Agilent U1191A User guide

Agilent U1190A Series Handheld Clamp Meters
User’s Guide
Agilent Technologies
Notices
CAUTION
WARNING
© Agilent Technologies, Inc., 2011
Manual Part Number
U1191-90003
Edition
First Edition, September 2011
Agilent Technologies, Inc. 5301, Stevens Creek Blvd. Santa Clara, CA 95051 USA
Warranty
The material contained in this document is provided “as is,” and is subject to change, without notice, in future editions. Further, to the maximum extent permitted by the applicable law, Agilent disclaims all war­ranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantabil­ity and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agi­lent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Safety Notices
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the likes of that, if not correctly performed or adhered to, could result in dam­age to the product or loss of impor­tant data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the likes of that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARN­ING notice until the indicated conditions are fully understood and met.
II U1190A Series User’s Guide
Safety Symbols
400 A
MAX
600 A
MAX
CAT III
600 V
CAT IV
300 V
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
DC (Direct current or voltage)
AC (Alternating current or voltage)
Both direct and alternating current
Earth (ground) terminal
Equipment protected throughout by double insulation or reinforced insulation
Caution, risk of electric shock Category IV 300 V overvoltage protection
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Application around and removal from HAZARDOUS LIVE conductors is permitted
U1191A/U1192A: Maximum allowable current measurement is 400 A
U1193A/U1194A: Maximum allowable current measurement is 600 A
Category III 600 V overvoltage protection
U1190A Series User’s Guide III
Safety Considerations
CAUTION
WARNING
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.
Use the proper terminals, function, and range for your measurements.
This device is for use at altitudes of up to 2,000 m.
Never measure voltage when current measurement is selected.
Always use the specified battery type. The power for the meter is
supplied with two standard AAA 1.5 V batteries. Observe the correct polarity markings before you insert the batteries to ensure proper insertion of the batteries in the meter.
IV U1190A Series User’s Guide
• Do not use the meter if it is damaged. Before you use the meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
• Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged test leads before you use the meter.
• Do not operate the meter around explosive gas, vapor, or wet environments.
• Do not apply more than the rated voltage and current (as marked on the meter) between the terminals or between the terminal and the earth ground.
WARNING
• Never use the meter in wet conditions or when there is water on the surface. If the meter is wet, ensure that the meter is dried only by trained personnel.
• Before use, verify the meter's operation by measuring a known voltage. Use caution when working above 60 V DC, 30 V AC RMS, or 42.4 V peak. Such voltages pose a shock hazard.
• When measuring current, turn off the circuit power before connecting the meter in the circuit. Remember to place the meter in series with the circuit.
• When measuring temperature, keep the thermocouple probe as close to the meter as possible, and avoid contact with surfaces above 60 V DC, 30 V AC RMS, or 42.4 V peak. Such voltages pose a shock hazard.
• When servicing the meter, use only the specified replacement parts.
• When using the probes, keep your fingers behind the finger guards on the probes.
• Connect the common test lead before you connect the live test lead. When you disconnect the leads, disconnect the live test lead first.
U1190A Series User’s Guide V
• Remove the test leads from the meter before you open the battery cover. Do not operate the meter with the battery cover or portions of the cover removed or loosened.
• To avoid false readings, which may lead to possible electric shock or personal injury, replace the battery as soon as the low battery indicator appears and flashes.
• Use the meter only as specified in this guide. Otherwise, the protection provided by the meter may be impaired.
• Individual protective equipment must be used if hazardous live parts in the installation are accessible where measurement is to be carried out.
• The tactile indicator or barrier, indicates the limit of safe access of the handheld part.
Environmental Conditions
NOTE
This instrument is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
Environmental conditions Requirements
Operating temperature –10 °C to 50 °C
Up to 80% RH (relative humidity) for temperature
Operating humidity
up to 30 °C, decreasing linearly to 50% RH at 50 °C
Storage temperature
Altitude Up to 2000 meters
Pollution degree Pollution degree 2
The U1190A Series Handheld Clamp Meter complies with the following safety and EMC requirements:
IEC 61010-1:2001/EN 61010-1:2001
IEC 61010-2-032:2002/EN 61010-2-032:2002
CAN/CSA-C22.2 No. 61010-1-04
CAN/CSA-C22.2 No. 61010-2-032-04
ANSI/UL Std. No. 61010-1:2004
IEC61326-1:2005/EN61326-1:2006
Canada: ICES/NMB-001: Issue 4, June 2006
Australia/New Zealand: AS/NZS CISPR 11:2004
–40 °C to 60 °C, 40% to 80% RH (without batteries)
VI U1190A Series User’s Guide
Regulatory Markings
The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
The C-tick mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est confomre a la norme NMB-001 du Canada.
The CSA mark is a registered trademark of the Canadian Standards Association.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
U1190A Series User’s Guide VII
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Agilent Service Center, or visit
www.agilent.com/environment/product
for more information.
VIII U1190A Series User’s Guide
Declaration of Conformity (DoC)
NOTE
The Declaration of Conformity (DoC) for this instrument is available on the Agilent website. You can search the DoC by its product model or description at the web address below.
http://regulations.corporate.agilent.com/DoC/search.htm
If you are unable to search for the respective DoC, please contact your local Agilent representative.
U1190A Series User’s Guide IX
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
X U1190A Series User’s Guide

Table of Contents

1 Introduction
About This Manual 2
Documentation map 2 Safety notes 2
Preparing Your Clamp Meter 3
Checking the shipment 3 Installing the batteries 3 Turning on your clamp meter 6 Automatic Power-Off (APO) 7 Enabling the backlight 7 Enabling the flashlight 8 Alerts and warnings during measurement 9 Power-on options 10
Your Clamp Meter in Brief 11
Dimensions 11 Overview 13 Rotary switch 15 Keypad 17 Display screen 18 Input terminals 21
Cleaning Your Clamp Meter 23
2 Making Measurements
Measuring AC or DC Current 26
Measuring AC Voltage 30
Measuring DC Voltage 32
Measuring Resistance 34
Testing for Continuity 36
U1190A Series User’s Guide XI
Testing Diodes 39
Measuring Capacitance 43
Measuring Temperature 45
Measuring AC or DC Current (up to µA) 48
Measuring Frequency 50
3 Clamp Meter Features
Detecting AC Voltage Presence (Vsense) 54
Making Relative Measurements (Null) 57
Capturing Maximum and Minimum Values (Max.Min) 58
Freezing the Display (Hold) 60
4 Characteristics and Specifications
Product Characteristics 62
Specification Assumptions 64
Measurement Category 65
Measurement category definition 65
Electrical Specifications 66
DC specifications 66 AC specifications 68 Capacitance specifications 69 Temperature specifications 70 Frequency specifications 71 Frequency sensitivity specifications 71 Display update rate (approximate) 72
XII U1190A Series User’s Guide

List of Figures

Figure 1-1 Installing the batteries 4 Figure 1-2 Powering on the clamp meter 6 Figure 1-3 Front panel 13 Figure 1-4 Rear panel 14 Figure 2-1 AC current display 27 Figure 2-2 Measuring AC current 27 Figure 2-3 Wire separator and hook design 28 Figure 2-4 AC voltage display 30 Figure 2-5 Measuring AC voltage 31 Figure 2-6 DC voltage display 32 Figure 2-7 Measuring DC voltage 33 Figure 2-8 Resistance display 34 Figure 2-9 Measuring resistance 35 Figure 2-10 Open continuity display 36 Figure 2-11 Closed continuity display 36 Figure 2-12 Testing for continuity 38 Figure 2-13 Diode display 39 Figure 2-14 Open diode display 40 Figure 2-15 Testing forward-bias diode 41 Figure 2-16 Testing reverse-bias diode 42 Figure 2-17 Capacitance display 43 Figure 2-18 Measuring capacitance 44 Figure 2-19 Temperature display 45 Figure 2-20 Measuring the surface temperature 47 Figure 2-21 DC current display 48 Figure 2-22 Measuring AC/DC current (up to µA) 49 Figure 2-23 Frequency display 51 Figure 2-24 Measuring frequency 51 Figure 3-1 Vsense (high sensitivity) display 55 Figure 3-2 Vsense (low sensitivity) display 55 Figure 3-3 Detecting voltage presence 56 Figure 3-4 Null display 57 Figure 3-5 Max.Min display 58 Figure 3-6 Hold display
60
U1190A Series User’s Guide XIII
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
XIV U1190A Series User’s Guide

