Alcon Accurus 300VS, Accurus 400VS, Accurus 600DS, Accurus 800CS, Accurus Cart User manual

Data
Sheet
for
Accurus®
300VS,
400VS, 600DS,
800CS
&
Cart
Customer
Model/Platform
NOTE:
Test
Check
2.1.1
Throughout Any
Equipment:
one
of
the
CONSOLE
2.1.1.1 O Tubing
2.12
2.1.3
2.14
21.5
2.1.6
2.17
2.1.8
2.2.1
CART
Is
there
2.1.2.1
CPU
2.1.3.1
system
214.3
FRONT
PRESSURE
2.1.6.3
2.1.6.4
CT
No
2.1.6.7 SYSTEM
21.710] ILLUMINATOR
2.1.8.4
2.1.8.5
LATCH
2.2.1.2
2.2.1.3
2.2.1.4
2.2.1.5
2.2.2
2.2.3
SYSTEM
222.60]
2.2.2.7]
SYSTEM
2.2.3.1
2.2.3.2
this
N/A’s
must
EQUIPMENT
following:
INSPECTION
INSPECTION
an
Accurus
L]
Console
BATTERY
LJCPU
BACKLIGHT
below
CILCD
PANEL
SOURCE
No
Is
pressure
Skip
section
Record
COOLING
Verify
DO
verity
SEAL
Latch
Hub
Connectors
Fan
POWER
System
Global
REL & ERROR
Record
XS4
PLATFORM
Data
Sheet
be
explained
and
Manifolds
Cart?
Secured
REPLACEMENT
PCB
battery
REPLACEMENT
IACL001377?
backlight
TOUCHSCREEN
TEST
audible
pressure
source
2.1.6
low
pressure
Cooling
“BULB
“Bulb
(install
REPLACEMENT
Seal
Roller
Filter
ON
TEST
successfully arrows
system
software
either
fill
in
the
Preventive
(complete
2.1.1.2
L]
No,
Skip
to
Cart
replacement
lamps
or
2.1.5.2]
leaks
adjustable
[Yes
threshold
Fan
Operation.
FAILURE”
Failure”
ONLY
CI
keys,
Message.
bulbs.
/
ACCURUS
Footswitch
Receptacles
completed
Standby
REL,
İİ
in
the
requested
notes
section
Maintenance
section
2.1,
L]
Filter 2.1.1.3
section
2.1.2.2
has
-
O
Skip
replacement
are
present.
between
Complete
+
2.1.8.6
boot-up
Switch
Record
any
Work
of
CALIBRATION
or
then
proceed
PREVENTIVE
2.1.2. [Lower
been
completed
CATHODE
section
Inspect
90-120
section
C1
Bulb
CONSOLE
sequence.
and
listed
Order
data
or
place a
this
document.
to
2.2)
MAINTENANCE
L]
CPC
LJ
Yes.
Complete
Compartment
in
less
ONLY
2.1.4
been
completed
and
clean
the
psi?
2.1.6 psi
(78 - 83)
is
properly
SERVICE
INSPECTION
Power
Cord.
O
Covers
cassette
eject
Error/Fault
S/N
mark
NUMBER
fittings
section
Door
than
two
O
Yes.
in
front
panel
aligned.
FUNCTIONAL
panels
are
backlit.
numbers
2.2.3.3
in
the
bracketed
Service
(start
Call/Installation/Upgrade
at
section
2.1.2
Opens & Closes
(2)
years.
Complete
less
section
than
three
Touchscreen.
TEST
L]
Clear
space,
CALIBRATION
2.2}
2.1.4
years.
error
log.
(
).
