The information in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including, but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Agilent Technologies shall not be liable for errors contained herein or
for incidental or consequential damages in connection with the furnishing, performance,
or use of this material.
This document contains or refers to proprietary information which is protected by
copyright. All rights are reserved. No part of this document may be photocopied,
reproduced, or translated to another language without the prior written consent of Agilent
Technologies Company.
Responsibility of the Manufacturer
Agilent Technologies only considers itself responsible for any effects on safety, reliability
and performance of the equipment if all the following are true:
• Assembly operations, extensions, re-adjustments, modifications or repairs are done by
persons authorized by Agilent Technologies.
• The electrical installation of the relevant room complies with the IEC or national requ irements.
• The instrument is used according to the instructions for use presented in this manual.
As with all electronic equipment, radio frequency interference between this
cardiograph and any existing RF transmitting or receiving equipment at the
installation site, including electrosurgical equipment, should be evaluated
carefully and any limitations noted before the equipment is placed in service.
Monitoring during electrosurgery should not be attempted and monitoring
electrodes should be removed from the patient to preclude the possibility of
burns. Radio frequency generation from electrosurgical equipment and close
proximity transmitters may seriously degrade cardiograph performance. Agilent
Technologies assumes no liability for failures resulting from RF interference
between Agilent medical electronics and any radio frequency generating
equipment at levels exceeding those established by applicable standards.
ii
Notice
CAUTION
CAUTION
Like all electronic devices, this cardiograph is susceptible to electrostatic discharge
(ESD). Electrostatic discharge typically occurs when electrostatic energy is transferred to
the patient, the electrodes, or the cardiograph. ESD may result in ECG artifact that may
appear as narrow spikes on the cardiograph display or on the printed report. When ESD
occurs, the cardiograph’s ECG interpretation may be inconsistent with the physician’s
interpretation.
ESD discharges to expose d m etal on th e rear of the cardiograph can o ccasion ally cau se an
error message to appear on the cardiograph display. The cardiograph returns to normal
operation after turning the power off, then on again.
The data transmission cable must have a suppression device attached to assure the
cardiograph’s compliance with the European Radiated Emissions Standard found in
CISPR 11. If your data transmission cable does not include a suppression device,
compliance can be achieved by attaching one of the following suppression devices to the
cable, near the cardiograph:
part number 0443164251Fair-Rite Products Corporation
P. O. Box J
One Commercial Row
Wallkill, New York 12589
telephone: (914) 895-2055
FAX: (914)895-2629
or
Euro-Schaffner, S. A.
1 B Avenue de Suisse - BP 16
68311 Illzach Cedex, France
telephone: 33-8-931-0400
FAX: 33-8-931-0401
part number 28B2025-0A0 Steward
East 36th Street
P. O. Box 510
Chattanooga, TN 37401
telephone: (615) 867-4100
FAX: (615) 867-4102
or
Steward EMC, S.A.
Rue Fritz-Couvoisier 40
Ch-2300 La Chaux-de-Fonds
Switzerland
telephone: 41-39-282-387
Fax: 41-39-280-277
This is to certify that this equipment is in accordance with the Radio Interference
Requirements of the EMC Directive.
iii
Notice
Medical Device Directive
The PageWriter M1770 series of cardiographs comply with the requirements of the Medical Device Directive 93/42/EEC and carries the
Agilent Technologies warrants this medical product against defects in materials and
workmanship for a period of three years in certain geographics, or one year with onsite
support.
If Agilent Technologies receives notice of such defects during the warranty period,
Agilent T echn olog ies shall, at its option, either repair or replace hardware pr odu cts which
prove to be defective.
Agilent Technologies software and firmware products that are designated by Agilent
Technologies for use with a hardware product, when properly installed on that hardware
product, are warranted not to fail to execute their programming instructions due to defects
during the warranty period. Agilent Technologies shall repair or replace software media
and firmware that do not execute their programming instructions due to such defects.
Agilent Technologies does not warrant that the operation of the software, firmware, or
hardware shall be uninterrupted or error free.
If Agilent Technologies is unable, within a reasonable time, to repair or replace any
product to a condition as warranted, Buyer shall be entitled to a refund of the purchase
upon return of the product to Agilent Technologies.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulti ng from any of the following:
1. Improper or inadequate maintenance by Buyer.
2. Buyer-supplied software or interfacing.
3. Unauthorized modification or misuse.
4. Operation outside of the environmental specifications for the product.
5. Improper site preparation and maintenance.
iv
Notice
THE WARRANTY SET FORTH ABOVE IS EXCLUSIVE AND NO OTHER
WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED.
AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE.
Printing History
October 1994Edition 1
May 1996Edition 2
October 1998Edition 3
December 1999Edition 4
v
Safety Summary
0Safety Summary
Safety Symbols Marked on the Cardiograph
The following symbols are used on the cardiograph or the cart:
Caution - See operating instructions
Alternating current.
Meets IEC type CF leakage current requirements and is defibrillator
protected. (Isolated ECG input.)
Equipotential (identifies independent protective earth conductor to the
cardiograph).
Fuse.
Indicates power control for cardiograph.
HzIndicates operating frequency in cycles per second.
The maximum weight that the cart can hold
Displays the configuration menu on the PageWriter
200/200i/300pi.
Please see Chapter 2, “Performance Verificati on and Maintenance,” for safety
requirements that apply to the cardiograph.
vi
Conventions Used in This Manual
WARNING
CAUTION
NOTE
0Conventions Used in This Manual
Warning statements describe conditions or actions that can result in personal
injury or loss of life.
Caution statements describe conditions or actions th at can result in damage to the
equipment or software.
Notes contain additional information on cardiog raph usage.
TEXT
.H\
Softkey
represents the labels that appear on the display.
represents keys on the key panel.
represents the temporary key labels that appear on the display.
vii
Preface
0Preface
This manual contains service information for the Agilent M1770A pageWriter 300pi,
PageWriter 200i, M1771A PageWriter 200, and M1772A PageWriter 100 cardiographs.
The information and procedures in this manual apply to all models unless otherwise
specified.
This manual is organized as follows:
• Chapter 1 — Introduction. Contains a gener al description of the cardiogr aphs, lists of
technical specifications, and lists of accessories and options.
• Chapter 2 — Performance V erification and Maintenance. Explains how to check the
cardiograph’ s perf ormance using buil t-in self-tes ts, and lists main tenance procedures and
safety requirements that apply to the cardiograph.
• Chapter 3 — Theory of Operation. Provides an overview of how the cardiograph
works and describes the operation of the m a jor subassemblies.
• Chapter 4 — Troubleshooting. Contains procedures and error codes to aid the service
person in localizing faults to a replaceable subassembly.
• Chapter 5 — Removal and Replacement. Contains procedures for removing and
replacing each of the cardiograph’s major subassemblies.
• Chapter 6 — Parts List. Lists part numbers for the cardiograph’s replaceable p arts, and
provides assembly drawings.
• Appendix A — Connector Pin Assign ment s. Identifies and defines the signals
This chapter introduces you to the M1770A PageWriter 200i/300pi, M1771A PageWriter
200, and M1772A PageWriter 100 cardiographs and lists their technical specifications.
These cardiographs are only to be serviced by Agilent qualified personnel.
Safe and effective use of medical instrumentation requires periodic inspection and
preventive maintenance. Perform the preventive maintenance procedures in Chapter 2 of
this manual at required intervals to ensure satisfactory instrument performance.
The cardiographs use a thermal printhead to record waveforms and label the ECG report.
The paper supplied with the cardiographs is a thermal paper designed to work with this
printhead and with the photo detector used to advancing the paper.
Agilent Technologies guarantees the performance of the cardiographs only when used
with Agilent Technologies supplies, accessories, and paper that meet or exceed Agilent
Technologies specifications.
0PageWriter 100, 200, 200i, and 300pi Series Cardiographs
The M1770A PageWriter 200i/300pi is Agilent Technologies’ economical, interpretive
cardiograph. The M1771A PageWriter 200 is the economical, non-interpretive
cardiograph. The M1772A PageWriter 100 is the most economical, non-interpretive
cardiograph. The cardiograph contains the user controls, the printer, and all the processing
circuitry. All models use the same enclosure. The M1705B cart is designed for these
cardiographs. Figure 1-1 shows the M1770A mounted on the cart. Option #A05 provides
transmission and storage capability for the M1770A and M1771A. The StressWriter
option provides capability for communication between a PageWriter 100/200/200i/300pi
cardiograph and a StressW riter system.
