Agilent U1251A Data Sheet

Page 1
Agilent U1251A and U1252A Handheld Digital Multimeter
Data Sheet
Features
• 50,000 count dual display
• Up to 0.025% basic DC voltage accuracy
• True RMS measurement
• J-type and K-type temperature measurements
• Data storage capability with optional IR-USB link to PC
• 20 MHz frequency counter
• Programmable square wave generator
–20 ° C to +55 ° C operating
temperature
• Safety certified with EN/IEC 61010­1
Category III 1000 V overvoltage
protection
• Built-In battery charger capability
• Multi function tilt stand
Introduction
The Agilent U1250A series handheld digital multimeter gives you 4.5 digit resolution with a 50,000 count full scale and basic DC voltage accuracy of up to 0.025%. These features give you the flexibility to perform quick validation measurements or perform tolerance checks and marginal failure troubleshooting. It also offers true RMS, dBm, and AC+DC readings to measure both sinusoidal and non­sinusoidal waveforms accurately.
Versatile functionality
The U1250A series handheld digital multimeter delivers many features that you would expect from a benchtop multimeter. These instruments come with all the basic measurement functions and added functionality needed in today’s changing maintenance environment such as temperature measurements, frequency measurements, 4-20 mA process loop measurement with % readout, 20 MHz frequency counter (U1252A only), and dBm measurements.
More than just a measurement tool
The U1250A series expands users’ capability beyond typical measurements. These instruments provide data storage capability which can be done both manually or automatedly with the optional PC
interface cable. The accompanying Agilent Graphical User Interface (GUI) further extends the potential of these instruments by allowing you to customize your data logging needs. Other features include sophisticated math functions that allow users to manipulate the data obtained. The U1252A can be used as a programmable square wave generator that allows you to stimulate electronic circuits for debugging purposes.
Uncompromising ruggedness and safety
Use The U1250A series handheld digital multimeter with confidence in almost any environment. These instruments come in a robust package with shock absorbing overmold and operate within their rated specifications from –20 ° C right up to +55 ° C. Furthermore, these instruments are rated at Cat III 1000 V (IEC 61010 Compliant) which allows you to confidently make measurements on building electrical installations, at locations between the main circuit breaker and the mains socket-outlets, including measurements performed on equipment that is permanently connected to the mains circuits.
Visit our Web for more information on Agilent’s Handheld Digital Multimeter at www.agilent.com
Page 2
U1251A & U1252A DC SPECIFICATIONS
FUNCTION RANGE RESOLUTION TEST CURRENT/
BURDEN VOLTAGE
VOLTAGE
(1)
50.000 mV 0.001 mV -
500.00 mV 0.01 mV -
1000.0 mV 0.1 mV -
5.0000 V 0.0001 V -
50.000 V 0.001 V -
500.00 V 0.01 V - 0.03+5
1000.0 V 0.1 V -
RESISTANCE
500.00
5.0000 k
(3)
(3)
0.01 1.04 mA
0.0001 k 416 µA
50.000 k 0.001 k 41.2 µA
500.00 k 0.01 k 4.12 µA
5.0000 M 0.0001 M 375 nA 0.2+5 0.15+5
50.000 M
(4)
0.001 M 187 nA
500.00 M 0.01 M 187 nA N/A 3+10<200 MΩ/8+10>200 M
500 nS
(5)
0.01 nS 187 nA
CURRENT 500.00 µA0.01 µA 0.06 V (100 Ω)
5000.0 µA0.1 µA0.6 V (100 Ω)
50.000 mA 0.001 mA 0.09 V (1 Ω)
440.00 mA 0.01 mA 0.9 V (1 Ω)
5.0000 A 0.0001 A 0.2 V (0.01 Ω) 0.3+10 0.3+10
10.000 A
(7)
0.001 A 0.4 V (0.01 Ω)
DIODE TEST - 0.1 mV 1.04 mA 0.05 + 5
ACCURACY ± (% of reading + No. of Least Significant Digit)
U1251A U1252A
(2)
0.05+50
0.03+5
0.08+10
0.08+5
1+10
1+20
(6)
0.05+5
0.1+5
(6)
0.1+5
0.05+5
(6)
0.2+5
(6)
0.2+5
0.3+10 0.3+5
0.05+50
0.025+5
0.05+10
0.05+5
1+5
1+10
0.15+5
0.15+5
(2)
(6)
(6)
(6)
(6)
TEMPERATURE AND CAPACITANCE SPECIFICATIONS
FUNCTION THERMOCOUPLE
TYPE
K –200 ~ 1372 ° C/
TEMPERATURE
(8)
J
- 10.000 nF 0.001 nF 1%+8
- 100.00 nF 0.01 nF
CAPACITANCE
- 1000.0 nF 0.1 nF
- 10.000 µF0.001 µF
- 100.00 µF0.01 µF
- 1000.0 µF0.1 µF1 time/sec.
