Agilent U2351A, U2352A, U2353A, U2354A, U2355A User guide

...

Agilent U2300A Series USB Multifunction Data Acquisition Devices

Service Guide
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2007 - 2010
Manual Part Number
U2351-90203
Edition
Second Edition, August 23, 2010
Printed in Malaysia
Agilent Technologies, Inc. Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
II U2300A Series DAQ Service Guide

Safety Information

The following general safety precautions must be observed during all phases of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the customer’s failure to comply with these requirements.
Safety Symbols
The following symbols indicate that precautions must be taken to maintain safe operation of the instrument.
Direct current
Warni ng
Regulatory Markings
The CE mark shows that the product complies with all the relevant European Legal Directives (if accompanied by a year, it signifies when the design was proven).
The CSA mark is a registered trademark of the Canadian Stan­dards Association. A CSA mark with the indicators "C" and "US" means that the product is certified for both the U.S. and Canadian markets, to the applicable American and Canadian standards.
The UL Mark is a registered trademark of Underwriters Laborato­ries Inc. UL listing mark with the indicators "C" and "US” indi­cates the product compliance with both Canadian and U.S. requirements.
The C-tick mark is a registered trademark of the Spectrum Man­agement Agency of Australia. This signifies compliance with the Australian EMC Framework regulations under the terms of the Radio Communications Act of 1992.
U2300A Series DAQ Service Guide III
General Safety Information
WARNING
CAUTION
• Do not use the device if it is damaged.Before you use the device, inspect the case. Look for cracks or missing plastic. Do not operate the device around explosive gas, vapor or dust.
• Do not apply more than the rated voltage (as marked on the device) between terminals, or between terminal and external ground.
• Always use the device with the cables provided.
• Observe all markings on the device before connecting to the device.
• Turn off the device and application system power before connecting to the I/O terminals.
• When servicing the device, use only specified replacement parts.
• Do not operate the device with the removable cover removed or loosened.
• Do not connect any cables and terminal block prior to performing self-test process.
• Use only the power adapter supplied by the manufacturer to avoid any unexpected hazards.
• Do not load the output terminals above the specified current limits. Applying excessive voltage or overloading the device will cause irreversible damage to the circuitry.
• Applying excessive voltage or overloading the input terminal will damage the device permanently.
• If the device is used in a manner not specified by the manufacturer, the protection provided by the device may be impaired.
• Always use dry cloth to clean the device. Do not use ethyl alcohol or any other volatile liquid to clean the device.
• Do not permit any blockage of the ventilation holes of the device.
IV U2300A Series DAQ Service Guide

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is shown as below:
Do not dispose in domestic household waste
To return this unwanted instrument, contact your nearest Agilent office, or visit:
http://www.agilent.com/environment/product
for more information.
U2300A Series DAQ Service Guide V

In This Guide...

1 Characteristics and Specifications 1
This chapter specifies the product specifications and electrical measurement specifications of the U2300A Series USB Multifunction DAQ devices.
2 Calibration 15
This chapter describes the step- by- step calibration procedures that covers hardware setup, self calibration, performance verification procedures and adjustment procedures.
3 Dismantle Procedures 41
This chapter shows the step- by- step disassemble procedures and list the available replacement parts together with its part number for U2300A Series USB Multifunction DAQ devices.
4 Troubleshooting and Self-Test Procedures 47
This chapter includes the information on general troubleshooting hints and self test procedures for the U2300A Series USB Multifunction DAQ devices.
A Connector Pins Configuration 51
All the U2300A Series USB Multifunction DAQ devices pins configuration are provided in the appendix.
B Test Limits 57
This appendix provides the test limits of analog input and analog output for the U2300A Series USB Multifunction DAQ devices.
VI U2300A Series DAQ Service Guide

Contents

Contents
1 Characteristics and Specifications 1
Product Specifications 2 Electrical Measurement Specifications 10
2 Calibration 15
Introduction 16 Types Of Signal Sources 17 Input Configurations 18 Hardware Connection 22 Self-Calibration 26 Adjustment Procedures 32 Performance Verification Procedure 34 A/D Data Conversion 38
3 Dismantle Procedures 41
General Disassemble 42 Replacement Parts 46
4 Troubleshooting and Self-Test Procedures 47
Troubleshooting 48 Self-Test Procedures 49
A Connector Pins Configuration 51
Pins Configuration for U2331A, U2355A and U2356A 52 Pins Configuration for U2351A, U2352A, U2353A and U2354A 53
B Tes t L im it s 57
Analog Input 58
U2300A Series DAQ Service Guide VII
Contents
Analog Output 62
VIII U2300A Series DAQ Service Guide
Agilent U2300A Series Multifunction USB DAQ Service Guide
1 Characteristics and Specifications
Product Specifications 2 Electrical Measurement Specifications 10
This chapter specifies the specifications of the U2300A DAQ devices.
Agilent Technologies
1
1 Characteristics and Specifications

