Ametek ITC 155, ITC 320, ITC 650 Operating Manual

Reference Manual
Temperature Calibrator
JOFRA ITC-155/320/650 A
Copyright 2002 AMETEK DENMARK A/S
The structure of the manual
This reference manual is aimed at users who are familiar with Ametek calibrators, as well as those who are not. The manual is divided into 10 chapters which describe how to set up, operate, service and maintain the calibrator. The technical specifications are described and accessories may be ordered from the list of accessories.
Along with the calibrator, you should have received a multi­lingual user manual which sets out the operating instructions for the instrument. It is designed to provide a quick reference guide for use in the field.
Safety symbols
This manual contains a number of safety symbols designed to draw your attention to instructions which must be followed when using the instrument, as well as any risks involved.
Warning
Events which may compromise the safe use of the instrument and result in considerable personal or material damage.
Caution…
Events which may compromise the safe use of the instrument and result in slight personal or material damage.
Note…
2 17-08-2004 123312 03
Special situations which demand the user’s attention.
List of contents
Introduction ..............................................................................4
1.0
2.0 Safety instructions ...................................................................6
3.0 Setting up the calibrator ........................................................11
3.1 Receipt of the calibrator.............................................................. 11
3.2 Preparing the calibrator .............................................................. 13
3.3 Choice of insertion tube .............................................................. 14
3.4 Inserting the sensor .................................................................... 16
4.0 Operating the calibrator.........................................................19
4.1 Keyboard, display and connections............................................ 19
4.2 Starting the calibrator.................................................................. 22
4.3 Selecting the set-temperature .................................................... 22
4.4 Using the SWITCH TEST ........................................................... 24
4.5 Using the AUTO STEP ............................................................... 28
4.6 Using the MENU ......................................................................... 31
4.6.1 Adjusting the temperature unit ....................................... 32
4.6.2 Adjusting the max-temperature ...................................... 33
4.6.3 Adjusting the SWITCH TEST slope rate ........................ 33
4.6.4 Adjusting the extra stability time..................................... 34
4.6.5 Adjusting the temperature resolution ............................. 34
4.7 Simulation/training ...................................................................... 35
5.0 Storing and transporting the calibrator ...............................36
6.0 Errors.......................................................................................38
7.0 Setting the main voltage and replacing the fuses..............40
7.1 Returning the calibrator for service............................................. 41
8.0 Maintenance............................................................................43
8.1 Cleaning...................................................................................... 43
8.2 Adjusting and calibrating the instrument..................................... 44
8.2.1 Adjusting the calibration date ......................................... 46
8.2.2 Calibrating/adjusting the instrument............................... 47
9.0 Technical specifications........................................................51
10.0 List of accessories.................................................................59
123312 03 17-08-2004 3
1.0 Introduction
Congratulations on your new Ametek Jofra ITC Calibrator!
With the Ametek Jofra calibrator, you have chosen an extremely effective instrument which we hope will live up to all your expectations. Over the past many years, we have acquired extensive knowledge of industrial temperature calibration. This expertise is reflected in our products which are all designed for daily use in an industrial environment. Please note that we would be very interested in hearing from you if you have any ideas or suggestions for changes to our products.
This reference manual applies to the following instruments:
Jofra ITC-155 A
Jofra ITC-320 A
Jofra ITC-650 A
ISO-9001 certified
)
4 17-08-2004 123312 03
Ametek Denmark A/S was awarded the ISO-9001 certificate in September 1994 by BVQI - Bureau Veritas Quality International.
) CE-label
Your new calibrator bears the CE label and conforms to the EMC directive and the Low­voltage Directive.
) Technical assistance
Please contact the dealer from whom you acquired the instrument if you require technical assistance.
) Guarantee
According to current terms of sale and delivery.
This guarantee only covers defects in manufacture and becomes void if the instrument has been subject to unauthorised intervention and/or misuse.
123312 03 17-08-2004 5
2.0 Safety instructions
Read this manual carefully before using the instrument!
Please follow the instructions and procedures described in this manual. They are designed to allow you to get the most out of your calibrator and avoid any personal injuries and/or damage to the instrument.
