Agilent N1914A Programming Guide

Page 1
Agilent N1913/1914A EPM Series Power Meters
Programming Guide
Agilent Technologies
Page 2

Notices

CAUTION
WARNING
No p art o f this manual may b e rep roduced in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
N1913-90008
Edition
First Edition, May 13, 2009
Printed in Malaysia
Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051 USA
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Safety Notices
A CAUTION notice denotes a haz- ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
ii N1913A/1914A EPM Series Power Meters Programming Guide
Page 3

General Warranty

The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Duration and conditions of warranty for this product may be superseded when the product is integrated into (becomes a part of) other Agilent products. During the warranty period, Agilent will, at its option, either repair or replace products which prove to be defective. The warranty period begins on the date of delivery or on the date of installation if installed by Agilent.

Restricted Rights Legend

The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as “commercial computer software” as defined in DFARS 252.227-7013 (Oct 1988), DFARS
252.211-7015 (May 1991), or DFARS 252.227-7014 (Jun 1995), as a “commercial item” as defined in FAR 2.101(a), or as “restricted computer software” as defined in FAR 52.227-19 (Jun 1987) (or any equivalent agency regulation or contract clause), whichever is applicable. You have only those rights provided for such Software and Documentation by the applicable FAR or DFARS clause or the Agilent standard software agreement for the product involved.
N1913A/1914A EPM Series Power Meters Programming Guide iii
Page 4

Equipment Operation

WARNING
CAUTION
Warnings and Cautions
This guide uses warnings and cautions to denote hazards.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or loss of life. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Personal Safety Considerations
This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited. If this instrument is not used as specified, the protection provided by the equipment could be impaired. This instrument must be used in a normal condition (in which all means of protection are intact) only.
iv N1913A/1914A EPM Series Power Meters Programming Guide
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers. For continued protection against fire hazard, replace the line fuse(s) only with fuses of the same type and rating (for example, normal blow, time delay, etc.). The use of other fuses or material is prohibited.
Page 5

General Safety Considerations

WARNING
CAUTION
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the customer’s failure to comply with these requirements.
• Before this instrument is switched on, make sure it has been properly grounded through the protective conductor of the ac power cable to a socket outlet provided with protective earth contact. Any interruption of the protective (grounding) conductor, inside or outside the instrument, or disconnection of the protective earth terminal can result in personal injury.
• Any adjustments or service procedures that require operation of the instrument with protective covers removed should be performed only by trained service personnel.

User Environment

N1913A/1914A EPM Series Power Meters Programming Guide v
This instrument is designed for indoor use only.
Page 6

In This Guide ...

Power Meter Remote Operation
1
that configure the power meter and helps you determine settings to optimize performance.
MEASurement Commands
2
group of instructions to acquire data using a set of high level instructions.
CALCulate Subsystem
3
subsystem is used to perform post acquisition data processing.
CALibration Subsystem
4
command subsystem is used to zero and calibrate the power meter.
DISPlay Subsystem
5
used to control the selection and presentation of the windows used on the power meter’s display.
FORMat Subsystem
6
used to set a data format for transferring numeric information.
MEMory Subsystem
7
subsystem is used to create, edit, and review sensor calibration tables.
OUTPut Subsystem
8
subsystem is used to switch the POWER REF output on and off.
Chapter 3 explains how the CALCulate
Chapter 4 explains how the CALibration
Chapter 5 explains how the DISPlay subsystem is
Chapter 6 explains how the FORMat subsystem is
Chapter 7 explains how the MEMory command
Chapter 8 explains how the OUTPut command
Chapter 1 describes the parameters
Chapter 2 explains how to use the MEASure
SENSe Subsystem
9
subsystem directly affects device specific settings that are used to make measurements.
STATus Subsystem
10
subsystem enables you to examine the status of the power meter by monitoring the “Device Status Register”, “Operation Status Register” and the “Questionable Status Register”.
vi N1913A/1914A EPM Series Power Meters Programming Guide
Chapter 9 explains how the SENSe command
Chapter 10 explains how the STATus command
Page 7
SYSTem Subsystem
11
subsystem is used to return error numbers and messages from the power meter, preset the power meter, set the remote address, and query the SCPI version.
Chapter 11 explains how the SYSTem command
TRIGger Subsystem
12
subsystem is used to synchronize device actions with events.
UNIT Subsystem
13
subsystem is used to set the power meter measurement units to Watts and % (linear), or dBm and dB (logarithmic).
SERVice Subsystem
14
subsystem is used to obtain and set information useful for servicing the power meter.
IEEE 488.2 Command Reference
15
the IEEE488.2 Common Commands that the power meter supports.
Programming Examples
16
and examples to remotely control the EPM Series power meters using SCPI commands.
Calibration Factor Block Layout
A
the calibration factor block layout for E4410 Series, E9300 Series, and N8480 Series sensors (excluding Option CFT).
Measurement Polling Example
B
example of VEE program in measurement polling.
Chapter 12 explains how the TRIGger command
Chapter 13 explains how the UNIT command
Chapter 14 explains how the SERVice command
Chapter 15 contains information about
Chapter 16 contains programming sequences
Appendix A contains information on
Appendix B contains information on
HP436A/HP437B/HP438A Command Summary
C
information on the HP436A/HP437B/HP438A command summary.
N1913A/1914A EPM Series Power Meters Programming Guide vii
Appendix C contains
Page 8
viii N1913A/1914A EPM Series Power Meters Programming Guide
Page 9

