No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
Windows
Adobe Acrobat Reader
®
®
Manual Part Number
N8201-90006
Edition
Edition, January 22, 2008
Printed in USA
Agilent Technologies, Inc.
1400 Fountaingrove Pkwy
Santa Rosa, CA 95403
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than
Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and met.
If software is for use in the performance of
a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted
Introducing the N8201A Performance Downconverter
The Agilent Technologies N8201A performance downconverter down converts a microwave
signal to an IF signal providing IF output frequencies of 7.5, 21.4, and 321.4 MHz to offer
three different signal bandwidth capabilities. External mixing can be utilized to down
convert microwave signals up to 110 GHz. The N8201A is based upon the industry’s most
accurate spectrum analyzer, the PSA Series.
Agilent's synthetic instrument family offers the highest-performing RF/MW LAN-based
modular instrumentation and the smallest footprint for automated test systems; providing
the maximum flexibility and minimizing the cost of an ATS over its lifetime.
Agilent’s synthetic instrument modules use LAN eXtension for Instrumentation (LXI)
modular format. LXI differs from other modular formats (such as VXI and PXI) by using an
external computer and local area network (LAN), rather than embedded computers, for
control.
The LXI standard supports the IEEE 1588 time synchronization and protocol standard, which
allows synchronous triggering of different instruments, even with different-length LAN
cables. The IEEE 1588 precision time protocol (PTP) enables a common sense of time over a
distributed system.
Synthetic instrument modules offered by Agilent Technologies include the following:
•N8201A performance downconverter, 3 Hz to 26.5 GHz
•N8211A performance analog upconverter, 250 kHz to 20 / 40 GHz
•N8212A performance vector upconverter, 250 kHz to 20 GHz
•N8221A IF digitizer, 30 MS/s
•N8241A arbitrary waveform generator, 15-Bit, 1.25 GS/s or 625 MS/s
•N8242A arbitrary waveform generator, 10-Bit, 1.25 GS/s or 625 MS/s
Introducing the N8201A Performance Downconverter 3
Installing Software and Instrument Drivers 6
Step 1. Install Microsoft .NET Version 1.1 7
Step 2. Install the Agilent I/O Libraries 8
Step 3. Install the IVI Shared Components 9
Step 4. Install the Agilent Synthetic Instrument Finder 11
Step 5. Install the Agilent Synthetic Instrument GUI 12
Step 6. Install the IVI-COM Drivers 13
Step 7. Install the Agilent N8201A Performance Downconverter User Interface 15
Installing Optional Software and Instrument Drivers 16
(Optional) Step 8. Install the Agilent N8201A Option H02 Spectrum Analyzer GUI 17
(Optional) Step 9. Install the Microsoft Virtual Machine (VM) 19
(Optional) Step 10. Install the Apache HTTP Server 21
(Optional) Step 11. Install the SA Remote Web Server 24
Step 1. Unpack the N8201A Performance Downconverter 28
Verify the Shipment 28
(Optional) Prepare the Instrument for Rack Mounting 28
Step 2. Connect LAN Cables and Turn On Power 29
(Optional) Connect to a LAN with a Cross-Over LAN Cable 31
Step 3. Verify Connection with Synthetic Instrument Finder 32
(Optional) Step 4. Connect to the Agilent N8201A Option H02 Spectrum Analyzer GUI 34
(Optional) Step 5. Connect to an SA Remote Web Server 36
(Optional) Step 6. Verify Operation < 3 GHz 43
Performing a Self-Test 43
(Optional) Step 7. Verify Operation > 3 GHz 47
Performing a Self-Test 47
Troubleshooting 50
Alternative Ways to Verify Connectivity to the PC 50
How to Use the Synthetic Instrument Finder 51
How to Reset the LAN Configuration 55
How to Set a Static IP Address 56
How to Troubleshoot Connectivity Problems on the Network 60
How to Determine a PCs Configuration Settings 60
If the Instrument was Unable to Join the LAN 62
If the LAN LED is Red 62
If the Instrument’s IP Address or Hostname Cannot be Found with Ping 63
If the Instrument is Not Found by the Synthetic Instrument Finder 63
If the Instrument’s Hostname and PC Cannot Communicate 63
If the Instrument Web Page is Not Visible 64
If the Software Driver Will Not Open the Connection 64
3 Using the Agilent Synthetic Instrument GUI
Starting the Agilent Synthetic Instrument GUI 67
Left Pane 68
Right Pane 68
Features of the Agilent Synthetic Instrument GUI 69
File Menu 69
Help 70
Left Pane 71
Right Pane 72
Settings on the Agilent Synthetic Instrument GUI 74
Frequency 74
Frequency List 74
Level 76
Input RF 76
IF Output 76
Reference Oscillator 76
Preselector (Option 123) 77
External Mixer 77
Calibration 78
Events (Settling Events and Event Logging) 78
Event Logging 79
Preset 80
Refresh All Values 80
4 Front and Rear Panel Features
N8201A Performance Downconverter Front Panel Features 82
RF INPUT 82
Power 82
Line Power LED 83
LAN LED 83
1588 LED 84
Safety 104
Safety Summary 104
Equipment Installation 105
Environmental Conditions 106
Before Applying Power 106
Magnetic Susceptibility 107
Vibration 107
Ground the Instrument or System 107
Fuses and Circuit Breakers 108
Maintenance 108
Safety symbols and Instrument Markings 108
Service and Support 111
Agilent on the Web 111
Return Procedure 112
Shipping the Instrument 112
This installation process installs the required software and instrument drivers used by the
N8201A performance downconverter:
“Verify the PC Meets Minimum Requirements" on page 6
“Step 1. Install Microsoft .NET Version 1.1" on page 7
“Step 2. Install the Agilent I/O Libraries" on page 8
“Step 3. Install the IVI Shared Components" on page 9
“Step 4. Install the Agilent Synthetic Instrument Finder" on page 11
“Step 5. Install the Agilent Synthetic Instrument GUI" on page 12
“Step 6. Install the IVI-COM Drivers" on page 13
“Step 7. Install the Agilent N8201A Performance Downconverter
User Interface" on page 15
Install the following only if the N8201A performance downconverter has Option H02
“(Optional) Step 8. Install the Agilent N8201A Option H02
Spectrum Analyzer GUI" on page 17
Install the following only if the N8201A performance downconverter has Option H02
(Optional Steps 9, 10, and 11 are available to support legacy installations and are not
required if the software from Step 8 is being used. This interface can be installed
along with the Agilent N8201A Option H02 Spectrum Analyzer GUI, but only one
interface can be used at any given time.)
