The U1583B current clamp is a dual range 40 A and 400 A clamp-on AC current clamp. This U1583B current
clamp is designed for Agilent handheld digital multimeter (DMM) and Agilent handheld oscilloscope. The
BNC-to-banana plug is required to connect the current clamp with DMM. While for handheld oscilloscope,
it needs to use with BNC connector.
Assistance
For technical assistance, contact your nearest Agilent Sales Offi ce or visit the Agilent website at
www.agilent.com/fi nd/assist for further information.
Regulatory Markings
The CE mark is a registered trademark of the European Community.This CE mark
shows that the product complies with all the relevant European Legal Directives. If
it was accompanied by a year, it indicates the year the design was approved. This
ISM device complies with Canadian ICES-001.
Product contains restricted substance(s) above the maximum value, with 40 yr
Environmental Protection Use Period.
The CSA mark is a registered trademark of the Canadian Standards Association.
The C-tick mark is a registered trademark of the Spectrum Management Agency of
Australia. This signifi es compliance with the Australia EMC Framework regulations
under the terms of the Radio Communication Act of 1992.
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affi xed product label indicates that you must not discard this
electrical/electronic product in domestic household waste.
Safety Information
Please use the Agilent U1583A current clamp only as specifi ed in this manual. Otherwise, the protection
provided may be impaired. A WARNING identifi es conditions and actions that pose hazards to the user. A
CAUTION identifi es conditions and actions that may be damaging to the equipment under test. To avoid
possible electric shock, personal injury or damage to this instrument, ensure that you use the adapter
safely, and refer to the guidelines below.
400A MAX
CAT III 600V
AC - Alternating Current
Caution, risk of danger (Refer to the
user’s and service guide for details)
Maximum allowable current
measurement is 400 A
Category III 600V over-voltage protectionDouble Insulation
Range button in lock mode.
Range ~ 40 A, Output ~ 10 mV/A
To be applied around or removed from
un-insulated hazardous live conductors
- Do not use the adapter if it is damaged. Inspect the case before you use the adapter.
Look for cracks or missing plastic. Pay particular attention to the insulation
surrounding the connectors.
- Inspect the clamp jaw before each use. It shall not have cracks or missing parts,
loose or weakened components. Be sure there is insulation surrounding the jaw.
Range button in release mode.
Range ~ 400 A, Output ~ 1 mV/A
DC - Direct Current
Ground
Caution, risk of electric shock (Refer to
the user’s and service guide for detail)
- Inspect output cable without exposing the metal to ensure insulation.
- Do not operate the adapter around explosive gas, vapor or dust.
- Do not exceed the rated voltage/current as marked on the adapter.
- Use with extreme caution when working around bare conductors or bus bars. It could
results in electric shock if have accidentally contact to the conductors.
- Always keep your hand behind the fi nger guard of clamp jaw.
- When servicing the adapter, use only specifi ed replacement parts.
- Use with caution when working above 30 V ac rms, 42 V peak or 60 V dc. Such
voltage poses a shock hazard.
- Avoid working alone.
- Do not operate the adapter if the cover is removed or loosened.
- Do not connect to BNC output or banana plug to any power sources.
- Use the proper terminals, function, and range for your measurements.
Standard Item Purchased Checklist
The following items are included when you make a purchase:
Specifi ed Current Range1 A to 400 A acLoad impedance> 1 M ohm, < 100 pF
Usable Current Range0.5 A to 400 AOperating Temperature–10 °C to 55 °C (14 °F to 131 °F)
AC crest factor< 3Storage Temperature–20 °C to 70 °C (–4 °F to 158 °F)
Bandwidth10 kHzMeasurement CategoryCAT III 600 V; Pollution degree II
Weight294 gramsDimensions (HxWxD)44 mm (H) x 92 mm (W) x 188 mm (L)
Cable length1500 +/– 20 mmMaximum conductor size30 mm or 16 mm x 2
Maximum Jaw Opening32 mmWarm-up timeImmediately upon power on
AltitudeUp to 2000 metersWarrantyOne year
Relative HumidityMax 80% RH for temperature up to 35 °C decreasing linearly to 50% RH at 55 °C
Safety ComplianceSafety Compliance Certifi ed by CSA (Canada & USA) for IEC/EN/UL 61010-1 2nd Edition &
EMC ComplianceCertifi ed to IEC/EN 61326:2002, CISPR 11, and equivalents for Group 1, Class A
EN/IEC 61010-2-032
Electrical Accuracy Specifi cations
Range
40 A10 mV/A0.5 A ~ 40 A2 % + 0.5 A5 % + 0.5 A10 % + 0.5 A
400 A1 mV/A0.5 A ~ 40 A2.5 % + 0.5 A4.5 % + 0.5 A12.5 % + 0.5 A
Output
Resolution
± (% of reading + digit) at 23 °C ±5 °C, with relative humidity less than 80% R.H.
Span48 Hz ~ 65 Hz
40 A ~ 200 A2 % + 0.5 A4 % + 0.5 A12 % + 0.5 A
200 A ~ 400 A1.5 % + 0.5 A3.5 % + 0.5 A11.5 % + 0.5 A
• The current conductor must be centered within the jaw aperture and no infl uence from adjacent
currents.
• Connect to load impedance > 1 MW ||100 pF.