Contents

Notices ii General Warranty iii Restricted Rights Legend iii Equipment Operation iv General Safety Considerations v User Environment v In This Guide ... vi
1 Power Meter Remote Operation
Introduction 4 Configuring the Remote Interface 5
• Interface Selection 5
• GPIB Address 5
• LAN Configuration 6
• USB Configuration 7
Zeroing and Calibrating 8
• Zeroing 8
• Calibration 9
Making Measurement 10
• Using MEASure? 11
• Using the CONFigure Command 16
• Using the Lower Level Commands 25
Using Frequency Dependent Offset Tables 27
• Overview 27
• Editing Frequency Dependent Offset Tables 29
• Selecting a Frequency Dependent Offset Table 32
• Enabling a Frequency Dependent Offset Table 32
• Making the Measurement 32
Setting the Range, Resolution and Averaging 34
• Resolution 34
• Averaging 34
• Auto Averaging Mode 35
• Filter Length 36
• Range 37
N1913/1914A EPM Series Power Meters Programming Guide ix
Page 10
Setting Offsets 38
• Channel Offsets 38
• Display Offsets 38
Setting Measurement Limits 40
• Setting Limits 40
• Checking for Limit Failures 42
• Using STATus 43
Getting the Best Speed Performance 44
• Measurement Rate 44
• Sensor 45
• Trigger Mode 45
• Output Format 47
• Units 47
• Command Used 47
• Fast Mode 48
How Measurements are Calculated 49 Status Reporting 50
• The General Status Register Model 50
• How to Use Register 53
• The Condition Polling Method 53
• The SRQ Method 54
• Device Status Register 65
• Using the Operation Complete Commands 67
Saving and Recalling Power Meter Configurations 70
• How to Save and Recall a Configuration 70
Using Device Clear to Halt Measurements 71 An Introduction to the SCPI Language 72
• Mnemonic Forms 72
• Using a Colon (:) 72
• Using a Semicolon (;) 73
• Using a Comma (,) 73
• Using Whitespace 73
• Using “?” Commands 73
• Using “*” Commands 74
• Syntax Conventions 74
x N1913/1914A EPM Series Power Meters Programming Guide
Page 11
• Syntax Diagram Conventions 74
• SCPI Data Types 75
• Input Message Terminators 80
SCPI Compliance Information 82 Summary of Commands 84
2 MEASurement Commands
MEASurement Commands 87 CONFigure[1] |2|3|4? 92 CONFigure [1] |2|3|4 Commands 97 CONFigure[1]|2|3|4[:SCALar][:POWer:AC] [<expected_value>[,<resolution>[,<source list>]]] 98 CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RELative [<expected_value>[,<resolution>[,<source
list>]]] 101 CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence [<expected_value>[,<resolution>[,<source
list>]]] 104 CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence: RELative
[<expected_value>[,<resolution>[,<source list>]]] 107 CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RATio [<expected_value>[,<resolution>[,<source
list>]]] 110 CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RATio: RELative[<expected_value>[,<resolution>[,<source
list>]]] 113 FETCh[1]|2|3|4 Queries 116 FETCh[1]|2|3|4[:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]] 117 FETCh[1]|2|3|4[:SCALar][:POWer:AC]:RELative? [<expected_value>[,<resolution>[,<source
list>]]] 120 FETCh[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence? [<expected_value>[,<resolution>[,<source
list>]]] 123 FETCh[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence: RELative?
[<expected_value>[,<resolution>[,<source list>]]] 126 FETCh[1]|2|3|4[:SCALar][:POWer:AC]:RATio? [<expected_value>[,<resolution>[,<source list>]]] 129 FETCh[1]|2|3|4[:SCALar][:POWer:AC]:RATio:RELative? [<expected_value>[,<resolution>[,<source
list>]]] 132 READ[1]|2|3|4 Commands 135 READ[1]|2|3|4[:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]] 136 READ[1]|2|3|4[:SCALar][:POWer:AC]:RELative? [<expected_value>[,<resolution>[,<source
list>]]] 139
N1913/1914A EPM Series Power Meters Programming Guide xi
Page 12
READ[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence? [<expected_value>[,<resolution>[,<source list>]]] 142
READ[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence: RELative? [<expected_value>[,<resolution>[,<source list>]]] 145
READ[1]|2|3|4[:SCALar][:POWer:AC]:RATio? [<expected_value>[,<resolution>[,<source list>]]] 148 READ[1]|2|3|4[:SCALar][:POWer:AC]:RATio:RELative? [<expected_value>[,<resolution>[,<source
list>]]] 151 MEASure[1]|2|3|4 Commands 154 MEASure[1]|2|3|4[:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]] 155 MEASure[1]|2|3|4[:SCALar][:POWer:AC]:RELative? [<expected_value>[,<resolution>[,<source
list>]]] 158 MEASure[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence? [<expected_value>[,<resolution>[,<source
list>]]] 161 MEASure[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence: RELative?
[<expected_value>[,<resolution>[,<source list>]]] 163 MEASure[1]|2|3|4[:SCALar][:POWer:AC]:RATio? [<expected_value>[,<resolution>[,<source
list>]]] 166 MEASure[1]|2|3|4[:SCALar][:POWer:AC]:RATio:RELative? [<expected_value>[,<resolution>[,<source
list>]]] 169
3 CALCulate Subsystem
CALCulate Subsystem 172 CALCulate[1]|2|3|4:HOLD:STAT <character_data> 175 CALCulate[1]|2|3|4:GAIN Commands 180 CALCulate[1]|2|3|4:GAIN[:MAGNitude] <numeric_value> 181 CALCulate[1]|2|3|4:GAIN:STATe <boolean> 183 CALCulate[1]|2|3|4:LIMit Commands 185 CALCulate[1]|2|3|4:LIMit:CLEar:AUTo <boolean>|ONCE 186 CALCulate[1]|2|3|4:LIMit:CLEar[:IMMediate] 188 CALCulate[1]|2|3|4:LIMit:FAIL? 189 CALCulate[1]|2|3|4:LIMit:FCOunt? 190 CALCulate[1]|2|3|4:LIMit:LOWer[:DATA] <numeric_value> 192 CALCulate[1]|2|3|4:LIMit:UPPer[:DATA] <numeric_value> 195 CALCulate[1]|2|3|4:LIMit:STATe <boolean> 198 CALCulate[1]|2|3|4:MATH Commands 200 CALCulate[1]|2|3|4:MATH[:EXPRession] <string> 201
xii N1913/1914A EPM Series Power Meters Programming Guide
Page 13
CALCulate[1]|2|3|4:MATH[:EXPRession]:CATalog? 205 CALCulate[1]|2|3|4:RELative Commands 207 CALCulate[1]|2|3|4:RELative[:MAGNitude]:AUTO <boolean>|ONCE 208 CALCulate[1]|2|3|4:RELative:STATe <boolean> 210
4 CALibration Subsystem
CALibration Subsystem 214 CALibration[1]|2|3|4[:ALL] 216 CALibration[1]|2|3|4[:ALL]? 218 CALibration[1]|2|3|4:AUTO <boolean>|ONCE 220 CALibration[1]|2|3|4:RCALibration <boolean> 222 CALibration[1]|2|3|4:RCFactor <numeric_value> 224 CALibration[1]|2|3|4:ZERO:AUTO <boolean>|ONCE 226 CALibration3|4:ZERO:TYPE EXTernal|INTernal 228
5 DISPlay Subsystem
DISPlay Subsystem 232 DISPlay:ENABle <boolean> 233 DISPlay:SCReen:FORMat <character_data> 234 DISPlay[:WINDow[1]|2] Commands 236 DISPlay[:WINDow[1]|2]:ANALog Commands 237 DISPlay[:WINDow[1]|2]:ANALog:LOWer <numeric_value> 238 DISPlay[:WINDow[1]|2]:ANALog:UPPer <numeric_value> 241 DISPlay[:WINDow[1]|2]:METer Commands 244 DISPlay[:WINDow[1]|2]:METer:LOWer <numeric_value> 245 DISPlay[:WINDow[1]|2]:METer:UPPer <numeric_value> 248 DISPlay[:WINDow[1]|2][:NUMeric[1]|2]:RESolution <numeric_value> 251 DISPlay[:WINDow[1]|2]:SELect[1]|2 253 DISPlay[:WINDow[1]|2][:STATe] <boolean> 255 DISPlay[:WINDow[1]|2]:TRACe:FEED <character_data> 257
6 FORMat Subsystem
FORMat Subsystem 260 FORMat[:READings]:BORDer <character_data> 261 FORMat[:READings][:DATA] <character_data> 263
N1913/1914A EPM Series Power Meters Programming Guide xiii
Page 14
7 MEMory Subsystem
MEMory Subsystem 266 MEMory[:METer]|:USB[1]|2:CATalog Commands 268 MEMory[:METer]|:USB[1]|2:CATalog[:ALL]? 269 MEMory[:METer]|:USB[1]|2:CATalog:STATe? 272 MEMory[:METer]|:USB[1]|2:CATalog:TABLe? 274 MEMory[:METer]|:USB[1]|2:CLEar Commands 277 MEMory[:METer]|:USB[1]|2:CLEar[:NAME] <character_data> 278 MEMory[:METer]|:USB[1]|2:CLEar:TABLe 280 MEMory[:METer]|:USB[1]|2:FREE Commands 281 MEMory[:METer]|:USB[1]|2:FREE[:ALL]? 282 MEMory[:METer]|:USB[1]|2:FREE:STATe? 283 MEMory[:METer]|:USB[1]|2:FREE:TABLe? 284 MEMory[:METer]|:USB[1]|2:NSTates? 285 MEMory[:METer]|:USB[1]|2:STATe Commands 286 MEMory[:METer]|:USB[1]|2:STATe:CATalog? 287 MEMory[:METer]|:USB[1]|2:STATe:DEFine <character_data>,<numeric_value> 288 MEMory[:METer]|:USB[1]|2:TABLe Commands 290 MEMory[:METer]|:USB[1]|2:TABLe:FREQuency <numeric_value>{,<numeric_value>} 291 MEMory[:METer]|:USB[1]|2:TABLe:FREQuency:POINts? 295 MEMory[:METer]|:USB[1]|2:TABLe:GAIN[:MAGNitude] <numeric_value>{,<numeric_value>} 296 MEMory[:METer]|:USB[1]|2:TABLe:GAIN[:MAGNitude]:POINts? 299 MEMory[:METer]|:USB[1]|2:TABLe:MOVE <character_data>,<character_data> 300 MEMory[:METer]|:USB[1]|2:TABLe:SELect <character_data> 302
8 OUTPut Subsystem
OUTPut Subsystem 304 OUTPut:RECorder[1]|2:FEED <data_handle> 305 OUTPut:RECorder[1]|2:LIMit:LOWer <numeric_value> 307 OUTPut:RECorder[1]|2:LIMit:UPPer <numeric_value> 309 OUTPut:RECorder[1]|2:STATe <boolean> 311 OUTPut:ROSCillator[:STATe] <boolean> 313 OUTPut:TRIGger[:STATe] <boolean> 315
xiv N1913/1914A EPM Series Power Meters Programming Guide
Page 15
9 SENSe Subsystem
[SENSe] Subsystem 319 [SENSe[1]]|2|3|4:AVERage Commands 322 [SENSe[1]]|2|3|4:AVERage:COUNt <numeric_value> 323 [SENSe[1]]|2|3|4:AVERage:COUNt:AUTO <boolean> 326 [SENSe[1]]|2|3|4:AVERage:SDETect <boolean> 329 [SENSe[1]]|2|3|4:AVERage[:STATe] <boolean> 331 [SENSe[1]]|SENSe2:BUFFer:COUNt <numeric_value> 333 [SENSe[1]]|SENSe2:CORRection Commands 335 [SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1][:INPut][:MAGNitude] <numeric_value> 336 [SENSe[1]]|SENSe2:CORRection:CSET[1]Commands 339 [SENSe[1]]|2|3|4:CORRection:CSET2 Commands 340 [SENSe[1]]|SENSe2:CORRection:CSET[1][:SELect] <string> 341 [SENSe[1]]|2|3|4:CORRection:CSET2[:SELect] <string> 343 [SENSe[1]]|SENSe2:CORRection:CSET[1]:STATe <boolean> 345 [SENSe[1]]|2|3|4:CORRection:DCYCle|GAIN3 Commands 349 [SENSe[1]]|2|3|4:CORRection:DCYCle|GAIN3[:INPut] [:MAGNitude] <numeric_value> 350 [SENSe[1]]|2|3|4:CORRection:DCYCle|GAIN3:STATe <boolean> 353 [SENSe[1]]|2|3|4:CORRection:FDOFfset|GAIN4[:INPut][:MAGNitude]? 355 [SENSe[1]]|2|3|4:CORRection:FDOFfset:UNIT <character_data> 356 [SENSe[1]]|2|3|4:CORRection:GAIN2 Commands 358 [SENSe[1]]|2|3|4:CORRection:GAIN2:STATe <boolean> 359 [SENSe[1]]|2|3|4:CORRection:GAIN2[:INPut][:MAGNitude] <numeric_value> 361 [SENSe[1]]|2|3|4:FREQuency[:CW|:FIXed] <numeric_value> 363 [SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STARt <numeric_value> 366 [SENSe[1]]|SENSe2:FREQuency[:CW|:FIXEd]:STEP <numeric_value> 369 [SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STOP <numeric_value> 372 [SENSe[1]]|2|3|4:MRATe <character_data> 375 [SENSe[1]]|2|3|4:POWer:AC:RANGe <numeric_value> 378 [SENSe[1]]|2|3|4:POWer:AC:RANGe:AUTO <boolean> 380 [SENSe[1]]|SENSe2:SPEed <numeric_value> 382 [SENSe[1]]|SENSe2:V2P ATYPe|DTYPe 385
10 STATus Subsystem
STATus Subsystem 388
N1913/1914A EPM Series Power Meters Programming Guide xv
Page 16
Status Register Set Commands 390 Device Status Register Sets 395 Operation Status Register Sets 397 STATus:OPERation 398 STATus:OPERation:CALibration[:SUMMary] 399 STATus:OPERation:LLFail[:SUMMary] 400 STATus:OPERation:MEASuring[:SUMMary] 401 STATus:OPERation:SENSe[:SUMMary] 402 STATus:OPERation:TRIGger[:SUMMary] 403 STATus:OPERation:ULFail[:SUMMary] 404 STATus:PRESet 405 Questionable Status Register Sets 406 STATus:QUEStionable 407 STATus:QUEStionable:CALibration[:SUMMary] 408 STATus:QUEStionable:POWer[:SUMMary] 409
11 SYSTem Subsystem
SYSTem Subsystem 413 SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric_value> 415 SYSTem:COMMunicate:LAN:AIP[:STATe] <boolean> 417 SYSTem:COMMunicate:LAN:CURRent:ADDRess? 418 SYSTem:COMMunicate:LAN:CURRent:DGATeway? 419 SYSTem:COMMunicate:LAN:CURRent:DNAMe? 420 SYSTem:COMMunicate:LAN:CURRent:SMASk? 421 SYSTem:COMMunicate:LAN:ADDRess <character_data> 422 SYSTem:COMMunicate:LAN:DGATeway <character_data> 424 SYSTem:COMMunicate:LAN:DHCP[:STATe] <boolean> 426 SYSTem:COMMunicate:LAN:DNAMe <character_data> 427 SYSTem:COMMunicate:LAN:HNAMe <character_data> 429 SYSTem:COMMunicate:LAN:MAC? 431 SYSTem:COMMunicate:LAN:RESTart 432 SYSTem:COMMunicate:LAN:SMASk <character_data> 433 SYSTem:DISPlay:BMP 435 SYSTem:ERRor? 436 SYSTem:HELP:HEADers? 444
xvi N1913/1914A EPM Series Power Meters Programming Guide
Page 17
SYSTem:LOCal 451 SYSTem:REMote 452 SYSTem:RWLock 453 SYSTem:VERSion? 454
12 TRIGger Subsystem
TRIGger Subsystem 458 ABORt[1]|2|3|4: 460 INITiate Commands 461 INITiate[1]|2|3|4:CONTinuous <boolean> 462 INITiate[1]|2|3|4[:IMMediate] 464 INITiate:CONTinuous:ALL <boolean> 465 INITiate:CONTinuous:SEQuence[1]|2|3|4 <boolean> 467 INITiate[:IMMediate]:ALL 469 INITiate[:IMMediate]:SEQuence[1]|2 |3|4 470 TRIGger Commands 471 TRIGger[1]|2|3|4:DELay:AUTO <boolean> 472 TRIGger[1]|2|3|4[:IMMediate] 474 TRIGger[1]|2|3|4:SOURce BUS|EXTernal|HOLD|IMMediate 475 TRIGger[:SEQuence[1]|2|3|4]:SLOPe <character_data> 478 TRIGger[:SEQuence[1]|2|3|4]:COUNt <numeric_value> 480 TRIGger[:SEQuence[1]|2|3|4]:DELay:AUTO <boolean> 482 TRIGger[:SEQuence[1]|2|3|4]:IMMediate 484 TRIGger[:SEQuence[1]|2|3|4]:SOURce BUS|EXTernal|HOLD|IMMediate 485
13 UNIT Subsystem
UNIT Subsystem 488 UNIT[1]|2|3|4:POWer <amplitude_unit> 489 UNIT[1]|2|3|4:POWer:RATio <ratio_unit> 491
14 SERVice Subsystem
SERVice Subsystem 495 SERVice:BIST:CALibrator <boolean> 505 SERVice:BIST:CW[1]|2:LINearity 507 SERVice:BIST:CW[1]|2:LINearity:PERRor? 508
N1913/1914A EPM Series Power Meters Programming Guide xvii
Page 18
SERVice:BIST:CW[1]|2:ZSET 509 SERVice:CALibrator:ADJ:COUR <numeric_value> 511 SERVice:CALibrator:ADJ:FINE <numeric_value> 512 SERVice:LAN:PHOStname 521 SERVice:SECure:ERASe 522 SERVice:SENSor[1]|2:CALFactor <cal_factor_data> 523 SERVice:SENSor[1]|2|3|4:CDATe? 525 SERVice:SENSor[1]|2|3|4:CPLace? 526 SERVice:SENSor[1]|2|3|4:FREQuency:MAXimum? 527 SERVice:SENSor[1]|2|3|4:FREQuency:MINimum? 528 SERVice:SENSor[1]|2|3|4:POWer:AVERage:MAXimum? 529 SERVice:SENSor[1]|2|3|4:POWer:USABle:MAXimum? 530 SERVice:SENSor[1]|2|3|4:POWer:USABle:MINimum? 531 SERVice:SENSor[1]|2|3|4:RADC? 532 SERVice:SENSor[1]|2|3|4:TNUMber? 534 SERVice:SENSor[1]|2|3|4:TYPE? 535 SERVice:SNUMber? 536 SERVice:STATe <boolean> 537 SERVice:VERSion:PROCessor <character_data> 539 SERVice:VERSion:SYSTem <character_data> 540
15 IEEE 488.2 Command Reference
SCPI Compliance Information 542 *CLS 543 *DDT <arbitrary block program data>|<string program data> 544 *ESE <NRf> 546 *ESR? 548 *IDN? 549 *OPC 550 *OPT? 551 *RCL <NRf> 552 *RST 553 *SAV <NRf> 554 *SRE <NRf> 555 *STB? 557
xviii N1913/1914A EPM Series Power Meters Programming Guide
Page 19
*TRG 559 *TST? 560 *WAI 561 GPIB Universal Commands 562
16 Programming Examples
Example 1: Identifying the EPM Series Power Meter In Use 568 Example 2: FETCh, MEASure, and READ Queries 569 Example 3: Making Repetitively Pulsed RF Power Measurement With Duty Cycle Correction 571 Example 4: CW Power Measurement 572 Example 5: Low Power Measurement 573 Example 6: Power Sweep Operation 574 Example 7: Frequency Sweep Operation 575 Example 8: Status of Various Settings 577 Example 9: Window Offset, Min/Max Hold, and Limits Settings 578
Appendix A Calibration Factor Block Layout
Calibration Factor Block Layout A-580
Appendix B Measurement Polling Example
Measurement Polling Example using VEE program B-586
Appendix C HP436A/HP437B/HP438A Command Summary
HP436A Command Summary C-590 HP437B Command Summary C-591 HP438A Command Summary C-594
N1913/1914A EPM Series Power Meters Programming Guide xix
Page 20

List of Figures

Figure B-1-1 Frequency Dependent Offset Tables 28 Figure B-1-2 Typical Averaged Readings on 8480 Series Sensors 35 Figure B-1-3 Averaging Range Hysteresis 36 Figure B-1-4 Limits Checking Application 40 Figure B-1-5 Limits Checking Results 41 Figure B-1-6 How Measurement are Calculated 49 Figure B-1-7 Generalized Status Register Model 51 Figure B-1-8 Typical Status Register Bit Changes 52 Figure B-1-9 Status System 58 Figure B-1-10 Hierarchical structure of SCPI 72 Figure B-1-11 Format of <character_data> 75 Figure B-1-12 Format of <non-decimal numeric> 77 Figure B-1-13 Format of <NR1> 78 Figure B-1-14 Format of <NR2> 78 Figure B-1-15 Format of <NR3> 79 Figure B-1-16 Format of <string> 80 Figure B-2-1 Measurement Display CALCulate Block Window 88 Figure B-3-1 Measurement Display CALCulate Block Window 172 Figure B-3-2 CALCulate Block 173 Figure B-9-1 Example of Averaged Readings 326 Figure B-11-1 IEEE 488.2 Arbitrary Block Program Data Format 444 Figure B-1 Example of VEE program used in measurement polling 587
xx N1913/1914A EPM Series Power Meters Programming Guide
Page 21

