Although for Vision procedures the Cautions service
information
available
If
you information, representative
Alcon
15800
Irvine,
(949) (800)
If
you
contact
this
manual
maintaining
System,
or
functional
operator's
in
the
manual.
manual.
provided
on-site
have
any
please
or
Laboratories Alton
Parkway
CA
92718
753-1393
832-7827
are
located
your
local
provides
optimum
it
does
performance
not
descriptions
In
addition,
operator's
The
with
questions
manual
operator's
in
this
the
system.
or
contact
the
Technical
outside
the
authorized
the
contain
manual
manual,
require
your
local
Services
United Alcon
necessary
of the
ail
of
the
operating
contained
the
Warnings
also
apply
supplements
and
should
additional
service
Department
States,
please
distributor.
information
Infiniti™*
in
and
for
this
be
at:
The
system
prior
to
shipment. damaged, the
for
parts). any Alcon and
necessary,
separately,
ask
system
external
of
is
If
damage
the
functional
representative.
packing
file a claim
with
was
inspected
If
that the
unpacked.
damage
is
material
the
insurance
mechanically
the
shipping
carrier’s
(i.e.
agent
The
system
scratches,
discovered
tests
notify
Retain
for the
with
the
carrier’s
the
carrier
company.
container
be
should
dents,
or
if
the
the
carrier
shipping
inspection.
or,
and
electrically
appears
present
system
be
or
and
broken
when
inspected
fails
an
container
if
insured
As
Federal order
Law
of a physician.
CAUTION
restricts
this
device
to
sale
by
or
on
the
ONFINITI
UNPACKING
1.
2.
3.
4.
5.
6.
With
the
carton
pallet,
lid
Remove
two
(see
Ensure tightly up
Remove
roll
protective
‘Inspect
Open cable footswitch.
(see
top
Figure
into
the
1-4).
in
this
cut
and
Figure
accessories
foam
that the
bound,
an
upright
the
system
Remove
cover.
console
footswitch
(in
bubble
section
AND
remove
1-2).
pieces
1-3).
velcro
then
velcro
away
all
Plug
of
SETTING
lying
horizontally
the
Lift the
from
and
straps
carefully
position.
straps
from
remaining
for
signs
drawer
wrap).
in
the
the
manual.
Connect
footswitch
UP
binding
lid
up
the
carton,
the
outer
securing tip
to
release
the
container
foam
of
shipping
and
remove
THE
SYSTEM
on
its
wooden
straps
securing
and
off the
then
remove
cardboard
the
console
the
shipping
the
console,
(see
and
the
plastic
damage.
footswitch
footswitch
as
instructed
carton.
sleeve
carton
then
Figure
cable
later
the
are
to
x
gi
>
<
—
~~
per
E,
Top
Outer
-
Accessories
Foam
Pieces
Cardboard
Sleeve
Č
`
7.
i
8.
9.
10.
Open
side
drawer
and
small
setscrew
bubble
away extension
Remove
supplied
Unwrap it
Perform Procedure
wrap).
from
onto
the
batteries.
the
into a power
the
(STP).
and
in
With
the
system
when
the
TV
remote
power
Infiniti™*
from
Place
cord
outlet.
remove
zip
lock
hanger
extended,
pole
with
its
remote
from
Vision
IV
pole
bag
(both
in
position
the
carton
in
the
rear
System
Binding
extension
wrapped
to
secure
a
the
setscrew.
and
insert
side
drawer.
panel
Service
Strap
point
IV
the
and
plug
Test
(2)
in
pole
Figure
1-3.
Accessories
and
foam
Velcro
Tightened
removed
Strap
Securely
from
carton.
Figure
8065750238.
1-2
Carton
lying
horizontally
—™—
on
Pallet
wooden
pallet.
Figure
1-4
Roll
the
system
away
from
the
container.
13
ONFINI
GENERAL
The
Injiniti™*
anterior
cataract
well coagulation. make tremendous designed of
Following
*
segment
lens
as
associated
it
simple
versatility
to
allow
every
patient.
are
Customized
-
OZil™
torsional amplitude with
traditional
-
AquaLase®
It
which
technology,
-
Infiniti™*
features
-
High
of a phaco
performance
piezoelectric,
+
Advanced peristaltic
+
Fully
+
Modularized
fluidics
aspiration.
programmable,
disposable
*
Emulation
»
Ability
of
to
drive a high
vitrectomy
¢
Bipolar
¢
Several
control
coagulation traditional
including
application
management.
*
Automated footswitch,
*
Linear U/S
control).
¢
Linear
in
¢
Linear
lens
*
On-demand
¢
Programmable,
*
Ability desired
*
Ability screen,
*
Emission Operational
footswitch
steps
footswitch
I/A,
VIT,
footswitch
removal
to
levels
to
remote,
(sophisticated
set
switch
of
INFORMATION
Vision
extraction,
was
to
key
cataract
and
NeoSoniX®
fluidic
Fluidic
venturi-like
guillotine
of
IV or
and
modes.
continuous
vacuum
variable
status.
System
procedures
irrigation,
procedures
developed
operate,
and
and
the
surgeon
features
of
lens
handpiece
can
be
used
phaco.
liquefaction
accessories.
handpiece
Infiniti™*
slim,
lightweight,
with
quick,
multi-microprocessor
connections
Management
performance
cutter.
capability.
modalities
continuous,
ultrasonic
pole,
controlled
remote
control.
control
control
lens
removal
control
irrigation.
pressurized
levels
in
phaco,
I/A,
between
or
footpedal.
tones
is
designed
that
require
and
such
as
with a dual
to
allow
control.
the
The
to
customize
Infiniti™*
removal
with
exclusively
device
handpiece,
handpiece
with
U/S
autoclavable.
smooth
achieved
System
fluidic
performance.
of
ultrasonic
pulsed,
power,
as
via the
of
ultrasonic
control
loop
of
aspiration
modes.
of
vacuum
reflux
and
aspiration
and
VIT
surgical
for
confirmation
for
use
simultaneous
aspiration,
vitrectomy
purpose:
the
surgeon
system
the
Vision
options:
ultrasonic
torsional
or
coupled
combining
sonic
oscillations.
handpiece:
control
control.
with
(FMS).
Infiniti™*
power
and
“burst”
well
as
duty
front
panel,
power
offers
low-end
flow
rate
in
I/A,
VIT,
via the
footswitch.
flow
steps.
steps
using
of
in
as
and
to
is treatment
System:
the
40
kHz,
of
the
cycle
in
(AFR)
and
rates
to
touch
system
»
Voice
confirmation
during
changes.
»
Fiat
screen,
display
+
High-tech
+
Multi-channel
*
This
product
Accessory Accessory equipment
IEC
Standard
equipment,
Furthermore,
System additional system
configuration
responsible
of
System
the
Technical
active
that
is
graphical
uses
Equipment equipment must
(e.g.,
and
IEC
ali
Standard
equipment
for
continued
Standard
Services
matrix
tiltable
user
wireless
SmartPhaco™
connected
be
certified
IEC
60601-1
configurations
TEC
60601-1-1.
or
than
TEC
department
representative.
Follow regarding and
User The of may potential disposed
In into conservation, take-back or equipment Alcon Alcon
Thecrossed-bin reminds also waste of
local
governing
disposal
ordinances
or
recycling
packaging.
Information — Environmental
equipment
natural
also
order
to
our
environment
recycle
that
resources
contain effect of
hazardous
on
improperly.
avoid
the
we
encourage systems. many
of
you
for
its
the
environment
entry
and
Such
the
materials
in a beneficial
office
for
assistance
or
other
providers.
you
to
emphasizing
equipment
it
as
unsorted
symbol
use
located
take-back
the
requirement
separately,
municipal
P
If
you
need
more
information
or
recycle
local
local
systems
or
regional
Alcon
waste
office
available
administration,
for
more
surgical
color
LCD
and
rotatable.
interface.
remote
control.
to
according
950
for
data
for
medical
shall
Anyone
otherwise
provided
compliance
60601-1-1.
of
have
purchased
production.
substances
of
any
such
to
promote
you
to
take-back
in
way.