DATE
Form
ITCDOC-000255,
T
Version
1.0,
Page 1 of
4
E982T22S0+
Parent
Document:
ITCSOP-000578
эчеочзтеен
39
Supercedes:
PEF:60
S0578-01
80
90
REY
2.2.4
2.2.5
2.2.6
2.2.7
2.2.8
2.2.9
2.2.10
2.2.11
2.2.12
2.2.13
2.2.14
2.2.15
DIATHERMY
2.2.4.1
2.2.4.3
2.2.4.4 - 2.2.4.5 MEASUREMENT
Frequency
Scope
2.2.4.6
2.2.4.7]
2.2.4.8
PRESSURE
2.2.5.1
Location
Press/Vac
Press/Vac Press/Vac
Manifold
Manifold Manifold
2.1.5.2
FOOTPEDAL
22610
GLOBAL
2.2.7.3
CASSETTE
2.2.8.1
2.2.8.2 TOUCH
2.2.9.2] EXTRUDE
2.2.10.3
MAX
LIMIT
200 400 600
REFLUX
22.11.10 221120
FLUID
2.2.12.3 0 Verify CASSETTE
2.2.13.3 VGFI
TEST
YGFI
2.2.143[]
BACKUP
2.2.15.2
Data
TEST
DIATHERMY
2)
Audible
Measure
Amplitude
LJ
With
Diathermy
Diathermy
Voltage
LI]
Feedback
REGULATORS
Measure
TP3
<62
TEST
Verify
DEFAULTS
UU
Verify
INSERT
1]
Verify
1)
Verify
SCREEN
Verify
(VACUUM)
Measure
PRESET
TEST
Verify Verify
LEVEL
SENSOR
SENSOR
Using
PRESET
022
070
120
verify
PRESSURE
Measure
2.2.14.2
Sheet
and
faotpedal
Global
Cassette
Cassette
TEST
LOAD
tone
when
and
linearity
Reading
record
Test
Port
ΤΡΙ
TP2
TP3
psi
with
AND
Display
EJECT
Loading
Unloading
for
VERIFICATION
record
power
the
T2
icon
KEY
Touchscreen
MODE
and
record
the
aspiration aspiration
Sensor
the
Actual
and
#1 #2
TEST
Operation
RECOGNITION
sensor
test
Measure
Display
record
backup
Accurus®
footpedal
is
the
following:
set
to
0,
increases
from
following:
clamped.
changes
TEST
Default
icons
TEST
following:
Reflux
Reflux
Values
Sequence
Sequence
and
handpiece
ACCEPTABLE
(5-BAR
pincher, pincher,
as
stated
-
MODEL
tool,
verify
and
record
Stability.
pressure
300VS,
Cautery
pressed.
ACCEPTABLE
the
amplitude
2.8 - 4.24
Test
323 - 357
38.9 - 42.5
with
Vdc
Acceptable
82 93
Psi
(5-Bar
correspondingly
LEDs
191 209
387 413
583 617
Pressure
cassette
the
ACCEPTABLE
+
select
hammer hammer
in
Table
800CS
sensor
following:
18—
65-75
113 127
3.
400VS,
Load
RANGE KHz Vrms
is < 0.7
Vrms.
each
increment
(3.96 - 4.04)
Range
Pressure
27-33 54 60
RANGE
sequence. sequence.
ONLY
Psi Psi
to
footpedal
function.
is
N/A)
recognition.
RANGE
26
600DS,
Select
treadle
of
Is
ohm
10.
N/A)
and
RECORD
-
-
-
RECORD
+
+
+
mmHg
800CS
(75.0
+/-
RECORD
switches.
mmHg
mmHg
mmHg
MEASURED
mmHg
mmHg mmHg
(25-35)
&
4.0)
MEASURED
Record
+
+ +
MEASURED
Cart
Work
KHz
Vrms
Measured
pst
psi psi
VALUE
VALUE
Order
VALUE
Value
No.
Form
ITCDOC-000255,
Version
1.0,
Page 2 of
EsBcTLESO+F
4
Parent
Document:
ITCSOP-000578
эчеэочзтеен
39
Supercedes:
8€+:60
S0578-01
80 90
REH
2.2.16
2.2.17
2.2.18
22.19
2.2.20
2.2.21
2.2.22,
STEPS
2.2.23
2.2.24.