List of Tables

Ta bl e 1 -1 MEASure? and CONFigure Preset States 10 Ta bl e 1 -2 Possibilities of the defaulted source list parameter 15 Ta bl e 1 -3 Range of Values for Window Limits 42 Ta bl e 1 -4 Model of Sensor and Measurement Rates 45 Ta bl e 1 -5 Bit Definitions - Status Byte Register 59 Ta bl e 1 -6 Bit Definitions - Standard Event Register 60 Ta bl e 1 -7 Bit Definitions - Questionable Status Registers 62 Ta bl e 1 -8 Bit change conditions for Questionable Status Register 62 Ta bl e 1 -9 Bit Definitions - Operation Status 63 Table 1-10 Bit change conditions for Operation Status 64 Table 1-11 Bit Definitions - Device Status Register 65 Table 1-12 Bit change conditions for Device Status Register 66 Ta bl e 3 -1 Measurement Units 192 Ta bl e 3 -2 Measurement Units 195 Ta bl e 5 -1 Measurement Units 238 Ta bl e 5 -2 Measurement Units 241 Ta bl e 5 -3 Measurement Units 245 Ta bl e 5 -4 Measurement Units 248 Ta bl e 7 -1 8480 Series Power Sensor Tables 270 Ta bl e 7 -2 8480 Series Power Sensor Tables 275 Ta bl e 7 -3 Frequency and Calibration/Offset Factor List 292 Ta bl e 7 -4 Frequency and Calibration/Offset Factor List 296 Ta bl e 1 0- 1 Commands and events affecting Status Register 388 Ta bl e A- 1 Calibration Factor Block Layout: E4410 Series Sensors A-2 Ta bl e A- 2 Calibration Factor Block Layout: E9300 Series Sensors A-3 Ta bl e A- 3 Calibration Factor Block Layout: N8480 Series Sensors
N1913/1914A EPM Series Power Meters Programming Guide xxi
Page 22
xxii N1913/1914A EPM Series Power Meters Programming Guide
Page 23
N1913/1914A EPM Series Power Meters Programming Guide
1 Power Meter Remote Operation
Introduction 4 Configuring the Remote Interface 5
• Interface Selection 5
• GPIB Address 5
• LAN Configuration 6
• USB Configuration 7
Zeroing and Calibrating 8
• Zeroing 8
• Calibration 9
Making Measurement 10
• Using MEASure? 11
• Using the CONFigure Command 16
• Using the Lower Level Commands 25
Using Frequency Dependent Offset Tables 27
• Overview 27
• Editing Frequency Dependent Offset Tables 29
• Selecting a Frequency Dependent Offset Table 32
• Enabling a Frequency Dependent Offset Table 32
• Making the Measurement 32
Setting the Range, Resolution and Averaging 34
• Resolution 34
• Averaging 34
• Auto Averaging Mode 35
• Filter Length 36
Range 37
Agilent Technologies
1
Page 24
1 Power Meter Remote Operation
Configuring the Remote Interface 5
• Interface Selection 5
• GPIB Address 5
• LAN Configuration 6
Setting Offsets 38
• Channel Offsets 38
• Display Offsets 38
Setting Measurement Limits 40
• Setting Limits 40
• Checking for Limit Failures 42
• Using STATus 43
Getting the Best Speed Performance 44
• Measurement Rate 44
• Sensor 45
• Trigger Mode 45
• Output Format 47
• Units 47
• Command Used 47
• Fast Mode 48
How Measurements are Calculated 49
2 N1913/1914A EPM Series Power Meters Programming Guide
Page 25
Power Meter Remote Operation 1
Status Reporting 50
• The General Status Register Model 50
• How to Use Register 53
• The Condition Polling Method 53
• The SRQ Method 54
• Device Status Register 65
• Using the Operation Complete Commands 67
Saving and Recalling Power Meter Configurations 70
• How to Save and Recall a Configuration 70
Using Device Clear to Halt Measurements 71
An Introduction to the SCPI Language 72
• Mnemonic Forms 72
• Using a Colon (:) 72
• Using a Semicolon (;) 73
• Using a Comma (,) 73
• Using Whitespace 73
• Using “?” Commands 73
• Using “*” Commands 74
• Syntax Conventions 74
• Syntax Diagram Conventions 74
• SCPI Data Types 75
• Input Message Terminators 80
SCPI Compliance Information 82
Summary of Commands 84
This chapter describes the parameters that configure the power meter and helps you determine settings to optimize performance.
N1913/1914A EPM Series Power Meters Programming Guide 3
Page 26
1 Power Meter Remote Operation

Introduction

This chapter describes the parameters which configure the power meter and help you determine settings to optimize performance. It contains the following sections:
• “Configuring the Remote Interface” on page 5.
• “Zeroing and Calibrating” on page 8.
• “Making Measurement” on page 10.
• “Using Frequency Dependent Offset Tables” on page 27.
• “Setting the Range, Resolution and Averaging” on page 34.
• “Setting Offsets” on page 38.
• “Setting Measurement Limits” on page 40.
• “Getting the Best Speed Performance” on page 44.
• “How Measurements are Calculated” on page 49.
• “Status Reporting” on page 50.
• “Saving and Recalling Power Meter Configurations” on page 70.
• “Using Device Clear to Halt Measurements” on page 71.
• “An Introduction to the SCPI Language” on page 72.
• “SCPI Compliance Information” on page 82.
• “Summary of Commands” on page 84.
4 N1913/1914A EPM Series Power Meters Programming Guide
Page 27

Configuring the Remote Interface

NOTE
NOTE
This section briefly describes how to configure the GPIB, LAN and USB remote interfaces.
For more information on configuring the remote interface connectivity, refer to the Agilent Technologies USB/LAN/GPIB Interfaces Connectivity Guide. If you have installed the IO Libraries Suite, you can access the Connectivity Guide via the Agilent IO Libraries Control
icon. Alternatively, you can access the Connectivity Guide via the Web at www.agilent.com/find/connectivity.

Interface Selection

You can choose to control the power meter remotely using the GPIB, LAN or USB interfaces.
For information on selecting and configuring the remote interface manually from the front panel, refer to the EPM Series Power Meters Installation Guide.
Power Meter Remote Operation 1
It is expected that most users will use the front panel keys to set up the remote interfaces. The remote interface commands are provided for completeness (for the front panel operation).

GPIB Address

Each device on the GPIB (IEEE- 488) interface must have a unique address. You can set the power meter’s address to any value between 0 and 30. The power meter is shipped with a default address set to 13. The GPIB address is stored in non- volatile memory, and does not change when the power meter is switched off, or after a remote interface reset.
Your GPIB bus controller has its own address. Avoid using the bus controller’s address for any instrument on the interface bus. Agilent Technologies controllers generally use address 21.
N1913/1914A EPM Series Power Meters Programming Guide 5
Page 28
1 Power Meter Remote Operation
For information on setting the GPIB address manually from the front panel, refer to the EPM Series Power Meters Installation Guide.
• To set the GPIB address from the remote interface use the:
SYSTem:COMMunicate:GPIB:ADDRess command.
• To query the GPIB address from the remote interface use the:
SYSTem:COMMunicate:GPIB:ADDRess? query.

LAN Configuration

The power meter has three LAN operating modes:
• Dynamic IP (Dynamic Host Configuration Protocol or DHCP)
• Auto IP (Local PC Control or isolated (non-site) LAN)
•Static IP (Manual mode)
These three modes can be set up from the front panel. For front panel operation refer to the EPM Series Power Meter Installation Guide.
Configuring the LAN Remotely
To automatically configure the LAN settings, enable DHCP operation using the SYSTem:COMMunicate:LAN:DHCP[:STATe] command.
In this Dynamic IP mode the IP Address, Subnet Mask, and Default Gateway values are obtained from a DHCP server. Using this Dynamic IP mode does not require a detailed knowledge of your network configuration.
The IP Address, Subnet Mask, Default Gateway, and Host settings can be changed manually or remotely. To individually specify the LAN settings, use the following commands:
• IP Address -SYSTem:COMMuniucate:LAN:ADDRess
• Subnet Mask - SYSTem:COMMunicate:LAN:SMASk
• Default Gateway - SYSTem:COMMunicate:LAN:DGATeway
• Domain Name -SYSTem:COMMunicate:LAN:DNAMe
6 N1913/1914A EPM Series Power Meters Programming Guide
Page 29
Power Meter Remote Operation 1
NOTE
NOTE
NOTE
NOTE
• Hostname - SYSTem:COMMunicate:LAN:HNAMe
• Restart Network -SYSTem:COMMunicate:LAN:RESTart
The character_data values for the IP address, Subnet Mask, and Default Gateway can range between 0.0.0.0 and 255.255.255.255.
If you configure an invalid IP Address or an IP address that is used by another device or host, an error message is generated. This error can be read by using the SYSTem:ERRor? command.
The LAN setting values are stored in non- volatile memory and are not part of the save- recall function.

USB Configuration

The USB interface requires no front panel or remote configuration.
The USB address cannot be changed - it is set at the factory and is unique for each power meter.
For further information about the USB configuration refer to the EPM Series Power Meters Installation Guide.
Before connecting the USB cable, make sure that I/O software is installed on your computer.
For more information about Agilent IO Libraries software refer to the Connectivity Guide.
If you have installed other I/O Software, refer to documentation that accompanies the software
N1913/1914A EPM Series Power Meters Programming Guide 7
.
Page 30
1 Power Meter Remote Operation

Zeroing and Calibrating

Zeroing

Zeroing adjusts the power meter’s specified channel for a zero power reading.
The command CALibration[1]|2|3|4:ZERO:AUTO [ONCE|ON|OFF|0|1] causes the power meter to perform its zeroing routine on the specified channel when enabled. This adjusts the power meter for a zero power reading with no power supplied to the power sensor.
1|ON can only be used with a U2000 Series USB sensor. When 1|ON is enabled the the zero is maintained by a combination of zero on- the- fly for measurements and temperature compensation.
8 N1913/1914A EPM Series Power Meters Programming Guide
Page 31
Power Meter Remote Operation 1
NOTE

Calibration

The command used to calibrate the power meter is:
CALibration[1|2]:AUTO ONCE
It is recommended that you zero the power meter before calibrating.
Calibration Sequence
This feature allows you to perform a complete calibration sequence with a single query. The query is:
CALibration[1|2][:ALL]?
The query assumes that the power sensor is connected to the power reference oscillator. It turns the power reference oscillator on, then after calibrating, returns the power reference oscillator to the same state it was in prior to the command being received. The calibration sequence consists of:
1 Zeroing the power meter (CALibration[1|2]:ZERO:AUTO ONCE)
2 Calibrating the power meter (CALibration[1|2]:AUTO ONCE)
The query enters a number into the output buffer when the sequence is complete. If the result is 0 the sequence was successful. If the result is 1 the sequence failed. Refer to “CALibration[1]|2|3|4[:ALL]?” on page 218 for further information.
The CALibration[1]|2|3|4[:ALL] command is identical to the CALibration[1]|2|3|4[:ALL]? query except that no number is returned to
indicate the outcome of the sequence. You can examine the Questionable Status Register or the error queue to discover if the sequence has passed or failed. Refer to “Status
Reporting” on page 50 for further information.
N1913/1914A EPM Series Power Meters Programming Guide 9
Page 32
1 Power Meter Remote Operation

Making Measurement

The MEASure? and CONFigure commands provide a straight-forward method to program the power meter for measurements. You can select the measurement’s expected power level, resolution and with the N1914A the measurement type (that is single channel, difference or ratio measurements) all in one command. The power meter automatically presets other measurement parameters to default values as shown in
Table 1- 1 below.
Ta b le 1 -1 MEASure? and CONFigure Preset States
Command MEASure? and CONFigure Setting
Tri g g e r so u rc e (TRIGger:SOURce)
Filter (SENSe:AVERage:COUNt:AUTO)
Filter state(SENSe:AVERage:STATe)
Tri g g e r cy c le (INITiate:CONTinuous)
Tr ig g e rD e l a y (TRIGger:DELay:AUTO)
Immediate
On
On
Off
On
An alternative method to program the power meter is to use the lower level commands. The advantage of using the lower level commands over the CONFigure command is that they give you more precise control of the power meter. As shown in Table 1- 1, the CONFigure command presets various states in the power meter. It may be likely that you do not want to preset these states. Refer to “Using the Lower Level Commands” on page 25 for further information.
10 N1913/1914A EPM Series Power Meters Programming Guide
Page 33
Power Meter Remote Operation 1

Using MEASure?

The simplest way to program the power meter for measurements is by using the MEASure? query. However, this command does not offer much flexibility. When you execute the command, the power meter selects the best settings for the requested configuration and immediately performs the measurement. You cannot change any settings (other than the expected power value, resolution and with the N1914A the measurement type) before the measurement is taken. This means you cannot fine tune the measurement, for example, you cannot change the filter length. To make more f lexible and accurate measurements use the CONFIGure command. The measurement results are sent to the output buffer. MEASure? is a compound command which is equivalent to an ABORT, followed by a CONFigure, followed by a READ?.
MEASure? Examples
The following commands show a few examples of how to use the MEASure? query to make a measurement. It is advisable to read through these examples in order as they become increasingly more detailed. These examples configure the power meter for a measurement (as described in each individual example), automatically place the power meter in the “wait-for- trigger” state, internally trigger the power meter to take one reading, and then sends the reading to the output buffer.
These examples give an overview of the MEASure? query. For further information on the MEASure? commands refer to the section
“MEASure[1]|2|3|4 Commands” on page 154.
Example 1 - The Simplest Method
The following commands show the simplest method of making single channel (for example A or B) measurements. Using MEAS1? results in an upper window measurement, and MEAS2? in a lower window measurement. The channel associated with the window can be set using the source list parameter (see “Example 2 - Specifying the Source List Parameter”), or
N1913/1914A EPM Series Power Meters Programming Guide 11
Page 34
1 Power Meter Remote Operation
MEAS1?
specifies window
MEAS2?
MEAS1? DEF,DEF,(@1)
specifies window
specifies channel
NOTE
defaults as in this example (see “Agilent N1914A Only” on page 14).
Example 2 - Specifying the Source List Parameter
The MEASure command has three optional parameters, an expected power value, a resolution and a source list. These parameters must be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
The following example uses the source list parameter to specify the measurement channel as Channel A. The expected power and resolution parameters are defaulted, leaving them at their current settings. The measurement is carried out on the upper window.
The operation of the MEAS1? command when the source list parameter is defaulted is described in the note “Agilent N1914A Only” on page 14.
For the N1913A it is not necessary to specify a channel as only one channel is available.
Example 3 - Specifying the Expected Power Parameter
The previous example details the three optional parameters which can be used with the MEASure? command. The first optional parameter is used to enter an expected power value. Entering this parameter is only relevant if you are using an E-Series power sensor or N8480 Series power sensor (excluding Option CFT). The value entered determines which of the power sensor’s two ranges is used for the measurement. If the current setting of the power sensor’s range is no longer valid for the new measurement, specifying the expected power value decreases the time taken to obtain a result.
12 N1913/1914A EPM Series Power Meters Programming Guide
Page 35
Power Meter Remote Operation 1
MEAS2? -50,DEF,(@2)
specifies window specifies channel
specifies expected power value
MEAS1? DEF,3
specifies window
specifies resolution setting
The following example uses the expected value parameter to specify a value of –50 dBm. This selects the power sensor’s lower range (refer to
“Range” on page 37 for details of the range breaks). The resolution
parameter is defaulted, leaving it at its current setting. The source list parameter specifies a Channel B measurement. The measurement is displayed on the lower window.
Example 4 - Specifying the Resolution Parameter
The previous examples detailed the use of the expected value and source list parameters. The resolution parameter is used to set the resolution of the specified window. This parameter does not affect the resolution of the data, however it does affect the auto averaging setting (refer to
Figure 1- 2).
Since the filter length used for a channel with auto- averaging enabled is dependent on the window resolution setting, a conflict arises when a given channel is set up in both windows and the resolution settings are different. In this case, the higher resolution setting is used to determine the filter length.
The following example uses the resolution parameter to specify a resolution setting of 3. This setting represents 3 significant digits if the measurement suffix is W or %, and 0.01 dB if the suffix is dB or dBm.
Refer to Chapter 2, “MEASurement Commands”on page 85, for further
details on the resolution parameter. The expected power and source list parameters are defaulted in the example. The expected power value remains unchanged at its current setting. The source list parameter
defaults as described in the note “Agilent N1914A Only” on page 14. Note
that as the source list parameter is the last specified parameter you do not have to specify DEF. The measurement is carried out on the upper
window.
N1913/1914A EPM Series Power Meters Programming Guide 13
Page 36
1 Power Meter Remote Operation
MEAS2:POW:AC:DIFF? DEF,DEF,(@2),(@1)
specifies window
specifies between which channels the difference is calculated
Channel B - A
MEAS1:POW:AC:RAT? DEF,DEF,(@1),(@2)
specifies window
specifies the relationship of the channels in the ratio
Channel A / B
NOTE
Example 5 - Making a Difference Measurement
The following command is performed on the N1914A. It queries the lower window to make a difference measurement of Channel B - Channel A. The expected power and resolution parameters are defaulted, leaving them at their current settings.
Example 6 - Making a Ratio Measurement
The following command is performed on the N1914A. It queries the upper window to make a ratio measurement of Channel A/B. The expected power and resolution parameters are defaulted, leaving them at their current settings.
Agilent N1914A Only
The operation of the MEASure? command when the source list parameter is defaulted depends on the current setup of the window concerned (for example, A, B, A/B, A-B etc.) and on the particular command used (for example, MEAS[:POW][:AC]? and MEAS:POW:AC:RAT?).
This means that when the source list parameter is defaulted, there are a number of possibilities.
14 N1913/1914A EPM Series Power Meters Programming Guide
Page 37
Power Meter Remote Operation 1
Ta b le 1 -2 Possibilities of the defaulted source list parameter
Command Current Window Setup Measurement
MEAS1[:POW][AC]? Upper Window: A A
BB
Any Other Any Other A
MEAS2[:POW][AC]? Lower Window: A A
BB
Any Other B
MEAS1:POW:AC:RAT Upper Window: A/B A/B
B/A B/A
Any Other A/B
MEAS2:POW:AC:RAT Lower Window: A/B A/B
B/A B/A
Any Other A/B
MEAS1:POW:AC:DIFF? Upper Window: A-B A-B
B-A B-A
Any Other A-B
MEAS2:POW:AC:DIFF? Lower Window: A-B A-B
B-A B-A
Any Other A-B
N1913/1914A EPM Series Power Meters Programming Guide 15
Page 38
1 Power Meter Remote Operation