List of Tables

Tab le 1- 1 Battery level indicator 5 Tab le 1- 2 Power-on options 10 Tab le 1- 3 Rotary switch functions 16 Tab le 1- 4 Keypad functions 17 Tab le 1- 5 General annunciators 18 Tab le 1- 6 Measurement units display 20 Tab le 1- 7 Terminal connections for different measuring
functions 21
Tab le 4- 1 DC specifications 66 Tab le 4- 2 AC voltage specifications 68 Tab le 4- 3 AC current specifications 68 Tab le 4- 4 Capacitance specifications 69 Tab le 4- 5 Temperature specifications 70 Tab le 4- 6 Frequency specifications 71 Tab le 4- 7 Frequency sensitivity and trigger-level specifications for
voltage measurements 71
Tab le 4- 8 Frequency sensitivity and trigger level specifications for
current measurements 72
Tab le 4- 9 Display update rate (approximate) 72
U1190A Series User’s Guide XV
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
XVI U1190A Series User’s Guide
U1190A Series Handheld Clamp Meter User’s Guide

1 Introduction

About This Manual 2
Documentation map 2 Safety notes 2
Preparing Your Clamp Meter 3
Checking the shipment 3 Installing the batteries 3 Turning on your clamp meter 6 Automatic Power-Off (APO) 7 Enabling the backlight 7 Enabling the flashlight 8 Alerts and warnings during measurement 9 Power-on options 10
Your Clamp Meter in Brief 11
Dimensions 11 Overview 13 Rotary switch 15 Keypad 17 Display screen 18 Input terminals 21
Cleaning Your Clamp Meter 23
This chapter teaches you how to set up your clamp meter for the first time. An introduction to all the features of the clamp meter is also given.
Agilent Technologies
1
1 Introduction

About This Manual

About This Manual
The descriptions and instructions in this manual apply to the Agilent U1191A, U1192A, U1193A, and U1194A handheld clamp meters (hereafter referred to as the clamp meter).
The model U1194A appears in all illustrations.

Documentation map

The following manuals and software are available for your clamp meter. For the very latest version, please visit our website at: http://www.agilent.com/find/hhTechLib.
Check the manual revision on the first page of each manual.
• User’s Guide. This manual.
•Quick Start Guide. Printed copy for outdoor use,
included with shipment.
• Service Guide. Free download at the Agilent website.

Safety notes

Safety notes are used throughout this manual (see the
“Safety Notices” section for format examples). Familiarize
yourself with each of the notes and its meaning before operating your clamp meter.
More pertinent safety notes for using this product are located under the “Safety Considerations” section.
Do not proceed beyond a safety notice until the indicated conditions are fully understood and met.
2 U1190A Series User’s Guide

Preparing Your Clamp Meter

CAUTION

Checking the shipment

When you receive your clamp meter, check the shipment according to the following procedure.
1 Inspect the shipping container for damage. Signs of
damage may include a dented or torn shipping container or cushioning material that indicates signs of unusual stress or compacting. Save the packaging material in case the clamp meter needs to be returned.
2 Carefully remove the contents from the shipping
container, and verify that the standard accessories and your ordered options are included in the shipment according to the standard shipped items list found in the printed copy of the U1190A Series Quick Start Guide.
3 For any question or problems, refer to the Agilent contact
numbers on the back of this manual.
Introduction 1
Preparing Your Clamp Meter
U1190A Series User’s Guide 3

Installing the batteries

Your clamp meter is powered by two 1.5 V AAA alkaline batteries (included with the shipment). When you receive your clamp meter, the AAA alkaline batteries are not installed.
Use the following procedure to install the batteries.
Before you proceed with the batteries installation, remove all cable connections to the terminals and ensure that the rotary switch is at the OFF position. Use only the battery type specified in the “Product
Characteristics” on page 62.
1 Introduction
Preparing Your Clamp Meter
1 Open the battery cover. Loosen the screw with a suitable
Phillips screwdriver and remove the battery cover as shown in Figure 1-1.
2 Insert the batteries. Observe the proper battery polarity.
The terminal ends of each battery are indicated inside the battery compartment.
3 Close the battery cover. Place the battery cover back in
its original position and tighten the screw.
Figure 1-1 Installing the batteries
The battery level indicator in the lower left- hand corner of the display indicates the relative condition of the batteries.
Table 1- 1 describes the various battery levels the indicator
represents.
4 U1190A Series User’s Guide
Tab l e 1 - 1 Battery level indicator
WARNING
Indication Battery capacity
Full capacity
2/3 capacity
1/3 capacity
Introduction 1
Preparing Your Clamp Meter
(Flashing periodically)
[1] Batteries change advised. Always use the specified battery type listed in the
“Product Characteristics” on page 62.
Almost empty
[1]
To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the low battery indicator appears. Do not discharge the batteries by shorting the batteries or reversing the polarity of the batteries.
U1190A Series User’s Guide 5
1 Introduction
Preparing Your Clamp Meter

Turning on your clamp meter

To power ON your clamp meter, turn the rotary switch from the OFF position to any other position.
Figure 1-2 Powering on the clamp meter
To power OFF your clamp meter, turn the rotary switch to the OFF position.
6 U1190A Series User’s Guide

Automatic Power-Off (APO)

NOTE
Your clamp meter automatically turns off if the rotary switch is not moved or a key is not pressed for 15 minutes (default). The clamp meter will beep thrice before it powers off. Pressing any key or turning the rotary switch to a new position will turn the clamp meter back on after it is powered off automatically.
The symbol is shown on the bottom left of the display when the automatic power-off function is enabled.
Follow the steps below to change the timer period or completely disable the automatic power-off.
1 Press and hold while powering on the clamp meter to
enter the Setup menu.
2 The automatic power- off (A #) timer period is the first
Setup item shown.
3 Press to make changes to the A # value.
Press or to change the timer period (from A 01 to A 99 minutes) or to completely disable the automatic power-off function (AoFF).
4 Press to save the changes, or press to discard
the changes and proceed to the next Setup item.
5 Cycle the clamp meter’s power to exit the Setup menu.
Introduction 1
Preparing Your Clamp Meter

Enabling the backlight

If viewing the display becomes difficult in low- light conditions, press (on the U1192A/U1193A/U1194A models) or (on the U1191A model) to activate the LCD backlight.
Press / again to deactivate the LCD backlight.
To conserve battery life, a user-adjustable timer controls how long the backlight stays on. The default timer period is 15 seconds.
U1190A Series User’s Guide 7
1 Introduction
WARNING
NOTE
Preparing Your Clamp Meter
Follow the steps below to change the timer period or completely disable the backlight timer.
1 Press and hold while powering on the clamp meter to
enter the Setup menu.
2 Press again. The backlight (b #) timer period is the
second Setup item shown.
3 Press to make changes to the b # value.
Press or to change the timer period (from b 01 to b 99 seconds) or to completely disable the backlight timer function (boFF).
4 Press to save the changes, or press to discard
the changes and proceed to the next Setup item.
5 Cycle the clamp meter’s power to exit the Setup menu.