Product Specifications

Basic Multifunction DAQ Device Specifications
Tab l e 1 - 1 Product specifications for basic multifunction DAQ device (U2351A,
U2352A, U2353A, and U2354A)
Analog Input
Model Number U2351A U2352A U2353A U2354A
Resolution 16 bits, no missing codes
Number of channels 16 SE/8 DI (software selectable/channel)
Maximum sampling rate 250 kSa/s 500 kSa/s
Scan list memory Up to 100 selectable channels entries
Programmable bipolar input range ±10 V, ±5 V, ±2.5 V, ±1.25 V
Programmable unipolar input range 0 to 10 V, 0 to 5 V, 0 to 2.5 V, 0 to 1.25 V
Input coupling DC Input impedance 1 GΩ / 100 pF
Operational common mode voltage range
Overvoltage protection Power on: Continuous ±30 V, Power off: Continuous ±15 V
Trig g e r s o u rc e s
Trigger modes Pre- trigger, delay-trigger, post-trigger and middle-trigger
FIFO buffer size Up to 8 MSa
External analog/digital trigger, SSI/star trigger
±7.5 V maximum
[4]
[1]
2 U2300A Series DAQ Service Guide
Characteristics and Specifications 1
Analog Output
Model Number U2351A U2352A U2353A U2354A
Resolution 16 bits N/A 16 bits N/A Number of channels 2 N/A 2 N/A Maximum update rate 1 MSa/s N/A 1 MSa/s N/A Output ranges 0 to 10 V, ±10 V,
0 to AO_EXT_REF,
±AO_EXT_REF
[2]
N/A 0 to 10 V, ±10 V,
0 to AO_EXT_REF,
±AO_EXT_REF
[2]
Output coupling DC N/A DC N/A Output impedance 0.1 Ω Typical N/A 0.1 Ω Typical N/A Stability Any passive load
up to 1500 pF
N/A Any passive load
up to 1500 pF Power-on state 0 V steady state N/A 0 V steady state N/A Trigger sources External
analog/digital trigger,
SSI/star trigger
[1]
Trigger modes Post-trigger and
delay-trigger
FIFO buffer size 1 channel: Maximum 8 MSa
2 channels: Maximum 4 MSa/ch
Function generation mode Sine-wave, square-wave, triangle,
sawtooth and noise waveform
N/A External
analog/digital trigger,
SSI/star trigger
[1]
N/A Post-trigger and
delay-trigger
N/A 1 channel: Maximum 8 MSa
2 channels: Maximum 4 MSa/ch
N/A Sine-wave, square-wave, triangle,
sawtooth and noise waveform
N/A
N/A
N/A
N/A
N/A
N/A
Digital I/O
Model Number U2351A | U2352A | U2353A | U2354A
Number of bits 24-bit programmable input/output Compatibility TTL Input voltage V
= 0.7 V maximum, IIL = 10 µA maximum
IL
VIH = 2.0 V minimum, IIH = 10 µA maximum
Input voltage range –0.5 V to +5.5 V Output voltage V
= 0.45 V maximum, IOL = 8 mA maximum
OL
= 2.4 V minimum, IOH = 400 µA maximum
V
OH
U2300A Series DAQ Service Guide 3
1 Characteristics and Specifications
General Purpose Digital Counter
Model Number U2351A | U2352A | U2353A | U2354A
Maximum count (2³¹-1) bits Number of channels Two independent up/down counter Compatibility TTL Clock source Internal or external Base clock available 48 MHz Maximum clock source frequency 12 MHz Input frequency range 0.1 Hz to 6 MHz at 50% duty cycle Pulse width measurement range 0.167 µs to 178.956 s
Analog Trigger
Model Number U2351A | U2352A | U2353A | U2354A
Trigger source All analog input channels, External analog trigger (EXTA_TRIG) Trigger level ±Full Scale for internal; ±10 V for external Trigger conditions Above high, below low and window (software selectable) Trigger level resolution 8 bits Bandwidth 400 kHz Input Impedance for EXTA_TRIG 20 kΩ Coupling DC Overvoltage Protection Continuous for ± 35 V maximum
Digital Trigger
Model Number U2351A | U2352A | U2353A | U2354A
Compatibility TTL/CMOS Response Rising or falling edge Pulse width 20 ns minimum
Calibration
[3]
Model Number U2351A | U2352A | U2353A | U2354A
On board reference voltage 5 V Temperature drift ±2 ppm/°C Stability ±6 ppm/1000 hours
4 U2300A Series DAQ Service Guide
Characteristics and Specifications 1
General
Model Number U2351A | U2352A | U2353A | U2354A
Remote interface Hi-Speed USB 2.0 Device class USBTMC Class Device Programmable interface Standard Commands for Programmable Instruments (SCPI) and IVI-COM
[1] System Synchronous Interface (SSI) and Star-trigger commands are used when modular devices are used in instrument chassis.
[2] Maximum external reference voltage for analog output (AO_EXT_REF) is ±10 V.
[3] 20 minutes warm-up time is recommended.
[4] Refer to Figure 1-1 for more information.
U2300A Series DAQ Service Guide 5
1 Characteristics and Specifications
7.5
-7.5
Figure 1-1 Operational common mode voltage range
This graph shows that the common mode voltage range is tightly linked with the output voltage. The output voltage range of the DAQ devices is ±10 V. Therefore, the common mode voltage range is ±7.5 V. Any operation beyond these voltage ranges may produce unexpected and unreliable results, and should be avoided.
6 U2300A Series DAQ Service Guide
Characteristics and Specifications 1
High Density Multifunction DAQ Device Specifications
Tab l e 1 - 2 Product specifications for high density multifunction DAQ device
(U2355A, U2356A and U2331A)
Analog Input Model Number U2355A U2356A U2331A
Resolution 16 bits, no missing codes 12 bits, no missing codes Number of channels 64 SE/32 DI (software selectable/channel) Maximum sampling rate 250 kSa/s 500 kSa/s 3 MSa/s (single channel)
1 MSa/s (multi channels) Scan list memory Up to 100 selectable channels entries Programmable bipolar input range ±10 V, ±5 V, ±2.5 V, ±1.25 V ±10 V, ±5 V, ±2.5 V,
±1.25 V, ±1 V, ±0.5 V,
±0 .25 V, ±0.2 V, ±0.0 5 V
Programmable unipolar input range 0 to 10 V, 0-5 V, 0-2.5 V, 0-1.25 V 0-10 V, 0-5 V, 0-4 V, 0-2.5 V, 0-2 V,
0-1 V, 0-0.5 V, 0-0.4 V, 0-0.1V
Input coupling DC Input impedance 1 GΩ / 100 pF Operational common mode voltage range ±7.5 V maximum Overvoltage protection Power on: Continuous ±30 V, Power off: Continuous ±15 V Trig g e r s o u rc e s
External analog/digital trigger, SSI/star trigger Trigger modes Pre-trigger, delay-trigger, post-trigger and middle-trigger FIFO buffer size Up to 8 MSa
[1]
Analog Output Model Number U2355A U2356A U2331A
Resolution 12 bits Number of channels 2 Maximum update rate 1 MSa/s Output ranges
0 to 10 V, ±10 V, 0 to AO_EXT_REF, ±AO_EXT_REF
[2]
Output coupling DC Output impedance 0.1 Ω Typical Stability Any passive load up to 1500 pF Power on state 0 V steady state Trigger sources
External analog/digital trigger, SSI/star trigger
[1]
Trigger modes Post-trigger and delay-trigger FIFO buffer size 1 channel: Maximum 8 MSa, 2 channels: Maximum 4 MSa/ch Function generation mode Sine-wave, square-wave, triangle, sawtooth and noise waveform
U2300A Series DAQ Service Guide 7
1 Characteristics and Specifications
Digital I/O Model Number U2355A | U2356A | U2331A
Number of bits 24-bit programmable input/output Compatibility TTL Input voltage V
Input voltage range –0.5 V to +5.5 V Output voltage V
General Purpose Digital Counter Model Number U2355A | U2356A | U2331A
Maximum count (2³¹-1) bits Number of channels Two independent up/down counter Compatibility TTL Clock source Internal or external Base clock available 48 MHz Maximum clock source frequency 12 MHz Input frequency range 0.1 Hz to 6 MHz at 50% duty cycle Pulse width measurement range 0.167 µs to 178.956 s
= 0.7 V max, IIL = 10 µA max
IL
= 2.0 V min, IIH = 10 µA max
V
IH
= 0.45 V max, I
OL
= 2.4 V min, IOH = 400 µA max
V
OH
= 8 mA max
OL
Analog trigger Model Number U2355A | U2356A | U2331A
Trigger source All analog input channels, External analog trigger (EXTA_TRIG) Trigger level ±Full Scale for internal; ±10 V for external Trigger conditions Above high, below low and window (software selectable) Trigger level resolution 8 Bits Bandwidth 400 kHz Input Impedance for EXTA_TRIG 20 kΩ Coupling DC Overvoltage Protection Continuous for ±35 V maximum
8 U2300A Series DAQ Service Guide
Characteristics and Specifications 1
Digital Trigger Model Number U2355A | U2356A | U2331A
Compatibility TTL/CMOS Response Rising or falling edge Pulse width 20 ns minimum
Calibration
[3]
Model Number U2355A | U2356A | U2331A
On board reference 5 V Temperature drift ±2 ppm/°C Stability ±6 ppm/1000 hours
General Model Number U2355A | U2356A | U2331A
Remote interface Hi-Speed USB 2.0 Device class USBTMC Class Device Programmable interface Standard Commands for Programmable Instruments (SCPI) and IVI-COM
[1] System Synchronous Interface (SSI) and Star-trigger commands are used when modular devices are used in instrument chassis.
[2] Maximum external reference voltage for analog output (AO_EXT_REF) is ±10 V.
[3] 20 minutes warm-up time is recommended.
U2300A Series DAQ Service Guide 9
1 Characteristics and Specifications