1-1
PageWriter 100, 200, 200i, and 300pi Series Cardiographs
The PageWriter 100, 200, and 300 family of cardiographs print ECGs in Auto and
Manual formats. All PageWriter cardiographs use a continuous feed, high-resolution,
thermal array printer. This produces electrocardiograms on standard- size paper that can
be included in patient records without cutting or mountin g. ECG reports clearly show the
ECG waveforms. The PageWriter 200 and 300 series cardiographs also annotate records
with patient information and ECG measurements. The PageWriter 200i and 300pi also
includes ECG interpretation for both adult and pediatric patients.
From a service standpoint, the PageWriter 100, 200, and 300 family of cardiographs are
similar except for two major hardware differences:
• The M1770A has additional firmware for ECG measurements and interpretation.
• The M1771A has additional firmware for ECG measurements.
• The M1772A keypanel do es no t hav e an L CD preview dis play and uses operation al k eys
instead of alphanumeric keys for user input.
All other model differences are found in the cardiograph’s software.
The PageWriter 100, 200, and 300 family of cardiographs is designed for low-cost, longterm reliability. The modular design makes extensive use of VLSI and gate array
technology, resulting in a minimum number of subassemblies. The modular approach
means less down-time for the user, since replacing subassemblies allows faster field
repairs. The internal self-test efficiently identifies faulty subassemblies, further speedin g
the repair process.
1-2
Inquiries
M1705B Cart
The optional M1705B Cart provides mobility for the PageWriter 100, 200, and 300 family
of cardiographs. The large wheels make the cart easy to move, yet steady. Slots in the
cardiograph’s feet and a thumbscrew secure the cardiograph to the cart rails. The tray
directly below the cardiograph provides storage for the user’ s guide. Storage for the power
cable is built in. Other built-in cable retainers hold the patient data cable ou t of the way.
Two compartmented trays provide storage for spare patient electrodes and patient cables,
additional thermal paper, and consumable supplies.
Transmission and Storage, Option #A05
Transmission and storage features are available for the M1770A and M1771A with
purchase of Option #A05. Up to th irty Auto ECGs can be stored in the car diograph’s
internal flash memory. Stored ECGs can be recalled later for editing, re-analyzing,
printing, or transmi ssi on to another PageWri te r 200 /2 00i/300pi equipped with Opt i on
#A05, a PageWriter XLi cardiograph, a TraceMaster ECG Management System, or
facsimile machine. Manual ECGs cannot be transmitted or st ored.
1Inquiries
Refer any questions or comments regarding these instruments to the nearest Agilent
Technologies Sales/Service Office or to one of Agilent Technologies’ Service Dispatch
Centers. Always identify the instrument by model number and serial number in all
correspondence. Telephone numbers for Service Dispatch Centers and Sales/Service
Offices are listed in Chapter 6, “Parts List.”
1Specification Data
The following tables list the technical specifications for the cardiographs and the mobile
cart. Specifications are the same for all models except as noted
Table 1-1Physical Specifications
Parameter Specification
Dimensions (h × w × l)
× 15 in × 17 in (11.5 cm × 39.2 cm × 43.7 cm)
cardiograph
cart
Weight
cardiograph
cart
4 in
× 17 in × 33 in (91.4 cm × 43.5 cm × 84.8 cm)
36 in
18.7 lbs. (8.5 kg) or less (includes batt ery, 200 sheets of paper, patient cable, and po w er co r d)
38 lbs. (17.24 kg)
Chemical resistance,
cleaners
Withstands the following: isopropyl alcohol (except patient cable), mild soap and water, chlorine
bleach and water (30 ml/l of water).
1-3
Specification Data
Table 1-2Electrical Specifications
Parameter Specification
Resolution
ECG (internal)
display
5 µV
128 row pixels by 24 0 column pixels
display sweep sp eed
Input impedancegreater than 2.5 MΩ@ 10 Hz, typical ly gr eat er than 100MΩ @ DC
Gain accuracy
Input biasless than 40 nA for in put leadwire, less th an 500 nA for Right Leg output lead-
Common mode rejec t io n110 dB or greater with AAMI test ci r cui t
Defibrillator recovery
Crosstalk rejection
Sample rate4.096M samples per second per input leadwire (unfiltered)
DC offset tole rance
Noise
Standardizing vo lt age
50/60 Hz notch filter50/60 Hz AC line rejection filters are always active.
