Agilent Quick Start Guide Datasheet

Page 1
Quick Start Guide
Agilent Technologies AC Source/ Power Analyzer Graphical User Interface
for Windows 95-98 and Windows NT 4.0
Page 2
Description...........................................................................................................3
System Requirements ...........................................................................................3
Installing and Running the Software.....................................................................4
Configuring the I/O ..............................................................................................4
Performing Basic Operations ...............................................................................5
Printing and Saving Measurement Data ...............................................................7
Sample Tests ........................................................................................................ 8
Inrush Current Measurement............................................................................8
AC Line Dropout..............................................................................................9
AC Line Surge or Sag.....................................................................................10
Telephone Ring Generator.............................................................................11
Using the Transient List Table ...........................................................................12
Creating User-Defined output Waveforms.........................................................12
Generating AM or FM Output Signals...............................................................13
Advanced Controls ............................................................................................. 14
Warranty.............................................................................................................16
2
Page 3

Description

What the GUI Will Do
The Agilent AC Source Graphical User Interface (GUI) is an easy to use soft front panel for the Agilent Technologies 6800-series AC Power Source/Analyzers. With it, you can: ñ Control output and measurement functions from a single
screen.
ñ View an oscilloscope-like display of actual output
waveforms.
ñ View the harmonic content of the output waveform. ñ Easily generate complex output transients and user-defined
waveforms by clicking and dragging the mouse .
ñ View the instrument commands generated by the graphical
user interface.
ñ Operate in either simulation mode (no ac source hardware
required), or in instrument control mode.
What the GUI Will Not Do
The Graphical User Interface does not:
ñ Automatically generate test programs ñ Control instruments other than Agilent a c sources ñ Provide drivers for other GPIB instruments ñ Link with the Agilent HFTS software ñ Operate with Windows 3.1

System Requirements

PC
32-bit Platforms
Supported I/O
Supported Ac Source Models
ñ 486DX66 minimum (Pentium 100 or better recommended) ñ 16 Mbytes RAM ñ 4 Mbytes of disk space
ñ Windows 95 - 98 ñ Windows NT 4.0
ñ Agilent 82340B, 82341C, 82341D, 82335B
1
ñ National AT GPIB/TNT, AT GPIB/TNT PnP (Windows 95) ñ RS-232
1
The Agilent 82335B is only supported for Windows 95, provided you have
either the E2094F or E2094E I/O library.
ñ Agilent 6811A/B, 6812A/B, 6813A/B, 6814B, 6834B ñ Agilent 6841A, 6842A, 6843A (in Normal mode only)
3
Page 4
Installing and Running the Software
g
(
y
Automatically searches
NOTE: Before running the Agilent AC Source GUI, you must have installed
and connected your ac source to the pc using the appropriate interface cable. If you are using a National Instruments GPIB interface, you must have the appropriate card installed and configured on your pc. If you are using an Agilent GPIB interface card, you must also have the appropriate 32-bit SICL library drivers installed on your pc.
1. Place Disk #1 in the A: drive of your computer and run SETUP.EXE.
2. Follow the directions on the screen to install the software. The README.TXT file contains product updates or corrections that are not documented in the help file. Use any text editor to open and read this file.
3. To run the Agilent AC Source GUI, click on its desktop icon: You can also click on the Start button and select: Programs | Agilent AC Source | AC Source GUI.

Configuring the I/O

To configure the I/O, select the I/O Configuration command in the Edit menu. This step is necessary to establish communication with the ac source. The I/O configuration screen comes up automatically when you turn Simulation mode off in the Source menu. Note that operating the ac source over a serial port is much slower than GPIB operation.
Select I/O type.
(GPIB, serial port, or
simulation mode)
Address must match
Agilent GPIB cards SICL name used in
setting of
instrument.
only. Must match
interface card’s
confi
GPIB
uration.
If there seems to be no communication between the Agilent AC Source GUI and the ac source, select Auto Detect to detect any ac sources connected to your pc. Also try resetting the I/O interface by clicking the red reset button. This also resets the ac source to its default settings.
RS-232 only. Specifies the com port on the pc.
RS-232 only. Ac source must match Baud rate and Parit
National cards only. Specifies board index. (1 = GPIB1)
all interfaces for ac sources.
settings.
NOTE: If you program the ac source from the front panel of the unit, you will
need to refresh or update the Agilent AC Source GUI. Go to the Source menu and select the Refresh Instrument State command.
4
Page 5