- 10.000 mF 0.001 mF 0.1 times/sec.
- 100.00 mF 0.01 mF 3%+10 0.01 times/sec.
[1] Input impedance: >1 GΩ for 50 mV~1000 mV ranges. For U1251A, input impedance is 10 M (nominal) for 5 V~1000 V ranges. For U1252A, input
impedance is 10 MΩ (nominal) in parallel with 1.1 MΩ at dual display.
[2] The accuracy could be 0.05%+10 for U1251A and 0.05%+5 for U1252A. Always use NULL function to zero out the thermal effect before measuring the
signal. [3] The accuracy of 500 Ω and 5 kΩ is specified after NULL function, which is used to subtract the test lead resistance and the thermal effect [4] For the range of 50 MΩ, the R.H. is specified for <60%. [5] The accuracy is specified for <50 nS and after NULL function with open test lead.
[6] Always use NULL function to zero out thermal effect with open test lead before measuring the signal. If NULL function is not used, an additional 20
counts needs to be added to the DC current accuracy. Thermal effect could occur due to the following:
• Wrong operation to measure the high voltage of 50 V ~ 1000 V for resistance, diode, and mV measurements.
• After battery-charging has completed.
• After measuring a current greater than 440 mA, it is suggested that the meter be left to cool down for twice the measuring time used. [7] Current can be measured up to 10 A continuously. An additional 0.5% needs to be added to the specified accuracy if the signal measured is in the range of 10 A~20 A for 30 seconds maximum. After measuring a current of > 10 A, leave the meter to cool down for twice the measuring time used before application of low current measurement. [8] Only available in U1252A.
RANGE RESOLUTION ACCURACY MEASURING RATE AT
MAX. DISPLAY
FULL SCALE
–328 ~ 2502 ° F –210 ~ 1200 ° C/ –346 ~ 2192 ° F
0.1 ° C/0.1 ° F0.3% +3 ° C/
0.3%+6 ° F
0.1 ° C/0.1 ° F0.3% +3 ° C/
0.3%+6 ° F
N/A N/A
N/A N/A
4 times/sec.