Using the CONFigure Command

When you execute this command, the power meter presets the optimum settings for the requested configuration (like the MEASure? query). However, the measurement is not automatically started and you can change measurement parameters before making measurements. This allows you to change the power meter’s configuration from the preset conditions. The power meter offers a variety of low-level commands in the SENSe, CALCulate, and TRIGger subsystems. For example, if you want to change the averaging use the [SENSe[1]]|SENSe2:AVERage:COUNt command.
Use the INITiate or READ? query to initiate the measurement.
Using READ?
CONFigure does not take the measurement. One method of obtaining a result is to use the READ? query. The READ? query takes the measurement using the parameters set by the CONFigure command then sends the reading to the output buffer. Using the READ? query obtains new data.
Using INITiate and FETCh?
CONFigure does not take the measurement. One method of obtaining the
result is to use the INITiate and FETCh? commands. The INITiate command causes the measurement to be taken. The FETCh? query retrieves a reading when the measurement is complete, and sends the reading to the output buffer. FETCh? can be used to display the measurement results in a number of different formats (for example, A/B and B/A) without taking fresh data for each measurement.
CONFigure Examples
The following program segments show how to use the commands READ?, INITiate and FETCh? and CONFigure to make measurements.
It is advisable to read through these examples in order as they become increasingly more detailed.
These examples give an overview of the CONFigure command. For further information on the CONFigure commands refer to Chapter 2, “MEASurement Commands”.
16 N1913/1914A EPM Series Power Meters Programming Guide
Page 39
Power Meter Remote Operation 1
Example 1 - The Simplest Method
The following program segments show the simplest method of querying the upper and lower window’s measurement results respectively.
Using READ?
*RST Reset instrument CONF1 Configure upper window - defaults to a Channel A
measurement
READ1? Take upper window (Channel A) measurement
*RST Reset instrument
CONF2 Configure lower window - defaults to a Channel A (N1913A),
Channel B (N1914A) measurement
READ2? Take lower window measurement (Channel A on N1913A, B
on N1914A)
Using INITiate and FETCh?
*RST Reset instrument
CONF1 Configure upper window - defaults to a Channel A
measurement
INIT1? Causes Channel A to make a measurement
FETC1? Retrieves the upper window’s measurement
For the N1913A only:
*RST Reset instrument
CONF2 Configure lower window - N1913A defaults to Channel A INIT1 Causes Channel A to make a measurement
FETC2? Retrieves the lower window’s measurement
For the N1914A only:
*RST Reset instrument
CONF2 Configure lower window INIT2? Causes Channel B to make a measurement
N1913/1914A EPM Series Power Meters Programming Guide 17
Page 40
1 Power Meter Remote Operation
NOTE
FETC2? Retrieves the lower window’s measurement
Example 2 - Specifying the Source List Parameter
The CONFigure and READ? commands have three optional parameters, an expected power value, a resolution and a source list. These parameters must be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
The following examples use the source list parameter to specify the measurement channel as Channel A. The expected power and resolution parameters are defaulted, leaving them at their current settings. The measurement is carried out on the upper window.
Although the READ? and FETCh? queries have three optional parameters it is not necessary to define them as shown in these examples. If they are defined they must be identical to those defined in the CONFigure command otherwise an error occurs.
For the N1913A it is not necessary to specify a channel as only one channel is available.
Using READ?
ABOR1 Aborts Channel A CONF1 DEF,DEF,(@1) Configures the upper window to make a
Channel A measurement using the current expected power and resolution settings
READ1? Takes the upper window’s measurement
18 N1913/1914A EPM Series Power Meters Programming Guide
Page 41
Power Meter Remote Operation 1
Using INITiate and FETCh?
ABOR1 Aborts Channel A
CONF1 DEF,DEF,(@1) Configures the upper window to make a
Channel A measurement using the current expected power and resolution settings
INIT1 Causes Channel A to make a measurement
FETC1? DEF,DEF,(@1) Retrieves the upper window’s
measurement
Example 3 - Specifying the Expected Power Parameter
The previous example details the three optional parameters which can be used with the CONFigure and READ? commands. The first optional parameter is used to enter an expected power value. Entering this parameter is only relevant if you are using an E- Series power sensor or N8480 Series power sensor (excluding Option CFT). The value entered determines which of the power sensor’s two ranges is used for the measurement. If the current setting of the power sensor’s range is no longer valid for the new measurement, specifying the expected power value decreases the time taken to obtain a result.
The following example uses the expected value parameter to specify a value of –50 dBm. This selects the power meter’s lower range (refer to
“Range” on page 37 for details of the range breaks). The resolution
parameter is defaulted, leaving it at its current setting. The source list parameter specifies a Channel B measurement. The measurement is carried out on the upper window.
N1913/1914A EPM Series Power Meters Programming Guide 19
Page 42
1 Power Meter Remote Operation
Using READ?
ABOR2 Aborts Channel B
CONF1 -50,DEF,(@2) Configures the upper window to make a
READ1? Takes the upper window’s measurement
Some fine tuning of measurements can be performed using the CONFigure and READ? commands. For example, in the above program segment some fine tuning can be performed by setting the filter length to 1024 and the trigger delay off.
1 ABOR2
2 CONF1 -50,DEF,(@2)
3 SENS2:AVER:COUN 1024
4 TRIG2:DEL:AUTO OFF
5 READ1?
Channel B measurement using an expected power of –50 dBm and the current resolution setting
Using INITiate and FETCh?
ABOR2 Aborts Channel B CONF1 -50,DEF,(@2) Configures the upper window to make a
Channel B measurement using an expected power of –50 dBm and the current resolution setting
INIT2 Causes Channel B to make a measurement FETC1? -50,DEF,(@2) Retrieves the upper window’s measurement
Some fine tuning of measurements can be carried out using the CONFigure command and INITiate and FETCh? commands. For example, in the above program segment some fine tuning can be carried out by
20 N1913/1914A EPM Series Power Meters Programming Guide
Page 43
Power Meter Remote Operation 1
setting the filter length to 1024 and the trigger delay off.
1 ABOR2
2 CONF1 -50,DEF,(@2)
3 SENS2:AVER:COUN 1024
4 TRIG2:DEL:AUTO OFF
5 INIT2
6 FETC1? -50,DEF,(@2)
Example 4 - Specifying the Resolution Parameter
The previous examples detailed the use of the expected value and source list parameters. The resolution parameter is used to set the resolution of the specified window. This parameter does not affect the resolution of the data, however it does affect the auto averaging setting (refer to Figure 1- 2 on page 35).
Since the filter length used for a channel with auto- averaging enabled is dependent on the window resolution setting, a conflict arises when a given channel is set up in both windows and the resolution settings are different. In this case, the higher resolution setting is used to determine the filter length.
The following example uses the resolution parameter to specify a resolution setting of 3. This setting represents 3 significant digits if the measurement suffix is W or %, and 0.01 dB if the suffix is dB or dBm (for further details on the resolution parameter refer to the commands in
Chapter 2, “MEASurement Commands”). Also, in this example the
expected power and source list parameters are defaulted. The expected power value is left unchanged at its current setting. The source list parameter is defaulted as described in the note “Agilent N1914A Only” on page 14. Note that as the source list parameter is the last specified parameter you do not have to specify DEF.
N1913/1914A EPM Series Power Meters Programming Guide 21
Page 44
1 Power Meter Remote Operation
Using READ?
ABOR1 Aborts Channel A
CONF1 DEF,3 Configures the upper window to make a measurement
READ1? Takes the upper window’s measurement. This is
Some fine tuning of the above program segment can be carried out for example, by setting the trigger delay off. The following program segment assumes that Channel A is currently being measured on the upper window.
1 ABOR1
2 CONF1 DEF,3
3 TRIG1:DEL:AUTO OFF
4 READ1?
using the current setting of the expected power and source list and a resolution setting of 3
Channel A or B measurement depending on current window setup.
Using INITiate and FETCh?
The following program segment assumes that Channel A is currently being measured on the upper window.
ABOR1 Aborts Channel A
CONF1 DEF,3 Configures the upper window to make a
measurement using the current setting of the expected power and source list and a resolution setting of 3
INIT1 Causes Channel A to make a measurement
FETC1? DEF,3 Retrieves the upper window’s measurement
Some fine tuning of the above program segment can be carried out for example, by setting the trigger delay off.
1 ABOR1
2 CONF1 DEF,3
22 N1913/1914A EPM Series Power Meters Programming Guide
Page 45
Power Meter Remote Operation 1
3 TRIG1:DEL:AUTO OFF
4 INIT1:IMM
5 FETC1? DEF,3
Example 5 - Making a Difference Measurement
The following program segment can be carried out on the N1914A. It queries the lower window to make a difference measurement of Channel A - Channel B. The expected power level and resolution parameters are defaulted, leaving them at their current settings. Some fine tuning of the measurement is carried out by setting the averaging, and the trigger delay to off.
Using READ?
ABOR1 ABOR2 CONF2:POW:AC:DIFF DEF,DEF,(@1),(@2) SENS1:AVER:COUN 1024 SENS2:AVER:COUN 1024 TRIG1:DEL:AUTO OFF TRIG2:DEL:AUTO OFF READ2:POW:AC:DIFF?
READ2:POW:AC:DIFF? DEF,DEF,(@2),(@1) (A second READ? query is sent
to make a Channel B - Channel A measurement using fresh measurement data).
Using INITiate and FETCh?
ABOR1 ABOR2 CONF2:POW:AC:DIFF DEF,DEF,(@1),(@2) SENS1:AVER:COUN 1024 SENS2:AVER:COUN 1024 TRIG1:DEL:AUTO OFF TRIG2:DEL:AUTO OFF INIT1:IMM INIT2:IMM FETC2:POW:AC:DIFF?
N1913/1914A EPM Series Power Meters Programming Guide 23
Page 46
1 Power Meter Remote Operation
FETC2:POW:AC:DIFF? DEF,DEF,(@2),(@1) (A second FETCh? query is sent to make a Channel B - Channel A measurement using the current measurement data).
Example 6 - Making a Ratio Measurement
The following program segment can be carried out on the N1914A. It queries the lower window to make a ratio measurement of Channel A/B. The expected power level and resolution parameters are defaulted, leaving them at their current settings. Some fine tuning of the measurement is carried out by setting the averaging.
Using READ?
ABOR1 ABOR2 CONF2:POW:AC:RAT DEF,DEF,(@1),(@2) SENS1:AVER:COUN 512 SENS2:AVER:COUN 256 READ2:POW:AC:RAT?
READ2:POW:AC:RAT? DEF,DEF,(@2),(@1) (A second R EAD? query is sent to
make a Channel B - Channel A ratio measurement using fresh measurement data.)
Using INITiate and FETCh?
ABOR1 ABOR2 CONF2:POW:AC:RAT DEF,DEF,(@1),(@2) SENS1:AVER:COUN 512 SENS2:AVER:COUN 256 INIT1:IMM INIT2:IMM FETC2:POW:AC:RAT?
FETC2:POW:AC:RAT? DEF,DEF,(@2),(@1) (A second FETCh? query is sent to
make a Channel B - Channel A measurement using the current measurement data.)
24 N1913/1914A EPM Series Power Meters Programming Guide
Page 47
Power Meter Remote Operation 1

Using the Lower Level Commands

An alternative method of making measurements is to use the lower level commands to set up the expected range and resolution. This can be done using the following commands:
[SENSe[1]]|SENSe2:POWER:AC:RANGe DISPlay[:WINDow[1|2]]:RESolution
The measurement type can be set using the following commands in the
CALCulate subsystem:
CALCulate[1|2]:MATH[:EXPRession] CALCulate[1|2]:RELative[:MAGNitude]
The advantage of using the lower level commands over the CONFigure command is that they give you more precise control of the power meter. As shown in Table 1- 1 the CONFigure command presets various states in the power meter. It may be likely that you do not want to preset these states.
Example
The following example sets the expected power value to –50 dBm and the resolution setting to 3 using the lower level commands. The measurement is a single Channel A measurement carried out on the lower window.
ABOR1 Aborts Channel A CALC2:MATH:EXPR
"(SENS1)"
SENS1:POW:AC:RANGE 0 Sets lower range (E- Series sensors and
DISP:WIND2:RES 3 Sets the lower window’s resolution to setting
INIT1 Causes Channel A to make a measurement
N1913/1914A EPM Series Power Meters Programming Guide 25
Displays Channel A on lower window
N8480 Series sensors (excluding Option CFT) only)
3
Page 48
1 Power Meter Remote Operation
FETC2? Retrieves the lower window’s measurement
26 N1913/1914A EPM Series Power Meters Programming Guide
Page 49

Using Frequency Dependent Offset Tables

This section describes how to use frequency dependent offset tables. These tables give you the ability to compensate for frequency effects in your test setup.