Electrical Measurement Specifications

Basic Multifunction USB DAQ Device
Analog Input Measurement
[1]
Model Number U2351A | U2352A U2353A | U2354A
Function 23 °C ± 5 °C
0 °C to 18 °C
28 °C to 45 °C
23°C ± 5°C
0°C to 18°C
28°C to 45°C
Offset Error ±1 mV ±5 mV ±1 mV ±5 mV Gain Error ±2 mV ±5 mV ±2 mV ±5 mV
–3 dB small signal bandwidth
1% THD large signal bandwidth
[2]
[2]
760 kHz 1.5 MHz
300 kHz 300 kHz
System noise 1 mVrms 2 mVrms 1 mVrms 2.5 mVrms CMRR 62 dB 62 dB
Spurious-free dynamic range (SFDR)
[3]
Signal-to-noise and distortion ratio (SINAD)
Total harmonic distortion (THD)
Signal-to-noise ratio (SNR)
Effective number of bits (ENOB)
Analog Output Measurement
[3]
[3]
[3]
[1]
[3]
88 dB 82 dB
80 dB 78 dB
–90 dB –82 dB
80 dB 78 dB
13 12.6
Model Number U2351A | U2353A
0 °C to 18 °C
Function 23 °C ± 5 °C
28 °C to 45 °C
Offset Error ±1 mV ±4 mV Gain Error ±4 mV ±5 mV Slew rate 19 V/µs Rise time 0.9 µs Fall time 0.9 µs Settling time to 1% output error 4 µs Driving capability 5 mA Glitch energy 5 ns-V (Typical), 80 ns-V (Maximum)
10 U2300A Series DAQ Service Guide
Characteristics and Specifications 1
[1] Specifications are for 20 minutes of warm-up time, calibration temperature at 23 °C and
input range of ±10 V.
[2] Specifications are based on the following test conditions
Bandwidth Test Model Number Test Conditions
–3 dB small
signal bandwidth
1% THD large signal bandwidth
U2351A U2352A
U2353A U2354A
(DUT setting at ±10 V bipolar)
Sampling Rate: Input voltage:
–3 dB small signal bandwidth
1% THD large signal
bandwidth
Sampling Rate: Input voltage:
–3 dB small signal bandwidth
1% THD large signal
bandwidth
[3] Specifications are based on the following test conditions.
Dynamic Range Test Model Number Test Conditions
(DUT setting at ±10 V bipolar)
SFDR, THD, SINAD, SNR, ENOB
U2351A U2352A
U2353A U2354A
Sampling Rate: Fundamental Frequency: Number of points: Fundamental input voltage: Sampling Rate: Fundamental Frequency: Number of points: Fundamental input voltage:
.
250 kSa/s
10% FSR FSR –1 dB FS
500 kSa/s
10% FSR FSR –1 dB FS
250 kSa/s
2.4109 kHz 8192 FSR –1 dB FS
500 kSa/s
4.974 kHz 16384 FSR –1 dB FS
U2300A Series DAQ Service Guide 11
1 Characteristics and Specifications
High Density Multifunction USB DAQ Device
Analog Input Measurement
[1]
Model Number U2355A U2356A U2331A
Function 23 °C ± 5 °C
0 °C - 18 °C
28 °C - 45 °C
23 °C ± 5 °C
0 °C to 18 °C
28 °C to 45 °C
23 °C ± 5 °C
0 °C to 18 °C
28 °C to 45 °C
Offset Error ±1 mV ±2 mV ±1 mV ±2 mV ±2 mV ±3 mV Gain Error ±2 mV ±3 mV ±2 mV ±6 mV ±6 mV ±7.5 mV
–3 dB small signal bandwidth
1% THD large signal bandwidth
[2]
[2]
760 kHz 1.3 MHz 1.2 MHz
400 kHz 400 kHz N/A
System noise 1 mVrms 2 mVrms 1 mVrms 4 mVrms 3 mVrms 5 mVrms CMRR 64 dB 61 dB 62 dB Spurious-free dynamic range
[3]
(SFDR) Signal-to-noise and distortion
ratio (SINAD)
[3]
Total harmonic distortion (THD)
Signal-to-noise ratio (SNR)
[3]
Effective number of bits
[3]
(ENOB)
Analog Output Measurement
[3]
[1]
88 dB 86 dB 71 dB
80 dB 78 dB 72 dB
–90 dB –84 dB –76 dB
80 dB 78 dB 72 dB
13 12.6 11.6
Model Number U2355A | U2356A U2331A
Function 23 °C ± 5 °C
0 °C to 18 °C
28 °C to 45 °C 23 °C ± 5 °C
0 °C to 18 °C
28 °C to 45 °C
Offset Error ±1 mV ±4 mV ±1.5 mV ±3mV Gain Error ±4 mV ±5 mV ±4 mV ±5 mV Slew rate 19 V/µs 19 V/µs Rise time 0.9 µs 0.9 µs Fall time 0.9 µs 0.9 µs Settling time to 1% output error 4 µs 4 µs Driving capability 5 mA 5 mA Glitch energy 5 ns-V (Typical),
80 ns-V (Maximum)
5 ns-V (Typical),
80 ns-V (Maximum)
12 U2300A Series DAQ Service Guide
Characteristics and Specifications 1
[1] Specifications are for 20 minutes of warm-up time, calibration temperature at 23 °C and
input range of ±10 V.
[2] Specifications are based on the following test conditions
Bandwidth Test Model Number Test Conditions
–3 dB small
signal bandwidth
1% THD large signal bandwidth
U2355A Sampling Rate:
U2356A Sampling Rate:
U2331A Sampling Rate:
(DUT setting at ±10 V bipolar)
Input voltage:
–3 dB small signal bandwidth
1% THD large signal
bandwidth
Input voltage:
–3 dB small signal bandwidth
1% THD large signal
bandwidth
Input voltage:
–3 dB small signal bandwidth
1% THD large signal
bandwidth
[3] Specifications are based on the following test conditions.
Dynamic Range Test Model Number Test Conditions
(DUT setting at ±10 V bipolar)
SFDR, THD, SINAD, SNR, ENOB
U2355A Sampling Rate:
Fundamental Frequency: Number of points: Fundamental input voltage:
U2356A Sampling Rate:
Fundamental Frequency: Number of points: Fundamental input voltage:
U2331A Sampling Rate:
Fundamental Frequency: Number of points: Fundamental input voltage:
.
250 kSa/s
10% FSR FSR –1 dB FS
500 kSa/s
10% FSR FSR –1 dB FS
3 MSa/s
10% FSR FSR –1 dB FS
250 kSa/s
2.4109 kHz 8192 FSR –1 dB FS
500 kSa/s
4.974 kHz 16384 FSR –1 dB FS
3 MSa/s
29.892 kHz 65536 FSR –1 dB FS
U2300A Series DAQ Service Guide 13
1 Characteristics and Specifications
14 U2300A Series DAQ Service Guide
Agilent U2300A Series USB Multifunction DAQ Service Guide
2 Calibration
Introduction 16 Types Of Signal Sources 17
Ground-referenced signal sources 17 Floating signal sources 17
Input Configurations 18
Single-ended connections 18
Differential Input Mode 20 Hardware Connection 22 Self-Calibration 26 Adjustment Procedures 32 Performance Verification Procedure 34 A/D Data Conversion 38
This chapter includes the equipments required for the calibration procedure and describe the hardware connection for analog input and analog output. The self- calibration procedure, performance verification procdure and adjustment procedure are also included.
Agilent Technologies
15
2 Calibration

Introduction

Prior to checking the performace of the instrument, ensure that you have all the equipments listed in the following to perform the verification procedure for analog input and analog output.
Equipment Checklist
Agilent U2300A Series USB Multifunction DAQ devices
Agilent Digital Multimeter (DMM) 34401A/34410A
Fluke Calibrator 5520A/5720A
Computer with Agilent IO Library Suites installed
U2901A Terminal Block
USB Mini- B Cable
U2901A SCSI Cable
Follow the following steps to ensure the instrument gives accurate analog input readings and analog outputs.
1 Perform the self- calibration procedure. See
Self- Calibration”.
2 Perform the performance verification procedure. See
Performance Verification Procedure”, “Analog Input
Connection” for information on analog input connection
and “Analog Output Connection” for information on analog output connection.
3 If the instrument gives accurate readings or accurate
outputs, adjustment procedure is not necessary. If the instrument does not give accurate readings or accurate outputs, perform the adjustment procedure. See “Adjustment Procedures”.
4 After perform the adjustment procedure, repeat the
self- calibration procedure and then perform the verification procedure.
5 If the instrument still does not give accurate readings or
accurate outputs, repeat step 3 and step 4.
16 U2300A Series DAQ Service Guide

Types Of Signal Sources

Ground-referenced signal sources
A ground- referenced signal source is defined as a signal source that is connected in some way to the building’s grounding system. This means that the signal source is connected to a common ground point with respect to the U2300A series DAQ (assume the host PC which is connected with DAQ is in the same power ground).
Floating signal sources
A floating signal source is a signal that is not connected to the building’s grounding system. It is also a device with an isolated output. Example of floating signal sources are optical isolator output, transformer output, and thermocouple.
Calibration 2
U2300A Series DAQ Service Guide 17
2 Calibration

