The compact, low-cost
alternative to conventional
power measurement solutions
Page 2
Why Agilents power meters and sensors?
Reliable, high-performing solutions
Every power meter and sensor from Agilent consistently delivers great results.
A sure investment for many years to come
Code-compatibility between power meters reduces the need for re-coding. Not only that, all Agilent
power meters are backward-compatible with most legacy power sensors.
One specific application: One right solution
Agilent offers a wide selection of power meters and sensors for practically all application needs
wireless communications, radar pulse measurements, component test and more.
Global network support
No matter where you are, Agilent is committed to giving you the 24-hour support you need regarding
our products, applications or services.
Agilents power meters have long been recognized as
the industry standard for RF and microwave
power measurements.
2
Page 3
Compact solutions for testing today's RF and microwave communication systems
For installation and maintenance
of base stations
Lightweight and rugged
Simple set-up and usage
For production testing of wireless
components
Compact build saves rack space
Simple set-up and usage
For R&D of wireless components
Compact build saves bench space
Simple set-up and usage
Portable with low power consumption
Wide dynamic and frequency ranges
Quick and easy testing with large display
of readings
Wide dynamic and frequency ranges
Fast reading speed
Internal zeroing reduces test time and
sensor wear-and-tear
Quick and easy multiple channels testing
with simultaneous display of readings, limits
and alerts
Seamless integration to system with
industry-standard SCPI
Wide dynamic and frequency ranges
High accuracy
Advanced troubleshooting of designs with
simultaneous display of multiple readings,
measurement math and data recording
3
Page 4
Introducing the U2000 Series USB power sensors
The U2000 Series enable simpler, lower-cost
power measurements versus conventional
power meter and sensor combinations. Now
with nine high-performance models, the U2000
Series USB power sensors offer compact, highperformance solutions for todays CW and
modulated signals.
Key features
Compact, lightweight solutions
Quick, simple set-up
High accuracy, high power
Internal zeroing capability
Fast reading speed
Wide frequency range: 9 kHz
to 24 GHz
Wide dynamic range: 60 dBm
to +44 dBm
Allows remote measurements
beyond cable length
Enables monitoring of more than
20 channels simultaneously
Converts select Agilent
instruments to power meters
Feature-packed software provides
various capabilities for easy testing
and analysis
Average power measurements of
CW and modulated signals, including
GSM, EDGE, WLAN and WiMAX
TM
Compact power meters, simple
set-up
The U2000 Series are standalone sensors. That
means they essentially operate like power
meters, just in smaller forms. No reference
calibrator is required. The fact that each sensor
draws minimal power from a USB portand
that it doesnt need additional triggering
modules or power adaptors to operatemakes
it more portable, especially for base station
testing. Setting up is easy: just plug it to the
USB port of your PC or laptopor even select
network or handheld spectrum analyzerand
start your power measurements. The figure
below illustrates the very simple,
straightforward setup of the U2000 Series.
High accuracy
Each U2000 Series sensor provides excellent
linearity, SWR and uncertainty specifications,
so you can be confident in every measurement
you make.
Wide range, high power
The U2000 Series dynamic range spans across
a wide 80 dB, taking on high power up to +44
dBm.
Remote monitoring and tests
With the U2000 Series sensor plugged to a
networked USB hub, you can conveniently
monitor power measurements of an antenna
tower from the control room, beyond the limits
of USB cable lengths.
USB sensor
DUT
Trigger signal
4
Page 5
Introducing the U2000 Series USB power sensors (continued)
Faster production testing of
multiple channels
The U2000 Series fast measurement speed
of 1000 readings/s in buffered mode helps
reduce test time. This, coupled with the
capability to enable monitoring of more than
20 channels simultaneously, is an advantage
in the production line where efficiency is of
utmost priority.
The U2000 Series has both internal and
external zeroing capabilities. With internal
zeroing, high isolation switches in the sensor
are opened to isolate the sensor from the
device-under-test (DUT) it is connected to. As
such, you dont need to power-off the DUT or
disconnect the sensors. This speeds up testing
and reduces sensor wear-and-tear.
No manual input of calibration data is required.
All calibration factors, as well as temperature
and linearity corrections, are stored in the
sensors EEPROM, auto-downloaded at
calibration.
Often times, youd need to automate your tests.
The U2000 Series sensors are well-equipped
for seamless integration to your system with
industry-standard SCPI compatibility. They also
come with built-in triggering capability to allow
receipt of external triggers from other
instruments.