Enabling the flashlight

This feature is applicable for U1192A, U1193A, and U1194A models only.
8 U1190A Series User’s Guide
If you are using the clamp meter in dark places, press and hold for more than 1 second to activate the LED flashlight for greater visibility on your test points. This is not applicable to the U1191A model.
Press for more than 1 second to deactivate the LED flashlight.
VISION ADVISORY CLAIM
It is advised that you do not look directly into the light source of the LED flashlight. As with any source of bright light, prolonged exposure can damage the eye.
To conserve battery life, a user-adjustable timer can be set to control how long the flashlight stays on. This function is disabled by default.
Preparing Your Clamp Meter
WARNING
Follow the steps below to change the timer period or completely disable the flashlight timer.
1 Press and hold while powering on the clamp meter to
enter the Setup menu.
2 Press twice. The f lashlight (t #) timer period is the
third Setup item shown.
3 Press to make changes to the t # value.
Press or to change the timer period (from t 01 to t 99 seconds) or to completely disable the flashlight timer function (toFF).
4 Press to save the changes, or press to discard
the changes and proceed to the next Setup item.
5 Cycle the clamp meter’s power to exit the Setup menu.

Alerts and warnings during measurement

Voltage alert
Introduction 1
U1190A Series User’s Guide 9
For your own safety, please do not ignore the voltage alert. When the clamp meter cautions you with a voltage alert, you are advised to take note of the existence of high voltage and pay closer attention when performing measurements.
Your clamp meter provides a voltage alert for voltage measurements. The clamp meter starts beeping periodically once the measured voltage exceeds the alert value (regardless of polarity) set.
Hazardous voltage indication
The clamp meter will also display the hazardous voltage ( ) symbol as an early precaution when the measured voltage is equal to or greater than 30 V in all voltage measurement modes.
1 Introduction
+
Preparing Your Clamp Meter

Power-on options

Some options can be selected only while you turn the clamp meter on. These power- on options are listed in the table below.
To select a power- on option, press and hold the specified key in Table 1- 2 while turning the rotary switch from the
OFF position to any other position.
Tab l e 1 - 2 Power-on options
Key Description
Tests the LCD.
All annunciators are displayed in the LCD for 10 seconds. Cycle the clamp meter’s power to exit this mode, or wait 10 seconds for the clamp meter to return to normal operation.
Checks the firmware version.
The clamp meter’s firmware version will be shown on the primary display. Cycle the clamp meter’s power to exit this mode.
Enters the clamp meter’s Setup menu.
See the following topics for more information on each respective Setup menu item.
“Automatic Power-Off (APO)” on page 7
“Enabling the backlight” on page 7
“Enabling the flashlight” on page 8
“Changing the continuity visual alert” on page 37
Cycle the clamp meter’s power to exit the Setup menu.
Enters the unit selection menu for temperature measurements (U1194A only).
See “Changing the default temperature unit” on page 46 for more information. Cycle the clamp meter’s power to exit this menu.
10 U1190A Series User’s Guide

Your Clamp Meter in Brief

77.14 mm
77.14 mm

Dimensions

Front view
Introduction 1
Your Clamp Meter in Brief
U1190A Series User’s Guide 11
1 Introduction
225 mm
238 mm
Your Clamp Meter in Brief
Rear view
12 U1190A Series User’s Guide

Overview

Clamp jaw
Hold/MaxMin button
Rotary switch
Display screen
Function buttons
Input terminals
Jaw markings
Introduction 1
Your Clamp Meter in Brief
Front panel
The front panel parts of your clamp meter are described in this section.
U1190A Series User’s Guide 13
Figure 1-3 Front panel
1 Introduction
Flashlight
Hand guard
Handle
Battery cover
Your Clamp Meter in Brief
Rear panel
The rear panel parts of your clamp meter are described in this section.
Figure 1-4 Rear panel
14 U1190A Series User’s Guide

Rotary switch

NOTE
WARNING
NOTE
Introduction 1
Your Clamp Meter in Brief
The measurement functions for each rotary switch position are described in Table 1- 3 on page 16. Turning the rotary switch changes the measurement function and resets all other measurement options.
Some rotary switch positions have a shifted function printed in orange. Press to switch between the shifted and primary function.
Remove the test leads from the measuring source or target before changing the rotary switch position.
Each position of the U1191A, U1192A, U1193A, and U1194A rotar y switches (shown in Figure 1- 3) is described in
Table 1- 3. Click the respective “Learn more” pages for more
information on each function.
U1190A Series User’s Guide 15
A list of some of the abbreviations used in Table 1-3 is given below.
•AC A: AC current measurement
•DC A: DC current measurement
•AC V: AC voltage measurement
•DC V: DC voltage measurement
•AC A: AC current measurement (up to microamperes)
•DC A: DC current measurement (up to microamperes)
1 Introduction
Your Clamp Meter in Brief
Tab l e 1 - 3 Rotary switch functions
Legend Functions shown in the primary display U1194A U1193A U1192A U1191A Learn more on:
Off
AC A
DC A
Frequency (current path)
AC V
Frequency (voltage path)
DC V
Resistance
Continuity
Diode
Capacitance
Non-contact voltage detector
Te m p e r a t u r e
DC A
AC A
✔✔✔✔
✔✔✔✔
✔✔✔
---
- page 50
✔✔✔✔
✔✔✔
- page 50
✔✔✔✔
✔✔✔✔
✔✔✔✔
✔✔✔✔
✔✔✔
✔✔✔
---page 45
---
---
- page 43
- page 54
page 6
page 26
page 30
page 32
page 34
page 36
page 39
page 48
16 U1190A Series User’s Guide

Keypad

Tab l e 1 - 4 Keypad functions
Introduction 1
Your Clamp Meter in Brief
The operation of each key is explained below. Pressing a key enables a function, displays a related symbol, and emits a beep. Turning the rotary switch to another position resets the current operation of the key.
Click the respective “Learn more” pages for more information on each function.
Legend
Function when pressed for:
Less than 1 second More than 1 second
Freezes the present reading in the display.
Switches between the primary and
shifted (icons printed in orange)
functions.
Turns the LCD backlight on or off. Turns the LED flashlight on or off.
U1191A only:
Turns the LCD backlight on or off.
Sets the null/relative mode. - page 57
Records the maximum, minimum, or average value.
- page 15
- page 7
Learn more on:
page 60
page 7
and
page 8
U1190A Series User’s Guide 17
1 Introduction
Your Clamp Meter in Brief

Display screen

The display annunciators of your clamp meter are described in this section. See also “Measurement units” on page 20 for a list of available measurement signs and notations.
General display annunciators
The general display annunciators of your clamp meter are described in the table below.
Tab l e 1 - 5 General annunciators
Legend Description Learn more on:
Hazardous voltage sign for measuring voltage 30 V or overload
Hold enabled page 60
Auto-ranging enabled -
Maximum reading shown on primary display
Minimum reading shown on primary display
Averaged reading shown on primary display
Present reading shown on primary display
Relative (Null) enabled page 57
Diode test selected page 39
Audible continuity test selected page 36
DC (direct current) indication
AC (alternating current) indication
page 9
page 58
page 26
and page 32
page 26
and page 30
18 U1190A Series User’s Guide
Introduction 1
Your Clamp Meter in Brief
Tab l e 1 - 5 General annunciators (continued)
Legend Description Learn more on:
Battery capacity indication page 5
APO (Auto Power-Off) enabled page 7
Primary measurement display -
Measuring units page 20
Overload (the reading exceeds the display range)
-
U1190A Series User’s Guide 19
1 Introduction
Your Clamp Meter in Brief
Measurement units
The available signs and notations for each measurement function in your clamp meter are described in Table 1- 6. The units listed below are applicable to the primary display measurements of your clamp meter.
Tab l e 1 - 6 Measurement units display
Sign/Notation Description
M Mega 1E+06 (1000000)
k kilo 1E+03 (1000)
n nano 1E–09 (0.000000001)
micro 1E–06 (0.000001)
m milli 1E–03 (0.001)
mV, V Voltage units for voltage measurement
A, mA, A Ampere units for current measurement
nF, F, mF Farad units for capacitance measurement
, k, M Ohm units for resistance measurement
Hz, kHz, MHz Hertz units for frequency measurement
°C Degree Celsius, unit for temperature measurement
°F Degree Fahrenheit, unit for temperature measurement
20 U1190A Series User’s Guide