Data
ILLUMINATOR
2.2.16.4
D)
Sheet
TEST
Verify
2.2.16,5 C Verify
PROBE
2.2.17.5/2.2.17.12
2.2.17.7/2.2.17.14
2.2.17.18/2.2.18.23
2.2.17.10
2.2.17.25/
2.2.17.27/2.2.17.36
SCISSORS
2.2.18.2
MPC
2.2.19.1
2.2.19.5
2.2.19.6
2.2.19.7 VFC
2.2.20.2
2.2.20.3
2.2.20.5
2.2.20.6
XS3 & XS4
XS3 & XS4 FRAGMATOME
2.2.21.1
2.2.21.2
2.2.21.7,
2.2.21.12, REMOTE
Does
22.22.10
2.2.23 2.2.25
VA
2.2.23.6
CONTINUOUS
2.2.24.1 Verify
2.2.24.2
DRIVE
Verify
SCISSORS X83
MPC
Verify
Verify
Verify
TEST DOES NOT
VFC
MAX
LIMIT 40 80
LI VFC
MAX
LIMIT
600
400
200
L]
Is
D)
2.2.21.8
the
Accurus®
MAX
Record
0]
TESTS
XS4
PROBE
2.2.17.34
DRIVE
pressure
LOAD
peak
hold
voltage
INJECT:
Verify
EXTRACT:
Verify
PLATFORM
PLATFORM
the
customer's
Yes
Complete
Customer's
LİFragmatome
2.2.21.13 CONTROL
Verify
APPLY
measured
REFLUX
Verify
MODELS
MODEL
pedal
pedal
TEST
have a remote
that
that
that
for
Accurus®
ILLUM 1 operation.
ILLUM 2 operation.
XS2 & XS3
MODELS
Verify
Verify
VERIFICATION:
DRIVE
is
20
&XS4
EXCEPT
EXCEPT 300VS
TESTS
"CLOSE"
"CLOSE"
+/- 2 +,
ONLY
ONLY ACCURUS
2.2.16.9 0 Bulb
PLATFORMS
300VS-
300VS
output
>29 psi
output
>29
psi
MPC
Test
voltage
voltage
with
hold
disabled
APPLY
Measure
77 83
control
Measure
583 - 617
control
ONLY
ONLY
handpiece
Handpiece
[1]
Fragmatome
the
ONLY
vacuum
continuous
continuous
TO
and
ACCEPTABLE
(5-BAR
of
pressure 0 -
and
ACCEPTABLE
(5-BAR
of
vacuum.
2.2.20.100]
2.2.20,110]
section
2.2.21 tunes.
Proportional
control?)
Remote
TO
Control
MODEL
reading
reflux
reflux
300VS
record
the
RANGE
38-42
Pressure
record
RANGE
Pressure
387 - 413
191 209
available? Ne
and
2.2.21.4,
Pulse
functions
No
is
600DS
-
pincher
pincher
300VS,
Verify
Verify
+
psi
+.
(18-22).2.2.18.4
Load
Vde
Vde
following:
select
is
80
psi.
the
following:
select
Verify
Verify
record
handpiece
2.2.21.5 Fragmatome
functions
Skip
capable
of
or
800CS
mmHg
momentarily
remains
400VS,
is
properly
output > to
output > to
N/A)
(0-70
is
N/A) | -
pressure
pressure
Skip
2.2,21.10 O Fragmatome
section
selecting
23
Verify output
psi
“OPEN”
gsi."OPEN"
L]
(10.0 11.6)
(1.0
1.4)
for
5-BAR
RECORD
-
and
leakage
steps
2.2.21.2 - 2.2.21.13
s/n
2.2.22 O ves
(387 - 413)
retracts.
retracted.
600DS,
aligned.
23
psi
+
psi
+.
is
>29
output
output
MultiCut
Vde
(0—
RECORD + +
Pressure
mmHg
-
vacuum
modes
mmHg
mmHg
is < 10
and
psi.
psi.
psi+
$9
<15
output
ohm
0.1)
MEASURED psi psi
select)
MEASURED
readings
psi.
Momentary
Complete
changing
800CS
psi.
psi
+___psi.
psi
+___psi
pulses.
(4.0
+/-
VALUE
VALUE
on
LCD
functions
Fixed
functions
section
preset
&
0.2)
(actual).
2.2,22
values:
Cart
Work
_
Order
No.
x
Form
ITCDOC-000255,
g'd
Version
1.0,
Page 3 of
E982T2250+
4
Parent
Document:
ITCSOP-000578
эчеочЗзтеен
39
Supercedes:
РЕФ:
БО
$0578-01
BO
90
REY
2.2.25
2.2.26
2.2.27
2.2.28
Data
PHACO
2.2.25.2 Ú Verify
Record Record Record
2.2.25.4 I Verify
2.2.25.5 222570
2.2.25.8
ACCURUS
Is
2.2.26.1
2.2.26.4
Handpiece Handpiece Handpiece
there
an
(7
0]
[|]
22.26.71)
Does
the
Accurus
C]
Yes.
