Agilent Increasing Productivity and Simplifying Sample Preparation With the SIPS Accessory Technical manual

Technical Overview
Increasing Productivity and Simplifying Sample Preparation With the SIPS Accessory
Flame Atomic Absorption Spectrometry (FAAS) typically involves element-specific sample preparation which can be time consuming and prone to operator-introduced errors. The SIPS 10 and SIPS 20 Sample Introduction Pump System (SIPS) are optional accessories for Agilent’s SpectrAA FAAS instruments. They offer a range of powerful features that automate many tedious and error prone sample preparation tasks. These greatly simplify and speed up the sample preparation process, improving productivity and reducing costs.
Whether you are using an autosampler, or doing manual sample introduction, a SIPS accessory, combined with the SpectrAA software, will automatically:
Prepare up to 10 calibration standards from a single bulk
standard, automating the instrument calibration process
Intelligently dilute over range samples
Prepare standard additions calibration from a single
standard
Introduce analytical spikes, using a single standard
Add chemical modifiers, e.g. ionization suppressants
Extend the dynamic range of FAAS to that approaching
ICP-OES
The SIPS accessory is available in two versions:
The single pump SIPS 10 provides in-line calibration and
sample dilution.
The dual pump SIPS 20 has the additional capabilities of
in-line spiking of samples and in-line sample modification.
Figure 2. How the SIPS 10 works
How SIPS works
The sample flow rate of conventional peristaltic pumps changes non-linearly with the pump speed. This makes them unsuitable for use to automatically dilute samples. The innovative design of the SIPS’ peristaltic pump system ensures that the solution flow is proportional to the pump speed, so the volume of solution being pumped is accurately controlled. To extend the lifetime of the peristaltic pump tubing (and thus lower operating costs), the SIPS will automatically release the pressure on the pump tubing at the end of run.
The SIPS 10 (Figure 1) combines a small flow of diluent with the sample flowing to the nebulizer. If the pump speed is reduced, the flow of sample is reduced. With a constant nebulizer uptake rate, the diluent flow is thus increased effectively diluting the sample. This simple approach makes automatic calibration and in-line dilution possible, by accurately controlling the pump speed with the SpectrAA software.
Figure 2. How the SIPS 20 works
2
The SIPS 20 comprises two pumps (Figure 2). One delivers the sample and the second delivers a standard or modifier to the nebulizer. Controlling the relative speeds of both pumps allows the amount of standard or modifier added to the sample to be precisely regulated. This allows in-line ‘spiking’ of samples.
No more tedious standard preparation
The SIPS 10 and 20 are capable of preparing calibration stan­dards from a single multi-element bulk standard. This reduces sample preparation time and the risk of operator error during standard preparation, which can lead to costly re-runs.
Using the concentration of the bulk standard, the SpectrAA software can automatically determine standard concentrations to cover the required concentration range.
Figure 3 shows an example of a calibration curve for Cu pre­pared from a single standard using the SIPS system. It has a correlation coefficient of 1.0000, demonstrating the excellent calibration accuracy that can be achieved with the SIPS.
The dilution accuracy of a SIPS is shown in Table1, demon­strating the ability of the SIPS system to provide accurate results, even at high dilution factors. Figure 4 shows the dilu­tion performance of a SIPS with increasing dilution factors. Excellent dilution accuracy was obtained across all dilution factors with a maximum dilution error of just 1.2%.
Table 1. Typical dilution performance of a SIPS system, obtained by repeating a number of standard measurements on different days under different sequence conditions.
Dilution Factor 1 3-20 20-50 50-200
Standard Conc (mg/L) 5-10 10-100 100-320 200-1000
# of determinations 53 65 25 58
Dilution Error (%) -1 2 1 -1
Precision (%RSD) 1 1 1 3
Reproducibility (%) 1 1 2 2
Figure 3. Calibration curve for Cu prepared from a single standard, using a SIPS. The concentrations of the standards were automatically set by the SpectrAA software to cover the required concentration range.
Automatic in-line dilution
The dilution capability of a SIPS extends the working range of AA by automatically diluting samples that are outside the calibrated range. The SpectrAA software calculates the dilu­tion required to bring over range samples into the top half of the calibrated range, where maximum precision and accuracy usually occur. This ensures the integrity of results and eliminates the need for costly re-analysis of samples at the end of a run.
Figure 4. Dilution performance of a SIPS 20 at different dilution factors. Excellent dilution accuracy was obtained with a maximum of only 1.2% dilution error.
Modifier addition
Flame AA methods routinely require the addition of chemical modifiers for accurate results. These modifiers are normally added during the sample preparation process. Using the second pump of the SIPS 20, modifiers can be automatically added during the analysis, taking a repetitive step out of the sample preparation process. The amount of modifier added is controlled by the pump speed, and this pump speed remains constant during sample dilutions to ensure consistent blank correction and matrix modification; something that must be accounted for when adding modifiers manually.
3
In-line spiking of samples
The SIPS 20 automatically ‘spikes’ samples in-line, eliminating sample preparation and reducing the risk of sample contami­nation. One pump introduces the sample, and the second spikes the sample with standard.
In-line standard additions calibration
The dual pump SIPS 20 automatically adds addition stan­dards during the analysis. One pump introduces the sample and the second introduces the standard. The SIPS system au­tomatically adjusts the speed of the second pump to achieve the required additions—eliminating time-consuming sample preparation and reducing the risk of contamination.
When maximum accuracy is required, the SIPS system en­sures that the calibration is completed using the linear portion of the curve. Where necessary, the addition standards will be diluted to ensure the absorbance signals fall within the linear range.
www.agilent.com/chem
This information is subject to change without notice.
© Agilent Technologies, Inc. 2021 Printed in the USA, January 5, 2021 5991-6613EN
Loading...