Input terminals

WARNING
CAUTION
The terminal connections for the different measurement functions of your clamp meter are described in the table below. Observe the rotary switch position of your clamp meter before connecting the test leads to the connector terminals.
Ensure that the probe accessories are connected to the correct input terminals for the selected measurement function before starting any measurement.
To avoid damaging this device, do not exceed the rated input limit.
Tab l e 1 - 7 Terminal connections for different measuring functions
Introduction 1
Your Clamp Meter in Brief
Legend Functions U1194A U1193A U1192A U1191A Input terminals Overload protection
U1190A Series User’s Guide 21
AC V
Frequency (voltage path)
DC V
Diode
Capacitance
Resistance
Continuity
Non-contact voltage detector
Te m p e r a t u r e
✔✔✔✔
✔✔✔
-
✔✔✔✔
✔✔✔✔
✔✔✔
-
✔✔✔✔
✔✔✔✔
✔✔✔
---
-
600 Vrms
600 Vrms for short
circuit current <0.4 A
600 Vrms for short
circuit current <0.1 A
1 Introduction
Your Clamp Meter in Brief
Tab l e 1 - 7 Terminal connections for different measuring functions
Legend Functions U1194A U1193A U1192A U1191A Input terminals Overload protection
DC A
AC A
AC A
DC A
Frequency (current path)
---
---
✔✔✔✔
✔✔✔
---
-
CAT III 600 V
600 Arms
22 U1190A Series User’s Guide

Cleaning Your Clamp Meter

WARNING
To avoid electrical shock or damage to the clamp meter, ensure that the insides of the casing stay dry at all times.
Dirt or moisture in the terminals can distort readings. Follow the steps below to clean your clamp meter.
1 Turn the clamp meter off, and remove the test leads.
2 Turn the clamp meter over, and shake out any dirt that
may have accumulated in the terminals.
Wipe the case with a damp cloth and mild detergent — do not use abrasives or solvents. Wipe the contacts in each terminal with a clean swab dipped in alcohol.
Introduction 1
Cleaning Your Clamp Meter
U1190A Series User’s Guide 23
1 Introduction
Cleaning Your Clamp Meter
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24 U1190A Series User’s Guide
U1190A Series Handheld Clamp Meter User’s Guide

2 Making Measurements

Measuring AC or DC Current 26 Measuring AC Voltage 30 Measuring DC Voltage 32 Measuring Resistance 34 Testing for Continuity 36 Tes ti n g D io de s 39 Measuring Capacitance 43 Measuring Temperature 45 Measuring AC or DC Current (up to µA) 48 Measuring Frequency 50
This chapter describes how to take measurements with your clamp meter.
Agilent Technologies
25
2 Making Measurements
WARNING
CAUTION
Current flow
Current flow
For best accuracy during current measurements, align the cable with the jaw markings.
Correct Incorrect
NOTE

Measuring AC or DC Current

Measuring AC or DC Current
Ensure that the test leads are disconnected from the input terminals when measuring current with the clamp jaws.
Ensure that the clamp meter measures only one conductor at a time. Measuring multiple conductors may cause inaccuracy in measurement readings due to the vector sum of currents flowing in the conductors.
Use the wire separator to separate individual wires or cables from a mesh of wires or cables. See “Using the wire separator and hook” on page 28 for more information.
Set up your clamp meter to measure AC current or DC current (U1194A model only) as shown in Figure 2- 2. Clamp the wire/cable, and read the display.
Press to measure the frequency of the AC current source (U1192A, U1193A, and U1194A models only). See “Measuring Frequency” on page 50 to learn more.
26 U1190A Series User’s Guide
Making Measurements 2
3
1
2
4
5
Conductor
Measuring AC or DC Current
Figure 2-1 AC current display
Figure 2-2 Measuring AC current
U1190A Series User’s Guide 27
2 Making Measurements
Wire separator tip
Hook
Measuring AC or DC Current
Using the wire separator and hook
Your clamp meter’s design has a wire separator and hook feature (see Figure 2- 3) that can be used to separate individual wires or cables for measurements from a mesh of wires or cables.
Use the wire separator and hook to avoid touching live wires or cables without the necessary insulation protection or powering off the voltage or current source.
Follow the instructions below to use the wire separator and hook feature.
Figure 2-3 Wire separator and hook design
28 U1190A Series User’s Guide
Making Measurements 2
Mesh of wires or cables
Measuring AC or DC Current
1 Open the clamp jaw slightly to reveal the wire separator
tip. Use the wire separator tip to locate the desired wire or cable.
2 Separate the wire or cable by hooking it in the clamp jaw
securely and pulling it back.
3 Close the clamp jaw to secure the wire or cable, and read
the measurement.
U1190A Series User’s Guide 29
2 Making Measurements
NOTE
NOTE
NOTE

Measuring AC Voltage

Measuring AC Voltage
Reversing the leads will produce a negative reading, but it will not damage the clamp meter.
Set up your clamp meter to measure AC voltage as shown in
Figure 2- 5. Probe the test points, and read the display.
For U1193A and U1194A models:
AC voltage measurements measured with this clamp meter are returned as true RMS (root mean square) readings. These readings are accurate for sinusoidal waves. For non-sinusoidal waveforms, please refer to the
“Specification Assumptions” on page 64.
Figure 2-4 AC voltage display
Press to measure the frequency of the AC voltage source (U1192A, U1193A, and U1194A models only). See “Measuring Frequency” on page 50 to learn more.
30 U1190A Series User’s Guide
Making Measurements 2
2
4
AC
Voltage source
1
3
Measuring AC Voltage
Figure 2-5 Measuring AC voltage
U1190A Series User’s Guide 31
2 Making Measurements
NOTE

Measuring DC Voltage

Measuring DC Voltage
Set up your clamp meter to measure DC voltage as shown in
Figure 2- 7. Probe the test points, and read the display.
This clamp meter displays DC voltage values as well as their polarity. Negative DC voltages will return a negative sign on the left of the display.
Figure 2-6 DC voltage display
32 U1190A Series User’s Guide
Making Measurements 2
DC
3
Voltage source
1
2
Measuring DC Voltage
Figure 2-7 Measuring DC voltage
U1190A Series User’s Guide 33
2 Making Measurements
CAUTION
NOTE
NOTE

Measuring Resistance

Measuring Resistance
Set up your clamp meter to measure resistance as shown in
Figure 2- 9. Probe the test points, and read the display.
To avoid possible damage to your clamp meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before measuring resistance.
Resistance (opposition to the current flow) is measured by sending a small current out through the test leads to the circuit under test. Because this current flows through all possible paths between the leads, the resistance reading represents the total resistance of all paths between the leads. Resistance is measured in ohms ().
34 U1190A Series User’s Guide
Figure 2-8 Resistance display
Keep the following in mind when measuring resistance.
The test leads can add 0.1  to 0.2 of error to resistance
measurements. To test the leads, touch the probe tips together and read the resistance of the leads.
Because the clamp meter’s test current flows through all possible
paths between the probe tips, the measured value of a resistor in a circuit is often different from the resistor’s rated value.
The resistance function can produce enough voltage to forward-bias
silicon diodes or transistor junctions, causing them to conduct.
Making Measurements 2
3
Resistor
1
2
Measuring Resistance
Figure 2-9 Measuring resistance
U1190A Series User’s Guide 35
2 Making Measurements
CAUTION
NOTE