“(Optional) Step 9. Install the Microsoft Virtual Machine (VM)" on page 19
“(Optional) Step 10. Install the Apache HTTP Server" on page 21
“(Optional) Step 11. Install the SA Remote Web Server" on page 24
Agilent Technologies
5
1Software Installation
NOTE
CAUTION
Installing Software and Instrument Drivers
1 Verify the PC Meets Minimum Requirements
• 1 GHz Intel Pentium processor
• Microsoft Windows XP Professional or Home Edition (Service Pack 1
or 2), Windows 2000 (Service Pack 2)
• 512 MB of RAM
• Up to 40 MB of available hard- disk space
• Microsoft Internet Explorer 6.0 (or higher), or Netscape 7.1 or 8.0
2 Place the CD with the Instrument Drivers and Documentation in the CD-ROM drive.
a Adobe Acrobat Reader 5.0 or later is used during this installation process and can be
installed from the CD described above.
b Click Adobe Acrobat Reader 5.0 to install the software, if needed.
3 Click Instrument Drivers.
If the following software or instrument drivers are installed on the PC to be used, uninstall
them and install the software and instrument drivers shipped on the CD described above:
Step 1. Microsoft .NET 1.1
Step 2. Agilent I/O Libraries
Step 3. IVI Shared Components
Step 4. Agilent Synthetic Instrument Finder
Step 5. Agilent Synthetic Instrument GUI
Step 6. IVI-COM Drivers (for N8201A with Option H02 and without Option H02)
Step 7. Agilent N8201A Performance Downconverter User Interface
(Optional) Install the following only if the downconverter has Option H02
(Optional) Steps 9, 10, and 11 are available to support legacy installations and are not
required if the software from Step 8 is being used. This interface can be installed along with
the Agilent N8201A Option H02 Spectrum Analyzer GUI, but only one interface can be used
at any given time.
Step 9. Microsoft Virtual Machine (VM)
Step 10. Apache HTTP Server
Step 11. SA Remote Web Server
There are two different IVI-COM drivers available for the N8201A performance
downconverter.
•If the N8201A performance downconverter is equipped with Option H02, a different
IVI-COM driver is available that enables functionality similar to a PSA spectrum analyzer.
•If the N8201A performance downconverter is not equipped with Option H02, only
downconverter functions are available.
You should see the above dialog when the installation is complete for IVI-COM that
supports Option H02 and the below dialog when the installation is complete for IVI-COM
that supports N8201A performance downconverters that are not equipped with Option H02.
Follow the installation instructions and accept the default settings.
Installing Optional Software and Instrument Drivers
Software for steps 8, 9, 10, and 11 can be installed on your PC, but Option H02 is required to
use these software applications with the N8201A performance downconverter; prior to
continuing, verify that Option H02 is installed.
To Verify that Option H02 is Installed
1 Start the Synthetic Instrument Finder (from the Windows Desktop,
click Start > All Programs > Agilent SI Tools > Synthetic Instrument Finder).
2 Select an instrument, from the left-hand pane of the Synthetic Instrument Finder, and
right-click on the instrument with the mouse.
3 Select Interactive IO.
4 Type *OPT? at the Command prompt and click Send & Read.
5 Read the response in the Instrument Session History box; the required option should be
listed as H02.
If Option H02 is Installed
•If Option H02 is installed, proceed to “(Optional) Step 8. Install the Agilent N8201A
Option H02 Spectrum Analyzer GUI" on page 17.
•If Option H02 is installed, you can also use the SA Remote Web Server. To use this
optional interface, you must perform “(Optional) Step 9. Install the Microsoft Virtual
Machine (VM)" on page 19, “(Optional) Step 10. Install the Apache HTTP Server" on
page 21, and “(Optional) Step 11. Install the SA Remote Web Server" on page 24; this
interface can be installed along with the Agilent N8201A Option H02 Spectrum Analyzer
GUI, but only one interface can be used at any given time.
If Option H02 is Not Installed
•If Option H02 is not installed, software installation is complete!
The N8201A performance downconverter can be manually controlled on instruments
without Option H02 by using the Agilent Synthetic Instrument GUI.
(Refer to “Starting the Agilent Synthetic Instrument GUI" on page 67.)
•Without Option H02, the N8201A performance downconverter cannot be used with
the SA Remote Web Server and does not need software installed for the Microsoft
Virtual Machine (VM), the Apache HTTP Server, or the SA Remote Web Server.
a Close the Interactive IO dialog box.
b Close the Synthetic Instrument Finder dialog box.
c Click Exit CD-ROM. The software and driver installation is complete!
d Restart the computer and continue to “Hardware Setup and Configuration" on
“Step 1. Unpack the N8201A Performance Downconverter" on page 28
“Step 2. Connect LAN Cables and Turn On Power" on page 29
“Step 3. Verify Connection with Synthetic Instrument Finder" on page 32
“(Optional) Step 4. Connect to the Agilent N8201A Option H02
Spectrum Analyzer GUI" on page 34
“(Optional) Step 5. Connect to an SA Remote Web Server" on page 36
“(Optional) Step 6. Verify Operation < 3 GHz" on page 43
•“Alternative Ways to Verify Connectivity to the PC" on page 50
•“How to Reset the LAN Configuration" on page 55
•“How to Set a Static IP Address" on page 56
•“How to Troubleshoot Connectivity Problems on the Network" on page 60
•“How to Determine a PCs Configuration Settings" on page 60
•“If the Instrument was Unable to Join the LAN" on page 62
•“If the LAN LED is Red" on page 62”
•“If the Instrument’s IP Address or Hostname Cannot be Found with Ping" on page 63
•“If the Instrument is Not Found by the Synthetic Instrument Finder" on page 63”
•“If the Instrument’s Hostname and PC Cannot Communicate" on page 63”
•“If the Instrument Web Page is Not Visible" on page 64”
•“If the Software Driver Will Not Open the Connection" on page 64”
Agilent Technologies
27
2Hardware Setup and Configuration
WARNING
NOTE
CAUTION
Step 1. Unpack the N8201A Performance Downconverter
The unique shape of the N8201A performance downconverter was intended to allow
multiple instruments to reside in a compact system that is both modular and
transportable. With instruments adjacent to each other, handles could not be installed on
the respective instruments. Exercise caution when lifting and carrying the instrument to
avoid personal injury. At 25.9 kilograms (57 pounds) shipping weight and 19 kilograms
(42 pounds) net weight, it is recommended that two people be utilized for instrument
lifting and transport.
