HPC600-2A 2 bara / 30 psia
HPC600-2C 2 bar / 30 psi
HPC600-10A 10 bara / 150 psia
HPC600-10C 10 bar / 150 psi
HPC600-020C 20 bar / 300 psi
» High accuracy
±0.020% of reading +0.015% F.S.
» True field calibration
Fully temperature compensated
Handheld Pressure Calibrator
HPC600
Electric pump
to 20 bar / 300 psi
» Unique built-in electrical pump
HPC600 has a built-in automatic pressure
system, based on an electrical pressure and
vacuum pump.
» Complete 4-20 mA loop calibrator
HPC600 features a complete mA loop calibrator, measure, source, sing and 24 volt
loop supply.
» Comprehensive set of features
Wide selection of useful functions and features. Damping, leak test, error calculation,
min/max, switch test, etc.
» Thermometer
High accuracy thermometer, Pt100 sensor.
» External pressure modules
Many different pressure modules, accuracy
up to 0.01% F.S.
The JOFRA HPC600 calibrator features a built-in electrical pump system,
for generation of pressure and vacuum, with a touch of a single key!
Besides the unique pump system, the HPC600 offers a complete suite of
features and functions, making it an ideal tool and calibrator.
A true multifunction solution, designed to save time, money and space…
HPC600 is a pressure indicator, a pressure calibrator, a complete mA
loop calibrator, including 24 volt supply, a voltmeter, a thermometer and
a pressure/vacuum generator. All in a ruggedized housing, with an easy
intuitive operating system, in the palm of your hand.
The HPC600 is delivered as a ready-to-test system, complete with hose,
fittings, and padded soft case, equipped to meet any need for pressure
calibration. International traceable calibration certificate, including vacuum and electrical parameters are a part of the standard delivery, to make
the HPC600 solution truly ready-to-use.
The HPC600 can be used in a very broad range of applications from
simple tool type jobs to complex calibration jobs in custody transfer systems. A time and money saver!
HPC600 is designed to meet high accuracy pressure calibration applications and is a multifunction tool. The HPC600 offers features such as
user configurable information display, 15 different pressure units, transmitter supply, mA input/output, % error calculation, voltage measurement, serial communication, and external pressure module capability.
The accuracy of the HPC600 calibrators is specified in % of reading to
ensure an even better accuracy and wider applicable pressure range.
The HPC600 is temperature compensated from -10 to 50ºC / 14 to 122ºF
for on-site operation. It is a truly superior pressure calibrator for laboratory and field use, bringing laboratory accuracy into the field.
The JOFRA APM series of pressure modules extends the application
base of the HPC calibrators by allowing calibrations in additional ranges.
ISO 9001 Manufacturer
Specification Sheet
SS-HPC600
Temperature measurement
High accuracy temperature measurement port by use of
the optional Pt100 sensor. A superb instrument for easy
and convenient temperature measurement.
Pressure connections
1/8” Combined NPT/BSP female. 300 series stainless steel.
1/4” NPT male and 1/4” BSP female adapters are supplied
as standard.
Intuitive menu system
Combination of “soft keys” and cursor keypad ensures an
easy operation.
ON / OFF
Auto shut-off to expand battery life time (user programmable).
Clear graphical display
Large backlit graphical display shows current status and
mode, 1, 2 or 3 measurement windows, to suit your applications and demands.
HPC calibrators are well equipped and include inputs for
mA, voltage, including 24 VDC supply for mA transmitters,
and source and sink for mA loop calibration and test.
Serial interface
RS232 connection for JOFRA APM pressure module or a
computer.
Units
15 different pressure engineering units.
Full set of features
Perform semi-automatic pressure switch tests, leak test,
% error calculation, switch in HART resistor or other calibrations tasks through the outstanding functions and features of the HPC600.
JOFRA APM pressure module
The JOFRA APM external pressure modules extend the
range of the JOFRA HPC. There are more than 35 models
available with gauge, absolute, differential, and vacuum
pressure references, in metric and imperial engineering
units.
The modules are engineered for in-plant, field, or laboratory use. They are ready-to-use with the JOFRA HPC and
the protocol allows for immediate recognition and use of
the module once it is plugged into the calibrator.
APM
Mk.II
Type
BarPSI%rdg%FS
FromToFromTo
Differential
025MD-0.0250.025-0.40.4-0.100%
075MD-0.0700.07-11-0.050%
350MD-0.3500.35-55-0.050%
Compound
001C-0.9601-14150.025% 0.010%
002C-0.9602-14300.025% 0.010%
007C*-0.8207-121000.025% 0.010%
020C*-0.82020-123000.025% 0.010%
035C*-0.82035-125000.025% 0.010%
Gauge
001G*010150.025% 0.010%
002G*020300.025% 0.010%
007G*0701000.025% 0.010%
020G*02003000.025% 0.010%
035G*03505000.025% 0.010%
070G*070010000.025% 0.010%
100G*0100015000.025% 0.010%
200G*0200030000.025% 0.010%
350G*0350050000.025% 0.010%
400G*0400060000.025% 0.010%
700G*07000100000.025% 0.015%
Absolute
001A0.0251.10.35160.025% 0.010%
003A0.0253.50.35500.025% 0.010%
007A*0.070711000.025% 0.010%
020A*0.0702013000.025% 0.010%
Pressure range
12 month
accuracy
* Stainless steel isolated pressure sensor
Specified temperature range 18°C to 28°C / 65°F to 82°F
F.S. (full scale) is the numerical value of the positive pressure
range.Accuracy includes hysteresis, nonlinearity, repeatability and
reference standard uncertainty, 1 Year typical long-term stability,
operated inside the rated temperature span and pressure range.