Please
in
take-back
on
equipment
systems,
to
and
not
waste.
on
the
to
you,
information.
step
or
mode
with
touch
screen
technology.
or
used
with
this
to
the
respective
processing
equipment).
comply
with
connecting
causing a different
by
Alcon
is
to
the
requirements
If
in
doubt,
consult
or
your
local
Alcon
and
recycling
device
components
plans
Considerations
This
and
requires
which
human
the
use
equipment
could
have
health
substances
natural
use
systems your
contact
resource
the
appropriate
reuse
end-of-life
your
options
through
local
while
collect
dispose
m-
collection,
please
or
reuse
contact
contact
your
your
if
ONE
Universal Universal
who
IMT!
Precautions
precautions
come
in
accessories
pathogens
In
any
or
body
be
uniformly
handled
and/or
circumstance,
fluids/tissues
in
accordance
guidelines.
Table
1-1
contact
to
help
prevent
other
considered
Guidance
is
intended
of
the
Infiniti™*
shall
be
observed
with
the
instrument
their
potentially
wherein
encountered
the
potentially
with
OSHA
and
Manufacturer's
for
use
Vision
by
and/or
exposure
to’
infectious
exact
status
are
unknown,
infectious and
or
your
in
the
electromagnetic
System
EMC
Statement
all
people
It is
important
accordance
blood-borne
materials.
of
blood
it
shall
harmful If
this
equipment devices and
on),
interference
(determined
the
interference
own
national
*
Reorient
*
Increase
+
Connect
different
the this
connected.
*
Consult
the
engineer
Declaration - Electromagnetic
environment
should
assure
that
it
is
used
to
instal!
with
the
instructions
with
causes
by
user
is
encouraged
by
one
or
more
or
relocate
the
distance
equipment
from
that
to
manufacturer
for
help.
Emissions - The
specified
in
below.
such
and
use
other
harmful
turning
of
other
between
into
which
the
or
your
The
an
environment.
the
equipment
in
order
to
devices
in
interference
the
equipment
to
try
to
correct
the
following
device(s).
the
equipment.
an
outlet
on a circuit
other
device(s)
Alcon
Infiniti™*
customer
in
prevent
the
vicinity.
to
other
off
the
measures:
field
service
Vision
or
the
is
System
user
RF
emissions
CISPR
RF
emissions
CISPR
Harmonic
IEC
61000-3-2
Voltage
Flicker EC
61000-3-3
11
11
emissions
fluctuations/
emissions
Group
1
Class
A
Class
A
Complies
The
Infiniti™*
Therefore,
interference
Based
for
use
to
the
for
domestic
The electromagnetic
its
in
on
extensive
in
all
establishments
public
low
purposes.
EMC
Statement
Vision
System
RF
emissions
nearby
electronic
field
experience
voltage
interference.
power
provides
uses
RF
are
very
eguipment.
the
including
supply
guidance
energy
low
and
Infiniti™*
domestic
network
on
only
are
Vision
and
that
steps
for
its
internal
not
likely
System
those
directly
supplies
to
take
function.
to
cause
is
suitable
connected
buildings
in
case
any
used
of
8065750238
15
Table
1-2
Electrostatic
discharge
IEC
Electrical
transient/burst
IEC
Surge
IEC
Voltage
interruptions,
voltage
on
input
IEC
Power (50/60 magnetic
IEC
Conducted
IEC
(ESD)
61000-4-2
fast
61000-4-4
61000-4-5
dips,
variations
power
supply
lines
61000-4-11
frequency
Hz)
field
61000-4-8
61000-4-6
Guidance
is
intended
of
the
Infiniti™*
short
and
RF
and
Manufacturer's
for
use
Vision
*
+6
KV
*
*
*
+
*
*
+
»
*
3A/m
3
150
contact
+8
kV
air
+2
kV
for lines #1
kV
for lines
+1
kV
differential +2
KV
common
<5%
Ur
Ur)
for
0.5
40%
Ur
Ur)
for 5 cycles
70%
(30%
for
25
cycles <5%
(>95%
for 5 sec
Vins
kHz
to
(>95%
(60%
Declaration - Electromagnetic
in
the
electromagnetic
System
power
input/output | +
mode
dip
cycle
dip
dip
in
dip
in
80
MHz
should
ㆍ・
supply | +
mode | +
{| + +2 KV
in
in
Uy)
Uy)
*
*
*
3A/m
3
assure
+4
KV
contact
は 8
kV
ar
+1
kV
for lines +7
kV
for lines
+1
kV
differential
common
<5%
Ur
Ur)
for
0.5
40%
Ur
Ur)
for 5 cycles
70%
(30%
for
25
cycles <5%
(>95%
for 5 sec
Vrms
environment
that
power
supply | Mains
input/output | mature
mode | Mains
mode | commercial
(>95%
dip
cycle
(60%
dip
in
dip
in
Ur)
dip
in
Ur)
it
in
is
specified
used
Floors not
use
material
commercial
the circuit (inductive
Mains
commercial
Infiniti™*
during that
an
uninterruptible
Power levels commercial
Portable
should
Vision separation applicable
Immunity - The
below.
in
such
an
should
be
around
to
power
shutdown
Infiniti™*
with
power
power
power
the
Infiniti™*
freguency
characteristic
and
be
System,
wood,
floors
avoid
laser
quality
or
hospital
due
Vision
sources
switching;
quality
or
hospital
quality
or
hospital
Vision
System
mains
Vision
or
hospital
mobile
used
no
closer
including
distance
to
the
frequency
power
Q@NF
Infiniti™*
The
customer
environment.
concrete,
that
stoppage
should
environment.
to
fast
System
that
can
e.g.,
should
environment.
should
environment.
requires
interruptions,
magnetic
of a typical
environment.
RF
communications
to any
cables,
calculated
or
are
covered
due
be
transients
on
the
generate
high
current
be
be
continued
System
supply
itis
be
or a battery.
fields
location
part
than
from
to
the
INIT)
Vision
ceramictile.
that
that
that
the
transmitter.
System
or
the
with
synthetic
to
ESD.
of a typical
To
avoid
avoid
powering
same
branch
fast
transients
motors).
of a typical
of a typical
Ifthe
use
of
operation
recommended
powered
should
of
recommended
be
in a typical
eguipment
the
infiniti™*
the
equation
user
Do
pre-
the
from
at
Radiated IEC
RF
61000-4-3
Note: Note
1:
Note
2:
a
Field FM
radio
to
fixed
{equipment verify the
Infiniti™*
b
Over
3
Vim
80
MHz
Uris
the
a.c,
At
80 These structures,
strengths
broadcast,
RF
transmitters,
or
normal
the
frequency
mains
MHz
and
guidelines
objects,
from
fixed
and
system) operation. Vision
System.
range
to
2.5
GHz
voltage
800
MHz,
the
may
not
and
people.
transmitters,
TV
broadcast
an
electromagnetic
is
used
exceeds
!f
abnormal
150
kHz
3V/m
prior
to
application
higher
apply
to
frequency
in
all
such
as
cannot
site
the
applicable
performance
80
MHz,
be
of
the
range
situations.
base
stations
predicted
survey
should
RF
is
observed,
field
strengths
test
Electromagnetic
for
theoretically
compliance
level.
applies.
radio
(cellular/cordless)
be
considered.
additional
should
be
Recommended
d=
1.2YP
d=
1.2YP
d=2.3YP
where P is transmitter
manufacturer distance
Field determined
should
frequency
Interference
equipment
in
strength
be
separation
the
maximum
in
watts and dis
meters
from
by an
less
than
rangeb.
may
marked
80
MHz
800
MHz
(W)
according
the
(m).
fixed
electromagnetic
the
occur
in
with
following
propagation
with
accuracy.
level
above,
measures
less
than 3 V/m.
