Voltage dips 60% - 100ms
Short interruption >95% - 5s
I, the undersigned, hereby declare that the equipment specified above complies with the requirements of the aforementioned Directives and Standards and carries the "CE" mark accordingly.
Gregory J. LaughonSeptember 12, 2002
Quality Assurance Manager
This manual contains the operation and maintenance instructions for the
American Magnetics, Inc. Model 185/186 Liquid Level Instrument. The
manual outlines the instructions for instrument use in various system
designs. Since it is impossible to cover all possible system/sensor designs,
the most common configuration is discussed and the user is encouraged to
contact an authorized AMI Technical Support Representative for
information regarding specific configurations not explicitly covered in this
manual.
Contents of This Manual
Introduction introduces the reader to the functions and characteristics of
the instrument. It provides the primary illustrations of the front and rear
panel layouts as well as documenting the performance specifications.
Installation describes how the instrument is unpacked and installed in
conjunction with ancillary equipment in a typical cryogenic system.
Operation describes how the instrument is used to measure and control
liquid level. All instrument controls are documented.
Remote Interface Reference documents all remote commands and
queries available through the serial and IEEE-488 interfaces. A quickreference summary of commands is provided as well as a detailed
description of each.
Service provides guidelines to assist Qualified Service Personnel in
troubleshooting possible system and instrument malfunctions.
Information for contacting AMI Technical Support personnel is also
provided.
The Appendix documents the rear panel connectors.
Rev. 3vii
Foreword
Applicable Hardware
Applicable Hardware
The Model 185/186 has been designed to operate with an AMI Liquid Level
Sensor. Operation with other equipment is not recommended and may
void the warranty.
General Precautions
Cryogen Safety
Personnel handling cryogenic liquids should be thoroughly instructed and
trained as to the nature of the liquids. Training is essential to minimize
accidental spilling. Due to the coldness of these materials, a cryogen
spilled on many objects or surfaces may damage the surface or cause the
object to shatter, often in an explosive manner.
Inert gases released into a confined or inadequately ventilated space can
displace sufficient oxygen to make the local atmosphere incapable of
sustaining life. Cryogenic liquefied gases are potentially extreme
suffocation hazards since a small amount of liquid will vaporize and yield
a very large volume of oxygen-displacing gas. Always ensure the location
where the cryogen is used is well ventilated. Breathing air with
insufficient oxygen content may cause unconsciousness without warning.
If a space is suspect, purge the space completely with air and test before
entry. If this is not possible, wear a forced-air respirator and enter only
with a co-worker standing by wearing a forced-air respirator.
Cryogenic liquids, due to their extremely low temperatures, will burn the
skin in a similar manner as would hot liquids. Never permit cryogenic
liquids to come into contact with the skin or allow liquid nitrogen to soak
clothing. Serious burns may result from careless handling. Never touch
uninsulated pipes or vessels containing cryogenic liquids. Flesh will stick
to extremely cold materials. Even nonmetallic materials are dangerous to
touch at low temperatures. The vapors expelled during the venting process
are sufficiently cold to burn flesh or freeze optic tissues. Insulated gloves
should be used to prevent frost-bite when operating valves on cryogenic
tanks. Be suspicious of valves on cryogenic systems; the extremes of
temperature they undergo causes seals to fail frequently.
In the event a person is burned by a cryogen or material cooled to
cryogenic temperatures, the following first aid treatment should be given
pending the arrival and treatment of a physician or other medical care
worker:
1. If any cryogenic liquid contacts the skin or eyes, immediately flush
the affected area gently with tepid water (102°F
40.5°C) and then apply cold compresses.
− 105°F, 38.9°C −
viiiRev. 3
Foreword
Safety Summary
2. Do not apply heat. Loosen any clothing that may restrict
circulation. Apply a sterile protective dressing to the affected area.
3. If the skin is blistered or there is any chance that the eyes have
been affected, get the patient immediately to a physician for
treatment.