STP
ACCURUS
2.2.27.1 VV
ACCURUS
Does
2.2.28.1
Pole
the
[]
Max
Accurus®
1]
Sheet
that
s/n s/n s/n
that
Verify
Verify
pulsed
JI
Verify
that
CART
VISUAL
Accurus
CART
CART
Cart?
Casters
Mayo
Tray
Cable
Cleats
Cart
is
Completed
FUNCTIONAL
Check
here
(113
to
REMOTE CONTROL
have a remote
Verify
Remote
that
117)
for
Accurus®
Phaco
Handpiece
detents I and
U/S
power
is
U/S
power
U/S
pulse
rate
INSPECTION
[7]
No.
STP
2.2.26.2
2.2.26.5(]
have a Manual
O
No.
TESTS
if
ceiling
height
control?
Control
raises
tunes
2 are
felt.
linear
while
and
linear
varies
from 1 to
(if
applicable)
is
Completed
C]
Cables
VGFVAVGFI
IV
Pole?
Continue
prohibits
cmH;0.
VERIFICATION
and
lowers
300VS,
Is Is Is
in
footswitch
vacuum.
L]
Pole
to
step
maximum
I/V
Pole
No.
STP
I/V
400VS, 600DS,
this a customer this 2 customer
this a customer
position
15
pps.
Yes
Complete
2.2.26.3 ] Drawers
2.2.26.6
2:2.27
of
115
Min
Complete
Pole.
Section
O]
cmH,O.
(5 to 9)
handpiece? handpiece? handpiece?
3.
2.2.26
Writing
Record
Yes.
Complete
800CS & Cart
[]
Ne O Yes
[]
No
LU)
Yes
D]
No
CJ
Yes
Tablet
Max
allowed.
<mH,O
section
2.2.28
Work
Order
No.
_____
Test
Performed
Notes:
Was
By
(PRINT)
system
tested
using
5-BAR
pressure
selection? O ne
Signature Date
Yes-
N/A
listed
for
all
pressures
exceeding
450
mmHg.
above
70psi and
vacuums
Form
ITCDOC-000255,
pod
Version
1.0,
Page 4 of
E982T2250+
4
Parent
Document:
ITCSOP-000578
еэчеэчзтеен
39
Supercedes:
Ebb:60
50578-01
80 90
REY
Alcon
Title: | STP-
1
SOP
ITCSOP-000578
PURPOSE
1.1
1.2
1.3
14
1.5
1.6
Irvine
Technology
Number:
Accurus
This
document
Accurus® Preventive
scheduled The
Service
procedure,
after
unit,
Sheet
for
and
signed
The
check
Sheet
for requested noted
at
Cart.
If a PM
Maintenance
600DS,
REL
2.07
BAR
pressure
should
made
The CPU
be
on
recording
battery
Center - Service
300VS,
is
Maintenance
Preventive
shall
repairs
Accurus®
at mark
Accurus®
data. A N/A,
the
800CS & Cart.
and
listed
Form
400VS, 600DS,
defines
performing
Maintenance.
Functional
be
used
made,
are
300VS,
the
time
the
in
box
300VS,
end
of
Form
is
performed
box
is
to
be
above
selection,
S0578-01,
of
allows
as
N/A. A checkbox
error
log
changes.
CURRENT,
800CS
the
step-by-step
to
specifications.
(PM),
Test,
to
test
WI
to
S0578-01,
tests
Data
located
found
ensure
and
400VS,
or
operation
symbol
400VS,
be
used
in
checked
for a 5-BAR
for
Sheet
at
step
the
after
600DS,
indicates
when a verification
Data
conjunction
on
pressures
for
2.2.3.2
Test
Procedure
Version:
and
Cart
mechanical
in
section
in
sections
Accurus®
routine
800CS & Cart,
is
performed.
that
600DS,
Form
entry
800CS & Cart.
Sheet
S0578-01, Data
pressure
above
that
Accurus®
should
1.0
and
electrical
2.1
of
this
2.2
to
(but not
meets
Preventive
specifications
Maintenance.
must
an
entry
is
step
could
for
Accurus®
with a repair
Sheet
selection.