Input Configurations

Single-ended connections
A single- ended connection is applicable when the analog input signal is referenced to a ground and can be shared with other analog input signals. There are two different types of single- ended connections, which are RSE and NRSE configuration.
• Referenced Single- Ended (RSE) mode
In referenced single- ended mode, all the input signals are connected to the ground provided by the U2300A series DAQ and suitable for connections with floating signal sources. The following figure illustrates the RSE mode.
Figure 2-1 Floating source and RSE input connections
NOTE
18 U2300A Series DAQ Service Guide
When more than two floating sources are connected, these sources are referenced to the same common ground.
Calibration 2
• Non- Referenced Single- Ended (NRSE) Mode
In NRSE mode, the DAQ device does not provide the grounding point. The ground reference point is provided by the external analog input signal. You can connect the signals in NRSE mode to measure ground- referenced signal sources, which are connected to the same grounding point. The following figure illustrates the connection. The signal local ground reference is connected to the negative input of the instrumentation Amplifier (AI_SENSE pin on connector1). Hence, any potential difference of the common mode ground between signal ground and the signal ground on DAQ board will be rejected by the instrumentation amplifier.
Figure 2-2 Ground-referenced sources and NRSE input connections
U2300A Series DAQ Service Guide 19
2 Calibration
Differential Input Mode
The differential input mode provides two inputs that respond to the difference of the signal voltage. The analog input of the U2300A series DAQ has its own reference ground or signal return path. The differential mode can be used for the common- mode noise rejection if the signal source is ground- referenced. The following figure shows the connection of ground- referenced signal sources under differential input mode.
Figure 2-3 Ground-referenced source and differential input mode
20 U2300A Series DAQ Service Guide
Calibration 2
The following figure illustrates the connection of a floating signal source to the U2300A series DAQ in differential input mode. For floating signal sources, additional resistor is needed at each channel to provide a bias return path. The resistor value is equivalent to about 100 times the source impedance. If the source impedance is less than 100 W, you can connect the negative polarity of the signal directly to AI_GND, as well as the negative input of the Instrumentation Amplifier. The noise couples in differential input mode are less compared to the single- ended mode.
Figure 2-4 Floating source and differential input
NOTE
U2300A Series DAQ Service Guide 21
Agilent U2300A series DAQ is designed with high input impedance.
Please ensure that all the connection are connected properly before acquiring any data. Failing to do so may cause data fluctuation or erroneous readings.
Unused pins at multiplexing DAQ inputs can be treated as floating
source with infinite output impedance. Therefore, necessary grounding system is required in user application system.
2 Calibration

Hardware Connection

The connection to verify the analog input readings and analog outputs are different. To verify the analog input readings, see “Analog Input Connection” for descriptions on the way to connect the instruments. To verify the analog outputs, see “Analog Output Connection” for descriptions on hardware setup.
Analog Input Connection
The equipments required for analog input connection are the DAQ device, Fluke calibrator, terminal block, USB mini- B cable and SCSI cable. Follow the follwing step- by- step instruction for analog input connection.
1 Connect the DAQ device to a PC with a USB mini- B cable
and connect the DAQ device to a U2901A terminal block using the U2901A SCSI cable.
NOTE
22 U2300A Series DAQ Service Guide
Ensure that the PC has the DAQ device’s driver and the Agilent IO
Libraries 14.2 or higher installed. Note that the Agilent Measurement Manager software comes with the standard purchase of the U2300A Series DAQ devices.
If you do not have the DAQ device’s driver and the Agilent IO Libraries
14.2 or higher installed, refer to the Agilent U2300A Series Data
Acquisition Devices and Agilent Measurement Manager Quick Start Guide, for more information on the installation.
2 Short all the DAQ device’s analog input channels. Refer to
Connector Pins Configuration” for the pins assignment.
3 Connect the DAQ device’s analog input channels to the
calibrator output.
4 Connect the DAQ device’s analog input ground (AI_GND)
to the calibrator ground. See Figure 2- 1. Note that the analog inputs shown in the figure uses U2351A as an example.
Calibration 2
DAQ
AI101
...
AI164
AIGND
Calibrator
Output
GND
Figure 2-5 The analog input and calibrator connection
Setup diagram
DAQ
SCSI Cable
Mini-B
USB Cable
Terminal Block
(e.g. AI101 to AI164
Analog Input
for U2351A)
PC
Calibrator
Figure 2-6 The analog input and calibrator connection
U2300A Series DAQ Service Guide 23
2 Calibration
Analog Output Connection
The equipments required for analog output connection are the DAQ device, DMM, terminal block, USB mini- B cable and SCSI cable. Follow the follwing step- by- step instruction for analog input connection.
1 Connect the DAQ device to a PC with a USB mini- B cable
and connect the DAQ device to a U2901A terminal block using the U2901A SCSI cable.
NOTE
Ensure that the PC has the DAQ device’s driver and the Agilent IO
Libraries 14.2 or higher installed. Note that the Agilent Measurement Manager software comes with the standard purchase of the U2300A Series DAQ devices.
If you do not have the DAQ device’s driver and the Agilent IO Libraries
14.2 or higher installed, refer to the Agilent U2300A Series Data
Acquisition Devices and Agilent Measurement Manager Quick Start Guide, for more information on the installation.
2 Connect a wire at each pin you would like to test on the
U2901A terminal block. (Refer to “Connector Pins
Configuration” for the pins assignment)
3 Connect the DAQ device’s analog output channel 1
(AO201) to the DMM input.
4 Connect the DAQ device’s analog output ground
(AO_GND) to DMM GND.
5 Repeat steps 3 and 4 for analog output channel 2
(AO202).
24 U2300A Series DAQ Service Guide
Calibration 2
DAQ
AO201
AOGND
DMM
Input
GND
Figure 2-7 The analog input and DMM connection
Setup diagram
DAQ
SCSI Cable
Mini-B
USB Cable
Terminal Block
Analog Output
(e.g. AO201 to AO202
for U2351A)
PC DMM
Figure 2-8 The analog output and DMM connection
U2300A Series DAQ Service Guide 25
2 Calibration

Self-Calibration

Self- calibration can be operated using the following SCPI command via Agilent Connection Expert.
CALibration:BEGin
In calibration mode, the command will initiate a voltage adjustment in sequence for the specified Digital Analog Converter (DAC) channel. This sequence sets a zero and gain adjustment constant for each DAC output.
The function of DAQ device will not carry on until the self- calibration has completed. You can query the status of the self- calibration performed using the following SCPI command.
*OPC?
Two ways of performing the self- calibration will be introduced in this section. The first option is to use the Agilent Connection Expert to send the SCPI commands and the second option is to use the Agilent Measurement Manager application software.
Option 1: Self-calibration with Agilent Connection Expert
WARNING
• Unplug all cables that are connected to the DAQ device before
performing self-calibration.
• Any cables connected to the DAQ device may cause the failure of
the self-calibration process.
1 Power on the DAQ and disconnect all connections from
DAQ device. Warm it up for 30 minutes to ensure that it is operating at stable condition.
2 Go to Start > All Programs > Agilent IO Libraries Suite > Agilent
Connection Expert to launch the Agilent Connection Expert.
26 U2300A Series DAQ Service Guide
Calibration 2
3 Connect the DAQ device to the PC with mini- B type USB
cable. The connected DAQ device will be visible in the
Instrument I/O on this PC panel as illustrated in Figure 2- 5.
4 Select the DAQ device that you wish to send the SCPI
commands to and then click the Interactive IO icon on the toolbar to launch the Agilent Interactive IO. See Figure 2- 5.
Step 2: Click the Interactive IO icon
Step 1:
Select the instrument
Figure 2-9 Launch the Interactive IO in Agilent Connection Expert
5 The Agilent Interactive IO dialog box will appear as shown in
Figure 2- 6. Click Send & Read to send the “*IDN?” default command. This instrument’s response should appear in the Instrument Session History panel.
U2300A Series DAQ Service Guide 27
2 Calibration
6 Successful communication between the Agilent Connection
Expert and the connected hardware will be shown in the Instrument Session History panel. The users may now send other SCPI commands to the instrument.
Figure 2-10 Interactive IO dialog box
7 Ensure that the DAQ device has been warmed up for 30
minutes. Send the SCPI commands “*RST” and “*CLS” to clear the register in DAQ device.
8 Send “CAL:BEG” to start the self- calibration process. This
process may take a few minutes to complete.
9 Send “*OPC?” to check the operation complete status.
10 If “*OPC?” return 1, send “SYST:ERR?” to check if any
system error has occurred during the self- calibration process. If there is no system error, the self- calibration process is done. Otherwise, the self- calibration process is failed.
28 U2300A Series DAQ Service Guide
Calibration 2
Option 2: Self-calibration with Agilent Measurement Manager
WARNING
• Unplug all cables that are connected to the DAQ device before
performing self-calibration.
• Any cables connected to the DAQ device may cause the failure of
the self-calibration process.
1 Power on the DAQ device and disconnect all connections
from it. Warm it up for 30 minutes to ensure that it is operating at stable condition.
2 Connect the DAQ device to the PC with mini- B type USB
cable. Launch the Agilent Measurement Manager and select the DAQ device you wish to do the self-calibration process.
3 Go to Tools and select Self Calibration.
4 The Self Calibration Form dialog box will appear as shown
below.
Figure 2-11 Self Calibration Form dialog box in Agilent Measurement
Manager
U2300A Series DAQ Service Guide 29
2 Calibration
5 Select the instrument that you would like to perform
self- calibration and the Start button will be enabled. Click Start to proceed. See Figure 2- 8.
6 The calibration process will take a few minutes to be
completed. Once done, the status and results of the process will be displayed as shown in Figure 2- 9.
Figure 2-12 Self Calibration Form dialog box in Agilent Measurement
Manager with a device being selected
30 U2300A Series DAQ Service Guide
Calibration 2
Figure 2-13 Self Calibration Form dialog box in Agilent Measurement Ma-
naer showing the status and resultof the self-calibration pro­cess
U2300A Series DAQ Service Guide 31
2 Calibration