If
the
and
To
measured
the
(equipment
may
be
is
affected
land
mobile
access
field
necessary,
by
the
electromagnetic
strength
or
such
distance:
to
800
MHz
to
2.5
GHz
output
power
RF
the
vicinity
symbol.
amateur
in
the
should
as
re-orienting
to
transmitters,
and
location
recommended
compliance
absorption
radios,
system)
rating
the
transmitter
separation
site
surveya,
level
in
of
(인
reflection
radio,
environment
in
which
be
observed
or
relocating
to
the
as
each
from
AM
and due
the
to
ンミュ
ONFINITI
Table
1-3
Recommended
the
Infiniti™*
environment
infiniti™* distance
Vision
equipment.
10
100
For
transmitters
be
estimated transmitter Note
1-
Note
At
2-
These structures,
between
System
rates
using
in
watts
80
MHz
Separation
Vision
in
which
Vision
System
as
recommended
at a maximum
the
equation
(W)
according
and
guidelines
800
objects,
System - The
radiated
can
portable
output
applicable
to
the
MHz,
the
may
not
apply
and
people.
Distances
help
and
3.8 3.8 12 12
Between
Infiniti™*
RF
disturbances
prevent
mobile
below,
Portable
Vision
electromagnetic
RF
communications
according
to
are
the
System
power
not
listed
above,
the
to
the
in
all
situations.
frequency
transmitter
separation
of
manufacturer.
distance
the
for
Electromagnetic
recommended
transmitter,
the
higher
frequency
and
Mobile
is
intended
controlled.
interference
equipment
maximum
separafion
where P is
propagation
range
is
RF
Communications
for
The
customer
(transmitters)
output
power
distance d in
the
maximum
applies.
affected
by
Equipment
use
in
an
electromagnetic
or
the
user
by
maintaining a minimum
and
the
of
the
communications
7.3 23
meters
(m)
output
power
absorption
and
can
rating
reflection
and
of
the
Infiniti™*
of
the
from
8065750238
17
WARNINGS
Most
of
these manual; greater please the
however,
detail
contact
Technical
The
Infiniti"
be
serviced
engineer.
lead
to
Good
clinical
irrigation,
applicable
eye.
Ensure
any
phase
AND
CAUTIONS
warnings
here.
your
Services
by a factory-trained
Access
injury.
practice
aspiration
for
that
the
of
operation.
are
for
easy
If
additional
local
Department.
Vision
by
flow,
each
handpiece
tubings
Alcon
dictates
stated
elsewhere
reference
information
service
System
untrained
testing
reflux,
are
not
in
this
they
are
repeated
is
required,
representative,
battery
and
prior
occluded
can
Alcon
personnel
service
for
adequate
operation
to
entering
during
There
or
Infiniti™* all engineer.
in
WARNINGS!
only
can
as
:
are
no
service
Do
not
use
IV
pole.
Keep
clear
prevent
trapped
moves
height
Keep
from
clothing
during
adjustment.
clear
stored
user Vision issues
Legacy®
of
skin,
in
the
of
position
from
serviceable
System
to
the
hair,
power
display
being
console
your
factory-trained
pole
IV
pole
and/or
IV
pole
mechanism.
on/off,
base
to
prevent
trapped
components
or
footswitch.
extender
when
clothing
when
with
it
is
priming,
raising
skin,
at
the
base.
inside
Alcon
Infiniti™*
in
motion
from
The
and
hair,
the
Refer
service
to
being
IV
pole
bottle
display
and
/or
Do
notexceed
ml).
Excessive
drain
bag
in a hazardous
Inadvertent handpiece condition
Inadvertent
system
If
the
handpiece
tuning, there through
imbalance. or
collapsing
If
the
Infiniti™*
-
240V
range
should
phase
be
circuit.
maximum
maximum
actuation
is in
that
pressing
is
active
is a potential
the
handpiece
This,
of
in
used
pressure
capacity
condition
the
eye
may
result
will
cause
test
chamber
in
turn,
the
anterior
Vision
System
the
United
on a center-tapped,
capacity
can
resultfrom
and
for
the
of
Prime
can
in
patient
of
Standby
unit
of
and
may
may
chamber.
States
of
drain
bag
(500
exceeding
potentially
patient.
or
Tune
create a hazardous
injury.
switch
to
shut
is
collapsed
low
irrigation
result
cause a shallowing
is
used
or
result
while
when
down.
after
flow
in a fluidic
at
the
220V
Canada,
240V
single
a
it
A
gualified
inspection
twelve
+
*
*
In
Alcon
A
continuity
to
(for
be standards, do
Services.
Use
provided
decreased
mobile
this
months:
Warning Power
Fuses
case
ensure
Cord
of a deficiency,
Technical
qualified
they
example:
recorded,
not
use
of
accessories
RF
medical
technician
of
the
Labels
Services.
technician
for
leakage
are
within
EN60601-1/1EC601-1).
and
if
or
50%
the
system;
may
result
immunity
communications
electrical
must
following
do
current
the
they
are
above
and
cables
in
increased
of
the
equipment.
perform
components
not
use
must
every
applicable
above
initial
call
Alcon
system.
equipment
the
system;
check
twelve
standards
Values
the
applicable
measurement,
Technical
other
than
emissions
Portable
can
a
visual
every
:
ground
months
must
those
and
affect
call
or
1.8
ÓN
EIDE
Handpiece The
Infiniti™*
NeoSoniX®,
.
surgical
The
handpiece
while
in handpiece is
completely
cleaning
for
Use
The
Infiniti™*
handpieces
Allow
the
after
autoclaving;
when
hot.
Never OZil™ irreparable
Prior
OZil™ should and damage handling,
Do handpieces irrigating use. result
Ensure irrigating
NeoSoniX®,
dry
ultrasonically
to
placed
not
Irreparable
if
may
Care
AquaLase®,
and
high
instruments
operation.
must
dry
and
sterilization
(DFU)
must
handpiece
torsional,
damage
sterilization,
torsional,
always
in
to
the
and
operate
unless
solution
run
dry.
that
test
solution
or
result
and
tip
should
Immediately
be
thoroughly
before
supplied
NeoSoniX®,
be
at
to
never
NeoSoniX®,
NeoSoniX®,
have
the
the
sterilization connectors especially
OZil™
the
or
damage
chamber
before
U/S
handpieces.
in
premature
OZil™
performance
must
be
handled
not
touch
following
cleaned.
connecting
procedures,
with
the
handpiece.
OZil™
room
temperature
air
cool
immerse
CAUTIONS
clean
may
the
connector
torsional,
tip
distilled
for
the
the
Infiniti™*
result.
Infiniti™*
and
tray.
and
handpieces
during
is
immersed
water
to
the
is
filled
tuning
tip
torsional,
U/S
handpieces
with
any
solid
surgery
Be
sure
it
to
console.
see
the
torsional, and
just
at
Jeast
15
handpiece
AquaLase®,
or
U/S
handpieces;
AquaLase®,
U/S
handpieces
end
cap
This
will
autoclaving.
NeoSoniX®,
in
BSS®
or
is
in
handpiece
with
BSS®
OZil™
Tuning a handpiece
failure
torsional,
and
are
care.
object
the
cord plug
For
Directions
U/S
before
secured
during
and
use.
minutes
in
liquid
prevent
or
U/S
sterile
surgical
tip
sterile
breakage.
can
Use
of
the
U/S,
or
AquaLase®
irrigation
or
lost
aspiration
heating
eye
Appropriate
Parameters successful low settings, during failure to combinations
significant
site
eye
Use OZiMm
or
authorization,
patient patient
The
pak
Infiniti
U/S
andthen
ordinances.