70
psi
and
5-BAR
be
pressure
300VS, 400VS,
performed
procedure,
including)
be
filled
required
The
entry
not
300VS,
or
upgrade
for
Accurus®
If
system
vacuums
selection
600DS,
prior
to
Effective
tests
on
be
400VS,
any
12-May-2005
to
verify
shall
be
section
after
installation
S0578-01,
Form
out
completely,
Form
50578-01,
can
be
completed,
600DS,
only
is
being
exceeding
was
utilized
800CS & Cart.
software
Page 1 of
Date:
that
used
for
3.0
of
of
Data
dated
Data
either a Y,
MUST
800CS
the
Preventive
300VS,
tested
450
400VS,
using
mmHg
should
upgrades
18
the
all
this
the
or
be
&
5-
be
or
2
PROCEDURE
NOTE: meet
Repair
2.1
Refer
specifications,
or
replace
PREVENTIVE
NOTE:
follows:
Models
Cataract
2.1.1
2.1.1.1
21.12
2.1.1.3
|
Status: | Effective
to
the
as
The
Accurus®
without
Surgery
CONSOLE
MW debris.
M1
required. MM
for
Service
refer
needed.
to
Manual
the
Accurus®
MAINTENANCE
Preventative
Cataract
Surgery
modes: A minimum
INSPECTION
Remove
Inspect
Replace
signs
top
Clean
water
top
of
wear.
cover.
or
replace
CPC
for
recommended
Service
Maintenance
modes: A minimum
of
twice
trap
Inspect
filter
pneumatic
as
required.
for
connector.
Replace
connector
tools
Manual
for
cycle
per
of
every
twelve
lines
signs
of
moisture
Inspect
other
as
required.
|
Supersedes: | SOP-0578
and
equipment.
adjustment
SOP-0579, SMP,
once
every
months.
and
manifolds
and/or
CPC
connectors
If
any
checks
or
calibration
12
months.
for
signs
discoloration.
and
procedures.
Accurus
Models
of
Replace
connector O rings
fail
®,
saline
to
is
as
with
and
as
gcd
E982T2250+
SJEDUJTESH
39
Pbb:6O
BO 90
REH
Alcon
Title: | STP - Accurus
irvine
SOP
ITCSOP-000578
Technology
Number:
2.12
VISUAL
NOTE:
Accurus®
2.1.2.1
2.1.2.2 M Check
2.13
REPLACEMENT
2.1.3.1 M Replace
2.14
2.1.5
REPLACEMENT
2.1.4.1
2.142
2.1.4.3
INSPECT
2.1.5.1
2.1.5.2.
2.1.6
PRESSURE
2.1.6.1
2.1.6.2
2.1.6.3
2.1.64
2.1.6.5
2.1.6.6
2.1.6.7
2.1.6.8
Center - Service
300VS,
M clamps.
system
previous
M
IACL001377
M
Display
Replace
less
Look necessary.
M1
based
Plug Set
psi.
M
on
and Power
Decrease
System advisory air
WM
pressure/vacuum removed
Set
400VS, 600DS,
INSPECTION
If
system
300VS,
Check
to
to
date
two
Does
Has
it
have
than
the
been
the
three AND for
cracks
Inspect
and
window
SOURCE
unit
into
pressure
Verify
Ifa
Form
there
regulated,
S0578-01,
proceed
On
system.
air
Advisory
“Press 5 bar
pressure”
Using
air
blue
from
source
does 400VS,
ensure
ensure
OF
the
of
years,
OF
system
and
been
backlight
years
CLEAN
cleaner.
standard
source
to
step
source
appears,
screen.
Pressure
CURRENT,
800CS
OF
ACCURUS
NOT
have a Cart,
600DS,
that
Accurus®
that
the
CPU
BATTERY
battery
per
manufacture
then,
skip
LCD
BACKLIGHT
use a LCD
below)?
three
replaced?
years
lamps
from
If
date
THE
or
other
physical
clean,
as
necessary,
TEST
outlet
Regulator
are
no
leaks
adjustable
Data
Sheet
2.1.7.
Pressure
“Please
metric
manifold
adjust
for low
select
tubing,
Verify
Regulator
Test
Procedure
Version:
1.0
and
Cart
CART
indicate
800CS & Cart,
console
lower
compartment
instruction
or
the
CPU
to
step
2.1.4.
display
yes,
complete
or
more
since
If
yes,
proceed
per
instruction
of
this
PM).
FRONT
PANEL
damage
touchscreen
and
connect
to
120
psi,
by
confirming
90-120
“Standard”.
assy.
low
psi
for
Accurus®
Regulator
supply
pressure
connect
Slowly
pressure
back
to
(if
applicable)
on
and
is
attached
door
in
the
Accurus
battery
LAMPS
with
step
2.1.4.
the
LCD
to
next
in
the
TOUCHSCREEN
to
the
footswitch
or
the
highest
no
audible
pressure
(to
source
operation
pressure
increase
threshold
90
source
300VS, 400VS,
about
pressure
to
120
Form
proceed
replacement
(Hot
hot-cathode
Accurus
front
with
pressure
psi.