Transform your signal generators and spectrum analyzers into accurate
power meters
You could literally have a power meter next to youor instead, turn your Agilent MXG signal
generator or N9340A/B handheld spectrum analyzer into a power meter for accurate power
measurements. Even with the U2000 connected, you can switch between power measurements
and the devices original function at any time. You can also use the U2000 with your Agilent PNA
network analyzer for source power calibration.
Intuitive power analysis software
The N1918A Power Analysis Manager software not only displays measurements with the U2000
Series, it also provides various features and functions to help you monitor and troubleshoot
signals efficiently and effectively.
Multiple display types
1
Multiple channel list and measurement math
1. Software capability differs between the two versions of the software: Power Panel and Power Analyzer. Please refer to
Page 17 for detailed comparison
5
Data recording with limit and alert settings
Min/Max readings
1
Page 6
Take a closer look
Built-in triggering port for
synchronization with external
instruments or events
Small and lightweight build
LED indicator for PC communication
Secure locking mechanism
Diode-based sensors frequently rely on the
application of correction factors to extend their
dynamic range beyond their square-law region,
typically in the range of 70 dBm to 20 dBm.
While this technique achieves measurement
of CW signals over a wide dynamic range, it
fails to do so for modulated signals when the
signal level is above the square-law region.
Modulated signals must be padded down, with
their average and peak power levels within
the diode square-law region, for accurate
average power measurement.
The U2000 Series USB power sensors are trueaverage, wide-dynamic-range RF/microwave
power sensors, based on a dual-sensor diode
pair/attenuator/diode pair topology as
proposed by Szente et. al. in 1990
1
. The
simplified block diagram shown here illustrates
this technique.
+
Low Sense
RF in
Low Sense
High Sense
High Sense
+
This technique ensures diodes in the selected
signal path are kept in their square law
regionwith output current and voltage
proportional to input power. The diode
pair/attenuator/diode pair assembly can then
yield the average of complex modulation
formats across a wide dynamic range,
irrespective of signal bandwidth.
USB 2.0-compliant cable
for connection to PC or
other instruments
The dual-range Modified Barrier Integrated
Diode (MBID)
2
package incorporates diode
stacks in place of single diodes. This further
improves measurement accuracy of high-level
signals with high crest factors without
3
incurring damage
to the sensor.
Implementation of both techniques in the
U2000 Series USB sensors enable effective
average power measurements of a wide range
of signals, including multitone and spread
spectrum signals used in CDMA, W-CDMA
and digital television sytems.
1. US Patent #4943764, assigned to Hewlett-Packard Company.
2. November 1986 Hewlett-Packard Journal pages 14-2, Diode Integrated Circuits for Millimeter-Wave Applications.
3. Refer to Maximum Power on page 9 for maximum power handling specifications.
6
Page 7
Specifications
Specifications contained in this chapter are
valid ONLY after proper calibration of the power
sensor and apply to continuous wave (CW)
signals unless otherwise stated. The
recommended calibration interval for this
product is 1 year. Specifications apply over a
temperature range 0 °C to +55 °C unless
otherwise stated. Specifications quoted over
a temperature range of 25 °C ± 10 °C apply to
a relative humidity of 15% to 75% and conform
to the standard environmental test conditions.
Specifications are valid after a 30-minute
warm-up period.
Supplemental characteristics, shown in italics,
are intended to provide useful information with
regard to applying the power sensors in that
they contain typical, but non-warranted
performance parameters. These characteristics
are shown in italics or denoted as typical,nominal or approximate.
The U2000 Series power sensors have two measurement paths: a low power path and a high power path, as shown in the table below.
Models
U2000/1/2/4A
U2000/1/2H
U2000/1B
AUTO (default) range
60 dBm to +20 dBm
50 dBm to +30 dBm
30 dBm to +44 dBm
Each power sensor automatically selects the
proper power level path. To avoid unnecessary
switching when the power level is close to
the switching point, switching point hysteresis
has been added.
Offset at switching point:
±0.5% ( ±0.02
dB) typical
Switching point hysteresis:
1 dB typical
Power Accuracy
Model
U2000/1/2/4A
U2000/1/2H
U2000/1B
Power range
60 dBm to +20 dBm
50 dBm to +30 dBm
30 dBm to +44 dBm
Low power path
60 dBm to 10 dBm
50 dBm to 0 dBm
30 dBm to +20 dBm
Example with U2000 A suffix sensors:
Switching point for the U2000/1/2/4A sensors
is at 10 dBm. Hysteresis causes the low
power path to remain selected until
approximately 9 dBm as the power level is
increased. Above this power, the high power
path is selected. The high power path remains
selected until approximately 11 dBm is
reached as the signal level decreases. Below
this power, the low power path is selected.
