Alcon Accurus 300VS, Accurus 400VS, Accurus 600DS, Accurus 800CS, Accurus Cart User manual

4.7 (6)

Data Sheet for Accurus® 300VS, 400VS, 600DS, 800CS & Cart

Customer

 

 

 

Work

Order

No.

 

 

Model/Platform

 

 

 

 

S/N

 

 

NOTE:

Throughout this Data Sheet either

fill

in the

requested

data

or place a “ mark

in the bracketed space, (

).

 

Any N/A’s must be explained in

the

notes

section of

this document.

 

 

Test

Equipment:

 

 

 

 

 

 

 

 

EQUIPMENT

 

 

CALIBRATION NUMBER

CALIBRATION

DUE DATE

Check one of the following:

Preventive Maintenance

or

Service Call/Installation/Upgrade

 

(complete section 2.1, then

proceed to 2.2)

(start at section 2.2}

2.1 PREVENTIVE MAINTENANCE

2.1.1CONSOLE INSPECTION

2.1.1.1 O Tubing and Manifolds 2.1.1.2 L] Filter 2.1.1.3 L] CPC fittings

2.12CART INSPECTION

Is there

an

Accurus

Cart?

L]

No,

Skip section 2.1.2.

LJ

Yes. Complete

section

2.1.2

2.1.2.1

L]

Console

Secured

to

Cart

2.1.2.2 [Lower

Compartment

Door

Opens & Closes

2.1.3CPU BATTERY REPLACEMENT

2.1.3.1 LJCPU PCB battery replacement has been completed in less than two (2) years.

2.14

LCD BACKLIGHT REPLACEMENT - HOT

CATHODE

ONLY

 

 

 

 

 

Is system below s/n IACL001377? O

Ne.

Skip section

2.1.4

O Yes.

Complete

section

2.1.4

 

214.3 CILCD backlight lamps or LCD

replacement has been

completed in

less than

three (3)

years.

21.5 FRONT PANEL TOUCHSCREEN 2.1.5.2] Inspect and clean the front panel Touchscreen.

2.1.6PRESSURE SOURCE TEST

2.1.6.3 No audible pressure leaks are present.

2.1.6.4

Is pressure source adjustable between

90-120

psi?

CT No

Skip section 2.1.6

[Yes Complete

section

2.1.6

2.1.6.7

Record low pressure threshold +

 

psi (78 - 83)

2.17SYSTEM COOLING

21.710] Verify Cooling Fan Operation.

2.1.8ILLUMINATOR “BULB FAILURE”

2.1.8.4DO verity “Bulb Failure” Message.

2.1.8.5 (install new bulbs. 2.1.8.6 C1 Bulb is properly aligned.

2.2SERVICE FUNCTIONAL TEST

2.2.1LATCH SEAL REPLACEMENT / ACCURUS CONSOLE INSPECTION

2.2.1.2Latch Seal

2.2.1.3 Lİ

Hub Roller

CI

Footswitch

Power

Cord.

2.2.1.4

Connectors

 

Receptacles

O Covers

and panels

2.2.1.5 Fan Filter

2.2.2SYSTEM POWER ON TEST

222.60] System successfully completed boot-up sequence.

2.2.2.7] Global arrows keys, Standby Switch and cassette eject are backlit.

2.2.3SYSTEM REL & ERROR LOG

2.2.3.1 Record system software REL,

2.2.3.2 XS4 PLATFORM ONLY İİ Record any listed Error/Fault numbers 2.2.3.3 L] Clear error log.

Form ITCDOC-000255, Version 1.0, Page 1 of 4

Parent Document: ITCSOP-000578

Supercedes: S0578-01

T

E982T22S0+

эчеочзтеен

39

PEF:60

80

90

REY

ACCEPTABLE RANGE
323 - 357 KHz
38.9 - 42.5 Vrms

Data Sheet for Accurus® 300VS, 400VS, 600DS, 800CS & Cart

Work Order No.

2.2.4DIATHERMY TEST

2.2.4.1DIATHERMY LOAD VERIFICATION Cautery Test Load

2.2.4.32) Audible tone when footpedal is pressed.

ohm (75.0 +/- 4.0)

2.2.4.4 - 2.2.4.5 Measure and record the following: MEASUREMENT

Frequency

Scope Amplitude

2.2.4.6 LJ With Diathermy power set to 0, the amplitude is < 0.7 Vrms.

2.2.4.7] Diathermy linearity increases from 2.8 - 4.24 with each increment of 10.

