EXCELLENCE IN MAGNETICS AND CRYOGENICS
MODEL 110A LIQUID HELIUM
AMI
LEVEL MONITOR
INSTALLATION, OPERATION, AND
MAINTENANCE INSTRUCTIONS
American Magnetics, Inc.
PO Box 2509, 112 Flint Road, Oak Ridge, TN 37831-2509, Tel: 865 482-1056, Fax: 865 482-5472
Rev. 5, June 1999
Declaration of Conformity
Application of Council Directives: Low Voltage Directive 72/23/EEC
EMC Directive 89/336/EEC
Manufacturer’s Name: American Magnetics, Inc.
Manufacturer’s Address: 112 Flint Road,
P.O. Box 2509
Oak Ridge, TN 37831-2509
U.S.A.
Type of Equipment: Liquid Level Instruments
Model Numbers: Model 110A
Standar ds to which Conformity is Declared:
Safety: EN 61010-1 (1993) w/A1, A2
EMC: EN55011 (1991) Group 1, Class A
EN50082-1 (1997) / EN61000-4-2 (1995) 8kV AD, 4kV CD
EN50082-1 (1997) / EN61000-4-3 (1996) 3V/m
EN50082-1 (1997) / EN61000-4-4 (1995) 1kV Power Supply
0.5kV I/O cables
EN50082-1 (1997) / EN61000-4-5 (1995) 2kV CM, 1kV DM
EN50082-1 (1997) / EN61000-4-6 (1996) 3V
EN50082-1 (1997) / EN61000-4-11 (1994) Voltage dips 30% - 10ms
Voltage dips 60% - 100ms
Short interruption >95% - 5s
I, the undersigned, hereby declare that the equipment specified above complies with the requirements of the aforementioned Directives and Standards and carries the "CE" mark accordingly.
Gregory J. Laughon September 12, 2002
Quality Assurance Manager
American Magnetics, Inc.
Oak Ridge, TN, U.S.A.
Table of Contents
................................................................
Model 110A Front Panel Layout
..............................................
Model 110A Rear Panel Layout
................................................
Model 110A Specifications @ 25
°C..........................................
.......................................................
Rack Mounting the Instrument
................................................
Preparing the Sensor for Connection
.......................................
Connecting any Communications Option
................................
Connecting the Instrument to Power
.......................................
........................................................
..............................................................
Troubleshooting Procedures
...................................................
Custom Instrument Configurations
.......................................
Additional Technical Support
.................................................
..............................................................
4-20 ma Current Loop Option
.................................................
Sensor/Analog Outputs Connector J1 Wiring
1 Introduction.............................................................................. 1
2 Installation................................................................................ 5
3 Operation.................................................................................. 9
4 Service Guide......................................................................... 11
Appendix................................................................................. 17
Index........................................................................................ 19
i
Foreword
This manual contains the operation and maintenance instructions for the
American Magnetics, Inc. Model 110A Liquid Level Monitor. The manual
outlines the instructions for instrument use in typical system designs.
Since it is impossible to cover all possible system/sensor designs, the most
common configuration is discussed and the user is encouraged to contact
an authorized AMI Technical Support Representative for infor mat i on
regarding specific configurations not explicitly covered in this manual.
introduces the reader to the functions and char acteristics of
the instrument. It provides the primary illustrations of the front and rear
panel layouts as well as documenting the performance specifications.
describes how the instrument is unpacked and installed in
conjunction with ancillary equipment in a typical cryogenic system.
describes how the instrument is used to measure liquid level.
provides guidelines to assist Qualified Service Personnel in
troubleshooting possible system and instrument malfunctions.
Information for contacting AMI Technical Support personnel is also
documents the rear panel con n e ctors.
Purpose and Scope
Contents of This Manual
iii
Foreword
The Model 110A has been designed t o opera te with an AMI Li quid Helium
Level Sensor. Operation with other equipment is not recommended and
Personnel handling cryogenic liquids should be thoroughly instructed and
trained as to the nature of the liquids. Training is essential to minimize
accidental spilling. Due to the coldness of these materials, a cryogen
spilled on many objects or surfaces may damage the surface or cause the
object to shatter, often in an explosive manner.
