American Magnetics 286 User Manual

EXCELLENCE IN MAGNETICS AND CRYOGENICS
AMI
MODEL 286 LIQUID LEVEL CONTROLLER
INSTALLATION, OPERATION, AND
MAINTENANCE INSTRUCTIONS
American Magnetics, Inc.
PO Box 2509, 112 Flint Road, Oak Ridge, TN 37831-2509, Tel: 865 482-1056, Fax: 865 482-5472
May 2005
Application of Council Directives: Low Voltage Directive 72/23/EEC
EMC Directive 89/336/EEC
Manufacturer’s Name: American Magnetics, Inc.
Manufacturer’s Address: 112 Flint Road,
P.O. Box 2509 Oak Ridge, TN 37831-2509 U.S.A.
Type of Equipment: Liquid Level Instruments Model Numbers: Model 286
Standar ds to which Conformity is Declared:
Safety: EN 61010-1 (1993) w/A1, A2 EMC: EN55022 (1998) Class A
EN61326 (1997) / EN61000-4-2 (1995) 8kV AD, 4kV CD EN61326 (1997) / EN61000-4-3 (1996) 1V/m EN61326 (1997) / EN61000-4-4 (1995) 1kV Power Supply
0.5kV I/O cables EN61326 (1997) / EN61000-4-5 (1995) 1kV CM, 0.5kV DM EN61326 (1997) / EN61000-4-6 (1996) 1V EN61326 (1997) / EN61000-4-11 (1994) 100% interruption - 10ms
I, the undersigned, hereby declare that the equipment specified above complies with the require­ments of the aforementioned Directives and Standards and carries the "CE" mark accordingly.
Gregory J. Laughon September 12, 2002 Quality Assurance Manager
American Magnetics, Inc. Oak Ridge, TN, U.S.A.
Model 286 Liquid Level Controller
Model 286 Configuration AMI Order Number:_____________________ Shipping Date:_________________________ Model 286 Serial #:______________________ Firmware Revision:_____________________ Input Power Requirements:___________________________________________________________ Calibration Passcode:____________________ Configuration Notes: _____________________________________________________________________________________ _____________________________________________________________________________________ _____________________________________________________________________________________ _____________________________________________________________________________________ _____________________________________________________________________________________ _____________________________________________________________________________________ _____________________________________________________________________________________ _____________________________________________________________________________________
AMI Warranty
All products manufactured by AMI are warranted to be free of defects in materials and workmanship and to perform as specified for a period of one year from dat e of shipme nt. In the event of failure occur ring during n ormal use, AMI, at its o ption, will r epai r or r eplace al l products or components that fail under warranty, and such repair or replacement shall constitute a fulfillment of all AMI liabilities with respect to its products. Since, however, AMI does not have control over the installation conditions or th e use to which its products are put, no warranty can be made of fitness for a particular purpose, and AMI cannot be liable for special or consequential damages. All warranty repairs are F.O.B. Oak Ridge, Tennessee, USA.
Copyright © 2005 by American Magnetics, Inc., Oak Ridge, Tennessee, USA
Rev. 1
Table of Contents
1 Introduction.............................................................................. 1
1.1 Model 286 Fea tu res........ .. ............... ................ ............... ........... 1
1.2 Front Panel Layout ................................................................... 3
1.3 Rear Panel La yo u t.......... .. .. ................ ............... ................ .. ...... 4
1.4 Model 286 Spe ci f i ca t i on s @ 25 °C .............. ............... ............. 6
1.5 Controller Modes Description................................................... 7
1.5.1 Normal Mode ................................................................. 7
1.5.2 Auto-Changeover Mode................................................. 8
1.5.3 Pre-Cool Mode..... ............... .. ................ ............... ........... 9
2 Installation.............................................................................. 11
2.1 Unpacking the Instrument..................................................... 11
2.2 Rack Mounting the Instrument ............................................. 12
2.3 Installing th e Se n sors in the Vess el s ............................... ...... 12
2.4 Connecting the Oscillator Cables to the AMI Sensors..........13
2.5 Connecting C and D Sensor Inputs to the Oscillator............ 13
2.6 Installing the Optional Solenoid-operated Fill Valves..........14
2.7 Connecting to Optional Analog Outputs................................ 15
2.8 Connecting to Communication Options ................................. 15
2.9 Connecting th e In st r u me n t to P ower.... ................ .. ............... 15
3 Calibration .............................................................................. 17
3.1 Calibration C on c e pt s............... ............... .. ................ ............... 17
3.2 Calibration Me th o d s........................... .. ............... ................ .... 19
3.2.1 Open Dewar Ca l i b ra t i on ............ .. ................ ............... 21
3.2.2 Closed Dewar Cal i b ra tion........ ............... ................ .... 22
3.2.3 Approximate Calibratio n ............. ............................. .. 26
3.3 Calibration Me n u Reference........... ... .. ............... ................ .. .. 29
3.3.1 CAL Menu and Passcode Entry..................................30
3.3.2 Sensor Input Se l ec t io n ........ ... ............... ...................... 3 0
3.3.3 Action for Sele ct ed Input..... ................ ............... ......... 3 0
3.3.4 Select Calib ration Sub-M e n u..................... ................. 31
3.3.5 Perform Calibration Sub-Menu .................................. 33