2.2.4.8 Voltage Feedback Reading

Vdc (3.96 - 4.04)

2.2.5PRESSURE REGULATORS

2.2.5.1 Measure and record the following:

Location

Test Port

Acceptable Range

Press/Vac Manifold

ΤΡΙ

82 — 93 Psi (5-Bar Pressure Select Is N/A)

RECORD MEASURED VALUE

KHz

Vrms

Record Measured Value

+pst

Press/Vac

Manifold

TP2

27-33

Psi

+

psi

Press/Vac

Manifold

TP3

54 — 60

Psi

+

psi

2.1.5.2 LI] TP3 <62 psi with T2 clamped.

2.2.6FOOTPEDAL TEST

22610 Verify faotpedal icon changes correspondingly to footpedal treadle and switches.

2.2.7GLOBAL DEFAULTS AND KEY TEST

2.2.7.3UU Verify Global Display Default Values

2.2.8CASSETTE INSERT EJECT

2.2.8.11] Verify Cassette Loading Sequence

2.2.8.21) Verify Cassette Unloading Sequence

2.2.9TOUCH SCREEN TEST

2.2.9.2] Verify Touchscreen icons and handpiece LEDs function.

2.2.10EXTRUDE (VACUUM) MODE TEST 2.2.10.3 Measure and record the following:

MAX LIMIT PRESET

ACCEPTABLE

RANGE

RECORD MEASURED VALUE

200

191 — 209

-

mmHg

400

387—413

-

mmHg

600

583—617

-

mmHg

 

(5-BAR Pressure

select is

N/A)

 

2.2.11REFLUX TEST

22.11.10 Verify aspiration #1 Reflux pincher, hammer sequence.

221120 Verify aspiration #2 Reflux pincher, hammer sequence.

2.2.12FLUID LEVEL SENSOR TEST

2.2.12.3 0 Verify Sensor Operation as stated in Table 3.

2.2.13CASSETTE SENSOR RECOGNITION - MODEL 800CS ONLY

 

2.2.13.3 Lİ

Using the

sensor test

tool, verify cassette sensor

recognition.

 

 

2.2.14

VGFI TEST

 

2.2.14.2 Measure

and

record

the following:

 

 

 

 

 

YGFI

PRESET

 

 

 

 

ACCEPTABLE

RANGE

RECORD MEASURED VALUE

 

 

022

 

 

 

 

18—

26

 

+

mmHg

 

 

070

 

 

 

 

65-75

 

+

mmHg

 

 

120

 

 

 

 

113—

127

 

+

mmHg

 

2.2.143[]

 

verify Actual

Display

Stability.

 

 

 

 

 

2.2.15

BACKUP

PRESSURE

 

 

 

 

 

 

 

 

 

 

2.2.15.2 Measure and record backup

pressure

+

 

 

mmHg

(25-35)

Form

ITCDOC-000255,

Version

1.0,

Page 2 of 4

 

Parent Document:

ITCSOP-000578

 

Supercedes: S0578-01

EsBcTLESO+F

эчеэочзтеен 39

8€+:60 80 90 REH

Data Sheet for Accurus® 300VS, 400VS, 600DS, 800CS & Cart

Work Order No.

2.2.16ILLUMINATOR TEST

2.2.16.4D) Verify ILLUM 1 operation.

2.2.16,5 C Verify ILLUM 2 operation.

2.2.16.9 0 Bulb is properly aligned.

2.2.17PROBE DRIVE TESTS XS2 & XS3 PLATFORMS

2.2.17.5/2.2.17.12 MODELS EXCEPT 300VSVerify

output>

to 23

psi

+

psi.

2.2.17.7/2.2.17.14

MODELS

EXCEPT

300VS

Verify

output> to

23

psi

+.

psi.

 

2.2.17.18/2.2.18.23

MODEL

300VS ONLY

ACCURUS Verify

output

is >29

psi+

psi.

2.2.17.10

XS4 PROBE

DRIVE TESTS

 

 

 

 

 

 

 

 

 

2.2.17.25/

2.2.17.34 Verify

"CLOSE"

output

>29 psi +

psi

“OPEN”

output

$9 psi +___psi.