1
Accuracy
(25 °C ± 10 °C)
±3.0%
±4.0%
±3.5%
(0 °C to 55 °C)
Accuracy
±3.5%
±5.0%
±4.0%
1
High power path
10 dBm to +20 dBm
0 dBm to +30 dBm
+20 dBm to +44 dBm
Switching point
10 dBm
0 dBm
+20 dBm
Specifications valid with the following conditions:
After zeroingNumber of averages = 1024After 30 minutes of power-on warm-up
1. This accuracy is essentially a combination of linearity, instrumentation accuracy, and traceability to absolute accuracy
at 50 MHz, 0 dBm. Note: Mismatch uncertainty, calibration factor uncertainty, and power level dependent terms (zero
set, drift, and noise) are excluded in this specification and specified elsewhere in the data sheet.
10
Page 11
Specifications (continued)
Typical power accuracy at 25 °C for
U2000/1/2/4A sensors
Typical power accuracy at 25 °C for
U2000/1B sensors
1
1
Error (%)
U2000/1/2/4A Models
1
0.75
0.5
0.25
0
0.25
0.5
0.75
1
40 30 20 10 0 10 20
Power Level (dBm)
Error (%)
U2000/1B Models
1
0.75
0.5
0.25
0
0.25
0.5
0.75
1
10 5 0 5 10 15 20 25 30
Power Level (dBm)
Typical power accuracy at 25 °C for
U2000/1/2H sensors
1
Error (%)
U2000/1/2H Models
1
0.75
0.5
0.25
0
0.25
0.5
0.75
1
30 20 10 0 10 20 30
1. Measurement uncertainty 1.6%. At room temperature and excluding power level dependent terms (zero set, drift and noise)
Power Level (dBm)
11
Page 12
Specifications (continued)
Zero and measurement noise
Power range
U2000/1/2/4A sensors
60 dBm to 35 dBm
38 dBm to 15 dBm
20 dBm to 9 dBm
11 dBm to 5 dBm
7 dBm to 15 dBm
10 dBm to 20 dBm
U2000/1/2H sensors
50 dBm to 25 dBm
28 dBm to 5 dBm
10 dBm to 1 dBm
1 dBm to 5 dBm
3 dBm to 25 dBm
20 dBm to 30 dBm
U2000/1B sensors
30 dBm to 5 dBm
8 dBm to 15 dBm
10 dBm to 21 dBm
19 dBm to 25 dBm
23 dBm to 44 dBm
1
Zero set (internal)
±1.5 nW (±2.8 nW for U2004A)
±2 nW (±3 nW for U2004A)
±12 nW
±2 µW
±4 µW
±6 µW
±1.5 nW
±20 nW
±120 nW
±20 µW
±40 µW
±60 µW
±1.8 µW
±2 µW
±12 µW
±2 mW
±4 mW
Zero set (external)
±600 pW
±1.5 nW
±10 nW
±500 nW
±1 µW
±5 µW
±8 nW
±20 nW
±100 nW
±20 µW
±30 µW
±60 µW
±800 nW
±2 µW
±10 µW
±1 mW
±2 mW
Zero drift
200 pW
400 pW
1.5 nW
50 nW
500 nW
2 µW
2 nW
4 nW
15 nW
500 nW
5 µW
20 µW
200 nW
400 nW
1.5 µW
50 nW
500 µW
2
Measurement noise
1 nW
1.5 nW
15 nW
650 nW
1 µW
10 µW
10 nW
15 nW
150 nW
6.5 µW
10 µW
100 µW
1 µW
1.5 µW
15 µW
650 µW
1 mW
3
Effects of averaging on noise: Averaging over
1 to 1024 readings is available for reducing
noise. The table below provides the
Example:
U2000A power sensor, 60 dBm to 35 dBm,
normal mode, number of averages = 4
measurement noise for a prticular sensor with
the number of averages set at 16 (for normal
mode) and 32 (for x2 mode). Use the noise
Measurement noise calculation:
1 nW x 1.7 = 1.7 nW
multiplier, for the appropriate of averages, to
determine the total measurement noise value.