4 Operation................................................................................ 37
4.1 Operational Guidelines........................................................... 37
4.1.1 Energize the In strument..... ................ .. ................ ...... 37
4.1.2 Assign Sensor In p ut s to the Channels ... ................ .. .. 38
i
Table of Contents
4.1.3 Select the Appropriate Uni t s.......... ................ .............3 8
4.1.4 Configure the HI an d L O Ala r ms................... .. ...........38
4.1.5 Configure the A SETPOINT and the B SETPOINT...39
4.1.6 Select the Con t ro l l e r Mo de ..... ............................ .........40
4.2 Sensor Contamination .............................................................42
4.3 Menu Reference........................................................................43
4.3.1 Accessing Menus ..........................................................43
4.3.2 HI Menu...... .. ................ .. ............... ................ .. .............4 4
4.3.3 LO Menu. .. ............... ................ .. ............... ................ ....45
4.3.4 A Menu..........................................................................45
4.3.5 B Menu..........................................................................46
4.3.6 RATE Menu..................................................................46
4.3.7 FILL Menu in Norm a l M od e........... ... .. ............... ... ......46
4.3.8 FILL Menu in Auto-Changeover Mode. .. ... ............... .. 4 8
4.3.9 FILL Menu in Pre- C ool M o de... .. .. ................ .. .............50
4.3.10 ASN Menu .................... ............... ................ ............... .. 5 3
4.3.11 MODE Menu ...............................................................53
4.3.12 MISC Menu... ................ .. ............... ................ ............... 5 4
4.3.13 CAL Menu.....................................................................55
4.4 Feature Matrix.........................................................................56
5 Remote Interface Reference..................................................57
5.1 SCPI Command Summary ......................................................57
5.2 Programming Overview...........................................................61
5.2.1 SCPI Language Introduction.......................................61
5.2.2 SCPI Status System.....................................................61
5.2.3 Standard Event Registers............................................64
5.2.4 Alarm Registers............................................................65
5.2.5 Measurement Re gi st e rs ............ .. ................ .. ...............67
5.2.6 Command Handshaking ..............................................68
5.2.7 Remote Units.................. ............... ................ ............... 6 9
5.3 RS-232/422 Configuratio n.................................. .....................69
5.3.1 Serial Port Connector...................................................70
5.3.2 Termination Characters...............................................70
5.3.3 Device Clear.................... ............... ............................. ..71
5.4 Command Reference ................................................................71
5.4.1 System-Related Commands.........................................71
5.4.2 Status System Commands...........................................72
5.4.3 Alarm Configuration Commands and Queries...........74
5.4.4 Level and Usag e Que r i es .............. ................ ............... 7 5
5.4.5 Fill Modes Control and Query Commands .................75
5.4.6 Assignment C o mmands and Queri e s ...... ... .. ............... 7 8
5.4.7 Calibration C ommands and Qu er i es... .. ......................79
5.4.8 Remote Units Commands and Qu er i e s................. .. ....80
ii
Table of Contents
5.5 Error Messages........................................................................ 82
5.5.1 Command Errors......................................................... 82
5.5.2 Query Errors................................................................ 84
5.5.3 Execution Errors.......................................................... 84
5.5.4 Device Error s .... .. ............... ............................. ............. 85
6 Service .................................................................................... 87
6.1 Model 286 Maintenance. ......................................................... 87
6.2 Model 286 Trou b l e sh ooting Hints .. ................ ............... ... ...... 87
6.3 Modifying the Line Voltage Input Range............................... 92
6.4 Additional Technical Support................................................. 93
6.5 Return Authorization.............................................................. 93
Appendix................................................................................. 95
A.1 4-20 mA Current Loop Option................................................ 95
A.2 J5A and J5B Connector Pinout .............................................. 96
A.3 RS-232 Connector (optional) .................. .. ........................... 98
A.4 RS-422 Connector (optional)................................................ ... 99
Index...................................................................................... 101
iii
Table of Contents
iv
Foreword
Purpose and Scope
This manual contains the operation and maintenance instructions for the American Magnetics, Inc. Model 286 Liquid Level Controller. The manual outlines the instructions for instrument use in typical s ystem designs. Since it is impossible to cover all possible system/sensor designs, the most common configurations are discussed and the user is encouraged to contact an authorized AMI Technical Support Represent ative for information regarding specific configurations not explicitly covered in this manual.