Verify that any options ordered are included with the shipment by checking the packing
literature included with the shipment.
The serial number label on the N8201A performance downconverter only verifies
hardware/firmware options. The packing literature verifies all items shipped.
Verify the Shipment
1 Inspect the shipping container for damage.
Signs of damage may include a dented or torn shipping container or cushioning material
that shows signs of unusual stress or compacting.
2 Carefully remove the contents from the shipping container and verify that the order is
complete.
The following items are shipped standard with each N8201A performance
downconverter:
•instrument drivers, synthetic graphical user interface, and documentation CD-ROM
(p/n N8200-90004)
•three-prong AC power cord specific to geographic location
(Optional) Prepare the Instrument for Rack Mounting
If the N8201A performance downconverter is to be placed in a system rack, the feet
currently attached can be replaced with the rack mount feet (part number
W1312-40032) supplied with the accessories. Failure to do so can result in a safety
issue. For further information, refer to the Agilent N8200A Series Synthetic Instrument Modules, Rack Configuration Guide, N8200-90003.
Install the N8201A performance downconverter so that the detachable power cord is
readily identifiable and is easily reached by the operator. The detachable power cord is the
instrument disconnecting device. It disconnects the mains circuits from the mains supply
before other parts of the instrument. The front panel switch is only a standby switch and is
not a LINE switch. Alternatively, an externally installed switch or circuit breaker (which is
readily identifiable and is easily reached by the operator) may be used as a disconnecting
device.
Before connecting to a LAN, verify your local policy by contacting the system administrator
in your Information Technology (IT) department and inquire about connecting instruments
to the LAN.
Hardware Setup and Configuration2
•If the network uses DHCP [
assigned to the device automatically. If you need to know what the IP address is, it can
be determined using the Synthetic Instrument Finder. (Refer to “Step 3. Verify
Connection with Synthetic Instrument Finder" on page 32.)
•If DHCP is not present, but the instrument is set to use DHCP (the default), the
instrument waits two minutes for its DHCP request to time out. When the
N8201A performance downconverter is used in this situation, there is a time delay of
approximately three minutes between the time of when the N8201A performance
downconverter’s power is turned on and when it is available for use.
•If the network does not use DHCP, you can use Auto IP or configure your LAN
settings manually. Although you can also manually configure LAN settings in a
network with DHCP, it is recommended that you do so with the assistance of your
system administrator.
•If the network uses Auto IP (does not use DHCP), the N8201A performance
downconverter acquires a 169.254.xxx.xxx address. (Refer to “How to Set a Static IP
Address" on page 56.)
If you wish to communicate directly between the N8201A performance downconverter and
your PC without the use of a LAN hub, you can connect directly to your PC. (Refer to
“(Optional) Connect to a LAN with a Cross-Over LAN Cable" on page 31.)
Dynamic Host Configuration Protocol], an address is
Connect a LAN cable from the LAN connector on your PC to an empty connector on your
internal local area network or LAN hub.
2 Connect a LAN cable from the LAN connector on the rear panel of the
N8201A performance downconverter to an empty connector on your internal local area
network or LAN hub.
3 Turn on power to the PC.
4 Turn on power to the N8201A performance downconverter and wait until the LAN LED
turns solid green or until you hear an attenuator click from within the
N8201A performance downconverter; this can take up to four minutes depending on
whether the instrument is using DHCP or Auto IP.
(Optional) Connect to a LAN with a Cross-Over LAN Cable
If you wish to communicate directly between the N8201A performance downconverter and
your PC without the use of a LAN hub, you can connect directly to your PC.
1 Connect a cross-over LAN cable from the LAN connector on your PC to the LAN
connector on the rear panel of the N8201A performance downconverter.
2 Turn on power to the PC.
3 Turn on power to the N8201A performance downconverter and wait until the LAN LED
turns solid green or until you hear an attenuator click from within the
N8201A performance downconverter; this can take up to four minutes depending on
whether the instrument is using DHCP or Auto IP.
Step 3. Verify Connection with Synthetic Instrument Finder
Agilent supplies a program named the Synthetic Instrument Finder that enables
connection between a PC and instruments that are connected on a LAN [Local Area
Network].
1 From the Windows Desktop,
click Start > All Programs > Agilent SI Tools > Synthetic Instrument Finder.
The Synthetic Instrument Finder should appear and look similar to the following.
2 Select an instrument, from the left-hand pane of the Synthetic Instrument Finder, and
(Optional) Step 4. Connect to the Agilent N8201A Option H02
Spectrum
Analyzer GUI
If Option H02 is not installed, the Agilent N8201A Option H02 Spectrum Analyzer GUI
software cannot be used!
In addition to Option H02, you must have performed “(Optional) Step 8. Install the Agilent
N8201A Option H02 Spectrum Analyzer GUI" on page 17; this interface can be installed
along with the SA Web Remote Server, but only one interface can be used at any given time.
(For further software installation information, refer to “Installing Optional Software and
Instrument Drivers" on page 16.)
The N8201A performance downconverter can be manually controlled on instruments
without Option H02 by using the Agilent Synthetic Instrument GUI.
(Refer to “Starting the Agilent Synthetic Instrument GUI" on page 67.)
This section describes how to access and use the Agilent N8201A Option H02
Spectrum Analyzer GUI.
1 From the Windows Desktop, click
Start > All Programs > Agilent SI Tools > N8201A Option H02 Spectrum Analyzer GUI.
2 Click File > Connect (upper-left corner)
on the Agilent N8201A Option H02 Spectrum Analyzer GUI.
3 Enter the VISA Address (for example, TCPIP0::141.121.87.18::inst0::instr) of the
instrument being connected to and click OK.
•If you don’t know the IP address of the instrument, that is used as part of the
VISA Address connection string, refer to the Default IP Address that is displayed
when verifying connection with the Synthetic Instrument Finder. (Refer to “Step 3.
Verify Connection with Synthetic Instrument Finder" on page 32.)
The Agilent N8201A Option H02 Spectrum Analyzer GUI controls the N8201A performance
downconverter that is equipped with Option H02 and simulates the functionality of an
Agilent PSA spectrum analyzer. To learn how to use the controls of this GUI interface, refer
to the PSA documentation.
PSA documentation is available from the Agilent N8201A Performance Downconverter
Synthetic Instrument Module, 3 Hz to 26.5 GHz, Instrument Drivers and Documentation CD
(refer to “Installing Software and Instrument Drivers" on page 6) or from the Web at:
(Optional) Step 5. Connect to an SA Remote Web Server
If Option H02 is not installed, the SA Remote Web Server software cannot be used!
