Agilent Protein A, rProtein A, Protein G User Manual

Data Sheet
Agilent Bio-Monolith Protein A, rProtein A, and Protein G Columns
Characteristics and shipping conditions
The information in this data sheet is provided to ensure proper product care and maximum product lifetime.
Agilent Bio-Monolith Protein A (p/n 5069-3639)
Catalog Number
Immobilized Ligand Agilent Bio-Monolith Protein A Agilent Bio-Monolith rProtein A Agilent Bio-Monolith Protein G
Support Matrix
Column Dimensions
Dynamic Binding Capacity Agilent Bio-Monolith Protein A Agilent Bio-Monolith rProtein A Agilent Bio-Monolith Protein G
Maximum Loading Capacity
Working Flow Rates
Maximum Allowed Pressure Over the Column
Temperature Stability
Recommended pH*
Shipping Solvent
* At high pH, ligands can slowly degrade, therefore, only use pH higher than 11 for regeneration, cleaning-in-place,
and sanitization procedures.
Immunoaffinity Protein A from Staphylococcus aureusis Immunoaffinity r-Protein A from Escherichia coli Immunoaffinity r-Protein G from Escherichia coli
Poly(glycidyl methacrylate-co-ethylene dimethacrylate) highly porous monolith
Diameter 5.2 mm, length 4.95 mm, bed volume (CV) 0.10 mL
>8 mg hIgG/mL wet support >5 mg hIgG/mL wet support >9 mg hIgG/mL wet support
(Conditions: hIgG 0.5 mg/mL, PBS buffer, pH 7.4, flow rate 1 mL/min)
0.4 to 0.5 mg
Recommended: 0.2 to 2 mL/min (1 to 10 cm/min, 2 to 20 CV/min) Maximum: 3 mL/min (15 cm/min, 30 CV/min)
150 bar (15 MPa, 2,100 psi) WARNING: Do not exceed the maximum allowed pressure, as this might seriously damage your column.
Working 4 to 40 °C Storage 4 to 8 °C
WARNING: Avoid prolonged use at elevatedtemperatures.
Working range: 2 to 11 Cleaning in place: 2 to 13
20 mM Tris, pH 7.4 and 20 % ethanol
Agilent Bio-Monolith Protein A, rProteinA,
Flat end
direction
Pointed end
and Protein G columns are analytical affinity columns designed for fast separation, isolation, and quantitation of antibodies in fermentation (cell culture supernatants and lysates). The columns are compatible with HPLC and preparative LC systems. The monolithic structure enables high reproduciblity and long lifetimes for separation and quantitation of immunoglobulin G (IgG) at high speeds.
Preparing the column before first use
Connecting the column
The flow direction is indicated in
Figure1. The flat end is the inlet, and the
pointed end is the outlet.
Flow
Figure 1. Flow direction through the flat end of the column.
1. Remove both end caps.
2. Connect the flat end* to the nut of the capillary, which is already connected to the HPLC system.
3. Tighten the fitting as instructed by
themanufacturer.
4. Flow liquid into the column.
5. Connect outlet (pointed end) to
thedetector.
* Long to extra long fittings and ferrules
(p/n 5065-9967) should be used.
Column equilibration
1. The columns should be washed with at least 2 mL (20 CV) of a 0.1M
buffer such as sodium phosphate
buffer or Tris buffer containing 1 M
NaCl, pH 7.0 to 8.0 at a flow rate of
0.5 to 1.0 mL/min.
2. Wash the columns with at least 2 mL (20 CV) of eluting buffer (see eluting buffers below).
3. Equilibrate the column with 30 CV of the binding buffer (see binding and washing buffers below) at a flow rate
of 0.5 to 1.0 mL/min.
4. The column is ready for injection.
Binding and washing buffers
Bio-Monolith Protein A, rProtein A, and Protein G columns are compatible with many buffers for binding and washing. The most common ones are:
1. Sodium phosphate buffer and Trisbuffer, 25 to 100 mM. We
recommend using 50 mM as the starting concentration.
2. Binding and washing buffers with pH 6 to 9. We recommend starting with pH 7.4. This pH seems to be the optimal for the columns to absorb/bind most antibodies.
3. Room temperature or 24 to 25°C
is recommended for optimal performance and extended column lifetime. Higher temperature can be also used, but might reduce column lifetime.
4. The columns are compatible with neutral salts such as sodium chloride (NaCl) or potassium chloride (KCl).
Salt concentrations of 0.1 to 0.2
M can improve the peak shape and recovery in some cases for
somesamples.
Note: Filtration of buffers through a 0.22 or 0.45-µm membrane is recommended to reduce buffer impurities that build up on the frits inside the column. This will prevent column blockage.
Note: Binding/washing buffers should be freshly made.
Eluting buffers
Bio-Monolith Protein A, rProtein A, and Protein G are compatible with many low pH buffers. The buffers commonly used are citric, glycine, HCl, and acetate
acid. Table 1 provides details for these
commonly used buffers.
Table 1. Some common eluting buffers for use with AgilentBio-Monolith Protein A, rProtein A, and
Protein G columns.
Column Buffer Conc. pH
Agilent Bio-Monolith Protein A and rProtein A
Agilent Bio-Monolith Protein G
Citric acid 0.1 M 2.5 to 3.0
Glycine 0.1 M 2.5 to 3.0
Acetic acid 5 to 20 %
Citric acid 0.1 M 2.0 to 2.5
Glycine 0.1 M 2.0 to 2.5
Acetic acid 5 to 20 %
Note: Commonly, elution buffers for affinity columns have a refractive index
(RI) very different from binding/washing
buffers; therefore, baseline noise and an artifact peak could appear when the eluents start flowing. This peak could interfere with the quantitation of low concentration samples. To minimize this effect, high-quality chemicals are recommended to be used and blank runs should be included to establish the artifact peak. Blank runs can be used for baseline subtraction if desired.
Note: HCl has a lower RI compared to
other eluents. If a low concentration sample is used and baseline noise and artifact peaks are of concern, HCl can be used as an eluent.
2
Loading...
+ 1 hidden pages