Contents of This Manual
Introduction introduces the reader to the functions and characteristic s of the instrument. It provides the primary illustrations of the front and rear panel layouts as well as documenting the performance specifications.
Installation describes how the instrument is unpacked and installe d in conjunction with ancillary equipment in a typical liquid delivery system.
Calibration details the available calibration methods and discusses all related display menu items.
Operation describes how the instrument is used to measure and control the liqui d level. All instrument displays and controls are documented.
Remote Interface Reference documents all remote commands and queries available through the serial interface. A quick-reference summary of commands is provided as well as a detailed description of each.
Service provides guidelines to assist the user in troubleshooting possible system and instrument malfunctions. Information for contacting AMI Technical Support personnel is also provided.
The Appendix documents the rear panel connectors.
i
Foreword
Applicable Hardware
Applicable Hardware
The Model 286 has been designed to operate with AMI Capacitance-Based Liquid Level Sensors. Operation with other sensors is not recommended and may void the warranty.
General Precautions
Cryogen Safety
Personnel handling cryogenic liquids should be thoroughly instructed and trained as to the nature of the liquids. Training is essential to minimize accidental spilling. Due to the low temperature of these materials, a cryogen spilled on many objects or surfaces may damage the surface or cause the object to shatter, often in an explosive manner.
Inert gases released into a confined or inadequately ventilated space can displace sufficient oxygen to make the local atmosphere incapable of sustaining life. Cryogenic liquefied gases are potentially extreme suffocation hazards since a small amount of liquid will vaporize an d yi eld a very large volume of oxygen-displacing gas. Always ensure the location where the cryogen is used is well ventilated. Breathing air with insufficient oxygen content may cause unconsciousness without warning. If a space is suspect, purge the space completely with air and tes t before entry. If this is not possible, wear a forced-air respirator and enter only with a co-worker standing by wearing a forced-air respirator.
Cryogenic liquids, due to their extremely low temperatures, will burn the skin in a similar manner as would hot liquids. Never permit cryogenic liquids to come into contact with the skin or allow liquid nitrogen to soak clothing. Serious burns may result from careless handling. Never touch uninsulated pipes or vessels containing cryogenic liquids. Flesh will stick to extremely cold materials. Even nonmetallic materials are dangerous to touch at low temperatures. The vapors expe lled during the venting p rocess are sufficiently cold to burn flesh or freeze optic tissues. Insulated gloves should be used to prevent frost-bite when operating valves on cryogenic tanks. Be suspicious of valves on cryogenic systems; the extremes of temperature they undergo cause seals to fail frequent ly.
In the event a person is burned by a cryogen or material cooled to cryogenic temperatures, the following first aid treatment should be given pending the arrival and treatment of a physician or other medical care worker:
1. If any cryogenic liquid contacts the skin or eyes, immediately flush the affected area gently with tepid water (102°F 105°F, 38.9°C
40.5°C) and then apply cold compresses.
ii
Foreword
Safety Summary
2. Do not apply heat. Loosen any clothing that may restrict circulation. Apply a sterile protective dressing to the affected area.
3. If the skin is blistered or there is any chance that the eyes have been affected, get the patient immediately to a physician for treatment.
Containers of cryogenic liquids are self pressuri zing (as the li quid boils off, vapor pressure i n cr e a se s) . Ho se s or li n e s us e d t o tr a ns f e r th e se liq u i d s should never be sealed at both ends (i.e. by closing valves at both ends).
When pouring cryogenic liquids from one container to another, the receiving container should be cooled gradually to prevent damage by thermal shock. The liquid should be poured slowly to avoid spattering due to rapid boil off. The receiving vessel should be vented during the transfer.
Introduction of a substance at or near room temperature into a cryogenic liquid should be done with great caution. There may be a violent gas boil off and a considerable amount of splas hing as a result of this rapid boil ing. There is also a chance that the material may crack or catastrophically fail due to forces caused by large differences i n thermal cont raction of diff erent regions of the material. Personnel engaged in this type of activity should be instructed concerning this hazard and should always wear a full face shield and protective clothing. If sever e sp raying or splashing could occur, safety glasses or chemical goggles along with body length protective aprons will provide additional protect ion.