In addition to Option H02, you must have installed the Microsoft Virtual Machine (VM), the
Apache HTTP Server, and the SA Remote Web Server software; this interface can be
installed along with the Agilent N8201A Option H02 Spectrum Analyzer GUI, but only one
interface can be used at any given time. (For further software installation information, refer
to “Installing Optional Software and Instrument Drivers" on page 16.)
The N8201A performance downconverter can be manually controlled on instruments
without Option H02 by using the Agilent Synthetic Instrument GUI.
(Refer to “Starting the Agilent Synthetic Instrument GUI" on page 67.)
1 From the Windows Desktop,
select Start > All Programs > Agilent I/O Libraries Suite > Agilent Connection Expert.
If you are attempting to access the web server from another machine, you need to make
sure to use the hostname or IP Address of the adapter that is on the same network as
the machine you are using. (In regards to the example above, if you are trying to access
the webpage from a machine on the network that is connected to the adapter with the
169.254.241.150 address, you need to use 169.254.241.150 and not 141.121.83.141 as the
IP Address.)
Operation verification is a test that, when completed, will ensure that the downconverter is
operating correctly in the low band (< 3 GHz).
Make sure that the total power of all signals at the downconverter input does not
exceed +30 dBm (1 watt).
Performing a Self-Test
Instrument Connections
•Downconverter: 7.5 MHz Out
•Spectrum Analyzer: RF Input
Hardware Setup and Configuration2
The N8201A performance downconverter can be controlled with either the
SA Remote Web Server or the Agilent Synthetic Instrument GUI, but only one interface can
be used at any given time.
•If Option H02 is not installed, the N8201A performance downconverter can be manually
controlled using the Agilent Synthetic Instrument GUI.
•If Option H02 is installed, you can use either the SA Remote Web Server or the
Agilent Synthetic Instrument GUI.
(For information on using these different interfaces, refer to “(Optional) Step 4. Connect
to the Agilent N8201A Option H02 Spectrum Analyzer GUI" on page 34 or “Starting the
Agilent Synthetic Instrument GUI" on page 67.)
Operation verification is a test that, when completed, will ensure that the
N8201A performance downconverter is operating correctly in the high band (> 3 GHz).
Make sure that the total power of all signals at the N8201A performance
downconverter
Performing a Self-Test
Downconverter to Spectrum Analyzer Connections
•Downconverter: 7.5 MHz Out
•Spectrum Analyzer: RF Input
Downconverter to Source Connections
input does not exceed +30 dBm (1 watt).
Hardware Setup and Configuration2
•Downconverter: RF Input
•Source: RF Output
Perform the following procedure to run a self-test:
1 From the Windows Desktop,
select Start > All Programs > Agilent SI Tools > Agilent Synthetic Instrument GUI.
In addition to using “Step 3. Verify Connection with Synthetic Instrument Finder" on
page 32 and “(Optional) Step 4. Connect to the Agilent N8201A Option H02
Spectrum Analyzer GUI" on page 34, connectivity can be verified between the
N8201A performance downconverter and the PC with the following:
•Verify that the LAN LED on the N8201A performance downconverter’s rear panel is
green or blinking green. This indicates a good connection.
If the LED is off, there is a problem with your LAN connection.
•Verify that the LAN LED on the N8201A performance downconverter’s front panel (next
to the LAN port) is solid green.
If the LED turns red, this indicates a problem with your LAN connection. This takes
approximately 60 seconds.
•Ping the N8201A performance downconverter from the PC.
a From the Windows Desktop, select Start > Run.
b At the Open prompt, type CMD and press Enter to open a command window.
c At the command prompt, type Ping and the instrument’s IP address (for example,
Ping 141.121.84.108.) or type Ping and the instruments hostname (for
example, Ping a-n8201a-00179).
Agilent supplies a program named the Synthetic Instrument Finder that enables
connection between a PC and instruments that are connected on a LAN [Local Area
Network].
1 From the Windows Desktop,
click Start > All Programs > Agilent SI Tools > Synthetic Instrument Finder.
The Synthetic Instrument Finder should appear and look similar to the following.
Hardware Setup and Configuration2
2 Select an instrument, from left-hand pane of the Synthetic Instrument Finder, and
On the instrument front panel, near the power switch, is a recessed button labeled
“RESET”. This button enables you to place the LAN configuration of the instrument
into a known state.
When this button is pressed (a straightened paper clip will do the job) the following
settings are made and the system reboots.
•Subnet Mask is set to 255.255.0.0
•DHCP is set to on
•Auto IP is set to on
•If DHCP and Auto IP are set to off, the IP address will be set to 192.168.EE.FF,
where EE and FF are the last two parts of the MAC address (AA.BB.CC.DD.EE.FF).
This is designed to prevent multiple instruments from using the same default IP
address (refer to the instrument label).
If you had manually configured LAN settings before, you may have to reconfigure your
instrument to reset DHCP and Auto IP to OFF. Refer to “How to Set a Static IP Address" on
page 56.
Hardware Setup and Configuration2
•The instrument hostname is set to A-N82XXA-NNNNN, where N82XXA is the
instrument model number (such as N8201A) and NNNNN represents the last five
digits of the instrument serial number.
If the instrument is in an environment with a DHCP server, it is assigned an IP address
through DHCP. The IP address can be found by using the instrument hostname as the
URL in a web browser.
Without DHCP, the instrument will use Auto IP and acquire a 169.254.X.X address. If
no DHCP is present, but the instrument is set to use DHCP (the default), the instrument
will wait two minutes for its DHCP request to time out. In this case, there is a time delay
of approximately three minutes between when the instrument is powered on and when it
is usable.
The DHCP server automates the process of setting up the IP addresses on your network
by default. When the N8201A performance downconverter is turned on, it searches for a
DHCP server on the network and selects a “dynamic IP address”. Each time the
N8201A performance downconverter is rebooted, the N8201A performance
downconverter may get a different IP address. To set the N8201A performance
downconverter to a static IP address, rather than allowing the DHCP server to select an
auto IP address:
1 Assign a N8201A performance downconverter instrument IP address that will work with
your computer.
F
For a company wide network, your system administrator will have to assign an IP address
that is compatible with your PC. If you have a private LAN network or a direct connection
from your PC to the instrument, you can assign the IP address. Refer to “Step 1. Unpack the
N8201A Performance Downconverter" on page 28.