Pacemaker pulse display> 0.2 mV indica tion for pace puls es be tween 0.5 an d 2 ms dur ati on at amplit ude
23.1 mm/sec
±5% of input signal or ±40 µV, whichever is greater
wire
System recovery 8 seconds after 360 j o ule discharge, no dam age
Less than 2% channel crosstalk
≥ ±300 mVdc with less than 5% gain change
≤ 30 µV peak-to-peak RTI
1 mV
2 to 250 mV
± 1%
± 5% for 100 mSec ± 5 mSec
Sampling Characteristics of Cardiograph
Physiological factors such as breathing can cause variation in amplitudes of heart beats,
independent of the heart rate. In addition, sampled systems (as op posed to continuous
systems) may show changes in the apparent height of the R-wave when sampling of the Rwave occurs slightly off-peak. Because the sampling rate and the heart rate are
asynchronous, the time between the peak and an adjacent sample can vary from one QRS
complex to the next. This results in a slight variation of displayed QRS amplitu de. This
effect is more pronounced wit h narrower signals, more commonly foun d in some pediatric
electrocardiograms and in pacemaker pulses. The M177XA family of cardiographs
minimizes this effect by:
• using an integrating-type A/D converter
• sampling all leadwires simultaneously
• oversampling the signals
• processing the oversampled signals with appropriate digital signal processing techniques
1-4
Specification Data
Table 1-3ECG to Paper Specifications
ParameterSpecification
Frequency respons e of PageWriter
200/200i/300pi
(-3dB bandwidth)
Auto ECG
Manual ECG
Frequency response of PageWriter 100
Auto ECG
Manual ECG
ECG resolution on paper
voltage axis
time axis
Recorder speed5, 10, 25, and 50 mm/sec + 1.5%
ECG visibility with pace pulseMeets AMMI EC11-1991 standard for Diagnostic Electrocardiographic
<3 dB down at (user selectable):
<3 dB down at (user selectable):
Approximately 8 dots/mm (200 dots/in)
Approximately 20 dots/mm (500 dots/in) at 25mm/sec and slower
Approximately 10 dots/mm (250 dots/in) at 50mm/sec
Devices
High PassLow Pass
0.05 Hz
0.15 Hz
0.5 Hz
0.05 Hz
0.15 Hz
0.5 Hz
0.15-150 Hz or 0.15-40 Hz
0.05-150 Hz or 0.05-40 Hz
150 Hz
100 Hz
40 Hz
150 Hz
100 Hz
40 Hz
1-5
Specification Data
Table 1-4Power Supply and Battery Specifications
Parameter Specification
AC line frequenc y
115 V setting
230 V setting
AC input power rating
Battery voltage6 V DC, 6.5 Ah
Battery recharge time with unit off
(battery fully discharged)
to 90% capacity
to full capacity
Battery capacity
Auto
continuous rhythm (Manual mode)
Battery life
AC and battery operation
battery only operation
Low battery warningAt least 2 Auto ECGs or at least 2 minutes of Man-
Table 1-5Safety Specifications
Parameter Specification
50 and 60 Hz (nomina l)
90–132 Vac
198–264 Vac
50 VA
7 hours typical
16 hours typical
40 ECGs
40 minutes
Typically 24 months
Typically 14–16 months
ual ECG is allowed after the Low Battery indication is given.
ECG leads source current to groundLess than 10 µA RMS
ECG leads sink current
120 Vac, 60 Hz
240 Vac, 50 Hz
Ground wire integrity
Shunting of defibrillator energy
Operator safety during defibrillation
Less than 20 µA RMS with patient cable
Less than 50 µA RMS with patient cable
Less than 100 mΩ @ 25 A AC
< 10%
< 100 µC
1-6
Specification Data
Table 1-6Environmental Specifications
ParameterSpecification
Temperature
operating
storage
Humidity
operating
storage
Pressure (altitude)
operating
storage
Table 1-7Miscellaneous Specifications
ParameterSpecification
Real time clock accuracy
10 to 40
° C
50
° to 104° F (10° to 40° C)
° to 122° F (0° to 50° C)
32
15 to 80% RH, non-c ondensing
15 to 90% RH, non-c ondensing
15,000 ft. (4600 m) for 2 hours
15,000 ft. (4600 m)
Less than 3 minutes deviation per month.