The properties of many materials at extremely low temperatures may be quite different from the properties that these same materials exhibit at room temperatures. Exercise extreme care when handlin g materials co oled to cryogenic temperatures until the properties of these materials under these conditions are known.
Metals for use in cryogenic equipment application must posses s ufficient physical properties at these low temperatures. Since ordinary carbon steels, and to somewhat a lesser extent, alloy steels, lose much of their ductility at low temperatures, they are considered unsatisfactory and sometimes unsafe for these applications. The austinetic Ni-Cr alloys exhibit good ductility at these low temperatures and the most widely used is 18-8 stainless steel. Copper, Monel considered satisfactory materials for cryogenic service.
®
, brass and aluminum are also
Safety Summary
Cryogenic storage systems are complex systems with the potential to seriously injure personnel or equipment if not operated according to procedures. Proper use of safety mechanisms (pressure relief valves,
iii
Foreword
Safety Summary
rupture disks, etc.) included in the cryostat and top plate assembly are necessary.
Recommended Safety Equipment
First Aid kit
Fire extinguisher rated for class C fires
Leather gloves
Face shield
Signs to indicate that there are potentially dangero us cr yo ge ns in use in the area.
Safety Legend
Instruction manual symbol: the product is marked with this symbol when it is necessary for you to refer to the instruction manual in order to protect against damage to the product or personal injury.
Ch 1 & 2
Hazardous voltage symbol. Alternating Current (Refer to IEC 417, No. 5032). Off (Supply) (Refer to IEC 417, No. 5008). On (Supply) (Refer to IEC 417, No. 5007).
Warning
The Warning sign denotes a hazard. It calls attention to a procedure or practice, which if not corre ctly adhered to, could result in personal inju ry. Do not proceed beyond a Warning sign until the indicated conditi ons are fully understood and met.
Caution
The Caution sign denotes a hazard. It calls attention to an operating procedure or practice, which if not adhered to, could cause damage or destruction of a part or all of the product. Do not proce ed beyond a Caution sign until the indicated conditions are fully understood and met.
This notation in the margin indicates a feature that is only available for Channels 1 and 2 of the Model 286.
iv
1 Introduction
1.1 Model 286 Features
The American Magnetics, Inc. (AMI) Model 286 Liquid Level Controller is a microprocessor-based instrume n t designed to provide multiple sensor capacitance-based monitoring and control of liquid levels. In addition to the brief feature descriptions provided below, a detailed feature matrix is provided on page 54.
1.1.1 Capacitance-Based Level Sensing
A typical control system consists of a Model 286 Liquid Level Controller , up to four AMI liquid level sensor(s), and connecting coaxial cable(s). The primary instrument sensing element is typically a 3/8 inch (9.5 mm) OD concentric-tube cylindrical capacitor constructed of stainless steel which allows the fluid to become the dielectric between the concentric cylinders. The instrument measures the sensor c apacitanc e, which is di rectly re lated to the perce n ta g e of the sensor immerse d i n the liquid.
1.1.2 Multiple Sensor Monitoring and Control
The Model 286 is capable of monitoring four independent, capacitance level sensors and controlling liquid level in up to two vessels by directly controlling the actuation of two solenoid-operat ed flow control valves.
1.1.2.1 Factory calibrated
The Model 286 is typically calibrated by AMI for matched sensors in liquid nitrogen, and is ready for immediate use upon delivery. If the working fluid is something other than nitrogen, an approximate calibration can be performed or the customer must perform an open or closed dewar calibration after installation when the sensor can be immersed in the target fluid.
1.1.2.2 Multiple calibrations per sensor
Each of the four sensor inputs to the Model 286 can ha ve up to four independent calibrations, which are user-selectable. The user simply selects the desired calibration using the menu-type interface. All calibration data is passcode protected and stored in nonvolatile memory.
1.1.3 Convenient Display and Direct Keypad Entry
The instrument is equipped with a 16-character x 2-line backlit charact er­mode LCD display which provides liquid level and setpoint indication in inches, centimeters, or percent as sele cted by the user. The default dis play
1
Introduction
Model 286 Features
indicates liquid level and fill/loss rate. The current liquid level is updated and visible in many of the display modes used in configuring various settings of the instrument.
The 4 x 4 keypad is provided for d irec t entry of values f or setpo ints, sen sor length, and other functions. The keypad also provides a convenient me nu­type interf a ce fo r co n figuring vari o u s i nstrument se tt i n g s th a t require selection from a list of options.