2.2.17.27/2.2.17.36 Verify "CLOSE" output >29 psi +.

gsi."OPEN"

output

<15

psi +___psi

2.2.18 SCISSORS

DRIVE

 

 

 

 

 

 

 

 

 

 

 

2.2.18.2 Verify pressure

is 20

+/- 2 +,

psi

(18-22).2.2.18.4 L]

MultiCut output

pulses.

22.19MPC SCISSORS — X83 &XS4 ONLY

2.2.19.1 MPC LOAD VERIFICATION: MPC Test Load

 

 

 

ohm (4.0 +/- 0.2)

2.2.19.5

Verify

peak

voltage

Vde

(10.0

11.6)

2.2.19.6

Verify

hold

voltage

Vde

(1.0

1.4)

2.2.19.7

Verify

voltage with hold disabled

 

 

 

Vde (0— 0.1)

2.2.20VFC TEST— DOES NOT APPLY TO 300VS

2.2.20.2 VFC INJECT: Measure and record the following:

MAX

LIMIT

ACCEPTABLE

RANGE

 

RECORD MEASURED

VALUE

40

 

38-42

 

 

 

+

psi

 

80

77

— 83 (5-BAR Pressure

select

is N/A)

+

psi

 

2.2.20.3

LI

Verify pedal

control of pressure 0

-

80 psi.

(0-70 for

5-BAR

Pressure select)

 

2.2.20.5

VFC

EXTRACT:

Measure and record the following:

 

 

 

MAX

LIMIT

ACCEPTABLE

RANGE

 

RECORD MEASURED

VALUE

600

583 -

617 (5-BAR Pressure select

is N/A)

| -

mmHg

 

400

 

387

-

413

 

 

 

-

mmHg

 

 

200

 

191

209

 

 

 

-

mmHg

 

2.2.20.6 L] Verify pedal control of vacuum.

XS3 & XS4 PLATFORM ONLY 2.2.20.100] Verify pressure and vacuum readings on LCD (actual). XS3 & XS4 PLATFORM ONLY 2.2.20,110] Verify pressure leakage is < 10 psi.

2.2.21FRAGMATOME TEST

2.2.21.1 Is the customer's handpiece available? Ne Skip steps 2.2.21.2 - 2.2.21.13

Yes Complete section 2.2.21 and record handpiece s/n

2.2.21.2D) Customer's Handpiece tunes. 2.2.21.4, 2.2.21.5 Fragmatome Momentary functions

 

2.2.21.7, 2.2.21.8 LİFragmatome

Proportional

functions 2.2,21.10

O

Fragmatome Fixed functions

 

 

2.2.21.12,

2.2.21.13 [1] Fragmatome Pulse

functions

 

 

 

 

 

 

 

 

2.2.22,

REMOTE

CONTROL

 

 

 

 

 

 

 

 

 

 

 

 

 

Does the Accurus® have a remote

control?)

No Skip

section

2.2.22

O

ves

Complete

section 2.2,22

 

 

22.22.10

Verify that

the Remote

Control

is

capable of selecting

modes

and

changing

preset values:

_

STEPS

2.2.23 — 2.2.25 APPLY

ONLY TO

MODEL

600DS or

800CS

 

 

 

 

 

 

x

2.2.23

VA MAX

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.2.23.6 Record measured vacuum

reading-

 

 

mmHg

(387

- 413)

 

 

 

 

 

2.2.24.CONTINUOUS REFLUX

2.2.24.1Verify that continuous reflux pincher momentarily retracts.

2.2.24.20] Verify that continuous reflux pincher remains retracted.

Form ITCDOC-000255, Version 1.0, Page 3 of 4

Parent Document: ITCSOP-000578

Supercedes: $0578-01

g'd

E982T2250+

эчеочЗзтеен

39

РЕФ:

БО

BO

90

REY

Data Sheet for Accurus® 300VS, 400VS, 600DS, 800CS & Cart

Work Order No. _____

2.2.25PHACO

2.2.25.2Ú Verify that Phaco Handpiece tunes

Record

Handpiece

s/n

Is

this

a

customer

handpiece?

[]

Ne

O

Yes

Record

Handpiece

s/n

Is

this

2

customer

handpiece?

[]

No

LU)

Yes

Record

Handpiece

s/n

Is

this

a

customer

handpiece?

D]

No

CJ

Yes

2.2.25.4I Verify that detents I and 2 are felt.

2.2.25.5[|] Verify that U/S power is linear while in footswitch position 3.