No. of averages
Noise multiplier
Normal mode
x2 mode
1. Condition: (i) 0 °C to 55 °C and (ii) 95% relative humidity at 40 °C non-condensing.
2. Within one hour after zero set, at a constant temperature, after a 24-hour warm-up of the power sensor.
3. The number of averages at 16 for Normal mode, measured over a one-minute interval and two standard deviations.
1
2.0
2.7
2
1.8
2.4
4
1.7
2.0
8
1.5
1.6
16
1.0
1.0
32
0.95
0.91
64
0.74
0.78
128
0.55
0.53
256
0.39
0.34
512
0.29
0.29
1024
0.21
0.20
12
Page 13
Specifications (continued)
Settling time
Manual filter, 10-dB decreasing power step (not across switching points)
No. of averages
Settling time (s)
Normal mode
x2 mode
1
0.045
0.042
2
0.09
0.05
4
0.17
0.09
8
0.34
0.17
Auto filter, default resolution, 10-dB decreasing power step
Typical
x2
mode
45 ms
82 ms
1.3 s
1.5 s
Normal
mode
45 ms
90 ms
2.6 s
2.7 s
U2000/1/2/4AU2000/1/2HU2000/1B
settling
times
450 ms
1.6 s
20 s
24 s
460 ms
2.8 s
39 s
42 s
+10 dBm
+2 dBm
4 dBm
10 dBm
20 dBm
30 dBm
40 dBm
16
0.66
0.34
32
1.3
0.66
+20 dBm
+12 dBm
+6 dBm
10 dBm
20 dBm
30 dBm
0 dBm
64
2.6
1.3
128
5.2
2.6
+40 dBm
+32 dBm
+26 dBm
+20 dBm
+10 dBm
0 dBm
10 dBm
256
10.4
5.2
Max dBm
High Power Path
Sensor
dynamic
range
512
20.9
10.4
1024
41.9
20.9
50 dBm
24 s
42 s
In Fast mode (using free run trigger), at 10-dB decreasing power step
Settling time = 25 ms*
* When a power step crosses through the sensors auto-range switching point, add 25 ms.
13
40 dBm
20 dBm
Low Power Path
Min dBm
Page 14
Specifications (continued)
Calibration factor and reflection
coefficient
Calibration factor (CF) and reflection coefficient
(Rho) data is unique to each sensor. The CF
corrects for the frequency response of the
sensor. The reflection coefficient (Rho or )
relates to the SWR based on the following
formula:
SWR =
Maximum uncertainties of the CF data are
listed in the following table. There is only one
set of CF data used for both high and low
power paths of each sensor.
The uncertainty analysis for the calibration
data was done in accordance with the ISO
Guide. The uncertainty data reported on the
calibration certificate is the expanded
uncertainty with a 95% confidence level and
a coverage factor of 2.
Frequency
U2000A sensor
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
U2001A sensor
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 6 GHz
U2002A sensor
50 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 24 GHz
U2004A sensor
9 kHz to 2 GHz
2 GHz to 6 GHz
U2000B sensor
10 MHz to 2 GHz
2 GHz to 12.4 GHz
12.4 GHz to 18 GHz
U2001B sensor
10 MHz to 2 GHz
2 GHz to 6 GHz
U2000H sensor
10 MHz to 8 GHz
8 GHz to 12.4 GHz
12.4 GHz to 18 GHz
U2001H sensor
10 MHz to 6 GHz
U2002H sensor
50 MHz to 8 GHz
8 GHz to 12.4 GHz
12.4 GHz to 18 GHz
18 GHz to 24 GHz
Uncertainty (%)
(25 °C ± 10 °C)
1.8
1.6
2.0
2.2
2.2
1.8
1.6
2.0
2.0
2.5
2.7
2.7
3.0
1.8
1.8
1.8
2.0
2.2
1.8
2.0
2.0
2.0
2.2
2.0
2.5
2.5
2.7
3.0
14
Page 15
Specifications (continued)
Trigger
Parameter
Trigger low
Trigger high
Minimum pulse width
Maximum repetition period
Latency
1
Gate
Parameter
Sweep and Offset
Resolution
Power
Measurement speed
1. Not applicable for the U2004A model
2. At conditions: Gate sweep time
2
2 ms; Gate sweep time + Gate offset < 2.3 ms; FAST mode
External TTL input
< 1.1 V
> 1.9 V
35 ns
80 ns
11 µs ± 2 µs
Performance
0 s < Sweep time < 0.15 s
0 s < Offset time < 0.15 s
Note: Sweep time + Offset time < 0.15 s
±1 µs
25 dBm to +20 dBm
10 ms/reading
15
Page 16
Specifications (continued)
General specifications
Physical characteristics
Dimensions (LxWxH)
Weight
Operating environment
Temperature
Humidity
Altitude
Pollution
Storage and shipment
Environment
Temperature
Humidity
Altitude
Pollution