Containers of cryogenic liquids are self pressurizing (as the liquid boils off,
vapor pressure increases). Hoses or lines used to transfer these liquids
should never be sealed at both ends (i.e. by closing valves at both ends).
When pouring cryogenic liquids from one container to another, the
receiving container should be cooled gradually to prevent damage by
thermal shock. The liquid should be poured slowly to avoid spattering due
to rapid boil off. The receiving vessel should be vented during the transfer.
Introduction of a substance at or near room temperature into a cryogenic
liquid should be done with great caution. There may be a violent gas boil
off and a considerable amount of splashing as a result of this rapid boiling.
There is also a chance that the material may crack or catastrophically fail
due to forces caused by large differences in thermal contraction of different
regions of the material. Personnel engaged in this type of activity should
be instructed concerning this hazard and should always wear a full face
shield and protective clothing. If severe spraying or splashing could occur,
safety glasses or chemical goggles along with body length protective
aprons will provide additional protection.
The properties of many materials at extremely low temperatures may be
quite different from the properties that these same materials exhibit at
room temperatures. Exercise extreme care when handling materials cooled
to cryogenic temperatures until the properties of these materials under
these conditions are known.
Metals to be used for use in cryogenic equipment application must posses
sufficient physical properties at these low temperatures. Since ordinary
carbon steels, and to somewhat a lesser extent, alloy steels, lose much of
their ductility at low temperatures, they are considered unsatisfactory and
sometimes unsafe for these applications. The austinetic Ni-Cr alloys
exhibit good ductility at these low temperatures and the most widely used
®
is 18-8 stainless steel. Copper, Monel
, brass and aluminum are also
considered satisfactory materials for cryogenic service.
Safety Summary
Cryogenic storage systems are complex systems with the potential to
seriously injure personnel or equipment if not operated according to
procedures. Proper use of safety mechanisms (pressure relief valves,
Rev. 3ix
Foreword
Safety/Manual Legend
rupture disks, etc.) included in the cryostat and top plate assembly are
necessary.
Recommended Safety Equipment
First Aid kit
Fire extinguisher rated for class C fires
Leather gloves
Face shield
Signs to indicate that there are potentially dangerous cryogens in
use in the area.
Safety/Manual Legend
Instruction manual symbol: the product is marked with this
symbol when it is necessary for you to refer to the instruction
manual in order to protect against damage to the product or
personal injury.
Hazardous voltage symbol.
Alternating Current (Refer to IEC 417, No. 5032).
Off (Supply) (Refer to IEC 417, No. 5008).
On (Supply) (Refer to IEC 417, No. 5007).
Warning
The Warning sign denotes a hazard. It calls attention to a procedure or
practice, which if not correctly adhered to, could result in personal injury.
Do not proceed beyond a Warning sign until the indicated conditions are
fully understood and met.
Caution
The Caution sign denotes a hazard. It calls attention to an operating
procedure or practice, which if not adhered to, could cause damage or
destruction of a part or all of the product. Do not proceed beyond a Caution
sign until the indicated conditions are fully understood and met.
0RGHO
This marking in the left margin of the manual designates a feature,
procedure, or specification that is unique to the Model 186.
xRev. 3
1Introduction
1.1 Model 185/186 Feature s
The American Magnetics, Inc. (AMI) Model 186 Liquid Level Controller
system is an advanced, microprocessor-based solution designed to provide
accurate and reliable level monitoring and control of virtually any
cryogenic liquid.
1.1.1Capacitance-based level sensing
The system consists of a Model 185/186 Liquid Level Instrument, sensor,
connecting cables, and an optional solenoid-operated fill valve. The
instrument sensing element is typically a 3/8 inch (9.5 mm) OD cylindrical
capacitor constructed of stainless steel which allows a cryogenic fluid to
become the dielectric between the concentric plates. The instrument
measures the sensor capacitance which is directly related to the
percentage of the sensor immersed in the cryogenic liquid. The sensors are
normally constructed in overall lengths of up to 20 feet (6.1 m). The
maximum active length is typically 7 inches less than the overall sensor
length.