1.1.4 HI/LO Alarms
Two user-defined setpoints, HI and LO, are provided for all four channels with front panel LED indications for Channels 1 and 2. The HI and LO setpoints may also operate two pairs of low-current, rear panel relay contacts and a built-in audible alarm. The HI/LO relay contact pairs may be assigned to Channels 1 and 2, or to Channels 3 and 4.
1.1.5 Controller Modes
The Model 286 provides three controller modes: normal, auto-changeover, and pre-cool. The preferred mo de can be selected via the front panel keypad or remote interface. The le vel co ntrol s equence varie s based on the selected mode. See the discussion beginning on page 7 for more information regarding the function of the controlle r modes .
1.1.6 Remote Operation
The Model 286 is capable of providing up to two analog 010 VDC signals, which corresponds to 0100% liquid level, accessible from the rear panel of the instrument for use with a voltage recorder. This is a fac tory-installed option and should be specified at the time the instrument is purchased.
Up to two self-powered (i.e. does n ot require an external power supply) 4­20 mA current loop outputs can also be ordered as an option.
Additional digital interface options, including RS-232 or RS-422, are available as options. The serial remote int erface provid es a comprehensive command set for remote monitoring and configuration of all functions of the instr u m e nt.
2
Introduction
Front Panel Layout
1.2 Front Panel Layout
POWER
I
O
OFF
3
1
$Ch 1: 100.0 % Ch 2: 50.0 %
AMI
Model 286
Liquid Level
Controller
4
1: HI/LO 1: FILL
2: HI/LO 2: FILL
5
2
HI LO
12
B RATE FILL
456
ASN MODE MISC
789
CAL
0
.
A
3
UNITS ENTER
ESC
MENU
OPTION
1 16 Character x 2 Line LCD Display
with LED Backlighting
4 HI/LO Alarm and Fill LED
Indicators for Channel 1
2 4 x 4 Keypad 5 HI/LO Alarm and Fill LED
Indicators for Channel 2
3 Powe r Switc h
3
Introduction
Rear Panel Layout
1.3 Rear Panel Layout
1 2 3
COMMUNICATIONS
RS-232
J8
J5A
CD
J5B
74
SENSOR INPUTS
5 6
1 Optional Digital Communications
Port (RS-232 option illustrated; RS-422 also available)
12
!
LINE VOLTAGE, EACH 2A MAX
BA
(OSC REQ'D)(OSC REQ'D)
4 Auxiliary DB-9 Connectors
(see
Appendix
LINE: 50-60 Hz, 4.2A MAX 100-120 V 200-240 V
for pinout)
2 Switched AC Output 1
(IEC-320 female connector)
3 Switched AC Output 2
(IEC-320 female connector)
4
5 Four BNC Coaxial Sensor Inputs
[denoted as A - D]
6 Power cord connector
(IEC-320 male)
Introduction
System Diagram
FRONT PANEL
$Ch 1: 100.0 % Ch 2: 50.0 %
POWER
O
OFF
MODEL 286 LIQUID LEVEL CONTROLLER
I
AMI
Model 286
Liquid Level
Controller
1: HI/LO 1: FILL
2: HI/LO 2: FILL
ASN MODE MISC
CAL
.
A
3
UNITS ENTER
HI LO
12
B RATE FILL
456
789
0
OSCILLATOR CABLES RG-59/U COAXI A L CABLE, 6 FT. LENGTH (FIXED)
ESC
MENU
OPTION
REAR PANEL
J5A
J5B
SOLENOID-OPERATED
FLOW VALVES
COMMUNICATIONS
CD
SENSOR INPUTS
12
!
BA
(OSC REQ'D)(OSC REQ'D)
(OPTIONAL)
LINE VOLTAGE, EACH 2A MAX
LINE: 50-60 Hz, 4.2A MAX 100-120 V 200-240 V
TOTAL SENSOR LENGTH
(SPECIFY)
ACTIVE SENSOR LENGTH
3/8 NPT NYLON FEED-THROUGH (OTHER MOUNTI NG OPTIONS AVAILABLE )
SENSOR
EXTENSION CABLES RG-59/U COAXI A L CABLE, 6 FT. LENGTH STANDARD (VARIABLE UP TO 500 FT.)
OSCILLATOR CABLES RG-59/U COAXI A L CABLE, 6 FT. LENGTH (FIXED)
A PROPERLY CONNECTED W ARM SENSOR CAN READ ~ +2% UNTIL COOL
Figure 1-1.
Model 286 instrument, fill valves, and sensor system diagram.
