American Magnetics 150A User Manual

INSTALLATION, OPERATION AND MAINTENANCE
INSTRUCTIONS FOR THE AMI
LIQUID HELIUM LEVEL METER
I. INTRODUCTION
The American Magnetics, Inc. (AMI) Model 150A is a compact, fully portable Liquid Helium Level Meter providing simple and easy operation. The instrument is battery operated and when used in conjunction with an AMI LHe Level Sensor, provides a truly portable means of measuring liquid helium levels at any location. The sensor connects to the instrument via a standard DB9 connector. The instrument displays 0-100% of sensor maximum active length immersed in LHe. The instrument can be used with AMI sensors with active lengths of up to 60 inches. The instrument is powered by sealed lead-acid batteries. A standard AC-DC adapter (either 120VAC or 230VAC) is supplied with the unit for charging the batteries. Since battery energy consumption is strongly dependent on the LHe level being measured, AMI has developed a special "battery saver" circuit which reduces the power drain from the batteries as more of the sensor becomes superconducting. Power requirements decrease as LHe level increases. A front panel POWER LED is provided to indicate sufficient battery power exists to make an accurate level reading.
MODEL 150A
II. SPECIFICATIONS (at 25 °C unless stated otherwise)
Input Power (instrument) Sealed lead-acid battery (6VDC)
Input Power (AC adapter)
Operating temperature (instrument) 0 to 50 °C non-condensing
Operating temperature (AC adapter) 0 to 40 °C non-condensing
Sensor active length range 1 inch to 60 inches
Linearity
Absolute Calibration 0.5% of full scale
Probe Current 75 mA nominal
Battery Life Typically 1,000 readings with a 60" sensor.
0.5% of full scale
90VAC to 132VAC 50/60 Hz, or 180VAC to 264VAC 50/60 Hz (depending on specified AC adapter model)
2% of full scale (from 0 to 50 ºC)
1 March 6, 2001
III. INSTALLATION
A. Carefully remove the instrument from the shipping carton and remove all the packing
material. Inspect the equipment for any physical damage that may have occurred during shipment.
NOTE:
If there is any shipping damage save all packing material and contact the shipping representative to file a damage claim. Do not return the instrument to AMI unless prior authorization has been received. (See Section IX)
B. Install the liquid helium sensor as per the AMI document "Installation, Operation And
Maintenance Instructions for the AMI Liquid Helium Level Sensor."
C. To facilitate mobility, the Model 150A is equipped with a single standard DB9 connector
which connects the instrument to a sensor. Connect the LHe sensor wires to the DB9 connector terminals as follows:
Wire Color DB9 Pin Number Signal Name
Red 1 I+
Yellow 6 V-
Black 7 I-
Blue 8 V+
IV. OPERATION
A. Connect the instrument to the sensor via the sensor (DB9) connector. Refer to Figure 1. B. Depress the READ push-button switch on the front panel to read the LHe level. The
button should be held for sufficient time to allow the meter reading to stabilize.
NOTE:
The front panel LED should light to indicate sufficient sensor current exists to make an accurate reading. The instrument is calibrated to read the helium level directly on the panel meter. If the LED does not light during the level measurement, recharge the batteries. When the READ button is released, the unit is automatically deenergized.
2 March 6, 2001
Fig. 1. Model 150A
3 March 6, 2001