Requiring frequently zeroing (Gauge/diff.) or entering of reference
pressure (Absolute).
F.S. (full scale) is the numerical value of the positive pressure range.
Accuracy includes hysteresis, nonlinearity, repeatability and reference
standard uncertainty, 1 Year typical long-term stability, operated inside the
rated temperature span and pressure range. Requiring frequently zeroing
(gauge/compound) or entering of reference pressure (absolute).
Pressure accuracy ambient temp. (-10 to 50°C / 14 to 122°F)
All pressure ranges ...............±0.030% RDG + 0.015% F.S.
F.S. (full scale) is the numerical value of the positive pressure range.
Accuracy includes hysteresis, nonlinearity, repeatability and reference
standard uncertainty, 1 Year typical long-term stability, operated inside the
rated temperature span and pressure range. Requiring frequently zeroing
(gauge/compound) or entering of reference pressure (absolute).
This table shows the resolutions that can be obtained by the HPC calibrators throughout all engineering units.
HPC600 Ranges and Display Resolution
Range 2.0 bar / 30 psi10 bar / 150 psi20 bar / 300 psi
Engineering unitFactorFS valueFS valueFS value
psi130.000150.00300.00
bar0.068947572.068410.342120.684
mbar68.947572068.410342.120684
kPa6.894757206.841034.212068.4
MPa.006894760.20681.034212.0684
2
kg/cm
O@4°C70.30892109.310546.321093
cmH
2
O@20°C70.43362113.010565.021130
cmH
2
O@4°C703.08921093n/an/a
mmH
2
O@20°C704.33621130n/an/a
mmH
2
O@4°C27.68067830.424152.18304.2
inH
2
O@20°C27.72977831.894159.58318.9
inH
2
O@60°F27.70759831.234156.18312.3
inH
2
mmHg@0°C51.715081551.57757.315515
inHg@0°C2.0360261.081305.40610.81
Burst pressure20 bar / 300 psi20 bar / 300 psi40 bar / 600 psi
Proof pressure4 bar / 60psi14 bar / 200 psi28 bar / 400 psi
0.070306972.109210.546021.092
Proof pressure - maximum allowable pressure without a shift in calibration
Burst pressure - sensor damaged or destroyed, some risk of personnel injury
Absolute ranges - the data for the 2 bar / 30 psi and 10 bar / 50 psi ranges also applies to the absolute pressure versions
of those ranges
002C-0.82 bar / -12 psi to 2 bar / 30 psi gauge/compound
010A0.2 bar / 3 psi to 10 bar / 150 psi absolute
010C-0.82 bar / -12 psi to 10 bar / 150 psi gauge/compound
020C-0.82 bar / -12 psi to 20 bar / 300 psi gauge/compound
Calibration certificate
GNIST traceable calibration certificate
Accessories (Optional)
BRechargeable battery pack
IInterface kit LEMO to RS232
TTemperature sensor Pt100
Sample order number
HPC600010CGBITHPC600, 10 bar, NIST traceable certificate, rechargeable
battery package, interface kit, temperature sensor
STANDARD DELIVERY (HPC Calibrator)
• HPC600 calibrator
• Padded soft carrying case
• Traceable calibration certificate (NIST) with pressure
vacuum and electrical performance
• 8 x 1.5 volt batteries
• Adapter to 1/4” NPT male from 1/8” NPT male*
• Adapter to 1/4” BSP female from 1/8” NPT male*
• Quick connector set and 1 meter / 3 feet hose
• Test leads: red and black
• User manual
ACCESSORIES
Rechargable battery kit - Option B
The HPC600 is available with a set of 8 rechargeable AA
batteries and a charger for 4 batteries.
Interface kit - Option I
Special LEMO to Sub D, RS232 interface cable for
HPC600.
Temperature sensor - Option T
To utilize the themometer utility of the HPC600, a ruggedized handheld temperature sensor is available. The
sensor comes as a ready-to-use probe with handle and
coiled cable with LEMO connector for HPC600. The sensor is buildt with a high accuracy Pt100 element, accuracy
±0.15ºC / ±0.27ºF. Sensor dimensions, without handle and
cable, Ø4x200 mm/stainless steel.
Complete calibration systems
The HPC600 is delivered as a ready-to-test system, complete and equipped to meet your pressure calibration
needs.
JOFRACAL calibration software ensures easy calibration of RTD´s,
thermocouples, transmitters, thermoswithes, pressure gauges and
pressure switches. JOFRACAL can be used with all JOFRA calibration
instruments. JOFRACAL does not support the 20 bar version. For date of
availability contact your local representative. When used with ASM-800
signal multi scanner, JOFRACAL can perform a simultaneous semi
automatic calibration on up to 24 pressure and/or temperature devices
under test in any combination.
JOFRACAL software controls the complete calibration procedure, stores
the results and provides a calibration audit trail through hard-copy
certificates. All calibration data are stored for each sensor to monitor drift
and optimise recalibration intervals. A scheduler feature allows planning
of future calibrations.
Dirt, moisture or liquid?
If particles or liquid is expected in the system to be calibrated, we
recommend using our”cleaning filter”. The filter is attached with out use
of tools, the build in o-rings ensure a tight system. The clear see-through
design, makes it easy to see when cleaning is needed.
Part number 127887
A business unit of AMETEK Measurement & Calibration
AMETEK Test & Calibration Instruments
Technologies Division offering the following industry
leading brands for test and calibration instrumentation.