Adjustment Procedures

DAQ on- board 5 V calibration
If in the performance verification procedure is not accurate, adjustment procedure is required. The following flowchart shows the steps for 5 V calibration.
Step 1:
Open the DAQ device metal casing. Power on DAQ device and warm up for 30 minutes. Connect the DMM to U37 on the daughter board.
Step 2: 5 V calibration
Connect the DAQ device to the computer with USB cable.
Monitor the DMM readings.
Tune the 5 V until you get the voltage in between 4.9999 to 5.0001 V.
Send “CAL:VREF:TUNE MINUS” (to decrease voltage),
Send “CAL:VREF:TUNE PLUS” (to increase voltage),
Send “CAL:VREF” (to save to EPROM)
Step 3:
After save to EPROM, power cycle the unit.
Step 4:
Reconfirm the 5 V reading.
Figure 2-14 The flowchart showing the steps for 5 V calibration
32 U2300A Series DAQ Service Guide
Calibration 2
Tab l e 2 - 3 Step-by-step descriptions for Figure 2-10
Step Descriptions
Step 1 Before power on the DAQ device, open the DAQ device’s metal casing. Refer to “Dismantle Procedures” for
more information to disassemble the unit. Then, power on the DAQ device and warm it up for approximately 30 minutes. Connect the DMM to U37 on the daughter board. The following images shows the location of U37 and the way to connect it.
U37
Step 2 Ensure that the DAQ device’s driver is installed in the computer before connecting it to the computer. The driver
can be obtained from the Agilent USB Modular Instrument U2300A & U2700A Series Product Reference CD-ROM. Measure the voltage difference at U37 with the DMM. If the measured voltage is more than 5 V, send the command “
CAL:VREF:TUNE MINUS” to decrease the voltage, otherwise if the measured voltage is less than 5 V, send the command “CAL:VREF:TUNE PLUS” to increase the voltage. Tune the voltage until the measured voltage is in the range of 4.9999 V and 5.0001 V. When the tuning process is done, send “CAL:VREF” to save it in the EPROM.
Step 3 Power cycle the DAQ device by turning on and off the DAQ device.
Step 4 Measure the voltage at U37 again to ensure the measured voltage is in the range of 4.9999V to 5.0001 V. If there
is any problem occurs, the DMM will not measure the voltage in the specified range. Proceed to self calibration procedure.
U2300A Series DAQ Service Guide 33
2 Calibration

Performance Verification Procedure

Prior to calibrate the instrument, check the performance of the instrument to see if any adjustment is required, “Step 2:
DAQ configuration and verification test”.
This is to ensure that the DAQ device gives accurate readings or outputs. The performance verification procedure for analog input and analog output are automated and are verify using the Automated Calibration software.
The performance verification procedure involve two steps:
Step 1: DAQ self- calibration
Step 2: DAQ configuration and verification test
Analog Output
34 U2300A Series DAQ Service Guide
Calibration 2
Step 1: DAQ self- calibration
Step 1:
Power on DAQ device and disconnect all connections from DAQ. Warm up for 30 minutes.
Step 2: Reset DAQ
Send “*RST”
Step 3: Clear DAQ
Send “*CLS”
Step 4:
Wait: 1 second
Step 5: Start self-calibration
Send “CAL:BEG”
Wait: 60 seconds
Step 6: Check operation complete status
Send “*OPC?”
No
U2300A Series DAQ Service Guide 35
OPC status = 1?
Ye s
Step 7: Read system error
“SYST:ERR?”
System error?
No
Calibration DONE!
Ye s
Calibration FAIL!
2 Calibration
Step 2: DAQ configuration and verification test
Step 1:
Step 1:
Power on DAQ device and connect the calibrator to DAQ. Warm up for
Power on DAQ device and connect the calibrator to DAQ. Warm up for 30 minutes.
30 minutes.
Step 2: DAQ configuration
Step 2: DAQ configuration
Send “ROUT:SCAN(@Channel))”, Send “ROUT:CHAN:RANG
Send “ROUT:SCAN(@Channel)”, Send “ROUT:CHAN:RANG Range,(Channel))”, Send “ROUT:CHAN:RANG Range,((@Channel)),
Range,(@Channel)”, Send “ROUT:CHAN:POL Polarity,(@Channel)”, Send “ACQ:SRAT MaxSamplingRate”
Send “ACQ:SRAT MaxSamplingRate”
Step 3: Calibrator
Step 3: Calibrator
Source the desired test voltage to DAQ
Source the desired test voltage to DAQ
Step 4: DAQ data acquisition
Send “ACQ:POIN 100”, Send “DIG”
Wait: 0.2 second
Step 5: Check waveform complete status
Send “WAV:COMP?”
No
Step 6: DAQ fetch data
Send “WAV:DATA?” Read data as int 16 array
Step 7: Data conversion
Convert the data into voltage and get the average value for 100 points (please see Data conversion section for the formula)
Repeat steps above for all channels, polarities and ranges The test points are shown as table
36 U2300A Series DAQ Service Guide
WAV complete status = YES?
Ye s
Compare the reading with the test specification
Calibration 2
NOTE
Byte ordering must use LSB first.
Analog Output
Step 1:
Power on DAQ device and conenct the DMM to DAQ. Warm up for 30 minutes
Step 2: DAQ configuration
Send “SOUR:VOLT:POL Polarity,(@Channel)”, Send “SOUR:VOLT”RSRC INT,(@Channel)”, Send “SOUR:VOLT Voltage,(@Channel)”
Step 3: DMM
Measure the output voltage of DAQ
Repeat steps above for all channels and polarities. The test points are shown as table
Compare the reading with the test specification
U2300A Series DAQ Service Guide 37
2 Calibration