Testing for Continuity

Testing for Continuity
Set up your clamp meter to test for continuity as shown in
Figure 2- 12. Probe the test points, and read the display.
To avoid possible damage to your clamp meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before testing for continuity.
Continuity is the presence of a complete path for current flow. The continuity test features a beeper that sounds and a backlight that flashes as long as a circuit is complete. The audible and visual alert allows you to perform quick continuity tests without having to watch the display.
Press to switch between resistance measurement, or continuity test. See Figure 2- 12 to learn more.
36 U1190A Series User’s Guide
Figure 2-10 Open continuity display
Figure 2-11 Closed continuity display
Making Measurements 2
NOTE
Testing for Continuity
You can set the beeper to sound and the backlight to flash as a
continuity indication whether the circuit-under-test is less than (short) the threshold resistance.
The continuity function detects intermittent shorts lasting as short as 1
ms. A brief short causes the multimeter to emit a short beep and flash.
You can enable or disable the visual alert via the Setup menu. See
“Changing the continuity visual alert” on page 37 for more information.
Changing the continuity visual alert
You can set the backlight to flash along with the beeper sound as a continuity indication whether the circuit- under-test is less than the threshold resistance.
Follow the steps below to enable or disable the continuity visual alert.
1 Press and hold while powering on the clamp meter to
enter the Setup menu.
2 Press again. The continuity visual alert ( ) is the
fourth Setup item shown.
3 Press to make changes to the continuity visual alert.
Press or to enable or to disable the continuity visual alert (the backlight turns on or off).
4 Press to save the changes, or press to discard
the changes and proceed to the next Setup item.
5 Cycle the clamp meter’s power to exit the Setup menu.
U1190A Series User’s Guide 37
2 Making Measurements
4
1
3
CLOSED
OPEN
2
Testing for Continuity
Figure 2-12 Testing for continuity
38 U1190A Series User’s Guide

Testing Dio d e s

CAUTION
NOTE
NOTE
Making Measurements 2
Testing Diodes
Set up your clamp meter to test diodes as shown in
Figure 2- 15. Probe the test points, and read the display.
To avoid possible damage to your clamp meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before testing diodes.
Use the diode test to check diodes, transistors, silicon controlled
rectifiers (SCRs), and other semiconductor devices. A good diode allows current to flow in one direction only.
This test sends a current through a semiconductor junction, and then
measures the junction’s voltage drop.
Connect the red test lead to the positive terminal (anode) of the diode
and the black test lead to the negative terminal (cathode). The cathode of a diode is indicated with a band.
U1190A Series User’s Guide 39
Figure 2-13 Diode display
Your clamp meter can display the forward-bias of a diode up to approximately 1.8 V. The forward-bias of a typical diode is within the range of 0.3 V to 0.8 V; however, the reading can vary depending on the resistance of other pathways between the probe tips.
2 Making Measurements
NOTE
Testing Diodes
If the beeper is enabled during diode test, the clamp meter will beep briefly for a normal junction and sound continuously for a shorted junction.
Reverse the probes (as shown in Figure 2- 16) and measure the voltage across the diode again. Assess the diode according to the following guidelines:
A diode is considered good if the clamp meter displays
in reverse- bias mode.
A diode is considered shorted if the clamp meter
displays approximately 0 V in both forward and reverse- bias modes, and the clamp meter beeps continuously.
A diode is considered open if the clamp meter displays
in both forward- and reverse- bias modes.
Figure 2-14 Open diode display
40 U1190A Series User’s Guide
Making Measurements 2
3
Forward-bias diode
1
2
Testing Diodes
Figure 2-15 Testing forward-bias diode
U1190A Series User’s Guide 41
2 Making Measurements
3
Reverse-bias diode
1
2
Testing Diodes
Figure 2-16 Testing reverse-bias diode
42 U1190A Series User’s Guide

Measuring Capacitance

CAUTION
NOTE
NOTE
This measurement function is applicable for U1192A, U1193A, and U1194A models only.
Set up your clamp meter to measure capacitance as shown in Figure 2- 18. Probe the test points, and read the display.
To avoid possible damage to the clamp meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance. Use the DC voltage function to confirm that the capacitor is fully discharged.
The clamp meter measures capacitance by charging the capacitor with a known current for a known period of time, measuring the resulting voltage, and then calculating the capacitance.
Making Measurements 2
Measuring Capacitance
U1190A Series User’s Guide 43
Figure 2-17 Capacitance display
For measuring capacitance values greater than 1000 F, discharge the capacitor first, then select a suitable range for measurement. This will speed up the measurement time and also ensure that the correct capacitance value is obtained.
2 Making Measurements
2
4
Capacitor
1
3
Measuring Capacitance
Figure 2-18 Measuring capacitance
44 U1190A Series User’s Guide

Measuring Temperature

WARNING
CAUTION
NOTE
This measurement function is applicable for the U1194A model only.
Set up your clamp meter to measure temperature as shown in Figure 2- 20. Probe the test points, and read the display.
Do not connect the thermocouple to electrically live circuits. Doing so will potentially cause fire or electric shock.
Do not bend the thermocouple leads at sharp angles. Repeated bending over a period of time can break the leads.
The clamp meter uses the type-K thermocouple probe (included in the
The approximate ambient temperature (cold-junction compensation) is
Shorting the terminal to the terminal will display the
Making Measurements 2
Measuring Temperature
standard shipped items of a U1194A model) for measuring temperature.
shown on the display when you have an open thermocouple. The open thermocouple message may be due to a broken (open) probe or because no probe is installed into the input jacks of the clamp meter.
temperature at the clamp meter’s terminals.
U1190A Series User’s Guide 45
Figure 2-19 Temperature display
2 Making Measurements
CAUTION
NOTE
Measuring Temperature
Changing the default temperature unit
Follow the steps below to change the temperature unit between Celsius (°C) or Fahrenheit (°F).
1 Press and hold and while powering on the clamp
meter to enter the temperature unit selection menu.
2 Press to make changes to the temperature unit.
Press or to change the temperature unit (°C or °F).
3 Press to save the changes.
4 Cycle the clamp meter’s power to exit the temperature
unit selection menu.
Always set the temperature unit display per the official requirements and in compliance with the national laws of your region.
The bead-type thermocouple probe is suitable for measuring temperatures from –40 °C to 204 °C (399 °F) in Teflon-compatible environments. Do not immerse this thermocouple probe in any liquid. For best results, use a thermocouple probe designed for each specific application — an immersion probe for liquid or gel, and an air probe for air measurement.
Observe the following measurement techniques:
Clean the surface to be measured, and ensure that the probe is
securely touching the surface. Remember to disable the applied power.
When measuring above ambient temperatures, move the thermocouple
along the surface until you get the highest temperature reading.
When measuring below ambient temperatures, move the thermocouple
along the surface until you get the lowest temperature reading.
46 U1190A Series User’s Guide
Place the clamp meter in the operating environment for at least 1 hour
as the clamp meter is using a non-compensation transfer adapter with miniature thermal probe.
Making Measurements 2
2
4
1
3
K-type thermocouple probe
Heat source
Measuring Temperature
Figure 2-20 Measuring the surface temperature
U1190A Series User’s Guide 47
2 Making Measurements
WARNING
CAUTION

Measuring AC or DC Current (up to µA)

Measuring AC or DC Current (up to µA)
Never attempt an in-circuit current measurement where the open-circuit potential to earth is greater than 1000 V. Doing so will cause damage to the clamp meter and possible electric shock or personal injury.
To avoid possible damage to the clamp meter or to the equipment
under test, use the proper terminals, function, and range for your measurement. Use the clamp jaw for currents above 600 µA.
To measure current, you must open the circuit under test, then place
the clamp meter in series with the circuit.
across (in parallel with) any circuit or component when the leads are plugged into the current terminals.
Placing the probes across (in parallel with) a powered circuit when
a lead is plugged into a current terminal can damage the circuit you are testing. This happens because the resistance through the clamp meter's current terminals is very low, resulting in a short circuit.
Never place the probes
48 U1190A Series User’s Guide
This measurement function is applicable for the U1194A model only.
Set up your clamp meter to measure AC or DC current (up to A) as shown in Figure 2- 22. Probe the test points, and read the display.
Figure 2-21 DC current display
Making Measurements 2
AC
2
4
1
Voltage source
DC
Voltage source
LOAD
3
LOAD
Measuring AC or DC Current (up to µA)
Figure 2-22 Measuring AC/DC current (up to µA)
U1190A Series User’s Guide 49
2 Making Measurements
WARNING
NOTE
Rise Time Fall Time
+ Width – Width
Period
90%
50%
10%

