Agilent Genomic DNA ScreenTape Assay Quick Start Guide

Genomic DNA ScreenTape Assay for TapeStation Systems
Quick Guide
The Agilent 4150 (G2992AA) and 4200 (G2991AA and G2991BA) TapeStation systems are automated platforms for scalable, flexible, fast, and reliable electrophoresis of nucleic acids. This Quick Guide is intended for use with the Agilent 4150 and 4200 TapeStation systems only. A Quick Guide specific for use with the Agilent 2200 TapeStation system is available online. The Genomic DNA ScreenTape assay is designed for assessing integrity of genomic DNA samples and analyzing double-stranded DNA molecules from 200 to > 60000 basepairs.

Specifications

Analytical specifications Genomic DNA ScreenTape assay
Sizing range 200 to >60000 bp
Sensitivity
Sizing precision
Sizing accuracy
Quantitative precision 15 % CV
Quantitative accuracy ±20 %
Quantitative range 10 – 100 ng/µL
DIN functional range
Maximum buffer concentration in sample
Physical specifications
Analysis time 15 samples: <25 min, 96 samples: <140 min
Samples per consumable 15
Sample volume required 1 µL
Kit stability 4 months
Kit size 105 samples
1 2 3
1
2
2
3
Signal-to-noise >3 (single peak) Determined using the Genomic DNA Ladder as sample DIN - DNA Integrity Number
0.5 ng/µL
200 – 15000 bp: 15 % CV
200 – 15000 bp: ±15 %
5 – 300 ng/µL
10 mM MgCl 1 µg/µL glycogen
, 50 mM NaCl, 10 mM NaOAc, 10 % ethanol, 10 % 2-propanol,
2
Agilent Genomic DNA ScreenTape Quick Guide for TapeStation Systems
Genomic DNA ScreenTape Assay for TapeStation Systems

Storage Conditions

Reagent vials and ScreenTape devices: 2 – 8 °C (36 – 46 °F).
Store partially used ScreenTape devices upright at 2 – 8 °C (36 – 46 °F) for a maximum of 2 weeks.
Never freeze ScreenTape devices. Discard any accidentally frozen ScreenTape devices.

Kit Components

Part Number Name Color Amount
5067-5365 Genomic DNA ScreenTape 7 ScreenTape devices
5067-5366 Genomic DNA Reagents
Genomic DNA Ladder
Genomic DNA Sample Buffer
2 vials 25 µL 1350 µL

Limited Use Label License

Some products within this assay contain SYBR™ Green I, which is licensed from Life Technologies Corporation for use in research and development only. SYBR™ is a registered trademark of Life Technologies Corporation.

For Research Use Only

Not for use in Diagnostic Procedures.

Additional Material Required for Analysis with the TapeStation Systems

Loading tips (5067-5598, 1 pk or 5067-5599, 10 pk)
Optical Tube 8x Strip (401428) and Optical Tube Cap 8x Strip (401425)
Vortex mixer IKA MS3 with 96-well sample plate adapter
96-well sample plates (5042-8502) and 96-well Plate Foil Seal (5067-5154) (4200 TapeStation systems only)

Additional Equipment Required (Not Supplied)

Volumetric micropipettes for handling volumes from 1 to 15 µL
Centrifuges for tube strips and 96-well sample plates
WARN IN G
CAU T IO N
2
Toxic agents
Refer to product material safety datasheets for further information.When working with the ScreenTape assay follow the appropriate safety procedures such as
wearing safety goggles, laboratory gloves and protective clothing.
Damage to the TapeStation systems
Only use the recommended consumables and reagents with the TapeStation systems.
Genomic DNA ScreenTape Assay for TapeStation Systems

Essential Measurement Practices

Read about good measurement practices in the Agilent Information Center and/or in the System Manual.
Environmental conditions
Steps before sample preparation
Ladder handling Do not shake or overmix ladder vial
Pipetting practice Pipette reagents carefully against the side of the 96-well sample plate or sample tube
Mixing and centrifugation recommendations
Ambient operating temperature: 15 – 30 °C (59 – 86 °F)
Keep reagents during sample preparation at room temperature
Allow Sample Buffer to equilibrate at room temperature for 30 min prior to use
Vortex each vial and briefly spin down
Flick ScreenTape device to eliminate bubbles in the buffer chamber
Ensure that no sample or Sample Buffer remains within or on the outside of the tip
Care must be taken due to viscosity of the Sample Buffer
Apply foil seal to 96-well sample plate or cap the tube strips prior to mixing and centrifugation
Centrifuge to collect liquid at the base; then vortex using the IKA MS3 vortexer and adaptor at
2000 rpm for 1 min
Briefly centrifuge and visually confirm that all liquid is collected at the bottom of the 96-well sample plate or tube strips and any air bubble is removed
Run samples immediately after preparation

Ladder Considerations

Ladder is exclusively loaded from location A1 on the tube strip holder.
Always use a complete tube strip when running ladder or samples from the tube strip holder.
The analysis of one ladder per ScreenTape device is required.
3
Genomic DNA ScreenTape Assay for TapeStation Systems
Sample
Buffer
DNA
10 µL
1 µL
Flick
Spin down 1 min
Samples, tips, and ScreenTape device
Vortex 1 min
Genomic DNA

Genomic DNA ScreenTape Assay Operating Procedure

1 Allow Genomic DNA Reagents to equilibrate at room temperature for 30 minutes. 2 Launch the Agilent TapeStation Controller software. 3 Flick the Genomic DNA ScreenTape device and insert it into the ScreenTape nest of the TapeStation instrument. 4 Select required sample positions in the TapeStation Controller software. 5 The required consumables (tips, further ScreenTape devices) are displayed in the TapeStation Controller software. 6 Vortex reagents and samples. Spin down before use. 7 Prepare ladder:
For 1 or 2 ScreenTape devices: pipette 10 µL Genomic DNA Sample Buffer ( ) and 1 µL Genomic DNA Ladder
( ) at position A1 in a tube strip.
For more than 2 ScreenTape devices
Ladder ( ) at position A1 in a tube strip.
8 For each sample, pipette 10 µL Genomic DNA Sample Buffer ( ) and 1 µL DNA sample in a tube strip or 96-well
sample plate
9 Apply caps to tube strips and/or foil seals to 96-well sample plates. 10 Mix liquids using the IKA MS3 vortexer at 2000 rpm for 1 min. 11 Spin down samples and ladder for 1 min.
1
.
1
: pipette 20 µL Genomic DNA Sample Buffer ( ) and 2 µL Genomic DNA

Sample Analysis

1 Load samples into the TapeStation instrument. Place ladder in position A1 on tube strip holder. 2 Carefully remove caps of tube strips. Visually confirm that liquid is positioned at the bottom. 3 Click Start. 4 The TapeStation Analysis software opens automatically after the run and displays results.

Technical Support and Further Information

For technical support, please visit www.agilent.com/chem/contactus. Visit Agilent Technologies’ web site. It offers useful information, support and current developments about the products and technology:
www.agilent.com/genomics/tapestation.