No. Description Base model number
DTI-1000A DTI-1000 A, -200 to 750°C / -328 to 1382°F (Pt-25 or Pt-100) DTI-1000B DTI-1000 B, -200 to 750°C / -328 to 1382°F (only Pt-25)
Calibration certificate DTI-1000
F NPL traceable calibration certificate H Accredited calibration certificate (Option)
Channels
C1 One channel C2 Two channels
STS reference sensor
100 Pt100 reference probe, solid,
-150 to 650°C (-238 to 1207°F)
Diameter of the probe
A Overall diameter 4 mm (0.16 in) B Overall diameter 6.35 mm (0.25 in)
Shape and length
250 Straight probe, 250 mm (9.8 in) in alu case 350 Straight probe, 350 mm (13.8 in) in alu case 500 Straight probe, 500 mm (19.7 in) in alu case 901 90° angled probe, 207 mm (8.1 in) in plastic case
Calibration certificate
F NPL traceable calibration certificate on sensor H Accredited calibration certificate on sensor (standard) FS NPL traceable system calibration certificate HS Accredited system calibration certificate
Options
C System carrying case (max. 350 mm. probe)
DTI1000AFC1100A250HC Sample order number
JOFRA DTI-1000 A with NPL traceable calibration certificate, One channel and 4 mm STS-100 reference sensor, straight 250 mm. with accredited calibration certificate and carrying case.
SYSTEM CALIBRATION DTI-1000 AND STS-100
If you need a system calibration certificate for the DTI-1000 and STS-100 reference sensor you can choose between:
FS, which is a standard traceable system calibration certificate HS , which is a standard accredited system calibration certificate
Please specify in the above ordering structure.
A standard system calibration of a DTI-1000 and a STS-100 probe includes the following calibration points:
-45, -20, 0, 50, 100, 200, 320, 450, 650°C / -49, -4, 32, 122, 212, 392, 608, 842, 1202°F
6 www.jofra.com
Digital Temperature Indicator • DTI-1000 & STS-100
Page 7
ORDERING JOFRA DTI-1000 AND STS-102
Order No. Description Base model number
DTI-1000A DTI-1000 A, -200 to 750°C / -328 to 1382°F (Pt-25 or Pt-100) DTI-1000B DTI-1000 B, -200 to 750°C / -328 to 1382°F (only Pt-25)
Calibration certificate DTI-1000
F NPL traceable calibration certificate H Accredited calibration certificate (Option)
Channels
C1 One channel C2 Two channels
STS reference sensor
102 Pt100 reference probe, cable,
-50 to 155°C (-58 to 311°F)
Diameter of the probe
A Overall diameter 4 mm (0.16 in)
Shape and length
030 Short sensor 30 mm / 1.18 in in plastic case
Calibration certificate
F NPL traceable calibration certificate on sensor H Accredited calibration certificate on sensor (standard) FS NPL traceable system calibration certificate HS Accredited system calibration certificate
Options
C System carrying case (max. 350 mm. probe)
DTI1000AFC1102A030HC Sample order number
JOFRA DTI-1000 A with NPL traceable calibration certificate, One channel and 4 mm STS-102 reference sensor, short 30 mm with accredited calibration certificate and carrying case.
SYSTEM ACCESSORIES DTI-1000 AND STS-102
If you need a system calibration certificate for the DTI-1000 and STS-100 reference sensor you can choose between:
FS, which is a standard traceable system calibration certificate HS , which is a standard accredited system calibration certificate
Please specify in the above ordering structure.
A standard accredited system calibration of a DTI-1000 and a STS-102 probe includes the following calibration points:
-45, -20, 0, 50, 100, 155°C / -49, -4, 32, 122, 212, 311°F
DTI-1000 & STS-100 • Digital Temperature Indicator
www.jofra.com 7
Page 8
ORDERING INFORMATION DTI-1000
Order No. Description Base model number
DTI-1000A DTI-1000 A, -200 to 750°C / -328 to 1382°F Pt-25 or Pt-100 DTI-1000B DTI-1000 B, -200 to 750°C / -328 to 1382°F only Pt-25
Calibration certificate
F NPL traceable calibration certificate H Accredited calibration certificate (Option)
Options
C Carrying case, aluminum
DTI1000AFC Sample order number
JOFRA DTI-1000 A with NPL traceable certificate and carrying case.
STANDARD DELIVERY
DTI-1000
JOFRACAL calibration software
AmeTrim ATC/DTI to adjust the DTI-1000
RS232 cable, 9 pin connector
8 batteries
User manual
Calibration certificate, traceable to International Standards
Mains adapter
ACCESSORIES
124716 4x 1,5 Volt rechargeable batteries * The DTI-1000 requires 8 x 1.5 Volt batteries 124718 Charger for rechargeable batteries 115/230 VAC 124944 Carrying Case for DTI-1000 * (May also hold two connected STS sensors up to 350 mm) 124315 Removable trolley for carrying case 125002 Edgeport Converter with 4 pcs of RS232 ports 124720 Mains adapter 9VDC/200mA - 230VAC/115VAC
AMETEK Calibration Instruments
is one of the world’s leading manufacturers and developers of calibration instruments for temperature, pressure and process signals as well as for temperature sensors both from a commercial and a technological point of view.
JOFRA Temperature Instruments
Portable precision thermometers. Dry-block and liquid bath calibrators : 5 series, with more than 25 models and temperature ranges from - 90° to 1205°C / -130° to 2200°F. All featuring speed, por tability, accuracy and advanced documenting functions with JOFRACAL calibration sof tware.
JOFRA Pressure Instruments
Convenient electronic systems ranging from -25 mbar to 1000 bar (0.4 to 15,000 psi) - multiple choices of pressure ranges, pumps and accuracies, fully temperature-compensated for problem-free and accurate field use.
JOFRA Signal Instruments
Process signal measurement and simulation for easy control loop calibration and measurement tasks - from handheld field instruments to laboratory reference level bench top instruments.
JOFRA / JF Marine Instrum ents
A complete range of calibration equipment for temperature, pressure and signal, approved for marine use.
FP Temperature Sensors
A complete range of temperature sensors for industrial and marine use.
M&G Pressure Testers
Pneumatic floating-ball or hydraulic piston dead weight testers with accuracies to 0.015% of reading.
M&G Pumps
Pressure generators from small pneumatic “bicycle” style pumps to hydraulic pumps generating up to 1,000 bar (15,000 psi).
AMETEK Denmark A/S Gydevang 32-34 | 3450 Allerød | Denmark T: +45 4816 8000 | ametek@ametek.dk
Sales & Service: Europe, Asia, Africa, Middle East and South America
Information in this document is subject to change without notice. ©2010, by AMETEK, Inc., www.ametek.com. All rights reserved.
8 www.jofra.com
.. .beca use calib ratio n is a ma tter of c onfid ence
AMETEK Mansfield & Green (North America)
T: +1 800 527 9999 | cal.info@ametek.com
AMETEK Singapore Pte. Ltd. (Singapore) T: +65 6 484 2388 | aspl@ametek.com.sg
AMETEK Inc. Beijing Rep. Office (China)
T: +86 10 8526 2111 | jofra@ametek.com.cn
AMETEK GmbH (Germany) T: +49 2159 9136 510 | info.mct-de@ametek.de
AMETEK Calibration Instruments (UK) T: +44 (0) 1243 833 302 | jofra@ametek.co.uk
www.jofra.com
Pub code SS-DTI-1000 Issue 1011
Loading...