Overview

If the [SENSe[1]]|SENSe2:CORRection:CSET2:STATe command is OFF, the frequency dependent offset tables are not used. When [SENSe[1]]|SENSe2:CORRection:CSET2:STATe is ON, the frequency dependent offset tables are used, providing you with a quick and convenient method of compensating for your external test setup over a range of frequencies. Note that when selected, frequency dependent offset correction is IN ADDITION to any correction applied for sensor frequency response. The power meter is capable of storing 10 frequency dependent offset tables of 80 frequency points each.
To use frequency dependent offset tables you:
Power Meter Remote Operation 1
1 Edit a frequency dependent offset table if necessary.
2 Select the frequency dependent offset table.
3 Enable the frequency dependent offset table.
4 Zero and calibrate the power meter.
If you are using an 8480 Series sensors or N8480 Series sensor with Option CFT, the reference calibration factor used during the calibration must be entered manually.
5 Specify the frequency of the signal you want to measure. The required
offset is automatically set by the power meter from the frequency dependent offset table.
6 Make the measurement.
Figure 1- 1 illustrates how frequency dependent offset tables operate.
N1913/1914A EPM Series Power Meters Programming Guide 27
Page 50
1 Power Meter Remote Operation
TABLE N
FREQ
FREQ
1
. . . . . . . . .
.
.
FREQ
2
80
OFFSET
OFFSET
1
. . . . . . . . .
.
.
80
TABLE 1
FREQ
FREQ
1
. . . . . . . . .
.
.
FREQ
2
80
OFFSET
OFFSET
1
. . . . . . . . .
.
.
2
80
TABLE 10
FREQ
FREQ
1
. . . . . . . . .
.
.
FREQ
2
80
OFFSET
OFFSET
1
. . . . . . . . .
.
.
OFFSET
2
80
OFFSET = Frequency Dependent Offset
FREQ
FREQ
1
. . . . . . . . .
.
.
FREQ
2
80
OFFSET
OFFSET
1
. . . . . . . . .
.
.
OFFSET
2
80
Frequen cy of the signal you want to measu re
TABLE SELECTED
OFFSET
2
OFFSET
Frequency dependent offset used to make Measurement. Calculated by the power meter using linear interpolation.
28 N1913/1914A EPM Series Power Meters Programming Guide
Figure 1-1 Frequency Dependent Offset Tables
Page 51
Power Meter Remote Operation 1

Editing Frequency Dependent Offset Tables

It is not possible to create any additional frequency dependent offset tables. However, the 10 existing tables can be edited using the MEMory subsystem. To do this:
1 Select one of the existing tables using:
MEMory:TABle:SELect <string>
For information on naming frequency dependent offset tables see
“Naming Frequency Dependent Offset Tables” on page 31. For
information on the current names which you can select refer to “Listing
the Frequency Dependent Offset Table Names” on page 30.
2 Enter the frequency data using:
MEMory:TABle:FREQuency <numeric_value> {,<numeric_value>}
3 Enter the offset factors as shown in the table below using:
MEMory:TABle:GAIN <numeric_value> {,<numeric_value>}
Frequency Offset
Frequency 1 Offset 1
Frequency 2 Offset 2
""
Frequency n Offset n
4 If required, rename the frequency dependent offset table using:
MEMory:TABLe:MOVE <string>,<string>. The first <string> parameter identifies the existing table name, and the second identifies the new table name.
N1913/1914A EPM Series Power Meters Programming Guide 29
Page 52
1 Power Meter Remote Operation
NOTE
The legal frequency suffix multipliers are any of the IEEE suffix multipliers, for example, KHZ, MHZ, and GHZ. If no units are specified the power meter assumes the data is Hz.
PCT is the only legal unit for offset factors and can be omitted.
The frequency and offset data must be within range. Refer to the individual commands in
Chapter 4 for their specified ranges.
Any offset values entered into the table should exclude the effect of the sensor. Characterization of the test setup independently of the sensor allows the same table to be used with any sensor.
Ensure that the frequency points you use cover the frequency range of the signals you want to measure. If you measure a signal with a frequency outside the frequency range defined in the frequency dependent offset table, then the power meter uses the highest or lowest frequency point in the table to calculate the offset.
To make subsequent editing of a frequency dependent offset table simpler, it is recommended that you retain a copy of your data in a program.
Listing the Frequency Dependent Offset Table Names
To list the frequency dependent offset tables currently stored in the power meter, use the following command:
MEMory:CATalog:TABLe?
Note that all tables are listed; including sensor calibration tables.
The power meter returns the data in the form of two numeric parameters and a string list representing all stored tables.
• <numeric_value>,<numeric_value>{,<string>}
The first numeric parameter indicates the amount of memory, in bytes, used for storage of tables. The second parameter indicates the memory, in bytes, available for tables.
Each string parameter returned indicates the name, type and size of a stored frequency dependent offset table:
• <string>,<type>,<size>
The <string>, <type> and <size> are all character data. The <type> is
always TABL. The <size> is displayed in bytes.
For example, a sample of the response may look like:
30 N1913/1914A EPM Series Power Meters Programming Guide
Page 53
Power Meter Remote Operation 1
560,8020,“Offset_1,TABL,220”,”Offset_2,TABL,340” ....
Naming Frequency Dependent Offset Tables
To rename a frequency dependent offset table use:
MEMory:TABLe:MOVE <string>,<string>
The first <string> parameter identifies the existing table name, and the second identifies the new table name.
The following rules apply to frequency dependent offset table names:
1 Table names use a maximum of 12 characters.
2 All characters must be upper or lower case alphabetic characters, or
numeric (0- 9), or an underscore (_).
No spaces are allowed in the name.
Reviewing Table Data
To review the data stored in a frequency dependent offset table, use the following commands:
MEMory:TABLe:SELect "Offset1"
Select the sensor calibration table named “Offset1”.
MEMory:TABLe:SELect?
Query command which returns the name of the currently selected table.
MEMory:TABLe:FREQuency:POINTs?
Query command which returns the number of stored frequency points.
MEMory:TABLe:FREQuency?
Query command which returns the frequencies stored in the frequency dependent offset table (in Hz).
MEMory:TABLe:GAIN[:MAGNitude]:POINTs?
Query command which returns the number of offset factor points stored in the frequency dependent offset table.
MEMory:TABLe:GAIN[:MAGNitude]?
Query command which returns the offset factors stored in the frequency dependent offset table.
Modifying Data
If you need to modify the frequency and offset factor data stored in a frequency dependent offset table you need to resend the complete data lists.
N1913/1914A EPM Series Power Meters Programming Guide 31
Page 54
1 Power Meter Remote Operation
If you have retained the original data in a program, edit the program and resend the data.

Selecting a Frequency Dependent Offset Table

After you have created the frequency dependent offset table, you can select it using the following command:
[SENSe[1]]|SENSe2:CORRection:CSET2[:SELect] <string>
To find out which frequency dependent offset table is currently selected, use the query:
[SENSe[1]]|SENSe2:CORRection:CSET2[:SELect]?

Enabling a Frequency Dependent Offset Table

To enable the frequency dependent offset table, use the following command:
[SENSe[1]]|SENSe2:CORRection:CSET2:STATe ON
If you set [SENSe[1]]|SENSe2:CORRection:CSET2:STATe to ON and no frequency dependent offset table is selected error –221, “Settings conflict” occurs.

Making the Measurement

To make the power measurement, set the power meter for the frequency of the signal you want to measure. The power meter automatically sets the calibration factor. Use either the INITiate, FETCh? or the READ? query to initiate the measurement as shown in the following program segments:
INITiate Example
ABORt1 CONFigure1:POWer:AC DEF,1,(@1) SENS1:CORR:CSET2:SEL "Offset1"
32 N1913/1914A EPM Series Power Meters Programming Guide
Page 55
Power Meter Remote Operation 1
NOTE
SENS1:CORR:CSET2:STAT ON SENSe1:FREQuency 500KHZ INITiate1:IMMediate FETCh1?
READ? Example
ABORt1 CONFigure1:POWer:AC DEF,2,(@1) SENS1:CORR:CSET2:SEL "Offset1" SENS1:CORR:CSET2:STAT ON SENSe1:FREQuency 500KHZ READ1?
If the measurement frequency does not correspond directly to a frequency in the frequency dependent offset table, the power meter calculates the offset using linear interpolation.
If you enter a frequency outside the frequency range defined in the frequency dependent offset table, then the power meter uses the highest or lowest frequency point in the table to set the offset.
To find out the value of the offset being used by the power meter to make a measurement, use the query command:
SENSe:CORRection:GAIN4|FDOFfset[:INPut][MAGNITUDE]?
The response may be an interpolated value.
N1913/1914A EPM Series Power Meters Programming Guide 33
Page 56
1 Power Meter Remote Operation

Setting the Range, Resolution and Averaging

This section provides an overview of setting the range, resolution and averaging. For more detailed information about these features refer to the individual commands in Chapter 9, “SENSe Subsystem”.

Resolution

You can set the window’s resolution using the following command:
DISPlay[:WINDow[1]|2][:NUMeric[1]|2] :RESolution <numeric_value>
There are four levels of resolution available (1 through 4).
When the measurement suffix is W or % this parameter represents the number of significant digits. When the measurement suffix is dB or dBM, 1 through 4 represents 1, 0.1, 0.01, and 0.001 dB respectively.
Refer to the :RESolution command on page 251 for further information.

Averaging

The power meter has a digital filter to average power readings. The number of readings averaged can range from 1 to 1024. This filter is used to reduce noise, obtain the desired resolution and to reduce the jitter in the measurement results. However, the time to take the measurement is increased. You can select the filter length or you can set the power meter to auto filter mode. To enable and disable averaging use the following command:
[SENSe[1]]|SENSe2:AVERage[:STATe] <boolean>
34 N1913/1914A EPM Series Power Meters Programming Guide
Page 57
Power Meter Remote Operation 1
NOTE
10 dB
10 dB
Minimum Sensor Power
Maximum Sensor Power
Power Sensor
Dynamic Range
10 dB
10 dB
1234
8 8 128 128
Resolution Setting
Number of Averages
1 1 16 256
11232
11116
1118

Auto Averaging Mode

To enable and disable auto filter mode, use the following command:
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <boolean>
When the auto filter mode is enabled, the power meter automatically sets the number of readings averaged together to satisfy the filtering requirements for most power measurements. The number of readings averaged together depends on the resolution and the power level currently being measured. Figure 1- 2 lists the number of readings averaged for each range and resolution when the power meter is in auto filter mode.
Figure 1-2 applies to 8480 Series only.
N1913/1914A EPM Series Power Meters Programming Guide 35
Figure 1-2 Typical Averaged Readings on 8480 Series Sensors
Figure 1- 3 illustrates part of the power sensor dynamic range hysteresis.
Page 58
1 Power Meter Remote Operation
Minimum Sensor Power Minimum Sensor Power + 10 dB
Range Hysteresis
10.5 dB9.5 dB
Figure 1-3 Averaging Range Hysteresis

Filter Length

You specify the filter length using the following command:
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
The range of values for the filter length is 1 to 1024. Specifying this command disables automatic filter length selection. Increasing the value of the filter length reduces measurement noise. However, the time to take the measurement is increased.
36 N1913/1914A EPM Series Power Meters Programming Guide
Page 59
Power Meter Remote Operation 1

Range

The power meter has no internal ranges which can be set. The only ranges that can be set are those of the E-Series power sensor and N8480 Series power sensors (excluding Option CFT). With an E- Series power sensor or N8480 Series power sensors (excluding Option CFT), the range can be set either automatically or manually. Use autoranging when you are not sure of the power level you will be measuring.
Setting the Range
To set the range manually use the following command:
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
If the <numeric_value> is set to:
• 0, the sensor’s lower range is selected. (For example, this range is –70 to –13.5 dBm for the E4412A power sensor.)
• 1, the sensor’s upper range is selected. (For example, this range is –14.5 to +20 dBm for the E4412A power sensor.)
For details on the range limits of other E- Series power sensor and N8480 Series power sensor (excluding Option CFT), refer to the appropriate power sensor manual.
For further information on this command refer to page 378.
To enable autoranging use the following command:
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO ON
Use autoranging when you are not sure of the power level you will be measuring.
N1913/1914A EPM Series Power Meters Programming Guide 37
Page 60
1 Power Meter Remote Operation
Loss
1
Gain
------------ -
=
Gain Loss–=
NOTE

Setting Offsets

Channel Offsets

The power meter can be configured to compensate for signal loss or gain in your test setup (for example, to compensate for the loss of a 10 dB attenuator). You use the SENSe command subsystem to configure the power meter. Gain and loss correction are a coupled system. This means that a gain set by [SENSe[1]]|SENSe2:CORRection:GAIN2 is represented in the [SENSe[1]]|SENSe2:CORRection:LOSS2? command. If you enter an offset value the state is automatically enabled. However it can be enabled and disabled using either the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe or [SENSe[1]]|SENSe2:CORRection:LOSS2:STATe commands.
LOSS2 is coupled to GAIN2 by the equation when the default
unit
is linear, and when the default is logarithmic.
You can only use LOSS2 and GAIN2 for external losses and gains. LOSS1 and GAIN1 are specifically for calibration factors.

Display Offsets

Display offset values can be entered using the
CALCulate[1|2]:GAIN[:MAGNitude] command. CALCulate[1|2]:GAIN:STATe must be set to ON to enable the offset
value. If you enter an offset value the state is automatically enabled. This offset is applied after any math calculations (refer to Figure 1- 6 on page 49).
Example
The following example program, in HP Basic, details how to use the channel and display offsets on an N1914A making a Channel A/B ratio measurement.
38 N1913/1914A EPM Series Power Meters Programming Guide
Page 61
Power Meter Remote Operation 1
A
dBm
10–
B
dBm
10–
------------------------ -
⎝⎠
⎛⎞
20–
⎝⎠
⎛⎞
dB
The final result is:
10 !Create I/O path name 20 ASSIGN @POWER TO 713 30 !Clear the power meter’s interface 40 CLEAR @POWER 50 !Set the power meter to a known state 60 OUTPUT @POWER;"*RST" 70 !Configure the Power Meter to make the measurement 80 OUTPUT @Power;"CONF:POW:AC:RAT 20DBM,2,(@1),(@2)" 90 !Set the measurement units to dBm 100 OUTPUT @POWER;"UNIT:POW DBM" 110 !Set the power meter for channel offsets of -10 dB 120 OUTPUT @POWER;"SENS1:CORR:GAIN2 -10" 130 OUTPUT @POWER;"SENS2:CORR:GAIN2 -10" 140 !Enable the gain correction 150 OUTPUT @POWER;"SENS:CORR:GAIN2:STATe ON" 160 OUTPUT @POWER;"SENS2:CORR:GAIN2:STATe ON" 170 !Set the power meter for a display offset of -20 dB 180 OUTPUT @POWER;"CALC1:GAIN -20 DB" 190 PRINT "MAKING THE MEASUREMENT" 200 !Initiate the measurement 210 OUTPUT @Power;"INIT1:IMM" 220 OUTPUT @Power;"INIT2:IMM" 230 ! ... and get the result 240 OUTPUT @Power;"FETC:POW:AC:RAT? 20DBM,2,(@1),(@2)" 250 ENTER @Power;Reading 260 ! 270 PRINT "The measurement result is ";Reading;"dB." 280 END
For further information on channel offsets refer to page 355. For further information on display offsets refer to page 180.
N1913/1914A EPM Series Power Meters Programming Guide 39
Page 62
1 Power Meter Remote Operation
Power Meter
Swept Source
CHANNEL A
INPUT
OUT
IN OUT
Device Under Test

Setting Measurement Limits

You can configure the power meter to detect when a measurement is outside of a predefined upper and/or lower limit value.
Limits are window or measurement display line based and can be applied to power, ratio or difference measurements.