1%+5
11000 counts
2
Page 3
U1251A AC SPECIFICATIONS
FUNCTION
RANGE
50.000 mV 0.001 mV 1+60 0.6+40 1.0+40 1.6+60
500.00 mV 0.01 mV 1+60 0.6+25 1.0+40 1.6+60
TRUE-RMS AC VOLTAGE
1000.0 mV 0.1 mV 1+60 0.6+25 1.0+25 1.6+40
5.0000 V 0.0001 V 1+60 0.6+25 1.0+25 1.6+40
50.000 V 0.001 V 1+60 0.6+25 1.0+25 1.6+40
500.00 V 0.01 V 1+60 0.6+25 1.0+25
1000.0 V 0.1 V 1+60 0.6+40 1.0+40 N/A
FUNCTION RANGE RESOLUTION
500.00 µA
(2)
5000.0 µA0.1 µA 1.5+40 0.8+20 3+60
AC CURRENT
50.000 mA 0.001 mA 1.5+40 0.8+20 3+60
440.00 mA 0.01 mA 1.5+40 0.8+20 3+60
5.0000 A 0.0001 A
(3)
10.000 A
RESOLUTION
0.01 µA 1.5+50 0.8+20 3+80
0.001 A
U1252A AC SPECIFICATIONS
FUNCTION RANGE RESOLUTION
50.000 mV 0.001 mV 1.5+60 0.4+40 0.7+40 0.75+40 3.5+120
500.00 mV 0.01 mV 1.5+60 0.4+25 0.4+25 0.75+40 3.5+120
TRUE-RMS AC VOLTAGE
1000.0 m V 0.1 mV 1.5+60 0.4+25 0.4+25 0.75+40 3.5+120
5.0000 V 0.0001 V 1.5+60 0.4+25 0.4+25 0.75+40 3.5+120
50.000 V 0.001 V 1.5+60 0.4+25 0.4+25 0.75+40 3.5+120
500.00 V 0.01 V 1.5+60 0.4+25 0.4+25 1.5+40
1000.0 V 0.1 V 1.5+60 0.4+40 0.4+40
ACCURACY ± (% of reading + No. of Least Significant Digit)
FREQUENCY
30 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 10 kHz 10 kHz ~ 30 kHz
(1)
1.6+40
ACCURACY ± (% of reading + No. of Least Significant Digit)
FREQUENCY
30 Hz ~ 45 Hz 45 Hz ~ 2 kHz 2 kHz ~ 20 kHz
2+40
2+40
(4)
(4)
0.8+20 3+60
0.8+20
<3 A/5 kHz
ACCURACY ± (% of reading + No. of Least Significant Digit)
FREQUENCY
20 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 10 kHz 10 kHz ~ 20 kHz 20 kHz~100 kHz
(1)
3.5+120
1.5+40
(1)
N/A
(5)
ACCURACY ± (% of reading + No. of Least Significant Digit)
FUNCTION RA NGE RESOLUTION
(2)
0.01 µA 1.0+20 0.7+20 0.75+20 5+80
AC CURRENT
500.00 µA
5000.0 µA0.1 µA 1.0+20 0.7+20 0.75+20 5+80
50.000 mA 0.001 mA 1.0+20 0.7+20 0.75+20 5+80
440.00 mA 0.01 mA 1.0+20 0.7+20 1.5+20 5+80
5.0000 A 0.0001 A
10.000 A
(3)
0.001 A
20 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 20 kHz 20 kHz ~ 100 kHz
(4)
1.5+20
(4)
1.5+20
FREQUENCY
0.7+20 3+60
0.7+20
<3 A/5 kHz
(5)
N/A
[1] The input signal is lower than the product of 20,000,000 V-Hz (product of voltage and frequency). [2] Input current >35 µArms. [3] Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5% needs to be added to the specified accuracy if the signal measured is in
the range of 10 A ~ 20 A for 30 seconds maximum. After measuring a current of >10 A, leave the meter to cool down for twice the measuring time used before application of low current measurement.
[4] Input current < 3 Arms. [5] The additional error to be added as frequency >20 kHz and signal input <10% of range: 3 counts of LSD per kHz.