Measuring Frequency

Measuring Frequency
Never measure the frequency where the voltage or current level exceeds the specified range.
This measurement function is applicable for U1192A, U1193A, and U1194A models only.
Your clamp meter allows simultaneous monitoring of real- time voltage or current with frequency measurements.
Measuring the frequency of a signal helps detect the presence of
harmonic currents in neutral conductors and determines whether these neutral currents are the result of unbalanced phases or non-linear loads.
Frequency is the number of cycles a signal completes each second.
Frequency is defined as 1/Period. Period is defined as the time between the middle threshold crossings of two consecutive, like-polarity edges, as shown in the figure below.
50 U1190A Series User’s Guide
The clamp meter measures the frequency of a voltage or current signal
by counting the number of times the signal crosses a threshold level within a specified period of time.
Making Measurements 2
3
1
2
4
Conductor
Measuring Frequency
Figure 2-23 Frequency display
Figure 2-24 Measuring frequency
U1190A Series User’s Guide 51
2 Making Measurements
Measuring Frequency
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
52 U1190A Series User’s Guide
U1190A Series Handheld Clamp Meter User’s Guide

3 Clamp Meter Features

Detecting AC Voltage Presence (Vsense) 54 Making Relative Measurements (Null) 57 Capturing Maximum and Minimum Values (Max.Min) 58 Freezing the Display (Hold) 60
The chapter describes the additional features available in your clamp meter.
Agilent Technologies
53
3Clamp Meter Features
WARNING
CAUTION
NOTE

Detecting AC Voltage Presence (Vsense)

Detecting AC Voltage Presence (Vsense)
• You are advised to test on a known live circuit within the rated AC
voltage range of this product before and after each use to ensure that the Vsense detector works.
• Voltage could still be present even if there is no Vsense alert
indication. Do not rely on the Vsense detector with shielded wire. Never touch live voltage or conductor without the necessary insulation protection or power off the voltage source.
• The Vsense detector may be affected by differences in socket
design, insulation thickness, and insulation type.
You are advised to measure voltage by using test leads through the AC V or DC V function after using the Vsense function, even if there is no alert indication.
This measurement function is applicable for U1192A, U1193A, and U1194A models only.
54 U1190A Series User’s Guide
The Vsense detector is a non- contact voltage detector that detects the presence of AC voltages nearby.
Set up your clamp meter to enable the Vsense function as shown in Figure 3-3.
If the presence of AC voltage is sensed, the clamp meter’s beeper will sound. The audible alert allows you to easily sense nearby AC voltage presence.
No resolution and accuracy of voltage measurement will be displayed in this mode.
Clamp Meter Features 3
NOTE
Detecting AC Voltage Presence (Vsense)
Place the top of the clamp meter close to a conductor when sensing for
AC voltages (as low as 24 V in the Hi.SE setting).
The low sensitivity setting can be used on flush mounted wall sockets
or outlets and various power strips or cords.
The high sensitivity setting allows for AC voltage sensing on other
styles of recessed power connectors or sockets where the actual AC voltage is recessed within the connector itself.
Press to toggle the Vsense detector’s sensitivity between
Hi.SE (high sensitivity) or Lo.SE (low sensitivity).
Figure 3-1 Vsense (high sensitivity) display
Figure 3-2 Vsense (low sensitivity) display
U1190A Series User’s Guide 55
3Clamp Meter Features
3
4
Voltage source
AC
2
1
Detecting AC Voltage Presence (Vsense)
Figure 3-3 Detecting voltage presence
56 U1190A Series User’s Guide

Making Relative Measurements (Null)

NOTE
When making Null measurements, also called relative, each reading is the difference between a stored (measured) null value and the input signal.
One possible application is to increase the accuracy of a resistance measurement by nulling the test lead resistance. Nulling the leads is also particularly important prior to making capacitance measurements.
1 To activate the relative mode, press the key. The
measurement value at the time, when Null is enabled, is stored as the reference value.
Clamp Meter Features 3
Making Relative Measurements (Null)
Figure 3-4 Null display
2 To disable the Null function, press again.
For any measurement function, you can directly measure and store the null value by pressing with the test leads open (nulls the test lead capacitance), shorted (nulls the test lead resistance), or across a desired null value circuit.
In resistance measurement, the clamp meter will read a non-zero value
even when the two test leads are in direct contact because of the resistance of these leads. Use the Null function to zero-adjust the display.
For DC voltage measurements, the thermal effect will influence the
accuracy of the measurements. Short the test leads and press when the displayed value is stable to zero-adjust the display.
U1190A Series User’s Guide 57
3Clamp Meter Features

Capturing Maximum and Minimum Values (Max.Min)

Capturing Maximum and Minimum Values (Max.Min)
The Max.Min operation stores the maximum, minimum, and average input values during a series of measurements.
When the input goes below the recorded minimum value or above the recorded maximum value, the clamp meter beeps and records the new value. The clamp meter also calculates an average of all readings taken since the Max.Min mode was activated.
From the clamp meter’s display, you can view the following statistical data for any set of readings:
Max: highest reading since the Max.Min function was
enabled
Min: lowest reading since the Max.Min function was
enabled
Avg: average or mean of all readings since the Max.Min
function was enabled
MaxMinAvg: present reading (actual input signal value)
1 Press and hold for more than 1 second to enable the
Max.Min operation.
2 Press again to cycle through the Max, Min, Avg, or
present (MaxMinAvg) input values.
3 Press for more than 1 second to disable the Max.Min
function.
Figure 3-5 Max.Min display
58 U1190A Series User’s Guide
Clamp Meter Features 3
NOTE
Capturing Maximum and Minimum Values (Max.Min)
If an overload is recorded, the averaging function will be stopped. is shown in place of the average value.
This mode is useful for capturing intermittent readings, recording minimum and maximum readings unattended, or recording readings while equipment operation keeps you from observing the clamp meter display.
The true average value displayed is the arithmetic mean of all readings taken since the start of recording. The average reading is useful for smoothing out unstable inputs, calculating power consumption, or estimating the percentage of time a circuit is active.
U1190A Series User’s Guide 59
3Clamp Meter Features

Freezing the Display (Hold)

Freezing the Display (Hold)
To freeze the display for any function, press the key.
Figure 3-6 Hold display
Press again to disable this function.
60 U1190A Series User’s Guide
U1190A Series Handheld Clamp Meter User’s Guide

4 Characteristics and Specifications

Product Characteristics 62 Specification Assumptions 64 Measurement Category 65
Measurement category definition 65
Electrical Specifications 66
DC specifications 66 AC specifications 68 Capacitance specifications 69 Temperature specifications 70 Frequency specifications 71 Frequency sensitivity specifications 71 Display update rate (approximate) 72
This chapter lists the characteristics, assumptions, and specifications of the U1191A, U1192A, U1193A, and U1194A handheld clamp meters.
Agilent Technologies
61
4 Characteristics and Specifications
NOTE