Use
infusion
liquefaction sleeves.
sleeves imbalances.
and
tissues.
flow
occlusion
to
using
power,
and
inside
tissue
of
an
torsional,
use
of a handpiece
injury,
and/or
U/Stips
are
only
"*
tip
is
intended
disposed
0.9
mm
sleeves.
may
WARNINGS!
OZi/™
flow
rates, extended
sufficiently
damage.
NeoSoniX®,
Mismatching
torsional,
handpiece
and/or
potential
use
of
and
accessories
procedures.
low
conditions
excessively
of
the
temperature
the
eye,
ultrasonic
Infiniti™*
is
not
including
operator.
supplied
to
be
used
to
ofaccording
U/S
tips
Use
tips
exclusively
create
potentially
Infiniti™*
in
the
presence
flow
can
thermal
Infiniti™*
Use
of
bottle
power
above
permitted,
inthe
be
heights,
usage,
aspirate
tight
actions
increases
and
lead
handpiece
NeoSoniX®,
repaired
potential
/nfiniti™*
on
the
or
U/S
used
only
exclusively
1.1
mm
with
U/S
NeoSonix®,
in
the
absence
of
cause
injury
is
low
(beeping
viscoelastic
to
and
handpieces.
to
U/S
tips
hazardous
excessive
to
Vision
important
vacuum
high
power
incisions,
may
at
the
severe
other
without
may
shock
Vision
OZi/™
once
local
governing
with
and
1.1
mm
and
of
reduced
adjacent
System
for limits, power usage
tones),
prior
and
result
than
result
hazard
torsional,
per
infusion
in
incision
thermal
the
or
UIS,
Alcon
in
to
System
Each
case,
0.9
mm
11
mm
infusion
fluidic
Quenching a hot damage
8065750238
and
will
handpiece void
warranty.
in
water
can
cause
Directing
cause
energy
tissue
toward
damage.
non-lens
material
may
1.9
Uliraflow™* Prior
to
the
tip
screws
damaged,
Technical
Use
of
VA
handpieces
specifications, specified may
resultin a fluidic
cause a shallowing
chamber.
Exceeding
with a 0.5
chamber
of
the
VA
tips
torsional,
Recommended It
is
important
when
operating
or
0.3
mm
above
100
(U/A)
each
procedure
onto replace Services
non-Alcon
for
use
the
mm
shallowing
posterior
are not
or
U/S
Vacuum
that
with
WA
tips
mmHg.
Handpiece
inspect
the
Ultraflow™*
the
o-rings.
Department.
WARNINGS!
surgical
that
do
or
use
with
the
imbalance.
or
recommended
or
larger
and/orincarceration
capsule.
to
be
used
handpieces.
only
the
maximum
should
the
If
in
reusable
not
of
an
Infiniti™*
collapsing
I/A tip
with
Range
proper
vacuum.
be
used
two
handpiece.
doubt,
meet
Alcon
Alcon
Vision
This,
level
may
NeoSonix®,
for
VA
size
with
O-rings
contact
or
disposable
surgical
handpiece
System,
inturn,
of
the
anterior
of
100
cause
I/A
vacuum
anterior
ortearing
Tips
tip
be
Only
0.2
where
If
Alcon's
not
may
mmHg
OZiI™
used
mm
limits
Infiniti™*
The guillotine
Do the solution
Irreparable result
After
the
With surgeon is *
*
*
*
Vitrectomy
Infiniti™*
vitreous
not
test
tip
is
immersed
or
if
run
priming
probe
visual Ifthe not
the The
position
port probe. If
air
or
exiting
the
Ifa
observed
immediately
is
the
probe should
verification
cutteris
move
probe.
port
in
is
partially
bubbles
probe.
reduction
Probe
vitrectomy
cutter,
WARNINGS!
or
operate
distilled
damage
dry.
and
properly
when
should
footpedal
the
during
to
before
tip
in
sterile
step
on that
observed
the
always
closed
are
observed
probe
of
cutting
the
and
replace
probe,
is
in
water
probe
position 0 or
tip
an
intended
vitrectomy
BSS®
or
is
the
vitrectomy
surgical
actuating
irrigating
the
foot
the
probe
to
not
is
remain
while
in
during
capability
surgical
the
oscillating
for
single
use
probes
sterile
in
surgical
use,
and
treadle
is
fully
close
actuated,
1.
idle,
the
aspiration
priming, replace
procedure,
probe.
unless
irrigating
probe
verify
aspirating.
solution,
until
there
cutting.
ordoes replace
in
an
open
Ifthe
cutting
replace
or
vacuum
only.
use.
can
that
the
the
line
is
stop
V/A
adjustable
Handpiece Ensure handpiece. generated that
use.
Use may
the
of a tool
cause
Poor
secured
that
and/or
tip
clinical
vacuum
Tips handpiece
If
not
securely inadequate
is
not too
other
than
damage
tightly
to
performance
to
range
is
tip
is
fully
attached,
tuning
tight
so
that
tip
wrenches
the
tip
and/or
WARNING!
the
handpiece.
0-650+.
tightened
an will
it
can
supplied
handpiece.
will
resuit
to
error
occur.
be
removed
if
the
may
Ensure
by
tip
be
after
Alcon
is
not
Aspiration/Vacuum Adjusting preset
values
exceeding
Dynamic in
shorter
engage
non-lens
Adjusting
the
preset
preset
collapse
aspiration
may
result
irrigation
Rise
values
periods
values,
of
material.
aspiration
values,
may
which
may
Adjustments
rates
or
vacuum
in
aspiration
inflow.
of
1,2,
3,
time.
Care must
WARNING!
rates
or
or
lowering
cause
result
chamber
in
limits
levels
or 4 will
be
vacuum
the
IV
pole
patient
above (volumes)
achieve
taken
shallowing
injury.
not
limits
belowthe
vacuum
above
the
to
or
Presurgical Presurgical outlined error
message
front
panel,
manual.
When
fluid
is
be
jeopardized.
testing
prior
Ensure
any
phase
Footswitch
Never
pick
Damage
High
Altitudes Vitrectomy Consult
Occlusion
\
Two
different
tones
during
or
at its stopped the
I/A
during
power
tone
type a occlusion or
The that value. the
8065750238
aspiration
or
is a lower,
of
higher,
AquaLase®
I/A
the
The
phaco
The
phaco
flow.
use
may
Use
AquaLase®
flow
aspiration
potential
Check-out
check-out
in
the
Operating
or
advisory
refer
to
If
the
problem
filling
to
Alcon
preset
to
occlusion
occlusion
occlusion
vacuum
Use
of
and/or
handpiece
weak
or
absent,
for
adequate
entering
that
the
of
operation.
up
or
move
may
result.
cutting
Technical
Tones
occlusion
occlusion)
limit,
avoid
exceeding
tone, only
AquaLase®
intermittent
tone,
intermittent
occurs
during
magnitude.
has
I/A
occlusion
occlusion
occlusion
of
high
lead
to
the
NeoSonixX®, handpiece
in
flow
thermal
Tests
tests
must
Instructions
message
the
Troubleshooting
persists,
WARNINGS!
test
good
Good
clinical
irrigation
the
eye.
tubings
performance
and
magnitude). The
double
and
reached
tone
WARNINGS!
U/S
thermal
the
can
injury
are not
the
footswitch
Service
tones
(intermittent
indicate
aspiration
the
sounds
(in
the
phaco
presence
when
the
absence
single
phaco
beep,
application
occlusion
its
tone
can
cannot
bell
indicates
settings
injury.