NOTE: UP TO FOUR LIQUID LEVEL SENSORS OF ANY LENGTH M AY BE CONNECTED
5
Introduction
Specifications
1.4 Model 286 Specifications @ 25 °C
Level Measurements
Resolution: 0.1 %, 0.1 cm, or 0.1 in
Linearity: ± 0.1 % or 1 mm (whichever is greater)
Operating Parameters
HI, A, B, and LO Setpoints: 0 % to 100 % adjustable
HI/LO Alarm Relay Contact Ratings: 30 VAC or 60 VDC, 10 VA (up to 0.5A
maximum) — 20 V at 0.5 A to 60 V at 0.167 A (normally open, closed on alarm)
Controller Outputs: AC line voltage at 2 A maximum each output
Analog Outputs
Integral Non-linearity: ± 0.012 %
Resolution: 16 bits
Total Error: ± 0.75 % for 4-20 mA output
± 1 % for 0-10 V output
Current Drift (4-20 mA): ± 75 ppm / °C maximum
Voltage Drift (0-10 V): ± 100 ppm / °C maximum
Power Requirements
Primary: 100-120 or 200-240 VAC ±10%, 50 - 60 Hz
4.2 A maximum (25 VA plus sum of controller output currents)
Physical
Dimensions (Standard): 97 mm H x 213 mm W x 290 mm D
(3.8" H x 8.4" W x 11.4" D)
Weight (Standard): 1.9 kg (4.2 lbs.)
Dimensions (Rack Mount): 89 mm H x 483 mm W x 290 mm D
(3.5" H x 19" W x 11.4" D)
Weight (Rack Mount): 2.2 kg (4.9 lbs.)
RG-59U Extension Cable Length: 500 ft. maximum
Environmental
Ambient Temperature: Operating: 0 °C to 40 °C (32 °F to 104 °F)
Nonoperating: −20 °C to 60 °C (−4 °F to 140 °F)
Relative Humidity: 0% to 95%; non-condensing
6
Introduction
Controller Modes Description: Normal Mode
1.5 Controller Modes Description
The Model 286 provides a unique feat ure in th e availability o f three modes for level control. The function of each mode is summarized below and a diagram is provided to help illustrate the function. The controller modes provide flexibility for solving a wide range of level control problems with a minimum of external hardware or logic.
1.5.1 Normal Mode
In the normal mode, as shown in Figure 1-2, Channels 1 and 2 of the Model 286 act as independent auto-fill systems. As each level falls below the “B” setpoint, an independent fill cycle is initiated and fills the controlled dewar to the “A” setpoint via two separate solid-state-relay­controlled AC outputs which can drive solenoi d-actuated valves. The A and B setpoints for Channels 1 and 2 operate as independent liquid level control bands.
Model 286
COMMUNICATIONS
12
!
LINE VOLTAGE, EACH 2A MAX
BA
J5A
CD
(OSC REQ'D)(OSC REQ'D)
LINE: 50-60 Hz, 4.2A MAX
100-120 V 200-240 V
SENSOR INPUTS
J5B
Controlled
Dewar #1
Auto-Fill System #1
Figure 1-2.
Fill Valve #1 Fill Valve #2
Storage Vessel #1
Storage Vessel #2
Auto-Fill System #2
Normal mode diagram illustrating two
independent
auto-fill systems.
Controlled
Dewar #2
7
Introduction
Controller Modes Description: Auto-Changeover Mode
1.5.2 Auto-Changeover Mode
In auto-changeover mode, as illustrated in Figure 1-3, the Model 286 monitors and controls liquid l evel measured via Channel 1, and us es liquid supplied from two storage vessels. The A and B setpoints for Channel 1 function as the liquid level cont rol band for the cont rolled d ewar. The d ual AC outputs control a fill valve for each of the two storage vessels. Channel 2 is unavailable in the auto-changeover mode.
The instrument either monitors dry contacts to determine availability of liquid from each of the two storage vessels or determines availability by fill timeouts. The Model 286 automa tically switches from one storage vessel t o the next. This allows one of the two storage vessels to be replaced when empty without interrupting the availability of liquid to the contro lled dewar.
Model 286
COMMUNICATIONS
12
!
LINE VOLTAGE, EACH 2A MAX
J5A
J5B
BA
CD
(OSC REQ'D)(OSC REQ'D)
SENSOR INPUTS
LINE: 50-60 Hz, 4.2A MAX
100-120 V 200-240 V
Controlled
Dewar
Figure 1-3.