Performing Basic Operations

The Main window appears when you first run the Agilent AC Source GUI. From this window you can control the output of the ac source as well as view all output measurements.
Click to start
a
Waveform
display area.
Measurement
summary
Click to configure measurements.
Turn on the ac source output.
Click Enter to apply all values to the output.
Set the output to 120Vac. Use
p/q
keyboard entry.
Select an output
To access an existing configuration,
ñ Click File, Open Configuration in the menu. Configur ation files contain
Agilent AC Source GUI settings but not actual ac source instrument settings.
ñ Select the desired file (files have an *.set extension) and click Open. ñ To create a new configuration, click File, New Configuration. This action
clears the present configuration, resets the ac source instrument to its default values, and sets the I/O to simulation mode.
keys or
To program the output of the ac source,
ñ Under Output Control, program the output rms voltage, dc offset voltage,
frequency, rms current limit, and peak current limit. You can either type in the values directly, or use the up/down arrows to adjust the present value. If you type in the values, you mst press the Enter button.
ñ For 3-phase ac sources, select the Phase that you want to control. You can
couple or control All phases simultaneously, or control each phase individually.
ñ Select a waveshape from the Waveshape list box - either Sinewave,
Squarewave, or Clipp ed sinewave. For clipped sinewave s, you can specify the clipping level in percent of amplitude, or in percent of total harmonic distortion.
5
Page 6
ñ Select Output On to enable the output of the unit. Select Output Off to
harmonic graph axes
turn the output off. Note that the selected waveshape will not appear on the Main window unless you click the Measure button.
To view output measurements,
ñ Select the Config Measurements command in the Measure menu. Select
the Configure Waveform Display folder.
Specify type of measurement.
Only applies to units with the AUX channel option.
Select units for
ñ To view voltage or current measurements, select Voltage, Current, or
Voltage + Current in the Waveform to Display dropdown list.
ñ To view harmonic measurements, select a harmonic display functions. ñ Click the Measure button to make an output measurement. The following
figure shows a sample voltage harmonics measurement. Note that for 3­phase ac sources, you can only measure one phase at a time. Only the measurement of the selected phase is returned.
Click
Displays the
specified
harmonics.
To save the configuration settings,
ñ Click File, Save Config in the menu. ñ Enter a filename (setup files have an .set extension) and click Save.
Note that Output Control settings are NOT saved in the setup file.
6
Page 7
Printing and Saving Measurement Data
g
waveform data logging.
g
To print the measurement screen, click on either the Print icon or select the
Print command in the File menu. You can preview and print the Main window only. The following is a sample preview/print of the Main window:
Click
To save the waveform that appears in the display, select Save Waveform in the
File menu. This saves the displayed waveform as an ASCII file, which can be easily imported into other application programs.
To log measurement or waveform data to an ASCII text file, select the Config Measurements command in the Measure menu. Select the Configure
Measurements folder. Check the appropriate Log to file box and provide a name for the data file. Data will be logged whenever an output measurement is made.
Check to enable measurement or
Specify a measurement data filename (file extension = *.ml
Specify a waveform data filename (file extension = *.wl
The measurement data file saves measurements in a comma-separated table of
values. The columns in the list correspond to all available measurements, with blanks for non-measured values. For 3-phase units, only information from the selected phase will be saved.
The waveform data file saves measurements in a 3-column format. The columns are Time, Voltage and Current. Harmonics are not saved.
7
).
).
Page 8

Sample Tests

adjust a
ude.)
gg
Several sample tests are provided with the Agilent AC Source Graphical User Interface. These pre-defined tests help you quickly perform some common ac source applications.

Inrush Current Measurement

To perform an inrush current measurement, select Inrush Current in the Tests menu. The test uses the parameters that you enter in the following dialog box. This box is displayed on the screen when the test is first selected.
Click here to continue. This displays the Output Transient
Enter a Final Line Voltage
Enter a turn-on
phase angle.
Editor. (see below)
Output goes to 120 Vac @ 90
phase angle. Start of output measurement
Output voltage
starts at 0 Vac.
Click to start sample test.
(Drag point to
Check to enable measurement tri
er.
You can use the transient editor to make any final adjustments to the inrush current measurment such as fine tuning the width of the transient step, or positioning and enabling the measurement trigger.
When measurements are enabled, you can view the inrush measurement in the Main window. If necessary, move the Transient editor out of the way to see the Main window. The following figure shows a sample inrush curent measurement.
°
mplit
8
Page 9
gg
gg
View the
inrush
current.