OZil™
in
the
cause
to
be
performed
section.
is
displayed
DO
NOT
chamber,
fluidics
practice
and
aspiration
occluded
by
may
vary
for
additional
that
the
flow
limit.
The
occlusion
of
YA
beep.
occlusion
and
sounds
of
ultrasonic
tones
maximum
be
turned
be
turned
and/or
torsional,
absence
of
reduced
excessive
adjacent
as
If
an
on
section
response
the
at
vacuum is
ultrasonic
The
no
eye
of
PROCEED.
if
stream
dictates
during
cable.
high
altitudes.
information.
beeping
is
reduced
first
type,
occurs
occlusion
second
tone,
is
when
power
indicate
allowed
off,
while
off.
aspiration
prolonged
U/S,
of
irrigation
or
heating
tissues.
the
this
of
will
the
flow
near or
preset
or
lost
and
Vacuum A to that
reduced
Coagulation
Listed
using *
+
*
*
*
*
*
+
*
Tone vacuum the
A
little OZ が torsional,
in
presence
cause injury
Do
ml).
drain
in a hazardous
To only used performance components
To
always surgical Interference frequency the Accessories cables damage Operation extraocular The always Skin-to-skin and by When and simultaneously electrodes surgical not In frequency
tone
is
amount
little
moderate
to
the
not
Excessive
below
the
ensure
approved
(See
reduce
operation
lowest
body
insertion
physiological
recommended.
all
of
to
no
flow
in
volume,
to
no
flow
absence
of
reduced
excessive
to
adjacent
exceed
bag
maximum
Function
are
Coagulation
safe
cables
your
can
or
the
risk
be
taken
equipment.
produced
surgical
of
should
should
HF
cases,
particularly
to
the
insulation.
of
the
uses
power
be
selected
contact
of
the of
(high
should
electrodes.
monitoring
current-limiting
provided.
vacuum. A high
is
maximum
pressure
condition
general
operation
Alcon
Alcon-endorsed
when
other
coagulation
only.
patient)
dry
frequency)
on
The
is
occuring.
but not
high
of
be
level
monitoring
be
turned
WARNINGS!
vacuum
occuring.
U/S,
irrigation
heating
eye
capacity
function:
and
representative).
guaranteed
of
accidental
operating
by
equipment
electronic
be
for
the
(for
gauze.
the
placed
Needle
Use
or
AquaLase®
or
lost
and
tissues.
capacity
can
for
precautions
of
the
accessories
the
operation
checked
be
checked
step
in
coagulation intended
example between
should
surgical
equipment
same
patient, as
monitoring
systems
devices
pitch
will
vary
vacuum
This
off.
tone
flow
aspiration
result
and
the
coagulation
only
components.
burns,
high-frequency
may
equipment.
regularly;
be
far
incorporating
can
tone
can
may
of
the
NeoSonix®,
handpiece
and/or
potential
of
drain
from
exceeding
potentially
patient.
to
be
followed
must
be
Coagulation
when
using
caution
of
high-
adversely
electrode
for
possible
is
limited
step
should
purpose.
the
avoided,
equipment
are
any
monitoring
as
possible
electrodes
are
recommended.
relative
indicate
be
indicate
in
the
flow
can
thermal
bag
(500
result
when
function,
Alcon
should
influence
to
arms
for
example
used
from
the
are
high
1.11
+
The
cables
positioned
or
other
*
Temporarily
so
that
*
The
such
avoided region are
sucked
e
Non-flammable
disinfection
+
Flammable
or
as evaporate materials, saturated produced
+
Do with
devices. implanted devices that
defibrillatory and with manufacturers
*
Failure
circuitry)
of
The
Infiniti™*
the
effects
Cautery,
The
InfinitiTM* “Coagulation” following
Coagulation - an conductors electrodes,
some
of
to
in
leads
unused
they
are
use
of
flammable
as
nitrous
if a surgical
of the
thorax
away.
wherever
agents
solvents
before
for
with
in
normal
not
use
pacemakers
Ifelectrosurgery
or
irreparable
lead
to
the
pacemaker
ofthe
could
output
power.
of
defibrillator
Diathermy,
in
definition:
(e.g.
halting
the
text
such a way
cardiac
Vision
the
surgical
that
is
avoided.
active
isolated
oxide
agents
of
adhesives,
the
example
oxygen
the
pacemaker
device
ventricular
HF
Vision
place
isolated
forceps).
bleeding.
of
this
from
anaesthetics
(N20)
procedure
or
the
should
possible.
used
for
application
cotton,
may
use
of
WARNINGS!
coagulation
or
pacemakers
damage
and
for
their
surgical
result
CAUTION
System
Coagulation
System
of
Cautery,
bipolar Current
operator's
electrodes
contact
electrodes
the
and
head,
cleaning should
be
the
implanted
is
electrodes,
its fibrillation. or
recommendations.
equipment
in
an
is
discharge.
uses
(Abbreviated
should
with
should
patient.
or
oxidizing
oxygen
is
unless
be
used
of wool ignited
HF
function
used
to
the
function
defibrillatory
unintended
not
the
based
current
passes
manual.)
should
carried
these
for
or
disinfecting,
be
allowed
HF
surgery.
and
gauze,
by
surgical
defibrillatory
on
patients
or
defibrillatory
pacemaker
Please
(coagulation
protected
Definition
word
on
supplied
between
“Coag”
be
the
patient
be
stored
gases
be
out
in
the
agents
cleaning
to
Some
when
sparks
equipment.
on
patients
with
be
aware
may
occur
check
device
increase
against
the
to
these
in
and
or
Consumable Consumable System then
discarded,
All
Infiniti™*
is
important
use.
NOTE: the
instructions
the
Directions
consumable
Mismatch
of
settings combination create a patient
Do
not
date.
Sterile
reused! These
time
The
equipment
disposables
system. disposables
create to
have
equipment voidance prevailing
In
all contained understood
configurations.
Read consumable
Paks
items
used
during
use
surgery
unless
paks
to
read
If
an
inconsistency
in
For
pak
or
of
consumable
not
specially
of
hazard.
use
paks
that
disposable
(Accreditation
components
only;
do
used
constitutes a complete
Use
of
may
potential
contributed
under
of
the
hourly
cases,
all
the
in
the
prior
package
paks
with
the
are
designed
labeled
contain
and
the Use
accessory,
WARNINGS!
consumable
medical
not
disposables
affect
hazards,
contract,
contract
rates.
instrument manual
label
prior
otherwise.
Directions
understand
exists
operator’s
(DFU)
adjusted
have
exceeded
Manual
have
been
reuse.
in
conjunction
system
and
to
the
to
using
material
to
their
Infiniti™*
to
be
for
the
DFU’s
between
manual supplied follow
components
devices
should
the
for a particular
components
for
Hospitals,
designed
other
performance
if
it
malfunction
could
and/or
setup
be
any
printed
use.
used
with
the
should
with
is
result
invoicing
instructions
of
Vision
once
and
Use
(DFU). prior
to
and
a
DFU.
and use
may
expiration
not
be
1982.)
for
one
the
Alcon
surgical
than
Alcon
and
determined
of
the
in
the
thoroughly
the
pak
on
the
It
at
Miscellaneous
*
Donotuse anesthetics.
*
Avoid
spilling
around
*
Do IV
are
be
door
*
Do
+
The located personnel
Tray when
the
not
push
pole.
Handles
provided
pulled
thresholds.
not
place
USB
connector
on
support
moving
the
Infiniti™*
BSS®
electrical
or
pull
the
located
for
moving
and
not
pushed,
more
than a 20
(e<»)
the
rear
only.
Failure
WARNING!
must instrument.