Current requirement
Connector
Cable
Programmability
Safety and EMC compliance
Calibration
Warranty
Compatible instruments
1
1
U2000/1/4A: 163.75 mm x 46.00 mm x 35.90 mm
U2002A: 134.37 mm x 46.00 mm x 35.90 mm
U2000/1B: 308.00 mm x 115.00 mm x 84.00 mm
U2000/1H: 207.00 mm x 46.00 mm x 36.00 mm
U2002H: 164.00 mm x 46.00 mm x 36.00 mm
U2000/1/4A: 0.262 kg
U2002A: 0.226 kg
U2000/1B: 0.762 kg
U2000/1H: 0.324 kg
U2002H: 0.274 kg
0 °C to 55 °C
Up to 95% relative humidity at 40 °C (non-condensing)
Up to 4600 m (15,000 ft)
Degree 2
Sensor should be stored in a clean, dry environment
30 °C to +70 °C
Up to 90% relative humidity at 65 °C (non-condensing)
Agilent PNA, PNA-L and PNA-X Series performance network
analyzers
®
Visual Basic
1. See Ordering information for available options
16
Page 17
Using the U2000 Series with the N1918A Power Analysis Manager
The N1918A software is available in two versions: the basic Power Panel and advanced Power
Analyzer. Power Analyzer provides full access to the softwares complete features and capabilities
and its license, N1918A-Option 100, is available for purchase separately. A free, fully functional
trial version of the Power Analyzer is accessible for 30 days upon installation from the bundled
N1918A Power Analysis Manager CD. The table here shows functions accessible with each
version when used with the U2000 Series.
Power Analyzer (advanced)Power Panel (basic)
Measurement display functions
Soft panel (digital) display
Gauge (analog) display
Strip chart display
Multiple tabs
Multiple displays per tab
Multilist
(List view of multiple channels)
Graph functions
Single marker
Dual marker
Graph autoscaling
Graph zooming
Measurement math
Save/Load file functions
Save measurement data
(with timestamp)
Load measurement data
Data recording
(with timestamp)
Limit and alert functions
Limit and alert notifications
Alert summary
Instrument setting options
Save/Restore instrument
settings
Gate settings
FDO table parameters
Print option
Print application screen
Enhanced with
and alert notifications
Enhanced with limit
and alert notifications
Up to 2 displays
Up to 2 markers per graph
Delta and ratio
Applies to strip chart displays;
up to 10,000 data points
Applies to strip chart displaysApplies to strip chart displays
Up to 3 displays
Up to 10 markers per graph
Up to 5 sets of markers per
graph
Delta and ratio
Applies to strip chart displays
Applies to soft panel, gauge
and strip chart displays; up
to 7 days
limit
Display units:
Absolute: Watts or dBm
Relative: Percent or dB
Display resolution:
Resolution of 1.0, 0.1, 0.01 and 0.001 dB in log
mode; one to four digits in linear mode.
Default resolution:
0.01 dB in log mode; three digits in linear mode
Zero: For performing internal and external
zeroing.
Range: Sensor-dependent, configurable in
1-kHz steps.
Relative: Displays all successive
measurements relative to the last referenced
value.
Offset: Allows power measurements to be
offset by 100 dB to +100 dB, configurable in
0.001 dB increments, to compensate for
external loss or gain.
Duty cycle: Duty cycle values between 0.001%
to 99.999% can be entered in increments of
0.01% to display a pulse power representation
of measured power. The following equation is
used to calculate the displayed pulse power
value: Pulse Power = Measured Power/
Duty Cycle
Limits: High and low limits can be set in the
range between 150.00 dBm to +230.000 dBm,
in 0.001 dBm increments.
Preset default values: Channel Offset (dB) =
0, Duty Cycle Off, Frequency 50 MHz, AUTO
Average, AUTO Range, Free Run Mode,
dBm mode.
For further details on the software, refer to N1918A Power Analysis Manager Data Sheet, literature
number 5989-6612EN.