AC Line Dropout

To perform an ac line dropout, select Dropout in the Tests menu. The test uses the parameters that you enter in the following dialog box. This box is displayed on the screen when the test is first selected.
Click here to continue. This displays the
Enter the number of
dropout cycles
Enter the dropout
voltage value
Enter a starting
phase angle.
Output Transient Editor.
Check to enable measurement tri
er.
Start of output measurement
Step #1,
output voltage
@ 120 Vac.
Step #2, output
voltage @ 0 Vac
for one cycle.
Click to start sample test.
Step #3, output voltage returns to 120 Vac. (Drag point to adjust amplitude .)
Drag pointer to adjust width of dropout.
Check to enable measurement tri
er.
9
Page 10
You can use the transient editor to make any final adjustments to the ac line
gg
dropout test such as fine tuning the width of the dropout, or positioning and enabling the measurement trigger.
When measurements are enabled, you can view the line dropout measurement in the Main window. If necessary, move the Transient editor out of the way to see the Main window.

AC Line Surge or Sag

To perform an ac line surge/sag, select Surge/Sag in the Tests menu. The test uses the parameters that you enter in the following dialog box. This box is displayed on the screen when the test is first selected.
Enter a Surge/Sa g
Voltage value
Enter T1 value.
Also enter T2 and
T3 values in the
appropriate fields.
Start of output measurement
Output voltage set to 120 Vac
at step #1.
Click here to continue. This displays the Output Transient Editor. (see below)
Output voltage returns to 120 Vac at step #3.
Output voltage
surges to 170
Vac at step #2.
Click to start sample test.
Check to enable measurement tri
er.
You can use the transient editor to make any final adjustments to the ac line surge/sag test such as fine tuning the slew rates of the surge or sag, or positioning and enabling the measurement trigger.
When measurement are enabled, you can view the line surge/sag measurement in the Main window. If necessary, move the Transient editor out of the way to see the Main window.
10
Page 11

Telephone Ring Generator

)
CAUTION: This test is only designed to test telephone ringer circuits. Do not
use this test to power any device that requires ac input power.
The telephone ring generator produces an output signal to test telephone ringers. The test uses the parameters that appear in the dialog box that is displayed on the screen when the test is first selected.
Enter the ring
amplitude in rms
Enter the dc level
of the waveform
Enter a ringer on-
time value.
Enter a ringer off-
time value.
Click here to continue. This displays the Output Transient Editor. (see below)
Enter a ringer frequency.
DC offset set to 80 V (Drag point to adjust amplitude.)
Ringer off-
time level.
Repeat the
number of
rings.
Start of step #2. (Drag pointer to adjust width of step.
Click to start
sample test.
You can use the transient editor to make any final adjustments to the telephone ringer waveform such as fine tuning the dc level of the waveform. Note that the transient editor only displays the rms-portion of the telephone ringer; it does not display the actual waveform.
When measurements are enabled, you can view the ringer waveform in the Main window. If necessary, move the Transient editor out of the way to see the Main window.
11
Page 12

Editing With the Transient List Table

y
The previous four sample tests introduced you to using the click and drag editing features of the Transient Editor.
A more precise method of creating and editing transients is available by using the Transient List table. Click the View Table button in the Output Transient editor. The Transient List table is simply another view of the transient editor. The values in the table correspond exactly to the corresponding transient graph. The following table shows the values for the sample Surge/Sag test.
Click
Menu commands
to simplify editing.
Select and edit
individual values
in each cell.

Creating User-Defined Output Waveforms

To create user-defined voltage waveforms, select either the waveform icon or the Arbitrary Waveforms command in the Source menu. This window lets you edit basic waveform shapes to create a user-defined arbitrary waveform.
Waveforms can be edited in either waveform or harmonic format. To create or edit a waveform, the Display Mode must match the Edit Mode. Download the completed waveform to the ac source. To output the downloaded waveform,
you must select it in the Waveshape box of the Main window.
Click
Drag
waveform
points to edit
the shape.
Edit Mode must
match Displa
Mode to edit the
waveform.
Download the waveform to the ac source.
Select New or Open to edit a
waveform
Save the waveform as an ASCII file.
Select to zoom the display in or out.
Indicates the of the active point.
12
Page 13
To create or edit waveforms in the harmonic format, set the Edit Mode and the Display Mode to Harmonics. Note that continually switching edit modes will result in cumulative waveform truncation errors. Therefore, select only one edit mode to work in. Switching display modes does not cause recalculation errors.
Drag harmonic
to change the
amplitude.
Edit Mode must
match Display
Mode to edit the
waveform.
Lets you edit individual waveform data points.