CAUTIONS
Vision
solution,
handpiece
unit
by
at
the
the
instrument.
especially
Ib.
and
panel
are
to
comply
be
set
System
or
moisture
connectors.
the
display,
rear
and
over
load
on
InfinitiM*
for
use
by
will
void
in
its
stored
near
flammable
of
the
tray, or
sides
of
The
unit
elevator
tray
support.
port(
Alcon
warranty.
position
any
kind,
the
should
«d+
trained
the
unit
and
)
[D
8065750238
ONFIN
INITITY*
Alcon's microprocessor-controlled instrument C/O)
circuitry.
its
Front
with
tones.
system
power
VISION
Infiniti™*
with
associated
The
Panel
display,
An
automatic
is
turned
SYSTEM
Vision
system
with
on.
System
ophthalmic
memory communicates
self-test
.
This
test
performs a variety
following:
*
Tests
the
+
Tests
the
*
Initializes
IV
Central RAM
Pole
the
with
system
Processing
and
ROM
Bottle
Hanger
D
is a multi
and
voice
confirmations, is
initiated
of
functions
Unit
(CPU)
memory,
and
TI
surgical
input/output
to
the
user
and
each
including
the
I/O
circuits
via
time
the
When
the
system
automatically
fails
the
self-test,
This
section
parts.
accessories. including interface,
part
of
is
where
of the
The
first
All
the
display
footswitch,
this
section
the
display
programming,
successfully
goes
into
an
error
manual
part
describes
the
parts
panel,
and
describes
screens
and
dialogs
the
Setup
message
is
of the
IV
remote
for
are
Front Touch
completes
mode.
is
broken
the
into
console
system
pole,
connectors,
control.
the
operator
system
shown.
|
Display
Screen
the
self-test,
If
the
system
displayed.
two
major
and
its
will
be
described,
The
second
interface.
setup,
surgery,
Panel
and
it
fluidic
This
Fluidic
Management
inserted
Figure
into
1-5
System
Fluidic
Module
Accessory
Caster
The
Console - The
by
the
(FMS)
Drawers
Wheels
surgeon
(4)
to
console
perform
contains
lens
extraction
all
the
controls,
surgery.
connectors,
Front
Instrument
Footswitch
Footswitch
and
communication
Connector
Tray
Drawer
Panel
devices
required
ÓN
FIN
CONSOLE
Fluidics The
panel. Fluidic module can
Front The maneuverability
and display allowing
Front The connector module.
connectors,
two AquaLase®/balanced
and
vitrectomy
facilitate
Module
fluidics
The
module
module
Management
contains
be
started
Display
front
display
transport
is
the
the
user's
fingertip
Connector
It
provides
one
connectors
one
luer
lock
handpiece.
handpiece
all
without
Panel
panel
during
front
command
Panel
panel
AquaLase®
for
pneumatic
is
located
allows
fast
at
the
and
System (FMS),
the
connections
delay.
and
Touch
tilts
setup
panel
main
source
is
located
two
self-locking
bipolar
salt
solution
Screen
and
rotates,
and
surgery.
folds
of
of
system
to
the
handpiece
coagulation handpieces,
connector
Symbols
near
identification.
top
of
easy
insertion
and
because
required,
allowing
down.
The
system
functions.
right
U/S
handpiece
connector,
bottle
receptacle,
for
the
connectors
the
front
of
the
surgery
easy
For
storage
front
control,
of
the
fluidics
the
anterior
the
an
Footswitch The
footswitch When the
footswitch. identified two
small
be
placed
insertion
Accurus®/Legacy®
insert
Two
footswitch
this
drawer.
enhanced
Legacy®
in,
and
Instrument Provides a movable field. that
allows
with
sterile tray
Drawer
not
in
use,
by
its
holes
in
the
and
removal
from
bottom
The
Infiniti™*
footswitch.
exits
from,
Tray
There
is a curved
for
drawer
The
ribbed
in
is
this
drawer
enhanced
rubber
its
heel,
bottom
of
footswitch
of
drawer.
cable
connectors
left
connector
footswitch; The
the
drawer.
instrument
creation
support
at
of
the
footswitch
metal
of a sterile
cover.
accommodating a variety
room
environment:
surgeon
as
well
as
right,
the
front
adjustable.
the
bottom
is
used
Infiniti™*
footpedal
requires
the
drawer.
enhanced
is
is
the
tray
rod
The
of
positions
left,
front
of
the
of
the
to
store
footswitch,
surface and
that a plastic
This
footswitch.
used,
remove
are
located
for
the
Infiniti™*
right
for
cord
is
within
on
pouch
tray
and
bed.
the
in
The
the
tray
when
is
capable
the
rear
front panel.
and
protect
insert
allows
easy
If
the
plastic
behind
and
Accurus®/
also
stored
sterile
arm
used
of
operating
of
the
tray
is
height
:
Figure
8065750238
a
Coagulation Connectors
U/S
and
“È
1-6
The pieces and
NeoSonix®
Vitrectomy Connector
Connectors
(2)
Front
Handpiece
(2)
Handpiece
Handpiece
Connector
consumables.
Panel - The
front
connector
panel
allows
+
i
AquaLase®
Aqualase®/BSS®
o
quick
and
Handpiece
Bottle
easy
connection
Connector
Receptacle
of
hand-
1.15
IV
Pole
with
A
bottle
of
from
the
to
raise
and
pressure
Do IV
to
not
pole.
Accessory Two
drawers
Bottle
BSS®
hook
on
lower
increase
use
Legacy®
Drawers
allow
Hanger
or
BSS
top
of
the
bottle
or
decrease.
WARNING!
storage
Plus®
this
pole.
height,
IV
pole
extender
of
miscellaneous
irrigating
The
IV
fluid
pole
causing
with
is
hung
is
used
irrigation
Infiniti™*
accessories.
Handles Handles
instrument,
For not
The system push pole. are be door
are
greater
pushed.
system
could
or
pull the
Handles
provided
pulled
and
thresholds.
located
and
should
safety
must
tip
unit
located
for
moving
not
on
the
always
and
control,
CAUTION
be
moved
over
and
by
at
pushed,
sides
and
be
the
carefully,
become
the
display,
the
rear
the
instrument.
especially
back
used
unit
and
ON
FINITI
of the
to
move
should
be
otherwise
damaged.
the
tray,
sides
of
The
unit
over
elevator
the
unit.
pulled,
Do
or
the
the
unit
should
and
the not
IV
Caster Four
System.
Wheels
large
The
mobility,
the
system
when
the
caster
wheels
and two
in
place.
unit
is
in
wheels
rotate
wheels
The
use,
support
360°
the
for
ease
have a locking
wheels
and
should
unlocked
Infiniti™*
of
system
lever
to
always
when
be
being
Vision
secure
locked moved.
Figure
1-7
The
cord
Rear
wrap,
Connectors
Equipotential
Panel - The
and
Fuse,
Power
AC
Power
standby
and
CD/DVD
Ground
Connector
rear
power
Connector
Switch,
panel
Drive
and
contains
switch
the
(shown
power
in
Figure
module,
1-8).
electrical
connectors,
CD/DVD
8065750238
drive,
ÖNE
REAR
Power
,
The
power
power A
INT
Master
PANEL
Module
power
switch,
module
standby
panel.
*
AC
Power
outlet
connects
be
used.
*
Main
Power
supply.
*
Fuse
Drawer - Holds
system
Equipotential For
Service
Cord
Wrap
Used
to
store
side
of
the
Connectors
“This
module,
contains
various electrical next
to
the
the
system.
e
e>
+
10101
ei
rita
module
power
to
identify
personnel
rear
contains
and a fuse
is
located
switch
an
drawer
at
the
is
located
Connector - Power
here. A hospital
Switch - Connects
fuse.
Refer
size
and
type.
Ground
Connector
use.
the
power
supply
panel.
and
CD/DVD
located
connectors
in
the
Drive
middle
and
AC
(see
bottom
at
cord
grade
AC
to
cord.
outlets
interconnections. A CD/DVD
connectors,
USB
Connector - Not
Serial
Infiniti™*
is
used
for
used.