A/D Data Conversion

A/D data conversion converts analog voltage into digital information. The following section illustrates the format of acquired raw data for the A/D conversion.
Below is the illustrated example of the acquired raw data scan list for CH 101, CH 102, and CH 103.
#800000200 <byte> <byte> <byte> <byte> <byte> <byte> <byte> <byte> ...
Data length indicator, The next 8 bytes (0000 0200) specifying the actual data length only, not actual data. Data length (200 bytes long)
1st data
LSB
1st data
MSB
CH 101 CH 102 CH 103 CH 101 ...
1st data
LSB
1st data
MSB
1st data
LSB
1st data
MSB
2nd data
LSB
16-bit Data Format
LSB MSB
DDDD DDDD DDDD DDDD
2nd data
MSB
...
12-bit Data Format
LSB MSB
DDDD XXXX DDDD DDDD
D - Data bits
X - Unused bits
Raw data conversion
To convert the data into an actual float number, we need the voltage range and polarity information. Below are the calculations on the raw data conversion for both bipolar and unipolar.
To perform a sample calculation of the conversion, take the U2356A as an example. The resolution of U2356A is 16 bits and the range is taken as 10 V. The Int16b value calculated using conversion algorithm is 12768.
38 U2300A Series DAQ Service Guide
Calibration 2
Hence, the 16 bits binary read back calculation will be as follows.
LSB MSB
<11100000> <00110001>
= 12768
NOTE
NOTE
The raw data provided by U2300A series DAQ devices is in the byte order of LSB first.
Bipolar:
2 Int16 value×
⎛⎞
Converted value
Example of converted value
--------------------------------------
⎝⎠
resolution
2
⎛⎞ ⎝⎠
Range×=
2 12768×
------------------------
16
2
10× 3.896 V==
Unipolar:
Int16 value
⎛⎞
Converted value
----------------------------
⎝⎠
resolution
2
Example of converted value
The converted value is of float type. As such, you may need to type cast the Int16 value to float in your programming environment.
0.5+
Range×=
12768
⎛⎞
---------------
⎝⎠
2
16
0.5+
10× 6.948 V==
To perform a sample calculation of the conversion, take the U2331A as an example. The resolution of U2331A is 12 bits and the range is taken as 10 V. The Int12b value calculated using conversion algorithm is 12768.
LSB MSB
<11100000> <00110001>
= 12768
U2300A Series DAQ Service Guide 39
2 Calibration
There are unused bits in the 12- bit data format. Therefore, there is a need to perform a 4- bit right shift operation. Hence, the 12 bits binary read back calculation will be as follows.
LSB MSB
<00011110> <00000011>
= 798
NOTE
NOTE
The raw data provided by U2300A series DAQ devices is in the byte order of LSB first.
Bipolar:
2 Int16 value×
⎛⎞
Converted value
Example of converted value
--------------------------------------
⎝⎠
resolution
2
Range×=
2798×
⎛⎞
------------------
⎝⎠
12
2
10× 3.896 V==
Unipolar:
Int16 value
⎛⎞
Converted value
----------------------------
⎝⎠
resolution
2
Example of converted value
The converted value is of float type. As such, you may need to type cast the Int12 value to float in your programming environment.
0.5+
Range×=
798
⎛⎞
---------
0.5+
⎝⎠
12
2
10× 6.948 V==
For the U2331A, there is a need to perform a 4-bit right shift operation. This is because it is equipped with 12-bit ADC, and the last 4 bits are truncated.
40 U2300A Series DAQ Service Guide
Agilent U2300A Series USB Multifunction DAQ Service Guide
3 Dismantle Procedures
General Disassemble 42
Mechanical Disassemble 42
Replacement Parts 46
This chapter describes the step- by- step disassemble procedures and list the available replacement parts for U2300A Series DAQ devices,
Agilent Technologies
41
3 Dismantle Procedures

General Disassemble

NOTE
This chapter provides the step- by- step guides on how to dismantle the module and install the replacement assembly. To assemble back the module, follow the instructions in reverse order.
The parts shown in the following figures are representative and may look different than what you have in your module.
The removable assemblies include:
Plastic casing
Metal casing
Rear metal casing
Front metal casing, which is attached to the carrier board
and measurement board
Mechanical Disassemble
Follow the instructions in this section for the instrument disassemble process.
Step 1: Pull the bumper out to remove the plastic casing.
Plastic casing
42 U2300A Series DAQ Service Guide
Dismantle Procedures 3
Step 2: Flip the plastic casing open.
Step 3: Slide the metal casing out of the plastic casing.
Metal casing
U2300A Series DAQ Service Guide 43
3 Dismantle Procedures
Front metal piece
Step 4: Unscrew all the following indicated screws from metal casing.
Step 5: Gently pull the front metal piece out, which is attached to the carrier and measurement boards.
Step 6: Unscrew all the following indicated screws from the metal casing and remove the rear metal piece.
Rear metal casing
44 U2300A Series DAQ Service Guide
Metal casing
Rear metal casing
Carrier board and
measurement board
Dismantle Procedures 3
Disassembled parts:
Front
metal casing
U2300A Series DAQ Service Guide 45
3 Dismantle Procedures

Replacement Parts

This section provides the information of orderable replacement parts for U2300A Series DAQ devices. The parts availabled for replacement are listed in the table below with the reference part numbers and the respective part names.
You can order the replacement parts from Agilent using the part number provided in the table below. To order replacement parts from Agilent, do the following:
1 Identify the Agilent part number of the required parts as
shown in the replacement parts list.
2 Contact your nearest Agilent Sales Office or Service Center.
3 Provide the instrument model number and serial number.
Tab l e 3 - 4 Replacement parts list
Part Number Part Name
U2351-60002 Metal Casing
5190-0013 Plastic Cover
U2351-00201 Rear Metal Piece
U2351-61201 L-shape Mount
46 U2300A Series DAQ Service Guide
Agilent U2300A Series USB Multifunction DAQ Service Guide
4 Troubleshooting and Self-Test Procedures
Troubleshooting 48 Self-Test Procedures 49
This chapter provides the information on general troubleshooting and self-test procedures.
Agilent Technologies
47
4 Troubleshooting and Self-Test Procedures

Troubleshooting

This section provides suggestions for solving general problems that you may encounter with the instrument. It guides you on what to check in the following situations:
1. Power Indicator LED is not lit
Verify that the ac power cord is connected to the power inlet in the DAQ device.
2. Power Indicator LED is lit but the AO/ AI Indicator LED is not lit
Verify that the USB cable is connected to the PC and the USB inlet in the DAQ device.
3. Power Indicator LED is lit and AO/ AI Indicator LED is lit
Verify if the SCPI commands are correct with “SYSTem:ERRor?” command.
Refer to U2300A Series USB Multifunction Programming Guide for SCPI error messages.
NOTE
48 U2300A Series DAQ Service Guide
If there are no response from the instrument, contact the nearest Agilent Service Center to obtain further assistance.

Self-Test Procedures

Troubleshooting and Self-Test Procedures 4
WARNING
4 Go to Start > All Programs > Agilent IO Libraries Suite > Agilent
5 Go to Start > All Programs > Agilent T&M Toolkit > Agilent Interactive IO
6 Send the SCPI command “*TST?” to the instrument to start
7 The command will return either "+0" to indicate all tests passes
8 If the command returns "+1" , apply SCPI command
NOTE
Do not connect any cables and terminal block prior to performing self-test procedures.
Connection Expert to launch the Agilent Connection Expert.
to launch the Interactive I/O dialog box.
perform the self- test of the instrument.
or "+1" to indicate one or more tests failed.
SYSTem:ERRor?” to enquire the error message.
Refer to Agilent U2300A Series USB Multifunction Data Acquisition Programming Guide for SCPI error messages.
U2300A Series DAQ Service Guide 49
4 Troubleshooting and Self-Test Procedures
50 U2300A Series DAQ Service Guide
Agilent U2300A Series USB Multifunction DAQ Service Guide
A Connector Pins Configuration
Pins Configuration for U2331A, U2355A and U2356A 52 Pins Configuration for U2351A, U2352A, U2353A and U2354A 53
This appendix attached the pins configuration for all the U2300A Series DAQ devices.
Agilent Technologies
51
A Connector Pins Configuration