3
Page 4
U1252A AC+DC SPECIFICATIONS
FUNCTION RANGE RESOLUTION
50.000 mV 0.001 mV 1.5+80 0.4+60 0.7+60 0.8+60 3.5+220
500.00 mV 0.01 mV 1.5+65 0.4+30 0.4+30 0.8+45 3.5+125
VOLTAGE
FUNCTION RANGE RESOLUTION
CURRENT
1000.0 mV 0.1 mV 1.5+65 0.4+30 0.4+30 0.8+45 3.5+125
5.0000 V 0.0001 V 1.5+65 0.4+30 0.4+30 0.8+45 3.5+125
50.000 V 0.001 V 1.5+65 0.4+30 0.4+30 0.8+45 3.5+125
500.00 V 0.01 V 1.5+65 0.4+30 0.4+30 1.5+45
1000.0 V 0.1 V 1.5+65 0.4+45 0.4+45
500.00 µA
(3)
0.01 µA 1.1+25 0.8+25 0.8+25
5000.0 µA0.1 µA 1.1+25 0.8+25 0.8+25
50.000 mA 0.001 mA 1.2+25 0.9+25 0.9+25
440.00 mA 0.01 mA 1.2+25 0.9+25 0.9+25
5.0000 A 0.0001 A
10.000 A
(4)
0.001 A
ACCURACY ± (% of reading + No. of Least Significant Digit)
FREQUENCY
30 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 10 kHz 10 kHz ~ 20 kHz 20kHz~100kHz
(2)
3.5+125 N/A
1.5+45
(2)
ACCU RACY ± (% of reading + No. of Least Significant Digit)
FREQUENCY
30 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 20 kHz
1.8+30
1.8+30
(5)
(5)
0.9+30 3.3+70
0.9+25
<3 A/5 kHz
(1)
U1251A & U1252A FREQUENCY SPECIFICATIONS
(2)
RANGE RESOLUTION ACCURACY MINIMUM INPUT FREQUENCY
99.999 Hz 0.001 Hz
999.99 Hz 0.01 Hz
9.9999 kHz 0.0001 kHz
0.02%+3
<600 kHz
1 Hz
99.999 kHz 0.001 kHz
999.99 kHz 0.01 kHz
[1] The additional error to be added as frequency >20 kHz and signal input <10%of range: 3 counts of LSD per kHz. [2] The input signal is lower than the product of 20,000,000 V-Hz (product of voltage and frequency). [3] Input current >35 µArms. [4] Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5% needs to be added to the specified accuracy if the signal measured is in
the range of 10 A ~ 20 A for 30 seconds maximum. After measuring a current of >10 A, leave the meter to cool down for twice the measuring time used before application of low current measurement.
[5] Input current < 3 Arms.
U1251A FREQUENCY SENSITIVITY DURING VOLTAGE MEASUREMENT
FREQUENCY SENSITIVITY AND TRIGGER LEVEL FOR U1251A
INPUT RANGE
(Maximum input for specified
accuracy = 10 x Range or
1000 V)
50.000 mV 10 mV 15 mV 10 mV 15 mV
500.00 mV 25 mV 35 mV 60 mV 70 mV
1000.0 mV 40 mV 50 mV 100 mV 150 mV
5.0000 V 0.25 V 0.5 V 0.5 V / 1.25 V (< 100 Hz) 0.6 V
50.000 V 2.5 V 5 V 5 V 6 V
500.00 V 25 V N/A 50 V N/A
1000.0 V 50 V N/A 300 V N/A
MINIMUM SENSITIVITY
20 Hz-100 kHz
(R.M.S. Sine-Wave)
>100 kHz ~ 200 kHz
TRIGGER LEVEL FOR DC COUPLING
< 100 kHz >100 kHz ~ 200 kHz
4
Page 5
U1252A FREQUENCY SENSITIVITY DURING VOLTAGE MEASUREMENT
FREQUENCY SENSITIVITY AND TRIGGER LEVEL FOR U1252A
INPUT RANGE
(Maximum input for specified
accuracy = 10 x Range or
1000 V)
50.000 mV 10 mV 25 mV 10 mV 25 mV
500.00 mV 70 mV 150 mV 70 mV 150 mV
1000.0 mV 120 mV 300 mV 120 mV 300 mV
5.0000 V 0.3 V 1.2 V 0.6 V 1.5 V
50.000 V 3 V 5 V 6 V 15 V
500.00 V 30 V < 100 kHz N/A 60 V N/A
1000.0 V 50 V < 100 kHz N/A 120 V N/A
20 Hz-200 kHz
MINIMUM SENSITIVITY
(R.M.S. Sine-Wave)
>200 kHz ~ 500 kHz
TRIGGER LEVEL FOR DC COUPLING
< 100 kHz >100 kHz ~ 500 kHz
DUTY CYCLE
MODE RAN GE ACCURACY AT FULL SCALE
DC Coupling 0.01% ~ 99.99% 0.3% per kHz + 0.3%
[1] Positive or negative pulse width must be greater than 10 µs and the range of duty cycle should be considered. The range of pulse width is determined by
the frequency of the signal.