Printhead lifeTypically 100,000 pages
Table 1-8Storage and Transmission Specifications (Option #A05 only)
ParameterSpecification
Auto ECG storage ca pacity30 ECGs
Transmission Protocols
DT
SCP
Modem command interfaces
Data modem
FAX modem
Modem protocols
Modulation
Error corre c tion
Compression
FAX modulation
Data transmission standard use d betwe en Agile nt Pag eWriter
cardiographs and A gi le nt TraceMa st er ECG Management Systems
Standard Communications Protocol
Standard AT command set
EIA/TIA-578 Service Class 1
V.34
V.42
V.42 bis
V.17
1-7
Options and Accessories
1Options and Accessories
PageWriter 100, 200, 200i, and 300pi Cardiographs
These tables list the options and accessories available for the PageWriter 100, 200, and
300 family of cardiographs.
Country/Region Options
Each country/region option includes the appropriate power cord, voltage, printer, patient
cable, and language. See T ab le 1-9, which shows the configuration of each country/region
option.
A05Adds storage and transmission capability
M2488A-#A70, A71, A72StressWriter system includes PageWriter cardiograph
M2488A-#C70StressWriter system includes interface card for PageW riter
Standard Accessories
Accessories included are based on model number and localization.
• Using the PageWriter 200/200i Cardiograph Operator Training Video (PageWriter
200/200i, also for use with PageWriter 300pi)
• Using the PageWriter 100 Cardiograph Operator Training Video (PageWriter 100 only)
1-12
Performance Verification and Maintenance
NOTE
1Introduction
This chapter describes how to verify the cardiograph’s performance, explains preventive
maintenance, presents patient safety information, and prov ides a p erfo rmance verification
checklist.
1Performance Verification
Several procedures make up performance verification: visual inspection of the
cardiograph exterior, execution of Extended Self-test, visual examination of an ECG
recorded from an ECG patient simulator, and system safety tests using a safety analyzer.
A Performance Verification matrix which appears at the end of this section specifies the
tests and inspections which must be performed following servicing of th e PageWriter
cardiographs. The Performance Verification test results must be recorded on
Customer Service order records
Make copies of the matrix. Fill out a copy each time the cardiograph is tested. Attach the
printer test output, and simulator ECG trace to the completed matrix and file with the
cardiograph’s permanent maintenance record.
If any of the performance verification tests fail, refer to Chapter 4, Troubleshooting.
2-1
Performance Verification
Visual Inspection and Power On Self Test
Before beginning the inspection, p res s the key to put the cardiograph in
2Q6WDQGE\
Standby mode, and unplug the power cord from the wall outlet. Inspect the cardiograph
for the following:
• Worn or damaged power cord
• Loose or missing hardware
• Mechanical damage
• Evidence of liquid spill
• W orn printer drive gear
• Worn printer roller
• Corroded or damaged reusable electrodes, if present
• Damaged patient cable
• Dirt/paper residue on the thermal printhead
• Frayed or damaged wiring
Replace any damaged or missing items, and clean the printhead and patient electrodes as
necessary. Cleaning instructions are listed under “Preventive Maintenance” later in this
chapter. Connect the cardiograph to AC power. Be sure that the AC indicator comes on.
Turn the cardiograph On and observe that the display turns on (PageWriter
200/200i/300pi) or that the LED’s flash on (PageWriter 100).
Extended Self-test
The cardiograph’s extended self-test checks each major subassembly. The extended selftest consists of two sections: the testing of internal circuitry and the testing of printer,
display, keyboard, and modem capabilities.
To begin the extended self-test, press and hold both the and key s while
turning on the cardiograph with thekey.
2Q6WDQGE\
T o st op or exit extended self-test, yo u must press to p lace the cardiogr aph in
$XWR0DQXDO
2Q6WDQGE\
Standby.
Internal Circuitry Testing of PageWriter 100
These tests are performed on the internal circuitry:
•Memory
• Gate Array
• Front-End Interface
• Serial Interface
Each of these tests is described in detail later in this chap ter.
When the internal circuitry test of the PageWriter 100 is performed, the LEDs show which
test is being performed. The following table shows which LED is associated with each
test.
2-2
Performance Verification
Table 2-1PageWriter 100 Test LED’s
Test Number Test Name LED(s) lit on Page-
Writer 100
1ROM-15 mm/mVOK
2ROM-210 mm/mVNormally fails for 100.
3ROM-X120 mm/mVOK for A#10.
4ROM-X2V LeadsOK for A#10.