Warning……
About the use:
The calibrator must not be used for any purposes
other than those described in this manual.
The calibrator is designed for interior use only and
should not be used in risk-prone areas, where vapour or gas leaks, etc. may constitute an explosives hazard.
The calibrator must be kept clear within an area of
20 cm on all sides and 1 metre above the calibrator.
Never use heat transfer fluids such as silicone, oil,
paste, etc.
These fluids may penetrate the calibrator and cause
damage or create poisonous fumes.
The calibrator must be switched off before any
attempt to service the instrument is made.
When cleaning the well, REMEMBER, wear goggles
when using compressed air!
About the frontpanel:
The switch-test connectors, on the front panel of the
calibrator, must NEVER be connected to a voltage source.
Thermostats must not be connected to any other
voltage source during a test.
6 17-08-2004 123312 03
About insertion tubes and insulation plugs:
Never leave hot insertion tubes which have been
removed from the calibrator unsupervised – they may constitute a fire hazard.
If you intend to store the calibrator in the optional aluminium carrying case after use, you must ensure that the instrument has cooled to a temperature below 100°C/212°F before placing it in the carrying case.
About the fuses:
The fuse box must not be removed from the power
control switch until the mains cable has been disconnected.
The two main fuses must be identical and correspond
to the chosen voltage.
123312 03 17-08-2004 7
Caution – Hot surface
This symbol is engraved in the grid plate.
Do not touch the grid plate, the well or the insertion
tube as the calibrator is heating up – they may be very hot.
Do not touch the tip of the sensor when it is
removed from the insertion tube/well – it may be very hot.
Do not touch the handle of the calibrator during use
– it may be very hot.
Over 100°C/212°F
If the calibrator has been heated up to temperatures above 100°C/212°F, you must wait until the instrument reaches a temperature below
100°C/212°F before you switch it off.
Below 0°C/32°F (applies only to the ITC-155 A model)
If the calibrator has reached a temperature below 0°C/32°F, ice crystals may form on the insertion tube and the well. This, in turn, may cause verdigris to form on the material. To prevent this from happening, simply heat up the calibrator to 50°C/122°F.
8 17-08-2004 123312 03
Caution…
About the use:
Do not use the instrument if the ventilator is out of
order.
Before cleaning the calibrator, you must switch it off,
allow it to cool down and remove all cables.
About the well, insertion tube and grid plate:
The well and the insertion tube must be clean before
use.
Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes carefully when not in use.
The insertion tube must never be forced into the
well. The well could be damaged as a result, and the insertion tube may get stuck.
Before using new insertion tubes for the calibration,
the insertion tubes must be heated up to maximum temperature - 320°C (608°F) / 650°C (1202°F) - for a period of minimum 30 minutes.
In order to ensure the best calibration of your sensors please avoid using insertion tubes in the ITC-320 calibrator which have been used in the ITC­ 650 calibrator.
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause verdigris to form
on the insertion tube inside the instrument. There is a risk that the insertion tube may become stuck if this is allowed to happen.
If the calibrator is to be transported, the insertion
tube must be removed to avoid damage to the instrument.
123312 03 17-08-2004 9
Note…
The product liability only applies if the instrument is subject to a manufacturing defect. This liability becomes void if the user fails to follow the maintenance instructions set out in this manual or uses unauthorised spare parts.
10 17-08-2004 123312 03
3.0 Setting up the calibrator
3.1 Receipt of the calibrator
When you receive the instrument…
Carefully unpack and check the calibrator and the accessories.
Check the parts off against the list shown below.
If any of the parts are missing or damaged, please contact the dealer who sold the calibrator.
You should receive:
1 calibrator
S T N
E
M U R
T
S N
I
N
O I T A
R
B I L A C
1 mains cable
1 set of test cables (1 black, 1 red)
1 insertion tube
3 pcs. insulation plugs for 5, 8, 11 mm sensors
(ITC-155 A only)
123312 03 17-08-2004 11
1 tool for insertion tube
1 traceable certificate
1 reference manual
Reference Ma nual
1 user manual
User Manual
1 RS 232 serial cable
1 CD-ROM containing software package
“JofraCal”.