Connector - Used
Port - Not
used.
software
power
connector,
Figure
the
from
of
the
top
AC
1-7).
of
power
power
to
label
on
Located
of
the
rear
used
drive,
upgrades
for
VideOverlay.
The
rear
panel.
the
power
cord
power
back
on
the
panel,
for
located
AC
rear
must
of
right
to
Data
Card
Slot
A
data
card
(e.g.,
inserted
back
the Copy/Delete the or
System
into
up
or
restore
Copy/Delete
Infiniti™*
copy
data
from a data card
(restore).
this
dialog
Standby Power This
pushbutton
ON
and
OFF.
to
operator
seconds
commands,
to
shut
Inadvertent system
Audio
The
console. conjunction Vision tones mode as
an generated increases are
is
active
Speaker
audio
speakers
These
with
System
are
generated
and
to
alert
occluded
to
audibly
as
the
adjustable
Multi
Media
data card
system
option
allows
Vision
slot
settings.
from
the
System
the
Switch
switch
If
system
is
used
freezes
press
down
system,
WARNING!
pressing
are
speakers
multiple
to
communicate
to
the
line.
vacuum
via
the
of
will
cause
located
produce
tones,
indicate a change
operator
Additionally, a varied
indicate
level
Custom
Card
(MMC))
when
the
This
Custom
user
to
copy
to a data
to
to
and
Standby
then
card
the
Infiniti™*
turn
secondary
is
unresponsive switch
re-boot.
Standby
unit
to
on
each
voice
confirmations,
to
allow
with
the
of
certain
vacuum
levels;
increases.
menus.
can
user
wants
is
done
by
drop
list.
data
from
(backup),
Vision
for
switch
shut
down.
side
of
the
the
Infiniti™*
user.
Audible
in
the
operating
conditions
pitch
tone
the
Speaker
be
to
using
The
power
five
when
in
such
is
pitch
volumes
„=
The
USB
located
personnel
Figure
8065750238
connector
on
the
rear
only.
Audio
Multi
Data
1-8
The
The
CAUTION
(e+)
panel
Failure
Speaker
Media Card
Slot
Right
Side
left
side
and are
to
comply
(2)
Card
in
Panel - The
panel
Infiniti™*
for
use
by
will
void
right
has
the
port
Alcon
warranty.
side
other
(+4)
trained
panel
speaker
contains
and two
the
data
accessory
card
slot
drawers.
Standby
and
Power
Switch
one-of-two
audio
speakers.
ON
FIN
FOOTSWITCH
The
Infiniti™*
Alcon
footswitches.
footpedal,
and Legacy® than a swivel
The graphical
When
3)
is arrow activated. position
on/off
on/off
footpedal
type
footswitch
representation
connected,
displayed
appears
If a footswitch
is
displayed
Vision
System
The
toe
switches
swivel
of
footswitch
footpedal.
icon
button
the
icon's
in
the
center
next
to
the
in
the
can
utilize
Infiniti™*
(horizontal
switches. contains
on
the
display
of
the
footswitch
footpedal
of
the
icon,
icon
each
is
not
connected,
icon.
two
different
footswitch
The
heel
has
and
vertical),
Accurus®/
switches
screen connected.
position
(0,
and a triangular
time a switch
no
footpedal
a
rather
is
a
1,
2,
is
or
Several are The
flow,
functions
controlled
footpedal
aspiration
AquaLase®
power. adjust
The
function
steps.
The
footswitch Footpedal footpedal are
listed
the
Custom
within
by
the
enables
rate;
energy,
switches
settings,
actions
positions
positions/functions
in
Table
the
surgeon
the
surgeon
OZil™,
vitrectomy
are
used
and
are
shown
are
shown
1-4.
To
program
system's
using
NeoSoniX®,
to
to
in
feature.
operating
the
footswitch.
to
control
cut
rate,
turn
functions
progress
in
Figure
in
Figure
each
mode
the
footswitch,
modes
irrigation
or
U/S
and
coagulation
on/off,
through
1-12.
1-11,
and
of
operation
power;
to
surgical
use
Figure
1.18
19
The
Accurus®/Legacy®
and
Infiniti
"*
Footswitches
Plugging
The behind Infiniti™* Legacy® directions
Nr
3.
in
footswitch
the
footswitch.
below.
Open
the
Simultaneously on
each
allow
access
Grasping
cable
into
on
the
the
red
connector
smoothly.
the
Footswitch
plugs
into
footswitch
footswitch;
drawer.
the
To
footswitch
press a metal
of
two
hinges
to
the
the
footswitch
one
of
the
cable
connector
dot
on
the
console
is
in
the
one-of-two
One
other
plug
in
connectors
connector
is
for
the
footswitch
drawer.
drawer
to
release
footswitch
cable
two connectors.
must
correct
the
cable
connector,
be
in
connector,
position
is
the
Accurus®/
follow
extension drawer connectors.
plug
The
red
alignment
and
when
it
will
slide
for
and
dot
with
the
the
latch
the
the
in
NOTE: be same functional.
4. À cable
5.
6.
Only
used
at a time.
time,
Loosen
and
place
cable
restraint
two
screws.
cable
exists
Loop
drawer, of
the
drawer.
the
drawer.
Shut
the
the
cable
one
footswitch
If
only
the
Infiniti™*
restraint
the
the
cable
then
There
footswitch
is
two
screws
cable
over
Ensure
between
through
route
management
are
connector
both
connectors
footswitch
located
through
the
that a slight
it
high
drawer.
on
securing
cable and
the
connector
the
slot
through
system
and low
its
the
back
the
center.
secure
amount
in
the
the
left
in
slots
is
are
of
cable
Replace
and
the
back
or
right
the
front
on
intended
used
at
connector
the
drawer.
restraint
the
it
with
the
of
excess
restraint.
of
the slot of
the
each
side
to
the
is
of
Figure
1-10
footswitch
cable
cable
connectors
restraint
footswitch
footswitch
(2)
cable
drawer
management
system
—
drawer
Footswitch
Cable
extension
Routing
latch
8065750238
1.19
ON
FINITE
Footpedal Depending
option
surgical
coagulation).
depression to
the very is
equal treadle output
Control
on
to
select
parameter
With
within
parameter
start
of the
to
the
range.
is
fixed
Figure
the
surgery
linear
linear
the
output.
treadle
limit
With
at
its
or
fixed
(i.e.,
aspiration,
pedal
The
range,
value
fixed
limit
1-11
step,
the
footpedal
footpedal
range
is
parameter
and
specified
footpedal
value
throughout
Diagram
user
may
control
vacuum,
control,
directly
output
the
parameter
at
the
end
control,
of
Footpedal
have
the
of
a
power,
the
angle
of
proportional
is
0 at
the
output
of
the
the
parameter
the
treadle
Positions
range.
Footpedal footpedal when footpedal also
The
are Footswitch Button
ーーー
position
slightly
from
be
accompanied
footswitch's
programmable
is
described
TN
ニー
プー
detents
more
one
Buttons
Button
POSITION
POSITION
ー
\
identify
to
another,
pressure
position
by
vibration
and
and
are
in
the
Main
later
in
0
1
ペーーーーー
POSITION
———
у
POSITION
-—
the
and
are
is
required
into
the
if
Treadle
available
Window.
this
section
Testing
position
detent
deleni2
3
{ull
1
footpedal
2
transition
felt
by
the
to
press
next.
Detents
programmed
adjustments
by
pressing
The
Footswitch
of
the
manual.
ο
depression
from
one
operator
the
can
to
do
the
so.