(1)
PULSE WIDTH
MODE RANGE ACCURACY AT FULL SCALE
500 ms 0.01 ms 0.2% + 3
2000 ms 0.1 ms 0.2% + 3
(1)
U1251A & U1252A FREQUENCY SENSITIVITY DURING
CURRENT MEASUREMENT
INPUT RANGE
500.00 µA100 µA
5000.0 µA250 µA
50.000 mA 10 mA
440.00 mA 25 mA
5.0000 A 1 A
10.000 A 2.5 A
MINIMUM SENSITIV ITY (R.M.S. Sine-Wave)
20 Hz - 20 kHz
PEAK HOLD (Capturing changes)
SIGNAL WIDTH ACCURACY FOR DC mV/VOLTAGE/CURRENT
Single event > 1 ms
Repetitive > 250 µs
2% + 400 for all ranges
2% + 1000 for all ranges
U1252A FREQUENCY COUNTER SPECIFICATIONS
Divide 1 (secondary display “-1-”)
RANGE RESOLUTION ACCURACY SENSITIVITY MIN. INPUT FREQUENCY
99.999 Hz 0.001 Hz 0.02%+3
999.99 Hz 0.01 Hz
9.9999 kHz 0.0001 kHz
99.999 kHz 0.001 kHz
999.99 kHz 0.01 kHz 200 mV R.M.S.
9.9999 MHz 0.0001 MHz
0.002%+5, <2 MHz
100 mV R.M.S.
Divide 100 (secondary display “-100-”)
RANGE RESOLUTION ACCURACY SENSITIVITY MIN. INPUT FREQUENCY
9.9999 MHz 0.0001 MHz 0.002%+5,
99.99 MHz 0.001 MHz
<20 MHz
400 mV R.M.S. 1 MHz 600 mV R.M.S.
0.5 Hz
5
Page 6
U1252A SQUARE WAVE OUTPUT
(1)
OUTPUT
FREQUENCY 0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150,
200, 240, 300, 400, 480, 600, 800, 1200, 1600, 2400, 4800 Hz
(2)
DUTY CYCLE
PULSE WIDTH
(2)
AMPLITUDE Fixed 0~+2.8 V 0.1 V 0.2 V
[1] Output impedance: 3.5 kΩ maximum. [2] The positive or negative pulse width must be greater than 50 µs for adjusting the duty cycle or pulse width under different frequency. Else, the accuracy
and range will be different from the definition.
[3] For signal frequencies greater than 1 kHz, an addition of 0.1% per kHz is added to the accuracy.
RANGE RESOLUTION ACCURACY
0.01Hz 0.005% +2
0.39% ~ 99.60% 0.390625%
0.4% of full scale
1/Frequency Range/256 0.2 ms+ Range/256
(3)
OPERATING CHARACTERISTICS
Measuring rate
Function Times/second
ACV 7
ACV + dB 7
DCV 7
ACV 7
AC + DC V 2
Ω/nS
14
Diode 14
Capacitance
4 (< 100 µF)
DCI 7
ACI 7
AC + DC I 2
Te m p e r at u r e 6
Frequency 2 (>10 Hz)
Duty cycle 1 (>10 Hz)
Pulse width 1 (>10 Hz)
DECIBEL (dB) CALCULATION
dB BASE REFERENCE DEFAULT REFERENCE
1 mW (dBm)
1 V (dBV) 1 V 1 V
1-9999 50
GENERAL SPECIFICATIONS
DISPLAY
Both primary and secondary displays are 5-digit liquid crystal display (LCD) with maximum reading of 50,000 counts. Automatic polarity indication.