When reordering, please specify the parts number found in the list of accessories, section 10.0.
12 17-08-2004 123312 03
3.2 Preparing the calibrator
Warning
The calibrator must not be used in areas prone to
explosives hazards.
The calibrator must be kept clear within an area of 20
cm on all sides and 1 metre above the calibrator.
Note…
The instrument must not be exposed to draughts.
S T N
E
M U R T
S
N I N
O I
T A R
B I
L A C
Fig. 1
When setting up the calibrator, you must…
Place the calibrator on an even horizontal surface in the spot you intend to use it.
123312 03 17-08-2004 13
Caution…
Do not use the instrument if the ventilator is out of order.
g
The calibrator is now ready for use.
Ensure a free supply of air to the ventilator located at the bottom of the instrument.
Check the voltage of the power control switch (on/off switch (230V/115V)). If the voltage of the power control switch differs from the line voltage, you must adjust the voltage of the power control switch as follows (see Fig. 1):
A. Open the fuse box lid using a screwdriver. B. Take out the fuse box. C. Remove both fuses and insertion tube two new fuses.
These must be identical and should correspond to the line voltage. See section 10.0.
B. Turn the fuse box 180° and slide it into place.
Check that the earth connection for the instrument is present and attach the cable.
Select an insertion tube with the correct bore diameter. See section 3.3 for information on how to select insertion tubes.
3.3 Choice of insertion tube
Caution…
Before using new insertion tubes for the calibration, the
14 17-08-2004 123312 03
insertion tubes must be heated up to maximum temperature
- 320°C (608°F) / 650°C (1202°F) - for a period of minimum 30 minutes.
In order to ensure the best calibration of your sensors please avoid using insertion tubes in the ITC-320 calibrator which have been used in the ITC-650 calibrator.
Insertion tubes are selected on the basis of the diameter of the sensor to be calibrated.
Use the table for insertion tubes in section 10.0 to find the correct part number.
Alternatively, you may order an undrilled insertion tube and drill the required hole yourself. The finished dimension should be as follows:
Sensor diameter +0.2
+0.05/-0 mm.
123312 03 17-08-2004 15
3.4 Inserting the sensor
Before inserting the sensor and switching on the calibrator, please note the following important warning:
Insert the sensor as shown in Fig. 2.
Fig. 2
In order to spare the sensor and its connections it is recommended to use a heat protection shield (104216) at high temperatures (see Fig.
3).
Warning
Never use heat transfer fluids such as silicone, oil,
paste, etc.
These fluids may penetrate the calibrator and cause
damage or create poisonous fumes.
S T N
E
M U R
T
S N I
N O I T
A
R B I L
A C
16 17-08-2004 123312 03
Sensor under test
Thermal protection shield
Insertion tube
S
T N E
M U R
T
S N I
N O
I
T A R
B I L
A C
Fig. 3
Caution…
The well and the insertion tube must be clean before
use.
Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes carefully when not in use.
The insertion tube must never be forced into the well.
The well could be damaged as a result, and the insertion tube may get stuck.
123312 03 17-08-2004 17
Caution – Hot surface
Do not touch the grid plate, the well or the insertion tube
as the calibrator is heating up – they may be very hot.
Do not touch the tip of the sensor when it is removed
from the insertion tube/well – it may be very hot.
Do not touch the handle of the calibrator during use – it
may be very hot.
18 17-08-2004 123312 03
4.0 Operating the calibrator
4.1 Keyboard, display and connections
Keyboard
Fig. 4
Pos. Description
CALIBRATION INSTRUMENTS
SWITCH
TEST
AUTO STEP
ESC
MENU
 
LCD.
SWITCH TEST button used to activate SWITCH TEST. The function automatically detects the opening/closing temperatures for thermostats.
AUTO STEP button used to activate AUTO STEP. The function is used to switch between a series of set­temperatures automatically.
ESC/MENU button used as Escape key or to activate the menu system (hold button down for min. 2 seconds).
 
ENTER button used to accept chosen options.
DOWN ARROW button used to adjust temperature values (value decreases) and to select menu options.
123312 03 17-08-2004 19
Loading...
+ 42 hidden pages