Pins Configuration for U2331A, U2355A and U2356A

The U2300A Series DAQ is equipped with 68–pin very high density cable interconnect (VHDCI) type connectors.
Figure A-1 Pins Configuration of Connector 1 for
U2331A, U2355A, U2356A
NOTE
52 U2300A Series DAQ Service Guide
(AIH101..132) and (AIL101..132) are for differential mode connection pair.
Figure A-2 Pins Configuration of Connector 2
for U2331A, U2355A, U2356A
Connector Pins Configuration A
Pins Configuration for U2351A, U2352A, U2353A and U2354A
Figure A-3 Pins Configuration for U2352A,
U2354A
NOTE
U2300A Series DAQ Service Guide 53
(AIH101..108) and (AIL101..108) are for differential mode connection pair.
Figure A-4 Pins Configuration for U2351A,
U2353A
A Connector Pins Configuration
Tab l e A- 5 68-pin VHDCI connector pins descriptions
Signal Name Direction Reference
Ground
AI_GND N/A N/A Analog input (AI) ground. All three ground
For 16 Channels: AI<101..116>
For 64 Channels: AI<101..164>
AI_SENSE Input AI_GND Analog input sense. The reference pin for any
EXTA_TRIG Input AI_GND External AI analog trigger AO201 Output AO_GND Analog output channel 1 AO202 Output AO_GND Analog output channel 2 AO_EXT_REF Input AO_GND External reference for AO channels AO_GND N/A N/A Analog ground for AO EXTD_AO_TRIG Input D_GND External AO waveform trigger EXTD_AI_TRIG Input D_GND External AI digital trigger RESERVED Output N/A Reserved pins. Do not connect them to any signal. COUNT<301,302>_CLK Input D_GND Source of counter <301,302> COUNT<301,302>_GATE Input D_GND Gate of counter <301,302> COUNT<301,302>_OUT Input D_GND Output of counter <301,302> COUNT<301,302>_UPDOWN Input D_GND Up/Down of counter <301,302> EXT_TIMEBASE Input D_GND External Timebase D_GND N/A N/A Digital ground DIO501<7,0> PIO D_GND Programmable DIO of Channel 501 DIO502<7,0> PIO D_GND Programmable DIO of Channel 502 DIO503<4,0> PIO D_GND Programmable DIO of Channel 503 DIO504<4,0> PIO D_GND Programmable DIO of Channel 504
Input AI_GND U2351A/U2352A/U2353A/U2354A
Description
references (AI_GND, AO_GND, and D_GND) are connected together on board.
Analog input channels 101~116. Each channel pair, AI<i, i+8>(i = 101..108), can be configured either as two single-ended inputs or one differential input (marked as AIH<101..108> and AIL<101..108>). U2331A/U2356A/U2355A Analog input channels 101~164). Each channel pair, AI<i, i+32> (i = 101..132), is configured either as two single-ended inputs or one differential input (marked as AIH<101..132> and AIL<101..132>)
AI<101..116> or AI<101..164> channels in NRSE input configuration.
54 U2300A Series DAQ Service Guide
Connector Pins Configuration A
U2300A Series DAQ Service Guide 55
A Connector Pins Configuration
56 U2300A Series DAQ Service Guide
Agilent U2300A Series USB Multifunction DAQ Service Guide
B Test Limits
Analog Input 58 Analog Output 62
This appendix provides the test limits for analog input and analog output when performing the verification procedure.
Agilent Technologies
57
BTest Limits
Setting Range (V) Test Point (V) Te st L im i t ( V ) Tes t Re s ul t
Polarity Min Max Location Val ue Lower Upper Reading
0.001 BIP –10 10 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.002 BIP –10 10 Positive FS 9.98779 9.98579 9.98979 DAQ reading – Offset
0.002 BIP –10 10 Negative FS –9.98779 –9.98979 –9.98579 DAQ reading – Offset
0.001 BIP –5 5 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.002 BIP –5 5 Positive FS 4.99390 4.99190 4.99590 DAQ reading – Offset
0.002 BIP –5 5 Negative FS –4.99390 –4.99590 –4.99190 DAQ reading – Offset
Offset/

Analog Input

Analog Input
Gain Error (V)
U2351A, U2352A, U2353A, U2354A:
0.001 BIP –2.5 2.5 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –1.25 1.25 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.0015 BIP –2.5 2.5 Positive FS 2.49695 2.49545 2.49845 DAQ reading – Offset
0.0015 BIP –2.5 2.5 Negative FS –2.49695 –2.49845 –2.49545 DAQ reading – Offset
0.0015 BIP –1.25 1.25 Positive FS 1.24847 1.24697 1.24997 DAQ reading – Offset
0.0015 BIP –1.25 1.25 Negative FS –1.24847 –1.24997 –1.24697 DAQ reading – Offset
0.002 UNIP 0 10 Positive FS 9.99390 9.99190 9.99590 DAQ reading – Offset
0.002 UNIP 0 10 Negative FS 0.00610 0.00410 0.00810 DAQ reading – Offset
0.002 UNIP 0 5 Positive FS 4.99695 4.99495 4.99895 DAQ reading – Offset
0.002 UNIP 0 5 Negative FS 0.00305 0.00105 0.00505 DAQ reading – Offset
0.001 UNIP 0 2.5 Center FS 1.25000 1.24900 1.25100 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 2.5 Positive FS 2.49847 2.49747 2.49947 DAQ reading – Offset
0.001 UNIP 0 2.5 Negative FS 0.00153 0.00053 0.00253 DAQ reading – Offset
0.001 UNIP 0 1.25 Center FS 0.62500 0.62400 0.62600 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 1.25 Positive FS 1.24924 1.24824 1.25024 DAQ reading – Offset
0.0015 UNIP 0 10 Center FS 5.00000 4.99850 5.00150 Offset = DAQ reading – Calibrator source voltage
0.0015 UNIP 0 5 Center FS 2.50000 2.49850 2.50150 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 1.25 Negative FS 0.00076 –0.00024 0.00176 DAQ reading – Offset
58 U2300A Series DAQ Service Guide
Test Limits B
Setting Range (V) Test Point (V) Te st L im i t ( V ) Tes t Re s ul t
Polarity Min Max Location Val ue Lower Upper Reading
0.001 BIP –10 10 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.002 BIP –10 10 Positive FS 9.98779 9.98579 9.98979 DAQ reading – Offset
0.002 BIP –10 10 Negative FS –9.98779 –9.98979 –9.98579 DAQ reading – Offset
0.001 BIP –5 5 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.002 BIP –5 5 Positive FS 4.99390 4.99190 4.99590 DAQ reading – Offset
0.002 BIP –5 5 Negative FS –4.99390 –4.99590 –4.99190 DAQ reading – Offset
Analog Input
U2355A, U2356A:
Offset/
Gain Error (V)
0.001 BIP –2.5 2.5 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –1.25 1.25 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.0015 BIP –2.5 2.5 Positive FS 2.49695 2.49545 2.49845 DAQ reading – Offset
0.0015 BIP –2.5 2.5 Negative FS –2.49695 –2.49845 –2.49545 DAQ reading – Offset
0.0015 BIP –1.25 1.25 Positive FS 1.24847 1.24697 1.24997 DAQ reading – Offset
0.001 UNIP 0 10 Center FS 5.00000 4.99900 5.00100 Offset = DAQ reading – Calibrator source voltage
0.0015 BIP –1.25 1.25 Negative FS –1.24847 –1.24997 –1.24697 DAQ reading – Offset
0.001 UNIP 0 5 Center FS 2.50000 2.49900 2.50100 Offset = DAQ reading – Calibrator source voltage
0.0015 UNIP 0 10 Positive FS 9.99390 9.99240 9.99540 DAQ reading – Offset
0.0015 UNIP 0 10 Negative FS 0.00610 0.00460 0.00760 DAQ reading – Offset
0.001 UNIP 0 2.5 Center FS 1.25000 1.24900 1.25100 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 2.5 Positive FS 2.49847 2.49747 2.49947 DAQ reading – Offset
0.001 UNIP 0 2.5 Negative FS 0.00153 0.00053 0.00253 DAQ reading – Offset
0.001 UNIP 0 1.25 Center FS 0.62500 0.62400 0.62600 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 1.25 Positive FS 1.24924 1.24824 1.25024 DAQ reading – Offset
0.0015 UNIP 0 5 Positive FS 4.99695 4.99545 4.99845 DAQ reading – Offset
0.0015 UNIP 0 5 Negative FS 0.00305 0.00155 0.00455 DAQ reading – Offset
0.001 UNIP 0 1.25 Negative FS 0.00076 –0.00024 0.00176 DAQ reading – Offset
U2300A Series DAQ Service Guide 59
BTest Limits
Setting Range (V) Test Point (V) Te st L im i t ( V ) Tes t Re s ul t
Polarity Min Max Location Val ue Lower Upper Reading
0.002 BIP –10 10 Center FS 0.00000 -0.00200 0.00200 Offset = DAQ reading – Calibrator source voltage
0.006 BIP –10 10 Positive FS 9.80469 9.79869 9.81069 DAQ reading – Offset
0.006 BIP –10 10 Negative FS –9.80469 –9.81069 –9.79869 DAQ reading – Offset
0.004 BIP –5 5 Positive FS 4.90234 4.89834 4.90634 DAQ reading – Offset
0.004 BIP –5 5 Negative FS –4.90234 –4.90634 –4.89834 DAQ reading – Offset
0.002 BIP –2.5 2.5 Positive FS 2.45117 2.44917 2.45317 DAQ reading – Offset
0.002 BIP –2.5 2.5 Negative FS –2.45117 –2.45317 –2.44917 DAQ reading – Offset
Analog Input
U2331A:
Offset/
Gain Error (V)
0.0015 BIP –5 5 Center FS 0.00000 –0.00150 0.00150 Offset = DAQ reading – Calibrator source voltage
0.0015 BIP –2.5 2.5 Center FS 0.00000 –0.00150 0.00150 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –1.25 1.25 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –1 1 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –1 1 Positive FS 0.98047 0.97947 0.98147 DAQ reading – Offset
0.001 BIP –1 1 Negative FS –0.98047 –0.98147 –0.97947 DAQ reading – Offset
0.001 BIP –0.5 0.5 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –0.5 0.5 Positive FS 0.49023 0.48923 0.49123 DAQ reading – Offset
0.001 BIP –0.5 0.5 Negative FS –0.49023 –0.49123 –0.48923 DAQ reading – Offset
0.001 BIP –0.25 0.25 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –0.25 0.25 Positive FS 0.24512 0.24412 0.24612 DAQ reading – Offset
0.001 BIP –0.25 0.25 Negative FS –0.24512 –0.24612 –0.24412 DAQ reading – Offset
0.001 BIP –0.2 0.2 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –0.2 0.2 Positive FS 0.19609 0.19509 0.19709 DAQ reading – Offset
0.001 BIP –0.2 0.2 Negative FS –0.19609 –0.19709 –0.19509 DAQ reading – Offset
0.001 BIP –0.05 0.05 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –0.05 0.05 Positive FS 0.04902 0.04802 0.05002 DAQ reading – Offset
0.0015 BIP –1.25 1.25 Positive FS 1.22559 1.22409 1.22709 DAQ reading – Offset
0.0015 BIP –1.25 1.25 Negative FS –1.22559 –1.22709 –1.22409 DAQ reading – Offset
0.001 BIP –0.05 0.05 Negative FS –0.04902 –0.05002 –0.04802 DAQ reading – Offset
60 U2300A Series DAQ Service Guide
Test Limits B
Setting Range (V) Test Point (V) Tes t Li m it ( V) Test Result
Polarity Min Max Location Val ue Lower Upper Reading
0.004 UNIP 0 10 Positive FS 9.90234 9.89834 9.90634 DAQ reading – Offset
0.004 UNIP 0 10 Negative FS 0.09766 0.09366 0.10166 DAQ reading – Offset
0.002 UNIP 0 5 Positive FS 4.95117 4.94917 4.95317 DAQ reading – Offset
0.002 UNIP 0 5 Negative FS 0.04883 0.04683 0.05083 DAQ reading – Offset
Analog Input
U2331A:
0.0015 UNIP 0 10 Center FS 5.00000 4.99850 5.00150 Offset = DAQ reading – Calibrator source voltage
Offset/
Gain Error (V)
0.0015 UNIP 0 5 Center FS 2.50000 2.49850 2.50150 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 4 Center FS 2.00000 1.99850 2.00150 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 2.5 Center FS 1.25000 1.24900 1.25100 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 2.5 Positive FS 2.47559 2.47409 2.47709 DAQ reading – Offset
0.001 UNIP 0 2.5 Negative FS 0.02441 0.02291 0.02591 DAQ reading – Offset
0.001 UNIP 0 2 Center FS 1.0000 0.99900 1.00100 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 2 Positive FS 1.98047 1.97947 1.98147 DAQ reading – Offset
0.001 UNIP 0 2 Negative FS 0.01953 0.01853 0.02053 DAQ reading – Offset
0.001 UNIP 0 1 Center FS 0.50000 0.49900 0.50100 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 1 Positive FS 0.99023 0.98923 0.99123 DAQ reading – Offset
0.001 UNIP 0 1 Negative FS 0.00977 0.00877 0.01077 DAQ reading – Offset
0.001 UNIP 0 0.5 Center FS 0.25000 0.24900 0.25100 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 0.5 Positive FS 0.49512 0.49412 0.49612 DAQ reading – Offset
0.001 UNIP 0 0.5 Negative FS 0.00488 0.00388 0.00588 DAQ reading – Offset
0.001 UNIP 0 0.4 Center FS 0.20000 0.19900 0.20100 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 0.4 Positive FS 0.39609 0.39509 0.39709 DAQ reading – Offset
0.001 UNIP 0 0.4 Negative FS 0.00391 0.00291 0.00491 DAQ reading – Offset
0.001 UNIP 0 0.1 Center FS 0.05000 0.04900 0.05100 Offset = DAQ reading – Calibrator source voltage
0.001 UNIP 0 0.1 Positive FS 0.09902 0.09802 0.10002 DAQ reading – Offset
0.0015 UNIP 0 4 Positive FS 3.96094 3.95894 3.96294 DAQ reading – Offset
0.0015 UNIP 0 4 Negative FS 0.03906 0.03706 0.04106 DAQ reading – Offset
0.001 UNIP 0 0.1 Negative FS 0.00098 –0.00002 0.00198 DAQ reading – Offset
U2300A Series DAQ Service Guide 61
BTest Limits
Setting Range (V) Test Point (V) Tes t L im i t ( V ) Tes t R esu l t
Polarity Min Max Location Val ue Lower Upper Reading
0.001 BIP –10 10 0 0.0000 –0.0010 0.0010 Offset = DMM reading – DAQ source voltage
Offset/