Generating AM or FM Output Signals

AM or FM ouput signals can be precisely generated using the AM amd FM Modulation table. Either click the Modulation icon, or select the Modulat ion command in the Source menu.
AM or FM output waveforms consist of a carrier signal (the fundamental) and a modulation signal. The voltage and frequency of the carrier signal is defined in the Output Control section of the Main window. Only the modulation signal is defined in this dialog box.
Click
Enter
parameters for
modulation
signal.
Lets you edit # of steps in modulation profile.
Modulation signals are comprised of from one to 20 individually programmed steps that may be combined into a modulation sequence and then run. Each step programs the modulation signal for the duration of the step. The amplitude and frequency of the modulation signal may be constant, swept linearly, or swept logarithmically from the starting value to a stop value. The sweep period is programmed in seconds.
13
Page 14

Advanced Controls

Applies to waveforms and
To access the less frequently used output controls, select the Output Control command in the Source menu. These output controls include slew rate settings, relative-phase on 3-phase units, output coupling, voltage range, and save/recall states.
Specify output
voltage, frequency, or
dc offset slew rates.
Select AC+DC to
dc-couple the
output.
Save or Recall instrument states in non-volatile memory
Resets the ac source to its power­on state.
To program Output impedance, select the Impedance command in the Source menu.You can program real or reactive impedance. These controls are not available on all ac source models.
Enable ouput
impedance control
Enter impedance
values
To configure output measurements, select the Config Measurements command in the Measure menu. Then select the Configure Measurements folder. You can select what measurements appear in the measurement summary area and how often the summary measurements are updated. You can also configure measurement and waveform logging, as previously discussed.
Check which
summary
measurements to
display on the Main
window.
Specifies the time between measurements.
summary measurements.
14
Page 15
To program minimum and maximum user-defined limits for critical ac source
g
parameters, select the User Defined Limits command in the Source menu. The controls in this dialog box can be used to prevent a user from accidentally programming dangero us voltage or frequency values in the Main window.
Enter min imum and
maximum user-
defined limit values.
Resets the limits to
the hardware default
values.
To display the maximum programmable values for the ac source that is presently being controlled by the Agilent AC Source GUI, select the Ratings command in the Source menu.
Displays maximum
programmable
settin
s of ac source.
To view the SCPI instrument commands sent to the ac source, select Instrument Commands in the View menu.
Select command
viewing options.
View the
programming
commands.
15
Page 16

Warranty

This Agilent Technologies software product is warranted against defects in materials and workmanship for a period of 90 days from date of delivery. During the warranty period, Agilent Technologies will, at its option either repair or replace parts which prove to be defective.
Agilent Technologies makes no express or implied warranty of any kind, including, but not limited to the implied warranties of merchantability or fitness for a particular purpose, with regard to the program material contained herein. Agilent Technologies shall not be liable for incidental or consequential damages in connection with or arising out of the furnishing, performance, or use of this software.
Use of the supplied manual and software is restricted to Agilent Technologies ac source products only. The software is copyrighted and may not be copied except for archival purposes, to rep l ace a defective copy, or for program error verification. Agilen t Technologies warrants that this software designed for use with a personal computer, will execute its programming instructions when properly installed on that personal computer. Agilent Technologies does not warrant that the operation of the personal computer , software, or ac source will be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from: misuse, unauthorized modification, operation or storage outside the environmental specifications for the product, in-transit damage, improper maintenance, or defects resulting from use of non­Agilent Technologies software, accessories, media, or such items not designed for use with the product.
Notice
This document contains proprietary information protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated into another language without the prior consent of Agilent Technologies. The information contained in this document is subject to change without notice.
Copyright 1997, 1999, 2000 Agilent Technologies, Inc. Windows is a registered trademark of Microsoft Corporation.
Manual part number: 5962-8191 Printed in USA, February 1999
Updated April 2000
Loading...