222570 Verify pulsed U/S power and linear vacuum. 2.2.25.8 JI Verify that U/S pulse rate varies from 1 to 15 pps.

2.2.26 ACCURUS CART VISUAL INSPECTION (if applicable)

Is there an Accurus Cart? [7] No. STP is Completed L] Yes Complete Section 2.2.26

2.2.26.1

(7

Casters

2.2.26.2 C]

Cables

2.2.26.3

]

Drawers

2.2.26.4

0]

Mayo

Tray

2.2.26.5(]

VGFVAVGFI Pole

2.2.26.6

O]

Writing Tablet

22.26.71)

Cable

Cleats

 

 

 

 

 

Does the Accurus Cart have a Manual IV Pole?

C] Yes. STP is Completed

O No. Continue to step 2:2.27

2.2.27ACCURUSCARTFUNCTIONAL TESTS

2.2.27.1

[]

Check

here if ceiling height

prohibits

maximum

of

115

cmH,O.

Record Max allowed.

VV Pole

Max

(113

to 117)

cmH;0.

I/V Pole

Min

(5

to 9)

<mH,O

2.2.28ACCURUS CART REMOTE CONTROL VERIFICATION

Does

the Accurus® have a remote

control?

No. STP

Complete

Lİ Yes.

Complete section 2.2.28

2.2.28.1 1] Verify Remote Control

raises and

lowers

I/V Pole.

 

 

 

 

Test Performed

By (PRINT)

 

Signature

 

 

 

 

Date

Was

system tested using 5-BAR

pressure

selection?

O

ne

Yes-

N/A

listed for all pressures above 70psi and vacuums

 

 

 

 

 

 

 

exceeding

450

mmHg.

Notes:

Form ITCDOC-000255, Version 1.0, Page 4 of 4

Parent Document: ITCSOP-000578

Supercedes: 50578-01

pod

E982T2250+

еэчеэчзтеен 39

Ebb:60 80 90 REY

Alcon

Irvine Technology

Center - Service Test Procedure

Page 1 of 18

 

SOP Number:

Version:

 

Effective Date:

ITCSOP-000578

CURRENT,

1.0

12-May-2005

Title: |

STPAccurus 300VS,

400VS, 600DS, 800CS and Cart

 

1 PURPOSE

1.1

This

document

 

defines

the

step-by-step

mechanical

and

electrical

tests

to

verify

that

the

 

Accurus®

is

performing

to

specifications.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.2

Preventive

Maintenance

(PM),

located

in

section

2.1

of

this

procedure,

shall

be

used

for

all

 

scheduled

 

Preventive

Maintenance.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.3

The

Service

Functional

Test,

found

 

in

sections

2.2

to

(but

not

including)

section

3.0

of

this

 

procedure,

shall

 

be

used

to

ensure

the

Accurus®

meets

specifications after

installation

of

the

 

unit, after repairs are made,

and

 

after

routine

Preventive

Maintenance.

Form

S0578-01,

Data

 

Sheet

for

Accurus®

300VS,

400VS,

 

600DS,

 

800CS

& Cart,

must

be filled

out

completely,

dated

 

and signed

at

the

time

the

test

or

operation

is

performed.

 

 

 

 

 

 

 

 

 

 

 

 

 

14

The

check

mark

in

box

WI

symbol

 

indicates

that

an

entry

is

required

on

Form 50578-01, Data

 

Sheet

for

 

Accurus®

300VS,

 

400VS,

 

600DS,

800CS

&

Cart.

 

The

entry

can be either a Y, or

 

requested

 

data.

 

A

N/A,

to be used when a

 

verification

step

could

not

be

completed,

 

MUST

be

 

noted

 

at

the

end

 

of

Form

S0578-01,

 

Data

Sheet

for

Accurus®

300VS,

400VS,

600DS,

800CS

&

 

Cart.

 

If

a

PM

 

is

performed

in

conjunction

with

a

repair

 

or

upgrade

only

the

Preventive

 

Maintenance

box

is

to

be

checked

on

Form

 

S0578-01,

Data

Sheet

for Accurus®

300VS,

400VS,

 

600DS,

800CS

&

Cart.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.5

REL

2.07

 

and

above

allows

for a 5-BAR

pressure

selection.

If system

is

being tested using 5-

 

BAR

 

pressure

selection,

tests

for

pressures

above

70

psi

and

vacuums

exceeding

450

mmHg

 

should

be

 

listed

 

as N/A.

