THE MATERIAL CONTAINED IN THIS DOCUMENT IS
PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED,
WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW,
AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS
OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY
INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY
INFORMATION CONTAINED HEREIN. SHOULD AGILENT
AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS COVERING THE MATERIAL IN
THIS DOCUMENT THAT CONFLICT WITH THESE TERMS,
THE WARRANTY TERMS IN THE SEPARATE AGREEMENT
WILL CONTROL.
R
ESTRICTED RIGHTS NOTICE
If software is for use in the performance of a U.S. Government prime
contract or subcontract, Software is delivered and licensed as “Commercial
computer software” as defined in DFAR 252.227-7014 (June 1995), or as a
“commercial item” as defined in FAR 2.101(a) or as “Restricted computer
software” as defined in FAR 52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use, duplication or disclosure of
Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and Agencies of the U.S. Government
will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no
greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or
DF AR 252.227- 7015 (b)( 2) (November 1995), as applicabl e in any t echnical
data.
ii Agilent R/Q/U/V281A, B Operating Note
In This Manual…
Description, page 1
•
•Specifications, page 3
•Care, Connection and Torque, page 5
AssistanceProduct maintenance agreements and other customer assistance agreements
are available for Agilent Technologies products.
For assistance, contact Agilent Technologies (refer to “Contacting Agilent”
Review this product and related documentation to familiarize yourself with
safety markings and instructions before you operate the instrument. This
product has been designed and tested in accordance with international
standards.
WARNING The WARNING notice denotes a hazard. It calls attention to a procedure,
practice, or the like, that, if not correctly performed or adhered to, could result
in personal injury. Do not proceed beyond a WARNING notice until the
indicated conditions are fully understood and met.
CAUTION The CAUTION notice denotes a hazard. It calls attention to an operating
procedure, practice, or th e like, whic h, if not cor rect ly performed or adhered
to, could result in damage to the product or loss of important data. Do not
proceed beyond a CAUTION notice until the indicated conditions are fully
understood and met.
Agilent R/Q/U/V281A, B Operating Note v
Instrument Markings
When you see this symbol on your instrument, you should refer to the instrument’s
instruction manual for important information.
!
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have a laser output.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered trademark of the European Community. If it is
accompanied by a year, it indicates the year the design was proven.
The CSA mark is a registered trademark of the Canadian Standards Association.
1SM1-AThis text indicates that the instrument is an Industrial Scientific and Medical Group 1
Class A product (CISPER 11, Clause 4).
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is OFF or in STANDBY position.
vi Agilent R/Q/U/V281A, B Operating Note
Description
The Agilent R/Q/U/V 281 adapters have:
•Excellent repeatability
•Measurement v ersatility
•WR-28, WR-22, WR-19, WR-15 (R, Q, U, V band) waveguide
•Precision 1.85 mm and 2.4 mm coax connectors
Increased Measurement
Versatility
Excellent AccuracyLow SWR transitions reduce mismatch uncertainty and excellent
Coax measurements to 65 GHz. The Agilent 281 family of precision
millimeter waveguide-to-coax adapters offers a simple, low SWR transition
between transmission media. Take advantage of stable, low loss and low
SWR waveguide when measuring coaxial devices at higher frequencies. Or
keep the convenience of coax, even when testing waveguide components.
Agilent R/Q/U/V 281 adapters increase the versatility of existing test sets to
measure both coax and waveguide components.
repeatability ens ures consistent measurements. Available with both male and
female connectors, these adapters help microwave engineers combine the
best test components from each transmission media. With the Agilent
R/Q/C1/V 281 adapters, engineers can balance the merits of coax and
waveguide to achie ve the bes t combinat ion of ac curacy, cost and ease-of-use.