Fill Valve #2
Fill Valve #1
Storage Vessel #1
Storage Vessel #2
Auto-changeover mode diagram illustrating the two
storage vessels and one controlled dewar.
8
Introduction
Controller Modes Description: Pre-Cool Mode
1.5.3 Pre-Cool Mode
The pre-cool mode, as illustrated in Figure 1-4, provides fo r cooling of a cryogen transfer line before opening the transfer line to the controlled dewar. The A and B setpoints for Channel 1 function as the liquid level control band for the controlled dewar. AC Output 1 controls a fill valve for the controlled dewar, while AC Output 2 controls a vent valve. Channel 2 is unavailable in the pre-cool mode.
When a fill cycle is initiated, the vent valve is initially opened for a user­programmed time, after which the vent valve is closed and the fill valve to the controlled dewar is opened. During the time the vent valve is open, the cryogen cools the transfer line, so that there is a minimal amount of cryogen gas entering the controlled dewar once the fill valve is opened.
Model 286
COMMUNICATIONS
12
!
LINE VOLTAGE, EACH 2A MAX
J5A
J5B
BA
CD
(OSC REQ'D)(OSC REQ'D)
SENSOR INPUTS
LINE: 50-60 Hz, 4.2A MAX
100-120 V 200-240 V
Figure 1-4.
Fill Valve
Controlled
Dewar
Vent Valve
Pre-Cooled Transfer Line
Storage Vessel
Vented Gas
Pre-cool mode diagram illustrating one storage vessel,
a vent valve, and one controlled dewar.
9
Introduction
Controller Modes Description: Pre-Cool Mode
10
2Installation
Warning
Before energizing the instrument, the earth ground of the power receptacle must be verified to be at earth potential and able to carry the rated current of the power ci rcuit. Using extension cords should be avoided; however, if one must be used, ensure the ground conductor is intact and capable of carrying the rated curr ent.
In the event that the ground path of the instrument becomes less than sufficient to carry the rated current of the power circuit, the instrument should be disconnected from power, la beled as unsafe, and removed from place of operation.
Do not operate this instrument in the presence of flammable gases. Doing so could result in a life-threatening explosion.
Do not modify this instrument in any way. If component replacement is required, return the instrument to AMI facilities as described in the Service section of this manual.
If the instrument is used in a manner not specified by AMI, the protection provided by the equipment may be impaired.
2.1 Unpacking the Instrument
Carefully remove the instrument, sensor(s) , os cillator(s) and interconnecting coaxial cables from the shipping carton and remove all packaging material. A rack mounting kit is supplied if the instrument was purchased with the rack mount option.
Note
If there is any shipping damage, save all packing material and contact the shipping representative to file a damage claim. Do not return the instrument to AMI unle ss prior authorization has been received.
If the chassis is a table top model, place the instrument on a flat, secure surface.
11
Installation
Installing the sensor
2.2 Rack Mounting the Instrument
If the instrument has a rack mount chassis, follow the following procedure :
a. Attach the rack mount a dapter piece s to the in strument by fi rst
removing the four screws on the side of the instrument that attach the cover to the chassis. Atta ch the rack mount adapter pieces to the sides of the instrument by reinstalling the screws.
b. Install the instrument in a 19" rack by securing the front panel
to the rail in each of the four corners with mounting hardware supplied by the cabinet manufacturer.
Warning
Do not remove the rubber feet from the bottom of the intsrument and then reinsert the origin al screws. Doing so could present a se vere life-threatening electrical hazard. Screws longer than 1/4" will contact and damage the printed circuit board inside the unit.
2.3 Installing the Sensors in the Vessels
Exercise care when installing the sensor since dents, crimps, bends or other physical distort ions in the cylindric al capacito r will ch ange elect rical characteristics, possibly caus ing calibration errors and/or disruption of proper instrument operation. Before installing the sensor, the user may want to review the Calibration and Operation sections to determine what, if any, calibration procedures may be necessary.
Note
The coaxial interconnecting cables and the oscillator are temperature sensitive and should be mounted in such a manner as to avoid large t emper ature c hanges such a s th ose encount ere d in t he path of dewar vents.
12
Installation
Sensor cabling
2.4 Connecting the Oscillator Cables to the AMI Sensors
For Model 286 inputs A and B, connect each sensor to the Model 286 using a supplied 6 foot RG-59/U coaxial cable.
For Model 286 inputs C and D, an external AMI oscillator must be used (note that inputs A and B do not require an external oscillator). For each sensor to be connected to either inpu t C or input D of the i nstrument, connect the sensor to the oscillator using a s upplied 6 foot RG-59/U coaxial cable. Ensure the oscillator is connected in the correct orientation (see page 5 for a system diagram).