Product Characteristics

Product Characteristics
Product characteristics specified in the table below are applicable for U1191A, U1192A, U1193A, and U1194A models unless stated otherwise.
POWER SUPPLY
Battery type:
2 × 1.5 V AAA Alkaline battery (IEC LR03) Battery life:
200 hours typical (based on new Alkaline batteries, for continuous DC voltage measurement, with backlight disabled)
40 hours typical (based on new Alkaline batteries, with backlight enabled)
Low battery indicator will flash when the battery voltage drops below 2.5 V
(approximately)
POWER CONSUMPTION
9 mVA maximum (based on new Alkaline batteries, for continuous DC voltage
measurement, with backlight disabled)
42 mVA maximum (based on new Alkaline batteries, for continuous DC voltage measurement, with backlight enabled)
DISPLAY
Liquid crystal display (LCD) with backlight (maximum reading of 6000 counts)
OPERATING ENVIRONMENT
Operating temperature from –10 °C to 50 °C, 0% to 80% RH
Up to 80% RH for temperatures up to 30 °C, decreasing linearly to 50% RH at
50 °C
Altitude up to 2000 meters
Pollution degree 2
STORAGE COMPLIANCE
–40 °C to 60 °C, 40% to 80% RH (without batteries)
SAFETY COMPLIANCE
IEC 61010-1:2001/EN 61010-1:2001
IEC 61010-2-032:2002/EN 61010-2-032:2002
CAN/CSA-C22.2 No. 61010-1-04
CAN/CSA-C22.2 No. 61010-2-032-04
ANSI/UL Std. No. 61010-1:2004
62 U1190A Series User’s Guide
Characteristics and Specifications 4
Product Characteristics
MEASUREMENT CATEGORY
CAT III 600 V and CAT IV 300 V (for digital multimeter and current clamp portions)
ELECTROMAGNETIC COMPATIBILITY (EMC)
IEC61326-1:2005/EN61326-1:2006
Canada: ICES/NMB-001: Issue 4, June 2006
Australia/New Zealand: AS/NZS CISPR 11:2004
TEMPERATURE COEFFICIENT
0.1 × (specified accuracy) / °C (from 0 °C to 18 °C, or 28 °C to 50 °C)
COMMON MODE REJECTION RATIO (CMRR)
>60 dB at DC, 60 Hz in the AC V function
>120 dB at DC, 50/60 Hz in the DC V function
NORMAL MODE REJECTION RATION (NMRR)
>60 dB at 50/60 Hz
DIMENSIONS (W × H × D)
U1191A/U1192A: 77.14 × 225 × 38.56 mm
U1193A/U1194A: 77.14 × 238 × 38.56 mm
WEIGHT
U1191A/U1192A: 320 grams (with batteries)
U1193A: 334 grams (with batteries)
U1194A: 348 grams (with batteries)
MAXIMUM CONDUCTOR SIZE
U1191A/U1192A: Up to 27 mm diameter for 500 MCM cable
U1193A/U1194A: Up to 35 mm diameter for 750 MCM cable
MAXIMUM JAW OPENING
U1191A/U1192A: Up to 31 mm
U1193A/U1194A: Up to 37 mm
U1190A Series User’s Guide 63
4 Characteristics and Specifications

Specification Assumptions

WARRANTY
Please refer to http://www.agilent.com/go/warranty_terms
Three years for the product
Three months for the product’s standard accessories, unless otherwise
specified
Please take note that for the product, the warranty does not cover:
Damage from contamination
Normal wear and tear of mechanical components
Manuals and batteries
CALIBRATION CYCLE
One year
Specification Assumptions
Accuracy is given as ± (% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH.
AC V and AC A specifications for U1193A and U1194A models are AC coupled, true RMS, and are valid from 5% of range to 100% of range.
The crest factor may be up to 3.0 at full- scale (4000 counts)
For non- sinusoidal waveforms, add (2% reading + 2% full scale) typically.
In the EMC RF field of 3 V/m, the total accuracy is specified as the specified accuracy ± 30 digits for all functions.
64 U1190A Series User’s Guide

Measurement Category

The Agilent U1190A Series Handheld Clamp Meters have a safety rating of CAT III, 600 V and CAT IV, 300 V.

Measurement category definition

Measurement CAT I are for measurements performed on
circuits not directly connected to the AC mains. Examples are measurements on circuits not derived from the AC mains and specially protected (internal) mains- derived circuits.
Measurement CAT II are measurements performed on
circuits directly connected to a low- voltage installation. Examples are measurements on household appliances, portable tools, and similar equipment.
Measurement CAT III are measurements performed in
building installations. Examples are measurements on distribution boards, circuit- breakers, wiring, including cables, bus- bars, junction boxes, switches, socket outlets in the fixed installation, and equipment for industrial use, and some other equipment including stationary motors with permanent connection to the fixed installation.
Characteristics and Specifications 4
Measurement Category
Measurement CAT IV are measurements performed at the
source of low-voltage installations. Examples are electricity meters and measurements on primary over current protection devices and ripple control units.
U1190A Series User’s Guide 65
4 Characteristics and Specifications
NOTE

Electrical Specifications

Electrical Specifications
Specification assumptions are given on page 64.

DC specifications

Tab l e 4 - 1 DC specifications
Function Range Resolution
U1191A U1192A U1193A U1194A (where applicable)
Voltage
Notes for DC voltage specifications:
1 Overload protection: 600 Vdc. 2 Input impedance: 10 M (nominal) in parallel with <100 pF
Resistance
Notes for resistance specifications:
1 Overload protection: 600 Vrms for short circuits with <0.1 mA current 2 Maximum open voltage is <1.4 V 3 The accuracy is specified after the Null function is used to subtract the test lead resistance and thermal effect (by shorting
the test leads).
60.00 V 0.01 V - 0.5% + 3 0.5% + 3 0.5% + 3 - -
600.0 V 0.1 V 0.5% + 3 0.5% + 3 0.5% + 3 0.5% + 3 - -
600.0 0.1 0.8% + 5 0.8% + 5 0.8% + 5 0.8% + 5 95 A1.4 V
6.000 k 0.001 k 0.8% + 3 0.8% + 3 0.8% + 3 0.8% + 3 95 A1.4 V
60.00 k 0.01 k - 0.8% + 3 0.8% + 3 0.8% + 3 95 A0.7 V
Accuracy Test current
Open
voltage
66 U1190A Series User’s Guide
Tab l e 4 - 1 DC specifications (continued)
Characteristics and Specifications 4
Electrical Specifications
Function Range Resolution
Accuracy Test current
U1191A U1192A U1193A U1194A (where applicable)
Continuity 600.0 0.1 0.8% + 5 0.8% + 5 0.8% + 5 0.8% + 5 95 A 1.4 V
Notes for continuity specifications:
1 Overload protection: 600 Vrms for short circuits with <0.1 A current 2 Built-in buzzer beeps continuously when the resistance measured is less than 30 . Resistance measurements above
200  are considered open. For resistance measured between 30  and 200 (30   reading 200 ), the built-in buzzer may beep depending on the device-under-test.
3 Continuity indicator: 2.7 kHz tone buzzer
Diode 1.500 V 0.001 V 1.0% + 3 1.0% + 3 1.0% + 3 1.0% + 3 0.3 mA 1.8 V
Notes for diode specifications:
1 Overload protection: 600 Vrms for short circuits with <0.4 mA current 2 Built-in buzzer beeps continuously when the voltage measured is less than 100 mV and beeps once for forward-biased
diode or semiconductor junctions measured between 0.3 V and 0.8 V (0.3 V reading 0.8 V).
60.00 µA 0.01 µA - - - 1.0% + 5 - -
600.0 µA 0.1 µA - - - 1.0% + 5 - -
Current
60.00 A 0.01 A - - - 2.0% + 5 - -
600.0 A 0.1 A - - - 2.0% + 5 - -
Open
voltage
Notes for DC current specifications:
1 60 A to 600 A ranges are for current clamp measurements. 2 60 µA to 600 µA ranges are for digital multimeter measurements. 3 Overload protection for 60 A to 600 A range: 600 Arms 4 Input impedance for 60 µA to 600 µA range: 4.2 k 5 Position error: 1% from reading 6 The accuracy is specified after the Null function is used to subtract the test lead resistance and thermal effect (by shorting
the test leads).
U1190A Series User’s Guide 67
4 Characteristics and Specifications
Electrical Specifications