Setting Limits

The power meter can be configured to verify the power being measured against an upper and/or lower limit value. The range of values that can be set for lower and upper limits is –150.00 dBm to +230.00 dBm. The default upper limit is +90.00 dBm and the default lower limit is –90.00 dBm.
A typical application for this feature is shown in Figure 1- 4.
Figure 1-4 Limits Checking Application
40 N1913/1914A EPM Series Power Meters Programming Guide
Page 63
Power Meter Remote Operation 1
+4 dBm
+10 dBm
Amplitude
Frequency
o
o
o
o
o
o
o
Fail
Fail
Figure 1-5 Limits Checking Results
The range of values that can be set for the upper and lower limits and the default values depends on the measurement units in the currently measurement line - see Table 1- 3.
N1913/1914A EPM Series Power Meters Programming Guide 41
Page 64
1 Power Meter Remote Operation
Ta b le 1 -3 Range of Values for Window Limits
Window Units
dB +200 dB
dBm +230 dBm
% 999.9 X% 100.0 a% 100.0 M% 100.0 p%
W 100.000 XW 1.000 aW 1.000 MW 1.000 pW
Maximum Minimum Maximum Minimum
–180 dB
–150 dBm
60 dB
90 dBm
Default
–120 dB
–90 dBm

Checking for Limit Failures

There are two ways to check for limit failures:
1 Use the SENSe:LIMit:FAIL? and SENSe:LIMit:FCOunt? commands
for channel limits or the
CALCulate[1|2]:LIMit:FAIL? and the CALCulate[1|2]:LIMit:FCOunt? for window limits
2 Use the STATus command subsystem
Using SENSe and CALCulate
Using SENSe to check the channel limit failures in Figure 1- 5 would return the following results:
SENSe:LIMit:FAIL? Returns 1 if there has been 1 or
more limit failures or 0 if there have been no limit failures. In this case 1 is returned.
SENSe:LIMit:FCOunt? Returns the total number of limit
failures, in this case 2.
Use the equivalent CALCulate commands for checking window limit failures.
42 N1913/1914A EPM Series Power Meters Programming Guide
Page 65
Power Meter Remote Operation 1
NOTE
If TRIGger:DELay:AUTO is set to ON, then the number of failures returned by SENSe:LIMit:FCOunt? or CALCulate[1|2]:LIMit:FCOunt? is affected by the
current filter settings.

Using STATus

If using GPIB, you can use the STATus subsystem to generate an SRQ to interrupt your program when a limit failure occurs. This is a more efficient method than using SENSe or CALCulate, since you do not need to check the limit failures after every power measurement.
Refer to “Status Reporting” on page 50 and “STATus Subsystem” on page 387 for further information.
N1913/1914A EPM Series Power Meters Programming Guide 43
Page 66
1 Power Meter Remote Operation

Getting the Best Speed Performance

This section discusses the factors that influence the speed of operation (number of readings/sec) of a EPM Series power meter.
The following factors are those which have the greatest effect upon measurement speed (in no particular order):
• The selected measurement rate, i.e. NORMal, DOUBle, FAST.
•The sensor being used.
• The trigger mode (for example, free run, trigger with delay etc.).
•The output format: ASCii or REAL.
• The units used for the measurement.
• The command used to take a measurement.
In addition, in FAST mode there are other influences which are described in “Fast Mode” on page 48.
The following paragraphs give a brief description of the above factors and how they are controlled from SCPI.

Measurement Rate

There are three possible speed settings NORMal, DOUBle and FAST. These are set using the SENSe:MRATe command and can be applied to each channel independently.
In NORMal and DOUBle modes, full instrument functionality is available and these settings can be used with all sensors. FAST mode is only available for the E- Series sensors. Also, in FAST mode averaging, limits and ratio/difference math functions are disabled.
Refer to “Specifications” in the EPM Series Power Meters User’s Guide to see the influence of these speed settings on the accuracy and noise performance of the power meter.
44 N1913/1914A EPM Series Power Meters Programming Guide
Page 67
Power Meter Remote Operation 1

Sensor

Different measurement rates are achievable depending on the sensor type being used, as shown in Table 1- 4:
Ta b le 1 -4 Model of Sensor and Measurement Rates
Sensor Measurement Rate
NORMal DOUBle FAST
8480 Series and N8480 Series 20 reading/s 40 reading/s NA
E-Series E4410 and E9300 50 ms
20 reading/s
U2000 Series 20 reading/s 40 reading/s Up to 110
25 ms
40 reading/s
Up to 400
readings/s

Tri gg er Mo de

The power meter has a very flexible triggering system. For simplicity, it can be described as having three modes:
•Free Run: When a channel is in Free Run, it continuously takes measurements on this channel. A channel is in free run when
INITiate:CONTinuous is set to ON and TRIGger:SOURce is set to IMMediate.
• Triggered Free Run: When a channel is in Triggered Free Run
Continuous Trigger, it takes a new measurement each time a trigger event is detected. A channel is in Triggered Free Run Continuous Trigger when INITiate:CONTinuous is set to ON and TRIGger:SOURce is not set to IMMediate.
• Single Shot: When a channel is in Single Shot, it takes a new
measurement when a trigger event is detected and then returns to the idle state. A channel is in Single Shot when INITiate:CONTinuous is set to OFF. Note that a measurement can take several INT/EXT triggers depending on the filter settings. Refer to
“TRIGger[1]|2|3|4:DELay:AUTO <boolean>” on page 472 for further
information.
N1913/1914A EPM Series Power Meters Programming Guide 45
Page 68
1 Power Meter Remote Operation
NOTE
A trigger event can be any of the following:
• The input signal meeting the trigger level criteria.
• Auto-level triggering being used.
•A TRIGger GET or *TRG command being sent.
• An external TTL level trigger being detected.
Trigger with Delay
This can be achieved using the same sequences above (apart from the second) with TRIG:DEL:AUTO set to ON. Also, the MEAS? command operates in trigger with delay mode.
In trigger with delay mode, a measurement is not completed until the power meter filter is full. In this way, the reading returned is guaranteed to be settled. In all other modes, the result returned is simply the current result from the filter and may or may not be settled. This depends on the current length of the filter and the number of readings that have been taken since a change in power level.
With trigger with delay enabled, the measurement speed can be calculated roughly using the following equation:
readings/sec = speed (as set by SENSe:SPEed) / filter length
For example, with a filter length of 4 and SENS:SPE set to 20, approximately 5 readings/sec is calculated by the power meter.
Typically, free run mode provides the best speed performance from the power meter (especially in 200 readings/sec mode).
46 N1913/1914A EPM Series Power Meters Programming Guide
Page 69
Power Meter Remote Operation 1

Output Format

The power meter has two output formats for measurement results: ASCii and REAL. These formats are selected using the FORMat command. When FORMat is set to REAL, the returned result is in IEEE 754 floating-point format (note that the byte order can be changed using FORMat:BORDer) plus <LF> as an end sentinel of the block.
The REAL format is likely to be required only for FAST mode as it reduces the amount of bus traffic.

Units

The power meter can output results in either linear or log units. The internal units are linear, therefore optimal performance is achieved when the results output are also in linear units (since the overhead of performing a log function is removed).

Command Used

In Free Run mode, FETCh? must be used to return a result.
In other trigger modes, there are a number of commands which can be used, for example, MEASure?, READ?, FETCh? Note that the MEAS? and READ? commands are compound commands—they perform a combination of other lower level commands. Typically, the best speed performance is achieved using the low level commands directly.
Trigger Count
To get the fastest measurement speed the a TRIG:COUNT must be set to return multiple measurements for each FETCh command. For average only measurements a count of 4 is required, however, 10 is recommended. In normal mode (peak measurements) a count of 50 is required to attain 1000 readings per second.
N1913/1914A EPM Series Power Meters Programming Guide 47
Page 70
1 Power Meter Remote Operation

Fast Mode

In the highest speed setting, the limiting factor tends to be the speed of the controller being used to retrieve results from the power meter, and to a certain extent, the volume of remote traffic. The latter can be reduced using the FORMat REAL command to return results in binary format. The former is a combination of two factors:
• the hardware platform being used
• the programming environment being used
48 N1913/1914A EPM Series Power Meters Programming Guide
Page 71

How Measurements are Calculated

TRIGger
DISPlay
WINDow2
:FORMat :METer : SELect [1]|2
: NUMeric[1 ]:RESolu tion
:NUMeric2:RESolution
Upper Meas
Lowe r Me a s
UNIT4
Conversion
:P OW
CAL C ul ate2
Maths O ffset Relat ive
Limits
:MA TH : GA IN :REL
:LIM
Swit ch Swit ch
:FEED
UNIT2
Conversion
:P OW
UNIT 3
Conv er sion
:POW
UNIT 1
Conv er sion
:POW
CAL C ul ate4
Maths O ffset Relat ive
Limits
:MA TH : GA IN :REL
:LIM
Swit ch Swit ch
:FEED
CAL C ul ate1
Maths O ffset Relat ive
Limits
:MA TH : GA IN :REL
:LIM
Swit ch Swit ch
:FEED
CAL C ul ate3
Maths O ffset Relat ive
Limits
:MA TH : GA IN :REL
:LIM
Swit ch Swit ch
:FEED
[WINDo w[ 1]]
:FORMat :METer : SELect [1]|2
: NUMeric[1 ]:RESolu tion
:NUMeric2:RESolution
Upper Meas
Lowe r Me a s
SENSe1
Sensor Filter
Freq.
Corr.
Offset
Duty
Cycle
:SPEe d
:POW:A C:RANG
:POW:AC:RANG:AUTO:DIR
:DET :FUNC
:AVER[1]
:FREQ :CORR:CFAC : CORR:CSET
:CORR:GAIN2
:CORR:LOSS2
:CORR:DCYC
SENSe2
Sensor Filter
Freq.
Corr .
Offset
Duty
Cycle
:SPEed
:PO W: AC:RANG
:POW:AC:RANG:AUTO:DIR
:DET:FU NC
:AVER[1]
:FR EQ :CORR:CFAC :CORR:CSET
:CORR:GAIN2
:CO RR:LOSS2
: CORR:DCYC
:CO NTrast :ENA Ble :FOR Mat
FO R M at
Swi tc h Swi tc h Swi tc h Swi tc h
MEAS? READ? FETC?
CONF
NOTE
Figure 1- 6 details how measurements are calculated. It shows the order in
which the various power meter functions are implemented in the measurement calculation.
Power Meter Remote Operation 1
N1913/1914A EPM Series Power Meters Programming Guide 49
Figure 1-6 How Measurement are Calculated
The MEASure commands in this figure can be replaced with the FETCh? and READ? commands.
All references to Channel B in the above diagram refer to the N1914A only.
Page 72
1 Power Meter Remote Operation

Status Reporting

Status reporting is used to monitor the power meter to determine when events have occurred. Status reporting is accomplished by configuring and reading status registers.
The power meter has the following main registers:
•Status Register
•Standard Event Register
•Operation Status Register
• Questionable Status Register
• Device Status Register
There are other registers that exist “behind” the main registers, and are described later in this chapter.
Status and Standard Event registers are read using the IEEE- 488.2 common commands.
Operation and Questionable Status registers are read using the SCPI STATus command subsystem.

The General Status Register Model

The generalized status register model shown in Figure 1-7 is the building block of the SCPI status system. This model consists of a condition register, a transition filter, an event register and an enable register. A set of these registers is called a status group.
50 N1913/1914A EPM Series Power Meters Programming Guide
Page 73
Power Meter Remote Operation 1
0 1
2
Bit 0
Bit 1
Bit 2
Bit 3
Cond ition
Register
Tra n si t io n
Filter
Event
Register
Enable
Register
Logical OR
Summary
Bit
Figure 1-7 Generalized Status Register Model
When a status group is implemented in an instrument, it always contains all of the component registers. However, there is not always a corresponding command to read or write to ever y register.
Condition Register
The condition register continuously monitors the hardware and firmware status of the power meter. There is no latching or buffering for this register, it is updated in real time. Condition registers are read- only.
Transition Filter
The transition filter specifies which types of bit state changes in the condition registers and set corresponding bits in the event register. Transition filter bits may be set for positive transitions (PTR), negative transitions (NTR), or both. Transition filters are read-write. They are unaffected by *CLS or queries. After STATus:PRESet the NTR register is set to 0 and all bits of the PTR are set to 1.
Event Register
The event register latches transition events from the condition register as specified by the transition filter. Bits in the event register are latched and on setting they remain set until cleared by a query or a *CLS. Also on setting, an event bit is no longer affected by condition changes. It remains set until the event register is cleared; either when you read the register or when you send the *CLS (clear status) command. Event registers are read- only.
N1913/1914A EPM Series Power Meters Programming Guide 51
Page 74
1 Power Meter Remote Operation
00 01
110
011
0 1
00 0
0
000
1
11
011
0
1
000
1
10
0
1
0
1
00 01
0
01
0 1
00
0
0
0
0
00 000 000
1
0
0
0
0
0
Condition
PTR
NTR
Enable
Condition
Event
Summary Bit
Condition
Event
Summary Bit
Condition
Event
Summary Bit
Condition
Event
Summary Bit
Condition
Event
Summary Bit
1
0
T1 T2 T3
T4
T5
00
0
0
00
***
*
marks when event register is read
Enable Register
The enable register specifies the bits in the event register that can generate a summary bit. The instrument logically ANDs corresponding bits in the event and enable registers and ORs all the resulting bits to obtain a summary bit. Enable registers are read- write. Querying an enable register does not affect it.
An Example Sequence
Figure 1- 8 illustrates the response of a single bit position in a typical
status group for various settings. The changing state of the condition in question is shown at the bottom of the figure. A small binary table shows the state of the chosen bit in each status register at the selected times T1 to T5.
Figure 1-8 Typical Status Register Bit Changes
52 N1913/1914A EPM Series Power Meters Programming Guide
Page 75
Power Meter Remote Operation 1

How to Use Register

There are two methods to access the information in status groups:
• the polling method, or
• the service request (SRQ) method. (GPIB mode only)
Use the polling method when:
• your language/development environment does not support SRQ interrupts.
• you want to write a simple, single purpose program and do not want to add the complexity of setting an SRQ handler.
Use the SRQ method when you:
• need time critical notification of changes.
• are monitoring more than one device which supports SRQ interrupts.
• need to have the controller do something else while it is waiting.
• cannot afford the performance penalty inherent to polling.