17
Page 18
Using the U2000 Series with the N1918A Power Analysis Manager (continued)
System requirements
Hardware
Processor
RAM
Hard disk space
Resolution
Operating system and browser
Operating system
Browser
Software
Agilent IO Libraries Suite
Microsoft .NET Framework
1
3
Microsoft Visual C++ 2005 Runtime Libraries
3
Desktop PC: 1.3 GHz Pentium® IV or higher recommended
Laptop PC: 900 MHz Pentium
®
M or higher recommended
512 MB (1.0 GB or higher recommended)
1.0 GB or more free disk space at runtime
800 x 600 or higher (1280 x 1024 recommended)
®
Windows
XP Professional, service pack 2 or higher
Microsoft Internet Explorer 5.1 (6.0 or higher recommended)
Version 14.2
2
or higher
Runtime version 2.0
Version 1.0 or higher
1. Available in Agilent Automation-Ready CD
2. Agilent IO Libraries Suite 15.0 is required if PC is running on Microsoft Windows Vista 32-bit edition
3. Bundled with N1918A Power Analysis Manager CD
18
Page 19
Ordering information
Power sensors
Models
U2000A
U2001A
U2002A
U2004A
U2000H
U2001H
U2002H
U2000B
U2001B
Description
Items shipped as standard with each sensor:
BNC (m) to SMB (f) trigger cable, 1.5 m, 50
USB 2.0 Type A to 5-pin Mini-B cable with secure locking mechanism, selectable lengths of
1.5 m, 3.0 m or 5.0 m
U2000 Series USB Power Sensor Programming Guide
U2000 Series USB Power Sensor Operating and Service Guide (English)
U2000 Series USB Power Sensor Documentation CD (contains Agilent RFPowerMeter IVI driver)
Certificate of Calibration
N1918A Power Analysis Manager Installation Guide
N1918A Power Analysis Manager CD
Agilent Automation-Ready CD (contains Agilent IO Libraries Suite)
Accessories, calibration and documentation options
Cables
U2031A
U2031B
U2031C
U2032A
Travel kits
U2000A-201
U2000A-202
U2000A-204
Hanging kit
U2000A-203
Calibration
Option 1A7
Option A6J
Warranty
Option R-51B-001-3C
Option R-51B-001-5C
Documentation
Option OB1
Option ABD
Option ABE
Option ABF
Option ABJ
Option ABZ
Option AB2
Software
N1918A-100
Complementary I/O connectivity hardware
1
E5813A
Description
USB 2.0 Type A to 5-pin Mini-B cable with secure locking mechanism, 1.5 m (5 ft)
USB 2.0 Type A to 5-pin Mini-B cable with secure locking mechanism, 3.0 m (10 ft)
USB 2.0 Type A to 5-pin Mini-B cable with secure locking mechanism, 5.0 m (16.4 ft)
BNC (m) to SMB (f) trigger cable, 1.5 m, 50
Description
Transit case
Soft carrying case
Soft carrying pouch
Description
Holster
Description
ISO 17025 calibration with test data
ANSI Z540 calibration with test data
Description
Extension of warranty and service plan from 1 year to 3 years
Extension of warranty and service plan from 1 year to 5 years
Description
English language Operating and Service Guide
German language Operating and Service Guide
Spanish language Operating and Service Guide
French language Operating and Service Guide
Japanese language Operating and Service Guide
Italian language Operating and Service Guide
Simplified Chinese language Operating and Service Guide
Description
N1918A Power Analysis Manager software, advanced Power Analyzer version
Description
Networked 5-port USB hub
1. Refer to www.agilent.com/find/e5813a for more information on the device.
19
Page 20
Ordering information (continued)
Transit case
Soft carrying pouch
Soft carrying case
Holster
20
Page 21
www.agilent.com
www.agilent.com/find/usbsensor
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
www.agilent.com/find/open
Agilent Open simplifies the process
of connecting and programming
test systems to help engineers
design, validate and manufacture
electronic products. Agilent offers
open connectivity for a broad range
of system-ready instruments, open
industry software, PC-standard I/O
and global support, which are
combined to more easily integrate
test system development.
WiMAX is a trademark of the WiMAX Forum. Windows and
Microsoft are U.S. registered trademarks of Microsoft
Corporation. Pentium is a U.S. registered trademark of Intel
Corporation. cdma2000 is a registered certification mark of
the Telecommunications Industry Association. Used under
license. Bluetooth is a trademark owned by Bluetooth SIG,
Inc., U.S.A. and licensed to Agilent Technologies, Inc.
Remove all doubt
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to
www.agilent.com/find/removealldoubt
Product specifications and descriptions
in this document subject to change
without notice.
For more information on Agilent Technologies
products, applications or services, please
contact your local Agilent office. The complete list is available at:
www.agilent.com/find/contactus
Phone or Fax
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand