Agilent Technologies N5161A/62A/81A/82A MXG Signal Generators
A3 RF Assembly Upgrade for Expanded License Key Upgradeability
Option 099 Upgrade Kit
Kit Part Numbers:
N5161AK-099
N5162AK-099
N5181AK-099
N5182AK-099
NOTEFor the latest revision of this installation note, go to the following
website:
http://www.agilent.com/find/mxg
Part Number N5180-90042
Printed in USA July 2008
Notice
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the
implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable
for errors contained herein or for incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
This upgrade kit adds standard specified performance down to 100 kHz, higher output power, and provides the capability to
1
install the following license key enabled hardware options: Option UNZ, UNT, UNU, UNV
NOTEThis upgrade requires LAN access and a PC with a CD drive.
, UNW, 1EQ, and 1EA.
Installation includes completing the following major steps, in the order given.
1. Run Self Test.
2. Install new firmware (>
3. Replace the A3 RF assembly.
4. Verify signal generator operation.
5. Activate Option 099.
6. Install new nameplate overlay. (For instruments with serial prefixes <US/MY/SG4818.)
A.01.43).
1. Not available on N5161A and N5181A instruments.
3
Installation Kit Parts List
ItemQuantityDescriptionPart Number
1
21
31
41
51
61
71
81
91
91
Tools Required
•TORX T-10 driver
1
Installation NoteN5180-90042
1 GHz Nameplate (N5181A)N5180-80027
3 GHz Nameplate (N5181A)N5180-80028
6 GHz Nameplate (N5181A)N5180-80029
3 GHz Nameplate (N5182A)N5180-80025
6 GHz Nameplate (N5182A)N5180-80026
Media Binder5958-5162
Documentation CDN5180-90007
For Option UNM Instruments
A3 RF assemblyN5180-60169
For Option UNM Instruments
A3 RF assemblyN5180-60170
•TORX T-15 driver
•TORX T-20 driver
•9/16” nut driver
•5 mm nut driver
•Torque Wrench (9-21 in-lb.)
•Small Phillips screw driver
•Long-nose pliers
Typical Safety Considerations
WARNINGBefore you disassemble the signal generator, turn the power switch off and unplug the
power cord. Failure to unplug the signal generator can result in personal injury.
CAUTIONElectrostatic discharge (ESD) can damage or destroy electronic components. All work on
electronic assemblies should be performed at a static-safe workstation.
4
Run Self Test
1. Turn on the signal generator and let it warm up for at least five minutes.
NOTEIf the ERR annunciator is on take note of the error(s) numbers, but do not resolve the error(s) at this time.
1. Open the MXG service guide (N5180-90006) on the documentation CD (N5180-90007) provided with this kit.
2. Go to “Running Self Test” in chapter 1, “Troubleshooting”, and follow the procedure for your instrument model.
3. Self test failures between 100 and 699 do not need to be resolved at this time.
Self test failures between 700 and 999 must be resolved before installing the new A3 RF assembly.
Install New Firmware
NOTEFirmware version >A.01.43 must be downloaded before installing the new A3 RF assembly.
Use the firmware CD provided in the 5958-5162 binder or go to www.agilent.com/find/mxg to download the latest
firmware.
Replace the A3 RF Assembly
1. If not already open, open the MXG service guide (N5180-90006) on the documentation CD (N5180-90007) provided
with this kit.
2. Use chapters 3 and 4 of the MXG service guide to remove and replace the A3 RF assembly in your instrument.
Verify Signal Generator Operation
To ensure the signal generator is meeting all published specifications review chapter 4 of the MXG service guide, then run
the recommended manual performance verification tests in the order listed.
1. I/Q Calibration (N5162A/82A only)
2. Self Test
3. Relative Frequency Range and Accuracy Verification
4. Leveled Output Power Verification
Activate Option 099
1. Follow the instructions on the Entitlement Certificate.
5
Install New Nameplate (For Instruments with Serial Prefixes <US/MY/SG4818)
Place nameplate here
NOTEThe nameplate is provided for customer convenience only. The actual model and option capability is
determined electronically, and viewed on the front panel LCD Display.
1. Use a sharp knife, (i.e. Exacto knife), to lift one end of the current nameplate.
Figure 0-1
2. Pull the lifted end to remove the nameplate.
3. Remove the backing from the new nameplate.
4. Place the nameplate in the nameplate area on the Front Panel Assembly, see Figure 0-2.
Figure 0-2
5. Press firmly.
6