A

checkbox

entry

that

5-BAR

pressure

selection

was

utilized

should

be

made on Form S0578-01, Data Sheet for Accurus® 300VS, 400VS, 600DS, 800CS & Cart.

1.6

The

recording

of error

log

at step

2.2.3.2 should be

performed

prior to

any

software

upgrades

or

 

CPU

battery

changes.

 

 

 

 

 

 

 

 

 

 

 

2 PROCEDURE

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE:

Refer to

the

Service

Manual

for

recommended

tools

and

equipment.

If any

checks fail

to

meet

specifications,

refer to

the

Accurus®

Service Manual for

adjustment

or calibration procedures.

Repair

or

replace

as

needed.

 

 

 

 

 

 

 

 

 

 

 

2.1PREVENTIVE MAINTENANCE

NOTE:

The Accurus® Preventative

Maintenance cycle

per

SOP-0579, SMP, Accurus ®, is as

follows:

 

 

 

 

 

 

 

Models

without

Cataract

Surgery

modes: A

minimum

of

once every 12 months. Models with

Cataract

Surgery

modes:

A minimum of twice

every twelve

months.

2.1.1CONSOLE INSPECTION

2.1.1.1

MW

Remove

top

cover.

Inspect pneumatic

lines

and

manifolds

for

signs

of saline

and

 

debris. Clean or

replace

as

required.

 

 

 

 

 

 

 

 

21.12

M1

Inspect

water trap

filter

for

signs

of

moisture

and/or discoloration.

Replace

as

 

required.

 

 

 

 

 

 

 

 

 

 

 

 

 

2.1.1.3

MM

Replace

top

CPC connector.

Inspect

other

CPC

connectors

and

connector O rings

for signs of wear. Replace connector as required.

| Status: | Effective

|

Supersedes: | SOP-0578

gcd

E982T2250+

SJEDUJTESH 39

Pbb:6O BO 90 REH

Alcon

irvine Technology

Center - Service Test Procedure

Page 2 of 18

 

SOP Number:

Version:

Effective Date:

ITCSOP-000578

CURRENT, 1.0

12-May-2005

Title: |

STP - Accurus 300VS,

400VS, 600DS, 800CS and Cart

 

2.12VISUAL INSPECTION OF ACCURUS CART (if applicable)

NOTE:

If

system

does NOT have a Cart,

indicate

on

Form

S0578-01,

Data

Sheet for

Accurus®

300VS,

400VS,

600DS, 800CS

& Cart,

and

proceed

to

step

2.1.3.

 

2.1.2.1

M

Check to

ensure

that Accurus®

console

is

attached

to

the

cart

with

two knob

 

clamps.

 

 

 

 

 

 

 

 

 

 

 

2.1.2.2 M Check to ensure that the lower compartment door opens and closes.

2.13REPLACEMENT OF CPU BATTERY

2.1.3.1

M

Replace

the

battery

per

instruction

in

the

Accurus

Service

Manual

unless

the

 

 

system

date

of

manufacture

or

the

CPU

battery

replacement

has

occurred

in

the

 

 

previous

two

years,

then,

skip

to

step

2.1.4.

 

 

 

 

 

 

 

 

 

 

2.14

REPLACEMENT

OF

LCD

BACKLIGHT

LAMPS

(Hot

Cathode

LCDs

only)

 

 

2.1.4.1

M

Does

the

system

use

a

LCD

display

with

hot-cathode

lamps

(systems

s/n

 

 

IACL001377

and

below)?

If

yes, complete

step

2.1.4. If no, skip

step

2.1.4.

 

 

2.142

M

Has

it been

three

years

or

more

since

the

LCD

backlight

lamps

or

the

LCD

 

 

Display

have

been

replaced?

If yes, proceed

to

next step. If no, proceed

to

step

2.1.5.

2.1.4.3 Replace the backlight lamps per instruction in the Accurus Service Manual (must be less than three years from date of this PM).

2.1.5INSPECT AND CLEAN THE FRONT PANEL TOUCHSCREEN

2.1.5.1

Look for cracks or other

physical damage to the

front

panel

touchscreen, replace if

 

necessary.

 

 

 

 

2.1.5.2.

M1 Inspect and clean, as

necessary, touchscreen

with

damp

cloth or non-ammonia

 

based window cleaner.

 

 

 

 

2.1.6PRESSURE SOURCE TEST

2.1.6.1

Plug unit

into

standard

outlet

and

connect

footswitch

to

rear

of

unit.