0RGHO
1.1.2HI/LO level alarms
The Model 185/186 provides two alarm setpoints for both HI and LO level
indication. The HI and LO level alarms activate front panel LEDs, an
audible alarm, and two independent sets of relay contacts accessible from
the rear panel. All setpoints are continuously adjustable from the front
panel.
1.1.3Level control
The Model 186 adds two additional setpoints which are used to specify a
control band for the liquid level. The Model 186 automatically energizes
and de-energizes a rear panel controller output receptacle which is
typically used to operate a solenoid valve. The controller output receptacle
state can be manually overridden from the front panel. A fill timeout
feature is also provided which can be used to terminate the fill function
after a user-specified period of time.
1.1.4Convenient display
The instrument is equipped with a 4-digit LED display which provides
liquid level indication in inches, centimeters, or percent as selected by a
front panel switch. A front panel switch allows the user to adjust the
instrument length quickly and easily for a specific active sensor length.
The sensor active length can be entered in either inches or centimeters.
Rev. 31
Introduction
The length adjustment only affects the scaling of the level display and does
not change the actual calibration of the instrument.
1.1.5Microprocessor-based electronics
Microprocessor-based electronics provide 0.1% readout accuracy.
Nonvolatile memory maintains instrument calibration without battery
backup. Watchdog timer circuitry and low line voltage (brownout) detector
prevent microprocessor lockup and provide fail-safe operation.
1.1.6Remote computer monitoring or controlled operation
The Model 185/186 can be optionally equipped with a 0-10 volt recorder
output. A 4-20 mA current loop option is available in lieu of the recorder
output. Available computer interface options include RS-232/422 Serial
Port/Data Logger or IEEE-488.
The Model 185/186 may be optionally configured for a maximum of one
analog output option and one computer interface option.
A PROPERLY CONNECTED WARM
SENSOR CAN READ ~ +2% UNTIL COOL
$0(5,&$10$*1(7,&6,1&2$.5,'*(7186$
/,1(+]$0$;
9999
+,/(9(/$%
0RGHO
/2/(9(/
/LTXLG/HYHO
&RQWUROOHU
5$,6(
6(732,176
}
/2:(5
$0,
,1&+
&0
Model 186 instrument, control valve, and sensor system diagram.
Rev. 37
Introduction
Specifications
1.6 Model 185/186 Specifications @ 25 °C
0RGHO
0RGHO
0RGHO
Level Measurements
Maximum Length Readout:650.0 cm (255.9 in)
a
Resolution:0.1%, 0.1 cm, or 0.1 in
Linearity: ± 0.1%
Operating Parameters
HI and LO Alarms:0% to 100% adjustable
HI/LO Alarm Relay Contact Ratings:10 VA, 30 VAC or 60 VDC, 0.5 A
(normally open, closed on alarm)
A and B Control Setpoints:
Controller Output:
Fill Timer:
0% to 100% adjustable
AC line voltage @ 2A max current
0.1 to 600.0 minutes
0-10 Volt Analog Output
Integral Non-linearity:± 0.012%
Resolution:16 bits
Total Error:± 1.1% for 0-10 V output
Voltage Drift (0-10 V):100 ppm / °C
Maximum Load:4 mA (2.5 kΩ @ 10 V output)
4-20 mA Analog Output @ 24 V
V
Supply Range:13-32 VDC (see Appendix for diagram)
ext
Voltage Compliance:
Integral Non-linearity:± 0.012%
Resolution:16 bits
Total Error:± 0.25% for 4-20 mA output
Current Drift (4-20 mA):75 ppm / °C
PSRR:10 µA / V
Power Requirements
Primaryb:
Maximum Current:30 VA for Model 185
V
− 3.5
ext
110-120 or 208-240 VAC ±10%
50 - 60 Hz
For Japan or S. Korea: 100 or 200 VAC ±10%
2.2 A for Model 186
8Rev. 3
Introduction
Specifications
Physical
Dimensions (Standard):97 mm H x 213 mm W x 282 mm D
(3.8" H x 8.4" W x 11.1" D)
Weight (Standard):1.6 kg (3.6 lbs.)