The cable length between the in strument and sensor for inputs A and B (or the oscillator and the sensor for inputs C and D) should not exceed 6 feet unless longer lengths were discussed with an Authorized AMI Technical Representative.
Caution
Moisture or contaminants in any of the BNC coaxial connectors can short out the sensor and cause a erroneous readings. A pack of non-conductive electrical connection lubricant (ECL or “Dielectric Tune-up Grease”) has been included with the liquid level sensor packaging to reduce the possibility of this occurring. If desired, apply a small amount of ECL to any of the BNC connectors that may be exposed to mo isture. Mate the ECL­coated connectors then remove any excess ECL from the outside of the connector. Added protection can be achieved by covering the ECL-coated connections with a short section of heat-shri nk tubing.
<No inp>
indication or other
Note
MSDS sheets for the ECL are available upon request.
2.5 Connecting C and D Sensor Inputs to the Oscillator
Caution
Operation of the AMI Model 286 Liquid Level Instrument with a device other than an AMI Liquid Level Sensor may void the instrument warranty.
Connect each oscillator to either the C or D input of the Model 286 using a RG-59/U coaxial cable. The length of the extension cable can be varied to suit the specific application. AMI has confirmed pro per operation for up to 500 feet of RG-59/U coaxial cabling between the instrument and oscilla tor.
13
Installation
Valve installation
2.6 Installing the Optional Solenoid-operated Fill Valves
Install each solenoid-operated fill valve by connecting the valve power cable to the AC controller output receptacle on the rear panel of the instrument (refer to the rear panel layout illustrated on page 4). The standard AMI supplied valve has a 9/32 inch orifice and the input and output are tapped for 3/8 NPT. Set ting the FILL function to AUTO should be avoided until the channels have been assigned and the setpoints have been specified. See t he Operation section for details on assigning channels, specifying setpoints, and selecting the fill modes for the AC output receptacles.
Note
When using the pre-cool mode, AC output 1 must be connected to the fill valve and AC output 2 must be connected to the vent valve.
Caution
When using a solenoid-operated control val ve with the Model 286, ensure the valve is configured for the operating voltage and line frequency of the Model 286. Failure to do so will result in faulty operation and may also result in valve damage.
Warning
Before touching any of the controller output receptacle terminals or touching the wiring connected to these terminals, remove power to the instrument by unplugging it or turning the powe r switch to the off position.
The controller output re ceptacle supplies hazardo us AC line voltage potentials. It is for use with equipment which has no live par ts which are accessible. Conductors conn ected to its terminals must be insulated from user contact by reinforced or double insulation capable of withstanding 4250 V (impulse ) for a 240 VAC Category II installation, or 2550 V (impulse) for a 120 VAC Category II installation.
This instrument is desi gned for opera tion from a single- phase power source for maximum safety. Th e controller output receptacle circuitry only switches the “line” (“hot”) connection to the AC mains. If two-phase power is applied, any equipment connected to the controller output receptacle conducts hazardous AC voltage even when the controller output receptacle is not energized.
14
Installation
Verifying power requirements
2.7 Connecting to Optional Analog Outputs
See paragraph A.1 on page 95 for connection details for the 4-20 mA analog output option. See paragraph A.2 on page 96 for connection details for the 0-10 VDC analog output option.
2.8 Connecting to Communication Options
See paragraph 5.3.1 on page 70 instructions concerning connection to the serial communications options (RS-232 and RS-422). Paragraph A.3 on page 98 shows RS-232 cable wiring. Paragraph A.4 on page 99 shows RS­422 cable wiring.
2.9 Connecting the Instrument to Power
Warning
The Model 286 operates on 50-60 Hz power and may be configured for 110-120 or 208-240 VAC ±10% (100 or 200 VAC ±10% for Japan and South Korea). The power requirement for each ins tr ume nt is marked on the ca l ib r ati o n st ic k e r on th e bo tto m o f th e in s t r u me n t . Be sure your instrument is co nfigu red f or your powe r so urce pr ior to plugging in the line cord. Do not fail to connect the input ground terminal securely to an external eart h ground.
Ensure the front panel switch is in the OFF position. Verify that the instrument is configured for the proper operat ing voltage by referring to the calibration sticker affixed to the bottom of the instrument. If the operating voltage is correct, plug the line cord into the appropriate power receptacle.
Warning
Do not install the instrument in a manner that prevents removal of the line cord from the rear panel of the instrument.
15
Installation
Verifying power requirements
16
Loading...
+ 86 hidden pages