• Crest factor < 3.
Accuracy
40 ~ 48 Hz/
65 Hz ~ 1 kHz
1 kHz ~ 10 kHz
Current Clamp with
BNC connector
BNC-to-banana
plug
Handheld Digital
Multimeter (DMM)
Current Clamp with
BNC connector
Handheld
Oscilloscope
Figure 1: Connection Diagram
Alignment Marks
Put one conductor only within the jaws on inside section of the
indicated marks as much as possible in order to meet the accuracy of
specifi cations. Make sure the clamp is vertical to the conductor.
Current Range Selection
To push and release the yellow button for 40 A range and lock the button for 400 A range.
ButtonRangeOutput
Release
Lock
~ 40 A~ 10 mV/A
~ 400 A~ 1 mV/A
Operation
AC current can be measured without removing the conductor out of the circuit by following the procedures
as shown below:
1. Plug the BNC of cable to the BNC with dual banana plugs, and then plug into the V/COM terminals on
multimeter. For oscilloscope, plug the BNC connector directly to oscilloscope.
2. Set the ACV measurement and range on multimeter you used.
3. Position the jaw to a single conductor and center it accordingly to alignment marks.
4. Ensure that the arrow marked on the clamp jaw point towards the load of phase measurements or
away from the load (toward the source) for neutral measurements.
5. Observe the AC value on a multimeter or the waveform on an oscilloscope which is proportional to the
current.
6. Select a lower range on the Clamp-on adapter and set the corresponding sensitivity (mV/ A setting)
on the oscilloscope if required.
Calibration Equipment
The pre-calibration guidelines are shown as follows:
• Be sure you are a qualifi ed person to perform the calibration
• The environment should be 23 °C ±2 °C and the relative humidity (RH) shall be < 80%.
The test equipment requirements listed in table below or equivalents are required to perform the calibration
and performance verifi cation test procedures. Alternative equipment may be used as long as the accuracy is
as good as or better than the specifi cations listed.
Standard SourceOperating Range
AC Current Calibrator33 mA – 329.99 mA at 10 Hz to 3 kHz
MultimeterAC 500.0 mV or 1000.0 mV
50 Turns Current Coil0.2 A – 20.5 A
0.33 A – 2.99999 A at 10 Hz to 3 kHz
3 A – 20.5 A at 10 Hz to 3 kHz
Accuracy
Required
≤ ± 0.2 %
≤ ± 0.6 %
≤ ± 3.0 %
≤ ± 1.5 %
≤ ± 1.0 %
Wavetek 9100 or Fluke 5520A or 5101B
or equivalent
Agilent U1251A or U1252A or
Agilent-34405A or equivalent
Fluke 5500A Coil or Wavetek 9100
Option 200 or equivalent
Recommended
Equipment
Adjustment Procedures
AC 40 A range
1. Lock RANGE button of U1583B current clamp to 40 A mode.
2. Connect the output BNC of U1583B to a BNC-to-dual banana converter plug and proceed to
connect it to the output of the V (HI) and COM (LO) terminals of the multi-meter.
3. Set multi-meter to AC 500.0 mV or 1000.0 mV.
4. Open the jaws of the current clamp and centrally place it around the 50 turns coil.
5. Set the calibrator output to 50 turns coil. Confi gure the calibrator to generate a current 20 A with 60 Hz frequency for adjustment of the current clamp.
6. Remove two Phillips screws on the back of the current clamp and proceed to adjust the VR1
until the display on multimeter indicates AC 200 mV ± 0.2 mV. Please refer to below fi gure for the position of VR1.
VR1 for AC 40 A
adjustment
VR2 for AC 400 A
adjustment
Figure 2: U1583B Current Clamp Circuit Board Diagram
AC 400 A range
1. Release the Range button of U1583B current clamp to 400 A mode.
2. Connect the output BNC of U1583B to a BNC-to-dual banana converter plug and proceed to
connect it to the output of the V (HI) and COM (LO) terminals of the multi-meter.
3. Set Multi-meter to AC 500.0 mV or 1000.0 mV.
4. Open the jaws of the current clamp and centrally place it around the 50 turns coil.
5. Set the calibrator output to 50 turns coil. Confi gure the calibrator to generate a current 20 A with 60 Hz frequency for adjustment of the current clamp.
6. Remove two Phillips screws on the back of the current clamp and proceed to adjust the VR2
until the display on multimeter indicates AC 200 mV ± 0.2 mV. Please refer to Figure 2 for the
position of VR2.
Remember to replace the screws to its original position after performing the calibration
adjustments.
Maintenance
Repair or service not cover in this sheet should be performed only by qualifi ed personnel.
To avoid electrical shock or damage to the Clamp-on adapter, do not get water inside
the case .
Cleaning
• Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasive or solvents.
• Open the jaws and wipe the metal of jaw with a lightly oiled cloth, and then wipe the oil with dry
cloth. Do not allow rust or corrosion to form on the metal ends of jaw.
Troubleshooting
If the current clamp does not perform properly, follow the steps below to identify the problem:
1. Inspect the mating surface of jaw for cleanliness. If any external material is present, the jaws
may not close properly and affects the result.
2. Verify that the function selection and range on the multimeter or oscilloscope are correct and
the range is adjusted on current clamp.