Dimensions (Rack Mount):89 mm H x 483 mm W x 282 mm D
(3.5" H x 19" W x 11.1" D)
Weight (Rack Mount):2.0 kg (4.3 lbs.)
Environmental
Ambient Temperature:Operating: 0 °C to 50 °C (32 °F to 122 °F)
Nonoperating: −20 °C to 60 °C (−4 °F to 140 °F)
Relative Humidity:0 to 95%; non-condensing
a. Under extreme radiated electromagnetic field conditions (3V/m at 450 MHz to 610
MHz), the accuracy may be degraded by an additional ±0.7%.
b. Units configured for Japan or South Korea cannot be configured for operation at
other voltages without an internal transformer change, and vice-versa.
Rev. 39
Introduction
Specifications
10Rev. 3
2Installation
Warning
Before energizing the instrument, the earth ground of the power
receptacle must be verified to be at earth potential and able to carry
the rated current of the power circuit. Using extension cords should
be avoided; however, if one must be used, ensure the ground
conductor is intact and capable of carryi ng the rated current.
In the event that the ground path of the instrument becomes less
than sufficient to carry the rated current of the power circuit, the
instrument should be disconnected from power, la beled as unsafe,
and removed from place of operation.
Do not operate this instrument in the presence of flammable gases.
Doing so could result in a life-threatening explosion.
Do not modify this instrument in any way. If component
replacement is required, return the instrument to AMI faciliti es as
described in the Service section of this manual.
If the instrument is used in a manner not specified by AMI, the
protection provided by the equipment may be impaired.
2.1 Unpacking the Instrument
Carefully remove the instrument, sensor, oscillator and interconnecting
coaxial cables from the shipping carton and remove all packaging
material. A rack mounting kit is supplied if the instrument was purchased
with the rack mount option.
Note
If there is any shipping damage, save all packing material and
contact the shipping representative to file a damage claim. Do not
return the instrument to AMI unless prior authorization has been
received.
If the chassis is a table top model, place the instrument on a flat, secure
surface.
Rev. 311
Installation
Installing the sensor
2.2 Rack Mounting the Instrument
If the instrument has a rack mount chassis, follow the following procedure:
a. Attach the rack mount adapter pieces to the instrument by first
removing the four screws on the side of the instrument that
attach the cover to the chassis. Attach the rack mount adapter
pieces to the sides of the instrument by reinstalling the screws.
b. Install the instrument in a 19" rack by securing the front panel
to the rail in each of the four corners with mounting hardware
supplied by the cabinet manufacturer.
Warning
Do not remove the cabinet feet and th en rein ser t t he origin al s cre ws.
Doing so could present a severe life-threatening electrical hazard. If
removal of the cabinet feet is desired, replace the original screws
with screws not to exceed 1/4" in length. Screws longer than 1/4"
will contact and damage the printed circuit board inside the unit.
2.3 Installing the Sensor in the Cryo-vessel
Exercise care when installing the sensor since dents, crimps, bends or
other physical distortions in the thin wall capacitor will change electrical
characteristics possibly causing calibration errors and/or disruption of
proper instrument operation. Before installing the sensor, the user may
want to review the Calibration and Operation sections to determine what,
if any, calibration procedures may be necessary.
Note
The coaxial interconnecting cables and the oscillator are
temperature sensitive and should be mounted in such a manner as
to avoid large t emper ature c hanges su ch as th ose enc ountere d i n the
path of dewar vents.
12Rev. 3
Installation
Interconnects with oscillator and valve
2.4 Connecting the Oscillator Cable to the AMI Sensor
Connect the oscillator to the sensor using a supplied 6 foot RG-59/U
coaxial cable. Ensure the oscillator is connected in the correct orientation
(see page 7 for a system diagram). The cable length between the oscillator
and the sensor should not exceed 6 feet unless longer lengths were
discussed with an Authorized AMI Technical Representative.
Caution
Moisture or contaminants in any of the BNC coaxial connectors can
short out the sensor and cause a false ‘full’ level indication or other
erroneous readings. A pack of non-conductive electrical connection
lubricant (ECL or “Dielectric Tune-up Grease”) has been included
with the liquid level senso r packaging to reduce the possibility of
this occurring. If desired, apply a small amount of ECL to any of the
BNC connectors that may be expos ed to m oisture. Mate the doped
connectors then remove any excess ECL from the outside of the
connector. Added protection can be achi eved by covering the doped
connections with a short section of heat-shrink tubing.
Note: MSDS sheets for the ECL are available upon request.
2.5 Connecting the Instrument to the Oscillator
Caution
Operation of the AMI Model 185/186 Liquid Level Instrument with
a device other than an AMI Liquid Level Sensor may void the
instrument warranty.
Using the J5 coaxial connector, connect the instrument to the oscillator
using a RG-59/U coaxial cable. The length of the extension cable can be
varied to suit the specific application. AMI has confirmed proper operation
for up to 500 feet of coaxial cabling between the instrument and oscillator.
Rev. 313
Installation
Interconnects with oscillator and valve
2.6 Installing an Optional Solenoid-operated Fill Valve
0RGHO
Install a solenoid-operated fill valve by connecting the valve power cable to
the AC controller output receptacle on the rear panel of the instrument.
The standard AMI supplied valve has a 9/32 inch orifice and the input and
output are tapped for 3/8 NPT. Operation of the controller output
receptacle in AUTO mode should be avoided until the instrument setpoints
have been specified. See the Operation section for details on specifying the
setpoints and selecting the operational mode for the controller output
receptacle.
Caution
When using a solenoid-operated control valve with the Model 186,
ensure the valve is configured for the operating voltage of the Model
186. Failure to do so will result in faulty operati on and may also
result in valve d a m a ge.
Warning
Before touching any of the controller output receptacle terminals or
touching the wiring connected to these terminals, remove power to
the instrument by unplugging it or turning the powe r switch to the
off position.
The controller output recep tacle conducts hazardous AC line volta ge
potentials. It is for use with equipment which has no live parts
which are accessible. Conductors connected to its terminals must be
insulated from user contact by reinforced or double insulation
capable of withstanding 4250 V (impulse ) for a 240 VAC Category II
installation, or 2550 V (impulse) for a 120 VAC Category II
installation.
This instrument is desi gned for opera tion from a single- phase power
source for maximum safety. Th e controller output receptacle
circuitry only switches the “line” (“hot”) connection to the AC mains.
If two-phase power is applied, any equipment connected to the
controller output receptacle conducts hazardous AC voltage even
when the controller output receptacle is not energized.
14Rev. 3
Installation
Verifying power requirements
2.7 Connecting the Instrument to Power
Warning
The Model 185/186 operates on 50-60 Hz power and may be
configured for 110-120 or 2 08-240 VAC ±10% (100 or 200 VAC ± 10%
for Japan and South Korea). The power requirements for each
instrument is marked on the calibration sticker on the bottom of the
instrument. Be sure your instrument is configured for your power
source prior to plugging in the line cord. Do not fail to connect the
input ground terminal securely to an external earth ground.
Ensure the front panel switch is in the OFF position. Verify that the
instrument is configured for the proper operating voltage by referring to
the calibration sticker affixed to the bottom of the instrument. If the
operating voltage is correct, plug the line cord into the appropriate power
receptacle.
Warning
Do not install the instrument in a manner that prevents removal of
the line cord from the rear panel of the instrument.
Rev. 315
Loading...
+ 57 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.