2VWorks Automation Control 14.0 Installation Guide
About computer requirements
The Agilent-supplied computer, if purchased, meets the computer requirements. The
computer is an HP Z2G4 or equivalent, with 16 GB RAM.
If you are installing the software on a different computer, refer to one of the following
for computer requirements:
•VWorks Software 14.0 Site Preparation Checklist
•VWorks Automation Control 14.0 Release Notes
Before you install the software
Before you start the installation, ensure that you have purchased a license for the
VWorks edition that you plan to install:
•VWorks Plus. In addition to the standard features, this edition includes features to
enable compliance with 21 CFR Part 11. All records and results are stored in
Content Management.
•VWorks Standard. The compliance-related features are unavailable in this edition.
The records and results are stored in the local file system.
Note: If you are upgrading from a previous version of Agilent OpenLab for VWorks, see
“Upgrading an installation from VWorks Standard to VWorks Plus” on page 20.
About computer requirements
Prerequisite checklist
ItemTask
1Decide on a computer name.
2If applicable, install your automation devices, including power and
The computer name will be reflected in the license. To avoid effort, Agilent
recommends that the computer name remain unchanged after installing
VWorks Plus or VWorks Standard.
IMPORTANT To make sure that a DNS server can resolve the computer
name, follow the internet standard for protocols (RFC952) and make sure the
name contains only approved characters (letters (a-z, A-Z), digits (0-9), and
hyphen (-)).
IMPORTANT Do not use an underscore! Installation is not possible if the
computer name contains an underscore.
communication cables.
VWorks Automation Control 14.0 Installation Guide
3
Before you install the software
ItemTask
3To enable the successful installation and activation of all components:
4If you use Trend Micro as an antivirus software, turn off Web Reputation to
5If applicable, uninstall any previous versions of the VWorks software. See
6Ensure the following:
•Ensure that the LAN-cable is connected and a network is available. A local
network is sufficient, such as the connection between the computer and
the automation device.
VWorks Plus only. If you install Agilent OpenLab for VWorks without this
connection, the activation of the Content Management component will
fail.
•Ensure that the Windows Sleep mode is set to Never.
allow the installation of all components.
“Uninstalling a previous VWorks version” on page 5.
•No Windows updates are performed during the installation.
•No system reboot is pending. Pending reboots are indicated in the
System Preparation Tool.
LAN power management
Microsoft Windows may be set to turn instruments/components off to save power
while sleeping or hibernating, which can interrupt VWorks processing.
To c h an g e t he W ind ows power management setting:
1In the Microsoft Control Panel, open the Network and Sharing Center.
Note: View the items by icon to see a list of all items
2Select Change adapter settings. Right-click Local Area Connection> Properties>
Configure.
3Select the Power Management tab.
4Clear the Allow the computer to turn off this device to save power check box.
4VWorks Automation Control 14.0 Installation Guide
Uninstalling a previous VWorks version
If the computer contains:
•A previous installation of VWorks Plus or VWorks Standard, see “Upgrading an
installation from VWorks Standard to VWorks Plus” on page 20 for instructions.
•VWorks 13.1.x or earlier versions, follow the instructions in this topic.
Before you uninstall
Create backup copies of the following items for safekeeping:
•Copy the C:\VWorks Workspace\ folder.
•Export the full set of Velocity11 registry keys to a file (.reg).
To export the Velocity11 registry keys:
1Open Registry Editor.
2In the Registry Editor window, right-click the following, and then click Export.
3In the Export Registry File dialog box, specify a file name and storage location, and
click Save.
Uninstalling a previous VWorks version
About migrating your VWorks files
Even though you must uninstall any previous VWorks versions before installing
VWorks 14.0 or later, the files and records created by the previous version remain
on the hard drive of the computer and will be accessible by the VWorks Migration
Wizard. The VWorks 14.0 Migration Wizard enables you to migrate the VWorks
protocols and associated files from VWorks 12.3, 13.0, or 13.1.x.
Uninstalling VWorks 13.1.x or earlier versions
Uninstall any previous versions of the following:
•Agilent Bravo IQOQ Test Utilities
•Agilent Bravo Metabolomics Workbench, Protein Sample Prep Workbench, and
BenchCel Workstations, if applicable
•Agilent Software Verification Tool
•VWorks software
To uninstall a previous version:
1In Windows, go to Settings > Apps & features.
2Select the program, for example, VWorks xx.x.x.xxx, and click Uninstall to uninstall
the selected software.
3Click Yes to allow the installer to make changes to your device.
After uninstalling the VWorks software, ensure that the following folder is not present.
If necessary, delete it.
Perform the following steps to install VWorks Plus or VWorks Standard 14.0.
StepFor this task...See...
1Prepare for the installation.“Before you install the software” on page 3
2Run the Agilent OpenLab for VWorks installer.“Installing Agilent OpenLab for VWorks” on page 6
3Run the Agilent VWorks 14.0 installer.“Installing VWorks 14.0 software” on page 12
4Get your license via SubscribeNet.
Log in to Control Panel and install your
license.
5Configure the software: Authentication,
Security policy, and User management
Administrator guide for your VWorks edition
Administrator guide for your VWorks edition
Installing Agilent OpenLab for VWorks
About the OpenLab for VWorks installer
This section describes a new installation of the software. For information on upgrading
an existing OpenLab for VWorks installation, see “Upgrading an installation from
VWorks Standard to VWorks Plus” on page 20.
The Agilent OpenLab for VWorks installation wizard installs and configures the
following required software tools and components:
•Control Panel and Shared Services
These components of the OpenLab software control the user access and storage
for the VWorks software.
•VWorks Plus only. Content Browser and Content Management
Note: Shared Services stores VWorks files in this secure repository.
•Agilent Software Verification Tool
Agilent support providers may use this tool to qualify the software installation.
•Database (PostgreSQL) and license tracking software (Flex-Net Publisher) required
by Shared Services
The installation wizard puts a Control Panel shortcut on the Windows desktop
and updates the Windows Start () menu under Agilent Technologies to include the
installed apps and utilities.
6VWorks Automation Control 14.0 Installation Guide
Installing Agilent OpenLab for VWorks
IMPORTANT
IMPORTANT
E
Do not attempt to install the software directly from the USB drive because the
installation may fail. Ensure that you install the software from C:\Temp\ on your local
computer.
To install Agilent OpenLab for VWorks:
1Copy the entire contents of the USB drive to the following location on your
computer and then remove the USB drive from the computer.
C:\Temp\
2Unzip the OpenLabForVWorks_2.5.0.927.zip archive to the following folder:
C:\Temp\
Installing Agilent OpenLab for VWorks
E
The installation wizard may trigger an reboot if the computer does not have all the
software prerequisites. Ensure that you save any work in process before continuing.
3Double-click the unzipped OpenLabForVWorks_2.5.0.927.exe file.
Note: If you have an antivirus software on your computer, it may block or
quarantine this installer. If that occurs, either disable the antivirus or restore the
file from the antivirus' quarantine to continue with the installation.