2 Connect the N8201A performance downconverter in one of the following two
configurations:
•Connect a LAN cable from the LAN connector on your PC to an empty connector on
your internal local area network or LAN hub. Connect a LAN cable from the LAN
connector on the rear panel of the N8201A performance downconverter to an empty
connector on your internal local area network or LAN hub.
Figure 1Connecting the PC LAN cable to a company/private LAN to the instrument LAN
•Connect a cross-over cable from the LAN connector on your PC to the LAN
connector on the rear panel of the N8201A performance downconverter.
How to Troubleshoot Connectivity Problems on the Network
The Synthetic Instrument Finder program is used to find instruments on a network when
the N8201A performance downconverter is connected through a router or cross-over
cable. There are three possible configurations:
•connecting the PC through a company wide site LAN connection to the
N8201A performance downconverter
•connecting the PC to the same private LAN network as the instrument
•connecting the PC directly to the instrument using a cross-over cable - this would
typically be used for troubleshooting and is not normally used to control an
instrument directly
The N8201A performance downconverter is shipped with a default IP address. This
default IP address is 192.168.EE.FF, where EE and FF are the last two parts of the
Media Access Control (MAC) address (AA.BB.CC.DD.EE.FF).
How to Determine a PCs Configuration Settings
From a DOS Window
1 From the Windows Desktop, click Start > Run.
2 At the Open: prompt, typeCMDand press Enter to open a DOS window.
3 At the command prompt, type ipconfig/all to display the PCs network connection
details.
Or,
From the PCs Control Panel
1 From the Windows Desktop, click Start > Settings > Control Panel > Network and
Internet Connections.
2 From the Network and Internet Connections window, double-click the Local Area
Connection.
3 In the Local Area Connection Status dialog, click the Support tab and click Details to
display the PCs Network Connection Details.
The Network Connection Details include:
•Physical Address
•DHCP status, enabled or disabled (displayed when using the DOS window ipconfig
command only)
•Auto configuration enabled or disabled (displayed when using the DOS window
If the Instrument’s IP Address or Hostname Cannot be Found with Ping
Possible CausesPossible Solutions
The instrument was unable to join the LAN.See “If the Instrument was Unable to Join the
LAN" on page 62.
The instrument’s LAN settings are incorrect.Verify that the instrument’s settings are appropriate
for your LAN.
A firewall is preventing communication between your
PC and your instrument.
The instrument is using Auto-IP (That is, the
instrument assigned itself a 169.254.x.x IP address)
and your PC is not using Auto IP (That is, PC does not
have a 169.254.x.x IP address.)
Make sure that your firewall settings allow
communication between your PC and other devices.
• Disable Auto-IP on the instrument.
• Configure your PC to use Auto-IP.
If the Instrument is Not Found by the Synthetic Instrument Finder
Possible CausesPossible Solutions
The instrument was unable to join the LAN.See “If the Instrument was Unable to Join the
LAN" on page 62.
The instrument and PC are on different
switches/hubs and different subnets.
• Put the instrument on the same switch or hub as
your PC.
• If the instrument is using DHCP, make sure that the
instrument and the PC are put on the same subnet.
• If the instrument is using a static IP address, make
sure that the instrument IP address and subnet
mask put the instrument on the same subnet as
your PC.
If the Instrument’s Hostname and PC Cannot Communicate
Possible CausesPossible Solutions
No DNS server is available.Communicate with the instrument using the
instrument’s IP address.
The DNS server has not been updated.Wait several minutes.
The PC cannot communicate with the device over
LAN.
“Starting the Agilent Synthetic Instrument GUI" on page 67
“Features of the Agilent Synthetic Instrument GUI" on page 69
“Settings on the Agilent Synthetic Instrument GUI" on page 74
Synthetic Instrument GUI
Agilent Technologies
65
3Using the Agilent Synthetic Instrument GUI
NOTE
Although only one interface can be used at any given time, the N8201A performance
downconverter can be controlled with any of the following:
•Agilent N8201A Option H02 Spectrum Analyzer GUI
•SA Remote Web Server
•Agilent Synthetic Instrument GUI
Which interface should be used:
•If Option H02 is not installed, the N8201A performance downconverter can be manually
controlled using the Agilent Synthetic Instrument GUI; without Option H02, the
N8201A performance downconverter cannot be controlled with SCPI commands.
•If Option H02 is installed, you can use either the Agilent N8201A Option H02 Spectrum
Analyzer GUI, the SA Remote Web Server, or the Agilent Synthetic Instrument GUI.
(For information on using these different interfaces, refer to “(Optional) Step 4. Connect
to the Agilent N8201A Option H02 Spectrum Analyzer GUI" on page 34, “(Optional) Step
5. Connect to an SA Remote Web Server" on page 36, or “Starting the Agilent Synthetic
Instrument GUI" on page 67.)
To verify that Option H02 is installed:
1 Start the Synthetic Instrument Finder (from the Windows Desktop,
click Start > All Programs > Agilent SI Tools > Synthetic Instrument Finder).
2 Select an instrument, from the left-hand pane of the Synthetic Instrument Finder, and
right-click on the instrument with the mouse.
3 Select Interactive IO.
4 Ty p e *OPT? at the Command prompt and click Send & Read.
5 Read the response in the Instrument Session History box; the required option should be
Select an N8201A performance downconverter and click Tes t (lower-left corner) on the
Connect to Instrument dialog box.
•If the bottom of the dialog box displays the message “Connection Succeeded“, the
instrument was found and communication has been established.
•If the bottom of the dialog box displays the message “N8201A is not supported”, the
instrument is not communicating. Refer to “Troubleshooting" on page 50.
5 Click Connect (lower-right corner) on the Connect to Instrument dialog box and the
following dialog box should appear.
If the Agilent Synthetic Instrument GUI appears, the N8201A performance downconverter
has successfully connected using a LAN connection!
Left Pane
The lower portion of the left pane displays the settings available for adjustment on the
N8201A performance downconverter. Click a function button to activate that function and
the related functions are displayed in the Properties area. For example, Center Frequency is
the active function and all settings associated with Center Frequency are available for
modification.
Right Pane
The upper portion of the right pane always displays the functions that are most commonly
used for a measurement. These functions are also accessible from the left pane. Changing
one of these parameters changes the setting in the left pane as well.
The lower portion of the right pane can have three tabs: Dynamic Help, Instrument
Information, and the Event Log. For more information on theses areas, refer to “Dynamic
Help" on page 72, “Instrument Information" on page 73, and “Event Log" on page 73.
The File menu accesses options for instrument connection, save and recall settings, and
exiting the application. Theses tasks are also available by clicking the icons on the tool bar.