Analog Output

Analog Output
Gain Error (V)
U2351A, U2353A:
0.004 BIP –10 10 Positive FS 9.9997 9.9957 10.0037 DMM reading – Offset
0.004 BIP –10 10 Negative FS –10.0000 –10.0040 –9.9960 DMM reading – Offset
0.001 UNIP 0 10 0 0.0000 –0.0010 0.0010 Offset = DMM reading – DAQ source voltage
0.002 UNIP 0 10 Positive FS 9.9998 9.9978 10.0018 DMM reading – Offset
Setting Range (V) Test Point (V) Tes t L im i t ( V ) Tes t R esu l t
Polarity Min Max Location Val ue Lower Upper Reading
Offset/
Analog Output
Gain Error (V)
U2355A, U2356A:
0.001 BIP –10 10 0 0.0000 –0.0010 0.0010 Offset = DMM reading – DAQ source voltage
0.004 BIP –10 10 Positive FS 9.9951 9.9911 9.9991 DMM reading – Offset
0.004 BIP –10 10 Negative FS –10.0000 –10.0040 –9.9960 DMM reading – Offset
0.001 UNIP 0 10 0 0.0000 –0.0010 0.0010 Offset = DMM reading – DAQ source voltage
0.002 UNIP 0 10 Positive FS 9.9976 9.9956 9.9996 DMM – Offset
Setting Range (V) Test Point (V) Tes t L im i t ( V ) Tes t R esu l t
Polarity Min Max Location Val ue Lower Upper Reading
Offset/
Analog Output
Gain Error (V)
U2331A:
0.004 BIP –10 10 Positive FS 9.9951 9.9911 9.9991 DMM reading – Offset
0.004 BIP –10 10 Negative FS –10.0000 –10.0040 –9.9960 DMM reading – Offset
0.0015 BIP –10 10 0 0.0000 –0.0015 0.0015 Offset = DMM reading – DAQ source voltage
62 U2300A Series DAQ Service Guide
0.004 UNIP 0 10 Positive FS 9.9976 9.9936 10.0016 DMM reading – Offset
0.0025 UNIP 0 10 0 0.0000 –0.0025 0.0025 Offset = DMM reading – DAQ source voltage
www.agilent.com
Contact us
o obtain service, warranty or technical
T
support assistance, contact us at the following phone numbers:
United States: (tel) 800 829 4444 (fax) 800 829 4433 Canada: (tel) 877 894 4414 (fax) 800 746 4866 China: (tel) 800 810 0189 (fax) 800 820 2816 Europe: (tel) 31 20 547 2111 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Korea: (tel) (080) 769 0800 (fax) (080) 769 0900 Latin America: (tel) (305) 269 7500 Ta i w a n : (tel) 0800 047 866 (fax) 0800 286 331 Other Asia Pacific Countries: (tel) (65) 6375 8100 (fax) (65) 6755 0042
Or visit Agilent worlwide web at: www.agilent.com/find/assist
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 2007 - 2010
Printed in Malaysia Second Edition, August 23, 2010
U2351-90203
Agilent Technologies
Loading...