Agilent R/Q/U/V281A,B Operating Note 1
Dimensions
Description
A=19.3 mm (R281)
A=28.7 mm (Q281)
A=28.7 mm (U281)
Figure 1Dimensions in millimeters
2 Agilent R/Q/U/V281A,B Operating Note
Specifications
Specifications
Specifications describe the instrument’s warranted performance over the
temperature range 0 to 55
Characteristics ar e typi cal but non-war rante d performanc e pa rameter s. They
are denoted as “typical”, “nominal”, or “approximate”.
Coax-to-Waveguide AdaptersSupplemental Characteristics
Model Coax Connector
Type
R281A2.4 mm (female)WR-2826.5 to 40≥24 dB≤0.3 dB≥ −50 dB
R281B2.4 mm (male)WR-2826.5 to 40≥24 dB≤0.3 dB≥ −50 dB
Q281A2.4 mm (female)WR-2233 to 50≥22 dB≤0.3 dB≥ −50 dB
Waveguide
Number
Frequency
(GHz)
°C (except where noted). The Supplemental
1
Return Loss
Insertion Loss2Repeatability
Typically Better
Than:
1,3
Q281B2.4 mm (male)WR-2233 to 50≥22 dB≤0.3 dB≥ −50 dB
U281A1.85 mm (female)WR-1940 to 60≥22 dB
U281B1.85 mm (male)WR-1940 to 60≥22 dB
V281A1.85 mm (female)WR-1550 to 67
67 to 70
V281B1.85 mm (male)WR-1550 to 67
67 to 70
1. At the coaxial port only.
2. Typical maximum insertion loss for single adapter.
3. Repeatability = 20 Log
disconnect/connect cycle. Repeatability depend upon proper torque and pin-depth.
4. Typical for single adapter. Mated pair return loss warranted to ≥18 dB.
|∆Γ| where |∆Γ| =|Γm1–Γm2|. This is the difference between two measurements Γm
≥20 dB
≥20 dB(typical)
≥20 dB
≥20 dB(typical)
4
4
≤0.5 dB≥ −50 dB
≤0.5 dB≥ −50 dB
≤0.5 dB
≤0.5 dB
≤0.5 dB
≤0.5 dB
, and Γm2 before and after one
1
≥ −50 dB
≥ −50 dB
≥ −50 dB
≥ −50 dB
Agilent R/Q/U/V281A,B Operating Note 3
Specifications
EnvironmentNon-operating environmental specifications apply to storage and shipment.
Products should be stored in a clean, dry environment. Operating
environmental specifications apply when the product is in use.
CharacteristicNon-OperatingOperating
Temperature–55 to +75 °C0 to +55 °C
Humidity <95% relative at +40 °C<95% relative at +40 °C
Altitude ≤15,300 m (50,000 ft)<4,600 m (15,000 ft)
NOTE Storage or operation within an environment other than that specified above
may cause damage to the product and may void the warranty.
4 Agilent R/Q/U/V281A,B Operating Note
Connector Torque and
Pin Depth
Care, Connection and Torque
Care, Connection and Torque
Connector
Type
1.85 mm8 lb-in
2.4 mm8 lb-in
Torque
(±0.5 lb-in)
(90 N-cm)
(90 N-cm)
Torque Wrench Part
Number
1250-1863U281:0 to 0.03 mm
1250-18630 to 0.038 mm
Mating 2.4 mm
connectors with 1.85
mm connectors
Connector Pin Depth
(male or female)
(0 to 0.0012 in)
V281:0 to 0.0152 mm
(0 to 0.0006 in)
(0 to 0.0015 in)
Protection Gage Kit
Part Number
85056-60018 (male)
85056-60017 (female)
11752-60108 (male)
11752-60107 (female)
End Cap Part Number
1401-0202 (female)
1401-0095 (male)
1401-0202 (female)
1401-0095 (male)
The 1.85 mm coaxial end of the Agilent U/V281AB adapters can be mated
with 2.4 mm connectors.
Agilent R/Q/U/V281A,B Operating Note 5
Care, Connection and Torque
6 Agilent R/Q/U/V281A,B Operating Note