Figure 3File sub menu
Connect
Accesses the Connect to Instrument dialog box which is used to connect to an instrument
on the LAN hub.
Figure 4Connect to Instrument Window
Reconnect
Re-establishes the connection to the instrument if the instrument has been disconnected.
The upper portion of the right pane always displays the functions that are most commonly
used for a measurement. These functions are also accessible from the left pane. Changing
one of these parameters changes the setting in the left pane as well.
The lower portion of the right pane can have the following tabs: Dynamic Help, Instrument
Information, and Event Log.
Dynamic Help
Provides information about the function currently selected in the left pane.
Sets the center frequency or frequency offset of the N8201A performance downconverter.
Center Frequency
Sets the center frequency while the span remains constant. The frequency range of the
N8201A performance downconverter is 3 Hz to 26.5 GHz plus the frequency offset. If the
External Mixer Enabled equals External, the frequency range is 3 Hz to 335 GHz.
Frequency Offset
Enables you to input a frequency offset value to account for external frequency
conversions. This value is added to the display readout of the marker frequency, center
frequency, start frequency, stop frequency, and all other absolute frequency settings. Limits
are
−500 THz to 500 THz.
Frequency List
Sets up a list of frequencies for the instrument to step through. The list begins with Start
Frequency, and adds the Step Frequency until the Stop Frequency is reached. The next
trigger causes the list to be repeated.
Disabled Prevents a frequency step sequence from being initiated.
Start Sets the first frequency that will be swept on a list sweep.
Stop Sets the last frequency that will be swept before the list is repeated.
Step Sets the difference between successive swept frequencies.
Trigger Se t u p Sets up the trigger that will be used to step through the frequency list.
Trigger Source
Sets the source of triggers. When the specified trigger occurs, the instrument will move to
the next frequency in its frequency list. Possible values are:
Table 2Controls available when Trigger Source is set to ALARM0
Alarm ModeDefines the way that Alarm Time will be interpreted. In absolute mode,
the alarm will begin firing at the time of day specified in Absolute Alarm
Time. In relative mode, the alarm will begin firing at a time relative to
when the alarm is set up.
Relative Alarm TimeDefines the number of seconds after pressing Execute Trigger that the
first alarm will occur.
Absolute Alarm TimeDefines the time that the first alarm will occur.
Alarm PeriodDefines the time between successive alarms.
Alarm RepeatDefines the number of times the alarm will be repeated.
A value of 1 means that the alarm will occur once and will not be
repeated. A value of 2 means that the alarm will occur twice.
A value of 0 means that the alarm will be repeated until the alarm is
disabled.
Trigger Detection
Controls the trigger polarity. It is Positive to trigger on a rising edge and Negative to trigger
on a falling edge.
Sets the input attenuation. The input attenuation can be set from 0 to 70 dB in 2 dB
steps.The input impedance is set to 50 ohms. Input attenuation is used to minimize
compression caused by a signal level that is too high in amplitude.
Input RF
Accesses the functions to select the input signal source (choices are RF, 50 MHz Cal, or
External Mixer), enable a preamplifier, and to set the coupling to either AC or DC.
IF Output
Accesses the functions to set the IF Bandwidth and IF Frequency.
Table 3Grouped Values
IF OutputIF Bandwidth
IF Output 10 to 3 MHz21.4 MHz50 Ohm
IF Output 2100 MHz321.4 MHz50 Ohm
IF Output 310 MHz7.5 MHz50 Ohm
Reference Oscillator
Allows you to use either the internal or an external reference oscillator for making
measurements.
Reference Output
Switches the 10 MHz out signal on the front panel of the N8201A performance
downconverter On and Off.
Reference Input Source
Specifies whether to use an External or Internal source for the reference oscillator. If set to
External, a value must be specified for External Frequency with limits from 1 MHz to
30 MHz.
Allows you to manually adjust the preselector filter center frequency to optimize its
response on the signal of interest. When enabled, the center frequency must be set to
3.045 GHz or greater.
Limits are: –250 MHz to 250 MHz.
PreSelector Enabled
Enables or disables the preselector.
Enabled: Can be set if the Center Frequency is greater than 3.045000000 GHz.
Possible Values: Enabled, Disabled
Peak
Using the Agilent Synthetic Instrument GUI3
Performs a peak search.
Enabled: Can run if Center Frequency is greater than 3.045000000 GHz.
Instrument Functions: A - L
External Mixer
Selects either the internal mixer or an external mm-wave mixer.
•When Internal mixing is selected, normal operation and all other external mixing
functions are unavailable.
•When External mixing is selected, you can analyze high frequency signals (that is,
higher than the spectrum analyzer’s maximum frequency) by using an appropriate
external mixer.
Band Selects a band, or sets the frequency band to be user defined.
Harmonic A user defined frequency band. Limits are −50 to −1 or 1 to 50.
Mixer Bias Turns on/off the mixer bias and adjusts an internal bias source for use with
external mixers. The bias signal is present on the center conductor of the IF INPUT
connector on the front panel.
Bias Level Sets the bias level. Limits are −10 mA to 10 mA.
Calibration
Accesses the Align system and immediately executes an alignment cycle of all the
subsystems (Align RF, Align IF, Align ADC, and Align Current Sys Gain). The instrument will
change any current system settings currently underway, perform the full alignment, and
then return to the current system settings that were in place before the calibration began.
A calibration nominally takes 35 seconds to perform.
The N8201A downconverter will meet its specifications when:
•It is stored a minimum of two hours within the operating temperature range and turned
on for at least 30 minutes followed by a calibration.
•When a calibration has been performed within the past 24 hours or when the
temperature changes 3 °C.
•The front panel 1
st
LO OUT connector is terminated with a 50 ohm load.
Events (Settling Events and Event Logging)
Settling Event
Allows you to define the parameters of the Settling Event. The Settling Event has a positive
edge when the frequency is set and a negative edge when the system is stabilized.
Settling Channel
Selects the channel to configure. Possible values are:
Selects the operating mode for the Settling Channel: Disabled, Driven, or Wired OR.
Wired OR is only available for LXI channels.Use Wired OR if more than one instrument is
driving the LXI channel. When only one instrument is driving the LXI channel then use the
Driven mode.
Settling Destination Path
The target of the LAN event. The Settling Destination Path can be either an IP address, a
hostname, or 'All'. (The function is only visible if Settling Channel is a LAN channel and can
be set if the Settling mode is not Disabled.)
Settling Slope
Determines whether an event will occur on positive or negative edges. Possible values are:
Positive or Negative.