POWER CONSUMPTION 105 mVA / 420 mVA (with backlit) maximum (U1251A)
165 mVA / 480 mVA (with backlit) maximum (U1252A)
OPERATING ENVIRONMENT
Full accuracy at –20 ° C to 55 ° C Full accuracy to 80% RH for temperature up to 35 ° C, decreasing linearly to 50%
RH at 55 °C Altitude:
0 - 2000 meters per IEC 61010-1 2
2000 - 3000 meters per IEC 61010-1 2
STORAGE COMPLIANCE
–40 ° C to 70 ° C
SAFETY COMPLIANCE
Certified by CSA for IEC/EN/CSA/UL 61010-1 2
MEASUREMENT CATEGORY CAT III 1000 V Overvoltage Protection up to 2000 m, Pollution degree 2
EMC COMPLIANCE
Certified to IEC/EN 61326: 2002, CISPR 11, and equivalents for Group 1, Class A
COMMON MODE REJECTION RATIO (CMRR)
> 90 dB at DC, 50/60 Hz + 0.1% (1 kΩ ubalanced)
NORMAL MODE REJECTION RATIO (NMRR)
> 60 dB at DC, 50/60 Hz + 0.1%
TEMPERATURE COEFFICIENT
0.15 x (specified accuracy) ° C (from –20 ° C to 18 ° C or 28 ° C to 55 °C)
SHOCK and VIBRATION Tested to IEC/EN 60068-2
DIMENSION (HxWxD)
203.5 mm x 94.4 mm x 59.0 mm
WEIGHT
504±5 grams with battery (U1251A) 527±5 grams with battery (U1252A)
CHARGING TIME (only U1252) <220 minutes approximately at the environment of 10 °C to 30 ° C.
WARRANTY
1 year factory + 2 years extended warranty
nd
Edition CAT III, 1000 V
nd
Edition CAT III, 600 V
nd
Edition
6
Page 7
Accessories Included
•Soft carrying case
• Alkaline 9 V battery (U1251A only)
• Rechargeable 7.2 V battery (U1252A only)
• Power cord & AC adaptor (U1252A only)
• U1160A Standard Test Lead Kit
•Quick Start Guide
• Reference CD containing the user’s guide, application
software and instrument drivers
•Certificate of Calibration (CoC)
• Test report
Optional Accessories (Sold seperately)
• U1161A Extension Test Lead Kit
• U1173A IR to USB cable
• U1180A Thermocouple Lead Kit
DIMENSION
7
Page 8
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system w e sell has a global warranty. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
www.agilent.com 800 123-4567
Sales, Service and Support
If you do not have access to the Internet, one of these centers can direct you to your nearest representative:
United States
1 800 123 4567
Our Promise
Our Promise means your Agilent test and measur ement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you receive your new Agilent equipment, we can help verify that it works properly and help with initial product operation .
Your Advantage
Your Advantage means that Agilent offers a
wide range of additional expert test and
measur ement serv ices, which you can
purchase according to your unique technical
and business needs. Solve problems
efficiently and gain a competitive edge by
contracting with us for calibration, extra-cost
upgrades, out-of-warranty repairs, and on-site
education and training, as w ell as design,
system integration, project management, and
other professional engineering services.
Experienced Agilent engineers and
technicians worldwide can help you maximize
your productivity, optimize the return on
investment of your Agilent instruments and
systems, and obtain dependable
measurement accuracy for the life of those
products.
Canada
1 877 123 4567 (905) 123 4567 (FAX)
Europe
(31 20) 123 4567 (31 20) 123 4567 (FAX)
Japan
(81) 123 45 6789 (81) 123 45 6789 (FAX)
Latin America
(305) 123 4567 (305) 123 4567 (FAX)
Australia
1 800 123 4567 (613) 123 4567 (FAX)
New Zealand
0800 123 4567 64 123 4567 (FAX)
Asia-Pacific
(852) 12 3 4567 (852) 12 3 4567 (FAX)
This informati on is subject to change with out notice.