 

 

2.1.6.2

Set

pressure source Regulator

to

120

psi,

or the

highest

pressure

available

below

120

 

psi.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.1.6.3

 

Verify

there

are

no

leaks by confirming

no audible

hissing

is

present.

 

 

2.1.64

M

Ifa regulated,

adjustable

90-120

psi pressure

source

is not

available,

indicate

so

on Form S0578-01, Data Sheet for Accurus® 300VS, 400VS, 600DS, 800CS & Cart, and proceed to step 2.1.7.

2.1.6.5Power On system.

2.1.6.6

Decrease

air

source Pressure

Regulator

(to

about

60-70 psi) until unit displays

 

System Advisory “Please adjust supply

source

pressure from 90 to 120 psi”. If

the

 

advisory

“Press

5 bar

for low

pressure

operation

or press Standard and

increase

 

air

pressure”

appears,

select “Standard”.

 

 

 

 

 

2.1.6.7

WM

Using

blue

metric

tubing,

connect

pressure

meter to 85 psi test port on

 

pressure/vacuum manifold assy. Slowly increase pressure until System Advisory

is

 

removed

from

screen. Verify low pressure threshold

is approximately 78-83

psi.

 

2.1.6.8

Set

air source

Pressure

Regulator back

to

90 to

120

psi.

 

 

| Status: | Effective

 

 

 

 

 

 

|

Supersedes: |

SOP-0578

 

 

gcd

E982T2250+

ΘΩΒΟΙΗΣΤΕΘΗ 39

eS+:60 80 90 Рен

Alcon

Irvine Technology

Center - Service Test Procedure

Page 3 of 18

 

SOP Number:

Version:

 

Effective Date:

ITCSOP-000578

CURRENT,

1.0

12-May-2005

Title: |

STP - Accurus 300VS,

400VS, 600DS, 800CS and Cart

 

 

2.1.7SYSTEM COOLING

2.1.7.1MM Verify cooling fan operation:

2.1.7.1.1

System fan

Flow

IN

2.1.7.1.2

U/S Fan

Flow

IN

2.1.8ILLUMINATOR “BULB FAILURE” INDICATOR

2.1.8.1 | Ensure system is in the Vitreoretinal Surgical setup screen.

2.1.8.2Press ILLUMINATOR RELEASE button to eject drawer.

2.183

Remove

illuminator

bulbs

and

insert

I!luminator drawer back into the Accurus.

2.1.8.4

M

Verify

 

that

message

appears on

screen

“Bulb

Failure detected....” and the Bulb

 

Out

icon

is

shown

at

the

illuminator

global

display.

 

 

 

2.1.8.5

HM

Install

2

new

bulbs,

using a

tissue

or cloth, and

taking care

not to

touch

bulbs with

 

fingers.

 

 

 

 

 

 

 

 

 

 

 

 

 

2.1.8.6

M

Place

a sheet of white paper

behind

bulbs

and ensure

each

bulb

js properly

 

aligned per

Figure

2.

 

 

 

 

 

 

 

 

 

Figure 2 - Bulb Alignment

22 SERVICE FUNCTIONAL TEST

22.1LATCH SEAL REPLACEMENT/ ACCURUS CONSOLE INSPECTION

2.2.1.1 Ensure system to be tested is turned OFF before proceeding.

22.121 Replace latch seal.

22,13 M Check that all four hub rollers spin freely on the Receiver Mechanism. Verify hub

 

roller is secure. Inspect Receiver Mechanism

optical

sensor areas

for BSS

residue.

 

Clean as required. Inspect footswitch and power cord

for

exposed

wires

or

damage.

 

Repair or replace as necessary.

 

 

 

 

 

 

 

22.14

Ы

Visually inspect all connectors, receptacles, panels

and

covers

for damage.

Repair

 

or

replace

as necessary.

 

 

 

 

 

 

 

2.2.1.5

ΠΜ

Inspect

intake fan filter on rear of unit and

clean if

necessary.

Record

service call

 

on

history

log.

 

 

 

 

 

 

 

222SYSTEM POWER ON TEST

2.2.2.1Plug unit into standard outlet.

2.2.2.2 Connect Footswitch to rear of unit and IV Pole Cable (if applicable).

| Status: | Effective

Supersedes: ] SOP-0578

L*d

EI8ZIZAZSOb

еэчеочзтеен

39

ES+:60

80

90

ReW

Loading...
+ 16 hidden pages