The Condition Polling Method

In this polling method, the power meter has a passive role. It only informs the controller that conditions have changed when the controller asks. When you monitor a condition with the polling method, you must:
1 Determine which register contains the bit that monitors the
condition.
2 Send the unique query that reads that register.
3 Examine the bit to see if the condition has changed.
The polling method works well if you do not need to know about the changes the moment they occur. The SRQ method is more effective if you must know immediately when a condition changes. Detecting an immediate change in a condition using the polling method requires your program to continuously read the registers at very short intervals. This is not particularly efficient and there is a possibility that an event may be missed.
N1913/1914A EPM Series Power Meters Programming Guide 53
Page 76
1 Power Meter Remote Operation
NOTE
For example on measurement polling, refer to Figure B- 1 on page B-2.

The SRQ Method

When a bit of the Status Register is set and has been enabled to assert SRQ (*SRE command), the power meter sets the GPIB SRQ line true. This interrupt can be used to interrupt your program, suspending its current operation, and find out what service the power meter requires. Refer to your computer and language manuals for information on how to program the computer to respond to the interrupt.
To allow any of the Status Register bits to set the SRQ line true, you must enable the appropriate bit(s) with the *SRE command. For example, if your application requires an interrupt whenever a message is available in the output queue (Status Register bit 4, decimal weight 16). To enable bit 4 to assert SRQ, use the command *SRE 16.
You can determine which bits are enabled in the Status Register using *SRE?. This command returns the decimal weighted sum of all the bits.
Procedure
• Send a bus device clear message
• Clear the event registers with the *CLS (clear status) command
• Set the *ESE (standard event register) and *SRE (status byte register) enable masks
• Enable your bus controller’s IEEE-488 SRQ interrupt
Examples
The following two examples are written in HP BASIC and illustrate possible uses for SRQ. In both cases, it is assumed that the power meter has been zeroed and calibrated.
54 N1913/1914A EPM Series Power Meters Programming Guide
Page 77
Power Meter Remote Operation 1
Example 1:
10 ! Program to generate an SRQ when a channel A sensor 20 ! connect or disconnect occurs 30 ! 40 ASSIGN @Pm TO 713 ! Power meter GPIB address 50 ON ON INTR 7 GOTO Srq_i! Define service request handler 60 CLEAR @Pm ! Selective device clear 70 OUTPUT @Pm;”*CLS;*RST” ! Clear registers and resetmeter 80 ! 90 ! Configure the device status register so that a sensor 100 ! connect or disconnect on channel A will cause an SRQ. 110 ! 120 OUTPUT @Pm;”STAT:DEV:ENAB 2” 130 OUTPUT @Pm;”STAT:DEV:NTR 2” 140 OUTPUT @Pm;”STAT:DEV:PTR 2” 150 OUTPUT @Pm;”*SRE 2” 160 ! 170 ENABLE INTR 7;2 ! Enable an SRQ to cause an interrupt 180 LOOP ! Idle loop 190 ! Forever 200 END LOOP 210 ! 220 ! When a SRQ is detected, the following routine will service it. 230 ! 240 Srq_i: ! 250 St=SPOLL(@Pm) ! Serial Poll (reads status byte)
260 IF BIT(St,1)=1 THEN ! Device status reg bit set ? 270 OUTPUT @Pm;”STAT:DEV:EVEN?” ! Yes , read register
280 ENTER @Pm;Event ! (this also clears it)
290 OUTPUT @Pm;”STAT:DEV:COND?” 300 ENTER @Pm;Cond 310 IF Cond=0 THEN 320 PRINT “Sensor disconnected” 330 ELSE 340 PRINT “Sensor connected” 350 END IF 360 END IF 370 GOTO 170 ! Return to idle loop
N1913/1914A EPM Series Power Meters Programming Guide 55
Page 78
1 Power Meter Remote Operation
380 END
Example 2:
10 ! Program to generate an SRQ when an over limit 20 ! condition occurs. 30 ! 40 ASSIGN @Pm TO 713 ! Power meter GPIB address 50 ON INTR 7 GOTO Srq_i ! Define service request handler 60 CLEAR @Pm ! Selective device clear 70 OUTPUT @Pm;”*CLS” ! Clear registers 80 OUTPUT @Pm;”SYST:PRES” ! Preset meter 90 ! 100 ! Set upper limit to 2dBm and configure the operation status 110 ! so that an over limit condition will cause an SRQ. 120 ! 130 OUTPUT @Pm;”CALC:LIM:UPP 2DBM” 140 OUTPUT @Pm;”CALC:LIM:STAT ON” 150 OUTPUT @Pm;”STAT:OPER:PTR 4096” 160 OUTPUT @Pm;”STAT:OPER:ENAB 4096” 170 OUTPUT @Pm;”*SRE 128” 180 ! 190 ENABLE INTR 7;2 ! Enable an SRQ to cause an interrupt 200 LOOP ! Idle loop 210 ! Forever 220 END LOOP 230 ! 240 ! When a SRQ is detected, the following routine will service it. 250 !
260 Srq_i: ! 270 St=SPOLL(@Pm) ! Serial Poll (reads status byte) 280 IF BIT(St,7)=1 THEN ! Operation status bit set? 290 OUTPUT @Pm;”STAT:OPER?”! Yes , read register 300 ENTER @Pm;Oper ! (this also clears it) 310 OUTPUT @Pm;”STAT:OPER:ULF?” 320 ENTER @Pm;Ulf 330 IF Ulf=2 THEN PRINT “Over limit detected” 340 END IF 350 GOTO 190 ! Return to idle loop
56 N1913/1914A EPM Series Power Meters Programming Guide
Page 79
Power Meter Remote Operation 1
360 END
Status Registers
The Status System in the power meter is shown in Figure 1-9. The Operation Status and Questionable Status groups are 16 bits wide, while the Status Byte and Standard Event groups are 8 bits wide. In all 16- bit groups, the most significant bit (bit 15) is not used and is always set to 0.
N1913/1914A EPM Series Power Meters Programming Guide 57
Page 80
1 Power Meter Remote Operation
Questionable Status
Logical OR
Condit ion E vent Enable
Operation Sta tus
Logical OR
Condition Event Enable
Standard Event
Logical OR
Event
Enable
*ESR
*ESE
Status Byte
Logical OR
*STB?
*SRE
MAV
0 1 2
QUE
ESB
RQS/MSS
OPR
MAV
0 1 2
QUE
ESB
X
OPR
Output Q ueue
Device Status
Logical OR
Condition Event Enable
Error/Event Queue
Figure 1-9 Status System
58 N1913/1914A EPM Series Power Meters Programming Guide
Page 81
Power Meter Remote Operation 1
The Status Byte Summary Register
The status byte summary register reports conditions from other status registers. Query data waiting in the power meter’s output buffer is immediately reported through the “message available” bit (bit 4). Clearing an event register clears the corresponding bits in the status byte summary register. Reading all messages in the output buffer, including any pending queries, clears the message available bit.
Ta b le 1 -5 Bit Definitions - Status Byte Register
Bit Number Decimal
Definition
Weight
0 1 Not Used (Always set to 0)
1 2 Device Status Register summary bit.
One or more bits are set in the Device Status Register (bits must be “enabled” in enable register)
2 4 Error/Event Queue
3 8 Questionable Status Register summary bit.
One or more bits are set in the Questionable Status Register (bits must be “enabled” in enable register).
4 16 Data Available
Data is available in the power meter’s output buffer.
532Standard Event
One or more bits are set in the Standard Event register (bits must be “enabled” in enable register).
664Request Service
The power meter is requesting service (serial poll).
7 128 Operation Status Register summary bit.
One or more bits are set in the Operation Status Register (bits must be “enabled” in enable register).
Particular bits in the status byte register are cleared when:
• The standard event, Questionable status, operation status and device status are queried.
• The error/event queue becomes empty.
• The output queue becomes empty.
N1913/1914A EPM Series Power Meters Programming Guide 59
Page 82
1 Power Meter Remote Operation
The status byte enable register (SRE, service request enable) is cleared when you:
•cycle the instrument power.
•execute a *SRE 0 command.
Using *STB? to Read the Status Byte
The *STB? (status byte query) command is similar to a serial poll except it is processed like any other power meter command. The *STB? command returns the same result as an IEEE- 488 serial poll except that the request service bit (bit 6) is not cleared if a serial poll has occurred. The *STB? command is not handled automatically by the IEEE-488 bus interface hardware and the command is executed only after previous commands have completed. Using the *STB? command does not clear the status byte summary register.
The Standard Event Register
The standard event register reports the following types of instrument events: power- on detected, command and syntax errors, command execution errors, self- test or calibration errors, query errors, or when an overlapped command completes following a *OPC command. Any or all of these conditions can be reported in the standard event summary bit through the enable register. You must write a decimal value using the *ESE (event status enable) command to set the enable register mask.
Ta b le 1 -6 Bit Definitions - Standard Event Register
Bit Number
0 1 Operation Complete
1 2 Not Used. (Always set to 0.)
2 4 Query Error
Decimal Value
Definition
All overlapped commands following an *OPC command have been completed.
A query error occurred, refer to error numbers 410 to 440 in the user’s guide.
60 N1913/1914A EPM Series Power Meters Programming Guide
Page 83
Power Meter Remote Operation 1
Bit Number
3 8 Device Error
4 16 Execution Error
5 32 Command Error
6 64 User request.
7128Power On
Decimal Value
Definition
A device error occurred, refer to error numbers 310 to 350 in the user’s guide.
An execution error occurred, refer to error numbers 211 to 241 in the user’s guide.
A command syntax error occurred, refer to error numbers 101 to 161 in the user’s guide.
Power has been turned off and on since the last time the event register was read or cleared.
The standard event register is cleared when you:
•send a *CLS (clear status) command.
• query the event register using the *ESR? (event status register) command.
The standard event enable register is cleared when you:
•cycle the instrument power.
•execute a *ESE 0 command.
Questionable Status Register
The questionable status register provides information about the quality of the power meter’s measurement results. Any or all of these conditions can be reported in the questionable data summary bit through the enable register. You must write a value using the STATus:QUEStionable:ENABle command to set the enable register mask.
The questionable status model is shown in the pullout at the end of this chapter.
The following bits in these registers are used by the power meter.
N1913/1914A EPM Series Power Meters Programming Guide 61
Page 84
1 Power Meter Remote Operation
Ta b le 1 -7 Bit Definitions - Questionable Status Registers
Bit Number
0 to 2 - Not used
38 POWer Summary
4 to 7 - Not used
8 256 CALibration Summary
9 512 Power On Self Test
10 to 14 - Not Used
15 - Not used (always 0)
Decimal Weight
Definition
The condition bits are set and cleared under the following conditions:
Ta b le 1 -8 Bit change conditions for Questionable Status Register
Bit
Meaning EVENts Causing Bit Changes
Number
3POWer
Summary
This is a summary bit for the Questionable POWer Register.
• SET: Error –230, “Data corrupt or stale” Error –231, “Data questionable;Input Overload”
Error –231, “Data questionable;Input Overload ChA”
Error –231, “Data questionable;Input Overload ChB” Error –231, “Data questionable;PLEASE ZERO”
Error –231, “Data questionable;PLEASE ZERO ChA”
Error –231, “Data questionable;PLEASE ZERO ChB”
Error –231, ”Data questionable;Lower window log error”
Error –231, ”Data questionable;Upper window log error”
• CLEARED: When no errors are detected by the power meter during a measurement covering the causes given for it to set.
*
*
*
*
*
*
62 N1913/1914A EPM Series Power Meters Programming Guide
Page 85
Power Meter Remote Operation 1
Bit
Meaning EVENts Causing Bit Changes
Number
8CALibration
Summary
9Power On Self
Te st
* N1914A only
This is a summary bit for the Questionable CALibration Register.
• SET:
• CLEARED: When any of the commands listed above
• SET: This bit is set when the power on self test fails.
• CLEARED: When the power on self test passes.
These may be caused by
CALibration[1|2]:ZERO:AUTO ONCE or CALibration[1|2]:AUTO ONCE or CALibration[1|2][:ALL] or CALibration[1|2][:ALL]?.
Error –231, “Data questionable; ZERO ERROR” Error –231, “Data questionable; ZERO ERROR ChA”
Error –231, “Data questionable; ZERO ERROR ChB” Error –231, “Data questionable; CAL ERROR”
Error –231, “Data questionable; CAL ERROR ChA”
Error –231, “Data questionable; CAL ERROR ChB”
succeed and no errors are placed on the error queue.
*
*
*
*
Operation Status
The Operation Status group monitors conditions in the power meter’s measurement process.
The Operation status model is shown in the pullout at the end of this chapter.
The following bits in these registers are used by the power meter:
Ta b le 1 -9 Bit Definitions - Operation Status
Bit Number Decimal
Weight
0 1 CALibrating Summary
1 - 3 - Not used
N1913/1914A EPM Series Power Meters Programming Guide 63
Definition
Page 86
1 Power Meter Remote Operation
Bit Number Decimal
Definition
Weight
416MEASuring Summary
5 32 Waiting for TRIGger Summary
6 - 9 - Not used
10 1024 SENSe Summary
11 2048 Lower Limit Fail Summary
12 4096 Upper Limit Fail Summary
13 to 14 - Not used
15 - Not used (always 0)
The condition bits are set and cleared under the following conditions:
Table 1-10 Bit change conditions for Operation Status
Bit
Meaning EVENts Causing Bit Changes
Number
0 CALibrating This is a summary bit for the Operation CALibrating Register.
• SET: At beginning of zeroing (CALibration:ZERO:AUTO ONCE) and at the beginning of calibration (CALibration:AUTO ONCE). Also for the compound command/query CALibration[:ALL]?, this bit is set when sensor zeroing begins.
• CLEARED: At the end of zeroing or calibration.
4 MEASuring This is a summary bit for the Operation MEASuring Register.
• SET: When the power meter is taking a measurement.
• CLEARED: When the measurement is finished.
5Waiting for
TRIGger
10 SENSe This is a summary bit for the Operation SENSe Register.
This is a summary bit for the Operation TRIGger Register.
• SET: When the power meter enters the “wait for trigger” state.
• CLEARED: When the power meter enters the “idle” state.
• SET: When the power meter is reading data from the power sensor’s EEPROM.
• CLEARED: When the power meter is not reading data from the power sensor’s EEPROM.
64 N1913/1914A EPM Series Power Meters Programming Guide
Page 87
Power Meter Remote Operation 1
Bit
Meaning EVENts Causing Bit Changes
Number
11 Lower Limit
Fail
12 Upper Limit
Fail