Settling mode must be enabled for Settling Slope to be set.
Execute Settling Event
Starts a settling event after all of the parameters are define.
Event Logging
When enabled, a history of all instrument errors and messages are displayed.
Enable Logging
Enables or disables event logging.
Event Log
Displays the event log in the right pane when the Event Log page is selected.
“N8201A Performance Downconverter Front Panel Features" on page 82
“N8201A Performance Downconverter Rear Panel Features" on page 87
“Interconnect Cabling" on page 89
“Operational Considerations" on page 90
Electrostatic discharge (ESD) can damage the highly sensitive components in your
instrument. ESD damage is most likely to occur as the instrument is being installed
or when cables are connected and disconnected. Protect the circuits from ESD
damage by wearing a grounding strap that provides a high resistance path to
ground. Alternately, ground yourself to discharge any static charge built-up by
touching the outer shell of any grounded instrument chassis before touching the
port connectors.
Agilent Technologies
81
4Front and Rear Panel Features
Option V7L
Option 219
CAUTION
N8201A Performance Downconverter Front Panel Features
Figure 10 N8201A performance downconverter front panel features
RF INPUT
The input for an external 3 Hz to 26.5 GHz RF signal. If AC Coupled, the range is
20 MHz to 26.5 GHz.
The total power of all signals at the downconverter input must not exceed +30 dBm
(1 watt).
Power
The front panel power switch is a standby switch only; it is not a LINE switch (power
disconnecting device). When in Standby, the LED is amber, When ON the LED is green.
The green LED indicates when the downconverter standby switch is set to the on position.
The green LED is off when the switch is in the standby position.
The yellow LED indicates when the downconverter standby switch is set to the standby
position. The yellow LED is off when the switch is in the on position
The LAN LED indicator works in the following states:
The IEEE 1588 LED clock status has the following states:
State of the ClockClock StatusIllumination
OFFNot synchronizedNone
ONSynchronized, clock is IEEE 1588 SlaveSolid green
ONSynchronized, clock is IEEE 1588 Master Blinking green (once
every second)
ONSynchronized, clock is IEEE 1588 to
OFFIEEE 1588 is in a fault stateSolid red
COHERENT CARRIERS
Do not remove the cables connecting the coherent carrier inputs and outputs. These cables
are connected at the factory and must remain so for proper instrument operation.
3-7 GHz 1st LO In Provides the 1st LO input signal (3 to 7 GHz) to the downconverters
internal LO distribution circuitry.
3-7 GHz 1st LO Out Provides the 1st LO output signal (3-7 GHz at +15 dBm max) as a
coherent carrier reference for phase locking two receivers.
3.6 GHz 2nd LO In Provides the 2st LO input signal (3.6 GHz) to the downconverters
internal LO distribution circuitry.
3.6 GHz 2nd LO Out Provides the 2nd LO output signal (3.6 GHz at +3 dBm) as a
coherent carrier reference for phase locking two receivers.
1st LO Out The 1st LO output (3-7 GHz at +15 dBm max) allows connections for external
mixing. The 1st LO output routes the internal first LO signal to an external mixer, which uses
the higher harmonics to mix with the high frequency signals. The external mixer’s IF output
connects to the downconverter’s IF input port.
Pre-Sel Out The Preselected external mixer tune output offers tuning voltage for a
preselected mixer.
IF In (Option AYZ) The IF input (321.4 MHz Nominal at -20 dBm, +10 dBm max) allows
connections for external mixing. This port is only functional when Option AYZ is ordered.
IF OUTPUTS
321.4 MHz The 321.4 MHz IF output signal is present. Conversion gain is nominally +2 to
+4 dB (with 0 dB input attenuation). Conversion gain varies nominally ±3 dB as a function
of tune frequency. The conversion gain drops at higher frequencies.
Front and Rear Panel Features4
21.4 MHz The 21.4 MHz IF output signal is present. Conversion gain is nominally
+10 dB (with 0 dB input attenuation). The conversion gain drops at higher frequencies.
7.5 MHz The 7.5 MHz IF output signal is present. The 7.5 MHz output signal is at
nominally at +13 dBm.
TRIGGERS
These SMB trigger input and output connectors are used to control the waveforms and
create event-based signal simulation. These connectors support LVTTL logic levels.
In allows external triggering of measurements.
Out used to synchronize other test equipment with the downconverter.
REFERENCES
1 - 30 MHz In Selects an external reference to phase lock all oscillators in the
instrument. You can select any external reference frequency between 1 and 30 MHz.
10 MHz Out A switched output of the analyzer’s internal 10 MHz reference signal locks
other test equipment to the same frequency reference used by the downconverter. The
10 MHz out signal is at +5 dB.
N8201A Performance Downconverter Rear Panel Features
Figure 11 N8201A performance downconverter rear panel features
AC Power Receptacle
The AC voltage is connected here. The power cord receptacle accepts a
three-pronged power cable that is shipped with the N8201A performance downconverter. The voltage range is 100/120/220/240 volts with a frequency range of
50 to 60 Hz and is automatically selected by the power supply.
LAN
This LAN interface allows ethernet local area network communication through a 100BaseT
LAN cable.
The LXI (LAN eXtensions for Instrumentation) Trigger Bus is a hardware bus providing eight
trigger channels using M-LVDS (low-voltage differential signaling). Cables connect various
instruments together in a daisy chain or star configuration. Any instrument in a cluster can
send or receive triggers on any of the channels.
For more information about the LXI Trigger Bus and LAN triggering, refer to the following
“LXI Consortium” white papers on your “Agilent N8201A Performance Downconverter
Documentation CD”:
This section includes instructions for configuring a LAN interface to the
Agilent Technologies N8221A IF digitizer and the N8201A performance downconverter. The
instructions assume that the computer being used has a LAN card installed and is
configured for TCP/IP protocol.
Agilent 89601A Vector Signal Analysis Software
The Agilent Technologies 89601A vector signal analysis (VSA) software is the virtual
instrument interface for the N8221A IF digitizer/N8201A performance downconverter and
offers a wide range of troubleshooting tools for analyzing signals.
Configuring the Local Area Network (LAN) Interface
You must ensure that the IP addresses for your computer and
N8221A IF digitizer/N8201A performance downconverter are compatible, then configure
the LAN interface. The IP address consists of four groups of numbers separated by periods
(for example 169.254.xxx.xxx).