5RAMFilterOK
6NVRAMAutoOK
7GAI II III Lead GroupWill also flash all LEDs. OK
8ECG FEaVR aVL aVF Lead GroupOK
9SIOII aVF V2 Lead GroupOK for A#10.
AREMOTEV LeadsFor future use - normally fails.
BMODEM5 mm/secOK for A#10, with modem or test tool
CFLASH110 mm/secOK for A#10.
DFLASH225 mm/secOK for A#10.
When the tests are completed, the results are also printed out in a pass/fail format
on the left-hand side of the first page of the test report.
Comments/ Ex pected Result s
attached
Table 2-2Test Results Format
Test Passes Test Fails
Printed ResultOK* *
In some cases, a test failure halts the test/test printout and an error code ind icatin g the
failure is flashed on the LED’s. See Chapter 4, “Troubleshooting,” for a list of self-test
failure codes and failure symptoms. The section “How the PageWriter 100 Communicates
Error Codes and Messages” in Chapter 4 describes how to read the LED’s.
2-3
Performance Verification
Internal Circuitry Testing of PageWriter 200/200i/300pi
These tests are performed on the internal circuitry:
•Memory
• Gate Array
• Front-End Interface
• Serial Interface
Each test is described detail later in this chapter.
When testing the internal circuitry of the PageWriter 200/200i/300pi, the numbers 1
through 9 and the characters A through D are displayed one at a time beginning with
number 1. Each number represents a particular test. If the test fails, the character X is
placed after the test number, f or example, 3X. The test descriptions below are listed in the
order the tests are performed.
Table 2-3PageWriter 200/200i/300pi Test Numbers
Test Number Test Name Character Displayed on Page-
Writer 200/200i/300pi
1ROM-11OK
2ROM-22Passes for 300p i, 200i, and 200 with
3ROM-X13Passes for Options #A05 and #A10
4ROM-X24Passes for Options #A05 and #A10
5RAM5OK
6NVRAM6OK
7GA7OK
8ECG FE8OK
9SIO9Passes for Options #A05 and #A10
AREMOTEAFor future use - normally fails
BMODEMBPasses for Options #A05 and #A10
Comments
software revision A.05.06 or later.
only.
only.
only.
only, with modem or test tool
attached
CFLASH1CPasses for Options #A05 and #A 10
only
DFLASH2DFor future use - normally fails
2-4
Performance Verification
CAUTION
If your cardiograph is equi p ped wi th Opt i on s # A05 or #A10, do not turn it off during Test
C or D. If you turn off the cardiograph while Test C or D are in process, you could lose
stored ECGs, the Log of ECGs Stored, or the Log of ECGs Taken.
When the tests are completed, the results are printed out in a pass /fai l format on the lefthand side of the first page of the test report.
Table 2-4Test Results Format
Test Passes Test Fails
Printed ResultOK* *
In some cases, a test failure halts the test/test printout and an error code ind icatin g the
failure appears on the display. See Chapter 4, “Troubleshooting,” for a list of self-test
failure codes and failure symptoms.
Memory Test
This test looks for failures in the following memory subsys tems:
• Read-Only Memory (ROM)
• Random Access Memory (RAM)
• Non-Volatile Random Access Memory (NVRAM)
Read-Only Memory (tests 1-4) . The cardiograph is designed to have up to four ROMs
installed, two on the main control board and two on an option board. The Read-Only
Memory test reads the contents and performs a 32-bit cyclic redundancy check (CRC) of
each of the four ROM address spaces. The test of a ROM address space will fail if the
ROM module is faulty or is not present. The four ROM address spaces are identified on
the report as
•ROM-1
•ROM-2
•ROM-X1
•ROM-X2
Depending on the serial number and model of the unit you are servicing, the
corresponding ROM part numbers vary. Refer to tables 2-5, 2-6, 2-7, and 2-8 for the
correct ROM replacement part numbers.
2-5
Performance Verification
Table 2-5ROM Replacement Part Numbers for PageWriter Cardiographs M1770, M1771A Serial Number
Prefix CNA, CNB, or CNC, XXXA or , US001
Test Part NumberLanguagePart Number
Labeled on
ROM
ROM 1
ROM 2U206 - Main CPU BoardM1770-89521
ROM X1U9 - Interface BoardM2488-17901
ROM X2U10 - Interface BoardM2488-17902