Device Status Register

The device status register set contains bits which give device dependent information.
The following bits in these registers are used by the power meter:
Table 1-11 Bit Definitions - Device Status Register
This is a summary bit for the Lower Limit Fail Register.
• SET: If a measurement is made and either a channel or window lower limit test fails.
• CLEARED: If a measurement is made and the lower limit test is not enabled or the test is enabled and passes.
This is a summary bit for the Upper Limit Fail Register.
• SET: If a measurement is made and either a channel or window upper limit test fails.
• CLEARED: If a measurement is made and the upper limit test is not enabled or the test is enabled and passes.
Bit Number
0- Not used
1 2 Channel A sensor connected
2 4 Channel B sensor connected (N1914A only)
38Channel A sensor error
4 16 Channel B sensor error (N1914A only)
5 32 Channel A sensor Front/Rear
6 64 Channel B sensor Front/Rear (N1914A only)
7 128 Channel C sensor connected
8 256 Channel D sensor connected
9512Not used
Decimal Weight
Definition
N1913/1914A EPM Series Power Meters Programming Guide 65
Page 88
1 Power Meter Remote Operation
Bit Number
10 1024 Not used
11 2048 Battery critical
12 4096 Over temperature
13 8192 Fan failed
14 16384 Front panel key press
15 - Bit 15 always 0
Decimal Weight
Definition
The condition bits are set and cleared under the following conditions:
Table 1-12 Bit change conditions for Device Status Register
Bit
Meaning EVENts Causing Bit Changes
Number
1 Channel A
sensor connected
2 Channel B
sensor connected
3 Channel A
error
4 Channel B
error
5 Channel A
Front/Rear
• SET: When a power sensor is connected to the Channel A input.
• CLEARED: When no power sensor is connected to the Channel A input.
• SET: When a power sensor is connected to the Channel B input.
• CLEARED: When no power sensor is connected to the Channel B input.
• SET: If the power sensor EEPROM on Channel A has failed or if there are power sensors connected to both the rear and front panel Channel A connectors.
• CLEARED: In every other condition.
• SET: If the power sensor EEPROM on Channel B has failed or if there are power sensors connected to both the rear and front panel Channel B connectors.
• CLEARED: In every other condition.
• SET: If a power sensor is connected to the Channel A rear panel.
• CLEARED: If a power sensor is connected to the Channel A front panel.
66 N1913/1914A EPM Series Power Meters Programming Guide
Page 89
Power Meter Remote Operation 1
Bit
Meaning EVENts Causing Bit Changes
Number
6 Channel B
Front/Rear
7 Channel C
sensor connected
8 Channel D
sensor connected
11 Battery critical This is an event, and DOES NOT set the condition register. The
12 Over
temperature
13 Fan failed This is an event, and DOES NOT set the condition register. The
14 Front Panel
Key Press
• SET: If a power sensor is connected to the Channel B rear panel.
• CLEARED: If a power sensor is connected to the Channel B front panel.
• SET: When a power sensor is connected to the Channel C input.
• CLEARED: When no power sensor is connected to the Channel C input.
• SET: When a power sensor is connected to the Channel D input.
• CLEARED: When no power sensor is connected to the Channel D input.
bit is set in the event register which is cleared when read. Note that the transition registers are of no use for this bit.
This is an event, and DOES NOT set the condition register. The bit is set in the event register which is cleared when read. Note that the transition registers are of no use for this bit.
bit is set in the event register which is cleared when read. Note that the transition registers are of no use for this bit.
This is an event, and DOES NOT set the condition register. The bit is set in the event register which is cleared when read. Note that the transition registers are of no use for this bit.

Using the Operation Complete Commands

The *OPC? and *OPC commands allow you to maintain synchronization between the computer and the power meter. The *OPC? query command places an ASCII character 1 into the power meter’s output queue when all pending power meter commands are complete. If your program reads this response before continuing program execution, you can ensure synchronization between one or more instruments and the computer.
The *OPC command sets bit 0 (Operation Complete) in the Standard Event
N1913/1914A EPM Series Power Meters Programming Guide 67
Page 90
1 Power Meter Remote Operation
NOTE
Status Register when all pending power meter operations are complete. By enabling this bit to be reflected in the Status Register, you can ensure synchronization using the GPIB serial poll.
For LAN and USB use the *STB? command. See “Using *STB? to Read the Status
Byte” on page 60.
Procedure
• Send a device clear message to clear the power meter’s output buffer.
• Clear the event registers with the *CLS (clear status) command.
• Enable operation complete using the *ESE 1 command (standard event
register).
• Send the *OPC? (operation complete query) command and enter the
result to assure synchronization.
• Send your programming command string, and place the *OPC
(operation complete) command as the last command.
• Send the *STB? (status byte query) command to poll the register. This
command does not clear the status byte summary register.
In GPIB mode only you can use a serial poll to check to see when bit 5 (standard event) is set in the status byte summary register. You could also configure the power meter for an SRQ interrupt by sending *SRE 32 (status byte enable register, bit 5).
Examples
This example program uses the *OPC? command to determine when the power meter has finished calibrating.
CAL:AUTO ONCE *OPC? MEAS:POW:AC?
This example GPIB program, in HP Basic, uses the *OPC command and serial poll to determine when the power meter has finished calibrating. The advantage to using this method over the *OPC? command is that the computer can perform other operations while it is waiting for the power meter to finish calibrating.
68 N1913/1914A EPM Series Power Meters Programming Guide
Page 91
Power Meter Remote Operation 1
10 ASSIGN @Power TO 713 20 OUTPUT @Power;“*CLS” 30 OUTPUT @Power;“*ESE 1” 40 OUTPUT @Power;“CAL:AUTO ONCE;*OPC” 50 WHILE NOT BIT(SPOLL(@Power),5) 60 !(Computer carries out other operations here) 70 END WHILE 80 OUTPUT @Power;“MEAS:POW:AC?” 90 ENTER @Power;Result 100 PRINT Result 110 END
N1913/1914A EPM Series Power Meters Programming Guide 69
Page 92
1 Power Meter Remote Operation

Saving and Recalling Power Meter Configurations

To reduce repeated programming, up to ten power meter configurations can be stored in the power meter’s non- volatile memory. The error list, remote addresses, sensor calibration table data, zeroing and calibration information are not stored.

How to Save and Recall a Configuration

Power meter configurations are saved and recalled with the following commands:
*SAV <NRf> *RCL <NRf>
The range of values for <NRf> in the above commands is 1 to 10.
Example Program
10 ASSIGN @POWER TO 713 20 !Configure the power meter 30 OUTPUT @POWER;“UNIT:POW W” 40 OUTPUT @POWER;“SENS:CORR:LOSS2 -10” 50 OUTPUT @POWER;“SENS:CORR:LOSS2:STAT ON” 60 !Save the configuration 70 OUTPUT @POWER;“*SAV 5” 80 PRINT “Configuration Saved” 90 !Now reset the power meter 100 OUTPUT @POWER;“*RST” 110 ! Recall the configuration 120 OUTPUT @POWER;”*RCL 5” 130 PRINT “Configuration Recalled” 140 PRINT “Save and Recall complete” 150 END
70 N1913/1914A EPM Series Power Meters Programming Guide
Page 93

Using Device Clear to Halt Measurements

NOTE
Device clear is an IEEE- 488 low- level bus message which can be used to halt measurements in progress. Different programming languages and IEEE- 488 interface cards provide access to this capability through their own unique commands. The status registers, the error queue, and all configuration states are left unchanged when a device clear message is received. Device clear performs the following actions.
• All measurements in progress are aborted.
• The power meter returns to the trigger “idle state”.
• The power meter’s input and output buffers are cleared.
• The power meter is prepared to accept a new command string.
For interfaces the that do not support a low-level device clear, use the ABORt command.
Power Meter Remote Operation 1
N1913/1914A EPM Series Power Meters Programming Guide 71
Page 94
1 Power Meter Remote Operation
“B” Subsystem
:G :I:H
:M
:N=:B:H:N
“A” S u bs ys tem
:D:F:E
“C” Subsystem
:J :L=:C:L:K

An Introduction to the SCPI Language

Standard Commands for Programmable Instruments (SCPI) defines how you communicate with an instrument from a bus controller. The SCPI language uses a hierarchical structure similar to the file systems used by many bus controllers. The command tree is organized with root- level commands (also called subsystems) positioned at the top, with multiple levels below each root- level command. You must specify the complete path to execute the individual lower- level commands.
Figure 1-10 Hierarchical structure of SCPI

Mnemonic Forms

Each keyword has both a long and a short form. A standard notation is used to differentiate the short form keyword from the long form keyword. The long form of the keyword is shown, with the short form portion shown in uppercase characters, and the rest of the keyword shown in lowercase characters. For example, the short form of TRIGger is TRIG.

Using a Colon (:)

When a colon is the first character of a command keyword, it indicates that the next command mnemonic is a root- level command. When a colon is inserted between two command mnemonics, the colon moves the path
72 N1913/1914A EPM Series Power Meters Programming Guide
Page 95
Power Meter Remote Operation 1
down one level in the present path (for the specified root-level command) of the command tree. You must separate command mnemonics from each other using a colon. You can omit the leading colon if the command is the
first of a new program line.

Using a Semicolon (;)

Use a semicolon to separate two commands within the same command string. The semicolon does not change the present path specified. For example, the following two statements are equivalent. Note that in the first statement the first colon is optional but the third is compulsory.
:DISP:FORM DIG;:DISP:RES 2 :DISP:FORM DIG;RES 2

Using a Comma (,)

If a command requires more than one parameter, you must separate adjacent parameters using a comma.

Using Whitespace

You must use whitespace characters, [tab], or [space] to separate a parameter from a command keyword. Whitespace characters are generally ignored only in parameter lists.

Using “?” Commands

The bus controller may send commands at any time, but a SCPI instrument may only send responses when specifically instructed to do so. Only query commands (commands that end with a “?”) instruct the instrument to send a response message. Queries return either measured values or internal instrument settings.
N1913/1914A EPM Series Power Meters Programming Guide 73
Page 96
1 Power Meter Remote Operation
NOTE
If you send two query commands without reading the response from the first, then attempt to read the second response, you may receive some data from the first response followed by the complete second response. To avoid this, do not send a query command without reading the response. When you cannot avoid this situation, send a device clear before sending the second query command.

Using “*” Commands

Commands starting with a “*” are called common commands. They are required to perform the identical function for all instruments that are compliant with the IEEE- 488.2 interface standard. The “*” commands are used to control reset, self-test, and status operations in the power meter.

Syntax Conventions

Throughout this guide, the following conventions are used for SCPI command syntax.
• Square brackets ([]) indicate optional keywords or parameters.
• Braces ({}) enclose one or more parameters that may be included zero or more times.
• Triangle brackets (<>) indicate that you must substitute a value for the enclosed parameter.
• Bars (|) can be read as “or” and are used to separate alternative parameter options.

Syntax Diagram Conventions

• Solid lines represent the recommended path.
• Ovals enclose command mnemonics. The command mnemonic must be entered exactly as shown.
74 N1913/1914A EPM Series Power Meters Programming Guide
Page 97
Power Meter Remote Operation 1
<digit>
<upper-case
alpha>
<upper-case
alpha>
• Dotted lines indicate an optional path for by passing secondary keywords.
• Arrows and curved intersections indicate command path direction.

SCPI Data Types

The SCPI language defines different data formats for use in program messages and response messages. Instruments are flexible listeners and can accept commands and parameters in various formats. However, SCPI instruments are precise talkers. This means that SCPI instruments always respond to a particular query in a predefined, rigid format.
<boolean> Definition
Throughout this document <boolean> is used to represent ON|OFF|<NRf>. boolean parameters have a value of 0 or 1 and are unitless. ON corresponds to 1 and OFF corresponds to 0.
On input, an <NRf> is rounded to an integer. A nonzero result is interpreted as 1.
Queries always return a 1 or 0, never ON or OFF.
<character_data> Definition
Throughout this document <character_data> is used to represent character data, that is, A - Z, a - z, 0 -9 and _ (underscore). For example: START and R6_5F. The format is defined as:
Figure 1-11 Format of <character_data>
N1913/1914A EPM Series Power Meters Programming Guide 75
Page 98
1 Power Meter Remote Operation
<NAN> Definition
Not a number (NAN) is represented as 9.91 E37. Not a number is defined
in IEEE 754.
<non-decimal numeric> Definition
Throughout this document <non-decimal numeric> is used to represent numeric information in bases other than ten (that is, hexadecimal, octal and binary). The following syntax diagram shows the standard for these three data structures. For examples, #HA2F, #ha4e, #Q62, #q15, #B01011.
76 N1913/1914A EPM Series Power Meters Programming Guide
Page 99
A/a
B/b C/c D/d
E/e
F/f
<digit>
H/h
1
2 3 4 5 6
0
7
Q/q
0
1
#
B/b
Power Meter Remote Operation 1
Figure 1-12 Format of <non-decimal numeric>
Refer to section 7.7.4.1 of IEEE 488.2 for further details.
<NRf> Definition
Throughout this document <NRf> is used to denote a flexible numeric representation. For example: +200; –56; +9.9E36. Refer to section 7.7.2.1 of
N1913/1914A EPM Series Power Meters Programming Guide 77
Page 100
1 Power Meter Remote Operation
digit
+
digit
+
digit
IEEE 488.2 for further details.
<NR1> Definition
Throughout this document <NR1> numeric response data is defined as:
Figure 1-13 Format of <NR1>
For example:
•146
•+146
• –12345
Refer to section 8.7.2 of IEEE 488.2 for further details.
<NR2> Definition
Throughout this document <NR2> numeric response data is defined as:
Figure 1-14 Format of <NR2>
For example:
•12.3
• +1.2345
• –0.123
Refer to section 8.7.3 of IEEE 488.2 for further details.
78 N1913/1914A EPM Series Power Meters Programming Guide
Loading...