Table
1-4
Table
tions.
selected.
of
Footpedal
This
table
As
Resting
Continuous
Resting
Continuous
Resting
Continuous
Resting
Continuous
Positions - The
shows
the
footpedal
Irrigation
Irrigation
Irrigation
Irrigation
Irrigation
Resting
the
functions
is
depressed
Irrigation
Irrigation
Irrigation
footpedal
controlled,
it
Irrigation/Aspiration | Irrigation/Aspiration
irrigation/Aspiration
Irrigation/Aspiration
Irrigation/Aspiration
Irrigation/Aspiration Irrigation/Aspiration
Irrigation/Cutting
Irrigation/Cutting
*
Phaco
*
NeoSonix
*
Torsional
+
AquaLase
Irrigation/Aspiration
*
» *
Torsional
*
AquaLase
irrigation/Aspiration
Power
Amplitude
Amplitude
Magnitude
Phaco
NeoSonix
Power
Amplitude
Amplitude
Magnitude
lirigation/Aspiration
Cutting
Cutting
Irrigation/Cutting
Aspiration
Irrigation/Cutting
Aspiration
is
used
dependent
travels
from
Coagulation
by
the
on
the
resting
surgeon
mode
Power
to
control
of
operation
position
into
several
and
type
its
active
surgical
of
func-
irrigation
positions.
4.20
8065750238
ONEFINITI
Switch
The to
footswitch
horizontally
activate
Accurus®
switches
switches
Switch footswitch
The
factory
listed
Control
footswitch
control
various
has
when
/Legacy® that that
functions
left
horizonal
default
as
None,
icon
has
six
surgical
left
and
and
vertically,
the
pedal
footswitch
operate
activate
are
and
making
switch
action:
their
and
must
be
programmed
switches
right toe
is
horizontally
when
programmable
Reflux.
functions
Pressing
to
medium
that
functions.
switches
and
footpedal
shifted
is
adjust
left
has
left
and
pressed
selections
the
only
The
other
are
mutually
this
button
the toe
(2),
or
wide
Left
toe
switch
left
for
one
down
for
can
be
programmed
The
Infinitit**
that
operate
switches
or
right.
and
right
vertically,
down.
by
pressing
on
the
display.
switch with
five
switches
exclusive,
allows
switch
another.
the
for a narrow
(1)
foot.
is
pressed
function,
and
that
The
toe
and
the
a
operator
(3),
heel
are
Footpedal
t
foo
for
given a function the
other
Reflux,
Grade+/-,
if
engaged; and disengaged. the be permitted
functions
and
les
switches.
witch
the
same
Cont.
the
footpedal
cannot
mode
allowed
the
swiveled
Accurus.
utilizes
heel
functions.
command
left
already
switch
however,
is
Irr.,
Irr.
Up,
is
be
engaged
If
the
of
operation,
to
engage.
to
be
engaged
are
not
available
and
right
switches
The
plugs
fension
are
given a None
Step+,
Irr.
not
the
footpedal
will
footswitch's
also on
by
the
designated
Step-,
Down,
depressed,
Furthermore,
into
knob
and
switches
until
all
is
depressed,
certain
not
this
with
when
be
end.
and
this
end.
removeable
carrying
switches
the
performed.
The
user.
When a switch
to
another
designation.
Step+/-,
are
other
the
footpedal
Grade+,
None.
any
switch
mutually
switches
may
even
treadle
treadle
cable
handle
depending
if a switch
depressed,
switch,
Choices
Grade-,
may
be
exclusive
are
on
or
may
not is some
is
depressed,
is
are
Left
Horizontal/Vertical
Switch
Pressing
{ο
adjust
for a short
size.
different
this
the
It
can
balance
Right
toe
one
function,
button allows
heel
rest
Sì
medium
also be
adjusted
and
Footpedal
switch
is
pressed
and
down
for
the
operator
forward
or
backward
(M),
or
long
(L}
to
attain
feel.
Right
Horizontal/Vertical
right
for
another.
foot
a
Switch
Figure
8065750238
1-12
Footswitches switch,
and
Left
Heel
below
Switch
Used
with
the
it
Infiniti™*
is
the
Accurus®/Legacy®
Vision
System - Shown
footswitch.
Right
at
Heel
the
Switch
top
is
the
Infiniti™*
foot-
4.21
The
following
function
permitted, when
the
assigned the
treadle
*
Reflux
The
default
height
pressure.
using
the
sections
is
permitted
and
the
user
treadle
is
depressed,
to a switch
depressed.
reflux
pressure
The
Reflux
Offset
indicate
with
intends
that
reflux
control
menu.
In
all
cases,
reflux
is
not
depressed,
*
Continuous
When a switch
is
toggled, activates/deactivates. available available
and
is
not
available
Irrigation
On/Off
assigned
the
continuous
Continuous
when
the
footpedal
in a Coagulation
whether
the
treadle
to
control
the
function
is
permitted
is
equal
pressure
in
the
available
in a Coagulation
to
Continuous
irrigation
irrigation
is
in
any
step.
each
depressed.
that
must
to
be
engaged
-
to
the
current
can
be
increased
Custom/Doctor
when
the
footpedal
Irrigation
status
immediately
toggling
position,
switch
If
it
function
be
with
bottle
step.
On/Off
but
is
is
is
is
not
*
Cataract
A
switch (Grade grade of
the
Grade
may
+),
cataract increase/decrease cataract
AquaLase®
footswitch
If
cataract assigned, cataract cataract then switch pressed
decrease
If
the cataract grade grade selected,
treadle
grade
when grade grade
cataract
is
pressed
for
more
function
highest
grade
will
be
is
currently
the
Increase,
be
assigned
grade
Decrease,
as
decrease
(Grade
grade
is
only
available
steps,
and
is
available
is
depressed.
increase
the
to
the
switch
current
or
is
increase/decrease
grade
increase
for
less
than
will
cataract
increase
selected.
selected
highest
cataract
will
than
1/2
second,
be
activated.
grade
is
selected,
Similarly,
and
cataract
(Grade
+/-).
Footswitch
whether
cataract
grade
pressed,
cataract
is
assigned
be
activated
1/2
second.
then
is
currently
the
if
the
cataract
grade
will
ONE
INET
Increase/Decrease
grade
increase
-),
or
cataract
control
in
phaco
and
or
not
the
the
next
grade
decrease
is
is
or
previous
selected.
If
to a switch,
if
the
If
the
the
cataract
selected,
lowest
cataract
lowest
grade
decrease
be
selected.
cataract
switch
is
grade
and
is
*
Step
Advance, A switch back
may
(Step
-), Coagulation, from
this
stepping
If
step
advance
is
pressed,
is
selected
assigned
if
the
switch
back
the in
to a switch,
switch
is
pressed
function
Step
be
assigned
or
step
and
Anterior
or
next
the
surgery
is
pressed
for
will
Back,
Step
as
advance/back
Vitrectomy
sequence.
step
back
is
or
previous
menu.
then
step
for
less
more
than
be
activated.
Advance/Back
step
advance
(Step
steps
assigned,
step
to
the
If
step
advance
than
1/2
1/2
second,
(Step
+),
step
+/-).The
when
current
are
excluded
the
Setup,
switch
step
advance/back
will
be
activated
second.
then
If
the
the
is
step
When a new
parameters
new
cataract
+
Irrigation
A switch
down.
released,
down. second,
until
function all
may
When
the
If
the
the
the
switch
is
footpedal
cataract
will
be
updated
grade.
Up,
Irrigation
be
assigned
the
switch
IV
pole
switch
IV
is
pole
will
is
released.
available
positions.
grade
is
with
Down
as
is
pressed
position
pressed
move
Control
in
all
steps
selected,
irrigation
will
and
surgical
those
specified
up
and
immediately increment held
for
continuously
of
this
but
coagulation,
for
the
or
irrigation
up
or
more
than
up
or
down
irrigation
and
1/2
in
1.22
8065750238
Loading...
+ 108 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.