AC specifications

AC voltage specifications
Tab l e 4 - 2 AC voltage specifications
Accuracy
Range Resolution
60.00 V 0.01 V - 1.2% + 5 1.2% + 5 1.2% + 5 10 M
600.0 V 0.1 V 1.2% + 5 1.2% + 5 1.2% + 5 1.2% + 5 10 M
Notes for AC voltage specifications:
1 Overload protection: 600 Vrms 2 Input impedance: 10 M (nominal) in parallel with <100 pF 3 Frequency response: 45 Hz to 500 Hz (sinusoidal waveform) 4 AC conversion type:
U1191A and U1192A: Average sensing, RMS indication
U1193A and U1194A: RMS sensing, RMS indication
U1191A U1192A U1193A U1194A
45 Hz to 500 Hz
Input
impedance
AC current specifications
Tab l e 4 - 3 AC current specifications
Accuracy
Range Resolution
60.00 µA0.01 µA------1.0% + 51.0% + 5
600.0 µA0.1 µA ------1.0% + 51.0% + 5
60.00 A 0.01 A - - 2.0% + 5 3.0% + 5 2.0% + 5 3.0% + 5 2.0% + 5 3.0% + 5
400.0 A0.1 A2.0% + 53.0% + 52.0% + 53.0% + 5----
68 U1190A Series User’s Guide
U1191A U1192A U1193A U1194A
45 Hz to
65 Hz
65 Hz to
500 Hz
45 Hz to
65 Hz
65 Hz to
500 Hz
45 Hz to
65 Hz
65 Hz to
500 Hz
45 Hz to
65 Hz
65 Hz to
500 Hz
Tab l e 4 - 3 AC current specifications (continued)
Characteristics and Specifications 4
Electrical Specifications
Accuracy
Range Resolution
600.0 A0.1 A----2.0% + 53.0% + 52.0% + 53.0% + 5
Notes for AC current specifications:
1 Overload protection:
U1191A and U1192A: 400 Arms
U1193A and U1194A: 600 Arms
2 Input impedance for 60 µA to 600 µA range: 4.2 k 3 Frequency response: 45 Hz to 500 Hz (sinusoidal waveform) 4 Position error: 1% from reading 5 AC conversion type:
U1191A and U1192A: Average sensing, RMS indication
U1193A and U1194A: RMS sensing, RMS indication
6 For non-sinusoidal waveform, add an additional accuracy of (2% of reading + 2% of full scale) typically for
crest factor 3.0.
U1191A U1192A U1193A U1194A
45 Hz to
65 Hz
65 Hz to
500 Hz
45 Hz to
65 Hz
65 Hz to
500 Hz
45 Hz to
65 Hz
65 Hz to
500 Hz
45 Hz to
65 Hz
65 Hz to
500 Hz

Capacitance specifications

Tab l e 4 - 4 Capacitance specifications
Range Resolution
Accuracy
U1191A U1192A U1193A U1194A
600.0 F0.1 F - 2.0% + 4 2.0% + 4 2.0% + 4 2 times/second
6.00 mF 0.001 mF - 2.0% + 4 2.0% + 4 2.0% + 4 1 time/9 seconds
Notes for capacitance specifications:
1 This function is only applicable for U1192A, U1193A, and U1194A models. 2 Overload protection: 600 Vrms for short circuits with <0.1 mA current 3 The accuracy of for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract
the test lead resistance and thermal effect (by opening the test leads).
Measuring rate
(at full scale)
U1190A Series User’s Guide 69
4 Characteristics and Specifications
Electrical Specifications

Temperature specifications

Tab l e 4 - 5 Temperature specifications
Thermal type Range Resolution
–40 °C to 400 °C 0.1 °C 1.0% + 2.0 °C
K
400 °C to 1200 °C 1.0 °C 1.0% + 2.0 °C
–40 °F to 752 °F 0.1 °F 1.0% + 3.6 °F
K
752 °F to 2192 °F 1.0 °F 1.0% + 3.6 °F
Notes for temperature specifications:
1 This function is only applicable for the U1194A model. 2 The specification above is specified after the multimeter has been left stationary in the same operating environment for
1 hour at least. If the unit is exposed during storage in a high humidity (condensing) environment, ensure that the multimeter has been in the same operating environment for 2 hours at least.
3 The accuracy does not include the tolerance of the thermocouple probe. 4 Do not allow the temperature sensor to contact a surface that is energized above 30 Vrms or 60 V DC. Such voltages pose
a shock hazard.
5 The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST175. 6 The accuracy specification assumes the surrounding temperature is stable with ±1 °C. For the surrounding temperature
changes of ±3 °C, the rated accuracy applies after two hours.
Accuracy
U1194A
70 U1190A Series User’s Guide

Frequency specifications

Tab l e 4 - 6 Frequency specifications
Characteristics and Specifications 4
Electrical Specifications
Range Resolution
Accuracy
U1191A U1192A U1193A U1194A
99.99 Hz 0.01 Hz 0.5% + 3 0.5% + 3 0.5% + 3 0.5% + 3
999.9 Hz 0.1 Hz 0.5% + 3 0.5% + 3 0.5% + 3 0.5% + 3
9.999kHz 0.001 kHz 0.5% + 3 0.5% + 3 0.5% + 3 0.5% + 3
99.99 kHz 0.01 kHz 0.5% + 3 0.5% + 3 0.5% + 3 0.5% + 3
Notes for frequency specifications:
1 This function is only applicable for U1192A, U1193A, and U1194A models. 2 Overload protection: 600 V
Minimum input

Frequency sensitivity specifications

For voltage measurements
Tab l e 4 - 7 Frequency sensitivity and trigger-level specifications for voltage measurements
Input range Minimum sensitivity (RMS sine wave)
Maximum input for specified accuracy 10 Hz to 10 kHz 10 kHz to 100 kHz
frequency
10 Hz
60 V6.0 V30 V
600 V 60 V 60 V
Notes for frequency sensitivity specifications for voltage measurements:
1 This function is only applicable for U1192A, U1193A, and U1194A models. 2 Maximum input for specified accuracy, refer to “AC specifications” on page 68.
U1190A Series User’s Guide 71
4 Characteristics and Specifications
Electrical Specifications
For current measurements
Tab l e 4 - 8 Frequency sensitivity and trigger level specifications for current measurements
Input range Minimum sensitivity (RMS sine wave)
Maximum input for specified accuracy 45 Hz to 1 kHz
60 A 6.0A
600 A 60 A
Notes for frequency sensitivity specifications for voltage measurements:
1 This function is only applicable for U1192A, U1193A, and U1194A models. 2 Maximum input for specified accuracy, refer to “AC specifications” on page 68.

Display update rate (approximate)

Tab l e 4 - 9 Display update rate (approximate)
(6 mF)
Times/second
2 times/1 second
1 time/9 seconds
(600 F)
(6 mF)
2 times/1 second
(600 F)
1 time/9 seconds
(6 mF)
Function
AC V 3 3 3 3
DC V 3 3 3 3
2222
Diode 3 3 3 3
Capacitance -
Te m p e r a t u r e - - - 2
DC A (A or A) - - - 3
AC A (A or A)3333
Frequency - 3 (>10 Hz) 3 (>10 Hz) 3 (>10 Hz)
U1191A U1192A U1193A U1194A
2 times/1 second
(600 F)
1 time/9 seconds
72 U1190A Series User’s Guide
www.agilent.com
Contact us
To obtain service, warranty, or technical assistance, contact us at the following phone or fax numbers:
United States: (tel) 800 829 4444 (fax) 800 829 4433 Canada: (tel) 877 894 4414 (fax) 800 746 4866 China: (tel) 800 810 0189 (fax) 800 820 2816 Europe: (tel) 31 20 547 2111 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Korea: (tel) (080) 769 0800 (fax) (080) 769 0900 Latin America: (tel) (305) 269 7500 Ta i w a n : (tel) 0800 047 866 (fax) 0800 286 331 Other Asia Pacific Countries: (tel) (65) 6375 8100 (fax) (65) 6755 0042
Or visit Agilent World Wide Web at:
www.agilent.com/find/assist
Product specifications and descriptions in this document are subject to change without notice. Always refer to the Agilent website for the latest revision.
© Agilent Technologies, Inc., 2011
First Edition, September 2011 U1191-90003
Agilent Technologies
Loading...