S0578-01,
to
to
the
opens
and
Service
Cathode
If
no,
skip
backlight
step.
If
no,
Service
panel
damp
to
rear
of
pressure
hissing
is
not
60-70
or
meter
is
psi)
from
press
Standard
to
until
approximately
Page 2 of
step
2.1.3.
cart
Effective
12-May-2005
Date:
Data
with
closes.
Manual
has
occurred
LCDs
lamps
proceed
step
2.1.4.
lamps
Manual
only)
(systems
or
to
step
touchscreen,
cloth
or
non-ammonia
unit.
available
is
present.
available,
600DS,
until
90
to
85
psi
System
below
indicate
800CS & Cart,
unit
120
psi”.
and
test
Advisory
78-83
18
Sheet
for
two
knob
unless
in
the
LCD
2.1.5.
(must
the
the
s/n
replace
120
displays
If
the
increase
port
on
psi.
be
if
so
is
|
Status: | Effective
gcd
E982T2250+
|
Supersedes: | SOP-0578
ΘΩΒΟΙΗΣΤΕΘΗ
39
eS+:60
80 90
Рен
Alcon
Title: | STP - Accurus
Irvine
SOP
ITCSOP-000578
Technology
Number:
300VS,
Center - Service
400VS,
600DS,
800CS
Test
Procedure
Version:
CURRENT,
and
Cart
1.0
Page 3 of
Effective
12-May-2005
Date:
18
2.1.7
2.1.8
SYSTEM
2.1.7.1
2.1.7.1.1
2.1.7.1.2
ILLUMINATOR
MM
2.1.8.1 | Ensure
2.1.8.2
2.183
Press Remove
2.1.8.4 M Verify
Out
2.1.8.5
2.1.8.6
HM
fingers. M
aligned
COOLING
Verify
cooling
System
U/S
Fan
system
fan
“BULB
is
ILLUMINATOR
illuminator
that
message
icon
is
shown
Install 2 new
Place a sheet
per
bulbs,
Figure
fan
operation:
Flow
Flow
IN
FAILURE”
in
the
Vitreoretinal RELEASE
bulbs
and
appears
at
the
illuminator
using a tissue
of
white
2.
IN
paper
INDICATOR
Surgical
button
insert
I!luminator
on
screen
global or
behind
to
eject
“Bulb
display.
cloth,
bulbs
setup
drawer.
drawer
Failure
and
taking
and
screen.
back
detected....”
care
ensure
into
the
not
each
Accurus.
and
to
touch
bulb
the
bulbs
js
properly
Bulb
with
22
SERVICE
22.1
LATCH
2.2.1.1
22.12
22,13
22.14
2.2.1.5
222
|
Status: | Effective
SYSTEM
2.2.2.1
2.2.2.2
FUNCTIONAL
SEAL
Ensure
1
M
roller Clean
Repair
Ы
or
ΠΜ
on
Plug
Connect
system
Replace
Check
Visually
replace
Inspect
history
that
is
secure.
as
required.
or
replace
POWER
unit
into
Footswitch
Figure 2 -
TEST
REPLACEMENT/
to
be
tested
latch
seal.
all
four
Inspect
Inspect
as
necessary.
inspect
as
intake
log.
all
necessary.
fan
filter
ON
TEST
standard
to
Bulb
ACCURUS
is
turned
hub
rollers
Receiver
footswitch
connectors,
on
rear
outlet.
rear
of
unit
Alignment
CONSOLE
OFF
before
spin
freely
Mechanism
and
receptacles,
of
unit
and
IV
Supersedes: ] SOP-0578
proceeding.
on
the
optical
power
and clean
Pole
cord
panels
Cable
INSPECTION
Receiver
Mechanism.
sensor
for
exposed
and
covers
if
necessary.
(if
applicable).
areas
for
Record
for
BSS
wires
damage.
Verify
residue.
or
damage.
Repair
service
hub
call
L*d
EI8ZIZAZSOb еэчеочзтеен
39
ES+:60
80 90
ReW
Loading...
+ 16 hidden pages