Inert gases released into a confined or inadequately ventilated space can
displace sufficient oxygen to make the local atmosphere incapable of
sustaining life. Cryogenic liquefied gases are potentially extreme
suffocation hazards since a small amount of liquid will vaporize and yield
a very large volume of oxygen-displacing gas. Always ensure the location
where the cryogen is used is well ventilated. Breathing air with
insufficient oxygen content may cause unconsciousness without warning.
If a space is suspect, purge the space completely with air and test before
entry. If this is not possible, wear a forced-air respirator and enter only
with a co-worker standing by wearing a forced-air respirator.
Cryogenic liquids, due to their extremely low temperatures, will burn the
skin in a similar manner as would hot liquids. Never permit cryogenic
liquids to come into contact with the skin or allow liquid nitrogen to soak
clothing. Serious burns may result from careless handling. Never touch
uninsulated pipes or vessels containing cryogenic liquids. Flesh will stick
to extremely cold materials. Even nonmetallic materials are dangerous to
touch at low temperatures. The vapors expelled during the venting process
are sufficiently cold to burn flesh or freeze optic tissues. Insulated gloves
should be used to prevent frost-bite when operating valves on cryogenic
tanks. Be suspicious of valves on cryogenic systems; the extremes of
temperature they undergo causes seals to fail frequently.
In the event a person is burned by a cryogen or material cooled to
cryogenic temperatures, the following first aid treatment should be given
pending the arrival and treatment of a physician or other medical care
If any cryogenic liquid contacts the skin or eyes, immediately flush
the affected area gently with tepid water (102°F
40.5°C) and then apply cold compresses.
Applicable Hardware
Applicable Hardware
General Precautions
Cryogen Safety
iv
−
−
Foreword
Do not apply heat. Loosen any clothing that may restrict
circulation. Apply a sterile protective dressing to the affected area.
If the skin is blistered or there is any chance that the eyes have
been affected, get the patient immediately to a physici an for
Containers of cryogenic liquids are self pressurizing (a s the liquid boils off,
vapor pressure increases). Hoses or lines used to transfer these liquids
should never be sealed at both ends (i.e. by closing valves at both ends).
When pouring cryogenic liquids from one container to another, the
receiving container should be cooled gradually to prevent damage by
thermal shock. The liquid should be poured slowly to avoid spattering due
to rapid boil off. The receiving vessel should be vent ed during the transfer.
Introduction of a substance at or near room temperature into a cryogenic
liquid should be done with great caution. There may be a violent gas boil
off and a considerable amount of splashing as a r esult of this rapid boi ling.
There is also a chance that the material may crack or catastrophically fail
due to forces caused by large diffe rences in ther mal contrac tion of differe nt
regions of the material. Personnel engaged in this type of activity should
be instructed concerning this hazard and should always wear a full face
shield and protective clothing. If severe spraying or splashing could occur,
safety glasses or chemical goggles along with body length protective
aprons will provide additional protection.
The properties of many materials at extremely low temperatures may be
quite different from the properties that these same materials exhibit at
room temperatures. Exercise extreme care when handling materials cooled
to cryogenic temperatures until the properties of these materials under
these conditions are known.
Metals to be used for use in cryogenic equipment application must posses
sufficient physical properties at these low temperatures. Since ordinary
carbon steels, and to somewhat a lesser extent, alloy steels, lose much of
their ductility at low temperature s, they are considered unsatisfactory and
sometimes unsafe for these applications. The austinetic Ni-Cr alloys
exhibit good ductility at these low temperatures and the most widely used
is 18-8 stainless steel. Copper, Monel
, brass and aluminum are also
considered satisfactory materials for cryogenic service .
Cryogenic storage systems are complex systems with the potential to
seriously injure personnel or equipment if not operated according to
procedures. Proper use of safety mechanisms (pressure relief valves,
Safety Summary
Safety Summary
v