“Using, Inspecting, and Cleaning RF Connectors" on page 92”
“General Procedures and Techniques" on page 95
“Instrument Removal" on page 98”
“Instrument Installation" on page 100
This chapter provides preventative maintenance information, which should be reviewed
prior to working with the Agilent system. This information applies to all Agilent-supplied
instruments in the system and the system as a whole.
Agilent Technologies
91
5Preventive Maintenance
CAUTION
Using, Inspecting, and Cleaning RF Connectors
Taking proper care of cables and connectors will protect your system’s ability to make
accurate measurements. One of the main sources of measurement inaccuracy can be
caused by improperly made connections or by dirty or damaged connectors.
The condition of system connectors affects measurement accuracy and repeatability. Worn,
out-of-tolerance, or dirty connectors degrade these measurement performance
characteristics.
Repeatability
If you make two identical measurements with your system, the differences should be so
small that they will not affect the value of the measurement. Repeatability (the amount of
similarity from one measurement to another of the same type) can be affected by:
•Dirty or damaged connectors
•Connections that have been made without using proper torque techniques (this applies
primarily when connectors in the system have been disconnected, then reconnected).
Static-Sensitive Devices
This system contains instruments and devices that are static-sensitive. Always take
proper electrostatic precautions before touching the center conductor of any
connector, or the center conductor of any cable that is connected to any system
instrument. Handle instruments and devices only when wearing a grounded wrist or
foot strap. When handling devices on a work bench, make sure you are working on
an anti-static worksurface.
RF Cable and Connector Care
Connectors are the most critical link in a precision measurement system. These devices are
manufactured to extremely precise tolerances and must be used and maintained with care
to protect the measurement accuracy and repeatability of your system.
To extend the life of your cables or connectors:
•Avoid repeated bending of cables—a single sharp bend can ruin a cable instantly.
•Avoid repeated connection and disconnection of cable connectors.
•Inspect the connectors before connection; look for dirt, nicks, and other signs of damage
or wear. A bad connector can ruin the good connector instantly.
•Clean dirty connectors. Dirt and foreign matter can cause poor electrical connections
and may damage the connector.
Look for metal particles from the connector threads and other signs of wear (such as
discoloration or roughness). Visible wear can affect measurement accuracy and
repeatability. Discard or repair any device with a damaged connector. A bad connector can
ruin a good connector on the first mating. A magnifying glass or jeweler’s loupe is useful
during inspection.
SMA Connector Precautions
Use caution when mating SMA connectors to any precision 3.5 mm RF connector. SMA
connectors are not precision devices and are often out of mechanical tolerances, even
when new. An out-of-tolerance SMA connector can ruin a 3.5 mm connector on the first mating. If in doubt, gauge the SMA connector before connecting it. The SMA center
conductor must never extend beyond the mating plane.
To prevent electrical shock, disconnect the module from mains before cleaning.
Use a dry cloth or one slightly dampened with water to clean the external case
parts. Do not attempt to clean internally.
1 Blow particulate matter from connectors using an environmentally-safe aerosol such as
Aero-Duster. (This product is recommended by the United States Environmental
Protection Agency and contains tetrafluoroethane. You can order this aerosol from
Agilent (see Tab l e 5).)
2 Use alcohol and a lint-free cloth to wipe connector surfaces. Wet a small swab with a
small quantity of alcohol and clean the connector with the swab.
3 Allow the alcohol to evaporate off of the connector before making connections.
Do not allow excessive alcohol to run into the connector. Excessive alcohol entering
the connector collects in pockets in the connector’s internal parts. The liquid will
cause random changes in the connector’s electrical performance. If excessive
alcohol gets into a connector, lay it aside to allow the alcohol to evaporate. This may
take up to three days. If you attach that connector to another device it can take
much longer for trapped alcohol to evaporate.
Table 5Cleaning Supplies Available from Agilent
Product Part Number
Aero-Duster8500-6460
Isopropyl alcohol8500-5344
Lint-Free cloths 9310-0039
Small polyurethane swabs9301-1243
Cleaning connectors with alcohol should only be performed with the
instruments’ mains power cord disconnected, in a well ventilated area.
Connector cleaning should be accomplished with the minimum amount of
alcohol. Prior to connector reuse, be sure that all alcohol used has dried, and that
the area is free of fumes.
If flammable cleaning materials are used, the material should not be stored, or
left open in the area of the equipment. Adequate ventilation should be assured to
prevent the combustion of fumes, or vapors.
This section introduces you to the various cable and connector types used in the system.
Read this section before attempting to remove or install an instrument! Each connector
type may have unique considerations.
Always use care when working with system cables and instruments.
Preventive Maintenance5
Figure 13 GPIB, 3.5 mm, Type-N, power sensor, and BNC connectors
These are removed by two captured screws, one on each end of the connector; these
usually can be turned by hand. Use a flathead screwdriver if necessary.
GPIB connectors often are stacked two or three deep. When you are removing multiple
GPIB connectors, disconnect each connector one at a time. It is a good practice to connect
them back together even if you have not yet replaced the instrument; this avoids confusion,
especially if more than one instrument has been removed.
When putting GPIB connectors back on, you must again detach them from one another and
put them on one at a time.
Precision 3.5 mm connectors
These are precision connectors. Always use care when connecting or disconnecting this
type of connector. When reconnecting, make sure you align the male connector properly.
Carefully join the connectors, being careful not to cross-thread them.
Loosen precision 3.5 mm connectors on flexible cables by turning the connector nut
counter-clockwise with a 5/16 inch wrench. Always reconnect using an 8 inch-lb torque
wrench (Agilent part number 8720-1765). Semirigid cables are metal tubes, custom-formed
for this system from semirigid coax cable stock.
3.5 mm connectors with a gold hex nut
The semirigid cables that go to the RF outputs of some devices have a gold connector nut.
These do not turn. Instead, the RF connector on the instrument has a cylindrical connector
body that turns. To disconnect this type of connector, turn the connector body on the
instrument clockwise. This action pushes the cable’s connector out of the instrument
connector.
To reconnect, align the cable with the connector on the instrument. Turn the connector
body counterclockwise. You may have to move the cable slightly until alignment is correct
for the connectors to mate. When the two connectors are properly aligned, turning the
instrument’s connector body will pull in the semirigid cable’s connector. Tighten firmly by
hand.
3.5 mm connectors with a silver hex nut
All other semirigid cable connectors use a silver-colored nut that can be turned. To remove
this type of connector, turn the silver nut counter-clockwise with a 5/16 inch wrench.
When reconnecting this type of cable:
•Carefully insert the male connector center pin into the female connector. (Make sure the
cable is aligned with the instrument connector properly before joining them.)